Read these Directions for Use carefully, and
become thoroughly familiar with the operation
and maintenance of the LSU before using it.
The LSU should only be used by persons trained
in the use of medical suction equipment.
The LSU is not suitable for use in the presence
of flammable liquids or gases;
there can be a danger of explosion or fire.
Do not use the LSU under environmental
conditions that are outside the ranges specified
under section 11.6.This can endanger safety and
adversely affect operation of the device.
Battery:
- To maintain satisfactory operation of the
battery, it is recommended to place the LSU
on continuous charge when not in use.
- The LSU must be placed on charge for a
minimum of 24 hours to reach full battery
capacity. The fast charging gives approximately
80% battery capacity after 3 hours (for a new
battery). Please note that repetitive 3 hour
charging is not recommended.
- If it is not possible to place the LSU on
continuous charge when not in use, make sure
the battery is charged for a minimum of
24 hours at least once a month.
- Do not store the battery when it is
discharged; always fully charge the batter y
before storage.
Do not immerse the LSU or allow it to stand in
water or other liquids.This might damage the
device, and cause electrical hazard.
Do not pump any cleaning solution or other liquids through the vacuum pump, i.e. through the
Vacuum Connector.This can damage the LSU.
Use only accessories supplied by Laerdal or one
of our authorised distributors to ensure that the
LSU operates satisfactorily.
Do not block the Exhaust Outlet during use. This
will lead to reduced flow and can also cause
damage to the LSU.
Do not use the Reusable version without a filter
or the Float Ball. Use only the disposable Abbott
liner with filter.
If overflow of liquid from the Canister into the
pump is suspected, the LSU must be returned for
service (see section 9). Overflow of suctioned
material can damage the device.
Disconnect the LSU from external power prior
to cleaning. Use a minimum of liquid to prevent
any electrical shock hazard.
- Remove the LSU from its packaging and
inspect all parts for damage. If there are any
signs of damage or if any parts are missing,
immediately notify Laerdal or one of our
authorised distributors.
- Insert the battery according to section 6.7.
- Place the LSU on charge for a minimum of
24 hours to make sure you have a fully
charged battery (see section 6.5).
- After being charged for 24 hours, perform the
Device Test according to section 7.
2.2 Indications for Use
The LSU is a portable, electrically powered,
medical suction equipment intended for field and
transport use. It is intended for intermittent
operation to remove secretions, blood or vomit
from a patient´s airway to allow ventilation.
Higher vacuum levels are generally selected for
oropharyngeal suctioning, and lower vacuum
levels are usually selected for tracheal suctioning
and the suctioning of children and infants.
2.3 Contraindications
None known.
2.4 Features and Benefits
- Splashproof, rugged and solid.
- Easy access to the tubing.
- Equipped with an internal battery that can
easily be replaced without using tools.
- Charging of battery can be done directly from
AC or DC mains - no external charger is
needed.
- Fast charge option of 3 hours, giving approx.
80% battery capacity.
- Operation possible from internal battery and
directly from AC or DC mains - no adapter is
needed.
- A large combined ON/OFF switch and
vacuum selector makes operation easy even
with heavy-duty gloves.
- A green LED bargraph makes it easy to
observe the actual vacuum level during
operation.
- Another green LED bargraph indicates approx.
remaining battery capacity.
2.5 Limited Warranty
The LSU comes with a five (5) year limited
warranty*. See the enclosed “Laerdal Global
Warranty” for terms and conditions.
The warranty is also available at
www.laerdal.com
* Excluding the canister and tubing systems.
4
3 Description
3.1 Quick Reference Guide
3.1.1 LSU w/Reusable Canister System
ENGLISH
Aerosol Filter
Vacuum Connector
VACUUM Inlet
User Interface
Canister
Canister Holder
Release Arm
3.1.2 LSU w/Disposable Canister System
Vacuum Connector
PATIENT Inlet
Patient Suction Tubing
Suction Catheter
Adapter
Holder for Water
Bottle/Patient
Suction Tubing.
Water Bottle
Canister Lid
with Float Ball
Battery Door
PATIENT Inlet
VACUUM Inlet
User Interface
Canister
Canister Holder
Release Arm
Patient
Suction
Tubing
Vacuum tubing
Abbott liner
with filter
T-bar
Battery Door
5
TEST-Button
Vacuum Indicator
Power ON Indicator
External Power Indicator
Failure Mode Indicator
3.2 User Interface
3.2.1 Operating Knob
The Operating Knob is a combined ON/OFF
switch and vacuum selector, located at the bottom of the User Interface.
The LSU is switched ON by turning the
Operating Knob to the required setting. Each setting indicates the maximum achievable vacuum
level for the selected position (80, 120, 200, 350,
500+ mmHg).The free air flow value changes
along with the increasing vacuum settings
(see section 11.4 for details).The LSU is switched
OFF by resetting the Operating Knob to "0".
Even if the Operating Knob is in “0” position,
voltage is present on some of the internal circuitry, when the LSU is connected to mains.To fully
remove power, pull out the Power Cord.
3.2.2 Indicators
3.2.2.1 Power ON Indicator
This green LED has 3 modes:
- It is lit continuously while the LSU is
switched ON.
- It flashes rapidly (approx. twice per second)
during the Device Test.
Battery Status Indicator
Operating Knob
- It flashes slowly (approx. once per second)
while the Automatic Power-save Function is
activated, if the Device Test is interrupted or
when the battery is discharged.
3.2.2.2 External Power Indicator
This green LED is continuously lit
while external AC or DC power is connected.
3.2.2.3 Failure Mode Indicator
The red LED is lit when a possible
malfunction of the LSU has been
detected. If lit, turn the LSU OFF, and then ON
again to check if the indication disappears.
Provided that the indication does not occur
again the LSU can be operated. If the Indicator
continues to lit after three OFF/ON cycles and
after replacing the battery with a fully charged
battery, discontinue use and return the LSU for
service. (See section 9).
6
3.2.2.4 Vacuum Indicator
This green LED bargraph displays
the actual vacuum level during
operation of the LSU.
Each fully lit segment represents
50 mmHg. If a segment has a
weak light, this represents
25 mmHg (e.g 125 mmHg
means 2 fully lit and 1 weakly
lit segments).
mmHg 80 120 200 350 500
kPa 1116274767
mBar107160267467667
CAPACITY for a
new battery stored in
room temperature OUTPUT
> 75 %4 LEDs lit.
50 - 75%3 LEDs lit.
25 - 50%2 LEDs lit.
< 25%1 LED lit.
Battery Low; (min. 2 minutes
of operation left based
on 500+ mmHg/free flow).1 LED flashes.
During charging: indicates approx. achieved
battery capacity.
CAPACITY OUTPUT*
< 75%The LED's will be
lit sequentially.
75 - 80%3 LED's lit and
4th flashing.
> 80%4 LED's lit.
ENGLISH
3.2.2.5 Battery Status Indicator
This green LED bargraph has 3 functions:
- Indicates remaining battery capacity
during operation from internal batter y
- Indicates approx. achieved battery
capacity during charging
- Indicates which test is currently
in progress during device test
During operation from internal batter y:
- indicates approx. remaining battery capacity.
During operation from internal batter y and during charging the displayed values must only be
used as indications.
Several parameters can influence them e.g.
device settings, state of battery, temperature etc.
Note:
Immediately after turning the LSU ON and
immediately after switching from external power to
internal battery operation, all 4 LED's will flash for
5 seconds before the remaining battery capacity
is displayed.
*Note:
If no battery is installed, the battery status indicator
will be fully lit for approx. 5 sec. until it is turned off.
During the Device Test: indicates which step of
the test is currently in progress or which corre-
sponding test result is being displayed:
LED 1 (lower LED) lit = STEP 1,
LED 2 lit = STEP 2 etc.
3.2.3 TEST-Button
This button allows you to run a 4-step
user initiated Device Test program to identify
whether the LSU is operating satisfactorily or if it
needs service (see section 7).
7
3.3 Canister Systems
The LSU can be delivered with a Reusable or a
Disposable Canister System as described below.
Both Canisters have a corresponding sized
Canister Holder.
3.3.1 LSU Reusable Canister System
Patient Suction Tubing
Angled Connector
Suction Catheter Adapter
Canister
Gasket
Vacuum seal
The LSU Reusable Canister System consists of:
- A 1000 ml transparent plastic Canister.
- A Canister Lid w/Float Ball that prevents
overflow if the Canister is full or the LSU
tips over.
- An Aerosol Filter with tubing that protects
the LSU by preventing aerosols from entering
the Pump Unit.*
The Canister Lid contains a Float Ball that shuts
off the vacuum if the canister is full or the LSU
tips over.To reset the valve, remove the Angled
Connector from the VACUUM Inlet.
The LSU Reusable Canister System must
not be used without the Aerosol Filter
or the Float Ball.
It is recommended always to have extra Aerosol
Filters with the LSU in case one has to be
discarded.
PATIENT Inlet
VACUUM Inlet.
Canister Lid w/Float Ball
Aerosol Filter
Decontamination must be performed in accordance with section 6.4.
For the Aerosol Filter the instructions below
must be followed:
- The Aerosol Filter is not designed for
decontamination.
- It is recommended that the Aerosol Filter be
replaced after each use or at least once every
shift. If the LSU is used on patients or in areas
where cross contamination is not an issue, the
Aerosol Filter should be replaced at least
once a month.
- If the Aerosol Filter becomes wet, it should
be replaced immediately or as soon as
possible after use.
*The Aersol Filter is not intended for microbiological or for particle filtration.
8
3.3.2 Disposable Canister System
Disposable Parts
VACUUM Inlet
Disposable Liner
including lid with
filter, tube and
connectors
PATIENT Inlet
Patient Suction Tubing
Reusable Parts
ENGLISH
Reusable Canister
T-bar
Vacuum Connector
Tubing
The Disposable Canister System consists of:
- A 1000 ml reusable transparent plastic
Canister including a Canister Connector.
- A disposable liner including lid with filter, tube
and connectors.
The Abbott Receptal® Canister contains a
hydrophilic filter that shuts off the vacuum if the
canister is full or the LSU tips over.To restore the
vacuum, replace the inner liner.
Note:
Do not use an Abbott liner without a filter!
The disposable parts must be replaced after each
use.The Reusable Canister must be cleaned as
stated on the canister. The Vacuum Connector
Tubing is reusable and should be cleaned in
accordance with section 6.3.
A Filter Tubing is available for the Abbott Canister
System (see section 8).
3.4 Power Sources
The LSU is delivered with an AC Power Cord for
connection to AC mains and a DC Power Cord
for connection to DC mains.
The device can be operated from the internal
battery, and can be operated or charged from
one of the following external Power Sources:
- AC mains when used with the AC Power
Cord: 100-240 VAC (50/60/400 Hz).
- DC mains when used with the DC Power
Cord: 12-28 VDC .
A Wall Bracket to hold the LSU during operation
and (optional) charging is available separately
(see section 8).
AC Connector Socket
DC Connector Socket
External power connection will cause the
External Power Indicator to be lit.
Always check that it is lit when
connecting external power or
inserting the LSU into the Wall Bracket.
3.5 Automatic Power-save Function
The LSU has an Automatic Power-save Function
that switches the pump motor off.
While in this mode, the Power ON
Indicator will flash slowly
(approx. once per second).
This function will be activated when the
Operating Knob is set to 200, 350 or 500+
mmHg and the actual vacuum level has been
continuously higher than 120 mmHg for more
than 2 minutes.
To exit the Power-save Mode and revert to
normal operation, set the Operating Knob to
any other position and then go back to required
setting.
9
Loading...
+ 19 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.