Instruction Manual
Digital Dry Bath
Models D1100 and D1200
Labnet International
PO Box 841
Woodbridge, NJ 07095
Phone: 732 417-0700
Fax: 732 417-1750 email: labnet@labnetlink.com
April 2003
1.0 General Specifications |
|
|
|
Temperature Range |
Ambient +5C to 150C |
||
Temperature display resolution |
0.1C |
4 digit LED |
|
Temperature uniformity |
+/-0.2C |
(at 37C in block) |
|
Temperature accuracy |
+/-0.3C |
|
|
Control |
|
PI Microprocessor controller |
|
I/O |
|
RS232 unidirectional |
|
Dimensions (W x D x H) / Weight |
20 x 26.5 x 8.3 cm 2.2 kg |
||
Electrical requirements: |
|
|
|
D1100 |
115V |
115V, 50/60Hz, 0.8A, Fuse 1.5A |
|
D1100 |
230V |
230V, 50/60Hz, 0.4A, Fuse 1.5A |
|
D1200 |
115V |
115V, 50/60Hz, 1.2A, Fuse 2.0A |
|
D1200 |
230V |
230V, 50/60Hz, 0.6A, Fuse 2.2A |
2.0 Safety Precautions:
-Do not use this product in an explosive environment
-Do not use in the presence of flammable or combustible material
-Do not heat substances that react violently when heated
-Do not touch block when hot or when unit is heating. Use block lifter.
-Do not touch area around block or block well when unit is hot.
-Do not spill liquids into the well area or into the unit side vent holes
-Connect unit only to a properly grounded outlet
Caution: HOT Surfaces that can burn or cause injury.
Use of this product in any manner not specified by the manufacturer or modification of the product may cause injury and/or may void the warranty.
3.0 Installation
Upon unpacking Digital Dry Bath, inspect for damages. Shipping damage is the responsibility of the carrier. Inspect that the following are present: Users manual, Line cord, Block Lifter.
Select a location that is dry and not subject to drafts or moving air from heating or air conditioning vents, or air blown by other equipment. Place the unit on a flat, preferably non-flammable surface. Allow sufficient room around the unit for access and cooling. Six inches minimum on all sides is suggested. Plug the unit into a properly grounded outlet. Using the lifter, insert the block(s) into the well. The unit is now ready for use.
4.0 Controls
-On / Off rocker switch. Back of unit. Turns primary power on and off
-“Start” LED, red. Illuminates when unit is in heating mode, off in temperature set mode.
-“Set” LED, green. Illuminates when unit is in temperature set mode and off when in heating mode.
-“Heating”LED, red. Illuminates when unit is in heating mode and unit is actually applying heat to the block. This LED is on continuously during heat-up and cycles on and off when the unit is at the set temperature
-“UP” arrow key. Raises set temperature when unit is in set mode.
-“Down” arrow key. Lowers set temperature when unit is in set mode
-“Start / Stop” key. Switches the unit between set and heating mode.
1