Labconco Protector XStream 11041, Protector XStream 11051, Protector XStream 11081, Protector XStream 11061 User Manual

Original instructions
Protector XStream
Models
11041 Series 11051 Series 11061 Series 11081 Series
Original instructions
Warranty
Labconco Corporation provides a warranty to the original buyer for the repair or replacement of parts and reasonable labor as a result of normal and proper use of the equipment with compatible chemicals. Broken glassware and maintenance items, such as filters, gaskets, light bulbs, finishes and lubrication are not warranted. Excluded from warranty are products with improper installation, erratic electrical or utility supply, unauthorized repair and products used with incompatible chemicals.
The warranty for Protector® XStream® Laboratory Fume Hoods will expire one year from date of installation or two years from date of shipment from Labconco, whichever is sooner. Warranty is non-transferable and only applies to the owner (organization) of record.
Buyer is exclusively responsible for the set-up, installation, verification, decontamination or calibration of equipment. This limited warranty covers parts and labor, but not transportation and insurance charges. If the failure is determined to be covered under this warranty, the dealer or Labconco Corporation will authorize repair or replacement of all defective parts to restore the unit to operation. Repairs may be completed by 3rd party service agents approved by Labconco Corporation. Labconco Corporation reserves the rights to limit this
warranty based on a service agent’s travel, working hours, the site’s entry restrictions and unobstructed access
to serviceable components of the product. Under no circumstances shall Labconco Corporation be liable for indirect, consequential, or special damages
of any kind. This warranty is exclusive and in lieu of all other warranties whether oral, or implied.
Copyright © 2019 Labconco Corporation. The information contained in this manual and the accompanying products are copyrighted and all rights reserved by Labconco Corporation. Labconco Corporation reserves the right to make periodic design changes without obligation to notify any person or entity of such change.
Returned or Damaged Goods
Do not return goods without the prior authorization from Labconco. Unauthorized returns will not be accepted. If your shipment was damaged in transit, you must file a claim directly with the freight carrier. Labconco Corporation and its dealers are not responsible for shipping damages.
The United States Interstate Commerce Commission rules require that claims be filed with the delivery carrier within fifteen (15) days of delivery.
Limitation of Liability
The disposal and/or emission of substances used in connection with this equipment may be governed by various federal, state, or local regulations. All users of this equipment are required to become familiar with any regulations that apply in the user’s area concerning the dumping of waste materials in or upon water, land, or air and to comply with such regulations. Labconco Corporation is held harmless with respect to user’s compliance with such regulations.
Contacting Labconco Corporation
If you have questions that are not addressed in this manual, or if you need technical assistance, contact Labconco’s Customer Service Department or Labconco’s Product Service Department at 1-800-821-5525 or 1-816-333-8811, between the hours of 7:30 a.m. and 5:30 p.m., Central Standard Time.
Part #9410600, Rev. I
ECO L864
Original instructions
TABLE OF CONTENTS
CHAPTER 1: INTRODUCTION 1
About This Manual 2 Typographical Conventions 3
(Conventions Typographiques)
Your Next Step 4
CHAPTER 2: PREREQUISITES 5
Location Requirements 6 Support Requirements 6 Exhaust Requirements 6 Electrical Requirements 7
(Exigences électriques)
Service Line Requirements 8 Space Requirements 8
CHAPTER 3: GETTING STARTED 9
Unpacking Your Fume Hood 10 Removing the Shipping Skid 10 Sash Weight Release 11 Installing the Hood on a Supporting Structure & Work Surface 11 Connecting to the Hood Exhaust System 13 Connecting the Electrical Supply Source to the Protector Fume Hood 14 Connecting the Service Lines to the Protector Fume Hood 16 Sealing the Protector Hood to the Work Surface 17 Certifying the Protector Fume Hood 17
CHAPTER 4: PERFORMANCE FEATURES AND SAFETY
PRECAUTIONS 19
Performance Features 19 Safety Precautions 23
(Consignes de sécurité)
CHAPTER 5: USING YOUR PROTECTOR XSTREAM 27
Operating the Vertical Rising Sash 27 Operating the A-Style Combination Sash 27 Operating the Blower 27 Operating the Lights 28 Working in your Protector Fume Hood 28
CHAPTER 6: MAINTAINING YOUR PROTECTOR XSTREAM 30
Service Safety Precautions 30
(Mesures de sécurité des services)
Routine Maintenance Schedule 31 Routine Service Operations 32
CHAPTER 7: MODIFYING YOUR PROTECTOR XSTREAM 33
Installing Work Surfaces 33 Installing Ceiling Enclosures Above the Fume Hood 34 Installing Rear Panels Behind the Fume Hood 36 Installing Additional Service Fixtures 37 Installing Guardian Airflow Monitors 37 Distillation Grids – Field Installation 37 Installing an Electrical Duplex Outlet 38
(Installation d'une sortie Duplex électrique)
CHAPTER 8: TROUBLESHOOTING 39 APPENDIX A: PROTECTOR XSTREAM COMPONENTS 42
APPENDIX B: PROTECTOR XSTREAM DIMENSIONS 45
APPENDIX C: PROTECTOR XSTREAM SPECIFICATIONS 46
(ANNEXE C: SPÉCIFICATIONS PROTECTEUR DU PREMIER MINISTRE)
APPENDIX D: PROTECTOR I-S HOOD 49
APPENDIX E: SERIAL NUMBER TAG DESCRIPTION 57
(ANNEXE D: DESCRIPTION DE SÉRIE TAG NUMBER)
APPENDIX F: REFERENCES 63
1
Original instructions
Figure 1-1
CHAPTER 1 INTRODUCTION
Congratulations on your purchase of a Labconco Protector® XStream® Laboratory Fume Hood. Your Protector Laboratory Fume Hood is designed to protect you by providing superior containment while conserving energy at OSHA approved “low flow” velocities as low as 50-60 feet per minute, with test limits as low as 40 fpm. It is the result of Labconco’s commitment to developing a high performance fume hood and more than 60 years experience in manufacturing fume hoods.
The Labconco Protector XStream Fume Hood has been engineered to provide maximum containment in a laboratory, and effectively contain toxic, noxious, or other harmful materials when properly installed. The Protector XStream offers many unique features to enhance safety, performance, and energy savings. To take full advantage of them, please acquaint yourself with this manual and keep it handy for future reference. If you are unfamiliar with how high performance fume hoods operate, please review Chapter 4: High Performance Features and Safety Precautions before you begin working in the fume hood. Even if you are an experienced fume hood user, please review Chapter 5: Using Your Protector XStream, which describes your Protector XStream Hood’s features so that you can use the hood efficiently.
Product Service 1-800-522-7658
Chapter 1: Introduction
2
Original instructions
About This Manual
This manual is designed to help you learn how to install, use, and maintain your laboratory fume hood. Instructions for installing optional equipment on your hood are also included.
Chapter 1: Introduction provides a brief overview of the laboratory fume hood, explains the organization of the manual, and defines the typographical conventions used in the manual.
Chapter 2: Prerequisites explains what you need to do to prepare your site before you install your laboratory fume hood. Electrical and service requirements are discussed.
Chapter 3: Getting Started contains the information you need to properly unpack, inspect, install, and certify your laboratory fume hood.
Chapter 4: High Performance Features and Safety Precautions explains how the Protector XStream operates and the appropriate precautions you should take when using the fume hood.
Chapter 5: Using Your Protector XStream discusses the basic operation of your fume hood. Information on how to prepare, use and shut down your Protector XStream Hood are included.
Chapter 6: Maintaining Your Protector XStream explains how to perform routine maintenance on your fume hood.
Chapter 7: Modifying Your Protector XStream explains how to modify the fume hood or add accessories.
Chapter 8: Troubleshooting contains a table of problems you may encounter while using your laboratory fume hood including the probable causes of the problems and suggested corrective actions.
Appendix A: Protector XStream Hood Components contains labeled diagrams of all of the components of the fume hoods.
Appendix B: Protector XStream Hood Dimensions contains comprehensive diagrams showing all of the dimensions for the laboratory fume hoods.
Appendix C: Protector XStream Hood Specifications contains the electrical requirements for laboratory fume hood. Wiring diagrams are also included.
Appendix D: Serial Number Tag Description provides current rating code used on serial number tag
Product Service 1-800-522-7658
Chapter 1: Introduction
3
Original instructions
4'
5'
6'
!
S A 8'
Appendix E: References lists the various resources available that deal with laboratory fume hoods.
Typographical Conventions
Recognizing the following typographical conventions will help you understand and use this manual:
Book, chapter, and section titles are shown in italic type (e.g., Chapter 3:
Getting Started).
Steps required to perform a task are presented in a numbered format. Comments located in the margins provide suggestions, reminders, and
references.
Critical information is presented in boldface type in paragraphs that are
preceded by the exclamation icon. Failure to comply with the information following an exclamation icon may result in injury to the user or permanent damage to fume hood.
Les informations critiques sont présentées en gras dans les paragraphes qui
sont précédés par l'icône d'exclamation. Ne pas se conformer aux informations qui suivent une icône d'exclamation peut résulter à la blessure de l'utilisateur ou à des dommages irréversibles de la hotte aspirante.
Critical information is presented in boldface type in paragraphs that are
preceded by the wrench icon. These operations should only be performed by a trained certifier or contractor. Failure to comply with the information following a wrench icon may result in injury to the user or permanent damage to your hood.
Les informations critiques sont présentées en gras dans les paragraphes qui
sont précédés par l'icône de clé plate. Ces opérations devraient être seulement exécutées par un professionnel agrée. L'échec pour se conformer aux informations qui suivent une icône de clé plate peut résulter à la blessure de l'utilisateur ou à des dommages irréversibles de la hotte.
Important information is presented in capitalized type in paragraphs that are
preceded by the pointer icon. It is imperative that the information contained in these paragraphs be thoroughly read and understood by the user.
A number icon precedes information that is specific to a particular model
of laboratory fume hood. The 4' icon indicates the text is specific to the 4-foot wide model. The 5' icon indicates the text is specific to the 5-foot model, etc.
The S icon indicates the text is specific to the standard model.
The A icon indicates the text is specific to the A-Style Combination Sash
Model.
Product Service 1-800-522-7658
Chapter 1: Introduction
4
Original instructions
!
CAUTION – See Manual. When this symbol is on a fume hood it indicates a
caution that is detailed in this manual.
PRUDENCE – Consulter le Manuel. Quand ce symbole est sur une hotte
aspirante, il indique une prudence qui est détaillée dans ce manuel.
CAUTION – Hot Surface.
AVERTIR – Surface Chaude
CAUTION – See Manual. This symbol on the fume hood indicates the
possibility of a pinch hazard.
PRUDENCE – Consulter le Manuel. Ce symbole sur la hotte indique la
possibilité d'un risque de pincement.
Your Next Step
If your Fume Hood needs to be installed, proceed to Chapter 2: Prerequisites to ensure your installation site meets all of the requirements. Then, go to Chapter 3: Getting Started for instructions on how to install your laboratory fume hood and make all of the necessary connections.
If you would like to review how Labconco’s high performance laboratory fume hoods operate, go to Chapter 4: Performance Features and Safety Precautions.
For information on the operational characteristics of your laboratory fume hood, go to Chapter 5: Using Your Protector XStream.
If your laboratory fume hood is installed and you need to perform routine maintenance on the cabinet, proceed to Chapter 6: Maintaining Your Protector XStream.
For information on making modifications to the configuration of your fume hood, go to Chapter 7: Modifying Your Protector XStream.
Refer to Chapter 8: Troubleshooting if you are experiencing problems with your fume hood.
Product Service 1-800-522-7658
5
Original instructions
CHAPTER 2 PREREQUISITES
Before you install your laboratory fume hood, you need to prepare your site for installation. Carefully examine the location where you intend to install your hood. You must be certain that the area is level and of solid construction. In addition, a dedicated source of electrical power must be located near the installation site.
Carefully read this chapter to learn the requirements for your installation site:
The location requirements. The support requirements. The exhaust requirements. The electrical power requirements. The service line requirements. The space requirements.
Refer to Appendix B: Protector XStream Dimensions for complete fume hood dimensions.
Refer to Appendix C: Protector XStream Specifications for complete laboratory fume hood electrical and environmental conditions, specifications and requirements.
Product Service 1-800-522-7658
Chapter 2: Prerequisites
6
Original instructions
!
!
Location Requirements
The fume hood should be located away from traffic patterns, doors, windows, fans, ventilation registers, and any other air­handling device that could disrupt its airflow patterns. All windows in the room should be closed.
La hotte aspirante devrait être localisé loin des voies de circulation, des portes, des fenêtres, des ventilateurs, des bouches de ventilation, et de tout appareil qui pourrait interrompre ses voies de flux d'air. Toutes les fenêtres dans la pièce devraient être fermées.
Support Requirements
DO NOT install the fume hood on a cart, dolly, or mobile bench. ALL Protector XStream Hood installations must be permanent and stationary. The supporting structure usually consists of a base cabinet and chemically resistant work surface.
NE PAS installer la hotte aspirante sur un chariot ou un banc mobile. TOUTES les installations de la Hotte Protecteur doivent être permanentes et fixes. La structure de soutien
consiste habituellement en un meuble doté d’une surface de
travail chimiquement résistante.
Exhaust Requirements
The exhaust duct connection has been designed for 12" nominal duct (12.75" OD) to allow for minimum static pressure loss while operating at 50 to 100 fpm face velocities. The 12" diameter exhaust duct also allows for proper transport velocities away from the hood with minimal static pressure loss. Should higher transport velocities of 1000 fpm to 2500 fpm be required, simply install a reducer after the exhaust outlet. The exhaust volume and fume hood static pressure loss are listed for each hood model at two sash heights and for face velocities from 40 to 100 fpm. When sizing the exhaust requirements, the choice must be made for air foil, sash height and face velocity.
Product Service 1-800-522-7658
Chapter 2: Prerequisites
7
Original instructions
Projected Airflows and Static Pressure
Eco Energy Saving Air Foil Flush Air Foil
Hood Type Face Velocity (fpm) Airflow Volumetric Rate (CFM) @ Airflow Volumetric Rate (CFM) @
Sash at Sash at 62.5% Static Pressure (inches of water) Static Pressure (inches of water)
Full open open 4' 5' 6' 8' 4' 5' 6' 8'
at 28" at 18" Hood Hood Hood Hood Hood Hood Hood Hood
XSTREAM 100 160 705, 0.26" 930, 0.32" 1150, 0.41" 1600, 0.29" 740, 0.28" 975, 0.34" 1205, 0.44" 1675, 0.31"
80 128 565, 0.17" 745, 0.20" 920, 0.26" 1280, 0.19" 590, 0.18" 780, 0.22" 965, 0.28" 1340, 0.20" 60 96 425, 0.09" 560, 0.12" 690, 0.15" 960, 0.10" 445, 0.10" 585, 0.12" 725, 0.16" 1005, 0.11" 50 80 350, 0.06" 465, 0.08" 575, 0.10" 800, 0.07" 370, 0.07" 490, 0.08" 605, 0.11" 840, 0.08"
40 64 280, 0.04" 370, 0.05" 460, 0.07" 640, 0.05" 295, 0.05" 390, 0.06" 480, 0.07" 670, 0.05"
Eco Energy Saving Air Foil Flush Air Foil
Hood Type Face Velocity (fpm) Airflow Volumetric Rate (CFM) @ Airflow Volumetric Rate (CFM) @
Sash at Sash at Static Pressure (inches of water) Static Pressure (inches of water)
62.50% Full open 4' 5' 6' 8' 4' 5' 6' 8'
open at 18" at 28" Hood Hood Hood Hood Hood Hood Hood Hood
XSTREAM 100 62 440, 0.10" 580, 0.12" 720, 0.16" 1000, 0.11" 475, 0.11" 625, 0.13" 775, 0.17" 1075, 0.12"
80 50 350, 0.06" 465, 0.08" 575, 0.10" 800, 0.07" 380, 0.07" 500, 0.08" 620, 0.11" 860, 0.08" 60 38 265, 0.04" 350, 0.05" 430, 0.06" 600, 0.04" 285, 0.04" 375, 0.05" 465, 0.06" 645, 0.04"
Proper blower selection can be determined from these exhaust requirements and the total system static pressure loss. Contact Labconco Customer Service for assistance in sizing a remote blower system.
Electrical Requirements
The Protector XStream Hood models feature internal wiring for the LED light assembly and light switch. All internal wiring is terminated at the single point wiring junction box for hook-up by a qualified electrician. The blower switch, and light switch wires are also terminated at the single point wiring junction box for hook-up by a qualified electrician. Refer to Chapter 3: Getting Started and Appendix C: Protector XStream Specifications for the wiring diagram for proper electrical installation.
Les modèles Protector XStream Hood sont dotés d'un câblage interne pour l'éclairage LED et l'interrupteur. Tout le câblage interne est terminé à la boîte de jonction point de câblage unique pour le raccordement par un électricien qualifié. Le bouton du ventilateur, et fils de l'interrupteur de lumière sont également mis fin à la boîte de jonction point de câblage unique pour le raccordement par un électricien qualifié. Reportez-vous au Chapitre 3: Mise en route et à l'Annexe C: Spécifications Premier Protector pour le schéma de câblage pour l'installation électrique correcte.
Product Service 1-800-522-7658
Chapter 2: Prerequisites
8
Original instructions
Service Line Requirements
All service lines to the laboratory fume hood should be ¼ inch outside diameter, copper (brass for natural gas), and equipped with an easily accessible shut-off valve, should disconnection be required. Recommended operating pressure is 40 PSI, with a maximum allowable pressure of 200 PSI. Consider a pressure regulator to reduce line pressure to 40 PSI. Please check with local codes for other requirements.
Space Requirements
The dimensions for the different models are shown in Appendix B: Protector XStream Dimensions.
Product Service 1-800-522-7658
9
Original instructions
!
CHAPTER 3 GETTING STARTED
Now that the site for your laboratory fume hood is properly prepared, you are ready to unpack, inspect, install, and certify your unit. Read this chapter to learn how to:
Unpack and move your Protector Hood. Set up the fume hood with the supporting structure and work surface. Connect to an exhaust system. Connect the electrical supply source. Connect the service lines. Seal the Protector Hood to the work surface. Arrange certification of your Protector Hood.
Depending upon which model you are installing, you may need common plumbing and electrical installation tools in addition to 5/16", 3/8", 7/16", and 1/2" wrenches, ratchets, sockets, a nut driver set, a flat-blade screwdriver, a Phillips screwdriver, and a carpenter level to complete the instructions in the chapter.
The Protector XStream Hood models weigh between 400 to 800 lbs. (182-363 kg). The shipping skid allows for lifting with a mechanical lift truck or floor jack. If you must lift the fume hood manually, follow safe-lifting guidelines. Normally, the fume hood can be slid off a hydraulic lift table and be placed
into position on top of the work surface. Do not lift by the
front air foil.
Les XStream modèles de la Hotte Protecteur pèsent entre 400 à 800 livres. (182-363 Kg). La palette bois d’envoi permet le
soulèvement par un camion muni d’un élévateur mécanique ou
par un cric rouleur. Si vous devez soulever manuellement la hotte aspirante, respectez les règles de sécurité du soulèvement. Normalement, la hotte aspirante peut être glissée d'une table munie d’un élévateur hydraulique et être placée en position sur
Product Service 1-800-522-7658
Chapter 3: Getting Started
10
Original instructions
The United States Interstate Commerce Commission rules require that claims be filed with the delivery carrier within fifteen (15) days of delivery.
!
la surface de travail. Ne pas soulever par l’écoulement d'air du devant.
Unpacking Your Laboratory Fume Hood
Carefully remove the shrink-wrap or carton on your fume hood and inspect it for damage that may have occurred in transit. If your unit is damaged, notify the delivery carrier immediately and retain the entire shipment intact for inspection by the carrier.
DO NOT RETURN GOODS WITHOUT THE PRIOR AUTHORIZATION OF LABCONCO. UNAUTHORIZED RETURNS WILL NOT BE ACCEPTED.
IF YOUR HOOD WAS DAMAGED IN TRANSIT, YOU MUST FILE A CLAIM DIRECTLY WITH THE FREIGHT CARRIER. LABCONCO CORPORATION AND ITS DEALERS ARE NOT RESPONSIBLE FOR SHIPPING DAMAGES.
Do not discard the shipping skid or packing material for your fume hood until you have checked all of the components and installed and tested the unit. The
XStream fume hood baffles are shipped loose behind the hood and do not discard. The lower baffle is 12.9" tall and the middle baffle is 27.4" tall. See Figure 1-1, Figure 3-1 and Figure B-1 that shows proper XStream baffle installation. Do not remove the fume hood from its shipping skid until it is ready
to be placed into its final location. Move the unit by placing a flat, low dolly under the shipping skid, or by using a floor jack.
Do not move the hood by tilting it onto a hand truck.
Ne pas déplacer la hotte en le penchant sur un diable.
Removing the Shipping Skid
LEAVE THE FUME HOOD ATTACHED TO ITS SHIPPING SKID UNTIL IT IS AS CLOSE TO ITS FINAL LOCATION AS POSSIBLE. MOVE THE HOOD BY USING A SUITABLE FLOOR JACK, OR BY PLACING A FURNITURE DOLLY UNDERNEATH THE SKID. DO NOT MOVE THE HOOD BY TILTING IT ONTO A HAND TRUCK.
After you verify the fume hood components, move your hood to the location where you want to install it. Follow the steps listed next to remove the shipping skid from your unit.
1. Remove the side panels by unscrewing the concealed screw. Then lift
off the side panels.
Product Service 1-800-522-7658
Chapter 3: Getting Started
11
Original instructions
!
!
2. Find the hardware (lag screws, etc.) that attach the fume hood to the
skid and remove the hardware. Some hardware is on the sides and some is on the back.
Sash Weight Release
To protect the fume hood from damage in shipment, the sash weight has been secured to the back of the fume hood with screws. Simply remove the screws and make sure the sash cables or chains are on the pulleys or sprockets before operating the sash. On models with more than one sash, the sash weights have been secured to the shipping skid with lag screws. Remove the weights from the skid and attach them to the respective sash cables or chains using the threaded connectors provided.
NOTE: THE SASH WEIGHT ITSELF WAS INDIVIDUALLY MATCHED FOR THIS SPECIFIC HOOD AND SHOULD NOT BE EXCHANGED ON ANY OTHER UNIT.
Installing the Hood on a Supporting
Structure and Work Surface
The Protector Hood is heavy! Use caution when lifting or moving the unit.
La Hotte Protecteur est lourd ! Prudence en soulevant ou en déplaçant l'objet.
When installing the Protector XStream Fume Hood onto a chemically-resistant work surface or benchtop, ensure that the structure can safely support the combined weight of the fume hood and any related equipment. The work surface should be at least as wide as the hood to properly support it. The work surface is aligned flush
with the back of the fume hood for good airflow: this will provide the correct spacing under the air foil for proper bypass airflow. The Protector XStream with
Eco-Foil performs best with a work surface having a large 1.0" leading radius to promote aerodynamic airflow at low velocities. The high performance XStream work surface should have a 36" deep dimension. The lower base cabinets are placed flush with the front of the work surface as shown in Figure 3-1.
WARNING: It is important to: 1). Support the rear of the work surface and fume hood. 2). The cross support provides support for the bottom of the work surface. 3). Install the cross support after the base cabinets and work surface are leveled and before installing the hood.
AVERTISSEMENT : Il est important de soutenir l'arrière de la surface de travail et la hotte aspirante. Le support
Product Service 1-800-522-7658
Chapter 3: Getting Started
12
Original instructions
Place Cross Support Here.
37.7
1.7
9.34
tranversal soutient le bas de la surface de travail. Installer le support transversal après que les meubles et la surface de travail soient nivelés et avant d'installer la hotte.
The following are instructions for mounting a cross support:
1. Level the base cabinets and the work surface. Work surface should be
placed flush with the back of the fume hood as shown in Figure 3-1.
2. Scribe a line on the wall or back of the base cabinet to locate the
support under the work surface.
3. Mount the support by attaching it to the wall or base cabinet.
4. Place the hood on top of the work surface and cross support.
The work surface should be smooth and durable, such as a chemical-resistant epoxy resin. The surface should be nonporous and resistant to the acids, solvents, and chemicals used in conjunction with the Protector XStream Fume Hood. The work surface should also contain a dished recessed area for containing primary spills.
Figure 3-1
Product Service 1-800-522-7658
Chapter 3: Getting Started
13
Original instructions
!
!
Connecting to the Hood Exhaust System
WARNING: The weight of the exhaust ductwork system must be supported independently of the hood superstructure. Do not allow this weight to be supported by the hood structure as damage to the hood may occur.
AVERTISSEMENT : Le poids du système d'aspiration de ductwork doit être soutenu d'une manière indépendante de la superstructure de la hotte. Au cas où ce poids est supporté par la structure de la hotte, des dommages à la hotte peuvent arriver.
The exhaust connection should be installed by a qualified HVAC contractor. The exhaust connection on your hood has been designed for 12" nominal pipe (12.75" OD) to allow for minimum static pressure loss with proper transport velocities away from the hood. Should higher transport velocities be required, simply install a reducer after the hood exhaust outlet. Consult Labconco Customer Service should you require help sizing your blower for the exhaust volume and total system static pressure loss.
La connexion d'échappement devrait être installée par un professionnel de CVC agrée. La connexion d'aspiration sur votre hotte a été conçue pour un tuyau nominal de 12 pouces
(12,75 pouces de diamètre externe) afin d’avoir une perte
minimale de pression statique avec les correctes flux de transport loin de la hotte. Consulter le Service Clientèle de Labconco si la calibration de votre soufflerie pour le volume d'aspiration et la perte de pression statique du système le requiert.
The selected exhaust duct material should match the hood procedures and chemicals used to ensure compatibility.
Le matériel de conduit d'aspiration choisi devrait être en conformité avec les procédures de la hotte et les produits chimiques qui sont utilisés pour garantir la compatibilité.
Product Service 1-800-522-7658
Chapter 3: Getting Started
14
Original instructions
!
Connecting the Electrical Supply Source
to the Protector Fume Hood
Prior to connecting any electrical wiring to the fume hood structure, refer to the hood identification plate for the proper electrical requirements of your specific model.
WARNING: The building electrical supply system for Protector XStream Hoods should include overload protection. A switch or circuit breaker should be in close proximity to the equipment and within easy reach of the operator. The switch or circuit breaker is to be marked as the disconnecting device for the equipment. Consult the NEC-2002 for proper installation.
AVERTISSEMENT : Le système d’alimentation électrique de la
Hotte Protecteur doit inclure la protection contre la surcharge. Un commutateur ou disjoncteur doit être tout près de l'équipement et à portée facile de l'opérateur. Le commutateur ou le disjoncteur doit être marqué comme l'appareil débranchant pour l'équipement. Consulter le NEC-2002 pour une installation correcte.
The identification plate, model number, serial number, and electrical connection boxes are accessible from the front of the fume hood by removing the front panel.
The Protector XStream Hood is wired for 115 Volt, 50/60 Hz, 20 Amp or 230Volt, 50/60 Hz, 10 Amp electrical service. Check the I.D. plate behind the front panel for voltage verification. The number of circuits varies depending on the model. All of the electrical connections are terminated at the field wiring terminal box for hook-up by a qualified electrician. We recommend each circuit be a dedicated branch circuit. However, if wired together the maximum load allowed is the sum of individual outlets plus the rating of the unit (i.e. 2 Amps). The single point internal junction box is used for the connection of the lights, blower, and duplex outlets. Refer to the wiring diagram for your Protector XStream in Appendix C: Protector XStream Fume Hood Specifications.
The fume hood is required to be grounded to the MAINS protective earthing ground for safe operation. Using a ring terminal sized for a 10-24 machine screw, connect the MAINS ground conductor to the grounding lug marked with the protective
earthing symbol, . Only MAINS ground conductors should be connected to the protective earthing ground lug, no other conductors should be connected to this grounding lug. Using wire nuts connect the MAINS supply conductors to the fume hood supply wires. Insure that the wires are connected as per the appropriate wire color codes for the input voltage. For 115V Phase (Hot) is black and Neutral is white, for 230V Phase1 is brown and Phase2 is blue. Refer to the wiring diagram for your Protector XStream in Appendix C: Protector XStream Specifications.
Product Service 1-800-522-7658
15
Original instructions
Internal Junction Box
Primary I.D. Plate
Front Panel
Front Panel
Secondary I.D. Label
Box Assembly,
and loose parts
Located behind
Chapter 3: Getting Started
holds manual
Figure 3-2
All wiring for the fume hood SHOULD be performed by a licensed electrician and conform to all local codes. In most cases, the hood will require the use of shielded conduit to protect the wiring into the hood. The grounding connection shall not be made to the terminal box cover.
Tout le câblage électrique pour la hotte aspirante devrait être exécuté par un électricien agrée et être conforme à tous les règles en vigueur. Dans la plupart des cas, la hotte exigera l'usage de conduit blindé pour protéger le câblage électrique dans la hotte. La prise de terre ne sera pas faite à la couverture de la boîte du terminal.
Product Service 1-800-522-7658
Chapter 3: Getting Started
16
Original instructions
Air
Hot Water
Vacuum
Cold Water
Natural Gas – See Caution Below
!
!
!
The LED light has been mounted outside the top liner panel and is sealed from vapors inside the hood structure. To change the LED light bulbs in your hood, you must first remove the front panel from the hood. Next remove the knock out plugs holding the light fixture in place. Lift fixture up and replace any defective bulbs. Reverse order to reassemble.
Connecting the Service Lines to the
Protector Fume Hood
The hoods with service fixtures have been plumbed from the valve to the hose connector or gooseneck for your installation convenience. A qualified installer shall provide the tubing. Tubing can enter the hood from above, through the back, or through the work surface to make these connections to the service fixtures.
NOTE: Inspect all fittings for leakage. Tighten the fittings slightly if needed.
NOTE : Inspecter toutes les installations à la recherche de fuite. Resserrer les installations légèrement si nécessaires.
CAUTION: Do not use oxygen with any standard service fixture. Contact Labconco Customer Service for oxygen fixture information.
PRUDENCE : Ne pas utiliser de l'oxygène avec l'accessoire de service standard. Contacter le Service Clientèle de Labconco pour les informations d'accessoire d'oxygène.
Should access to the hood plumbing fixture bodies be required, remove the service access plate on the hood front corner posts by loosening their individual screws (see item 11, Figure A-1 in Appendix A). The valve body will now be fully exposed for any service work that may be necessary. The service fixtures supplied on your laboratory hood are designed for use with the following services:
WARNING: Contact Labconco Customer Service directly before using any service other than those listed above in these valves to assure full compatibility.
AVERTISSEMENT : Contacter le Service Clientèle de Labconco directement avant d'utiliser n'importe quel service autre que ceux énumérés au-dessus dans ces soupapes pour assurer une pleine compatibilité.
Product Service 1-800-522-7658
Chapter 3: Getting Started
17
Original instructions
!
CAUTION: Natural gas should be used only in the service fixture that has been pre-plumbed with brass tubing. Sulfur content of the gas could cause deterioration of standard copper supply lines.
PRUDENCE : Le gaz naturel devrait être seulement utilisé dans l'accessoire de service qui a été pré soudé avec des tuyaux de cuivre. Le contenu soufré du gaz pourrait causer la détérioration des lignes d’alimentation en cuivre standard.
Sealing the Protector Hood to the Work
Surface
When the hood has been set in place, ducted, wired, and plumbed, it should be sealed at the work surface to prevent spilled materials from collecting under the walls of the hood. Materials such as silicone sealants are recommended to seal the hood structure.
Certifying the Protector Fume Hood
The combination of your laboratory hood, exhaust ductwork, and exhaust blower gives you the flexibility to change the airflow at the sash opening of your hood. To determine the actual face velocity at the sash opening, airflow velocity readings will need to be taken. This should be done across the sash opening of the hood in accordance with the Industrial Ventilation Manual section on laboratory hoods (see Appendix E: References). Labconco recommends an average face velocity at the sash opening of 50 to 100 feet per minute for Protector XStream high performance fume hoods. Consult Chapter 2 for proper airflow volumes for your particular model. To ensure the performance at 50-60 feet per minute, Labconco researchers successfully challenged the Protector XStream at face velocities lower than 50 feet per minute, under various adverse conditions.
Your Protector Fume Hood has been tested at the factory per ASHRAE 110-1995. All hoods achieve an “as manufactured rating” of less than 0.05 part per million (ppm) at 4 liters per minute (lpm); AM <0.05 (consult Labconco for individual fume hood ratings). For “field use” ASHRAE testing contact Labconco Sales Engineering Team or Customer Service for a certified on-site contractor.
Product Service 1-800-522-7658
Chapter 3: Getting Started
18
Original instructions
!
NOTE: Face velocity profiles and smoke testing should be done periodically to ensure safe performance.
NOTE : Les profils de flux frontal et les tests de fumée devraient être régulièrement faits pour garantir une utilisation en toute securité.
Product Service 1-800-522-7658
19
Original instructions
CHAPTER 4 PERFORMANCE FEATURES AND SAFETY PRECAUTIONS
Performance Features
The Protector XStream High Performance Laboratory Hood is designed to meet the needs of the laboratory scientist, and provide superior containment while conserving energy at OSHA approved “low flow” velocities as low as 50-60 feet per minute. The Protector XStream laboratory fume hood has been designed to effectively contain toxic, noxious, or other harmful materials when properly installed. What makes the Protector XStream so unique is the revolutionary way it directs air into and through the contaminated air chamber. Labconco engineered the Protector XStream to minimize the effects of turbulence. The innovative and aerodynamic designs of the sash handle, air foil, upper dilution air supply, and rear downflow baffle all work in concert to produce horizontal airflow patterns that significantly reduce chemical concentrations through the
work area. Depending on sash position, air turbulence, vortexing and “the roll”
frequently observed during fume hood smoke tests are virtually eliminated by the Protector XStream high performance system. Optional A-Style Combination Sash models are also available upon request. The hood features a by-pass airflow design that promotes full containment as the sash is moved. Airflow is diverted behind the front panel and under the air foil to help control fluctuations in face velocity, which occur as the sash is moved.
1. Unique sash provides maximum visibility of 37.5" high while
conserving energy by limiting sash travel to 28". Vertical-rising sash may be raised from a closed to 28" operating height. Exhaust volume, and blower sizing is based on the 18" height to maximize energy usage. Sash stops are included to limit sash height to 18" and reduce energy usage.
Product Service 1-800-522-7658
Chapter 4: Performance Features & Safety Precautions
20
Original instructions
2. Containment-Enhancing Sash Handle includes a perforated air passage
directly atop the handle to bleed air into the hood chamber and direct chemical fume concentrations away from the user breathing zone. The large radiused handle sweeps airflow into the hood with minimal turbulence.
3. Large usable interior work depth and interior height of 48" provides
ample working space and directs contaminants away from the operator.
4. Opti-Zone™ Rear Downflow Dual Baffle System directs horizontal
streams of airflow to the rear slots of the primary baffle in a single pass. Baffles are removable for cleaning. The secondary baffle, located between the primary baffle and the back wall, counteracts the upward air streams that create roll in traditional hoods.
5. Exterior access cover plates are removable for easy access to plumbing
valves and sash adjustment hardware when servicing through the sides is not possible.
6. Lift-Away™ front panel provides easy access to electrical wiring, sash
weights, and lighting fixtures.
7. Energy efficient LED lighting is located behind a laminated safety glass shield
mounted to the top of the hood. The factory-wired instant start T8 lighting is serviceable from outside the hood cavity. Additionally, the long lasting 50,000 hour direct wired LED T8 bulbs are more energy efficient, utilizing approximately ½ the power of fluorescent bulbs. See Appendix A for wattage.
8. Low mounted, factory-wired light and blower switches are ADA compliant.
9. Eco Energy Conservation Air Foil and Flush Foil allow air to sweep the
work surface for maximum containment. The perforated Clean-Sweep™ openings create a constant protective barrier from contaminants. In addition, should the operator inadvertently block the airflow entering, the air continues to enter from under the air foil and through the Clean-Sweep openings. See Appendix A for alternate PVC Eco-Air Foil as PVC material resists corrosion from mineral acids.
10. Streamlined corner posts provide maximum visibility and the flexibility
to add services after installation.
11. All hoods are factory prepared for up to 8 service fixtures.
12. Duplex electrical receptacles are mounted on the right and left corner
posts as requested. Receptacles are factory-wired to hood single point junction box.
13. Upper Dilution Air Supply (not shown) provides bypass air from above
the work area. This feature constantly bathes the sash interior with clean air and reduces chemical fumes along the sash plane, near the critical breathing zone. The upper dilution air supply perforations sweep the upper interior to reduce stagnant pockets of air in the upper interior.
14. Shipped fully assembled and eliminates the need for costly onsite
assembly. If needed, the fume hood easily breaks down for transport through doorways.
Product Service 1-800-522-7658
Chapter 4: Performance Features & Safety Precautions
21
Original instructions
15. Accessory Guardian™ Digital Airflow Monitor or Guardian Airflow
Monitor continuously monitors face velocity. An audio/visual alarm
alerts the user to low airflow conditions. The right corner post is factory prepared to accommodate the Guardian Monitor (sold separately).
16. Optional Sash Models. Sashes are offered in smooth anti-racking cable,
chain, auto-return, or Intelli-Sash.
17. Frame of Epoxy-Coated Steel and Aluminum is durable and corrosion
resistant.
18. Exhaust Connection. The hood features 12" (12.75" OD pipe) exhaust
connections sized to allow for a minimum static pressure loss through the hood structure while providing a good transport velocity through the exhaust system.
19. Spillstopper™ Solid Epoxy Work Surface is dished to contain spills.
Hoods with Eco-Foil utilize a large 1.0" radiused leading edge to promote aerodynamic airflow. (Work surface is sold separately).
20. Optional Ceiling Enclosure Kits are available for a decorative facade
between the hood and the ceiling.
21. Sash Stops located 18" off the work surface provide a means of
controlling the operating height of the sash, and further reducing energy requirements if so desired.
Product Service 1-800-522-7658
Chapter 4: Performance Features & Safety Precautions
22
Original instructions
23
15
18
7
4
19
10
8
9 2 12
11
5
21
1 6 17
16
22
24
22
Figure 4-1
Product Service 1-800-522-7658
Chapter 4: Performance Features & Safety Precautions
23
Original instructions
!
See Diagram on Page 22.
22. CAUTION – Current rating of receptacle is specified in Appendix D.
AVERTIR – Classification des prises de courant est spécifié dans l’Annexe D.
23. CAUTION – See Appendix C and D for complete current rating.
AVERTIR – Voir Annexe C et D pour la classification complete du courant.
24. CAUTION – Flammable Gas.
AVERTIR – Gaz Inflammable.
Safety Precautions
Although the laboratory hood has been engineered to maintain optimum operator safety, caution should always be used while working in the hood. Prior to using the hood, check to make sure that the exhaust blower is operating and that air is entering the hood at its specified face velocity.
Bien que la hotte de laboratoire ait été réglée pour maintenir la sécurité optimale de l’opérateur, la prudence devrait toujours être utilisée en travaillant sous la hotte. Avant utiliser la hotte, le contrôle pour s'assurer que la soufflerie d'aspiration fonctionne et que cet air entre dans la hotte au flux spécifié.
USE GOOD HOUSEKEEPING IN THE HOOD AT ALL TIMES. CLEAN UP SPILLS IMMEDIATELY WITH A MILD DETERGENT. PERIODICALLY CLEAN HOOD INTERIOR, INCLUDING LED LIGHT GLASS PANEL. REPLACE BURNED OUT LIGHT BULBS TO MAINTAIN MAXIMUM ILLUMINATION.
DO NOT OVERLOAD THE WORK SURFACE WITH APPARATUS OR WORK MATERIAL. THE SAFE OPERATION OF THE LABORATORY HOOD IS BASED UPON HAVING PROPER AIRFLOW THROUGH THE STRUCTURE. DO NOT PLACE LARGE, BULKY OBJECTS SUCH AS BLOCK HEATERS, DIRECTLY ON THE HOOD WORK SURFACE. INSTEAD, ELEVATE THE OBJECT 2" TO 3" ON BLOCKS TO ALLOW A FLOW OF AIR UNDER THE OBJECT AND INTO THE LOWER REAR BAFFLE EXHAUST SLOT. ENSURE BLOCKS ARE LEVEL AND SECURED IN PLACE.
Product Service 1-800-522-7658
Chapter 4: Performance Features & Safety Precautions
24
Original instructions
!
Blocking the bottom of the baffle at rear of hood will change the airflow pattern in the hood causing turbulence and possible
leakage at the face of the hood. (Don’t store containers or
supplies against baffles, as this will affect airflow through the hood).
Avoid placing your head inside hood. Keep hands out of hood as much as possible.
Bloquer le fond du déflecteur à l'arrière de la hotte changera le modèle du flux d'air dans la hotte causant de la turbulence et une fuite possible devant la hotte. (Ne pas emmagasiner des récipients ou des provisions contre les déflecteurs, car ceci affectera le flux d'air à travers la hotte).
Eviter de placer votre tête à l’intérieur de la hotte. Garder les
mains à l’extérieur de la hotte le plus possible.
Always work as far back in hood as possible. It is best to keep all chemicals and apparatus 6" inside the front of the hood.
Toujours travailler aussi loin que possible de la hotte. Il est recommandé de garder tous les produits chimiques et appareils à 6 pouces à l’intérieur de l’avant de la hotte.
This hood does not feature explosion-proof electrical components, unless ordered separately. Therefore, use of flammable or explosive materials in quantities above the explosive limit are not recommended.
Cette hotte ne possède pas de composants électriques anti­explosion, à moins que commandé séparément. Donc, l'usage de matériels inflammables ou explosifs dans les quantités au­dessus de la limite explosive n'est pas recommandé.
Do not work with chemicals in this hood without the exhaust system running. Do not store chemicals in a fume hood.
Ne pas travailler avec les produits chimiques sous cette hotte
sans le système de d’aspiration en marche. Ne pas stocker des
produits chimiques sous une hotte aspirante.
Perchloric acid use in this hood is prohibited.
L'usage d’acide perchlorique sous cette hotte est interdit.
Product Service 1-800-522-7658
Chapter 4: Performance Features & Safety Precautions
25
Original instructions
!
High level radioisotope materials are prohibited for usage in this hood.
Les matériels d'isotope radioactif de haut niveau sont interdits à l'usage sous cette hotte.
AVOID CROSS DRAFTS AND LIMIT TRAFFIC IN FRONT OF THE HOOD. AIR DISTURBANCES CREATED MAY DRAW FUMES OUT OF THE HOOD.
The use of heat-generating equipment in this hood without the exhaust system operating properly can cause damage to the hood.
L'usage d'équipement chaleur-produiant dans ce capuchon sans l'opération de système d'aspiration peut causer convenablement des dommages à la hotte.
The Laboratory Hood should be certified by a qualified certification technician before it is initially used. The hood should be re-certified whenever it is relocated, serviced or at least annually thereafter.
La Hotte de Laboratoire Protecteur devrait être certifié par un technicien de certification qualifié avant qu'elle soit utilisée au début. La hotte devrait être re-certifiée quand elle est réinstallée, entretenue ou du moins annuellement par la suite.
Ensure that the unit is connected to electrical service in accordance with local and national electrical codes. Failure to do so may create a fire or electrical hazard. Do not remove or service any electrical components without first disconnecting the hood from electrical service.
Garantir que l'unité est connectée au service électrique conformément aux règles électriques en vigueur. Le non­respect peut créer un feu ou un risque d'origine électrique. Ne pas enlever ou entretenir des composants électriques sans débrancher premièrement l’alimentation électrique de la hotte.
Proper operation of the fume hood depends largely upon the hood’s location and the operator’s work habits. Consult the Reference Manual in Appendix E.
Le correct fonctionnement de la hotte d'aspiration dépend principalement de son emplacement et les habitudes de travail
Product Service 1-800-522-7658
Chapter 4: Performance Features & Safety Precautions
26
Original instructions
!
!
de l'opérateur. Consulter le Manuel de Référence dans l'Annexe E.
If the unit is not operated as specified in this manual it may impair the protection provided by the unit.
Si l'unité n'est pas utilisée comme spécifié dans ce manuel il peut diminuer la protection fournie par l'unité.
Do not touch the blower motor. The surfaces of the motor can become hot and could cause burns.
Ne pas toucher le moteur de soufflerie. Les surfaces du moteur peuvent devenir chaudes et pourrait causer des brûlures.
Do not position the fume hood so that it is difficult to operate the main disconnect device.
Ne pas positionner la hotte de sorte qu'il est difficile de faire fonctionner le dispositif principal de déconnexion.
To prevent the possibility of minor injury keep hands and fingers clear of sprockets at the top of the four corners.
Pour éviter les risques de blessures mineures garder les mains et les doigts de pignons en haut de la quatre coins.
Product Service 1-800-522-7658
27
Original instructions
S
S
A
A
CHAPTER 5
USING YOUR
PROTECTOR XSTREAM
Operating the Vertical-Rising Sash
Because of the Protector XStream Hood counterbalanced sash mechanism, it will take only a few pounds of force to move the sash up or down, and you can operate the sash smoothly with one or two hands positioned any where along the handle. The vertical-rising sash may be raised to a maximum 28" operating height. The airflow requirements should be sized for the 28" operating height; if using sash stops provided at 18", then the airflow requirements can be reduced by approximately 40%. See Chapter 2 for airflows. Optional models with auto­return to 18" or electric Intelli-Sash are available.
Operating the A-Style Combination Sash
Some other hood models have sashes called A-Style Combination Sashes in place of vertical-rising sashes. These combination sashes allow the operator to use the hood with sashes either half open horizontally or vertically to conserve energy. The horizontal sashes are used in normal operating mode. Optional sets of sash stops can be installed to prevent raising the vertical sash above the half-open and fully closed positions unless manually defeated by the operator. The airflow requirements are sized for the 50% open sash condition.
Operating the Blower
Your Protector XStream Fume Hood utilizes a remote style blower, which can be
activated by turning the blower switch to “ON.” You can validate the hood
performance by watching smoke drawn into the hood face opening.
Product Service 1-800-522-7658
Chapter 5: Using Your Protector XStream
28
Original instructions
S
A
Operating the Lights
Your Protector XStream Fume Hood utilizes a factory-wired LED light to illuminate the hood interior (see Appendix A). Simply turn the light switch to
“ON” to operate.
Working in your Protector Fume Hood
Planning
Thoroughly understand procedures and equipment required before
beginning work.
Arrange for minimal disruptions, such as room traffic or entry into the
room while the hood is in use.
Start-up
Turn on light and hood blower. Slowly raise the sash. Check the baffle air slots for obstructions. Allow the hood to operate unobstructed for 5 minutes. Wear a long sleeved lab coat and rubber gloves. Use protective eyewear.
Wear a protective mask if appropriate.
Loading Materials and Equipment
Only load the materials required for the procedure. Do not overload the
hood.
Do not obstruct the front air foil, or rear baffle slots. Large objects should not be placed close together and spaced above the
work surface to permit airflow to sweep under the equipment.
After loading the hood, wait one minute to purge airborne contaminants
from the work area.
Work Techniques
Keep all materials at least 6 inches inside of the sash, and perform all
contaminated operations as far to the rear of the work area as possible.
Segregate all clean and contaminated materials in the work area. Avoid using techniques or procedures that disrupt the airflow patterns of
the hood.
Final Purging
Upon completion of work, the hood should be allowed to operate for two
to three minutes undisturbed, to purge airborne contaminants from the work area before shutting down blower.
Product Service 1-800-522-7658
Chapter 5: Using Your Protector XStream
29
Original instructions
Unloading Materials and Equipment
Objects in contact with contaminated material should be surface
decontaminated before removal from the hood.
All open trays or containers should be covered before being removed from
the hood.
Shutdown
Turn off the light and hood blower, then close the sash.
Product Service 1-800-522-7658
30
Original instructions
CHAPTER 6 MAINTAINING YOUR PROTECTOR XSTREAM
Now that you have an understanding of how to work in the fume hood, we will review the suggested maintenance schedule and the common service operations necessary to maintain your fume hood for peak performance.
Service Safety Precautions
If performing any electrical maintenance, always disconnect the power at
the main disconnect.
If performing decontamination inside the fume hood, consult your safety
officer for proper personal protective equipment and procedure.
Since some service operations require a step ladder, always use proper
safety and consult your safety officer
If performing maintenance on any service lines, always shut off the supply
first.
Some removable components may be heavy, follow safe-lifting guidelines. Verify all components are installed correctly with performance verified
before conducting normal operations.
Précautions de Sécurité pour l’Entretien
Lors de l'entretien électrique, toujours débrancher le courant du secteur
principal.
Durant la décontamination sous la hotte d'aspiration, consulter votre
responsable de sécurité pour le correct équipement de protection du personnel et la procédure.
Product Service 1-800-522-7658
Chapter 6: Maintaining Your Protector XStream
31
Original instructions
!
Puisque certaines opérations d’entretien exigent plusieurs étapes, toujours
utiliser la correcte sécurité et consulter votre responsable de sécurité
Lors de l'entretien sur n'importe quelles lignes de secteur, toujours éteindre
premièrement l’alimentation.
Quelques composants détachables peuvent être lourds, respecter les règles
de sécurité du soulèvement.
Vérifier tous les composants sont correctement installés avec un
fonctionnement vérifié avant de faire des opérations normales.
Only trained and experienced certification technicians should
perform some of the service operations after the fume hood has been properly decontaminated. DO NOT attempt to perform these operations if you are not properly trained. The wrench icon precedes the service operations that require qualified technicians.
Seulement les techniciens de certification expérimentés et entraînés
devraient exécuter certaines des opérations d’entretien après que la hotte d’aspiration ait été convenablement décontaminée. NE PAS
tenter d'exécuter ces opérations si vous n'êtes pas convenablement
entraîné. L'icône de clé plate précède les opérations d’entretien qui
exigent des techniciens qualifiés.
Routine Maintenance Schedule
Weekly
Using ordinary dish soap to clean the surface inside of the fume hood, and
the work surface.
Using an appropriate glass cleaner, clean the sash and all glass surfaces. Operate the fume hood blower, noting the airflow velocity through the
hood using a source of visible smoke.
Monthly (or more often as required) Mensuellement (ou plus régulièrement si nécessaire)
Determine the actual face velocity through the sash opening of the hood
where the average reading should be at the specified velocity. (Use calibrated thermal anemometer or other approved apparatus).
Déterminer l’actuel flux d’aspiration à travers le sas d’entrée de la hotte où
la valeur moyenne doit être égale à la valeur spécifiée. (Utiliser un anémomètre thermique calibré ou d’autres appareils approuvés).
Using a damp cloth, clean the exterior surfaces of the hood, particularly
the front of the hood, to remove any accumulated dust.
En utilisant un chiffon humidifié, nettoyer les parties extérieures de la
hotte, en particulier le devant de la hotte pour enlever la poussière accumulée.
Check all service valves, if so equipped, for proper operation. Contrôler toutes les soupapes, si présentes, pour le bon fonctionnement.
Product Service 1-800-522-7658
Chapter 6: Maintaining Your Protector XStream
32
Original instructions
S
A
The hood baffles should be checked for blockages behind them to ensure
that the hood is maintaining proper airflow.
Les déflecteurs de hotte devraient être contrôlés pour leur blocage arrière
afin d’assurer que la hotte maintient un flux d’air correct.
All weekly activities. Toutes les activités hebdomadaires.
Annually Annuellement
Replace the lamps if needed. Remplacer les lampes, si nécessaire. Have the fume hood recertified by a qualified certification technician. See
“Certifying the Protector Fume Hood” in Chapter 3.
Recertifier la hotte aspirante par un technicien agréé en certification.Voir
au Chapitre 3 « Certifier la Hotte Aspirante Protecteur ».
All monthly activities. Toutes les activités mensuelles.
Biannually
The sash assembly should be checked to ensure proper operation and to
make sure there are no signs of abnormal wear on the sash pulleys, cables and clamps.
Routine Service Operations
Front Panel Removal:
1. Simply lift the front panel up and then away from the hood to provide access
to the top.
Changing the Lamp:
1. Turn light switch to “OFF”.
2. Remove the front panel as noted earlier.
3. Reach over the front header of the hood and remove the knock out plugs at
both ends of fixture. Lift fixture up.
4. Remove the lamp by pushing it out of the spring-loaded lamp socket and
swinging it out of the other lamp socket.
5. Install the new lamp by reversing the removal procedure.
Product Service 1-800-522-7658
33
Original instructions
CHAPTER 7
MODIFYING YOUR
PROTECTOR XSTREAM
There are several ways to modify the fume hood for your individual requirements. These include the addition of work surfaces, service fixtures, air monitor, distillation grids, electrical duplex outlets, ceiling enclosures, and rear panels.
Installing Work Surfaces
Your Protector XStream Fume Hood with Eco-Foil requires a large 1.0" leading edge radiused work surface to work properly, and achieve high performance. The work surface used for the Flush Foil also has precise dimensions for optimal performance. Contact Labconco Customer Service for ordering information.
Figure 7-1
Product Service 1-800-522-7658
Chapter 7: Modifying Your Protector XStream
34
Original instructions
9 5 5 0 ___
___
___
STEP 1. Select the nominal width of your fume hood. This
number is the fifth digit of your catalog number.
3 = 3' (XL Benchtop only) 4 = 4' 5 = 5' 6 = 6' 7 = 7' (XL Benchtop only) 8 = 8'
STEP 2. Select the depth
based on the model chosen. Choose 0 for Premier and 1 for XStream. Choose 1 for PVC. Choose 0, 1 or 2 for the correct depth of the XL Benchtop model. Choose 0 or 1 for the correct depth of Stainless Steel. This is the sixth digit of your catalog number.
STEP 3. Choose the correct height that matches the
number. This height should match the height between the top of the hood and the building ceiling. This is the seventh digit of your catalog number.
No.
Height
0
6"
1
12" 2 18"
3
24"
4
30" 5 36"
No.
Ext. Depth
Model
0
31.7"
Premier,
XL Benchtop,
Stainless Steel
1
37.7"
XStream,
XL Benchtop,
XL Floor-Mounted,
Stainless Steel,
PVC
2
43.7"
XL Benchtop,
XL Floor-Mounted
Installing Ceiling Enclosures Above the
Fume Hood
Your Protector XStream Fume Hood has mounting holes to accept a ceiling enclosure to close off the area between the top of the hood and the ceiling. Contact Labconco Customer Service for ordering information. Labconco offers both non-adjustable fixed height ceiling enclosures as well as adjustable height ceiling enclosures per the charts below. Other special sizes are available upon request.
3-Sided Ceiling Enclosure Kits
The panels extend above the top of the hood to the ceiling to hide exposed ductwork, plumbing and wiring. For any 3 ft to 8 ft Protector XL Benchtop, Premier, XStream, Stainless Steel or PVC Fume Hood. Each kit includes three glacier white powder-coated steel ceiling enclosure panels, one for the front of the hood and one for each side of the hood.
Configure Your Non-Adjustable Fixed Height Ceiling Enclosure Kit Catalog Number
Use this key to configure the seven digit catalog number to order your 3-Sided Fixed Height Ceiling Enclosure Kit. For example, 9550511 is a 3-Sided Ceiling Enclosure Kit for a 5' Protector XStream Hood with a fixed height of 12".
Product Service 1-800-522-7658
Chapter 7: Modifying Your Protector XStream
35
Original instructions
9 4 1 4 ___
___
___
STEP 1. Select the nominal width of your fume hood. This
number is the fifth digit of your catalog number.
3 = 3' (XL Benchtop only) 4 = 4' 5 = 5' 6 = 6' 7 = 7' (XL Benchtop only) 8 = 8'
STEP 2. Select the depth
based on the model chosen. Choose 0 for Premier and 1 for XStream. Choose 1 for PVC. Choose 0, 1 or 2 for the correct depth of the XL Benchtop model. Choose 0 or 1 for the correct depth of Stainless Steel. This is the sixth digit of your catalog number.
STEP 3. Choose the correct height range that matches
the number. This height range should match the height between the top of the hood and the building ceiling. This is the seventh digit of your catalog number.
No.
Height Range
0
11-14"
1
14-18.6"
2
18.6-24.4"
No.
Ext. Depth
Model
0
31.7"
Premier,
XL Benchtop,
Stainless Steel
1
37.7"
XStream,
XL Benchtop,
XL Floor-Mounted,
Stainless Steel,
PVC
2
43.7"
XL Benchtop,
XL Floor-Mounted
Configure Your Adjustable Height Ceiling Enclosure Kit Catalog Number
Use this key to configure the seven digit catalog number to order your 3-Sided Adjustable Height Ceiling Enclosure Kit. For example, 9414511 is a 3-Sided Ceiling Enclosure Kit for a 5' Protector XStream Hood with a height range of 14-19".
Product Service 1-800-522-7658
Chapter 7: Modifying Your Protector XStream
36
Original instructions
Side Panel
(Ceiling Enclosure)
Rear Stiffener
Side Panel
(Ceiling Enclosure)
Hood Side Panel
Hood Front Panel
Front Panel
(Ceiling Enclosure)
Figure 7-2
Installing Rear Panels Behind the Fume
Hood
Your Protector XStream Fume Hood can be modified to add a rear panel behind the fume hood when the fume hood is placed on an island. Contact Labconco Customer Service for ordering information.
Product Service 1-800-522-7658
Chapter 7: Modifying Your Protector XStream
37
Original instructions
Figure 7-5 Hose Connector
Installing Additional Service Fixtures
Additional service fixtures can be installed in the available service fixture holes in both sidewalls and corner posts. The fume hood is factory set to accept up to four valves per side. Contact Labconco Customer Service for information.
Figure 7-3 Knob Figure 7-4 Valve
Installing Guardian™ Digital Airflow
Monitor or Guardian™ Airflow Monitor
The Guardian Airflow Monitors continuously monitor face velocity through the fume hood opening. The fume hood right corner post is factory prepared to mount either monitor. Contact Labconco Customer Service to order.
Distillation Grids – Field Installation
The distillation grids have been strategically placed with the vertical rod center lines in front of the lower baffle and middle baffle. The distillation grids allow the hood user to mount glassware, motors, stirrers, and other apparatus. Contact Labconco Customer Service for ordering information.
Figure 7-8
Product Service 1-800-522-7658
Chapter 7: Modifying Your Protector XStream
38
Original instructions
Installing an Electrical Duplex Outlet
Your Protector Fume Hood can be ordered with duplex outlets, however, if you ordered a model without an electrical duplex outlet you can have one installed in the field by a qualified electrician. Contact Labconco Customer Service for ordering information. (Not acceptable on explosion-proof hoods).
Votre Protège-capot de fumées peut être commandé avec prises doubles, cependant, si vous avez commandé un modèle sans prise de courant duplex, vous pouvez en faire installer sur le terrain par un électricien qualifié. Contactez le service clientèle Labconco pour commander. (Non acceptable sur antidéflagrants hottes).
Figure 7-9
Product Service 1-800-522-7658
39
Original instructions
PROBLEM
CAUSE
CORRECTIVE ACTION
Remote blower and lights won’t operate
Wires not connected at junction boxes or switches.
Check connection of switches. Check connection to control box on top of unit.
Circuit breakers tripped in building electrical supply.
Reset circuit breakers.
Remote blower
won’t operate, but
lights work
Blower wiring is disconnected. Belt broken.
Inspect blower wiring and switch.
Replace belt.
Blower motor is defective.
Replace blower motor.
Fume hood blower operates but lights will not operate
Lamp not installed correctly.
Inspect lamp installation.
Lamp is defective.
Replace lamp.
Lamp circuit breaker in building is tripped.
Reset the lamp circuit breaker.
CHAPTER 8
TROUBLESHOOTING
Refer to the following table if your fume hood fails to operate properly. If the suggested corrective actions do not solve your problem, contact Labconco for additional assistance.
Product Service 1-800-522-7658
Chapter 8: Troubleshooting
40
Original instructions
PROBLEM
CAUSE
CORRECTIVE ACTION
Fume hood blower operates, but lights will not operate
Lamp wiring is disconnected.
Inspect lamp wiring. Defective lamp ballasts.
Replace lamp ballasts.
Contaminants outside of fume hood
Improper user techniques for the fume hood.
See “Certifying the Hood” Chapter 3 and “Safety Precautions” Chapter 4 sections in
the manual. (Ref. Appendix E)
Restriction of the baffle air slots or – blockage of the exhaust outlet.
Remove baffles to ensure that all air slots, and the exhaust outlet are unobstructed.
External factors are disrupting the fume hood airflow patterns or acting as a source of contamination.
See “Location Requirements” Chapter 2, “Certifying the Hood” Chapter 3, and “Safety Precautions” Chapter 4 sections of
this manual. (Ref. Appendix E)
XStream Fume hood has improper face velocity.
Have fume hood re-certified and check remote blower exhaust system. XStream Hood should have average face velocity of 50-100 fpm.
Vertical sash no longer operates smoothly
Cable is frayed or plastic protection is damaged.
Inspect cable and replace cable if worn or damaged immediately; otherwise injury could result.
Pulley bearing is damaged.
Replace pulley, bearing or add grease.
Cable or chain has slipped off the pulleys or sprockets.
Re-install, cable or chain must be replaced immediately if damaged.
Weight has broken pulleys or sprockets.
Replace pulleys or sprockets.
Combination A­Style Sash no longer operates smoothly
Horizontal glass panels have come off the tracks.
Re-install horizontal glass on tracks.
Product Service 1-800-522-7658
Chapter 8: Troubleshooting
41
Original instructions
PROBLEM
CAUSE
CORRECTIVE ACTION
Vertical sash frame is distorted.
Place horizontal glass symmetrically and pull sash down to air foil. Straighten damaged frame.
Cable is frayed or has slipped off the pulleys.
Re-install, cable must be replaced immediately if damaged.
Electrical duplex outlets no longer have power
Wires not connected or faulty duplex.
Check wire connection or replace duplex.
Circuit breakers tripped in building electrical supply.
Reset circuit breakers.
Service valves no longer operate
Faulty building supply.
Inspect building supply shut off valves and appropriate pressures below the recommended 40 PSI.
Valve no longer operates.
Replace valve and check for leaks.
Supply line or outlet line has leaks.
Inspect line for leaks and fix any leaking plumbing connections.
Product Service 1-800-522-7658
42
Original instructions
Item
Quantity
Part No.
Description
1A 1 9823700
Valve, Labconco (Water) 1/4" Compression Fitting
1B 1 9823701
Valve, Labconco (Water) 3/8" Compression Fitting
1C 1 9817000
Valve, Labconco 1/4" Compression Fitting (AIR, GAS, VAC, NIT, etc.)
1D 1 9817001
Valve, Labconco 3/8" Compression Fitting (AIR, GAS, VAC, NIT, etc.)
1E 1 9823702
Valve, Labconco Deionized 1/4" Compression Fitting
1F 1 9823703
Valve, Labconco Deionized 3/8" Compression Fitting
1G 1 9818000
Nut, Valve Mtg. (Labconco)
2A 1 9826800
WaterSaver Valve/Gooseneck – GRN
2B 1 9826801
WaterSaver Valve/Connector (VAC) – YEL
2C 1 9826802
WaterSaver Valve/Connector (AIR) – ORG
2D 1 9826803
WaterSaver Valve/Connector (GAS) – BLU
2E 1 9826805
WaterSaver Valve/Connector (HOT WATER) – RED
2F 1 9826806
WaterSaver Valve/Connector (CW) – GRN
2G 1 9826807
WaterSaver Valve/Connector (STEAM) – BLK
2H 1 9826808
WaterSaver Valve/Connector (NITROGEN) – BRN
2I 1 9826809
WaterSaver Valve/Connector (OXYGEN) – LIGHT GREEN
2J 1 9826810
Swivel Gooseneck only – GRN
2K 1 9826812
Swivel Gooseneck only – WHITE
3 1 9818700 thru 08
Knobs (GRAY, GRN, BLU, ORG, YEL, RED, WHT, BLK, BRN)
4A 1 9818800
Hose Barb, GRAY (NEUTRAL OR ARGON) – NOT SHOWN
4B 1 9818801
Hose Barb, GREEN (COLD WATER) – NOT SHOWN
4C 1 9818802
Hose Barb, BLUE (GAS) – NOT SHOWN
4D 1 9818803
Hose Barb, ORANGE (AIR) – NOT SHOWN
4E 1 9818804
Hose Barb, YELLOW (VACUUM) – NOT SHOWN
4F 1 9818805
Hose Barb, RED (HOT WATER) – NOT SHOWN
4G 1 9818806
Hose Barb, WHITE (DEIONIZED WATER) – NOT SHOWN
4H 1 9818807
Hose Barb, BLACK (NEUTRAL OR STEAM) – NOT SHOWN
4I 1 9818808
Hose Barb, BROWN (NITROGEN) – NOT SHOWN
4J 1 9819000
Nut, Hose Barb – NOT SHOWN
5 1 9825500
Label, Knob (contains all the labels)
6 1 9818900
Lens, Knob
Protector XStream Replacement Parts
APPENDIX A PROTECTOR XSTREAM COMPONENTS
Illustration A-1 indicate the location of the following service parts:
Product Service 1-800-522-7658
Appendix A: Protector XStream Components
43
Original instructions
Item
Quantity
Part No.
Description
7A 1 9947100, 01, 02
115V Duplex Receptacle (GRAY) Right, Left 4' - 6', Left 8' w/ wires
7B 1 9818200
Cover Plate 115V Duplex
7C 1 9947103, 04, 05
115V GFCI Duplex Receptacle (GRAY) Right, 4' - 6', Left 8' w/ wires
7D 1 9818100
Cover Plate, 115V GFCI
7E 1 9818300
Cover Plate, Blank
8A 1 1297503
LED, Direct Wired T8 x 3' – newest models use on 4' & 8' Hoods – 12 watts each
8B 1 1297504
LED, Direct Wired T8 x 4' – newest models use on 5' & 6' Hoods – 22 watts each
9A 1 9945300
LED Light Fixture, 36" – use on 4' Hoods
9B 1 9945301
LED Light Fixture, 48" – use on 5', 6' Hoods
9C 1 9945302
LED Light Fixture, 36" Combo – use in 8' Hoods
10A 1 1302301
Switch, Rocker
10B 1 1327500
Switch, Plug (Fills cutout when switch is not used)
11A 1 9818400
Access Cover
12A 1 9409801
Side Panel, 30" internal deep hoods – NOT SHOWN
12B 4 1885308
Screw, Machine #10-24 x .50 Phillips
13A 1 9436501
Eco-Foil 4'
13B 1 9436502
Eco-Foil 5'
13C 1 9436503
Eco-Foil 6'
13D 1 9436505
Eco-Foil 8'
13E 1 9436521
PVC Eco-Foil, 4'
13F 1 9436522
PVC Eco-Foil, 5'
13G 1 9436523
PVC Eco-Foil, 6'
13H 1 9436525
PVC Eco-Foil, 8'
14A 2 1850000
Pulley, Front, 2" Dia. Nylon
14B 2 9709300
Plastic Pulley (Rear 2")
15A 2 4949902
Cable, Sash 130" – NOT SHOWN
15B 2 9414011
Cable Replacement Kit 30"
15C 1 9545800
Weight Support Bracket Kit
16 2 9713300
Bumper, Rubber (lower sash bumper)
17 4 1920100
Clamp, Cable Replacement (use two each side) – NOT SHOWN
18 2 9935800
Threaded Connector (to attach weight to cable) – NOT SHOWN
19 1 9946300 or 01
Wiring Harness, Main
20 1 9409601, 02, 03, 05
Front Panel, 4', 5', 6', 8'
21 1 9410300
Sash Stop Kit (XStream)
22 1 1598500
Hinged Screw Cap Cover – NOT SHOWN
Product Service 1-800-522-7658
Appendix A: Protector XStream Components
44
Original instructions
20
14
15
21
11
3
1
5, 6 7 13
2
10, 19
12
8, 9
4
16
Product Service 1-800-522-7658
45
Original instructions
HOOD SIZE
HOOD WIDTH
INTERNAL HOOD WIDTH
4'
48.00"
38.25"
5'
60.00"
50.25"
6'
72.00"
62.25"
8'
96.00"
86.25"
APPENDIX B PROTECTOR XSTREAM DIMENSIONS
Figure B-1
Product Service 1-800-522-7658
46
Original instructions
Volts AC
Phase
Cycle
AMP*
115 1 50/60
10
115 1 50
10
115 1 60
10
230 1 50/60
5
230 1 50
5
230 1 60
5
APPENDIX C PROTECTOR XSTREAM SPECIFICATIONS
Environmental Conditions
Indoor use only. Maximum altitude: 10,000 feet (3,048 meters). Ambient temperature range: 41° to 104°F (5° to 40°C). Maximum relative humidity: 80% for temperatures up to 88°F (31°C),
decreasing linearly to 50% relative humidity at 104°F (40°C).
Main supply voltage fluctuations not to exceed ±10% of the nominal
voltage.
Transient over-voltages according to Installation Categories II (Over-
voltage Categories per IEC 1010). Temporary voltage spikes on the AC input line that may be as high as 1500V for 115V models and 2500V for 230V models are allowed.
Used in an environment of Pollution degrees 2 (i.e., where normally only
non-conductive atmospheres are present). Occasionally, however, a temporary conductivity caused by condensation must be expected, in accordance with IEC 664.
Electrical Ratings
*Does not include current rating of receptacles.
Product Service 1-800-522-7658
Appendix C: Protector XStream Specifications
47
Original instructions
Product Service 1-800-522-7658
Appendix C: Protector XStream Specifications
48
Original instructions
Product Service 1-800-522-7658
49
Original instructions
APPENDIX D PROTECTOR I-S HOOD
Labconco offers Protector® I-S™ Fume Hood models installed with limit switches for automatic Intelli-Sense Blower operation that change 3 speeds automatically based off the sash position. The Intelli-Sense Blower low speed runs with the rising sash from 0"-3", with the medium speed sash from 4"-19" and the high speed sash from 20"-28"; the medium and high speed sash height positions can be altered by 1" increments in the field between 15" and 20". The limit switches are installed and wired per the automatic operation wiring diagrams in Appendix C. The limit switches replace the manually operated 3-position switch. A Digital Airflow Monitor on the Protector I-S Hood constantly monitors face velocity at all speeds and sash positions to ensure proper hood operation. Note: The integrated Protector I-S Hoods with limit switches are only available from the factory due to the wiring complexity.
The integrated Protector I-S Hood models with limit switches automatically regulate the Intelli-Sense Blower, but independent of any room controls. Room controls such as room pressurization controls depicted in Appendix E are sold separately and are not sold by Labconco. The Protector I-S Hood is primarily used for small hood projects when the number of fume hoods cannot justify a complete variable airflow volume system or VAV. The Protector I-S Hood alters the (RPM) speed of the electronically commutated motor (ECM) blower wheel automatically with the sash position to change speeds. Some VAV systems can degrade and usually have annual service issues, but the Protector I-S Hood with Intelli-Sense Blower is reliable with little maintenance as it has no belts, valves or sophisticated controls systems. The reliable Intelli-Sense Blower has a rated bearing life of 50,000 hours (6 years) at full speed and a test life of over 150,000 hours (over 17 years).
Product Service 1-800-522-7658
Appendix D: Protector I-S Hood
50
Original instructions
Item
Qty.
Part No.
Description – Protector I-S Unique Components
1A 1 7115010
MSB Control Box, 115V Automatic
1B 1 7115020
MSB Control Box, 230V Automatic
2 3 3832400
Limit Switch, Automatic Blower Operation
3 1 7117700
Accessory Cable, Airflow Monitor for Night Setback (when programmed)
4 1 7116000
Power Cable, Control Box Inlet Power (from hood lighting junction box)
Please keep in mind that room controls such as room pressurization controls (Appendix E), duct sensors, or hood exhaust volume controls are not included. The variety of room controls MUST BE selected by the building controls engineering department. Note: Room controls are not sold by Labconco. In typical environments the room controls are set to a negative pressure with a range of -0.01" to -0.05" depending on the application.
The following summarizes the Protector I-S Hood installed with limit switches for automatic Intelli-Sense Blower operation.
INTEGRATED – Minimal set up, pre-mounted, multi-speed MSB
controller, and pre-wired limit switches.
INDEPENDENT – No conscious control as blower speeds change
automatically with sash position and room controls work independently. Room controls are sold separate and not sold by Labconco.
INTELLIGENT – The Integrated Intelli-Sense Hood becomes a Smart
Hood that adjusts 3 speeds automatically based off sash position, has built­in over-amperage adjustment to prevent overheating, provides simple blower sizing, and provides 3 easy speed adjustments (MIN, MED, MAX) based off MIN sash height of 0"-3", MED sash height of 4"-19", and MAX sash height of 20"-28"; the MED and MAX sash heights can be altered by 1" increments in the field between 15" and 20".
Protector I-S Service Parts Only
Product Service 1-800-522-7658
Appendix D: Protector I-S Hood
51
Original instructions
Left side upper limit switch wired for high speed and set to activate with sash raised 20" above work surface. The upper limit switch can be lowered in 1" increments between 15" and 20".
Red wire to common.
White wire to normally open.
Intelli-Sense Control Box powered up and functional without lights. Control box is wire to incoming hood light circuit per wiring diagram 7115810 or 7115811.
Product Service 1-800-522-7658
Appendix D: Protector I-S Hood
52
Original instructions
Red wire to common.
Left side lower limit switch wired for low speed.
Black wire to normally open.
Product Service 1-800-522-7658
Appendix D: Protector I-S Hood
53
Original instructions
Airflow monitor in-line sensor, air pathway, and wiring to monitor.
Night setback lower switch shown on right. Switch is wired to normally open and common terminals, and wired to air monitor input I/P1. Programmed for “Night Setback – Input 1.”
Back of airflow monitor.
Power to air monitor.
Product Service 1-800-522-7658
Appendix D: Protector I-S Hood
54
Original instructions
Front of airflow monitor
showing “SETBACK”
when sash is closed.
Product Service 1-800-522-7658
Appendix D: Protector I-S Hood
55
Original instructions
Back of airflow monitor night setback switch wiring to I/P1 for Input 1.
Sensor wiring in to air monitor.
Power in to air monitor.
Product Service 1-800-522-7658
Appendix D: Protector I-S Hood
56
Original instructions
115V Wiring Diagram – Automatic Operation
230V Wiring Diagram – Automatic Operation
Product Service 1-800-522-7658
57
Original instructions
Volts AC
Phase
Cycle
AMP x
115
1
50/60
10X
115 1 50
10X
115 1 60
10X
230
1
50/60
5X
230 1 50
5X
230 1 60
5X
APPENDIX E SERIAL NUMBER TAG DESCRIPTION
Serial tag includes standard information with the following changes:
The portion of the tag for electrical information is labeled with the following units: Volts AC, Phase, Cycle, AMP.
Each hood will include one of the following:
Where may be a separate label indicating: Caution, See Manual.
The X above will be one of several possible alpha characters as defined in the manual as follows:
X: No additional duplexes on hood, noted amperage applies. B: This fume hood includes one additional electrical receptacle, individually
wired to the field wired box and individually rated as 115V, single phase, 60Hz, 20 Amps. Each duplex can be wired on a dedicated circuit rated at 20A or the duplexes may be ganged together on the same circuit with the total load of the ganged duplexes not exceeding the 20A electrical service they are wired to.
C: This fume hood includes two additional electrical receptacles, individually
wired to the field wired box and individually rated as 115V, single phase, 60Hz, 20 Amps. Each duplex can be wired on a dedicated circuit rated at 20A or the duplexes may be ganged together on the same circuit with the total load of the ganged duplexes not exceeding the 20A electrical service they are wired to.
Product Service 1-800-522-7658
Appendix E: Serial Number Tag Description
58
Original instructions
D: This fume hood includes three additional electrical receptacles,
individually wired to the field wired box and individually rated as 115V, single phase, 60Hz, 20 Amps. Each duplex can be wired on a dedicated circuit rated at 20A or the duplexes may be ganged together on the same circuit with the total load of the ganged duplexes not exceeding the 20A electrical service they are wired to.
E: This fume hood includes four additional electrical receptacles,
individually wired to the field wired box and individually rated as 115V, single phase, 60Hz, 20 Amps. Each duplex can be wired on a dedicated circuit rated at 20A or the duplexes may be ganged together on the same circuit with the total load of the ganged duplexes not exceeding the 20A electrical service they are wired to.
F: This fume hood includes one additional electrical receptacle, individually
wired to the field wired box and individually rated as 230V, single phase, 60Hz, 10 Amps. Each duplex can be wired on a dedicated circuit rated at 10A or the duplexes may be ganged together on the same circuit with the total load of the ganged duplexes not exceeding the 10A electrical service they are wired to.
G: This fume hood includes two additional electrical receptacles, individually
wired to the field wired box and individually rated as 230V, single phase, 60Hz, 10 Amps. Each duplex can be wired on a dedicated circuit rated at 10A or the duplexes may be ganged together on the same circuit with the total load of the ganged duplexes not exceeding the 10A electrical service they are wired to.
H: This fume hood includes three additional electrical receptacles,
individually wired to the field wired box and individually rated as 230V, single phase, 60Hz, 10 Amps. Each duplex can be wired on a dedicated circuit rated at 10A or the duplexes may be ganged together on the same circuit with the total load of the ganged duplexes not exceeding the 10A electrical service they are wired to.
J: This fume hood includes four additional electrical receptacles,
individually wired to the field wired box and individually rated as 230V, single phase, 60Hz, 10 Amps. Each duplex can be wired on a dedicated circuit rated at 10A or the duplexes may be ganged together on the same circuit with the total load of the ganged duplexes not exceeding the 10A electrical service they are wired to.
K: This fume hood includes one additional electrical receptacle, individually
wired to the field wired box and individually rated as 230V, single phase, 50Hz, 10 Amps. Each duplex can be wired on a dedicated circuit rated at 10A or the duplexes may be ganged together on the same circuit with the total load of the ganged duplexes not exceeding the 10A electrical service they are wired to.
L: This fume hood includes two additional electrical receptacles, individually
wired to the field wired box and individually rated as 230V, single phase, 50Hz, 10 Amps. Each duplex can be wired on a dedicated circuit rated at 10A or the duplexes may be ganged together on the same circuit with the total load of the ganged duplexes not exceeding the 10A electrical service they are wired to.
Product Service 1-800-522-7658
Appendix E: Serial Number Tag Description
59
Original instructions
M: This fume hood includes three additional electrical receptacles,
individually wired to the field wired box and individually rated as 230V, single phase, 50Hz, 10 Amps. Each duplex can be wired on a dedicated circuit rated at 10A or the duplexes may be ganged together on the same circuit with the total load of the ganged duplexes not exceeding the 10A electrical service they are wired to.
N: This fume hood includes four additional electrical receptacles,
individually wired to the field wired box and individually rated as 230V, single phase, 50Hz, 10 Amps. Each duplex can be wired on a dedicated circuit rated at 10A or the duplexes may be ganged together on the same circuit with the total load of the ganged duplexes not exceeding the 10A electrical service they are wired to.
Product Service 1-800-522-7658
Appendix E: Serial Number Tag Description
60
Original instructions
Volts AC
Phase
Cycle
AMP
115
1
50/60
10X
115 1 50
10X
115 1 60
10X
230
1
50/60
5X
230 1 50
5X
230 1 60
5X
ANNEXE E DESCRIPTION DE SÉRIE TAG NUMBER
Étiquette de série comprend des informations standard avec les modifications suivantes:
La partie de l'étiquette d'information électrique est étiqueté avec les unités suivantes: Volts AC, phase, cycle, AMP.
Chaque hotte comprendra une des opérations suivantes:
peut-être une étiquette distincte indiquant: Attention, consultez le manuel.
Le X ci-dessus sera l'un des plusieurs caractères alphabétiques possibles telles que définies dans le manuel comme suit:
X: Pas de duplex supplémentaires sur le capot, noté ampérage s'applique. B: Cette hotte comprend un réceptacle électrique supplémentaire,
individuellement raccordés aux boîtiers câblées individuellement et classé comme 115V, monophasé, 60 Hz, 20 ampères. Chaque duplex peut être branché sur un circuit dédié évalué à 20A ou les duplex peuvent être groupés ensemble sur le même circuit que la charge totale des duplex couplées ne dépassant pas le service électrique 20A ils sont câblés.
C: Cette hotte comprend deux autres prises électriques, individuellement
câblés à la boîte câblées individuellement et classé comme 115V, monophasé, 60 Hz, 20 A. Chaque duplex peut être branché sur un circuit dédié évalué à 20A ou les duplex peuvent être groupés ensemble sur le même circuit que la charge totale des duplex couplées ne dépassant pas le service électrique 20A ils sont câblés.
D: Cette hotte comporte trois prises électriques supplémentaires,
individuellement câblés à la boîte câblées individuellement et classé comme 115V, monophasé, 60 Hz, 20 A. Chaque duplex peut être branché sur un circuit dédié évalué à 20A ou les duplex peuvent être groupés ensemble sur le même circuit que la charge totale des duplex couplées ne dépassant pas le service électrique 20A ils sont câblés.
Product Service 1-800-522-7658
Appendix E: Serial Number Tag Description
61
Original instructions
E: Cette hotte comprend quatre autres prises électriques, individuellement
câblés à la boîte câblées individuellement et classé comme 115V, monophasé, 60 Hz, 20 A. Chaque duplex peut être branché sur un circuit dédié évalué à 20A ou les duplex peuvent être groupés ensemble sur le même circuit que la charge totale des duplex couplées ne dépassant pas le service électrique 20A ils sont câblés.
F: Cette hotte comprend un réceptacle électrique supplémentaire,
individuellement raccordés aux boîtiers câblées individuellement et classé comme 230V, monophasé, 60 Hz, 10 ampères. Chaque duplex peut être branché sur un circuit dédié classé au 10A ou les duplex peuvent être groupés ensemble sur le même circuit que la charge totale des duplex couplées ne dépassant pas le service électrique 10A ils sont câblés.
G: Cette hotte comprend deux autres prises électriques, individuellement
câblés à la boîte câblées individuellement et classé comme 230V, monophasé, 60 Hz, 10 ampères. Chaque duplex peut être branché sur un circuit dédié classé au 10A ou les duplex peuvent être groupés ensemble sur le même circuit que la charge totale des duplex couplées ne dépassant pas le service électrique 10A ils sont câblés.
H: Cette hotte comporte trois prises électriques supplémentaires,
individuellement câblés à la boîte câblées individuellement et classé comme 230V, monophasé, 60 Hz, 10 ampères. Chaque duplex peut être branché sur un circuit dédié classé au 10A ou les duplex peuvent être groupés ensemble sur le même circuit que la charge totale des duplex couplées ne dépassant pas le service électrique 10A ils sont câblés.
J: Cette hotte comprend quatre autres prises électriques, individuellement
câblés à la boîte câblées individuellement et classé comme 230V, monophasé, 60 Hz, 10 ampères. Chaque duplex peut être branché sur un circuit dédié classé au 10A ou les duplex peuvent être groupés ensemble sur le même circuit que la charge totale des duplex couplées ne dépassant pas le service électrique 10A ils sont câblés.
K: Cette hotte comprend un réceptacle électrique supplémentaire,
individuellement raccordés aux boîtiers câblées individuellement et classé comme 230V, monophasé, 50 Hz, 10 ampères. Chaque duplex peut être branché sur un circuit dédié classé au 10A ou les duplex peuvent être groupés ensemble sur le même circuit que la charge totale des duplex couplées ne dépassant pas le service électrique 10A ils sont câblés.
L: Cette hotte comprend deux autres prises électriques, individuellement
câblés à la boîte câblées individuellement et classé comme 230V, monophasé, 50Hz, 10 ampères. Chaque duplex peut être branché sur un circuit dédié classé au 10A ou les duplex peuvent être groupés ensemble sur le même circuit que la charge totale des duplex couplées ne dépassant pas le service électrique 10A ils sont câblés.
M: Cette hotte comporte trois prises électriques supplémentaires,
individuellement câblés à la boîte câblées individuellement et classé
Product Service 1-800-522-7658
Appendix E: Serial Number Tag Description
62
Original instructions
comme 230V, monophasé, 50Hz, 10 ampères. Chaque duplex peut être branché sur un circuit dédié classé au 10A ou les duplex peuvent être groupés ensemble sur le même circuit que la charge totale des duplex couplées ne dépassant pas le service électrique 10A ils sont câblés.
N: Cette hotte comprend quatre autres prises électriques, individuellement
câblés à la boîte câblées individuellement et classé comme 230V, monophasé, 50Hz, 10 ampères. Chaque duplex peut être branché sur un circuit dédié classé au 10A ou les duplex peuvent être groupés ensemble sur le même circuit que la charge totale des duplex couplées ne dépassant pas le service électrique 10A ils sont câblés.
Product Service 1-800-522-7658
63
Original instructions
APPENDIX F REFERENCES
Many excellent reference texts and booklets are currently available. The following is a brief listing:
Laboratory Ventilation Standards
Federal Register 29 CFR Part 1910
Non-mandatory recommendations from “Prudent Practices”.
Fume hoods should have a continuous monitoring device Face velocities should be between 60-100 linear feet per minute (lfpm) Average 2.5 linear feet of hood space per person
Occupational Health and Safety U.S. Department of Labor 200 Constitution Avenue N.W. Washington, DC 20210 (202) 523-1452
Industrial Ventilation-ACGIH
Fume hood face velocities between 60-100 lfpm Maximum of 125 lfpm for radioisotope hoods Duct velocities of 1000-2000 fpm for vapors, gasses and smoke Stack discharge height 1.3-2.0 x building height Well designed fume hood containment loss, <0.10 ppm
Industrial Ventilation, A Manual of Recommended Practice. 24th Edition, 2001 American Conference of Governmental Industrial Hygienists 1330 Kemper Meadow drive Cincinnati, OH 45240-1634 (513) 742-2020
ASHRAE 110-1995 Method of Testing Performance of Fume Hoods
Evaluates fume hood’s containment characteristics
Three part test: Smoke generation, Face velocity profile, Tracer gas release @ 4 liters
per minute
Rated As Manufactured (AM), As Installed (AI) and As Used (AU)
American Society of Heating, Refrigerating, and Air Conditioning Engineers 1791 Tullie Circle N.E. Atlanta, GA 30329 (404) 636-8400
Product Service 1-800-522-7658
Appendix F: References
64
Original instructions
ANSI Z9.5-2011 Laboratory Standard
Covers entire laboratory ventilation system.
New and remodeled hoods shall have a monitoring device Ductless hoods should only be used with non-hazardous materials
American Industrial Hygiene Association 2700 Prosperity Avenue, Suite 250 Fairfax, VA 22031 (703) 849-8888
SEFA 1-2002
Fume hood face velocities based on toxicity levels of chemicals
Class A – 125 to 150 fpm Class B – 80 to100 fpm Class C – 75-to 80 fpm
Test method – face velocity profile and smoke generation
Scientific Equipment & Furniture Association 1028 Duchess Drive McLean, VA 22102 (703) 538-6007
NFPA 45 – 2002 Fire Protection for Laboratories Using Chemicals
Laboratory hoods should not be relied on for explosion protection Exhaust air from fume hoods should not be recirculated Services should be external to the hood Canopy hoods only for non-hazardous applications Materials of construction should have flame spread of 25 or less 80 to 120 fpm to prevent escape
NFPA 30 – 2000 Flammable and Combustible Liquids Code
Approved cabinets may be metal or wood Vent location on cabinets are required Venting of cabinets not a requirement
National Fire Protection Association 1 Batterymarch Park P.O. Box 9101 Quincy, MA 02269-9101 (800) 344-3555
General References
American Conference of Governmental Industrial Hygienists. Industrial Ventilation, A Manual of Recommended Practice, Cincinnati, OH
ASHRAE Standard Committee. ASHRAE Standard Atlanta: ASHRAE Publications Sales Department, 1995
British Standards Institution, Laboratory Fume Cupboards. Parts 1, 2 and 3, London: 1990
Product Service 1-800-522-7658
Appendix F: References
65
Original instructions
Department of Labor, Occupational Safety and Health Administration, 29 CFR Part 1910, Occupational Exposures to Hazardous Chemicals in Laboratories, Final Rule. Vol.
55, No. 21. Washington D.C.:1990
DiBerardinis. L. et al. Guides for Laboratory Design, Health and Safety Considerations. Wiley & Sons, 1987
McDermott, Henry, Handbook of Ventilation for Contaminant Control, 2nd Edition. Butterworth Publishers, 1985.
Miller, Brinton M. et al. Laboratory Safety: Principles and Practices. American Society for Microbiology, Washington, D.C.: 1986
NIH Guidelines for the Laboratory Use of Chemical Carcinogens. NIH Publication No. 81-2385.
Rayburn, Stephen R. The Foundation of Laboratory Safety, A Guide for the Biomedical Laboratory. Springer-Verlag, New York: 1990
Sax, N. Irving and Lewis, JR., Richard J. Rapid Guide to Hazardous Chemicals in the Workplace. Van Nostrand Reinhold, 1987.
Schilt, Alfred A. Perchloric Acid and Perchlorates. The G. Frederick Smith Chemical Company, Columbus, OH: 1979.
Steere, Norman. CRC Handbook of Laboratory Safety, 2nd Edition. CRC Press, 1971.