Labnet MINI-CENTRIFUGE User Manual

OPERATING INSTRUCTIONS FOR
MINI-CENTRIFUGE (MODEL C1202)
General Information:
The Model C1202 Mini-centrifuge is a personal centrifuge supplied with a strip tube rotor. The strip tube rotor is designed to centrifuge two 8 x 0.2ml strip tubes. It will also accept up to 16 individual 0.2ml tubes. The Mini-centrifuge is designed for applications requiring relatively low g-forces, such as microfiltration, cell separation and quick spin downs from the walls of tubes. Please read this manual thoroughly before attempting to operate the centrifuge.
Operation:
To begin a run, simply close the lid of the centrifuge. No other controls are required. To stop rotation, press down on the lid release tab on the front of the unit. This will release the lid lock and the rotor will slow gradually to a complete stop. Caution: Do not attempt to open the lid or remove samples until the unit has come to a complete stop. After the rotor has stopped, the lid may be opened by grabbing it with the thumb on the front and fingers on the back then lifting the lid back on the hinge.
Safe operation of the centrifuge requires that the rotor be loaded in a balanced fashion. Failure to load the rotor properly presents a hazardous condition and may damage the centrifuge. Strip tubes and individual tubes must always be loaded symmetrically to ensure proper balance.
Safety Precautions:
NEVER use the centrifuge in any manner not specified in these instructions. NEVER operate the centrifuge without a rotor properly attached to the shaft. NEVER fill tubes while they are in the rotor. Liquid spillage may harm unit. NEVER put hands in the rotor area unless the rotor is completely stopped. NEVER move the centrifuge while the rotor is spinning. NEVER use solvents or flammables near this or other electrical equipment. NEVER centrifuge flammable, explosive or corrosive materials. NEVER centrifuge hazardous materials outside of a hood or proper containment facility . NEVER operate the centrifuge with just one strip tube in place. ALWA YS load the rotor symmetrically. Each strip tube or individual tube should be counterbalanced properly. Weight of
individual tubes may not exceed of 3.0g.
ALWA YS locate the centrifuge within easy access to an electrical outlet. ALWA YS use only microcentrifuge tubes made from plastic and designed to withstand centrifugal forces of at least
2,000 xg.
Cleaning:
Before using any cleaning or decontamination methods except those recommended by the manufacturer, check with the manufacturer that the proposed method will not damage the equipment. To clean the centrifuge, use a damp cloth and a mild, noncorrosive detergent (ph <8). After cleaning, ensure all parts are dried thoroughly before attempting to operate the unit. Do not immerse the centrifuge in liquid or pour liquids over it.
Questions and Service:
Should you have a question about the centrifuge or require service for the unit, please use the contact numbers listed below. Do not send in a unit for service without first calling to obtain a repair authorization number and a decontamination form. Should the centrifuge require service, be sure to decontaminate it first and enclose the decontamination form. The unit should be properly packed to avoid damage. Any damage resulting from improper packaging shall be the responsibil­ity of the user.
Specifications: Spare Parts:
Dimensions........... 5” x5” (127 x 127mm) Replacement fuses: C-1200-F
Electrical ...............115V, 60Hz, 130mA Replacement lid: C-1200-L
Fuse ..................... 0.5A, 250V (F) Replacement rotor: C-1200-RTS
Speed ...................6000rpm
RCF ...................... 1177 x g
or 240V, 50/60Hz, 65mA Replacement base: C-1200-C
PO Box 841 Woodbridge, NJ 07095 Phone 888-LABNET1 or 732 417-0700 Fax 732 417-1750 email labnet@labnetlink.com website www.labnetlink.com