Type C1 Biosafety Cabinet Models
30441 30448
30461 30468
Labconco Corporation
8811 Prospect Avenue
Kansas City, MO 64132-2696
800-821-5525, 816-333-8811
FAX 816-363-0130
E-MAIL labconco@labconco.com
HOME PAGE www.labconco.com
Axiom Technical Manual and Specifications
2
Labconco Training Programs:
In order to assist technicians gain a better understanding of our products,
Labconco Corporation offers Training Seminars at our Kansas City Facility on
a regular basis. For further details about these training programs, contact
Labconco's Product Service Department at 800-821-5525, or (816) 333-8811
from 7:30 a.m. to 5:30 p.m. C.S.T.
Additional Copies:
Additional copies of this document can be downloaded from Labconco’s
website at www.labconco.com. Electronic copies are available to members of
the service community free of charge. Please request electronic copies by
contacting our Product Service Department.
All Rights reserved. No part of this document may be translated or
reproduced in any form without written permission from Labconco
Corporation, 8811 Prospect, Kansas City, Missouri, 64132, U.S.A.
While the advice and information offered in this document are believed to be
true and accurate at the date of going to press, neither the author, nor
Labconco Corporation can accept any legal responsibly for any errors or
omissions that may be made.
P/N 3849911
Rev. G, ECO J633
Product Service 1-800-522-7658
3
Table of Contents
Introduction 7
Contacting Labconco 7
Section One – Background Information
QuickStarts 8
What’s the password for the Axiom? 9
How do I calibrate blower speed, filter life & airflow sensor? 10-11
How do I see the current settings? 12
How do I change the configuration? 13-14
How do I run a diagnostic on the: Sash position switches? 15
Keypad, relays and contact? 16
Inlet Relief Valve? 17
Airflow Sensor? 18
How do I set up automatic operation of the blower & lights? 19
Axiom Filter Life Gauge Operation and Resetting for Local Conditions 20
QuickCharts
Axiom Type C1 Purifiers 21-23
Identifying Axiom Models 24
Serial Number Tags 25
Model Numbering System 26
Axiom Data Plate 27
Final Inspection and Test Report 28
Theory of Operation Background 29
Airflow Patterns and Features 29
Chem-ZoneTM Directly Exhausted Work Zone 30
Inlet Relief Valve 31
Exhaust System 31
Exhaust System Alarm 32
Airflow Patterns of Axiom 33
ECM Motor 34-35
Electronics 35-36
Section Two – Installation Requirements
Prerequisites 37
Space Requirements 37
Clearance 37
Location Requirements 38
If You Intend to Connect the Axiom to an Exhaust System 39-41
Electrical Requirements 42
Service Line Requirements 43
Product Service 1-800-522-7658
4
Section Three – Installation 45
Unpacking the Biosafety Cabinet 46
Preparing the Biosafety Cabinet for operation 46
Moving and Lifting the Cabinet 47
Installing the Biosafety Cabinet on an Existing Work Surface 47
Installing the Cabinet on a Labconco Base Stand 48
Telescoping Base Stands 48
Manual or Electric Lift Base Stands 48
SoLo Electric Hydraulic Lift Base Stands 48
Optional Exhaust System Connection Requirements 49
Configuring for an Exhaust System Connection 50-53
Disconnecting the Axiom from an Exhaust System to Operate in
Recirculation 54-56
Connecting the Biosafety Cabinet to Utility Service Lines 57
Optional Vacu-Pass Cord and Cable Portal 58
Drain Valve Installation (optional) 59
Section Four – Initial Operation of the Axiom
System Reset Switch 61
Information Center 62
Alarm Screens 63
Operating the Sliding Sash 64
Starting the Biosafety Cabinet 64
The Axiom Touchpad 65
Navigating the Axiom Menu Screens 66
Navigating the MyLogic Menu Screens 67-69
Setting the Clock 67
Configuring the Axiom 68-69
Navigating the Settings Menu Screens 69-71
Display Options
Units of Measure 69
Startup Tone 70
Security Lock 70
RS-232 Output 70
UV Settings
UV Lamp Hour meter 71
Reset UV Lamp Hour Meter 71
Change UV Lamp Life 71
Timer Operation Interval Timer Operation 72
Stopwatch Timer Operation 72
If An Airflow Alert Activates 72
Resetting the Airflow Alert System 72
Product Service 1-800-522-7658
5
Section Five – Certification Procedures for the Axiom
Accessing the Certifier’s Menu73
Adjusting the Inflow, Downflow and Filter Life Gauges 73-74
Axiom Filter Life Gauge Operation and Resetting for Local Conditions 75
HEPA Filter Leak Test Information Supply HEPA Filter Leak Test 76
If the Axiom is contaminated 77
Supply HEPA Filter Leak Test Specifications 77
If the Axiom is uncontaminated 77-79
Exhaust HEPA Filter Leak Test 80-81
Exhaust HEPA Filter Leak Test Specifications 80
Downflow Velocities 81-82
Circuit Breaker Resetting a circuit breaker 117
Circuit breaker specifications 117-18
Testing a circuit breaker 118
Replacing a circuit breaker 118
Power Supply Board Service Operations Fuse Testing & Replacement 119
Display Board Service Operations 120
Touchpad Connections 120
Diagnostic Operations 121
Motor Service Operations Motor Testing 121-23
Motor Output Signal Testing 124
Wiring Diagrams 125-26
Section Eight – Purifier Axiom Accessories
Air Flow Sensor Kit 127-30
Service Fixture Kit 131-32
U.V. Lamp Kit 133-35
I.V. Bar Kit 136
Vacu-Pass Cord & Cable Portal 137
RS-232 Connection Kits 138-40
Using the RS-232 Receptacle 141
Connection for Computer Interface 141
Optional Axiom RS-232 Data Output 142-43
Axiom Relay Contact Outputs 144-48
Section Nine - Miscellaneous
Troubleshooting 149-52
Purifier Logic Replacement Parts 153
Dimensions 154-55
Product Service 1-800-522-7658
7
INTRODUCTION
This text is designed for you, the Biohazard Cabinet Certifier. Based on
your suggestions, comments, and ideas, Labconco has compiled this
information for you to use in the certification or servicing of our
Purifier Axiom Biosafety Cabinets.
When you use this document in your office or in the field, we would
appreciate your input so that we can make further improvements to
future editions.
As always, we at Labconco want to assist you in having a better
understanding of our products and their operation; if you have any
questions, or need additional information, please contact us.
Thanks again for all your support in the past, and in the future.
Labconco
CONTACTING LABCONCO
Labconco’s Address is:
LABCONCO Corporation
8811 Prospect
Kansas City, MO 64132
Our hours of operation are from 07:30 - 17:30 CST, Monday through Friday,
except national holidays
We can be reached at the following numbers:
(800) 821-5525
(816) 333-8811
Our Fax numbers are:
(816) 444-5343
(816) 363-0130
Should you require Technical Assistance, Service Parts, or have general
questions regarding the Purifiers, please direct them to our Product Service
Department, so that your call can be properly routed and answered.
Product Service 1-800-522-7658
8
SECTION ONE –
Background Information
Quickstart
The Axiom is a significant step forward in Biosafety Cabinet technology. While many
internal components are the same or similar to our Logic+-platform products, its
operation and certification are different. The
show you how to answer the questions asked in each case. Keystrokes are shown in
bold blue italics
, while screen selections are shown in
Quickstarts
that follow will help you
green italics
.
Product Service 1-800-522-7658
9
Quickstart
What’s the password for the Axiom?
Like the Logic, the Axiom models password protects the user from entering the
Calibration
the airflow sensor readings. New to the Axiom is a
Configuration
connection status, whether its sash height is 8” or 10” and airflow sensor optionsthere is more on this in the Quickstart “How do I change the BSC configuration?”
The password for all protected screens is:
1. Light
2. UV Light
3. Timer
4. Timer
5. OK/Mute
menu to prevent them from adjusting the blower speed, filter gauge or
Service
; entering this option allows you to configure the Axiom’s exhaust
menu selection entitled
Product Service 1-800-522-7658
10
Quickstart
MAIN MENU
1. MyLogic
TM
2. Settings
3. Service
or to choose,
OK to select, MENU to escape
or to choose,
OK to select, MENU to r eturn
Current Settings
Restore Default
Diagnostics
Configuration
Calibration
Service
ATTENTION!
The following screens are used
to alter the operation of the
Logic. They should only be used
by a qualified certifier as part
of the certification process.
Calibration
OK to proceed, MENU to return
PASSWORD:
Calibration
The ECM setting is: 80
The current RPM is: 1050
The average RPM is: 1020
9. If the password is properly entered, the first certification screen
will display the supply blower motor’s PWM setting. Press the ▲
or ▼ button to increase or decrease the blower speed as
needed. When the speed is set properly, press the
button to lock the new blower speed setting, and continue to the
filter life screen.
Menu
OK/Mute
OK/Mute
warning screen
OK/Mute
then
button.
Services
option is highlighted (it will
to enter the Services menu screen.
Calibration
option is highlighted (it
to start the calibration procedure.
alerting you that you are about to
to input the password.
Light, UV light
OK/Mute
.
OK/Mute
,
Product Service 1-800-522-7658
11
Quickstart
The Inflow displayed
value is: 102 FPM
Calibration
Inflow Display
or to change display value,
UV Light to reset to 105 FPM,
OK to select, MENU to finish
Calibration
Supply Filter Gauge
or to change filter gauge,
OK to select, MENU to escape
Reset Filter Gauge to 100%
Leave Gauge unchanged
Set Gauge to new setting
The ECM setting is: 80
The current RPM is: 1050
The average RPM is: 1020
Calibration
Exhaust Fan Speed
or to change motor speed,
OK to select, MENU to escape
10. If this is an initial certification or a recertification after supply
HEPA filter replacement, select
this is an annual recertification, select
If you want to set the filter gauge at a preset value, select
Gauge to new setting
. Then press
11. The next certification screen will display the exhaust blower motor’s PWM setting. Press the ▲ or ▼ button to increase or
decrease the blower speed as needed. When the speed is set
properly, press the
OK/Mute
speed setting, and continue to the filter life screen.
12. If this is an initial certification or a recertification after exhaust
HEPA filter replacement, select
this is an annual recertification, select
If you want to set the filter gauge at a preset value, select
Gauge to new setting
. Then press
13. If the unit is has no airflow sensor, after the filter gauge option
is selected and you press
down, and then restart with the new settings. If the Unit has an
airflow sensor, you will then be asked to calibrate the inflow
sensor to match the value you obtained when you measured the
inflow. After the inflow is calibrated, and you press
the Axiom will power down, and then restart with the new
settings.
Reset Filter Gauge to 100%
Leave Gauge unchanged
OK/Mute
.
button to lock the new blower
Reset Filter Gauge to 100%
Leave Gauge unchanged
OK/Mute
OK/Mute
, the Axiom will power
.
OK/Mute
. If
.
Set
. If
.
Set
,
Product Service 1-800-522-7658
12
Quickstart
MAIN MENU
1. MyLogic
TM
2. Settings
3. Service
or to choose,
OK to select, MENU to escape
or to choose,
OK to select, MENU to return
Current Settings
Restore Default
Diagnostics
Calibration
Configuration
Service
Current Settings
Supply System
or to change, OK to return
The ECM setting is: 65
The current RPM is: 0
The average RPM is: 0
The new filter RPM is: 1450
C1, with afs, with UV
10484 -
Current Settings
Exhaust System
or to change, OK to return
The ECM setting is: 65
The current RPM is: 0
The average RPM is: 0
The new filter RPM is: 1450
C1, with afs, with UV
10484 -
How do I see the Current Settings of the Axiom?
1. Press the
2. Press the ▼ button until the
turn yellow). Press
screen.
3. Press the ▼ button until the
highlighted (it will turn white). Press
configuration.
4. The first screen shows the current settings of the supply blower
system.
5. By pressing the ▲ or ▼ button, you can see the current
settings of the exhaust blower system. If you are recertifying the
unit, you may want to record the data on this screen. The
firmware revision is shown in the bottom right corner as a -, or
as a letter.
Menu
button.
OK/Mute
Services
option is highlighted (it will
to enter the Services menu
Current Settings
OK/Mute
option is
to see the
Product Service 1-800-522-7658
13
Quickstart
MAIN MENU
1. MyLogic
TM
2. Settings
3. Service
or to choose,
OK to select, MENU to escape
or to choose,
OK to select, MENU to return
Current Settings
Restore Default
Diagnostics
Calibration
Configuration
Service
Configuration
PASSWORD:
ATTENTION!
The following screens are used
to alter the operation of the
Logic. They should only be used
by a qualified certifier as part
of the certification process.
Configuration
OK to proceed, MENU to return
How do I change the Configuration of the Axiom?
1. Press the
2. Press the ▼ button until the
turn yellow).
3. Press
4. Press the ▼ button until the
(it will turn white).
5. Press
6. You will see a alter the BSC’s settings.
7. Press
8. When requested for the password press
Timer, Timer
Menu
OK/Mute
OK/Mute
warning screen
OK/Mute
then
button.
Services
to enter the Services menu screen.
Configuration
to alter the configuration.
alerting you that you are about to
to input the password.
OK/Mute
.
option is highlighted (it will
option is highlighted
Light, UV light
,
Product Service 1-800-522-7658
14
This cabinet is not configured
for a UV light
If you want to install a UV
light, contact Labconco Product
Service at 800-821-5525
Attention
MENU to return
Configuration
Type C Configuration
This cabinet has:
an
or to choose,
OK t o select , MENU to escape
no exhaust connect ion
exhaust connection
Configuration
Type C Configuration
If there is an exhaust system
failure, Active protection should
last:
or to choose,
OK t o select , MENU to escape
0 Seconds
The sash height is:
or to choose,
OK to select, MENU to escape
8 inches
10 inches
NOTE: If you change the sash
height, you must recertify
airflows.
Sash Height Settings
This Logic has:
or to choose,
OK to select, MENU to escape
a UV light
NO UV light
UV Settings
Quickstart
How do I change the Configuration of the Axiom? – continued
9. The first screen allows you to decide if the cabinet is connected to
an exhaust system or not. When the unit is configured for no
exhaust connection, the Axiom will ignore the inlet relief valve
sensor’s status. Selecting connected to an exhaust system
activates the sensor; if the valve is open, the unit operates
normally, if closed, an exhaust alarm is activated. Use the ▲or▼
buttons to select the exhaust configuration, and then press
OK/Mute.
10. If the unit is configured to be connected to an exhaust system,
then the next screen allows you to program the amount of time
interval of Active Protection. Active Protection will allow the Axiom
blowers to continue to operate after the detection of an exhaust
system failure. This will maintain a safe inflow to allow the user a
programmed time interval to secure any hazardous materials in
the cabinet before the BSC shuts off.
11. This screen configures the sash height. After your selection is
chosen, and you press OK/Mute.
12. This screen configures the unit for a UV light or not. After your
selection is chosen, and you press OK/Mute, the unit will shut off
and then reset in the new configuration.
13. If the unit is configured to NOT have a UV light, and the user
presses the UV Light button, they will get this message.
Product Service 1-800-522-7658
15
Quickstart
MAIN MENU
1. MyLogic
TM
2. Settings
3. Service
or to choose,
OK to select, MENU to escape
or to choose,
OK to select, MENU to return
Current Settings
Restore Default
Calibration
Configuration
Diagnostics
Service
Diagnostics
Test the alarm by closing the
inlet valve.
Note the alarm status below:
Open
OK to return
How do I run a diagnostic on the sash position switches?
1. Press the
2. Press the ▼ button until the
turn yellow).
3. Press
4. Press the ▼ button until the
will turn white).
5. Press
6. Press the ▼ button until the Diagnostic
highlighted (it will turn white). Press
7. Lower the sash, and note the status line – it should properly
report the sash as open too high, open, nearly closed or closed
when the sash is in the corresponding position.* Press
OK/Mute
Menu
button.
OK/Mute
OK/Mute
to enter the Services menu screen.
to enter the Diagnostic submenu.
to exit the diagnostic.
Services
option is highlighted (it will
Diagnostic
OK/Mute
option is highlighted (it
Sash sensors
.
option is
Product Service 1-800-522-7658
16
Quickstart
MAIN MENU
1. MyLogic
TM
2. Settings
3. Service
or to choose,
OK to select, MENU to escape
or to choose,
OK to select, MENU to return
Current Settings
Restore Default
Calibration
Configuration
Diagnostics
Service
How do I run a diagnostic on the keypad, relays and contacts?
1. Press the
2. Press the ▼ button until the
turn yellow).
3. Press
4. Press the ▼ button until the
will turn white).
5. Press
6. Press the ▼ button until the Diagnostic
(it will turn white). Press
7. Press any button on the keypad (except
display.
How do I run a diagnostic on the Inlet Relief Valve?
1. Press the
2. Press the ▼ button until the
turn yellow).
3. Press
4. Press the ▼ button until the
will turn white).
5. Press
6. Press the ▼ button until the Diagnostic
is highlighted (it will turn white). Press
7. The display shows the status of the Inlet Relief Valve. During
normal operation, it will be open, and during an exhaust alarm,
it will be closed. NOTE: The inlet relief valve function is
OPPOSITE OF THE MAGNETIC SWITCH FUNCTION! – WHEN
THE INLET VALVE IS OPEN, THE MAGNETIC SWITCH IS
CLOSED.
Menu
OK/Mute
OK/Mute
button.
Services
option is highlighted (it will
to enter the Services menu screen.
Diagnostic
option is highlighted (it
to enter the Diagnostic submenu.
Inlet Relief Valve
OK/Mute
.
Product Service 1-800-522-7658
option
18
MAIN MENU
1. MyLogic
TM
2. Settings
3. Service
or to choose,
OK to select, MENU to escape
or to choose,
OK to select, MENU to return
Current Settings
Restore Default
Calibration
Configuration
Diagnostics
Service
Quickstart
How do I run a diagnostic on the Airflow Sensor?
1. Press the
2. Press the ▼ button until the
turn yellow).
3. Press
4. Press the ▼ button until the
will turn white).
5. Press
6. Press the ▼ button until the Diagnostic
highlighted (it will turn white). Press
7. The display shows the status of the airflow sensor. Its current
output, alarm setting and zero point in Millivolts, along with the
corrected Inflow velocity currently being displayed.
Quickstart
Menu
OK/Mute
OK/Mute
button.
Services
option is highlighted (it will
to enter the Services menu screen.
Diagnostic
option is highlighted (it
to enter the Diagnostic submenu.
Airflow Sensor
OK/Mute
.
Product Service 1-800-522-7658
option is
19
How do I set up automatic operation of the blower and lights?
1. Press the
2. Press the ▼ button until the
turn blue).
3. Press
Menu
OK/Mute
button.
MyLogic
to enter the MyLogic menu screen.
option is highlighted it will
4. Press the ▼ button until the option
option is highlighted (it will turn white).
5. Press the ▼ button to select either Smart-Start or manual
operation of the cabinet blower. (In Smart-Start, the cabinet
blower will automatically start every time the sash is opened.
6. Press the ▼ button to select either Smart-Start or manual
operation of the fluorescent lights.
7. Press the ▼ button to select either Night-Smart or manual
operation of the cabinet blower. (In Night-Smart, the cabinet
blower will automatically operate at idle every time the sash is
closed.
8. Press the ▼ button to select either Night-Smart or manual
operation of the UV Light.*
9. Press the ▼ button to select the time interval the UV light will
operate for until it automatically shuts off.*
* - These screens will only appear if the unit is configured for a UV light.
configure my Logic for use
Axiom Filter Life Gauge Operation and Resetting
for Local Conditions
Product Service 1-800-522-7658
20
The filter life gauge calculates the remaining filter life by measuring and comparing
the current motor speed to the initial speed noted at the factory during assembly,
and the calculated maximum motor speed, using the formula:
100-(100x (current speed-original speed) ) = % filter life left
max. speed-original speed
So
anything
Blocking the grille.
Raising the air temperature.
Lowering the barometric pressure.
will increase the motor’s speed, causing the Filter Life Gauge to decrease. Electronic
dampening of the gauge will limit the change in readings to approximately
1%/minute.
Axiom Biosafety Cabinets being installed at altitudes or operating at temperatures
significantly different from those at the time of manufacture (900 +/- 500 ft above
sea level, 75o +/- 5o F)
conditions. Failure to do so will result in the users noticing a steady drop in Percent
(%) Filter Life remaining as the unit operates. Below are the instructions on how to
correct the Filter Life Gauge for local conditions.
that changes the motor’s current speed, like:
must
have their filter life gauge recalibrated for local
1. Ensure to correct all measurement equipment for local conditions
(temperature, absolute pressure). Anemometers may not have this
ability and might have to be done manually; whereas, a
ShortRidgeTM instrument has a setting to correct for local conditions.
2. As the elevation or the ambient temperature rises, the motor power
factor will have to be reduced to compensate. For an initial
certification, the motor speed should be reduced until it equals the
Axiom’s“Initial Motor Speed”as reported on the unit’s test sheet,
plus approximately 10-15 RPM per each additional 1000 feet of
elevation.
3. The Axiom MUST be recertified using instruments corrected for local
conditions.
4. When the unit is recertified, and operating properly, the certifier
MUST RESET THE FILTER LIFE GAUGE, INDICATING A NEW FILTER
HAS BEEN INSTALLED.
5. Only by performing Step #4 AFTER THE UNIT IS READJUSTED AND
RECERTIFIED FOR ITS LOCAL CONDITION, will the filter life gauge
register 100% (+/- 2%) during operation. As the Filters then load,
the filter life gauge will then begin to drop as appropriate.
Product Service 1-800-522-7658
21
QuickChart – Axiom Type C1 BSCs – Imperial Units
Model
30441
30448
30461
30468
Type
C1
C1
C1
C1
Cabinet Size (in feet)
4 4 6
6
Sash Opening (inches)
10 8 10
8
Starting Serial #
1503_
1503_
1503_
1503_
Downflow Data
Nominal Avg. Downflow (FPM)
65+/-5
65+/-5
55+/-5
55+/-5
grid # of points (rows x columns)
24 (3x8)
24 (3x8)
36(3x12)
36(3x12)
Grid distance from back & sides
6.0
6.0
6.0
6.0
Distance between rows
5.55
5.75
5.55
5.75
Distance between columns
5.21
5.21
5.50
5.50
Inflow Data
Nominal Average Inflow (FPM)
105+/-5
105+/-5
105+/-5
105+/-5
Sash Open Area (Sq. Ft)
3.37
2.69
5.03
4.03
Nominal Avg. Exhaust Vol. (CFM)
354
283
529
423
Avg. Exhaust Vol. Range (CFM)
337-371
269-296
503-553
403-443
Secondary Inflow Data
Sash Opening Template
1
4+
4+
6+
6+
Sensor distance (inches)2
3.25
3.25 4 4
# of Test points
8 8 12
12
Test point location
* * *
*
Nominal Avg. Inflow Vel. (FPM)
354
283
289
231
Avg. Inflow Vel. Range (FPM)
312-344
249-274
252-277
202-222
Correction Factor (CF)
1.08
1.08 2 2
Avg. Inflow Volume (CFM)
337-371
269-296
503-553
403-443
Sash Open Area (Sq. Ft)
3.37
2.69
5.03
4.03
Inflow Velocity Range (FPM)
100-110
100-110
100-110
100-110
Supply HEPA Data
Air Displacement (CFM)
530
530
660
660
Laskin Nozzles needed
1 1 1
1
Theoretical aerosol conc. (ug/l)3
25
25
20
20
Width x Depth x Height (in.)4
48x18x3.06
48x18x3.06
72x18x3.06
72x18x3.06
Performance (CFM)
445
445
665
665
Performance (Pressure in. H2O)
<.45”
<.45”
<.42”
<.42”
Labconco P/N
3838401
3838401
3838403
3838403
Exhaust HEPA Data
Air Displacement (CFM)
354
283
529
423
Laskin Nozzles needed
1 1 1
1
Theoretical aerosol conc. (ug/l)3
38
48
26
32
Width x Depth x Height (in.)
4
26x18x5.88
26x18x5.88
48x18x5.88
48x18x5.88
Performance (CFM)
354
354
529
52
Performance (Pressure in. H2O)
<0.3
<0.3
<0.3
<0.3
Labconco P/N
3838501
3838501
3838503
3838503
Exhaust Volume Data
DIM Exhaust Volume (CFM)5
400
323
570
463
Traverse Exhaust Volume (CFM)6
480
387
684
556
Differential Pressure (in. H20)7
0.3
0.3
0.3
0.3
IRV Inflow Range (CFM)8
20-100
20-100
20-100
20-100
Nominal IRV Slot Velocity (FPM)
240-260
240-260
240-260
240-260
Nominal IRV Slot Volume (CFM)
50
50
50
50
Supply Motor/Blower Data
Motor HP
1/3
1/3
1/3
1/3
Exhaust Motor/Blower Data
Motor HP
1/3
1/3
1/3
1/3
Fluorescent Lamp Data
Fluorescent Lamps (2 each)
F32T8 TL741
F32T8 TL741
F40T8 SP41
F40T8 SP41
Fluor. Lamp Labconco Part #
9721900
9721900
9721903
9721903
Fluor. Ballast Labconco Part #
3838100
3838100
3838100
3838100
Product Service 1-800-522-7658
22
QuickChart – Axiom Type C1 BSCs – Metric Units
Model
30441
30448
30461
30468
Type
C1
C1
C1
C1
Cabinet Size (in m)
1.22
1.22
1.83
1.83
Sash Opening (mm)
254
203
254
203
Starting Serial #
1503_
1503_
1503_
1503_
Downflow Data
Nominal Avg. Downflow (FPM)
65+/-5
65+/-5
55+/-5
55+/-5
grid # of points (rows x columns)
24 (3x8)
24 (3x8)
36(3x12)
36(3x12)
Grid distance from back & sides (mm)
152
152
152
152
Distance between rows (mm)
141
146
141
146
Distance between columns (mm)
132
132
140
140
Inflow Data
Nominal Average Inflow (m/s)
.533+/-.03
.533+/-.03
.533+/-.03
.533+/-.03
Sash Open Area (Sq. m)
.313
.250
5.03
4.03
Nominal Avg. Exhaust Vol. (m3/s)
354
283
.468
.374
Avg. Exhaust Vol. Range (m3/s)
337-371
269-296
503-553
403-443
Secondary Inflow Data
Sash Opening Template
1
4+
4+
6+
6+
Sensor distance (mm)2
83
83
102
102
# of Test points
8 8 12
12
Test point location
* * *
*
Nominal Avg. Inflow Vel. (m/s)
1.80
1.44
1.47
1.17
Avg. Inflow Vel. Range (m/s)
1.71-1.88
1.37-1.50
1.40-1.53
1.12-1.23
Correction Factor (CF)
.0929
.0929
.1700
.1700
Average Inflow Volume Range
(AIV) (Avg. velocity x CF)
.159-.175
.127-.140
.238-.261
.190-.209
Sash Open Area (Sq. m)
.313
.250
.468
.374
Inflow Velocity Range
(AIV / Sash open area)
.533+/-.03
.533+/-.03
.533+/-.03
.533+/-.03
Supply HEPA Data
Air Displacement (m3/s)
.250
.250
.311
.311
Laskin Nozzles needed
1 1 1
1
Theoretical aerosol conc. (ug/l)3
25
25
20
20
Width x Depth x Height (in.)4
48x18x3.06
48x18x3.06
72x18x3.06
72x18x3.06
Performance (CFM)
445
445
665
665
Performance (Pressure in. H2O)
<.45”
<.45”
<.42”
<.42”
Labconco P/N
3838401
3838401
3838403
3838403
Exhaust HEPA Data
Air Displacement (m3/s)
.167
.134
.250
.200
Laskin Nozzles needed
1 1 1
1
Theoretical aerosol conc. (ug/l)3
38
48
26
32
Width x Depth x Height (in.)
4
26x18x5.88
26x18x5.88
48x18x5.88
48x18x5.88
Performance (CFM)
283
283
423
423
Performance (Pressure in. H2O)
<0.3
<0.3
<0.3
<0.3
Labconco P/N
3838501
3838501
3838503
3838503
Exhaust Volume Data
DIM Exhaust Volume (m3/s)5
.189
.152
.269
.219
Traverse Exhaust Volume (m,3/s)6
.227
.183
.323
.262
Differential Pressure (Pascal)7
75
75
75
75
IRV Inflow Range (m/s)
.010-.047
.010-.047
.010-.047
.010-.047
Nominal IRV Slot Velocity (m/s)
1.22-1.32
1.22-1.32
1.22-1.32
1.22-1.32
Nominal IRV Slot Volume (m3/s)
.235
.235
.235
.235
Supply Motor/Blower Data
Motor HP
1/3
1/3
1/3
1/3
Exhaust Motor/Blower Data
Motor HP
1/3
1/3
1/3
1/3
Fluorescent Lamp Data
Fluorescent Lamps (2 each)
F32T8 TL741
F32T8 TL741
F40T8 SP41
F40T8 SP41
Fluor. Lamp Labconco Part #
9721900
9721900
9721903
9721903
Fluor. Ballast Labconco Part #
3838100
3838100
3838100
3838100
Product Service 1-800-522-7658
23
QuickChart – Axiom/Cell Axiom Type C1 BSCs footnotes
Axiom Model Identification
The primary serial tag is on the lower right edge of the front dress panel.
The secondary serial tag is located on the front of the electronics module on the top
right side of the cabinet.
The first two digits of the serial number are the year of production; the next two are
the month. The next 5 digits are the sequence of production, and the letter following
the serial number is the revision level of the cabinet.
Downflow Test Specifications
All models are classified as uniform downflow.
All tests performed as described in ANSI/NSF Standard 49:2012.
UV Lamp, IV bar and all other accessories must be removed before measuring
downflow.
Inflow Test Specifications
All tests performed as described in ANSI/NSF Standard 49:2012.
Secondary Inflow Test Specifications
Must use Labconco holder P/N 3836405 to perform this test properly.
1. Use the appropriate template included in Certifier Kit# 3858403.
2. Measured from the bottom edge of the probe holder to the center of the thermal
anemometer sensor element.
* - Locate the single row of holes at the front of the grille. Mark the 6th hole from the
side wall and subsequent test points every 9 holes until the number of test points
marked equals the width of the cabinet in feet (3-foot cabinet, mark the first 3 points).
Repeat for the opposite side.
HEPA Filter Leak Test Specifications
3. Based on mineral oil.
4. Without gasket
Aerosol generator should be placed in the left rear corner of the work area, pointing
toward the rear grille.
For uncontaminated units, the upstream concentration can be sampled from the
tube located under the work surface.
Metric calculation of concentration is:
# of Laskin nozzles @ 138K Pascal x 6.372/Volume of air in m3/sec. = Conc.
In ug/l
Exhaust Data
5. Total Exhaust Volume; The sum of the inflow and air drawn into the Inlet Relief
6. Measured as per ASHRAE methodology for measuring air volume in round ducts.
7. Measured at the exhaust transition sampling point, relative to atmosphere.
8. IRV = Inlet Relief Valve.
UV lamp Specifications
For all models, the UV lamp number is G30T8. The Labconco part number is
For all 115/230 VAC models, the ballast assembly is Labconco part number 3829901.
For all 100 VAC models, the ballast assembly is Labconco part number 3830600. The
Valve.
1271300.
The ballast is Robertson part number PSM2GPH18MVW.
ballast is Robertson part number RSO1GPH30100.
Product Service 1-800-522-7658
24
IDENTIFYING AXIOM MODELS
The Axiom platform Biosafety Cabinet can be identified readily by several
external features, as shown in Figure 1-1:
Model number Stars with 304xxxxxx
Serial number beginning with 1503xxxxx.
Exhaust System Cover.
Segmented Work Surfaces.
Figure 1-1
Exhaust System Cover
Axiom Logo
Segmented Work
Surfaces
Product Service 1-800-522-7658
25
SERIAL NUMBER TAGS
The model/serial number tag is located on the bottom right edge of the
dress panel. A second tag is located on the front of the electronics module
on the top of the unit. This tag includes the unit’s operating voltage and
amperage. The model and serial numbers of each unit are also recorded on
the unit test report attached beneath the front dress panel. These items are
shown graphically in the figures 1-3 to 1-5 below:
Figure 1-3
Primary Serial Tag
Figure 1-4
Secondary Serial Tag
Figure 1-5
Final Inspection and Test Report Form
Obviously, you can get confirmation of the Biosafety Cabinet's model by
contacting Labconco's Product Service Department, and providing us with
the serial number. From this number, we can tell the date of production and
its model.
Product Service 1-800-522-7658
26
AXIOM MODEL NUMBERING SYSTEM
Axiom units’ model numbers define the configuration of the cabinet. The first
five digits define the unit's configuration, while the last four are assigned
based on the unit's options and electrical configuration. Thus, a model
number of:
30448xxxx
Labconco Product Group
3=Laminar Flow Products
Indicates the Product Family
0= Logic
1= PuriCare
2-4= Cell Logic
The Axiom Data Plate contains much of the information required to certify
the unit. Information on the plate includes nominal inflow and downflow
velocity settings, a description of how to establish the downflow grid and the
secondary inflow methodology are also included on the data plate, as shown
below. The plate is located in the upper left corner of the dress panel.
Figure 1-6
Product Service 1-800-522-7658
28
Final Inspection and Test Report Form
A copy of the Final Inspection Test Report is attached to the blower plenum
cover, beneath the front dress panel. The report contains initial setup
information on inflow and downflow velocities, and the initial blower torque
setting. A copy of the report is shown in Figure 1-7, below
Figure 1-7
Model/Serial
Numbers
HEPA Filter
Information
Test Results
Downflow Test
Results
Product Service 1-800-522-7658
29
Theory of Operation
Background
The Axiom Type C1 BSC was the result of a customer request for a Type B
(directly exhausted) cabinet that could operate at a constant exhaust system
vacuum of only 0.25 inches of H2O. Realizing that this was impossible for
existing Type B technologies, Labconco developed a new type of BSC; The
proposed Type C1.
Figure 1-8
Airflow Patterns and Features
The overall airflow pattern in the Type C1 is similar to a Type B1, with three
important differences:
1. In the C1, the directly exhausted portion of downflow air is
taken from the center of the work area (The Chem-ZoneTM).
2. An inlet relief valve, unique to the Axiom, stabilizes the inflow
as exhaust system airflows fluctuates, and closes during an
exhaust system failure, signaling the Axiom to initiate a timed
“Active Protection” of the user and their work.
3. A constant volume exhaust blower upstream of the Exhaust
HEPA maintains constant inflow, and allows the unit’s blowers
to continue to operate during an exhaust system failure
(Active Protection), maintaining safe inflows and downflows
while the user secures any hazardous materials in the work
area.
Product Service 1-800-522-7658
30
Figure 1-8a
Chem-ZoneTM Directly Exhausted Work Zone
Unique to the Axiom is the Chem-Zone, a directly exhausted work zone. The
central portion of the work surface is surrounded by grilles on the sides,
front and back. Air entering these grilles is drawn to the Exhaust HEPA filter,
and when the cabinet is connected to an exhaust system, out of the
laboratory. This feature prevents the recirculation of volatile chemicals as
seen in Type A cabinets, while exhausting much less air than Type B2
models. This is illustrated in Figure 1-8a.
Product Service 1-800-522-7658
Loading...
+ 125 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.