Laerdal 7800 Suction Unit User manual

78 00 10 78 00 00
Directions for Use
Laerdal Suction Unit
www.laerdal.com
1 Cautions and Warnings
Read these Directions for Use carefully, and become thoroughly familiar with the operation and maintenance of the LSU before using it. The LSU should only be used by persons trained in the use of medical suction equipment. The LSU is not suitable for use in the presence of flammable liquids or gases; there can be a danger of explosion or fire. Do not use the LSU under environmental conditions that are outside the ranges specified under section 11.6.This can endanger safety and adversely affect operation of the device.
Battery:
- To maintain satisfactory operation of the battery, it is recommended to place the LSU on continuous charge when not in use.
- The LSU must be placed on charge for a minimum of 24 hours to reach full battery capacity. The fast charging gives approximately 80% battery capacity after 3 hours (for a new battery). Please note that repetitive 3 hour charging is not recommended.
- If it is not possible to place the LSU on continuous charge when not in use, make sure the battery is charged for a minimum of 24 hours at least once a month.
- Do not store the battery when it is discharged; always fully charge the batter y before storage.
Do not immerse the LSU or allow it to stand in water or other liquids.This might damage the device, and cause electrical hazard. Do not pump any cleaning solution or other liq­uids through the vacuum pump, i.e. through the Vacuum Connector.This can damage the LSU. Use only accessories supplied by Laerdal or one of our authorised distributors to ensure that the LSU operates satisfactorily.
Do not block the Exhaust Outlet during use. This will lead to reduced flow and can also cause damage to the LSU.
Do not use the Reusable version without a filter or the Float Ball. Use only the disposable Abbott liner with filter. If overflow of liquid from the Canister into the pump is suspected, the LSU must be returned for service (see section 9). Overflow of suctioned material can damage the device. Disconnect the LSU from external power prior to cleaning. Use a minimum of liquid to prevent any electrical shock hazard.
2
Contents
1 Cautions and Warnings . . . . . . . . . . . . . . . .2
2 Introduction . . . . . . . . . . . . . . . . . . . . . . . . .4
2.1 Unpacking and prepare
the LSU for use . . . . . . . . . . . . . . . . . . . . . 4
2.2 Indications for Use . . . . . . . . . . . . . . . . . . 4
2.3 Contraindications . . . . . . . . . . . . . . . . . . . 4
2.4 Features and Benefits . . . . . . . . . . . . . . . . 4
2.5 Limited Warranty . . . . . . . . . . . . . . . . . . . 4
3 Description . . . . . . . . . . . . . . . . . . . . . . . . 5
3.1 Quick Reference Guide . . . . . . . . . . . . . . 5
3.2 User Interface . . . . . . . . . . . . . . . . . . . . . . 6
3.3 Canister Systems . . . . . . . . . . . . . . . . . . . . 8
3.4 Power Sources . . . . . . . . . . . . . . . . . . . . . . 9
3.5 Automatic Power-save Function . . . . . . . . 9
4 Operation LSU
with Reusable Canister . . . . . . . . . . . . . . 10
4.1 Getting Started . . . . . . . . . . . . . . . . . . . . .10
4.2 Emptying the Canister . . . . . . . . . . . . . . .10
4.3 Assembly of the Reusable
Canister System . . . . . . . . . . . . . . . . . . . . .11
4.4 After Use . . . . . . . . . . . . . . . . . . . . . . . . . . 12
5 Operation LSU
with Disposable Canister. . . . . . . . . . . . . 13
5.1 Getting Started . . . . . . . . . . . . . . . . . . . . 13
5.2 Replace the Disposable Canister . . . . . . 13
5.3 Assembly of the Disposable
Canister System . . . . . . . . . . . . . . . . . . . . 14
5.4 After Use . . . . . . . . . . . . . . . . . . . . . . . . . . 15
6 Maintenance . . . . . . . . . . . . . . . . . . . . . . . 15
6.1 General . . . . . . . . . . . . . . . . . . . . . . . . . . .15
6.2 Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . .15
6.3 Cleaning the Cabinet . . . . . . . . . . . . . . . .15
6.4 Decontamination of Reusable Par ts . . . .16
6.5 Charge the Batter y . . . . . . . . . . . . . . . . . 17
6.6 Check the Batter y Quality . . . . . . . . . . . 17
6.7 Replace the Batter y . . . . . . . . . . . . . . . . . 18
6.8 Fastening brackets . . . . . . . . . . . . . . . . . . 18
6.9 Release arm for Canister holder . . . . . . 18
7 Device Test . . . . . . . . . . . . . . . . . . . . . . . . 18
7.1 Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
7.2 Run the test . . . . . . . . . . . . . . . . . . . . . . . 18
7.3 Evaluation of Device test Results . . . . . . 20
8 Accessories and Par ts . . . . . . . . . . . . . . . 21
9 Servicing . . . . . . . . . . . . . . . . . . . . . . . . . . 22
10 Trouble Shooting . . . . . . . . . . . . . . . . . . . 22
11 Specifications . . . . . . . . . . . . . . . . . . . . . . 23
11.1 Classification . . . . . . . . . . . . . . . . . . . . . . . 23
11.2 General tolerance . . . . . . . . . . . . . . . . . . 23
11.3 Physical Characteristics . . . . . . . . . . . . . . 23
11.4 Operation . . . . . . . . . . . . . . . . . . . . . . . . . 23
11.5 Power Requirements . . . . . . . . . . . . . . . . 24
11.6 Environmental Conditions . . . . . . . . . . . 24
11.7 Material Chart . . . . . . . . . . . . . . . . . . . . . 24
11.8 Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
11.9 EMC Specifications . . . . . . . . . . . . . . . . . 25
12 Addresses . . . . . . . . . . . . . . . . . . . . . . . . . 27
ENGLISH
3
2 Introduction
2.1 Unpacking and prepare the LSU for use
- Remove the LSU from its packaging and inspect all parts for damage. If there are any signs of damage or if any parts are missing, immediately notify Laerdal or one of our authorised distributors.
- Insert the battery according to section 6.7.
- Place the LSU on charge for a minimum of 24 hours to make sure you have a fully charged battery (see section 6.5).
- After being charged for 24 hours, perform the Device Test according to section 7.
2.2 Indications for Use
The LSU is a portable, electrically powered, medical suction equipment intended for field and transport use. It is intended for intermittent operation to remove secretions, blood or vomit from a patient´s airway to allow ventilation. Higher vacuum levels are generally selected for oropharyngeal suctioning, and lower vacuum levels are usually selected for tracheal suctioning and the suctioning of children and infants.
2.3 Contraindications
None known.
2.4 Features and Benefits
- Splashproof, rugged and solid.
- Easy access to the tubing.
- Equipped with an internal battery that can easily be replaced without using tools.
- Charging of battery can be done directly from AC or DC mains - no external charger is needed.
- Fast charge option of 3 hours, giving approx. 80% battery capacity.
- Operation possible from internal battery and directly from AC or DC mains - no adapter is needed.
- A large combined ON/OFF switch and vacuum selector makes operation easy even with heavy-duty gloves.
- A green LED bargraph makes it easy to observe the actual vacuum level during operation.
- Another green LED bargraph indicates approx. remaining battery capacity.
2.5 Limited Warranty
The LSU comes with a five (5) year limited warranty*. See the enclosed “Laerdal Global Warranty” for terms and conditions. The warranty is also available at
www.laerdal.com
* Excluding the canister and tubing systems.
4
3 Description
3.1 Quick Reference Guide
3.1.1 LSU w/Reusable Canister System
ENGLISH
Aerosol Filter
Vacuum Connector
VACUUM Inlet
User Interface
Canister
Canister Holder
Release Arm
3.1.2 LSU w/Disposable Canister System
Vacuum Connector
PATIENT Inlet
Patient Suction Tubing
Suction Catheter Adapter
Holder for Water Bottle/Patient Suction Tubing.
Water Bottle
Canister Lid with Float Ball
Battery Door
PATIENT Inlet
VACUUM Inlet
User Interface
Canister
Canister Holder
Release Arm
Patient Suction Tubing
Vacuum tubing
Abbott liner with filter
T-bar
Battery Door
5
TEST-Button
Vacuum Indicator
Power ON Indicator
External Power Indicator
Failure Mode Indicator
3.2 User Interface
3.2.1 Operating Knob
The Operating Knob is a combined ON/OFF switch and vacuum selector, located at the bot­tom of the User Interface. The LSU is switched ON by turning the Operating Knob to the required setting. Each set­ting indicates the maximum achievable vacuum level for the selected position (80, 120, 200, 350, 500+ mmHg).The free air flow value changes along with the increasing vacuum settings (see section 11.4 for details).The LSU is switched OFF by resetting the Operating Knob to "0". Even if the Operating Knob is in “0” position, voltage is present on some of the internal circuit­ry, when the LSU is connected to mains.To fully remove power, pull out the Power Cord.
3.2.2 Indicators
3.2.2.1 Power ON Indicator
This green LED has 3 modes:
- It is lit continuously while the LSU is switched ON.
- It flashes rapidly (approx. twice per second) during the Device Test.
Battery Status Indicator
Operating Knob
- It flashes slowly (approx. once per second) while the Automatic Power-save Function is activated, if the Device Test is interrupted or when the battery is discharged.
3.2.2.2 External Power Indicator
This green LED is continuously lit while external AC or DC power is connected.
3.2.2.3 Failure Mode Indicator
The red LED is lit when a possible malfunction of the LSU has been detected. If lit, turn the LSU OFF, and then ON again to check if the indication disappears. Provided that the indication does not occur again the LSU can be operated. If the Indicator continues to lit after three OFF/ON cycles and after replacing the battery with a fully charged battery, discontinue use and return the LSU for service. (See section 9).
6
3.2.2.4 Vacuum Indicator
This green LED bargraph displays the actual vacuum level during operation of the LSU. Each fully lit segment represents 50 mmHg. If a segment has a weak light, this represents 25 mmHg (e.g 125 mmHg means 2 fully lit and 1 weakly lit segments).
mmHg 80 120 200 350 500
kPa 1116274767
mBar 107 160 267 467 667
CAPACITY for a new battery stored in room temperature OUTPUT
> 75 % 4 LEDs lit.
50 - 75% 3 LEDs lit.
25 - 50% 2 LEDs lit.
< 25% 1 LED lit. Battery Low; (min. 2 minutes
of operation left based on 500+ mmHg/free flow). 1 LED flashes.
During charging: indicates approx. achieved
battery capacity.
CAPACITY OUTPUT*
< 75% The LED's will be
lit sequentially.
75 - 80% 3 LED's lit and
4th flashing.
> 80% 4 LED's lit.
ENGLISH
3.2.2.5 Battery Status Indicator
This green LED bargraph has 3 functions:
- Indicates remaining battery capacity during operation from internal batter y
- Indicates approx. achieved battery capacity during charging
- Indicates which test is currently in progress during device test
During operation from internal batter y:
- indicates approx. remaining battery capacity.
During operation from internal batter y and dur­ing charging the displayed values must only be used as indications. Several parameters can influence them e.g. device settings, state of battery, temperature etc.
Note: Immediately after turning the LSU ON and immediately after switching from external power to internal battery operation, all 4 LED's will flash for 5 seconds before the remaining battery capacity is displayed.
*Note:
If no battery is installed, the battery status indicator
will be fully lit for approx. 5 sec. until it is turned off.
During the Device Test: indicates which step of
the test is currently in progress or which corre-
sponding test result is being displayed:
LED 1 (lower LED) lit = STEP 1,
LED 2 lit = STEP 2 etc.
3.2.3 TEST-Button
This button allows you to run a 4-step
user initiated Device Test program to identify
whether the LSU is operating satisfactorily or if it
needs service (see section 7).
7
3.3 Canister Systems
The LSU can be delivered with a Reusable or a Disposable Canister System as described below. Both Canisters have a corresponding sized Canister Holder.
3.3.1 LSU Reusable Canister System
Patient Suction Tubing
Angled Connector
Suction Catheter Adapter
Canister
Gasket
Vacuum seal
The LSU Reusable Canister System consists of:
- A 1000 ml transparent plastic Canister.
- A Canister Lid w/Float Ball that prevents overflow if the Canister is full or the LSU tips over.
- An Aerosol Filter with tubing that protects the LSU by preventing aerosols from entering the Pump Unit.*
The Canister Lid contains a Float Ball that shuts off the vacuum if the canister is full or the LSU tips over.To reset the valve, remove the Angled Connector from the VACUUM Inlet.
The LSU Reusable Canister System must not be used without the Aerosol Filter or the Float Ball.
It is recommended always to have extra Aerosol Filters with the LSU in case one has to be discarded.
PATIENT Inlet
VACUUM Inlet.
Canister Lid w/Float Ball
Aerosol Filter
Decontamination must be performed in accor­dance with section 6.4. For the Aerosol Filter the instructions below must be followed:
- The Aerosol Filter is not designed for decontamination.
- It is recommended that the Aerosol Filter be replaced after each use or at least once every shift. If the LSU is used on patients or in areas where cross contamination is not an issue, the Aerosol Filter should be replaced at least once a month.
- If the Aerosol Filter becomes wet, it should
be replaced immediately or as soon as possible after use.
*The Aersol Filter is not intended for microbio­logical or for particle filtration.
8
3.3.2 Disposable Canister System
Disposable Parts
VACUUM Inlet
Disposable Liner including lid with
filter, tube and
connectors
PATIENT Inlet
Patient Suction Tubing
Reusable Parts
ENGLISH
Reusable Canister
T-bar
Vacuum Connector Tubing
The Disposable Canister System consists of:
- A 1000 ml reusable transparent plastic Canister including a Canister Connector.
- A disposable liner including lid with filter, tube and connectors.
The Abbott Receptal® Canister contains a hydrophilic filter that shuts off the vacuum if the canister is full or the LSU tips over.To restore the vacuum, replace the inner liner.
Note: Do not use an Abbott liner without a filter!
The disposable parts must be replaced after each use.The Reusable Canister must be cleaned as stated on the canister. The Vacuum Connector Tubing is reusable and should be cleaned in accordance with section 6.3.
A Filter Tubing is available for the Abbott Canister System (see section 8).
3.4 Power Sources
The LSU is delivered with an AC Power Cord for connection to AC mains and a DC Power Cord for connection to DC mains. The device can be operated from the internal battery, and can be operated or charged from one of the following external Power Sources:
- AC mains when used with the AC Power Cord: 100-240 VAC (50/60/400 Hz).
- DC mains when used with the DC Power Cord: 12-28 VDC .
A Wall Bracket to hold the LSU during operation and (optional) charging is available separately (see section 8).
AC Connector Socket
DC Connector Socket
External power connection will cause the External Power Indicator to be lit. Always check that it is lit when connecting external power or inserting the LSU into the Wall Bracket.
3.5 Automatic Power-save Function
The LSU has an Automatic Power-save Function that switches the pump motor off. While in this mode, the Power ON Indicator will flash slowly (approx. once per second). This function will be activated when the Operating Knob is set to 200, 350 or 500+ mmHg and the actual vacuum level has been continuously higher than 120 mmHg for more than 2 minutes. To exit the Power-save Mode and revert to normal operation, set the Operating Knob to any other position and then go back to required setting.
9
Loading...
+ 19 hidden pages