Labnet Labpette R Instruction Manual

1. Components and functions
(1) Plunger
Depressing the plunger dispenses the set volume of liquid
(2) Volume adjustment dial
Select position 1 to 5 to set desired volume (see table 1)
(3) Fill lever
Fills tip with liquid when pushed upwards
(4) Locking lever
(5) Reservoir tip
Reservoir for holding liquid to be dispensed. For best results use Labnet Combi-Syringes.
Labpette R
Instruction Manual
2. Selecting the volume
Locate the volume to be dispensed in table 1
Set the volume adjustment dial to the position
indicated at the top of the column.
Choose the appropriate reservoir tip from the left hand column of the chart.
The maximum number of dispensing steps for the selected tip and volume is located at the bottom of the column.
3. Inserting reservoir tip
Push the fill lever down until it stops.
Push the locking lever upwards.
Insert a reservoir tip with slight pressure to lock it in
place.
Push the fill lever down slightly and press down on the locking lever to secure the tip in place.
Use the volume adjustment dial to set the required volume.
4. Filling the reservoir tip
Immerse the end of the reservoir tip into the liquid to be drawn up.
Slowly raise the fill lever upwards to draw liquid into the tip.
Dispense the first step with the plunger.
Table 1 Dial position 12345
using 0.05 ml tip 1 µl2 µl3 µl4 µl5 µl using 0.50 ml tip 10 µl 20 µl 30 µl40 µl50 µl using 1.25 ml tip 25 µl 50 µl 75 µl 100 µl 125 µl using 2.50 ml tip 50 µl 100 µl 150 µl 200 µl 250 µl using 5.00 ml tip 100 µl 200 µl 300 µl 400 µl 500 µl using 12.50 ml tip 250 µl 500 µl 750 µl 1000 µl 1250 µl using 25.00 ml tip 500 µl 1000 µl 1500 µl 2000 µl 2500 µl using 50.00 ml tip 1000 µl 2000 µl 3000 µl 4000 µl 5000 µl Steps 48 23 15 1 1 8
Please note :
25 ml und 50 ml tips require the use of an adapter (adapters are supplied free with each Labpette R repeating pipettors).
Attach the adapter by placing it on the tip
as shown in the illustration and turning clockwise. After the tip has been used, the adapter may be removed and reused.
0,05 ml tips should be used with a 10 pipette tip on the liquid end.
The fill lever must be raised slowly to prevent negative pressure and air
bubbles in the liquid. A small air bubble in the upper part of the reservoir will not affect the accuracy as the last step is not carried out.
µl
The reservoir tip does not need to be filled completely.
The 0.5 ml, l, 1.25 ml and 2.5 ml tips may be used with a 100
affixed to the end. This facilitates filling from vessels with narrow necks and provides better accuracy with small volumes.
µl pipette tip
5. Dispensing
Check that the volume setting is correct.
Wipe the end of the tip with a lint free wipe or other
appropriate material.
Place the tip at an angle to the wall of the vessel. Press the plunger down to the stop to dispense.
After using the 25 ml and 50 ml tips, the adapter can be removed by turning it counter-clockwise and reused with another tip.
Avoid touching the end of the tip
when removing it from the pipette. When using viscous solutions, the fit of the tip may loosen after several pipettings.
6. Removing the tip
Empty the tip completely.
Push the locking lever upwards.
Remove the tip from the dispenser.
Do not attempt to disassemble the Labpette R repeating pipettor. Doing so will void the warranty.
7. Maintenance
Labpette R repeating pipettors requires no maintenance. Potential clogs are best removed with warm water or isopropanol.
8. Troubleshooting
Problem Possible cause Remedy
Tips will not fit in pipette Fill lever has not been pushed down completely Push fill lever down to the stop
Locking lever cannot be pushed down Tip has not been inserted Properly Remove the tip and push tip piston in. Be sure the filling lever is in the down position, replace tip and press locking lever down again
Fill lever cannot be moved upwards Locking lever has not been pushed Push the locking lever down completely down to the stop
Dispensed volume is not correct Volume adjustment wheel is not locked in place Lock adjustment wheel in place
Table 2
9. Technical data
Weight : Approx. 100 grams Accuracy and precision : see Table 2 Liquid : bidistilled water Reference temperature : 20° - 25° C
± 0.5° C constant
Measurement : according to
Subject to technical modifications Made in Germany
DIN 12650
Dial position 1 2 3 4 5 Inpre- Inaccur­Steps 48 23 15 11 8 cision (%) acy(%)
using 0.05 ml tip 1 using 0.50 ml tip 10 using 1.25 ml tip 25 using 2.50 ml tip 50 using 5.00 ml tip 100 using 12.50 ml tip 250 using 25.00 ml tip 500 using 50.00 ml tip 1000
10.Ordering information
Labnet Combi-Syringe tips may be ordered from your Labnet dealer. All sizes are supplied in stackable dispenser boxes and are available in sterile or non-sterile packaging.
11.Warranty
Labpette R repeating pipettors are guaranteed against defects in materials and workmanship for a period of 6 months from the data of purchase. This warranty is valid only if the dispenser has been used in compliance with these instructions. If a failure cannot be remedied using the troubleshooting guide above, contact Labnet for further assistance.
µl2 µl3 µl4 µl5 µl< 5 < ± 2,5
µl 20 µl 30 µl 40 µl 50 µl < 0,7 < ± 0,8 µl 50 µl 75 µl 100 µl 125 µl < 0,5 < ± 0,8 µl 100 µl 150 µl 200 µl 250 µl < 0,4 < ± 0,8
µl 200 µl 300 µl 400 µl 500 µl < 0,3 < ± 0,4 µl 500 µl 750 µl 1000 µl 1250 µl < 0,3 < ± 0,3 µl 1000 µl 1500 µl 2000 µl 2500 µl < 0,3 < ± 0,2
µl 2000 µl 3000 µl 4000 µl 5000 µl < 0,2 < ± 0,2