Labconco Type C1 User Manual

1
Original instructions
Type C1 Biosafety Cabinet Models 30441 30448 30461 30468
Axiom Technical Manual and Specifications
2
Labconco Training Programs:
In order to assist technicians gain a better understanding of our products, Labconco Corporation offers Training Seminars at our Kansas City Facility on a regular basis. For further details about these training programs, contact Labconco's Product Service Department at 800-821-5525, or (816) 333-8811 from 7:30 a.m. to 5:30 p.m. C.S.T.
Additional Copies: Additional copies of this document can be downloaded from Labconco’s website at www.labconco.com. Electronic copies are available to members of the service community free of charge. Please request electronic copies by contacting our Product Service Department.
Labconco Purifier Axiom Technical Manual and Specifications Printed in the U.S.A. ©2014 by Labconco Corporation, Kansas City, MO.
All Rights reserved. No part of this document may be translated or reproduced in any form without written permission from Labconco Corporation, 8811 Prospect, Kansas City, Missouri, 64132, U.S.A.
While the advice and information offered in this document are believed to be true and accurate at the date of going to press, neither the author, nor Labconco Corporation can accept any legal responsibly for any errors or omissions that may be made.
P/N 3849911
Rev. G, ECO J633
Product Service 1-800-522-7658
3
Table of Contents
Introduction 7 Contacting Labconco 7
Section One – Background Information
QuickStarts 8 What’s the password for the Axiom? 9 How do I calibrate blower speed, filter life & airflow sensor? 10-11 How do I see the current settings? 12 How do I change the configuration? 13-14 How do I run a diagnostic on the: Sash position switches? 15 Keypad, relays and contact? 16 Inlet Relief Valve? 17 Airflow Sensor? 18 How do I set up automatic operation of the blower & lights? 19
Axiom Filter Life Gauge Operation and Resetting for Local Conditions 20 QuickCharts Axiom Type C1 Purifiers 21-23 Identifying Axiom Models 24
Serial Number Tags 25
Model Numbering System 26 Axiom Data Plate 27 Final Inspection and Test Report 28 Theory of Operation Background 29
Airflow Patterns and Features 29 Chem-ZoneTM Directly Exhausted Work Zone 30 Inlet Relief Valve 31 Exhaust System 31 Exhaust System Alarm 32 Airflow Patterns of Axiom 33 ECM Motor 34-35 Electronics 35-36
Section Two – Installation Requirements
Prerequisites 37 Space Requirements 37 Clearance 37 Location Requirements 38 If You Intend to Connect the Axiom to an Exhaust System 39-41 Electrical Requirements 42 Service Line Requirements 43
Product Service 1-800-522-7658
4
Section Three – Installation 45 Unpacking the Biosafety Cabinet 46 Preparing the Biosafety Cabinet for operation 46 Moving and Lifting the Cabinet 47 Installing the Biosafety Cabinet on an Existing Work Surface 47 Installing the Cabinet on a Labconco Base Stand 48 Telescoping Base Stands 48 Manual or Electric Lift Base Stands 48
SoLo Electric Hydraulic Lift Base Stands 48 Optional Exhaust System Connection Requirements 49 Configuring for an Exhaust System Connection 50-53
Disconnecting the Axiom from an Exhaust System to Operate in
Recirculation 54-56
Connecting the Biosafety Cabinet to Utility Service Lines 57 Optional Vacu-Pass Cord and Cable Portal 58 Drain Valve Installation (optional) 59
Section Four – Initial Operation of the Axiom System Reset Switch 61 Information Center 62 Alarm Screens 63 Operating the Sliding Sash 64 Starting the Biosafety Cabinet 64 The Axiom Touchpad 65 Navigating the Axiom Menu Screens 66 Navigating the MyLogic Menu Screens 67-69 Setting the Clock 67 Configuring the Axiom 68-69 Navigating the Settings Menu Screens 69-71 Display Options
Units of Measure 69
Startup Tone 70 Security Lock 70 RS-232 Output 70 UV Settings UV Lamp Hour meter 71 Reset UV Lamp Hour Meter 71 Change UV Lamp Life 71 Timer Operation Interval Timer Operation 72 Stopwatch Timer Operation 72 If An Airflow Alert Activates 72 Resetting the Airflow Alert System 72
Product Service 1-800-522-7658
5
Section Five – Certification Procedures for the Axiom
Accessing the Certifier’s Menu 73 Adjusting the Inflow, Downflow and Filter Life Gauges 73-74 Axiom Filter Life Gauge Operation and Resetting for Local Conditions 75 HEPA Filter Leak Test Information Supply HEPA Filter Leak Test 76 If the Axiom is contaminated 77 Supply HEPA Filter Leak Test Specifications 77 If the Axiom is uncontaminated 77-79 Exhaust HEPA Filter Leak Test 80-81 Exhaust HEPA Filter Leak Test Specifications 80 Downflow Velocities 81-82
Downflow Test Specifications 81
Inflow Velocity Calculation Primary Inflow Method 83 Secondary Inflow Method 84-87 Concurrent Balance Values 88 Work Area Air Cleanliness Test (optional) 89 Fluorescent Light Level (optional) 90 Noise Level (optional) 90 Vibration (optional) 91 UV intensity (optional) 91 Ground Fault Circuit Interrupter 92
Section Six – Axiom Service Operations
Recommended Tools for Service 93
External Dress Components Front Dress Panel Removal and Installation 94 Side Panel Removal 95 Side Panel Installation 95
Corner Post Lower Trim 95-96
Sash Operation/Adjustment 97 Sash Position Switches Sash Closure Position Switch 98 Sash Restricted Opening Switch 98
8” and 10” Sash Operating Position Switches 99
Sash Cable Retainer 100
HEPA Filter Replacement 100-03
Supply Motor/Blower Replacement 103-05
Exhaust Motor/Blower Replacement 105-06
Center Work Surface Removal 107
Wing Work Surfaces Removal 107-08
Grille Removal 109 Diffuser Removal 110-11 Preparing the Logic for Gaseous Sterilization 112
Section Seven –Electrical System Service Operations
Fluorescent Lamp Lamp Identification 113 Changing the Fluorescent Lamp 113
Lamp Socket Replacement 114
Optional Ultraviolet Lamp
Product Service 1-800-522-7658
6
Lamp Identification 114 UV Lamp Specifications 114 Changing the UV Lamp 115
Lamp Socket Replacement 115
Electronics Module Access 116 Electronic Module Component Identification 116
Circuit Breaker Resetting a circuit breaker 117 Circuit breaker specifications 117-18 Testing a circuit breaker 118 Replacing a circuit breaker 118 Power Supply Board Service Operations Fuse Testing & Replacement 119 Display Board Service Operations 120
Touchpad Connections 120
Diagnostic Operations 121 Motor Service Operations Motor Testing 121-23 Motor Output Signal Testing 124
Wiring Diagrams 125-26
Section Eight – Purifier Axiom Accessories
Air Flow Sensor Kit 127-30
Service Fixture Kit 131-32 U.V. Lamp Kit 133-35 I.V. Bar Kit 136 Vacu-Pass Cord & Cable Portal 137
RS-232 Connection Kits 138-40
Using the RS-232 Receptacle 141
Connection for Computer Interface 141
Optional Axiom RS-232 Data Output 142-43
Axiom Relay Contact Outputs 144-48
Section Nine - Miscellaneous
Troubleshooting 149-52 Purifier Logic Replacement Parts 153
Dimensions 154-55
Product Service 1-800-522-7658
7
INTRODUCTION
This text is designed for you, the Biohazard Cabinet Certifier. Based on your suggestions, comments, and ideas, Labconco has compiled this information for you to use in the certification or servicing of our Purifier Axiom Biosafety Cabinets.
When you use this document in your office or in the field, we would appreciate your input so that we can make further improvements to future editions.
As always, we at Labconco want to assist you in having a better understanding of our products and their operation; if you have any questions, or need additional information, please contact us.
Thanks again for all your support in the past, and in the future.
Labconco
CONTACTING LABCONCO
Labconco’s Address is:
LABCONCO Corporation 8811 Prospect Kansas City, MO 64132
Our hours of operation are from 07:30 - 17:30 CST, Monday through Friday, except national holidays
We can be reached at the following numbers:
(800) 821-5525 (816) 333-8811
Our Fax numbers are:
(816) 444-5343 (816) 363-0130
Should you require Technical Assistance, Service Parts, or have general questions regarding the Purifiers, please direct them to our Product Service Department, so that your call can be properly routed and answered.
Product Service 1-800-522-7658
8
SECTION ONE –
Background Information
Quickstart
The Axiom is a significant step forward in Biosafety Cabinet technology. While many internal components are the same or similar to our Logic+-platform products, its operation and certification are different. The show you how to answer the questions asked in each case. Keystrokes are shown in
bold blue italics
, while screen selections are shown in
Quickstarts
that follow will help you
green italics
.
Product Service 1-800-522-7658
9
Quickstart
What’s the password for the Axiom?
Like the Logic, the Axiom models password protects the user from entering the
Calibration
the airflow sensor readings. New to the Axiom is a
Configuration
connection status, whether its sash height is 8” or 10” and airflow sensor options­there is more on this in the Quickstart “How do I change the BSC configuration?”
The password for all protected screens is:
1. Light
2. UV Light
3. Timer
4. Timer
5. OK/Mute
menu to prevent them from adjusting the blower speed, filter gauge or
Service
; entering this option allows you to configure the Axiom’s exhaust
menu selection entitled
Product Service 1-800-522-7658
10
Quickstart
MAIN MENU
1. MyLogic
TM
2. Settings
3. Service

or to choose,
OK to select, MENU to escape
or to choose,
OK to select, MENU to r eturn
Current Settings
Restore Default
Diagnostics
Configuration
Calibration

Service
ATTENTION!
The following screens are used
to alter the operation of the
Logic. They should only be used
by a qualified certifier as part
of the certification process.
Calibration
OK to proceed, MENU to return
PASSWORD:
Calibration
The ECM setting is: 80 The current RPM is: 1050 The average RPM is: 1020
Calibration
Supply Fan Speed
or to change motor speed,
OK to select, MENU to escape

How do I recalibrate the
blower speed filter life gauge airflow sensor reading
of the Axiom?
1. Press the
2. Press the button until the turn yellow).
3. Press
4. Press the button until the will turn white).
5. Press
6. You will see a alter the BSC’s settings.
7. Press
8. When requested for the password press
Timer, Timer
9. If the password is properly entered, the first certification screen will display the supply blower motor’s PWM setting. Press the or button to increase or decrease the blower speed as needed. When the speed is set properly, press the button to lock the new blower speed setting, and continue to the filter life screen.
Menu
OK/Mute
OK/Mute
warning screen
OK/Mute
then
button.
Services
option is highlighted (it will
to enter the Services menu screen.
Calibration
option is highlighted (it
to start the calibration procedure.
alerting you that you are about to
to input the password.
Light, UV light
OK/Mute
.
OK/Mute
,
Product Service 1-800-522-7658
11
Quickstart
The Inflow displayed value is: 102 FPM
Calibration
Inflow Display
or to change display value, UV Light to reset to 105 FPM, OK to select, MENU to finish

Calibration
Supply Filter Gauge
or to change filter gauge,
OK to select, MENU to escape

Reset Filter Gauge to 100% Leave Gauge unchanged
Set Gauge to new setting
The ECM setting is: 80 The current RPM is: 1050 The average RPM is: 1020
Calibration
Exhaust Fan Speed
or to change motor speed, OK to select, MENU to escape

Calibration
Exhaust Filter Gauge
or to change filter gauge,
OK to select, MENU to escape

Reset Filter Gauge to 100% Leave Gauge unchanged
Set Gauge to new setting
How do I recalibrate the
blower speed filter life gauge airflow sensor reading
of the Axiom? - Continued
10. If this is an initial certification or a recertification after supply HEPA filter replacement, select this is an annual recertification, select If you want to set the filter gauge at a preset value, select
Gauge to new setting
. Then press
11. The next certification screen will display the exhaust blower motor’s PWM setting. Press the or button to increase or decrease the blower speed as needed. When the speed is set properly, press the
OK/Mute
speed setting, and continue to the filter life screen.
12. If this is an initial certification or a recertification after exhaust HEPA filter replacement, select this is an annual recertification, select If you want to set the filter gauge at a preset value, select
Gauge to new setting
. Then press
13. If the unit is has no airflow sensor, after the filter gauge option is selected and you press down, and then restart with the new settings. If the Unit has an airflow sensor, you will then be asked to calibrate the inflow sensor to match the value you obtained when you measured the inflow. After the inflow is calibrated, and you press the Axiom will power down, and then restart with the new settings.
Reset Filter Gauge to 100%
Leave Gauge unchanged
OK/Mute
.
button to lock the new blower
Reset Filter Gauge to 100%
Leave Gauge unchanged
OK/Mute
OK/Mute
, the Axiom will power
.
OK/Mute
. If
.
Set
. If
.
Set
,
Product Service 1-800-522-7658
12
Quickstart
MAIN MENU
1. MyLogic
TM
2. Settings
3. Service

or to choose,
OK to select, MENU to escape
or to choose,
OK to select, MENU to return
Current Settings
Restore Default
Diagnostics Calibration Configuration

Service
Current Settings Supply System
or to change, OK to return
The ECM setting is: 65 The current RPM is: 0 The average RPM is: 0 The new filter RPM is: 1450 C1, with afs, with UV

10484 -
Current Settings Exhaust System
or to change, OK to return
The ECM setting is: 65 The current RPM is: 0 The average RPM is: 0 The new filter RPM is: 1450 C1, with afs, with UV

10484 -
How do I see the Current Settings of the Axiom?
1. Press the
2. Press the button until the turn yellow). Press
screen.
3. Press the button until the highlighted (it will turn white). Press configuration.
4. The first screen shows the current settings of the supply blower system.
5. By pressing the or button, you can see the current settings of the exhaust blower system. If you are recertifying the unit, you may want to record the data on this screen. The firmware revision is shown in the bottom right corner as a -, or as a letter.
Menu
button.
OK/Mute
Services
option is highlighted (it will
to enter the Services menu
Current Settings
OK/Mute
option is
to see the
Product Service 1-800-522-7658
13
Quickstart
MAIN MENU
1. MyLogic
TM
2. Settings
3. Service

or to choose,
OK to select, MENU to escape
or to choose,
OK to select, MENU to return
Current Settings
Restore Default
Diagnostics Calibration
Configuration

Service
Configuration
PASSWORD:
ATTENTION!
The following screens are used
to alter the operation of the
Logic. They should only be used
by a qualified certifier as part
of the certification process.
Configuration
OK to proceed, MENU to return
How do I change the Configuration of the Axiom?
1. Press the
2. Press the button until the turn yellow).
3. Press
4. Press the button until the (it will turn white).
5. Press
6. You will see a alter the BSC’s settings.
7. Press
8. When requested for the password press
Timer, Timer
Menu
OK/Mute
OK/Mute
warning screen
OK/Mute
then
button.
Services
to enter the Services menu screen.
Configuration
to alter the configuration.
alerting you that you are about to
to input the password.
OK/Mute
.
option is highlighted (it will
option is highlighted
Light, UV light
,
Product Service 1-800-522-7658
14
This cabinet is not configured for a UV light
If you want to install a UV light, contact Labconco Product Service at 800-821-5525
Attention
MENU to return
Configuration
Type C Configuration
This cabinet has:
an
or to choose,
OK t o select , MENU to escape
no exhaust connect ion
exhaust connection

Configuration
Type C Configuration
If there is an exhaust system failure, Active protection should last:
or to choose,
OK t o select , MENU to escape
0 Seconds

The sash height is:
or to choose,
OK to select, MENU to escape
8 inches
10 inches NOTE: If you change the sash height, you must recertify airflows.

Sash Height Settings
This Logic has:
or to choose,
OK to select, MENU to escape
a UV light
NO UV light

UV Settings
Quickstart
How do I change the Configuration of the Axiom? – continued
9. The first screen allows you to decide if the cabinet is connected to an exhaust system or not. When the unit is configured for no exhaust connection, the Axiom will ignore the inlet relief valve sensor’s status. Selecting connected to an exhaust system activates the sensor; if the valve is open, the unit operates normally, if closed, an exhaust alarm is activated. Use the or buttons to select the exhaust configuration, and then press
OK/Mute.
10. If the unit is configured to be connected to an exhaust system, then the next screen allows you to program the amount of time interval of Active Protection. Active Protection will allow the Axiom blowers to continue to operate after the detection of an exhaust system failure. This will maintain a safe inflow to allow the user a programmed time interval to secure any hazardous materials in the cabinet before the BSC shuts off.
11. This screen configures the sash height. After your selection is chosen, and you press OK/Mute.
12. This screen configures the unit for a UV light or not. After your selection is chosen, and you press OK/Mute, the unit will shut off and then reset in the new configuration.
13. If the unit is configured to NOT have a UV light, and the user presses the UV Light button, they will get this message.
Product Service 1-800-522-7658
15
Quickstart
MAIN MENU
1. MyLogic
TM
2. Settings
3. Service

or to choose,
OK to select, MENU to escape
or to choose,
OK to select, MENU to return
Current Settings
Restore Default
Calibration Configuration
Diagnostics

Service
Diagnostics
Test the alarm by closing the inlet valve.
Note the alarm status below:
Open
OK to return
How do I run a diagnostic on the sash position switches?
1. Press the
2. Press the button until the turn yellow).
3. Press
4. Press the button until the will turn white).
5. Press
6. Press the button until the Diagnostic highlighted (it will turn white). Press
7. Lower the sash, and note the status line – it should properly report the sash as open too high, open, nearly closed or closed when the sash is in the corresponding position.* Press
OK/Mute
Menu
button.
OK/Mute
OK/Mute
to enter the Services menu screen.
to enter the Diagnostic submenu.
to exit the diagnostic.
Services
option is highlighted (it will
Diagnostic
OK/Mute
option is highlighted (it
Sash sensors
.
option is
Product Service 1-800-522-7658
16
Quickstart
MAIN MENU
1. MyLogic
TM
2. Settings
3. Service

or to choose,
OK to select, MENU to escape
or to choose,
OK to select, MENU to return
Current Settings
Restore Default
Calibration Configuration
Diagnostics

Service
How do I run a diagnostic on the keypad, relays and contacts?
1. Press the
2. Press the button until the turn yellow).
3. Press
4. Press the button until the will turn white).
5. Press
6. Press the button until the Diagnostic (it will turn white). Press
7. Press any button on the keypad (except display.
Menu
OK/Mute
OK/Mute
button.
Services
option is highlighted (it will
to enter the Services menu screen.
Diagnostic
option is highlighted (it
to enter the Diagnostic submenu.
Keypad-Relays
OK/Mute
.
Menu
option is highlighted
), and note the
Product Service 1-800-522-7658
17
MAIN MENU
1. MyLogic
TM
2. Settings
3. Service

or to choose,
OK to select, MENU to escape
or to choose,
OK to select, MENU to return
Current Settings
Restore Default
Calibration Configuration
Diagnostics

Service
Diagnostics
or t o choose,
OK to select, MENU to escape
Sash sensors Keypad - Relays Inlet Relief Valve Airflow Sensor

Diagnostics
Test the alarm by closing the inlet valve.
Note the alarm status below:
Open
OK to return
Quickstart
How do I run a diagnostic on the Inlet Relief Valve?
1. Press the
2. Press the button until the turn yellow).
3. Press
4. Press the button until the will turn white).
5. Press
6. Press the button until the Diagnostic is highlighted (it will turn white). Press
7. The display shows the status of the Inlet Relief Valve. During normal operation, it will be open, and during an exhaust alarm, it will be closed. NOTE: The inlet relief valve function is OPPOSITE OF THE MAGNETIC SWITCH FUNCTION! – WHEN THE INLET VALVE IS OPEN, THE MAGNETIC SWITCH IS CLOSED.
Menu
OK/Mute
OK/Mute
button.
Services
option is highlighted (it will
to enter the Services menu screen.
Diagnostic
option is highlighted (it
to enter the Diagnostic submenu.
Inlet Relief Valve
OK/Mute
.
Product Service 1-800-522-7658
option
18
MAIN MENU
1. MyLogic
TM
2. Settings
3. Service

or to choose,
OK to select, MENU to escape
or to choose,
OK to select, MENU to return
Current Settings
Restore Default
Calibration Configuration
Diagnostics

Service
Quickstart
How do I run a diagnostic on the Airflow Sensor?
1. Press the
2. Press the button until the turn yellow).
3. Press
4. Press the button until the will turn white).
5. Press
6. Press the button until the Diagnostic highlighted (it will turn white). Press
7. The display shows the status of the airflow sensor. Its current output, alarm setting and zero point in Millivolts, along with the corrected Inflow velocity currently being displayed.
Quickstart
Menu
OK/Mute
OK/Mute
button.
Services
option is highlighted (it will
to enter the Services menu screen.
Diagnostic
option is highlighted (it
to enter the Diagnostic submenu.
Airflow Sensor
OK/Mute
.
Product Service 1-800-522-7658
option is
19
How do I set up automatic operation of the blower and lights?
1. Press the
2. Press the button until the turn blue).
3. Press
Menu
OK/Mute
button.
MyLogic
to enter the MyLogic menu screen.
option is highlighted it will
4. Press the button until the option option is highlighted (it will turn white).
5. Press the button to select either Smart-Start or manual operation of the cabinet blower. (In Smart-Start, the cabinet blower will automatically start every time the sash is opened.
6. Press the button to select either Smart-Start or manual operation of the fluorescent lights.
7. Press the button to select either Night-Smart or manual operation of the cabinet blower. (In Night-Smart, the cabinet blower will automatically operate at idle every time the sash is closed.
8. Press the button to select either Night-Smart or manual operation of the UV Light.*
9. Press the button to select the time interval the UV light will operate for until it automatically shuts off.*
* - These screens will only appear if the unit is configured for a UV light.
configure my Logic for use
Axiom Filter Life Gauge Operation and Resetting for Local Conditions
Product Service 1-800-522-7658
20
The filter life gauge calculates the remaining filter life by measuring and comparing the current motor speed to the initial speed noted at the factory during assembly, and the calculated maximum motor speed, using the formula:
100-(100x (current speed-original speed) ) = % filter life left max. speed-original speed
So
anything
Blocking the grille. Raising the air temperature. Lowering the barometric pressure.
will increase the motor’s speed, causing the Filter Life Gauge to decrease. Electronic dampening of the gauge will limit the change in readings to approximately 1%/minute.
Axiom Biosafety Cabinets being installed at altitudes or operating at temperatures significantly different from those at the time of manufacture (900 +/- 500 ft above sea level, 75o +/- 5o F) conditions. Failure to do so will result in the users noticing a steady drop in Percent (%) Filter Life remaining as the unit operates. Below are the instructions on how to correct the Filter Life Gauge for local conditions.
that changes the motor’s current speed, like:
must
have their filter life gauge recalibrated for local
1. Ensure to correct all measurement equipment for local conditions (temperature, absolute pressure). Anemometers may not have this ability and might have to be done manually; whereas, a ShortRidgeTM instrument has a setting to correct for local conditions.
2. As the elevation or the ambient temperature rises, the motor power factor will have to be reduced to compensate. For an initial certification, the motor speed should be reduced until it equals the Axiom’s “Initial Motor Speed” as reported on the unit’s test sheet, plus approximately 10-15 RPM per each additional 1000 feet of elevation.
3. The Axiom MUST be recertified using instruments corrected for local conditions.
4. When the unit is recertified, and operating properly, the certifier MUST RESET THE FILTER LIFE GAUGE, INDICATING A NEW FILTER HAS BEEN INSTALLED.
5. Only by performing Step #4 AFTER THE UNIT IS READJUSTED AND RECERTIFIED FOR ITS LOCAL CONDITION, will the filter life gauge register 100% (+/- 2%) during operation. As the Filters then load, the filter life gauge will then begin to drop as appropriate.
Product Service 1-800-522-7658
21
QuickChart – Axiom Type C1 BSCs – Imperial Units
Model
30441
30448
30461
30468
Type
C1
C1
C1
C1
Cabinet Size (in feet)
4 4 6
6
Sash Opening (inches)
10 8 10
8
Starting Serial #
1503_
1503_
1503_
1503_
Downflow Data
Nominal Avg. Downflow (FPM)
65+/-5
65+/-5
55+/-5
55+/-5
grid # of points (rows x columns)
24 (3x8)
24 (3x8)
36(3x12)
36(3x12)
Grid distance from back & sides
6.0
6.0
6.0
6.0
Distance between rows
5.55
5.75
5.55
5.75
Distance between columns
5.21
5.21
5.50
5.50
Inflow Data
Nominal Average Inflow (FPM)
105+/-5
105+/-5
105+/-5
105+/-5
Sash Open Area (Sq. Ft)
3.37
2.69
5.03
4.03
Nominal Avg. Exhaust Vol. (CFM)
354
283
529
423
Avg. Exhaust Vol. Range (CFM)
337-371
269-296
503-553
403-443
Secondary Inflow Data
Sash Opening Template
1
4+
4+
6+
6+
Sensor distance (inches)2
3.25
3.25 4 4
# of Test points
8 8 12
12
Test point location
* * *
*
Nominal Avg. Inflow Vel. (FPM)
354
283
289
231
Avg. Inflow Vel. Range (FPM)
312-344
249-274
252-277
202-222
Correction Factor (CF)
1.08
1.08 2 2
Avg. Inflow Volume (CFM)
337-371
269-296
503-553
403-443
Sash Open Area (Sq. Ft)
3.37
2.69
5.03
4.03
Inflow Velocity Range (FPM)
100-110
100-110
100-110
100-110
Supply HEPA Data
Air Displacement (CFM)
530
530
660
660
Laskin Nozzles needed
1 1 1
1
Theoretical aerosol conc. (ug/l)3
25
25
20
20
Width x Depth x Height (in.)4
48x18x3.06
48x18x3.06
72x18x3.06
72x18x3.06
Performance (CFM)
445
445
665
665
Performance (Pressure in. H2O)
<.45”
<.45”
<.42”
<.42”
Labconco P/N
3838401
3838401
3838403
3838403
Exhaust HEPA Data
Air Displacement (CFM)
354
283
529
423
Laskin Nozzles needed
1 1 1
1
Theoretical aerosol conc. (ug/l)3
38
48
26
32
Width x Depth x Height (in.)
4
26x18x5.88
26x18x5.88
48x18x5.88
48x18x5.88
Performance (CFM)
354
354
529
52
Performance (Pressure in. H2O)
<0.3
<0.3
<0.3
<0.3
Labconco P/N
3838501
3838501
3838503
3838503
Exhaust Volume Data
DIM Exhaust Volume (CFM)5
400
323
570
463
Traverse Exhaust Volume (CFM)6
480
387
684
556
Differential Pressure (in. H20)7
0.3
0.3
0.3
0.3
IRV Inflow Range (CFM)8
20-100
20-100
20-100
20-100
Nominal IRV Slot Velocity (FPM)
240-260
240-260
240-260
240-260
Nominal IRV Slot Volume (CFM)
50
50
50
50
Supply Motor/Blower Data
Motor HP
1/3
1/3
1/3
1/3
Exhaust Motor/Blower Data
Motor HP
1/3
1/3
1/3
1/3
Fluorescent Lamp Data
Fluorescent Lamps (2 each)
F32T8 TL741
F32T8 TL741
F40T8 SP41
F40T8 SP41
Fluor. Lamp Labconco Part #
9721900
9721900
9721903
9721903
Fluor. Ballast Labconco Part #
3838100
3838100
3838100
3838100
Product Service 1-800-522-7658
22
QuickChart – Axiom Type C1 BSCs – Metric Units
Model
30441
30448
30461
30468
Type
C1
C1
C1
C1
Cabinet Size (in m)
1.22
1.22
1.83
1.83
Sash Opening (mm)
254
203
254
203
Starting Serial #
1503_
1503_
1503_
1503_
Downflow Data
Nominal Avg. Downflow (FPM)
65+/-5
65+/-5
55+/-5
55+/-5
grid # of points (rows x columns)
24 (3x8)
24 (3x8)
36(3x12)
36(3x12)
Grid distance from back & sides (mm)
152
152
152
152
Distance between rows (mm)
141
146
141
146
Distance between columns (mm)
132
132
140
140
Inflow Data
Nominal Average Inflow (m/s)
.533+/-.03
.533+/-.03
.533+/-.03
.533+/-.03
Sash Open Area (Sq. m)
.313
.250
5.03
4.03
Nominal Avg. Exhaust Vol. (m3/s)
354
283
.468
.374
Avg. Exhaust Vol. Range (m3/s)
337-371
269-296
503-553
403-443
Secondary Inflow Data
Sash Opening Template
1
4+
4+
6+
6+
Sensor distance (mm)2
83
83
102
102
# of Test points
8 8 12
12
Test point location
* * *
*
Nominal Avg. Inflow Vel. (m/s)
1.80
1.44
1.47
1.17
Avg. Inflow Vel. Range (m/s)
1.71-1.88
1.37-1.50
1.40-1.53
1.12-1.23
Correction Factor (CF)
.0929
.0929
.1700
.1700
Average Inflow Volume Range
(AIV) (Avg. velocity x CF)
.159-.175
.127-.140
.238-.261
.190-.209
Sash Open Area (Sq. m)
.313
.250
.468
.374
Inflow Velocity Range
(AIV / Sash open area)
.533+/-.03
.533+/-.03
.533+/-.03
.533+/-.03
Supply HEPA Data
Air Displacement (m3/s)
.250
.250
.311
.311
Laskin Nozzles needed
1 1 1
1
Theoretical aerosol conc. (ug/l)3
25
25
20
20
Width x Depth x Height (in.)4
48x18x3.06
48x18x3.06
72x18x3.06
72x18x3.06
Performance (CFM)
445
445
665
665
Performance (Pressure in. H2O)
<.45”
<.45”
<.42”
<.42”
Labconco P/N
3838401
3838401
3838403
3838403
Exhaust HEPA Data
Air Displacement (m3/s)
.167
.134
.250
.200
Laskin Nozzles needed
1 1 1
1
Theoretical aerosol conc. (ug/l)3
38
48
26
32
Width x Depth x Height (in.)
4
26x18x5.88
26x18x5.88
48x18x5.88
48x18x5.88
Performance (CFM)
283
283
423
423
Performance (Pressure in. H2O)
<0.3
<0.3
<0.3
<0.3
Labconco P/N
3838501
3838501
3838503
3838503
Exhaust Volume Data
DIM Exhaust Volume (m3/s)5
.189
.152
.269
.219
Traverse Exhaust Volume (m,3/s)6
.227
.183
.323
.262
Differential Pressure (Pascal)7
75
75
75
75
IRV Inflow Range (m/s)
.010-.047
.010-.047
.010-.047
.010-.047
Nominal IRV Slot Velocity (m/s)
1.22-1.32
1.22-1.32
1.22-1.32
1.22-1.32
Nominal IRV Slot Volume (m3/s)
.235
.235
.235
.235
Supply Motor/Blower Data
Motor HP
1/3
1/3
1/3
1/3
Exhaust Motor/Blower Data
Motor HP
1/3
1/3
1/3
1/3
Fluorescent Lamp Data
Fluorescent Lamps (2 each)
F32T8 TL741
F32T8 TL741
F40T8 SP41
F40T8 SP41
Fluor. Lamp Labconco Part #
9721900
9721900
9721903
9721903
Fluor. Ballast Labconco Part #
3838100
3838100
3838100
3838100
Product Service 1-800-522-7658
23
QuickChart – Axiom/Cell Axiom Type C1 BSCs footnotes
Axiom Model Identification
The primary serial tag is on the lower right edge of the front dress panel. The secondary serial tag is located on the front of the electronics module on the top
right side of the cabinet.
The first two digits of the serial number are the year of production; the next two are
the month. The next 5 digits are the sequence of production, and the letter following the serial number is the revision level of the cabinet.
Downflow Test Specifications
All models are classified as uniform downflow. All tests performed as described in ANSI/NSF Standard 49:2012. UV Lamp, IV bar and all other accessories must be removed before measuring
downflow.
Inflow Test Specifications
All tests performed as described in ANSI/NSF Standard 49:2012.
Secondary Inflow Test Specifications
Must use Labconco holder P/N 3836405 to perform this test properly.
1. Use the appropriate template included in Certifier Kit# 3858403.
2. Measured from the bottom edge of the probe holder to the center of the thermal anemometer sensor element.
* - Locate the single row of holes at the front of the grille. Mark the 6th hole from the side wall and subsequent test points every 9 holes until the number of test points marked equals the width of the cabinet in feet (3-foot cabinet, mark the first 3 points). Repeat for the opposite side.
HEPA Filter Leak Test Specifications
3. Based on mineral oil.
4. Without gasket
Aerosol generator should be placed in the left rear corner of the work area, pointing
toward the rear grille.
For uncontaminated units, the upstream concentration can be sampled from the
tube located under the work surface.
Metric calculation of concentration is: # of Laskin nozzles @ 138K Pascal x 6.372/Volume of air in m3/sec. = Conc.
In ug/l
Exhaust Data
5. Total Exhaust Volume; The sum of the inflow and air drawn into the Inlet Relief
6. Measured as per ASHRAE methodology for measuring air volume in round ducts.
7. Measured at the exhaust transition sampling point, relative to atmosphere.
8. IRV = Inlet Relief Valve.
UV lamp Specifications
For all models, the UV lamp number is G30T8. The Labconco part number is
For all 115/230 VAC models, the ballast assembly is Labconco part number 3829901.
For all 100 VAC models, the ballast assembly is Labconco part number 3830600. The
Valve.
1271300.
The ballast is Robertson part number PSM2GPH18MVW.
ballast is Robertson part number RSO1GPH30100.
Product Service 1-800-522-7658
24
IDENTIFYING AXIOM MODELS
The Axiom platform Biosafety Cabinet can be identified readily by several external features, as shown in Figure 1-1:
Model number Stars with 304xxxxxx Serial number beginning with 1503xxxxx. Exhaust System Cover. Segmented Work Surfaces.
Figure 1-1
Exhaust System Cover
Axiom Logo
Segmented Work Surfaces
Product Service 1-800-522-7658
25
SERIAL NUMBER TAGS
The model/serial number tag is located on the bottom right edge of the dress panel. A second tag is located on the front of the electronics module on the top of the unit. This tag includes the unit’s operating voltage and amperage. The model and serial numbers of each unit are also recorded on the unit test report attached beneath the front dress panel. These items are shown graphically in the figures 1-3 to 1-5 below:
Figure 1-3
Primary Serial Tag
Figure 1-4
Secondary Serial Tag
Figure 1-5
Final Inspection and Test Report Form
Obviously, you can get confirmation of the Biosafety Cabinet's model by contacting Labconco's Product Service Department, and providing us with the serial number. From this number, we can tell the date of production and its model.
Product Service 1-800-522-7658
26
AXIOM MODEL NUMBERING SYSTEM
Axiom units’ model numbers define the configuration of the cabinet. The first five digits define the unit's configuration, while the last four are assigned based on the unit's options and electrical configuration. Thus, a model number of:
30448xxxx
Labconco Product Group 3=Laminar Flow Products
Indicates the Product Family 0= Logic 1= PuriCare 2-4= Cell Logic
Indicates the Cabinet Type 1= Type A1
2= Type A2 3= Type B2 4= Type C1
Indicates the Cabinet width
3= 3-foot Wide
4= 4-foot Wide
5= 5-foot Wide
6= 6-foot Wide
Indicates the sash height 1= 10” sash opening 2= 12” sash opening 8= 8” sash opening 9= 9” sash opening
Product Service 1-800-522-7658
27
Axiom Data Plate
The Axiom Data Plate contains much of the information required to certify the unit. Information on the plate includes nominal inflow and downflow velocity settings, a description of how to establish the downflow grid and the secondary inflow methodology are also included on the data plate, as shown below. The plate is located in the upper left corner of the dress panel.
Figure 1-6
Product Service 1-800-522-7658
28
Final Inspection and Test Report Form
A copy of the Final Inspection Test Report is attached to the blower plenum cover, beneath the front dress panel. The report contains initial setup information on inflow and downflow velocities, and the initial blower torque setting. A copy of the report is shown in Figure 1-7, below
Figure 1-7
Model/Serial Numbers
HEPA Filter Information
Test Results
Downflow Test Results
Product Service 1-800-522-7658
29
Theory of Operation
Background
The Axiom Type C1 BSC was the result of a customer request for a Type B (directly exhausted) cabinet that could operate at a constant exhaust system vacuum of only 0.25 inches of H2O. Realizing that this was impossible for existing Type B technologies, Labconco developed a new type of BSC; The proposed Type C1.
Figure 1-8
Airflow Patterns and Features
The overall airflow pattern in the Type C1 is similar to a Type B1, with three important differences:
1. In the C1, the directly exhausted portion of downflow air is taken from the center of the work area (The Chem-ZoneTM).
2. An inlet relief valve, unique to the Axiom, stabilizes the inflow as exhaust system airflows fluctuates, and closes during an exhaust system failure, signaling the Axiom to initiate a timed “Active Protection” of the user and their work.
3. A constant volume exhaust blower upstream of the Exhaust
HEPA maintains constant inflow, and allows the unit’s blowers
to continue to operate during an exhaust system failure (Active Protection), maintaining safe inflows and downflows while the user secures any hazardous materials in the work area.
Product Service 1-800-522-7658
30
Figure 1-8a
Chem-ZoneTM Directly Exhausted Work Zone
Unique to the Axiom is the Chem-Zone, a directly exhausted work zone. The central portion of the work surface is surrounded by grilles on the sides, front and back. Air entering these grilles is drawn to the Exhaust HEPA filter, and when the cabinet is connected to an exhaust system, out of the laboratory. This feature prevents the recirculation of volatile chemicals as seen in Type A cabinets, while exhausting much less air than Type B2 models. This is illustrated in Figure 1-8a.
Product Service 1-800-522-7658
Loading...
+ 125 hidden pages