LaMotte SL-1 User Manual

DPD CHLORINE KIT
OCTA-SLIDE 2, 0.2-3.0 ppm
CODE 3316-01
QUANTITY CONTENTS CODE
50 50
50 50
2 Test Tubes, plastic, w/caps 0106 1 Octa-Slide 2 Viewer 1101 1 Chlorine Octa-Slide 2 Bar, 0.2-3.0 ppm 3401-01
*Chlorine DPD #1 Tablets (*6999A) *Chlorine DPD #3 Tablets (*6905A)
6905A­6999ABOX
6904A­6906ABOX
*WARNING: Reagents marked with an * are considered to be potential health hazards. To view or print a Material Safety Data Sheet (MSDS) for these reagents go to www.lamotte.com. To obtain a printed copy, contact LaMotte by e-mail, phone or fax.
To order individual reagents or test kit components, use the specifi ed code number. This kit is used for testing water for Free Available Chlorine, Monochloramine,
Dichloramine, and Total Residual Chlorine.
USE OF THE OCTA-SLIDE 2 VIEWER
The Octa-Slide 2 Viewer should be held so non-direct light enters through the back of the Viewer. Insert the reacted sample into the top of the Viewer. Slide the Octa-Slide 2 Bar into the Viewer and match the color of the reaction to the color standards.
WARNING! This set contains chemicals
that may be harmful if misused. Read
cautions on individual containers
carefully. Not to be used by children
except under adult supervision
LaMOTTE COMPANY
Helping People Solve Analytical Challenges
PO Box 329 • Chestertown • Maryland • 21620 • USA
800-344-3100 • 410-778-3100 (Outside U.S.A.) • Fax 410-778-6394
Visit us on the web at www.lamotte.com
63316-01 5/11
®
PROCEDURE
FREE AVAILABLE CHLORINE
1. Fill a test tube (0106) to 5 mL line with sample water.
2. Add one *Chlorine DPD #1 Tablet (6999A). Cap and mix until tablet disintegrates. Solution will turn pink if chlorine is present.
3. Insert test tube into the Octa-Slide 2 Viewer (1101). Slide the 0.2 to 3.0 ppm Chlorine Octa-Slide 2 Bar (3401-01) into the viewer. Match sample color to a color standard. Record as Reading A. Save sample for further tests.
Free Available Chlorine (ppm) = Reading A
MONOCHLORAMINE
4. Add one Chlorine DPD #2 Tablet (6904A) to sample from Step 3. Cap and mix until tablet disintegrates.
5. Insert test tube into the Octa-Slide 2 Viewer (1101). Match sample color to a color standard. Record as Reading B.
Monochloramine (ppm Chlorine) = Reading B - Reading A
DICHLORAMINE & TOTAL RESIDUAL CHLORINE
6. Add one *Chlorine DPD #3 Tablet (6905A) to the sample from Step 5. Cap and mix until tablet disintegrates.
7. Insert test tube into the Octa-Slide 2 Viewer (1101). Match sample color to a color standard. Record as Reading C.
Dichloramine (ppm Chlorine) = Reading C - Reading B
Total Residual Chlorine, ppm = Reading C
TOTAL RESIDUAL CHLORINE & COMBINED CHLORINE
1. Rinse test tube (0106) with sample water. Fill to 5 mL line with sample water.
2. Add one *Chlorine DPD #1 Tablet (6999A). Cap and mix until tablet disintegrates.
3. Immediately insert test tube into the Octa-Slide Viewer (1101). Match sample color to a color standard. Record as ppm Free Available Chlorine.
4. Add one *Chlorine DPD #3 Tablet (6905A) to the same sample. Cap and mix until disintegrated.
5. Insert test tube into Octa-Slide 2 Viewer (1101). Match sample color to a color standard.
Record as ppm Total Residual Chlorine.
Combined Chlorine, ppm =
Total Residual Chlorine - Free Available Chlorine
TOTAL RESIDUAL CHLORINE
The *Chlorine DPD #4 Tablet determines Total Residual Chlorine and is used where a differentiation in the type of chlorine is not required.
1. Rinse the test tube (0106) with sample water. Fill to 5 mL line with sample water.
2. Add one *Chlorine DPD #4 Tablet (6906A). Cap and mix until tablet disintegrates.
3. Insert test tube into Octa-Slide 2 Viewer (1101). Match sample color to a color standard.
Record as ppm Total Residual Chlorine.
Loading...