CHLORINE KIT, DUAL RANGE
OCTA-SLIDE 2, 0.1-6.0 ppm
CODE 3314-01
QUANTITY CONTENTS CODE
100 *Chlorine DPD #1R Tablets *6999A-J
100 *Chlorine DPD #3R Tablets *6905A-J
2 Test Tubes 2.5-5.0-10.0 mL, plastic, w/caps 0106
1 Octa-Slide Viewer 1101
1 Chlorine Octa-Slide 2 Bar, 0.1-1.0 ppm 3405-01
1 Chlorine Octa-Slide 2 Bar, 1.0-6.0 ppm 3404-01
*WARNING: Reagents marked with an * are considered to be potential health hazards.
To view or print a Material Safety Data Sheet (MSDS) for these reagents go to
www.lamotte.com. To obtain a printed copy, contact LaMotte by e-mail, phone or fax.
To order individual reagents or test kit components, use the specifi ed code number.
This kit is for use in testing water for Free Available Chlorine, Total Residual
Chlorine and Combined Chlorine.
USE OF THE OCTA-SLIDE 2 VIEWER
The Octa-Slide 2 Viewer should be held so
non-direct light enters through the back of
the Viewer. Insert the reacted sample into
the top of the Viewer. Slide the Octa-Slide 2
Bar into the Viewer and match the color of
the reaction to the color standards.
WARNING! This set contains chemicals
that may be harmful if misused. Read
cautions on individual containers
carefully. Not to be used by children
except under adult supervision

EPA ACCEPTED PROCEDURE
EPA accepted for NIPDWR monitoring. For compliance monitoring a Check
Standard should be prepared.
CHECK STANDARD PREPARATION: 1 ppm Equivalent Solution
1. Dissolve 891 mg of potassium permanganate in 1000 mL of distilled water in a
volumetric fl ask (1000 ppm equivalent solution).
2. Dilute 1 mL of this solution to 1000 mL with distilled water in a volumetric fl ask.
This solution is equivalent to 1 ppm Free Available Chlorine.
PROCEDURE
FREE AVAILABLE CHLORINE
1. Fill a test tube (0106) to the 5 mL line with sample water.
2. Add one *Chlorine #1R Tablet (6999A). Cap and mix until tablet disintegrates.
A pink to red color indicates the presence of Chlorine.
3. Insert test tube into the top of the Octa-Slide 2 Viewer (1101). Slide the 0.1 to 1.0
ppm Chlorine Octa-Slide 2 Bar (3405-01) into the Octa-Slide 2 Viewer. Match
sample color to a color standard. Record as ppm Free Available Chlorine.
4. If the sample is darker than the 1.0 standard, remove 0.1 to 1.0 Octa-Slide 2 Bar,
replace it with the 1.0 to 6.0 Octa-Slide 2 Bar (3404-01). Match sample color to
a color standard. Record as ppm Free Available Chlorine.
NOTE: Save sample for Total Residual and Combined Chlorine tests.
TOTAL RESIDUAL & COMBINED CHLORINE
5. Add one *Chlorine DPD #3R Tablet (6905A) to the sample from Step 4. Cap tube
and mix until tablet disintegrates.
6. Insert test tube into the Octa-Slide 2 Viewer (1101). Match sample color to a color
standard. Record as ppm Total Residual Chlorine.
7. Subtract Free Available Chlorine from Total Residual Chlorine. Record as ppm
Combined Chlorine.
Combined Chlorine = Total Residual Chlorine – Free Available Chlorine
NOTE: Thoroughly clean and rinse test tubes after each test.
LaMOTTE COMPANY
Helping People Solve Analytical Challenges
PO Box 329 • Chestertown • Maryland • 21620 • USA
800-344-3100 • 410-778-3100 (Outside U.S.A.) • Fax 410-778-6394
Visit us on the web at www.lamotte.com
63314-01 5/11
®