Specifications are subject to change without notice.
1.1Bench requirements
Ideally, the modules are configured in the following left-to-right order: computer (with
peripherals), sampler, PDS200 Dilutor, reagent pump and system unit (refer to Figure 1).
Note: The sampler and dilutor can be placed on top of the optional shelf accessory in order to
conserve bench space.
Figure 1 QuikChem FIA overview
1Autosampler5Computer and peripherals
2Dilutor (in rear)6Power cords
3Pump7Communication cables
4System unit
The bench space required for a QuikChem® FIA System with up to five channels, dilutor,
sampler, computer and printer is:
•Width: 2.72 m (9 ft)
•Depth: 0.83 m (2.75 ft)
•Height: 0.58 m (2 ft)
5
Specifications
1.2Module dimensions
Table 1 contains the general dimensions for individual modules. Use Table 1 for any
system configuration that differs from the one explained above.
Table 1 Dimensions
ModuleWidth (cm)
PDS20012.7 22.5 23.44
ASX-500 sampler
ASX-400 sampler16 1416.57.3
Reagent pump (12 position)15 26135.6
System unit (5 channels)70.5 71.226.746.5
SPM69 9 264
1
Divide dimension by 2.54 to convert to inches.
2
Multiply weight by 2.2 to convert to pounds.
3
Allow an extra 15 cm of height for probe clearance.
3
1
52.5 462510
Depth (cm)
1
Height (cm)
1
Weight (kg)
Allow a minimum of 10 cm of free space in the back of the instrument and between
modules for cabling and tubing. This extra space also offers easy access to the
system components.
Note: To allow convenient access to the top and rear of the system, avoid putting the system
under shelving.
1.3Electrical and environmental specifications
•Connect all modules to a surge-protected power strip.
2
Note: Refer to the back of the modules for the proper requirement.
•This equipment is intended for use in a laboratory.
•Use cables and power cords supplied by the manufacturer only.
Figure 2 shows the single phase voltage and frequencies worldwide and Table 2
describes the electrical specifications for the instrument modules.
Table 3 describes the environmental specifications for the instrument modules.
Figure 2 General guide to single phase voltage and frequencies worldwide
6
Specifications
Table 2 Individual electrical specifications
Maximum active
power (W),
Module
PDS200110–240 VAC 50–60 Hz0.5 AT, ½ A, 250 V
ASX-500 sampler100–240 VAC50–60 Hz1.5 A
ASX-400 sampler100–120 VAC50–60 Hz200 VA
Reagent pump
System unit
AC input voltage
rating or range
100 or 120 VAC50–60 Hz 15 VA T, ½ A, 250 V
220 or 240 VAC50–60 Hz 15 VAT, ¼ A, 250 V
100–115 V ± 10% 50–60 Hz800 WT, 8 A, 250 V
230 V ± 10% 50–60 Hz800 WT, 4 A, 250 V
Table 3 Environmental specifications
AC input frequency
rating or range
apparent power
(VA) or current
(A)
rating
Fuse rating
Fuse not
user-replaceable
Fuse not
user-replaceable
Temperature 15–35 °C (60–95 °F)
Use Laboratory
Altitude Up to 10,000 feet (3 km) above sea level
Relative humidity Up to 92.5% non-condensing
Pollution degree II
Installation category II
Table 4 Certifications
Certification MarksStandards
CE
cETL CSA C22.2 No 61010-1
ETLus UL 61010-1
IEC/EN 61010-1 per LV Directive (2006/95/EC), IEC/EN 61326 per EMC: Directive
(2004/108/EC
7
Specifications
8
Section 2General Information
2.1Safety information
Read this entire manual before unpacking, setting up or operating this equipment. Pay
attention to all danger and caution statements. Failure to do so could result in serious
injury to the operator or damage to the equipment.
To make sure that the protection provided by this equipment is not impaired, do not use
or install this equipment in any manner other than that specified in this manual.
2.1.1 Use of hazard information
DANGER
Indicates a potentially or imminently hazardous situation which, if not avoided, will
result in death or serious injury.
WARNING
Indicates a potentially or imminently hazardous situation which, if not avoided,
could result in death or serious injury.
CAUTION
Indicates a potentially hazardous situation that may result in minor or moderate
injury.
Important Note: Indicates a situation which, if not avoided, may cause damage to the
instrument. Information that requires special emphasis.
2.1.2 Precautionary labels
Read all labels and tags attached to the instrument. Personal injury or damage to the
instrument could occur if not observed. A symbol, if noted on the instrument, will be
included with a danger or caution statement in the manual.
This is the safety alert symbol. Obey all safety messages that follow this symbol to avoid potential injury. If on the
instrument, refer to the instruction manual for operation or safety information.
This symbol indicates that a risk of electrical shock and/or electrocution exists.
This symbol identifies the location of a fuse or current limiting device.
This symbol, when noted on the product, indicated that the marked item can be hot and should not be touched
without care.
This symbol, when noted on the product, indicated the presence of devices sensitive to Electro-static Discharge
(ESD) and indicate that care must be taken to prevent damage with the equipment.
This symbol, when noted on the product, identifies a risk of chemical harm and indicates that only individuals
qualified and trained to work with chemicals should handle chemicals or perform maintenance on chemical delivery
systems associated with the equipment.
This symbol, if noted on the product, indicates the need for protective eye wear.
This symbol, when noted on the product, identifies the location of the connection for Protective Earth (ground).
Electrical equipment marked with this symbol may not be disposed of in European public disposal systems after
12 August of 2005. In conformity with European local and national regulations (EU Directive 2002/96/EC),
European electrical equipment users must now return old or end-of life equipment to the Producer for disposal at no
charge to the user.
Note: For return for recycling, please contact the equipment producer or supplier for instructions on how to return
end-of-life equipment, producer-supplied electrical accessories, and all auxiliary items for proper disposal.
9
General Information
2.2General product information
The system unit manages five sample processing modules or channels, configured for
performing flow injection analysis with photometric detection. The number of channels
can be expanded to a maximum of eight by adding a second core. Alternate
configurations, such as potentiometric and amperometric detection methods are
described in the analytical method where they are required. For complete information on
the software, refer to the QuikChem FIA Omnion
system in the software.
The QuikChem
the following modules:
•Dilutor (optional)—The PDS200 uses positive displacement to draw diluent water to
mix with non-flammable samples for analysis. Refer to section 3.4 on page 25.
•Sampler—The QC8500 Analyzer works with either the ASX-400 or ASX-500 series
autosamplers. Refer to section 3.3 on page 21.
•Reagent Pump (12 or 16 cartridges)—The multichannel peristaltic pump is used to
pump reagents into the manifold. It also aspirates sample aliquots from the sampler to
the valve. These portions are then loaded into the sample loop and swept into the
manifold by the carrier stream to mix with the reagents.
•QuikChem FIA System Unit—The System Unit consists of the core unit and the
sample processing modules (SPM). The core unit distributes power, light and
electronic signals to and from each SPM. A sample processing module, also known as
a channel, typically consists of an injection valve, a manifold, a photometric detector
and a heating module (optional). A System Unit can have up to five channels that run
simultaneously. A single computer can control two System Units in order to run up to
eight channels at once. A chemical reaction module or manifold is installed on a
channel for the quantitative analysis of an analyte. There are more than 450
QuikChem methods available for determining over 40 analytes; many of them in
several matrices.
®
8500 Series 2 analyzer is modular in design (Figure 3) and consists of
®
Software Manual or the on-line help
•Computer
•Printer (optional)
Note: If any items are missing or damaged, contact the manufacturer or sales representative
immediately. Ship the components returned for service in the original packaging material to protect
against damage during transportation.
Note: The Ion Chromatography option is available with this instrument. Refer to the QuikChem Ion
Chromatography User Manual for instructions.
10
Figure 3 QuikChem® FIA automated ion analyzer with shelf (Catalog No. 31079)
Section 3Installation
DANGER
Only qualified personnel should conduct the tasks described in this section of the
manual.
3.1System unit installation
3.1.1 Carrying instructions
CAUTION
Potential Lifting Hazard. A five-channel system weighs 47 kg (103 lb). Do not
unpack, carry or move the system without proper equipment or sufficient number
of people. Make sure that the system is lifted with leg support and not back
support. People with a history of back problems or cardiovascular problems are
prohibited to lift, move or unpack this system.
•Lift the System Unit and channels from the bottom. Do not lift by the front panel.
•Always support the entire instrument (e.g., with a cart or a board) during transport.
3.1.2 Connect the power strip
DANGER
Electrical Shock Hazard. The facility circuit breaker rating must not exceed 20 A.
DANGER
Electrical Shock Hazard. A good low impedance earth ground connection is
required to the equipment cord plug.
DANGER
Potential Electrocution Hazard. If equipment is to be used in wet or potentially wet
locations, always connect to power via an outlet with a Ground Fault interrupt
Circuit (GFIC) or a residual Current Circuit Breaker.
To connect the power strip:
1. Connect the power strip to a power supply.
2. Position the power strip on the bench so that other components can be connected.
3. Leave the power strip off during the installation of other system components.
3.1.3 Connect the system unit
Note: Do not position the equipment in any way that will prevent access to the On/Off switch. Refer
to Figure 4 on page 12, item 1.
To connect the system unit:
1. Connect the power cord to the power outlet on the left side of the System Unit
(Figure 4 on page 12). Connect the other end of the power cord to the power strip.
Leave the power strip switch off.
2. Locate the 9-pin connector of the cable attached to the reagent pump. Plug this
connector in the outlet for Pump 1 (Figure 4 on page 12, item 4) on the interface
panel, located on the left side of the System Unit.
3. Connect one USB cable to the outlet labeled USB In. The other end will connect to
the computer.
11
Installation
1
2
1
2
3
5
4
Figure 4 System unit connection
1On/Off switch4Pump connectors
2Power for system unit5USB in (from computer)
3High pressure pump (IC only)
3.1.4 Flow cell installation
To install the flow cell:
1. Insert the flow cell into the detector with the printed side (bottom) down. Since the
flow cell is keyed, insert the flow cell only in the corresponding slot of the detector
(Figure 5).
2. Make sure the flow cell is pushed all the way in the slot. The bottom flared tubing
connector on the flow cell is the inlet and the top connector is the outlet. During the
installation of a manifold (reaction module), it is very important to make the
connections accordingly, so air bubbles do not get trapped in the flow cell. Refer to
Figure 6 on page 13.
3. Connect the top connector (outlet) to the waste line.
Figure 5 Flow cell
1To waste3Inlet
2Flow cell body
12
Figure 6 Flow cell connectors
1
2
3
20 mm
20 mm
[0.79 in]
[0.79 in]
1To waste2Inlet (from manifold)
3.1.5 Waste and sample line installation
3.1.5.1 Sample line
A sample line consists of a union connector, 130 or 190 cm of Teflon® tubing 0.8 mm
(0.032 in.) ID, two pump tube adapters, a green-green pump tube with its ends trimmed
to 2 cm past both green tabs and 30 cm of Teflon tubing 0.8 mm (0.032 in.) ID.
To install the sample line:
Installation
1. Trim the ends of a green-green pump tube to 2 cm past both green tabs (Figure 7).
This procedure only applies to the green-green pump tube used in a sample line. All
other pump tubes do not need to be trimmed.
Note: Refer to the QuikChem method to find the pump tube color to be used as a sample line.
Most manifolds will use a green-green pump tube, but there are some exceptions.
Figure 7 Trimming the green-green pump tube
1Scalpel2Green-green pump tube3Green tab (2x)
2. Attach the sample line pump tube on a pump tube cartridge (Figure 8). Do not clamp
down the cartridge on the pump. Refer to Section 4 on page 29 for further details.
Figure 8 Sample line pump tube and cartridge
1From probe2To injection valve
3. Connect the 30 cm Teflon tubing to port 6 of the injection valve.
4. Connect the 130 or 190 cm Teflon tubing to the probe on the sampler.
13
Installation
3.1.5.2 Waste line
A waste line consists of a 13 cm of Teflon tubing 0.8 mm ID, a pump tube adapter, 150
cm of Tygon tubing 1.5 mm ID, a glass weight and a plastic nipple. Refer to Figure 9.
Figure 9 Waste line
113 cm Teflon 0.8 mm ID5To waste container
2Adapter6Nipple
3150 cm Tygon 1.5 mm ID7From flow cell or valve (port 5)
4Glass weight
To install a waste line:
1. Connect a waste line to port 5 of the injection valve and the top connector of the
flow cell.
2. When running two or more channels simultaneously, interconnect adjacent valves by
installing 15 cm of 0.8 mm Teflon tubing between ports 5 and 6 (Figure 10 on
page 15). The waste channel farthest from the System Unit will drain directly into the
waste container. Refer to Section 4 on page 29 for details.
14
Installation
Figure 10 Flow injection connections
1System unit5Waste lines
2Channel 16Waste container
3Channel 27Waste line
4Channel 38Sample line
3. Place the glass weight end of the waste line in a proper waste container. Never
submerge the waste line in the waste solution because this may cause back pressure
and flow restrictions. Make sure the waste solution always drips into the container.
3.1.6 Manifold installation
The manifold is the reaction module where the chemistry occurs.
Specific installation requirements are detailed in the Figure 28 on page 32 included with
each chemistry method. Refer to Figure 11 on page 16 and the following steps to install
the manifold for flow injection analysis.
To install the manifold:
1. On the valve, connect port 1 to port 4 with the sample loop.
2. Connect the manifold outlet to the flow cell inlet (Figure 11 on page 16, item 6).
3. Install the interference filter (Figure 11 on page 16, item 5).
4. Connect the carrier line to port 2 on the valve (Figure 11 on page 16, item 8).
5. Connect the tubing from port 3 to the connector labeled From Valve.
6. Attach one end of a waste line to the flow cell outlet. Drop the other end into a waste
container (Figure 11 on page 16, item 4).
15
Installation
Figure 11 Flow injection analysis
1Sample line6From manifold outlet
2Sample loop7Manifold configuration varies
3To manifold inlet8Carrier line
4To waste9Waste or next valve
5Interference filter
3.1.7 Data system installation
Follow the manufacturer’s instructions for the setup of the computer, keyboard, monitor
and mouse.
For Data Systems purchased from the manufacturer, the Omnion software is installed at
the factory.
For customer-supplied systems, refer to the Omnion User Guide for software
installation instructions.
3.2Pump installation and power up
3.2.1 Pump installation
To install the pump:
1. Set the pump on the bench (Figure 3 on page 10 and Figure 1 on page 5).
2. Connect the power cord to the power strip. Leave the power strip switch off.
3. Connect the power cord to the mains supply socket on the back of the pump
(Figure 12 on page 17).
16
Figure 12 Power cord connection
1Power cord2Main supply socket
4. Connect the analog cable to the back of the pump (Figure 13).
Installation
Figure 13 Analog cable connection
5. Connect the other side of the analog cable to the left panel of the QuikChem System
Unit. Refer to section 3.1.3 on page 11.
17
Installation
3.2.2 Apply power to the pump
To apply power to the pump:
1. Turn on the power strip, then turn on the power switch on the back of the pump
(Figure 14). In an RP-150 Model pump, the top panel will show the set speed of 35.
1Power switch
3.2.3 Pump tubing installation
To install the pump tubing:
1. Remove all of the pump cartridges. Press the cartridge holder against the pump, then
lift the cartridges out (Figure 15).
Figure 14 Power switch location
Figure 15 Pump cartridge removal
1Cartridge2Cartridge holder
18
2. Place the pump tubing on the cartridges (Figure 16). Make sure that the tabs lock into
12
3
5
7
6
4
the adaptor.
Figure 16 Placement of pump tubing on the cartridges
1Tension lever5To injection valve or
2Pump tube cartridge6Pump tube
3Adapter7From probe or reagent
4Pump tube tab
Installation
3. Move the tension lever to the far right position (Up position).
4. Install the pump cartridge. Engage one side of the cartridge on the cartridge holder,
then clamp down the other side (Figure 17). The tension levers are always on the left
side of the pump. An arrow indented on top of the cartridges shows the flow direction.
Make sure the pump is moving when the cartridge is clamped down.
Figure 17 Pump cartridge installation
19
Installation
5. Move the tension lever to the left until it clicks into place (Figure 18). Make sure the
tension lever is in the 12 o'clock position.
Note: Move the lever to the right to remove tension.
6. Store unused cartridges.
1One click back
3.2.4 Reagent pump initial test
To test the reagent RP-150 pump:
1. Make sure that the Omnion software is not active at this time. Press the
RUN
button. Make sure the pump is running at a normal speed of 35. The NORMAL
LED will be illuminated. If the display shows a different speed, press the
button to change the setting to 35.
2. Press the
of 4.
3. Hold down the
speed of 999. The pump will go back to the original speed when the
is released.
4. To stop the pump, press the
but the display will show “35” (the pump speed).
5. Press the
The pump will be running at normal speed of 35.
6. Press the
above, turn off the pump and start with step 1.
Figure 18 Tension lever setup
NORMAL
ARROW
MIN button. The MIN LED will be illuminated. The display will show a speed
MAX button until the MAX LED is illuminated. The display will show a
MAX button
MANUAL RUN/ STOP button. No LEDs will be illuminated,
MANUAL RUN/ STOP button again. The MANUAL LED will be illuminated.
MANUAL RUN/ STOP button to stop the pump. To repeat any of the steps
20
7. If any of the above steps fail, contact Technical Support or a local
sales representative.
Note: Contact Technical Support for queries regarding the installation and testing of a different
peristaltic pump model.
3.3Sampler installation
3.3.1 Sampler type
The QuikChem 8500 Series 2 can be used with either the ASX-500 Series Sampler or
the ASX-400 Series Sampler (Figure 19). Installation is the same for both samplers.
Installation
Figure 19 Autosamplers
1ASX-500 Autosampler2ASX-400 Autosampler
3.3.2 Unpack the ASX sampler
Note: If any of these items are missing or damaged, contact the manufacturer or sales
representative immediately. Ship the instruments returned for service in the original packaging
material to protect against damage during transportation.
ASX samplers are shipped with:
•Sample tray(s), assembled and installed
•Probe tube in styrofoam container
•Tubing kit with waste line, sample line and probe
To unpack the sampler:
1. Remove the sampler from the shipping carton and protective plastic and place on the
laboratory bench.
2. Remove the probe arm from the styrofoam container and lay carefully on top of the
sample racks.
3. Open the tubing kit and identify the components required for fluidic connections.
21
Installation
3.3.3 Sampler setup
To connect the sampler:
1. Connect the power cord to the power strip. Leave the power strip switched off.
2. Connect the supplied RS232C cable to the 9-pin connector (COM1) on the back of
the sampler.
3. Connect the other end of the RS232C cable to COM1 on the computer.
3.3.4 Mount Z-drive
Except for the Z-drive, the sampler is assembled at the factory.
To mount the Z-drive on the sampler arm:
1. Remove thumbscrews and bushings from the drive block located on the sampler arm
(Figure 20 on page 23).
2. Attach the x-axis slider block on the arm tube.
3. Secure the slider block finger-tight with the 12-mm thumbscrews (installed from
the top).
3.3.5 Probe installation
To install the probe:
1. Loosen the probe knob on the probe tube slider and remove the piece of tubing
installed for shipping. Lift the probe tube slider until it stops.
Note: The tan knob will come off without the tubing or probe to hold it in place.
2. The guide slider will move to the upper position with the probe tube slider. Use the
wheel in the back of the sampler to lower both sliders until the guide slider is about
1 mm from the top. Make sure that the two sliders are in contact.
Note: Always move the sliders using the wheel at the back of the sampler to avoid crimping the
PEEK tubing.
3. Install the sample probe through both sliders, the probe knob and the opening on the
probe guide plate. Keep the sliders about 1 mm from the top. Adjust the probe so the
tip is 3 to 6 mm below the guide plate. To secure the probe, tighten the probe knob on
the probe tube slider.
4. Move the sampler arm to the rinse station (extreme left position in the standards
rack). Verify if the probe tip clears the top of the rinse station when the sampler is in
the home position.
22
Installation
Figure 20 Probe tube assembly
1Sample transfer tubing 7Slider block guide rail
2Sipper assembly tube 8Home position flag
3Z-drive assembly 9Sample probe guide plate
4Thumbscrews with bushings 10 Z-axis slider
5X-axis block 11 O-ring
6Y-axis lead screw
23
Installation
3.3.6 Fluidic connections
3.3.6.1 Connect the rinse station
Rinse solution is pumped into the bottom of the rinse station and drains to a waste
container through drain tubing attached at the top of the station (Figure 21).
To connect the rinse station:
1. Connect the rinse-in end of the wash line assembly to the inlet fitting (bottom).
2. Install the pump tube with the purple pump tube tabs on the pump. Refer to
section 3.2.3 on page 18.
Note: The outlet of the tube on the right side of the pump attaches to the inlet of the wash bath
(Figure 21, item 1).
Important Note: Do not lock down the right side of the pump cartridge until ready to
perform the test.
3. Drop the other end of the wash line into the rinse solution.
4. Connect one end of the Tygon drain tubing to the drain fitting (top). Drop the other
end into a waste container.
Figure 21 Rinse station diagram
1Drain fitting 5Inlet fitting
2Wash bath 6Rinse-in tubing
3Mounting block 7From pump
4Flow 8Drain tubing for waste
3.3.6.2 Sample line installation (no dilutor)
To install the sample line:
1. Connect the end of the sample line with the union fitting to the probe.
2. Install the pump tubing (green collar locks) in the pump. Refer to section 3.2.3 on
page 18.
Note: Do not lock down the right end of the pump tubing.
3. Connect the sample line to the injection valve on the system unit.
24
3.3.7 Connect the sampler to a PDS200 dilutor
Systems that use a PDS200 dilutor require the ASX-500 Series Sampler and a dual
probe. Refer to section 3.4 for installation instructions.
3.4Dilutor installation
The PDS200 uses positive displacement to draw deionized water or other diluent to mix
with non-flammable samples for analysis.
For operation, the PDS200 dilutor requires:
•An ASX-500 Series autosampler with upgraded firmware, ASX-500 Series serial
number 080313 or greater. For older ASX-500 samplers, contact Technical Support.
•Installation of Omnion 3.0.218 or higher.
3.4.1 Cable connections
To install the dilutor:
1. Make sure that the dilutor power switch is off (Figure 22).
2. Turn off and disconnect the power strip.
3. Connect the Desktop Power Supply with power cable (Catalog No. 5011200) to the
dilutor (Figure 22). Plug the other end into the power strip.
Installation
4. Connect the male connector of the 9-pin serial cable to the back of the dilutor
(Figure 22). Connect the other end to the dilutor port on the back of the autosampler.
Figure 22 PDS200 dilutor
1Power indicator light4Power cable connector
2On/Off switch 5Solenoid valve
3Serial cable connector
25
Installation
5
3
4
1
2
3.4.2 PDS200 connections
Note: Probe speed on existing autosamplers has been reduced to improve performance. The
autosampler may produce a different sound as a result.
To connect the PDS200:
1. Connect the DI Water Line to Port A of the solenoid valve of the dilutor. Place the
other end of the line into the DI water (Figure 23).
2. Connect the Sample Line Assembly to port B of the solenoid valve
(Figure 23, item 3).
3. If necessary, attach the Dual Probe Assembly to the autosampler. The probe guide
may need to be rotated to accommodate the dual probe.
4. Connect the probe sample line (attached to the bent tube of the Dual Probe
Assembly) to port C of the solenoid valve (Figure 23, item 5).
5. Make sure the ferrule on the line from the straight tube of the Dual Probe Assembly is
flush with the end of the tubing (Figure 23, item 2).
6. Insert the connector from the straight tube of the Dual Probe Assembly into port D of
the dilutor (Figure 24 on page 27) and tighten.
7. Make sure the ferrule on the dilutor end of the Diluent Line Filter Assembly is flush
with the end of the tubing (Figure 23, item 2).
8. Insert the connector from the Diluent Line Filter Assembly into port E of the dilutor
and tighten (Figure 23, item 3).
9. Insert the filter end of the Diluent Line Filter Assembly into the diluent (Figure 23,
item 2).
Figure 23 Tubing connectors and ferrule placement
1Ferrule4To DI water port A
2Flush5To sample probe line port C
3To pump/analyzer port B
26
Installation
Figure 24 Dilutor and solenoid valve fluidic connections
1To sample probe7Dilutor power connection
2To diluent8Dilutor power supply
3Dual probe assembly9Sampler power supply
4Dilutor/sampler serial cable10 To pump/analyzer
5Sampler/computer connection11 DI water line
6Sampler power connection12 Diluent line filter assembly
27
Installation
28
Loading...
+ 74 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.