Labconco Protector Multi-Hazard Glove Box User Manual

Protector®
Multi-Hazard Glove Box
Models 5065000, 5065002
5065500, 5065502
INSTRUCTION MANUAL
Product designs are subject to change without notice
50937 Rev C, ECO 9999
Printed in U.S.A.
TABLE OF CONTENTS
Introduction
Preface/Important Notices .................................................................4
Components Shipped .........................................................................5
General Description ...........................................................................6
Optional Filters ..................................................................................6
Charcoal Filters..................................................................................6
Additional Exhaust HEPA Filter Kit .................................................6
Performance .......................................................................................7
Environmental Conditions .................................................................7
Component Identification ..................................................................8
Installation....................................................................................................11
Location .............................................................................................11
Connection Exhaust ...........................................................................11
Glove Attachment ..............................................................................11
Glove Replacement............................................................................11
Safety Precautions........................................................................................12
Normal Operation........................................................................................13
Blower Operation...............................................................................13
Routine Maintenance...................................................................................14
Estimation of Filter Condition ...........................................................14
Filter Replacement.............................................................................14
Suggested Plan for Decontamination.................................................14
Gloves ................................................................................................15
Transfer Chamber Inner and Outer Door...........................................15
Fluorescent Lamp Replacement.........................................................16
Replacement Parts .......................................................................................17
Dimensional Data.........................................................................................24
Electrical Specifications ..............................................................................25
Wiring Diagrams..........................................................................................26
Accessories....................................................................................................28
Warranty ......................................................................................................29
Shipping Claims ...........................................................................................30
Contacting Labconco...................................................................................31
Declaration of Conformity ..........................................................................32
PREFACE
Thank you for displaying confidence in us by selecting a Labconco Protector® Multi­Hazard Glove Box. Our design engineers, assemblers and inspectors have utilized their skills and years of experience to ensure that the new Labconco Controlled Atmosphere Glove Box meets our high standards of quality and performance.
IMPORTANT NOTICE
Labconco Protector® Glove Boxes are designed to contain chemical and some bio­hazardous materials. The Protector should not be considered a Class III Bio-Hazard Cabinet, nor should they be considered for use with Biosafety Level 4 agents. While the Protector Glove Box provides a physical barrier between the user and their work, it is the user’s responsibility to ensure that all materials are appropriately decontaminated, inactivated, or rendered harmless before removal from the unit. For further information regarding the handling of specific materials, please contact your facilities safety officer.
All the end users should read this manual carefully in order to become familiar with the operation of the glove box. Recommendations are made within the manual to help you obtain maximum performance and life from your product.
We have included sections on initial set up, operation, maintenance and troubleshooting to provide you with all the tools necessary to achieve maximum performance.
If you have questions or concerns, do not hesitate to call us at 1-800-821-5525 for assistance.
4
INTRODUCTION
Components Shipped
Carefully check the contents of the carton for damage that might have occurred in transit. Do not discard the carton or packing material until all components have been checked against the following component list and the equipment has been installed and tested.
As shipped, the carton should contain the following:
Part Number Description
5065000 Multi-Hazard Glove Box, Fiberglass
or
5065500 Multi-Hazard Glove Box, Stainless Steel
5005600 Glove Kit – Pair, Neoprene, size 9-3/4
(includes 2 - O-Rings)
5093700 Instruction Manual
5
INTRODUCTION
General Description
The Multi-Hazard Glove Box is equipped with a blower, air damper and filter system that
provide versatility for many particular applications. The variable speed blower provides air
volumes between 10 and 50 CFM and negative glove box pressure up to 1.5 inches H2O
gauge. The air damper provides a means for operating the glove box at a more negative
pressure when the blower is running at low speed (low air volume). The filtering system
consists of one inlet HEPA and one exhaust HEPA filter (both rated at 99.99% efficient on
0.3 micron particulates). A “roughing” prefilter is also provided to capture room air
particles and extend the life of the inlet HEPA filter.
Optional Filters: (for installation between exhaust HEPA and blower).
Charcoal Filter: (Cat. No. 5065300) contains 4 lbs. of Barnebey and Sutcliffe type 787
impregnated, activated carbon for removal of gaseous radioisotopes.
Additional Exhaust HEPA Filter Kit (Cat. No. 5065200) same as filter shipped with unit,
plus an extended length tube for connection to blower.
Figure 1
6
INTRODUCTION
Performance
The Labconco Protector Multi-Hazard Glove Box has been carefully designed, constructed
and tested in order to provide an effective physical barrier between the laboratory and the glove box interior. This barrier and the filtering system, provides a clean air enclosure for protection of specimens and materials. It also provides protection to the technician from contaminated materials and specimens.
Environmental Conditions
This equipment is designed to be safe under the following conditions:
Indoor use
Altitude up to 6562 Ft. (2000m)
Temperature 41° to 104° (5° to 40°)
Maximum relative humidity 80% for temperatures up to 88°F (31°C) decreasing linearly to
50% relative humidity at 104° (40°C)
Main supply voltage fluctuations not to exceed ± 10% of the nominal voltage
Transient over voltages according to Installation Categories II (Over voltage Categories per
IEC 1010)
Pollution degrees 2 (Normally only non-conductive foreign matter, solid, liquid, or gaseous
(ionized gases), that may produce a reduction of dielectric strength or surface resistivity occurs. Occasionally, however, a temporary conductivity caused by condensation must be expected), in accordance with IEC 664
7
INTRODUCTION
Component Identification
General components of the glove box are listed below and identified in Figure 2.
1. Glove Ports. Epoxy coated aluminum, 8" ID, spaced 17" apart, sealed to viewing window with one-piece molded neoprene gasket. Supplied with dual O-ring grooves, neoprene gloves, and retaining O-rings.
2. Neoprene Gloves. Supplied with the glove box, gloves are .015" thick neoprene, 30" long, one pair, and size 9-3/4. (Note: For replacement gloves, see Replacement Parts on page
18.
3. Window Frame. 12-gauge epoxy coated steel, attached with 30 stainless steel, ¼-20 screws, removable for full access to the interior.
4. Viewing Window. 3/8" thick laminated safety glass, mounted at 10° angle and sealed with one-piece molded neoprene gasket. Optional 3/8" thick polycarbonate with abrasion resistant coating available. See page 18.
5. Left Side Panel. 20-gauge epoxy coated steel, easily removable for access to glove box and left electrical outlet.
6. Fluorescent Lamp. 30 watts provides minimum of 60-foot candles illumination to interior work surface. Supplied with electrical cord and plug, the lamp may be easily removed for bulb replacement and for access to removable viewing window. For lamp replacement see page 16.
7. Façade. 18-gauge, epoxy coated steel, conceals filter damper, and blower.
8. Control Panel. Contains gauge for glove box pressure and switches for fluorescent lamp, interior receptacles, blower power, and blower speed control.
9. 115 Volt, 3-Wire Receptacles (Models 5065000 and 5065500). One right and one left interior receptacle, 9 amps combined rating. Exterior switches provide separate control of each receptacle.
230 Volt, 3-Wire Receptacles (Models 5065002 and 5065502). One right and one left interior receptacle, 8 amps combined rating. Exterior switches provide separate control of each receptacle.
10. Transfer Chamber Door. Counterweighted with quick operating latch.
8
INTRODUCTION
Figure 2
9
INTRODUCTION
Airflow system components are listed below and are identified in Figure 3.
1. Inlet Pre-Filter. Polyester fiber mat filter captures larger room air particles to extend life of inlet HEPA filter.
2. Air Damper. Adjust fully open (parallel with air flow) for maximum airflow volume. Adjust toward closed position to increase glove box negative pressure differential (if desired) when operating blower at low speed.
3. Inlet HEPA Filter. Rated 99.99% efficient on 0.3-micron particulates. NOTE: Inlet and exhaust filters are identical. Airflow direction is indicated on the filter.
4. Exhaust HEPA Filter. Rated 99.99% efficient on 0.3-micron particulates.
5. NOTE: Unit is normally supplied with a removable tube at this location which can easily be replaced with either of the two following optional filters:
Charcoal Filter (Cat. No. 5065300) contains 4 lbs. of Barnebey and Sutcliffe type 787 impregnated carbon (for removal of gaseous radioisotopes).
Additional Exhaust HEPA Filter Kit (Cat. No. 5065200)
6. Blower. Adjustable speed (control is located on front control panel) provides airflow between 10 and 50 CFM. The discharge opening measures 3 inch square.
Figure 3
10
INSTALLATION
Location
Glove box should be placed on a stable, level base stand near an appropriate electrical receptacle. The base stand must be capable of supporting 500 lbs minimum, and the height should provide a comfortable working position through glove ports (normally 35" to 37" height for standing position).
Connections, Exhaust
NOTE: It is the user’s responsibility to determine if the exhaust should be vented to outside the building. If it is ducted to the outside, duct restrictions should not exceed 0.5 inches water gauge static pressure at 50 CFM.
Glove Attachment
With glove thumbs up and right/left orientation, secure the gloves in place on the glove ports by stretching the beaded glove cuff into the groove nearest the window. Install the 8" diameter O-ring onto the outer groove over the glove surface.
Figure 4
Glove Replacement
If it becomes necessary to change gloves without disturbing the integrity of the glove box environment, follow these steps:
1. Remove the O-ring from the old glove.
2. Roll the old glove cuff bead from the inner groove nearest the window to the outside groove.
3. Insert the new glove through the old glove until the fingers of both are inside the glove box.
4. Stretch the new glove’s beaded cuff over the old glove cuff and into the groove nearest the window.
5. To remove the old glove, grasp the new glove surface and manipulate the old glove bead free from the outer groove.
6. With the other glove hand inside the box, pull the old glove into the glove box.
7. Install the O-ring over the new glove onto the outer groove.
11
SAFETY PRECAUTIONS
This product is neither designed nor intended to be an explosion-proof enclosure. It is the responsibility of the user to determine the lower explosive limits and flammability of the enclosed gases and other matter. The user is also responsible for using proper precautions to prevent equipment damage or injury due to explosion or combustion.
It is the responsibility of the user to determine the suitability of this product for the intended applications.
12
NORMAL OPERATION
Start Up
NOTE: All internal wiring is protected by two manual reset circuit breakers located on the rear panel.
Confirm that the glove box electrical power cord is plugged into an appropriate electrical power source.
Switch on the fluorescent lamp and check to make sure bulb is working.
Connect a test light to each interior receptacle and check for proper operation by
turning on the left and right outlet switches located on the control panel.
Turn on the blower switch and check blower operation.
NOTE: This should be done with gloves installed so that glove box negative pressure will be induced (observed by gauge on control panel). If the blower speed control is adjusted in the low range, you will notice that the blower starts at a higher speed, then slows down.
If unit is vented to the outside, confirm that ducting is sealed properly for positive pressure operation.
Blower Operation
The blower is operated on and off by a switch located on the control panel. The speed of the blower is regulated from 10% to 100% of maximum speed. Normal operation will be in the 30% to 50% area, depending on filter loading conditions and desired negative pressure within the enclosure. It may be necessary, during continued operation to increase the blower speed to compensate for possible restrictions to air flow caused by a dirty filter. The static pressure setting should be within .25 to .5 H2O. Pressure settings higher than this may cause excessive air flow, and are disruptive to operations within the enclosure.
13
ROUTINE MAINTENANCE
Estimation of Filter Condition
The negative pressure inside the box will give a good indication of filter and system condition. Set blower speed dial to the speed that gives the proper negative pressure conditions inside the box. Record both the blower speed dial setting and the pressure reading on the gauge. If box pressure begins to drop due to filter loading, increase the blower speed. To check filters, reduce the blower speed dial setting to the setting recorded when the filters were clean. Read the differential pressure gauge. When the gauge reading is one-half of the initial reading, replace filters.
Filter Replacement
When filters become contaminated or after a predetermined length of time in use, the filter should be replaced. Only experienced personnel, competent in the procedure of filter replacement should perform this procedure.
Suggested Plan for Decontamination
Activities such as filter changing, service work, moving the cabinet, or gross spillage generally require a gas decontamination procedure.
ONLY EXPERIENCED PERSONNEL COMPETENT IN THIS PROCEDURE SHOULD BE USED.
LABCONCO strongly recommends reference to the Laboratory Safety Monograph (a supplement to the NIH Guidelines for Recombinant DNA Research), published by the U.S. Department of Health, Education and Welfare, January 1979.
Reference may also be made to the Formaldehyde Decontamination booklet available from the U.S. Department of Health, Education and Welfare, Public Health Service, National Institute of Health, National Cancer Institute.
For decontamination procedures with formaldehyde, refer to the following information.
1. NCI Safety Standards for Research Involving Chemical Carcinogens. Publication No. (NIH) 76-900, 1975.
2. Carcinogen Safety Monographs, NCI Office of Research Safety.
3. “Potential Hazard of Tissue Culture Assays Arising from Carcinogenic Compounds Incompletely Removed by Washing” E. B. Sansone, J. A. Poiley, R. J. Pienta and W. B. Lebberz, III, Cancer Research 36.2455 – 1976.
4. “Recommended Decontamination Procedures for Aflatoxin”, L. Stoloff and W. Trager. Journal of the Association of Official Agricultural Chemists, 48.681 – 1965.
14
ROUTINE MAINTENANCE
Gloves
At least every three months (more often if subjected to heavy use), inspect the gloves for signs of damage or wear. For replacement, refer to page 11.
Transfer Chamber Inner and Outer Door
Periodically check the closing tension of the door latches. Closing the latch handle should require some firm force, indicating compression of the door seals. Seals, compression, and latching force can be adjusted as follows:
1. Close and latch the door.
2. Using a 1/2" open-end wrench, turn the door adjustment screw (located between the door and the latch bar) clockwise to increase seal compression and latching force. Turn screw counterclockwise to decrease compression. Refer to figure 5.
Figure 5
15
ROUTINE MAINTENANCE
Fluorescent Lamp Replacement
1. Turn the light switch off and unplug the lamp power cord from the receptacle located on the outside top of the cabinet.
2. From the top of the lamp housing, remove the two Phillips screws attaching the lamp assembly to the top edge of the window frame.
3. Remove the lamp assembly from the unit.
4. The lamp tube can then be easily removed and replaced. Replacement lamp part #12779 (fluorescent lamp) #F30T12/CW/RS may be purchased from most local lighting supply stores.
5. Re-install the lamp assembly and attach with the two Phillips screws. Plug in the power cord and switch on the lamp. Refer to figure 6.
Figure 6
16
REPLACEMENT PARTS
Figure 7
17
REPLACEMENT PARTS
Ref. No. Catalog No. Description
1 5066600 Glass Viewing Window 2 5066700 Polycarbonate Viewing Window (Optional) 3 5067200 Viewing Window Gasket (NOT SHOWN) – For Model 5065000 3A 5098400 Viewing Window Gasket (NOT SHOWN) – For Stainless Steel Model 5065500 4 5005500 Glove Kit 8-1/2 Size Neoprene Gloves (1 Pr.) Two O-Rings 5 1640500 8-1/2 Size Neoprene Gloves (1 Pr.) 6 5005501 Kit: 8-1/2 size Butyl Gloves (1 Pr.) 2 O-Rings 7 1640501 8-1/2 Size Butyl Gloves (1 Pr.) 8 5005502 Kit: 8-1/2 Size Hypalon Gloves (1 Pr.) Two O-Rings 9 1640502 8-1/2 Size Hypalon Gloves (1 Pr.) 10 5005600 Kit: 9-3/4 Neoprene Gloves (1 Pr.) Two O-Rings 11 1640600 9-3/4 Size Neoprene Gloves (1 Pr.) 12 5005601 Kit: 9-3/4 Size Butyl Gloves (1 Pr. ) Two O-Rings 13 1640601 9-3/4 Size Butyl Gloves (1 Pr.) 14 5005602 Kit: 9-3/4 Size Hypalon Gloves (1 Pr.) Two O-Rings 15 1640602 9-3/4 Size Hypalon Gloves (1 Pr.) 16 1640000 O-Rings (1 Pr.) for Gloves 17 1952500 Pressure Differential Gauge 18 1329700 Switch, 115V 18A 1301500 Switch, 230V 19 5092600 Blower Speed Controller 20 5064200 Transfer Chamber Door Gasket (1 inner, 1 outer) 20A 5098300 Transfer Chamber Door Gasket, (1 inner, 1 outer) Stainless Steel 21 1928800 Housing Filter 22 1928900 Pre-Filter 23 5065100 Filter, HEPA 24 5065300 Charcoal Filter (Optional) 25 1490700 Tube, Offset 26 1490600 Elbow – 90° 27 5094600 Motor – Blower Assembly 5109700 Motor Assembly 1327900 Relay 5094400 Circuit Board Assembly 28 5104500 Transformer Assembly (NOT SHOWN) 230V models only
18
(A) Glove Port Components (Detailed on Page 19) (B) Sealed Receptacle Components (Detailed on Page 20) (C) Fluorescent Lamp Assembly (Detailed on Page 21) (D) Inner Door Assembly (Detailed on Page 22) (E) Outer Door Assembly (Detailed on Page 23)
A. Glove Ports
Ref. No. Catalog No. Description
REPLACEMENT PARTS
1 5074201 Glove Port 2 5074301 Glove Port Mounting Ring 3 5066800 Glove Port Gasket 4 1889310 Glove Port Mounting Screw
Figure 8
19
REPLACEMENT PARTS
B. Sealed Receptacle Assembly
Ref. No. Catalog No. Description
1 5075600 Sealed Receptacle, 115V
1A 5102100 Sealed Receptacle, 230V
2 1645702 Receptacle O-Ring 3 1349100 Cupped Washer 4 1349200 Locknut
Figure 9
20
REPLACEMENT PARTS
C. Fluorescent Lamp Assembly (Complete with Bulb), No. 5072600
Ref. No. Catalog No. Description
1 1277900 Lamp Bulb #F30T12/CW/RS
(also available locally) 2 1279200 Lamp Fixture Assembly 3 5072901 Lamp Housing
Figure 11
21
REPLACEMENT PARTS
D. Inner Door Assembly, No. 5067700 (Used on 5065000)
No. 5098500 (Used on 5065500)
Catalog No.
Ref. No.
1 5102800 5097900 Inner Door Assembly 2 5064500 Door Fitting 3 1645701 Door O-Ring 4 1927000 Nut, Stainless Steel ¾-16 5 5064600 5064600 Door Adjusting Screw 6 5068600 5068600 Door Clamp 7 5085201 5085201 Inside Latch Bar Assembly 8 5074401 5074401 Inside Counterweight 9 5098100 Grab Knob
(Acrylic 5067700)
Catalog No.
(Stainless Steel 5098500)
Figure 11
Description
22
E. Outer Door Assembly, No. 5068400 (Used on 5065000)
No. 5098600 (Used on 5065500)
Catalog No.
Ref. No.
1 5102900 5097500 Outer Door Assembly 2 5064500 Door Fitting 3 1645701 Door O-Ring 4 1927000 Nut, Stainless Steel ¾-16 5 5064600 5064600 Door Adjusting Screw 6 5068600 5068600 Door Clamp 7 5085301 5085301 Outside Latch Bar Assembly 8 5074501 5074501 Outside Counterweight
(Acrylic 5068400)
Catalog No.
(Stainless Steel 5098600)
Figure 12
REPLACEMENT PARTS
Description
23
DIMENSIONAL DATA
24
ELECTRICAL SPECIFICATIONS
Electrical Specifications for both the 5065000 and 5065500 models are as follows:
115 Volts, 60 Hz, 15 Amps One 30-Watt Fluorescent Lamp Two 115 Volt, combined 9 Amp electrical outlets (front, left and right interior of cabinet) One blower motor rated 1.8 FLA, 115V
Electrical Specifications for both the 5065002 and 5065502 models are as follows:
230 Volts, 50 Hz, 10 Amps One 30-Watt Fluorescent Lamp Two 230 Volt, combined 8 Amp electrical outlets (front, left and right interior of cabinet) One blower motor rated 1.8 FLA, 115V One transformer 230/115V
25
WIRING DIAGRAMS
Model 5065000 and 5065500 – 115 VAC, 60 Hz
26
Models 5065002 and 5065502 – 230 VAC, 50 Hz
WIRING DIAGRAM
27
ACCESSORIES
Accessory Part # Description
5061600
5062000
5062001
5060400
5060500
5065300
5065200
Interior Storage Shelves
Three stainless steel storage shelves are attached to epoxy coated stainless steel upright supports. Shelves are adjustable height, measuring 6" D x 30" W x 21-1/2" H. Components may be passed through the transfer chamber and assembled within the glove box enclosure. Installation by customer.
Glove Box Mobile Base Stand
Welded steel construction, epoxy coated with black phenolic laminate top surface. 1-1/8" thick x 30" D x 60" W, height adjustable from 30" - 37". Lower shelf provides convenient storage space. Accessory includes 5" swivel casters with locking brakes.
Glove Box Stand
Same as above except furnished with adjustable leveling feet instead of casters.
Interior Glove Port Cover
12-gauge stainless steel with molded rubber self-centering gasket and spring loaded latches with adjustable closure knob. Use for internal sealing of glove port in event of worn or damaged gloves.
Exterior Glove Port Cover
12-gauge stainless steel with molded neoprene self-centering gasket. Adjustable tension knob with latching bar attaches to port mounting ring screw adaptor.
Charcoal Filter
Filter installs between exhaust HEPA filter and blower. Contains 4 lbs. of Barnebey and Stucliffe type 787 activated carbon impregnated for removal of gaseous radioisotopes.
Additional Exhaust HEPA Filter Kit
Filter installs between existing exhaust HEPA and blower. Rated 99.99% efficient on 0.3 micron particulates.
28
WARRANTY
We are committed to providing our customers with quality equipment and service after the sale. Part of this objective involves keeping you informed of changes and new product additions. We, therefore, request that you take a moment to fill out the product registration card so we many know your location as well as some of the reasons that prompted you to purchase our product.
Labconco provides a warranty on all parts and factory workmanship. The warranty includes areas of defective material and workmanship, provided such defect results from normal and proper use of the equipment.
The warranty for all Labconco products will expire one year from date of installation or two years from date of shipment from Labconco, whichever is sooner, except the following:
Purifier® Delta® Series Biological Safety Cabinets carry a three-year warranty from date of installation or four years from date of shipment from Labconco, whichever is sooner.
Carts carry a lifetime warranty.
Glassware is not warranted from breakage when dropped or mishandled.
This limited warranty covers parts and labor, but not transportation and insurance charges. In the event of a warranty claim, contact Labconco Corporation or the dealer who sold you the product. If the cause is determined to be a manufacturing fault, the dealer or Labconco Corporation will repair or replace all defective parts to restore the unit to operation. Under no circumstances shall Labconco Corporation be liable for indirect, consequential, or special damages of any kind. This statement may be altered by a specific published amendment. No individual has authorization to alter the provisions of this warranty policy or its amendments. Lamps and filters are not covered by this warranty. Damage due to corrosion or accidental breakage is not covered.
WARNING: The disposal and/or emission of substances used in connection with this equipment may be governed by various federal, state or local regulations. All users of this equipment are urged to become familiar with any regulations that apply in the user’s area concerning the dumping of waste materials in or upon water, land or air and to comply with such regulations.
29
SHIPPING CLAIMS
If a shipment is received in visibly damaged condition, be certain to make a notation on the delivering carrier’s receipt and have their agent confirm the damage on your receipt. Otherwise, the damage claim may be refused.
If concealed damage or pilferage is discovered, notify the carrier immediately and retain the entire shipment intact for inspection. Interstate Commerce Commission rules require that the claim be filed with the carrier within 15 days after delivery.
NOTE: Do not return goods. Goods returned without prior authorization will not be accepted. Labconco Corporation and its dealers are not responsible for shipping damage. Claims must be filed directly with the freight carrier by the recipient. If authorization has been received to return this product, by accepting this approval, the user assumes all responsibility and liability for biological and chemical decontamination and cleansing. Labconco reserves the right to refuse delivery of any products, which do not appear to have been properly cleaned and/or decontaminated prior to return.
30
CONTACTING LABCONCO
If you have any questions that are not addressed in this manual, or if you need technical assistance, please contact Labconco’s Sales Information Department at 1-800-821-5525, and Service Information at 1-800-522-7658 or 1-816-333-8811, between the hours of 7:00 a.m. and 6:00 p.m. Central Standard Time.
Labconco’s mailing address is:
Labconco Corporation 8811 Prospect Avenue Kansas City, Missouri 64132-2696
Fax # 816-363-0130
Visit Labconco through the Internet at:
http://www.labconco.com
or
email: labconco@labconco.com
31
32
Loading...