Labconco PuriCare Vertical Flow Stations User Manual

PuriCare
Ve rtical Flow Stations
Models 3830000 3830020
User’s Manual
LABCONCO CORPORATION 8811 PROSPECT AVENUE KANSAS CITY, MO 64132-2696 800-821-5525, 816-333-8811 FAX 816-363-0130 E-MAIL labconco@labconco.com HOME PAGE www.labconco.com
Labconco’s Mascot,
Labby the LABster
Register your product online at
www.labconco.com/productreg.html
or return the attached card and
receive FREE LabbyWear
!
Copyright Information
Copyright © 2003 Labconco Corporation. All rights reserved.
Labconco Corporation 8811 Prospect Avenue Kansas City, Missouri 64132
The information contained in this manual and the accompanying product are copyrighted and all rights are reserved by Labconco Corporation. Labconco Corporation reserves the right to make periodic design changes without obligation to notify any person or entity of such change.
PuriCare™ is a trademark of Labconco Corporation.
Warranty
Labconco provides a warranty on all parts and factory workmanship. The warranty includes areas of defective material and workmanship, provided such defect results from normal and proper use of the equipment.
The warranty for all Labconco products will expire one year from date of installation or two years from date of shipment from Labconco, whichever is sooner, except the following:
Purifier® Delta® Series Biological Safety Cabinets and PuriCare™ Vertical Flow Stations carry a three-year warranty from date of installation or four years from date of shipment from Labconco, whichever is sooner.
Carts carry a lifetime warranty.
Glassware is not warranted from breakage when dropped or mishandled.
This limited warranty covers parts and labor, but not transportation and insurance charges. In the event of a warranty claim, contact Labconco Corporation or the dealer who sold you the product. If the cause is determined to be a manufacturing fault, the dealer or Labconco Corporation will repair or replace all defective parts to restore the unit to operation. Under no circumstances shall Labconco Corporation be liable for indirect, consequential, or special damages of any kind. This statement may be altered by a specific published amendment. No individual has authorization to alter the provisions of this warranty policy or its amendments. Lamps and filters are not covered by this warranty. Damage due to corrosion or accidental breakage is also not covered.
Limitation of Liability
The disposal and/or emission of substances used in connection with this equipment may be governed by various federal, state, or local regulations. All users of this equipment are required to become familiar with any regulations that apply in the user’s area concerning the dumping of waste materials in or upon water, land, or air and to comply with such regulations. Labconco Corporation is held harmless with respect to user’s compliance with such regulations.
Contacting Labconco Corporation
If you have questions that are not addressed in this manual, or if you need technical assistance, contact Labconco’s Customer Service Department or Labconco’s Product
)
Visit Labconco’s web site at: http://www.labconco.com
Service Department at 1-800-821-5525 or 1-816-333-8811, between the hours of 7:00 a.m. and 6:00 p.m., Central Standard Time.
or email Labconco at: labconco@labconco.com.
Part #3794300, Rev A
ECO C722
T
AABBLLEE
T
CHAPTER 1: INTRODUCTION 1 About This Manual 2 Typographical Conventions 3
CHAPTER 2: PREREQUISITES 5 Location Requirements 6 Exhaust Requirements 6 Electrical Requirements 7 Space Requirements 8 Overhead Clearance 8
CHAPTER 3: GETTING STARTED 9 Unpacking Your Vertical Flow Station 10 PuriCare Components 11 Preparing the Station for Operation 11 Exhaust System Requirements 12 Initial Certification 12
CHAPTER 4: THEORY OF OPERATION AND
HEPA Filters 14 Directional Airflow 15 Impellers and Motor/Blower 16 Speed Controls 16 Station Air Intakes (Grilles), Ductwork & Air Balance Controls 17 Safety Precautions 17
CHAPTER 5: USING YOUR PURICARE 21 Operating the Sash 21 Working in Your PuriCare 23
CHAPTER 6: MAINTAINING YOUR PURICARE 26 Routine Maintenance Schedule 26 Service Operations 27
O
O
SAFETY PRECAUTIONS 13
FF
C
OONNTTEENNTTSS
C
Downflow Velocities 30 Downflow Velocity Grid Patterns 30 Inflow Velocity Calculation – Primary Method 30 Inflow Velocity Calculation – Secondary Method 31 Speed Control Adjustments 32 Supply HEPA Filter Replacement 33 Exhaust HEPA Filter Replacement 35 Impeller and Motor/Blower Maintenance & Replacement 36 Diffuser Removal 38 Storage 39
CHAPTER 7: MODIFYING YOUR VERTICAL FLOW
STATION 40
Installing a Shelf 40
CHAPTER 8: TROUBLESHOOTING 42
APPENDIX A: VERTICAL FLOW STATION COMPONENTS 44
APPENDIX B: VERTICAL FLOW STATION DIMENSIONS 46
APPENDIX C: VERTICAL FLOW STATION
SPECIFICATIONS 47
Electrical Data 47 Impeller Specifications 47 Motor Specifications 48
APPENDIX D: PURICARE ACCESSORIES 51 Shelf Kit 51
APPENDIX E: QUICK CHART FOR THE PURICARE VERTICAL FLOW STATIONS 52
APPENDIX F: REFERENCES 53
DECLARATION OF CONFORMITY 54
C
HHAAPPTTEERR
C
I
NNTTRROODDUUCCTTIIOONN
I
Congratulations on your purchase of a Labconco PuriCare Series Vertical Flow Station. Your PuriCare product is designed to protect you, your product and your laboratory environment from aerosols. It is the result of Labconco’s years of experience manufacturing laboratory equipment, and many of its features were suggested to us by users like you.
The PuriCare offers many unique features to enhance safety, performance and ergonomics. To take full advantage of them, please acquaint yourself with this manual and keep it handy for future reference. If you are unfamiliar with how Vertical Flow Stations operate, please review Chapter 4: Theory of Operation and Safety Precautions before you begin working in the station. Even if you are an experienced user, please review Chapter 5: Using Your Vertical Flow Station; it describes your PuriCare’s features so that you can use the station efficiently.
1
1
Product Service 1-800-522-7658
1
Chapter 1: Introduction
About This Manual
This manual is designed to help you learn how to install, use, and maintain your Vertical Flow Station. Instructions for installing optional equipment on your station are also included.
Chapter 1: Introduction provides a brief overview of the Vertical Flow Station, explains the organization of the manual, and defines the typographical conventions used in the manual.
Chapter 2: Prerequisites explains what you need to do to prepare your site before you install your Vertical Flow Station. Electrical and service requirements are discussed.
Chapter 3: Getting Started contains the information you need to properly unpack, inspect, install, and certify your Vertical Flow Station.
Chapter 4: Theory Of Operation And Safety Precautions explains how the PuriCare operates and the
appropriate precautions you should take when using the station.
Chapter 5: Using Your PuriCare discusses the basic operation of your station. Information on how to prepare, use and shut down your PuriCare are included.
Chapter 6: Maintaining Your PuriCare explains how to perform routine maintenance on your Vertical Flow Station. Information on how to safely disinfect the interior of your station and replace the lamps are included.
Chapter 7: Modifying Your Vertical Flow Station describes how to install the optional equipment on the station.
Chapter 8: Troubleshooting contains a table of problems you may encounter while using your Vertical Flow Station including the probable causes of the problems and suggested corrective actions.
2
Product Service 1-800-522-7658
Appendix A: Vertical Flow Station Components
contains labeled diagrams of all of its components.
Appendix B: Vertical Flow Station Dimensions contains comprehensive diagrams showing all of the dimensions for the PuriCare model.
Appendix C: Vertical Flow Station Specifications contains the electrical requirements for the Vertical Flow Station. Wiring diagrams for both the 115V and 230V models are also included.
Appendix D: PuriCare Accessories lists the part number and descriptions of all of the accessories available for your Vertical Flow Station.
Appendix E: Quick Chart for the PuriCare Vertical Flow Station provides useful operating specifications.
Appendix F: References lists the various resources
available that deal with biosafety.
Chapter 1: Introduction
Typographical Conventions
Recognizing the following typographical conventions will help you understand and use this manual:
Book, chapter, and section titles are shown in italic
type (e.g., Chapter 3: Getting Started).
Steps required to perform a task are presented in a
numbered format.
Comments located in the margins provide
suggestions, reminders, and references.
Critical biosafety information is presented in
boldface type in paragraphs that are preceded by the biosafety icon. Failure to comply with the information following a biosafety icon may result in illness or death.
Product Service 1-800-522-7658
3
Chapter 1: Introduction
!
)
115V
230V
Critical information is presented in boldface type in paragraphs that are preceded by the exclamation icon. Failure to comply with the information following an exclamation icon may result in injury to the user or permanent damage to your Vertical Flow Station.
Critical information is presented in boldface type in paragraphs that are preceded by the wrench icon. These operations should only be performed by a trained certifier or contractor. Failure to comply with the information following a wrench icon may result in injury to the user or permanent damage to your Vertical Flow Station.
Important information is presented in capitalized type in paragraphs that are preceded by the pointer icon. It is imperative that the information contained in these paragraphs be thoroughly read and understood by the user.
Information that is specific to a particular model of Vertical Flow Station is preceded by a number icon. The 115V icon indicates the text is specific to the 115 volt models. The 230V icon indicates the text is specific to the 230 volt models.
4
Product Service 1-800-522-7658
C
HHAAPPTTEERR
C
P
RREERREEQQUUIISSIITTEESS
P
Before you install your Vertical Flow Station, you need to prepare your site for installation. Carefully examine the location where you intend to install your station. You must be certain that the area is level and of solid construction. In addition, a dedicated source of electrical power must be located near the installation site.
Carefully read this chapter to learn:
the location requirements for your installation site.
the electrical power requirements for your installation site.
the exhaust requirements for your installation site.
the space requirements for your installation site.
Refer to Appendix C: Vertical Flow Station Specifications for complete station electrical and environmental conditions, specifications and requirements.
2
2
Product Service 1-800-522-7658
5
Chapter 2: Prerequisites
Location Requirements
The PuriCare should be located away
!
from traffic patterns, doors, fans, ventilation registers, fume hoods and any other air-handling device that could disrupt its airflow patterns. All windows in the room should be closed. Figure 2-1 shows the optimum location for the PuriCare.
Figure 2-1
6
Exhaust Requirements
THE VERTICAL FLOW STATION DISCHARGES ITS FILTERED
)
EXHAUST AIR UNDERNEATH THE UNIT. IT SHOULD BE PLACED AT LEAST 12 INCHES AWAY FROM ANY VERTICAL OBSTRUCTION TO OPERATE PROPERLY.
Product Service 1-800-522-7658
Electrical Requirements
The different PuriCare models have the following electrical requirements:
Model # Requirements
38300- 00-09 115 VAC, 60 Hz, 12 Amps 38300- 20-29 230 VAC, 50 Hz, 6 Amps
All PuriCare Stations with model numbers ending in – 00 to –09 are designed for operation at 115 volts, 60Hz, alternating current. PuriCare Stations with model numbers ending in –20 to –29 are designed for operation at 230 +/- 20 volts, 50 Hz alternating current. A dedicated outlet with a circuit breaker rated at 20 Amps for 115 volt models and 10 Amps for 230 volt models, should be located within 10 feet of the station.
Always follow the plug manufacturer’s instructions for the
!
115V
proper assembly and testing of the plug and power cord.
THE ELECTRICAL OUTLETS OF THE 115 VOLT PURICARE IS PROTECTED BY A GROUND FAULT INTERRUPTER CIRCUIT (GFIC). LABCONCO DOES NOT RECOMMEND INSTALLING A GFIC ON THE OUTLET THAT THE PURICARE PLUGS INTO.
Table 2-1
Chapter 2: Prerequisites
Product Service 1-800-522-7658
7
Chapter 2: Prerequisites
Space Requirements
The dimensions for the different models are shown in Appendix B: Vertical Flow Station Dimensions.
Overhead Clearance
In order for the Vertical Flow Station to operate properly, at least six inches (150 mm) clearance from any overhead obstructions is required when the station is in its final operating position.
8
Product Service 1-800-522-7658
C
HHAAPPTTEERR
C
G
EETTTTIINNGG
G
Now that the site for your Vertical Flow Station is properly prepared, you are ready to unpack, inspect, install, and certify it. Read this chapter to learn how to:
unpack and move your PuriCare.
set up the station.
connect the electrical supply source.
arrange certification.
Depending upon which model you are installing, you may need common tools in addition to a 3/4" wrench, a flat-blade screwdriver, and a Phillips screwdriver.
The PuriCare models weigh approximately 450 lbs. (205 kg). The
!
carton allows for lifting with a mechanical lift truck or floor jack.
S
S
3
3
TTAARRTTEEDD
Product Service 1-800-522-7658
9
Chapter 3: Getting Started
The United States Interstate Commerce Commission rules require that claims be filed with the delivery carrier within fifteen (15) days of delivery.
Unpacking Your Vertical Flow Station
Carefully unpack your PuriCare and inspect it for damage that may have occurred in transit. If your unit is damaged, notify the delivery carrier immediately and retain the entire shipment intact for inspection by the carrier.
DO NOT RETURN GOODS WITHOUT THE PRIOR
)
Do not discard the carton or packing material for your PuriCare until you have checked all of the components and installed and tested the unit.
!
AUTHORIZATION OF LABCONCO. UNAUTHORIZED RETURNS WILL NOT BE ACCEPTED.
IF YOUR STATION WAS DAMAGED IN TRANSIT, YOU MUST FILE A CLAIM DIRECTLY WITH THE FREIGHT CARRIER. LABCONCO CORPORATION AND ITS DEALERS ARE NOT RESPONSIBLE FOR SHIPPING DAMAGES.
Do not remove the PuriCare from its shipping skid until it is ready to be placed into its final location. Move the unit by placing a flat, low dolly under the shipping skid, or by using a floor jack.
Do not move the station by tilting it onto a hand truck.
10
Product Service 1-800-522-7658
PuriCare Components
Labconco manufactures Vertical Flow Stations for use with 115 volts or 230 volts.
Locate the station model you received in the following group of tables. Verify that the components listed are present and undamaged.
Catalog # PuriCare Description
3830000 Vertical Flow Station, 115 VAC, 60 Hz 3830020 Vertical Flow Station, 230 VAC, 50 Hz
Plus the Following:
Part # Component Description
3794300 User’s Manual Power Cord (4) Caster Wheels (8) Washers (4) Lock Washers (4) Acorn Nuts
If you did not receive one or more of the components listed for your PuriCare, or if any of the components are damaged, contact Labconco Corporation immediately for further instructions.
Chapter 3: Getting Started
Preparing the Station for Operation
ASSEMBLY INSTRUCTIONS FOR
)
Product Service 1-800-522-7658
THE STATION (LABCONCO P/N
1058300) ARE ATTACHED TO THE SASH OF THE PURICARE. IF THESE INSTRUCTIONS ARE MISSING OR UNCLEAR, CONTACT LABCONCO AT 800-821-5525.
11
Chapter 3: Getting Started
Exhaust System Requirements
The Vertical Flow Station discharges its HEPA filtered exhaust air beneath the unit.
Maintain at least 12 inches clearance
!
between the station and any vertical obstruction on all sides of the unit.
Initial Certification
Prior to use, all PuriCare stations should be certified by a qualified certifier. Under normal operating conditions, the PuriCare stations should be recertified at least annually or when serviced. The certifier should perform the following tests:
HEPA Filter Leak Tests
Downflow Velocity Profile Test*
Inflow Volume Test
Airflow Smoke Patterns
*There is no tolerance on the downflow profile values.
If you have any questions regarding certification agencies or need assistance in locating one, contact Labconco’s Product Service Department at 1-800-522­7658 or 816-333-8811.
12
Product Service 1-800-522-7658
C
HHAAPPTTEERR
C
T
HHEEOORRY
T
O
PPEERRAATTIIOONN
O
S
AAFFEETTYY
S
P
RREECCAAUUTTIIOONNSS
P
The Vertical Flow Station operates using the following principles:
Filtration and retention of particulates by High Efficiency Particulate Air (HEPA) filters Directional airflow
The major components in the Vertical Flow Station are:
The HEPA filters
The impellers and motor/blower to force air
through the unit
A speed controls for the impellers and motor
Station air intakes (grilles)
Y
O
O
4
4
F
F
A
A
NNDD
Product Service 1-800-522-7658
13
Chapter 4: Theory Of Operation And Safety Precautions
HEPA Filters
HEPA filters are disposable, dry-type particulate filters. The filter material or media is typically made of borosilicate microfibers that are made into a thin sheet, in a process similar to the production of paper. This sheet is folded, or pleated to increase its surface area. The pleats are held in place by aluminum diffusers or by beads of glue that add rigidity to the media pack. The media pack is then set into a suitable frame, and the perimeter sealed to the filter frame, as shown in Figure 4-1.
The HEPA filter media is very fragile. Do not touch or contact the media
!
!
surface. If you think the surface of a HEPA filter is damaged, DO NOT USE THE STATION. Have the HEPA filter integrity tested by a qualified certifier before using the station.
HEPA Filters are only effective against particulate material. Gases will pass through the filter.
14
Product Service 1-800-522-7658
Chapter 4: Theory Of Operation And Safety Precautions
Figure 4-1
Directional Airflow
Directional airflow plays a key role in the Vertical Flow Station’s performance. Sterile, HEPA filtered air flows downward throughout the work area, protecting the animals inside from external and cross contamination. All of this “supply” air and some room air is drawn into the front of the station at the grille. This “curtain” of air makes it difficult for aerosols to escape out of the work area and into the outside environment.
Product Service 1-800-522-7658
15
Chapter 4: Theory Of Operation And Safety Precautions
Figure 4-2
Impellers and Motor/Blower
The Vertical Flow Station uses two air handling systems to direct air through it. In the top of the station twin motorized impellers draw room air in through prefilters and then force the air through the upper (supply) HEPA filter. Below the worksurface there are another set of prefilters, a heavy duty motor blower and the exhaust HEPA filter.
Speed Controls
The speed controls should only be
!
adjusted by a qualified certifier.
16
Product Service 1-800-522-7658
Chapter 4: Theory Of Operation And Safety Precautions
The speed controls are electronic circuits that allow the certifier to set the speed of the motors by adjusting their voltage. The PuriCare speed controls are rated for 10 Amps current each, far in excess of the motor's normal current draw, to allow for greater reliability.
Station Air Intakes (Grilles), Ductwork and Air Balance Controls
The station’s containment and performance are affected by the location, size, and pattern of the grilles at the front and rear of the work area.
Never block or obstruct the grilles of
!
The internal ductwork of the PuriCare conveys the air from the work area to the blower, and then from the blower to the filters. The exhaust filter plenum of the PuriCare is unique in the industry, utilizing perfect manifold technology to deliver a more uniform airflow to the exhaust HEPA filter, optimizing filter loading and operational life.
the PuriCare.
Safety Precautions
The PuriCare Vertical Flow Station is NOT a biosafety cabinet. Do not use this device for the containment of biohazardous, toxic, flammable, or explosive materials. If you have questions regarding the operation of this device, contact Labconco at 800­821-5525 or www.labconco.com.
Product Service 1-800-522-7658
17
Chapter 4: Theory Of Operation And Safety Precautions
HEPA filters are only effective for the
!
!
entrapment of particulate matter. Manipulations which generate gases or vapors, i.e., toxic chemicals or radionuclides, should not be performed in this station.
The PuriCare Vertical Flow Station should be certified by a qualified certification technician before its initial use. The station should be recertified whenever it is relocated, serviced or at least annually thereafter.
Some components of the PuriCare Vertical Flow Station should only be serviced by a qualified certification technician. Some internal components of the PuriCare may become contaminated with allergenic materials during operation of the unit. Only experienced personnel competent in decontamination procedures should decontaminate the station before servicing contaminated components. If you have any questions regarding certification agencies, or need assistance in locating one, contact Labconco’s Product Service Department at 800-522-7658 or 816-333-8811.
Ensure that the unit is connected to electrical service in accordance with local and national electrical codes. Failure to do so may create a fire or electrical hazard. Do not remove or service any electrical components without first disconnecting the PuriCare from electrical service.
18
Product Service 1-800-522-7658
)
Chapter 4: Theory Of Operation And Safety Precautions
Avoid the use of flammable gases or solvents in the PuriCare. Care must be taken to ensure against the concentration of flammable or explosive gases or vapors. An open flame should NOT be used in the PuriCare. Open flames may disrupt the airflow patterns in the station, burn the HEPA filter and/or damage the filter’s adhesive. Gases under high pressure should not be used in the PuriCare station, as they may disrupt the airflow patterns of the station.
The surface of the HEPA filters are fragile and should not be touched. Care must be taken to avoid puncturing either HEPA filter during installation or normal operation. If you suspect that a HEPA filter has been damaged, DO NOT use the station; contact a local certification agency or Labconco at 800-821-5525 or 816-333-8811 for recertification information.
THE HEPA FILTERS IN THE PURICARE STATION WILL GRADUALLY ACCUMULATE AIRBORNE PARTICULATE MATTER FROM THE ROOM AND FROM WORK PERFORMED IN THE STATION. THE RATE OF ACCUMULATION WILL DEPEND UPON THE CLEANLINESS OF THE ROOM AIR, THE AMOUNT OF TIME THE STATION IS OPERATING AND THE NATURE OF WORK BEING DONE IN THE STATION.
Product Service 1-800-522-7658
19
Chapter 4: Theory Of Operation And Safety Precautions
Proper operation of the station depends largely upon the station’s
!
)
location and the operator’s work habits. Consult the Installation and Normal Operation sections of this manual for further details.
WHEN SURFACE DISINFECTING THE PURICARE STATION:
AVOID SPLASHING THE DISINFECTING SOLUTION ON SKIN OR CLOTHING.
ENSURE ADEQUATE VENTILATION.
CAREFULLY FOLLOW THE MANUFACTURER’S SAFETY INSTRUCTIONS WHEN HANDLING DISINFECTANTS AND ALWAYS DISPOSE OF DISINFECTING SOLUTIONS IN ACCORDANCE WITH LOCAL AND NATIONAL LAWS.
DO NOT ALLOW DISINFECTANTS WITH FREE CHLORINE TO CONTACT THE STAINLESS STEEL COMPONENTS OF THE PURICARE FOR A LONG PERIOD OF TIME. FREE CHLORINE WILL CORRODE STAINLESS STEEL AFTER EXTENDED CONTACT.
The electrical receptacle cover, because of its construction, may be difficult to surface decontaminate. In the event of gross contamination, the cover should be removed, sterilized and/or decontaminated as required and discarded. The receptacle cover should then be replaced with the repair part listed in Appendix A: PuriCare Components.
20
Product Service 1-800-522-7658
C
HHAAPPTTEERR
C
U
SSIINNG
U
P
UURRI
P
Operating the Sash
The sash of the Vertical Flow Station can be locked in the full open position. To lock the sash open:
1. Grasp the bottom edge of the sash, and pivot it up until the sash is fully opened.
2. Locate the sash locking bolt on the inside left side of the sash, as shown in Figure 5-1.
3. Restrain the sash by aligning the bolt with the hole in the left side wall of the station, and pressing the bolt to the left until it engages the hole.
4. To release the sash, support the bottom edge while pulling the bolt to the right. Allow the sash to pivot down to its closed position, ensuring that the sash is set in the closed position.
I
G
C
C
Y
Y
AARREE
5
5
OOUURR
Product Service 1-800-522-7658
21
Chapter 5: Using Your PuriCare
Figure 5-1
22
The sash must be completely closed
!
5. Turn the blower and light switches to the 'On' position, as shown in Figure 5-2.
for the PuriCare to contain properly.
Product Service 1-800-522-7658
Figure 5-2
Working In Your PuriCare
Planning
Thoroughly understand procedures and equipment required before beginning work.
Arrange for minimal disruptions, such as room traffic or entry into the room while the station is in use.
Start-up
Turn on fluorescent light and station blower.
Check the return air grilles for obstructions, and
note the pressure gauge reading.
Allow the station to operate unobstructed for 5 minutes.
Wash hands and arms thoroughly with germicidal soap.
Wear a long sleeved lab coat with knit cuffs and over-the- cuff rubber gloves. Use protective eyewear. Wear a protective mask if appropriate.
Wipe-Down
Pivot the sash to its full open position.
Wipe down the interior surfaces of the station
with 70% ethanol, or a suitable disinfectant, and allow to dry.
Slowly close the sash.
Chapter 5: Using Your PuriCare
Product Service 1-800-522-7658
23
Chapter 5: Using Your PuriCare
Loading Materials and Equipment
Work Techniques
Final Purging
Unloading Materials and Equipment
Only load the materials required for the procedure. Do not overload the station.
Do not obstruct the grilles on the front, back or sides of the work area.
Large objects should not be placed close together.
Keep all materials at least 4 inches inside of the sash, and perform all contaminated operations as close to the center of the work area as possible.
Segregate all clean and contaminated materials in the work area.
Arrange materials to minimize the movement of contaminated materials into clean areas.
Keep all discarded contaminated material to one zone work area.
Avoid the use of an open flame.
Use proper aseptic technique.
Avoid using techniques or procedures that
disrupt the airflow patterns of the station.
If there is a spill or splatter during use, all objects in the station should be surface decontaminated before removal. Thoroughly disinfect the working area of the station WHILE IT IS STILL IN OPERATION. This will prevent the release of contaminants from the station.
Upon completion of work, the station should be allowed to operate for two to three minutes undisturbed, to purge airborne contaminants from the work area.
Objects in contact with contaminated material should be surface decontaminated before removal from the station.
All open trays or containers should be covered before being removed from the station.
24
Product Service 1-800-522-7658
Wipe-Down
Wipe down the interior surfaces of the station with 70% ethanol, or a suitable disinfectant, and allow to dry.
Periodically lift the work surface and wipe down the area beneath it.
Inspect the prefilters located beneath the work surface, beneath the work pan. Replace them if necessary
Dispose of rubber gloves appropriately, and have lab coat laundered properly.
Wash hands and arms thoroughly with germicidal soap.
Shutdown
Turn off the fluorescent light and station blower, close the sashes.
Chapter 5: Using Your PuriCare
Product Service 1-800-522-7658
25
C
HHAAPPTTEERR
C
M P
Now that you have an understanding of how to work in the Vertical Flow Station, we will review the suggested maintenance schedule and the common service operations necessary to maintain your PuriCare for peak performance.
AAIINNTTAAIINNIINNG
M
C
P
!
UURRI
I
AARREE
C
Many of the service operations should be performed only by trained and experienced certification technicians after the station has been properly decontaminated. DO NOT attempt to perform these operations if you are not properly trained. The service operations that require qualified certifiers are preceded by the wrench icon.
6
6
G
Y
Y
OOUURR
26
Routine Maintenance Schedule
Weekly
Product Service 1-800-522-7658
Using 70% ethanol, or a suitable disinfectant, surface disinfect the inside of the station, and the work surface.
Using an appropriate optical glass cleaner and cloth, clean the sash.
Operate the station blowers, noting the pressure readings in an operational log.
Lift the work surface and inspect the prefilters.
Monthly (or more often as required)
Using a damp cloth, clean the exterior surfaces of the station, particularly the front, back and top of the unit, to remove any accumulated dust.
Disinfect and lift the work surface. Surface disinfect the lower plenum with a solution of 70% ethanol, or a suitable disinfectant. Check the prefilters for retained materials.
Replace the prefilters as needed.
All weekly activities.
Annually
Have the station recertified by a qualified certification technician.
All monthly activities.
Biannually
Replace the fluorescent lamp.
Chapter 6: Maintaining Your PuriCare
Service Operations
Work Surface Removal:
1. Loosen and remove the three thumbscrews located at the front of the work surface. Only loosen the three thumbscrews on the back of the work surface.
2. Lift the front edge of the work surface up and pull it out of the station work area.
Prefilter Removal and Cleaning:
The prefilters may be contaminated
!
Product Service 1-800-522-7658
with allergenic material. Take appropriate precautions when removing and handling the prefilters to prevent exposure. The prefilters are disposable and cannot be cleaned.
27
Chapter 6: Maintaining Your PuriCare
Upper Prefilter Removal:
1. Loosen the nuts that secure the two upper prefilter brackets as shown in Figure 6-1.
2. Swing the upper prefilter brackets until the prefilters can be lifted out of the top of the units.
3. Replace the prefilter elements, ensuring the air flow direction arrow on the prefilter frame points down.
Figure 6-1
28
Lower Prefilter Removal:
1. Remove the work surface as described earlier.
2. Loosen the nuts that secure the two lower prefilter brackets as shown in Figure 6-2.
3. Swing the lower prefilter brackets until the prefilters can be lifted out of the unit.
Product Service 1-800-522-7658
Chapter 6: Maintaining Your PuriCare
4. Replace the prefilter elements, ensuring the air flow direction arrow on the prefilter frame points down.
Figure 6-2
Changing the Fluorescent Lamp:
1. Unplug the station.
2. Swing up the sash.
3. Remove the fluorescent lamp by rotating it 90 degrees and pulling it straight down and out of its sockets.
3. Install the new lamp by reversing the removal procedure.
Resetting a Circuit Breaker:
To reset any of the circuit breakers located on the back of the unit near the power cord outlet, depress the white button until it sets.
Product Service 1-800-522-7658
29
Chapter 6: Maintaining Your PuriCare
!
Downflow Velocities
The service operations listed in the rest of this chapter should only be performed by a qualified certifier.
The average downflow velocity for all Vertical Flow Stations should be set at 50-55 FPM.
Downflow Velocity Grid Patterns
The downflow velocity test grid for all Vertical Flow Stations is as follows:
The perimeter of the down test grid is 6.00 inches from the side walls and 6.00 from the rear of the work area. The grid consists of three rows, 5.95 inches apart, from the back to the front, with eight test points on each row, spaced 5.21 inches apart. All points are read at the bottom edge of the sash.
Inflow Volume Calculation -
30
Primary Method
The Inflow Volume can be measured directly, using a Direct Inflow Measurement (DIM) meter, using the following method:
1. Turn the station off and raise the sash.
2. Remove the work surface and the lower prefilters as described earlier.
3. Using rigid cardboard, or fiberboard, cut two patterns as shown in Figure 6-3.
4. Place the solid pattern over the prefilter hole directly above the blower. Seal the pattern to the prefilter shelf with duct tape if necessary.
Product Service 1-800-522-7658
Chapter 6: Maintaining Your PuriCare
5. Place the pattern with the hole over the other prefilter hole. Seal the pattern to the prefilter shelf if necessary.
6. Place the DIM upside down over the hole in the second pattern. Seal the DIM to the pattern with duct tape if necessary.
7. Start the Station and let it run for at least two minutes. Turn on the DIM and take at least 10 readings of the inflow volume. Average the readings.
8. The average inflow volume, corrected for local conditions, should be 650-670 CFM.
Figure 6-3
Inflow Volume Calculation ­Secondary Method
1. Turn the station off and raise the sashes.
2. Remove the work surface and the lower prefilters as described earlier.
3. Using rigid cardboard or fiberboard, cut a 20 inch square.
4. Use the square to cover the prefilter hole directly above the blower.
5. At the other prefilter hole, mark a test grid at 4 ½ inch intervals around the perimeter of the hole, establishing a 16 square grid in the hold.
Product Service 1-800-522-7658
31
Chapter 6: Maintaining Your PuriCare
6. Using a thermal anemometer, measure the velocity at the center of each of the squares.
7. Average the readings. Multiply the average velocity by the correction factor 2.25 to obtain the average inflow volume.
8. The average inflow volume, corrected for local conditions, should be 650-670 CFM.
Speed Control Adjustments
!
1. Locate the two plugged speed control access holes located on the lower front panel as shown in Figure 6-4.
2. Gently pry out the chrome plugs to gain access to the speed control adjustment screws. The screw on the bottom controls the lower Motor/Blower, the one on the top controls the upper impellers.
3. Turn the screw(s) counterclockwise to increase the speed, clockwise to decrease it.
!
4. After adjusting the appropriate speed control, measure the downflow velocities or Inflow Volume as needed.
The Vertical Flow Station blower speeds should only be adjusted by a qualified certification technician. Adjustment of the impeller and/or blower speeds will affect the containment of the unit.
The Vertical Flow Station uses two independent speed controls to control the upper impellers (Supply Air) and the lower Motor/Blower (Exhaust Air)
Turning the speed control adjustment screw all the way counterclockwise will shut the speed control circuit off and the motor(s) will stop. If this occurs, turn the screw clockwise until the motor(s) start running again.
32
Product Service 1-800-522-7658
Chapter 6: Maintaining Your PuriCare
Figure 6-4
Supply HEPA Filter Replacement
The HEPA filters should only be
!
1. Unplug the station.
2. Remove the two upper prefilters as described earlier in this chapter.
3. Remove the five screws that secure the upper rear panel as shown in Figure 6-5.
4. Carefully swing the rear panel out.
serviced by a qualified certification technician. Following replacement of a HEPA filter, the station MUST be recertified by a qualified certification technician.
Product Service 1-800-522-7658
33
Chapter 6: Maintaining Your PuriCare
5. Locate the four locking bolts that secure the upper plenum. They are located in each corner of the plenum, and they thread through the upper bracket and down onto the top of the plenum. Loosen each bolt at least ½ inch.
6. Locate the four plenum lift bolts that raise the upper plenum. They are located in each corner of the plenum, next to the locking bolts, and they thread into the top of the plenum. Tighten each bolt until there is approximately ¼ inch clearance between the top of the HEPA filter and the bottom of the plenum.
7. Slide the Supply HEPA filter straight out of the station.
8. Ensure the replacement HEPA filter matches the performance specifications of the original filter, in particular the efficiency and the flow rate at a given pressure drop.
9. Install the replacement HEPA filter by reversing the disassembly steps.
10. Plug in the station and have it recertified before use.
Figure 6-5
34
Product Service 1-800-522-7658
Chapter 6: Maintaining Your PuriCare
To Remove the Exhaust HEPA Filter:
The exhaust HEPA filter is awkward
!
to handle and heavy. Use appropriate lifting techniques to remove and handle it. Use two people to remove the filter if possible.
Exhaust HEPA Filter Replacement
1. Remove the lower rear panel by removing the #2 Phillips screws on the perimeter of the panel.
2. Remove the work surface and two lower prefilters as described earlier in this chapter.
3. Locate the four locking bolts that secure the lower plenum. They are located in each corner of the plenum and they thread through the lower bracket and down onto the top of the plenum, as shown in Figure 6-6. Loosen each bolt at least ½ inch.
4. Locate the four plenum lift bolts that raise the lower plenum. They are located in each corner of the plenum, next to the locking bolts and they thread into the top of the plenum. Tighten each bolt until there is approximately ¼ inch clearance between the top of the HEPA filter and the bottom of the plenum.
5. Remove the two filter retainer brackets.
6. Slide the Exhaust HEPA filter straight out of the station.
7. Ensure the replacement HEPA filter matches the performance specifications of the original filter, in particular the efficiency and the flow rate at a given pressure drop.
8. Install the replacement HEPA filter by reversing the disassembly steps.
9. Plug in the station and have it recertified before use.
Product Service 1-800-522-7658
35
Chapter 6: Maintaining Your PuriCare
Figure 6-6
Impeller and Motor/Blower Maintenance and Replacement
)
To replace an Impeller:
1. Unplug the unit.
2. Remove the two upper prefilters as described earlier in this chapter.
3. Remove the five screws that secure the upper rear panel as shown in Figure 6-5.
4. Carefully swing the upper rear panel forward.
The motor/blower should be serviced by a qualified certification technician.
The stations blower motor and impeller bearings are sealed and require no lubrication. DO NOT attempt to lubricate them.
36
Product Service 1-800-522-7658
Chapter 6: Maintaining Your PuriCare
5. Locate the four locking bolts that secure the upper plenum. They are located in each corner of the plenum, and they thread through the upper bracket and down onto the top of the plenum. Loosen each bolt as far as it will go.
6. Locate the four plenum lift bolts that raise the upper plenum. They are located in corner of the plenum next to the locking bolts, and they thread into the top of the plenum. Tighten each bolt until the top of the plenum contacts the lift brackets.
7. Slide the Supply HEPA filter straight out of the station.
8. Disconnect the wiring for the impeller to be replaced.
9. Remove the four ¼ inch bolts and nuts that secure the impeller bracket to the plenum. Remove the impeller/bracket assembly.
10. Remove the impeller from the bracket by removing the four 5 mm screws that secure the impeller to the bracket.
11. Install the new impeller by reversing the disassembly steps.
To replace the Lower Motor/Blower:
1. Unplug the station.
2. Remove the lower rear panel.
3. Using a screwdriver, disconnect the motor ground wire from the blower housing.
4. Disconnect the motor leads from the wiring harness by pulling apart each of the 4 connectors.
5. Using a 1/2-inch wrench, remove the two upper and one lower blower mounting bolts, as shown in Figure 6-7.
The motor/blower assembly is heavy.
!
6. While supporting the blower assembly, pull the assembly straight out the plenum.
7. To replace the motor/blower assembly, reverse the above procedure. Before installing the lower rear panel, ensure that the motors rotate in the correct direction when turned on.
Handle with care.
Product Service 1-800-522-7658
37
Chapter 6: Maintaining Your PuriCare
8. Plug in the station and have it recertified before use.
Figure 6-7
Diffuser Removal
Bending or distorting the diffuser may
!
Upper Diffuser Removal
1. Swing the sash up until it is in the locked position.
2. Locate and remove the four acorn nuts that secure the upper diffuser brackets to the station. Support the diffuser when removing the brackets and lower the diffuser straight down.
Lower Diffuser Removal
1. Locate and remove the six screws that secure the lower diffuser to the station, starting with the corner screws. Support the diffuser when removing the last two center screws and lower the diffuser straight down.
affect the airflow in the station. Use care when handling it.
38
Product Service 1-800-522-7658
Chapter 6: Maintaining Your PuriCare
Storage
If the PuriCare is to be left unused for more than one month, the unit should be prepared for storage.
The station should not be stored in
!
1. Lower the sashes completely and seal work area, the bottom edge and the exhaust outlet with plastic sheeting.
2. Unplug the unit.
3. Ensure that the station will not be moved or disturbed while in storage.
areas of excess humidity or temperature extremes. If the station is moved during storage, it must be recertified before use.
Product Service 1-800-522-7658
39
C
HHAAPPTTEERR
C
M V S
Installing a Shelf
1. Ensure that the shelf kit you have is the proper one.
2. Remove the four plastic plugs on the appropriate
3. Lower the shelf brackets on the shelf and align them
4. Secure the shelf to the panel using the four screws
5. Identify which bracket has the trigger to release the
6. Place the dot on the lower panel, 1 ½ inches directly
OODDIIFFYYIINNGG
M
EERRTTIICCAAL
V
TTAATTIIOON
S
Kits for mounting on the operator’s left side are #3794900. Kits for mounting the shelf on the operator’s right side are #3794901.
side of the lower panel, as shown in figure 7-1.
with the holes in the panel.
included in the shelf kit.
shelf from its upright position. Remove the adhesive backing of the white plastic dot included in the shelf kit.
below the bottom of the bracket with the trigger. This will prevent the trigger from scratching the lower panel.
N
L
7
7
F
F
Y
OOUURR
Y
LLOOW
W
40
Product Service 1-800-522-7658
Chapter 7: Modifying Your Open Access Station
Figure 7-1
Product Service 1-800-522-7658
41
C
HHAAPPTTEERR
C
T
RROOUUBBLLEESSHHOOOOTTIINNGG
T
Refer to the following table if your Vertical Flow Station fails to operate properly. If the suggested corrective actions do not solve your problem, contact Labconco for additional assistance.
PROBLEM CAUSE CORRECTIVE ACTION
Station blower and lights won’t turn on
Circuit breakers
Blower(s) won’t turn on but lights work
Blower(s) motor is
Station blower turns on but lights don’t work
Lamp is defective Replace lamp.
Lamp wiring is
Defective lamp
Unit not plugged into outlet
tripped
Blower(s) wiring is disconnected
defective
Lamp(s) not installed correctly
disconnected
ballast
8
8
Plug the PuriCare into appropriate electrical service.
Reset circuit breakers.
Inspect blower(s) wiring.
Replace blower(s) motor.
Inspect lamp installation.
Inspect lamp wiring.
Replace lamp ballast.
42
Product Service 1-800-522-7658
Chapter 8: Troubleshooting
PROBLEM CAUSE CORRECTIVE ACTION
Slight increase in pressure reading
Prefilter loading Check the prefilters for obstruction(s).
Blockage or
Blockage of the
Contamination of work in the station
Restriction of the
External factors are
Station is out of
HEPA filter loading The pressure reading will steadily
increase as the unit is used.
Clean if necessary.
Ensure that the plenum beneath the restriction under the work surface
exhaust outlet
Improper technique or procedure for the Vertical Flow Station
grille – blockage of the exhaust outlet
disrupting the station’s airflow patterns or acting as a source of contamination
adjustment/HEPA filter(s) are defective
work surface is unobstructed.
Ensure that the exhaust outlet is not
blocked or restricted.
See “Use of the Vertical Flow Station”
section in the manual.
Ensure that all return air slots, grilles
and the exhaust outlet are unobstructed.
See “Installation” section of this
manual.
Have station recertified.
Product Service 1-800-522-7658
43
A
PPPPEENNDDIIXX
A
V
EERRTTIICCAALL
V
S
TTAATTIIOONN
S
C
OOMMPPOONNEENNTTSS
C
Illustration A-1 indicates the location of the following service parts:
Vertical Flow Station Replacement Parts
Item Quantity Part No. Description
1 2 3717400 Impeller, 115 Volt
1A 2 3717500 Impeller, 230 Volt
2 1 3803100 Supply HEPA Filter 3 1 1211200 Motor, 115 VAC 4-foot
3A 1 1210502 Motor, 230 VAC 4-foot
4 1 3731101 Motor/Blower Assembly, 115 VAC 4-foot
4A 1 3751901 Motor/Blower Assembly, 230 VAC 4-foot
5 1 3739501 Exhaust HEPA Filter 6 4 3789300 Prefilter 7 2 9721900 Lamp, Fluorescent (not shown) 8 2 1287900 Receptacle Cover (not shown)
A
F
A
LLOOW
F
W
44
Product Service 1-800-522-7658
Appendix A: Purifier Components
A-1
Product Service 1-800-522-7658
45
A
PPPPEENNDDIIXX
A
V
EERRTTIICCAAL
V
S
TTAATTIIOON
S
N
B F
L
F
D
D
B
LLOOW
W
IIMMEENNSSIIOONNSS
B-1
46
Product Service 1-800-522-7658
A
PPPPEENNDDIIXX
A
V
EERRTTIICCAAL
V
S
TTAATTIIOONN
S
S
PPEECCIIFFIICCAATTIIOONNSS
S
Electrical Data
Station Model Electrical Requirements
3830000 to –09 115 VAC – 60 Hz, 1 Phase – 12 Amps
Station Model Electrical Requirements
3830020 to –29 230 VAC – 50/60 Hz, 1 Phase – 7 Amps
Impeller Specifications
Station Model Electrical Requirements
3830000 to –09 115 Volt – 60 Hz, 100 Watt, 2550 RPM
Station Model Electrical Requirements
3830020 to –29
L
C
F
C
F
LLOOW
W
Automatic Thermal Protection
230 Volt – 50/60 Hz, 100 Watt, 2550 RPM Automatic Thermal Protection
Product Service 1-800-522-7658
47
Appendix C: Vertical Flow Station Specifications
Motor Specifications
Station Model Electrical Requirements
3830000 to –09 115 VAC – 60 Hz, 3.8 Full Load Amps
Station Model Electrical Requirements
3830020 to –29 230 VAC – 50/60 Hz, 2.8 Full Load Amps
Environmental Conditions
Indoor use only.
Maximum altitude: 6562 feet (2000 meters).
Ambient temperature range: 41° to 104°F (5° to
40°C).
Maximum relative humidity: 80% for
temperatures up to 88°F (31°C), decreasing linearly to 50% relative humidity at 104°F (40°C).
Main supply voltage fluctuations not to exceed
±10% of the nominal voltage.
Transient overvoltages according to Installation
Categories II (Overvoltage Categories per IEC
1010). Temporary voltage spikes on the AC input line that may be as high as 1500V for 115V models and 2500V for 230V models are allowed.
Used in an environment of Pollution degrees 2
(i.e., where normally only non-conductive atmospheres are present). Occasionally, however, a temporary conductivity caused by condensation must be expected, in accordance with IEC 664.
1/3 H.P. 1625 RPM Automatic Thermal Protection
1/2 H.P. 1625 RPM Automatic Thermal Protection
48
Product Service 1-800-522-7658
Appendix C: Vertical Flow Station Specifications
C-1
Product Service 1-800-522-7658
49
Appendix C: Vertical Flow Station Specifications
C-2
50
Product Service 1-800-522-7658
A
PPPPEENNDDIIXX
A
P
UURRI
P
A
CCCCEESSSSOORRIIEESS
A
Shelf Kit
Shelf kits are available for either the left side (part number 3794900) or the right side (part number
3794901). The shelves are suitable for mounting on either side of the station. The shelves have a capacity of 20 pounds, a trigger to release the shelf, and fold down flat when not in use.
C
I
C
AARREE
D
D
Product Service 1-800-522-7658
51
A
PPPPEENNDDIIXX
A
Q
UUIICCKK
Q
T
HHEE
T
V
EERRTTIICCAAL
V
S
TTAATTOONNSS
S
Supply Air Volume Displacement (CFM) 440
Exhaust Air Volume Displacement (CFM) 650-670
*For further details on the secondary inflow method, consult the inflow velocity calculation – Secondary Method section of Chapter 6: Maintaining Your PuriCare.
Use the downflow velocity test grid as described in this manual when checking the downflow air velocity.
P
Approximate Settings for new Filters
Number of Laskin Nozzles Needed 1
Number of Laskin Nozzles Needed 2
C
C
UURRI
P
Sash Height (inches) 14
Mag Gauge Reading 0.3-0.4 upper / 0.5-0.7 lower
Supply HEPA Filter Dims. (in.) 48x18x3 Exhaust HEPA Filter Dims (in.) 48x18x6
Fluorescent Light(s) 1 each F32T8/TL741
E
HHAARRTT
C
I
C
L
F
E
AARREE
F
LLOOW
F
F
W
OORR
Model 38300
Downflow (FPM) 50-55
52
Product Service 1-800-522-7658
A
PPPPEENNDDIIXX
A
R
EEFFEERREENNCCEESS
R
Many excellent reference texts and booklets are currently available. The following is a brief listing:
Biosafety in Microbiological and Biomedical Laboratories. 4th Edition, May 1999. U.S. Department
of Health, Education, and Welfare, National Institutes of Health, Division of Safety, Bethesda, MD 20892.
Laboratory Safety: Principles and Practices. 2nd Edition 1995. American Society for Microbiology, Washington, D.C.
The Foundations of Laboratory Safety. 1990. Stephen R. Rayburn. Springer-Verlag, New York.
Websites of Interest:
www.absa.org www.cdc.gov www.labconco.com
F
F
Product Service 1-800-522-7658
53
54
Product Service 1-800-522-7658
Loading...