Labconco provides a warranty on all parts and factory workmanship. The warranty includes areas of
defective material and workmanship, provided such defect results from normal and proper use of the
equipment. Glassware is not warranted from breakage when dropped or mishandled.
The warranty for Paramount® Ductless Enclosures will expire one year from date of installation or two
years from date of shipment from Labconco, whichever is sooner. Warranty is non-transferable and
only applies to the owner (organization) of record.
This limited warranty covers parts and labor, but not transportation and insurance charges. In the
event of a warranty claim, contact Labconco Corporation or the dealer who sold you the product. If
the cause is determined to be a manufacturing fault, the dealer or Labconco Corporation will repair or
replace all defective parts to restore the unit to operation. Under no circumstances shall Labconco
Corporation be liable for indirect, consequential, or special damages of any kind. This statement may
be altered by a specific published amendment. No individual has authorization to alter the provisions
of this warranty policy or its amendments. Lamps and filters are not covered by this warranty.
Damage due to corrosion or accidental breakage is not covered.
The information contained in this manual and the accompanying products are copyrighted and all rights
reserved by Labconco Corporation. Labconco Corporation reserves the right to make periodic design
changes without obligation to notify any person or entity of such change.
Returned or Damaged Goods
Do not return goods without the prior authorization from Labconco. Unauthorized returns will not be
accepted. If your shipment was damaged in transit, you must file a claim directly with the freight
carrier. Labconco Corporation and its dealers are not responsible for shipping damages.
The United States Interstate Commerce Commission rules require that claims be filed with the delivery
carrier within fifteen (15) days of delivery.
Limitation of Liability
The disposal and/or emission of substances used in connection with this equipment may be governed by
various federal, state, or local regulations. All users of this equipment are required to become familiar
with any regulations that apply in the user’s area concerning the dumping of waste materials in or upon
water, land, or air and to comply with such regulations. Labconco Corporation is held harmless with
respect to user’s compliance with such regulations.
Contacting Labconco Corporation
If you have questions that are not addressed in this manual, or if you need technical assistance, contact
Labconco’s Customer Service Department or Labconco’s Product Service Department at 1-800-8215525 or 1-816-333-8811, between the hours of 7:00 a.m. and 6:00 p.m., Central Standard Time.
Part #6968600, Rev. E
ECO J457
T
T
AABBLLEE
O
O
FF
C
C
OONNTTEENNTTSS
CHAPTER 1: INTRODUCTION 1
CHAPTER 2: PREREQUISITES 3
Support and Stability Requirements 3
Location and Air Current Requirements 4
Exhaust and Blower Requirements 4
Electrical Requirements 6
Space Requirements 6
CHAPTER 3: GETTING STARTED 7
Unpacking the Enclosure 8
Set Up the Ductless Enclosure on a Supporting Structure and
Work Surface 8
Install Carbon Filters and/or HEPA Filters 11
Connect the Electrical Supply 12
Set Filter Life and Filter Check Timers 12
Initial Safety-First™ Vapor Sensor Warm Up Period 13
Reset or Change the Airflow Face Velocity (If Necessary) 13
Validate the Enclosure 14
Seal the Enclosure to the Work Surface 14
Connecting to the Exhaust System (If Applicable) 14
Ergonomic Sloped Front 20
Convenient Filter Replacement 20
Durable Construction 20
Fluorescent Light with Electronic Ballast 20
Utility Iris Ports 20
PRECAUTIONS 15
Maximum Visibility 21
Simple Electrical Connection 21
Safety Precautions 21
Misapplications that Could Result in a Hazardous Situation 24
CHAPTER 5: USING THE DUCTLESS ENCLOSURE 25
Information Center 25
Status Area 26
Operating the Pivoting Sash 26
Starting the Ductless Enclosure 26
The Paramount Touchpad 27
All Program Screens 28
Calibration Program Screens 31
Diagnostic Program Screens 32
User Selectable Features 33
Clock Timer Operation 33
Interval Timer Operation 33
Elapsed Timer Operation 33
Light Lamp Timer and Blower Timer 34
Blower Operation 34
Enclosure Set Up 34
Units of Measure 34
Security Lock 34
Filter Setup Display Options 35
Restoring Software Default Settings 35
Calibration 35
Diagnostic Test Operation 35
If a Check Airflow Activates 36
Resetting the Airflow Alert System 36
Working in the Paramount Ductless Enclosure 36
CHAPTER 6: MAINTAINING THE DUCTLESS ENCLOSURE 38
Routine Maintenance Schedule 38
Determination of When to Replace Carbon Filters 39
Determination of When to Replace HEPA Filters 41
Install Carbon Filters and/or HEPA Filters 42
Set Filter Life and Filter Check Timers 42
HEPA Filter Leak Test 42
Change the Airflow Face Velocity (If Necessary) 44
Initial Certification 44
Re-Certification 44
Fluorescent Light Replacement 45
Motorized Impeller Replacement 47
Main Control Board Replacement 48
Organic Sensor Board Replacement 49
CHAPTER 7: ACCESSORIZING THE DUCTLESS
ENCLOSURE 50
Filters 50
Work Surfaces 51
Work Surface with Deck Mounted Valves, Gooseneck,
and Cupsink 52
Base Stands, Accessory Shelves, Seismic Supports and
Hydraulic Lift Base Stands 53
Storage Cabinets 54
Exhaust Transitions 55
Remote Blowers 56
Exhaust Dampers 57
Utility Shelf Kits and Holders 57
Syringe Pump Kit and Detector Tubes 57
Sash Closures 58
Congratulations on the purchase of a Labconco Paramount® Ductless
Enclosure, commonly called a ductless fume hood. Unlike traditional fume
hoods, the Paramount Ductless Enclosure is considered green as it conserves energy because it requires no ducting. The Paramount Ductless Enclosure
is a self-contained work station that protects the operator from specified
vapors and/or particulates released in the work area, through the use of
carbon filters and/or HEPA filters. Typical organic solvents, formaldehyde,
acid gases, ammonia, amines, particulates, and powders are contained inside
the enclosure to protect the operator and then filtered. Purified air is drawn
from the filters and returned to the room. Vapors are adsorbed on activated
carbon filters or impregnated carbon filters and particulates/powders are
filtered by the HEPA filter.
Each Paramount Ductless Enclosure includes the Safety-First™ Vapor
Sensor that detects filter saturation and alerts the operator to replace the
filters. The early warning provided by the Safety-First Vapor Sensor
provides the user ample time (5-10% remaining filter life) to complete work
in process before changing the filters. The Safety-First Vapor Sensor
includes a primary and secondary sensor for redundant monitoring and the
sensor detects typical organic solvents, smoke particulates, ammonia gases,
formaldehyde gases, and hydrogen sulfide gases. The Paramount Ductless
Enclosure is the result of Labconco’s years of experience in the design and
manufacture of fume hoods and ductless hoods.
The Paramount Ductless Enclosure offers many unique features to enhance
safety, performance, and ergonomics. Safety is illustrated with the SafetyFirst Vapor Sensor. Performance and Safety are illustrated with validated
containment and patented aerodynamic features (U.S. Patent No. 6,461,233).
Ergonomics is illustrated with the patented inclined design (U.S. Patent No.
D538,941). To take full advantage of all features, please acquaint yourself
with this manual and keep it handy for future reference. If you are unfamiliar
with how ductless enclosures operate, please review Chapter 4: Performance Features and Safety Precautions, before you begin working.
Product Service 1-800-522-7658
Chapter 1: Introduction
2
Even if you are an experienced ductless enclosure user, please review
Chapter 5: Using the Ductless Enclosure; it describes the Paramount
Ductless Enclosure features so that you can use it efficiently and effortlessly.
The Paramount Ductless Enclosure ships without any filters. Chapter 3:
Getting Started and Chapter 6: Maintaining the Ductless Enclosure, must
be followed to properly select and install filters.
This manual and other technical information is available in PDF format
at our website: www.labconco.com.
Product Service 1-800-522-7658
3
CChhaapptteerr 22::
PPrreerreeqquuiissiitteess
Before you install the Paramount Ductless Enclosure, you need to prepare
your site for installation. A dedicated source of electrical power should be
located near the installation site to power the ductless enclosure and other
apparatus. Additionally, the enclosure should be strategically placed in the
lab to provide efficient workflow.
Carefully read this chapter to learn the requirements for your installation site:
The support and stability requirements.
The location and air current requirements.
The exhaust and blower requirements.
The electrical power requirements.
Space requirements.
Refer to Appendix B: Dimensions for complete enclosure dimensions.
Refer to Appendix C: Specifications for complete enclosure electrical and
environmental conditions, specifications and requirements.
Support and Stability Requirements
At a minimum, the supporting structure usually consists of a base cabinet or
stand and a chemically resistant work surface. Please consider the following:
Work surfaces should be of a thick rigid material that remains stable.
A bench that is rigidly mounted to the floor or fixed to the wall may
be appropriate.
Tubular stands or mobile benches with locking casters may be
appropriate.
Product Service 1-800-522-7658
Chapter 2: Prerequisites
4
Location and Air Current Requirements
The Paramount Ductless Enclosures have been designed to contain hazards
by negating typical cross drafts and turbulence within the opening. However,
it is recommended that the enclosure be placed in an area to avoid:
High traffic areas where walking might cause an air disturbance or be a
nuisance.
Overhead or wall HVAC diffusers, fans, radiators or other lab equipment
producing air currents.
Near doorways or windows that may be opened.
Exhaust and Blower Requirements
The Paramount Ductless Enclosure uses a green, 95% energy efficient,
electronically-commutated motorized (ECM) impeller to draw room air past
the operator and through the enclosure (see Appendix D for power usage by
model). This contaminated air is passed through filters and exhausted out the
top of the enclosure. The filtered exhaust air is normally recirculated into the
laboratory or may be exhausted outside with the addition of the Canopy
Exhaust Connection and remote blower listed in Chapter 7. Contact
Labconco for blower sizing assistance for exhausting to the outside.
Electrical connections are covered in Chapter 2 for this configuration.
Data for the enclosure’s exhaust volume, noise pressure and face velocity are
listed on the next page in chart form for each Paramount Ductless Enclosure
model at face velocities of 60, 80, and 100 fpm. All models are shipped to
operate at 75-80 fpm, but can be reset by the Safety Officer or certification
technician through a password protected menu.
Important Note: If the enclosure is connected directly to a house exhaust system,
an adjustable damper (or valve) must be installed to control the airflow properly.
This is equally important when a house exhaust system is controlling multiple
ductless enclosures. See Chapter 7 for information on accessories.
Product Service 1-800-522-7658
Chapter 2: Prerequisites
5
Model No.
Voltage
Enclosure
Width
Internal
Working
Height
Sash
Opening
Exterior
Depth
Exterior
Height
Face
Velocity
(fpm)
Airflow
Exhaust
Volume
(CFM)
Noise
Pressure
(dbA)
6963200
100-115V
2'
22.7"
9.4"
29"
43.7"
60
80
100
85
115
145
50
54
59
6963220
208-230V
2'
22.7"
9.4"
29"
43.7”
60
80
100
85
115
145
50
54
59
6963300
100-115V
3'
22.7"
9.4"
29"
43.7"
60
80
100
130
175
220
46
52
56
6963301
100-115V
3'
32"
12"
29"
53"
60
80
100
170
230
285
52
57
62
6963302
100-115V
3'
32" x 360°
12"
29"
53"
60
80
100
170
230
285
52
57
62
6963320
208-230V
3'
22.7"
9.4"
29"
43.7"
60
80
100
130
175
220
46
52
56
6963321
208-230V
3'
32"
12"
29"
53"
60
80
100
170
230
285
52
57
62
6963322
208-230V
3'
32" x 360°
12"
29"
53"
60
80
100
170
230
285
52
57
62
6963400
100-115V
4'
22.7"
9.4"
29"
43.7"
60
80
100
175
235
295
53
57
62
6963401
100-115V
4'
32"
12"
29"
53"
60
80
100
230
305
380
57
62
67
6963402
100-115V
4'
32" x 360°
12"
29"
53"
60
80
100
230
305
380
57
62
67
6963403
100-115V
4'
35"
12"
35.5"
56"
60
80
100
230
305
380
59
62
67
6963420
208-230V
4'
22.7"
9.4"
29"
43.7"
60
80
100
175
235
295
53
57
62
6963421
208-230V
4'
32"
12"
29"
53"
60
80
100
230
305
380
57
62
67
6963422
208-230V
4'
32" x 360°
12"
29"
53"
60
80
100
230
305
380
57
62
67
6963423
208-230V
4'
35"
12"
35.5"
56"
60
80
100
230
305
380
59
62
67
6963501
100-115V
5'
32"
9.4"
29"
53"
60
80
100
220
295
370
57
62
67
6963503
100-115V
5'
35"
9.4"
35.5"
56"
60
80
100
220
295
370
57
62
67
6963521
208-230V
5'
32"
9.4"
29"
53"
60
80
100
220
295
370
57
62
67
6963523
208-230V
5'
35"
9.4"
35.5"
56"
60
80
100
220
295
370
57
62
67
6963601
100-115V
6'
32"
12"
29"
53"
60
80
100
350
465
580
57
61
66
6963603
100-115V
6'
35"
12"
35.5"
56"
60
80
100
350
465
580
57
61
66
6963621
208-230V
6'
32"
12"
29"
53"
60
80
100
350
465
580
57
61
66
6963623
208-230V
6'
35"
12"
35.5"
56"
60
80
100
350
465
580
57
61
66
Product Service 1-800-522-7658
Chapter 2: Prerequisites
6
Electrical Requirements
Standard duplex electrical receptacles should be nearby for connecting the
Paramount Ductless Enclosure and other equipment. The enclosures include iris
pass-through ports to allow electrical cords through the back of the enclosure
without leaving a large hole for contaminants to escape. (See Appendix C for
amperage.)
There is no provision for controlling a remote blower for exhausting air from the
Paramount Ductless Enclosure. If remote blower control is required, a qualified
electrician may connect a wall-mounted power outlet switch receptacle nearby so
that the switch controls both the remote blower and the outlet. The power cord
for the Paramount is then plugged into this outlet.
Space Requirements
The dimensions for the different models are shown in the chart in Chapter 2 and
detailed in Appendix B: Dimensions. To convert inches to millimeters, multiply
by 25.4.
Product Service 1-800-522-7658
7
CChhaapptteerr 33::
GGeettttiinngg SSttaarrtteedd
Now that the site for your Paramount Ductless Enclosure is properly
prepared, you are ready to unpack, inspect, install, and validate your system.
Read this chapter to learn how to:
Unpack and move the enclosure.
Set up the enclosure with the proper supporting structure and work
surface.
Install carbon filters and/or HEPA filters.
Connect the electrical supply.
Set Filter Life and Filter Check Timers.
Reset or change the face velocity, if necessary.
Validate the enclosure.
Seal the enclosure to the work surface.
If applicable, connect to an exhaust system.
Depending upon which model you are installing, you may need common
mechanical and electrical installation tools in addition to 5/16", 3/8", 7/16",
and 1/2" wrenches, ratchets, sockets, a nut driver set, a flat-blade screwdriver,
a Phillips screwdriver, and a carpenter level to complete the instructions in
the chapter.
Note: Each enclosure model weighs between 125 to 350 lbs. (55 to 152 kg).
The shipping container allows for lifting with a mechanical lift truck or
floor jack. If you must lift the enclosure manually, follow safe-lifting
guidelines. Do not lift by the front air foil.
Product Service 1-800-522-7658
Chapter 3: Getting Started
8
Unpacking the Enclosure
We recommend that you do not remove the enclosure from its shipping
container until it is ready to be placed into its final location. Move the unit
by placing a flat, low dolly under the shipping skid, or by using a floor jack.
Carefully remove the shrink-wrap or carton on the enclosure and inspect it
for damage that may have occurred in transit. If damaged, notify the delivery
carrier immediately and retain the entire shipment intact for inspection by the
carrier.
Note: The United States Interstate Commerce Commission rules require that
claims be filed with the delivery carrier within fifteen (15) days of delivery.
Note: Do not return goods without the prior authorization of Labconco.
Unauthorized returns will not be accepted.
Note: If the enclosure was damaged in transit, you must file a claim directly
with the freight carrier. Labconco Corporation and its dealers are not
responsible for shipping damages.
Do not discard the packing material until you have checked all of the
components and tested the enclosure.
Do not move the enclosure by tilting it onto a hand truck.
Set Up the Ductless Enclosure on a Supporting
Structure and Work Surface
When installing the enclosure onto a chemical-resistant work surface or
benchtop, ensure that the structure can safely support the combined weight of
the enclosure and any related equipment. See Chapter 7: Accessorizing the
Paramount Ductless Enclosure for appropriate work surfaces, stands,
and cabinets. The work surface should be at least as wide as the enclosure
to properly support it. The front of the enclosure should be aligned within
0.3" of the front of the work surface as shown in Figure 3-1. Mounting holes
are provided in the Labconco accessory work surfaces to secure the
enclosure. RTV Silicone sealant can be applied to seal the work surface to
the enclosure.
Work Surface Specifications
The work surface should be smooth, rigid, and durable. The surface should
be non-porous and resistant to the powders, solvents and chemicals used in
conjunction with the Paramount Ductless Enclosure.
Product Service 1-800-522-7658
Chapter 3: Getting Started
9
Work Surface and Enclosure Installation
1. Level the base cabinets and the work surface. Work surface should be
placed flush with the front of the stand or base cabinet as shown in
Figure 3-1.
2. Identify and position the work surface ensuring the mounting holes
will align with the four screw locations on the base of the enclosure.
(Rear mounting holes are located close to the rear edge.)
3. Secure the work surface to the base cabinet with a structural adhesive
or silicone sealant.
4. Before moving the enclosure, insert the supplied mounting screws in
the four holes. Allow a minimum of 1/8" clearance under the head of
the screw for positioning the enclosure. For the stainless work
surfaces, use a #10-24 x 3/8" hex washer head screw (Labconco part
number 1885806).
5. Place the enclosure on the work surface and slide the rear flange and
front air foil flanges under the mounting screw heads.
6. Tighten the four screws to complete the installation.
7. If desired, apply RTV silicone to seal the work surface and enclosure.
Product Service 1-800-522-7658
Chapter 3: Getting Started
10
Figure 3-1
Paramount Ductless Enclosure Installation
Product Service 1-800-522-7658
Chapter 3: Getting Started
11
6938100
HEPA Filter, 15" x 18" x 1.5", 99.99% efficient particulate on 0.3 micron
Mixed Bed Carbon Filter, Org. Vap., Acid-Sul., Ammon, Form (25% Mix) 10 lbs.
6938205
Impregnated Carbon Filter, Radioisotope 10 lbs.
Paramount
Size
Filter
Compartments
Required
No. of Filters
Total Organic
Vapor Filter Weight
2' 1 2
15 lbs.
3' 2 4
30 lbs.
4' 3 6
45 lbs.
5' 3 6
45 lbs.
6' 4 8
60 lbs.
Install Carbon Filters and/or HEPA Filters
Refer to Figure 3-2 for a view of the filter loading compartment with the
front panel removed. The Paramount Ductless Enclosure ships without
filters that must be ordered separately. It is highly recommended that
an extra set of new filters be on hand on a continuous basis.
To remove the front panel, use a Phillips screwdriver to loosen the two
screws holding the front panel in position; slide the front panel forward and
then up to remove. The filter loading compartment is hinged and supported
by gas spring cylinders. Lift the handle up and remove the cardboard
packing. The cardboard packing will be replaced with filters that always
stack two high.
The Paramount Ductless Enclosure has seven different filter types identified
below:
All impregnated Carbon Filters, specific chemical treated, have an organic
vapor capacity of 25%. Mixed Bed Carbon Filters have an organic vapor
capacity of 50%. For example, the Impregnated Acid Filter has 25% of 10
pounds or 2.5 pounds of organic vapor capacity. Thus, a common application
using solvents with acids would require one Activated Carbon Filter for
organic vapors and one Impregnated Acid Filter for acid vapors.
Any combination of two filter types may be used. Filters are always loaded
with the gasket side up. If placing the Paramount Ductless Enclosure in a
clean room, always place the HEPA filter above the carbon filter to eliminate
any trace carbon particles from entering the clean room. Load the stacked
filters into the individual filter compartments. The Paramount model sizes,
filter compartments, and number of filters are listed below for installation
convenience.
Product Service 1-800-522-7658
Chapter 3: Getting Started
12
To install the filters, push the stacked filters to the stop in the rear of each
filter compartment. After loading the stacked filters into the compartments,
close the gas spring cylinders by pulling the handle down and compressing
the filter gaskets. Filter installation is complete.
Figure 3-2
Filter Installation
Connect the Electrical Supply
Simply connect the 115V power cord supplied to the IEC electrical supply
plug on the back of the enclosure. For 230V, the same procedure applies
except it is shipped without a plug. Install the appropriate plug for your
electrical specifications per local codes. (See Appendix C for amperage.)
Set Filter Life and Filter Check Timers
Once the filters are installed and electrical power is connected, it is necessary
to set the Filter Life Timers in the Filter Setup menus described in Chapter 5:
Using the Ductless Enclosure. Consult the separate Paramount Chemical
Guide, Appendix E: Estimated Filter Life Calculation, and/or Labconco
Specialist to determine the filter capacity. Through the use of all three
sources, set the Filter Life Timers.
Product Service 1-800-522-7658
Chapter 3: Getting Started
13
Initial Safety-First™ Vapor Sensor Warm Up Period
The Safety-First™ Vapor Sensor includes a 5V, 0.28W integrated heater and
any detectable gas that accumulates during shipment must be driven off the
organic sensor over an initial 10-30 minute warm up period with the blower
activated. Refer to the Diagnostic Program Screen for Organic Sensor on
page 32 in Chapter 5 to view the display of the millivolt levels of the Vapor
Sensor. With new filters installed, the Organic Sensor should read between
1300-1500 millivolts after the initial 10-30 minute warm up period with the
blower running. All filters should be changed upon completion of work once
the Organic Sensor sees an increase of 400-500 millivolts with the display
indicating “Replace Filters.”
Reset or Change the Airflow Face Velocity
(If Necessary)
The face velocity for the Paramount Ductless Enclosure has been preset at the
factory between 75-80 fpm. Important Note: Transport, altitude, and
temperature differences may require a reset of the “Calibration” of the
“Pressure Sensor” as outlined in Chapter 5: Using the Ductless Enclosure.
When “Calibration” of the “Pressure Sensor” is reset, it is best to try the
previously calibrated values of motor speed and airflow face velocity and
then confirm the average face velocity through the enclosure sash opening.
Adjust the motor speed and airflow face velocity as necessary after
confirmation. Average face velocity measurements are made using a
calibrated thermal anemometer and taking an average across the sash opening
at 5" directly below the sash handle.
Additionally, the same “Calibration” menu in Chapter 5 should be used to
change the face velocity to another value. For most applications, 75-80 fpm
is ideal. However, there are some requirements for 100 fpm. Furthermore,
the face velocity alarm has been preset at the factory at 60 fpm. If needed,
follow the “Calibration” menu for the “Airflow Alarm” in Chapter 5 and
adjust the alarm value between 10-20 fpm below the actual face velocity.
The Paramount Ductless Enclosure has been set up to accept the SmartFlow™ feedback system. The Smart-Flow feature adjusts the motor speed
automatically for filter pressure drop, barometric pressure, and room
temperature, and maintains constant airflow velocity at the calibrated value.
Note: At altitudes above or below 900 feet, the Paramount must be
recalibrated to account for significant pressure changes that will have an
effect on face velocity. Face velocity measurements are always made using a
calibrated thermal anemometer and taking an average across the sash
opening. Follow the “Calibration” menu for “Pressure Sensor” as outlined in
Chapter 5: Using the Ductless Enclosure for recalibration of face velocity.
Product Service 1-800-522-7658
Chapter 3: Getting Started
14
Validate the Enclosure
To determine the actual face velocity at the sash opening, airflow velocity
readings are taken. Each Paramount Ductless Enclosure has been preset at
75-80 fpm. The “average face velocity” is achieved by taking readings
across the enclosure with the readings 6" from the ends and evenly spaced
every 12". Readings are measured at 5" directly below the upper sash handle.
Refer to Chapter 2 for proper airflow volumes for your particular model.
The Paramount Ductless Enclosures have been tested at Labconco’s airflow test
facility per ASHRAE 110-1995. All enclosures achieve an “as manufactured
rating” of less than 0.05 part per million (ppm) at 4 liters per minute (lpm); AM
<0.05 (Consult Labconco for individual ratings). For particulate powder
validation, Labconco had containment testing performed to validate the
enclosures for naproxen sodium powders. The Paramount Ductless Enclosures
demonstrated excellent containment when used by an operator using excellent
technique and good containment when used by an operator using marginal
technique. While no enclosure can compensate for improper technique, these
tests confirm that the Paramount Ductless Enclosures provide a safe working
environment. For copies of these validation reports, contact Labconco at 800821-5525 or 816-333-8811.
Seal the Enclosure to the Work Surface
When the enclosure has been set in place, it can be sealed at the work surface
to prevent spilled materials from collecting under the walls. A bead of
silicone sealant is recommended to seal the enclosure to the work surface.
Connecting to the Exhaust System (If Applicable)
WARNING: The weight of the exhaust ductwork system must be
supported independently of the enclosure superstructure or damage may
occur.
The exhaust system should be installed by a qualified HVAC contractor.
Normal operations with the Paramount Ductless Enclosure are exhausted into
the room and the accessory Canopy Exhaust Connection is not required. The
Canopy Exhaust Connection aids in the further removal of chemicals beyond
the filters or applications where a higher degree of powder and particulate
removal is required. However, if desired, the Canopy Exhaust Connection
(see Chapter 7) accessory has been designed to accept 6" diameter ductwork.
Review Chapter 2 for exhaust prerequisites and review Chapter 7 for ordering
blower exhaust equipment. Consult Labconco should you require help sizing
your blower for the exhaust volume and system static pressure loss. To
ensure compatibility, the selected exhaust duct material should match the
enclosure, procedures and chemical applications.
Product Service 1-800-522-7658
15
CChhaapptteerr 44::
PPeerrffoorrm
maannccee FFeeaattuurreess aanndd
SSaaffeettyy PPrreeccaauuttiioonnss
The Paramount Ductless Enclosure is a self-contained filtered enclosure that
protects the operator from specified vapors and particulates through the
revolutionary combination of features to enhance safety, performance, and
ergonomics. Carbon filters and/or HEPA filters must be installed.
All Paramount Ductless Enclosures have the following performance features
as illustrated in Figure 4-1:
1. High Performance Containment
2. Safety-First™ Vapor Sensor
3. Smart-Flow™ Airflow Monitor
4. Quiet, Energy-Efficient ECM Impeller
5. Audible/Visual Alarms
6. Large Control Panel LCD Display
7. Ergonomic Sloped Front
8. Convenient Filter Replacement
9. Durable Construction
10. Fluorescent Light with Electronic Ballast
11. Utility Iris Ports
12. Maximum Visibility
13. Simple Electrical Connection
Product Service 1-800-522-7658
Chapter 4: Performance Features and Safety Precautions
16
Figure 4-1
Performance Features
Product Service 1-800-522-7658
Chapter 4: Performance Features and Safety Precautions
17
High Performance Containment
The Paramount Ductless Enclosures use patented technology to direct air into
and through the contaminated air chamber (U.S. Patent 6,461,233). The
containment enhancing and aerodynamic designs of the lower Clean-
Sweep™ Air Foil, upper sash foil, side air foils, upper dilution air supply,
and zoned rear perforated baffle all work in concert to produce horizontal
airflow patterns that significantly reduce chemical concentrations through the
work area (illustrated in Figure 4-2). The unique lower air foil shape and
Clean-Sweep™ openings sweep the work surface and create a constant
protective barrier from contaminants. The radiused upper sash foil includes
an open air passage directly atop the sash foil into the enclosure chamber and
directs chemical concentrations away from the sash opening. The side entry
air foils allow turbulence-free air to enter the enclosure from the sides and
allow clean air to sweep the interior walls. The upper dilution air supply
provides by-pass air from above the work surface to constantly bathe the
inside of the sash and upper chamber with clean air to reduce chemical
concentrations. The zoned rear perforated baffle directs horizontal laminar
air streams to the three zones to minimize the potential for air to roll forward
preventing contaminants from moving toward the sash opening.
Figure 4-2
Airflow Diagram
Product Service 1-800-522-7658
Chapter 4: Performance Features and Safety Precautions
18
Chemical Family
Chemical
Sensitivity Alert
Concentration
Odor
Threshold
(ppm)
Exposure
Limit TWA
(ppm)
Filter
Capacity
(%W)
Filter Type
Aldehydes & Ketones
Cyclohexanone
0.2-0.5 ppm
0.068
50
22%
OV
Mixture of Aliphatic
Hydrocarbons
Gasoline
0.3-1.0 ppm
0.3
300
11%
OV
Particulates
Cigarette smoke
0.4-1.0 ppm
N/A
N/A
N/A
OV & HEPA
Aldehydes & Ketones
Acetone
0.5-1.0 ppm
4.58
250
2%
OV
Aromatic Hydrocarbons
Toluene
0.5-1.0 ppm
0.16
100
20%
OV
Ethers
Methyl Tert-Butyl Ether
0.5-1.0 ppm
0.053
50
9%
OV
Sulfur Compounds
Hydrogen Sulfide
0.5-2.0 ppm
0.0005
10
10%
AG, Acid-Sulfur
Nitrogen Compounds
Diethylamine
1.5-2.0 ppm
0.186
10
7%
OV
Esters
Ethyl Acetate
1.5-4 ppm
0.61
400
9%
OV
Ethers
Diethyl Ether
2-4 ppm
2.29
400
4%
OV
Aldehydes
Formaldehyde
2-4 ppm, best to use
other detector methods
0.87
0.1 ceiling,
0.016
10%
FORM
Nitrogen Compounds
Ammonia Solution
(Ammonium Hydroxide)
2-5 ppm
5.75
25
10%
AM, Ammonia-
Amine
Alcohols
Ethyl Alcohol*
2.5-4 ppm
0.136
1000
1.3%
OV
Acids
Acetic Acid
5-6 ppm
0.016
10
4%
OV
Halogens
Chlorobenzene
5-8 ppm
0.741
10
20%
OV
Alcohols
Isopropyl Alcohol
8-11 ppm
22
200
7%
OV
Aliphatic Hydrocarbons
Hexane
9-15 ppm
21.9
50
11%
OV
Alcohols
Methanol*
15-25 ppm
141
200
0.1%,
very low
not recommended
Mineral Acids
Hydrochloric Acid
Not Detected, use other
detector means
0.77 5 17%
AG, Acid-Sulfur
*
Safety-First™ Vapor Sensor
The Safety-First™ Vapor Sensor detects filter saturation in the exhaust to
alert the operator to replace the filters. The early warning provided by the
Safety-First Vapor Sensor provides the user ample time (5-10% remaining
filter life) to complete work in process before replacing the filters. The early
warning signals with an intermittent beep and the LCD display message alerts
the user to replace filters. The final warning signals with a constant alarm
and the LCD display message alerts the user to replace filters. The SafetyFirst Vapor Sensor includes a primary and secondary backup sensor for
redundant monitoring. Both sensors detect typical organic vapor solvents,
smoke particulates, ammonia gases, formaldehyde gases and hydrogen
sulfide gases. The Safety-First Vapor Sensor has an optimal detection range
of 1-10 ppm. Typical concentrations tested by Labconco for chemicals in
separate chemical families are listed in order of sensor sensitivity as follows:
Important Note: Clean up procedures using alcohols or volatile
chemicals with low filter capacity could saturate the filters quickly.
Important Note: The vapor sensor does not detect mineral acid gases such
as hydrochloric acid, nitric acid, or sulfuric acid and other detector means
such as a mineral acid sensor or interval timed sampling with sampling tubes
must be used.
Chemical Safety Note: The chart above is only a guideline! Frequent
chemical testing or filter monitoring is recommended for other chemicals. If
chemical suitability is ever in question, always work below the acceptable
exposure limit/TWA to maximize both safety and filter performance.
Product Service 1-800-522-7658
Chapter 4: Performance Features and Safety Precautions
19
Smart-Flow™ Airflow Monitor
The Smart-Flow™ Airflow Monitor continuously monitors airflow and
displays face velocity. An audible/visual alarm alerts the user to low airflow
conditions. The Smart-Flow Airflow Monitor enables the motor speed to
automatically adjust for conditions such as temperature, barometric pressure,
and filter loading and maintain constant programmed airflow speed. The
Smart-Flow feedback maintains constant airflow speed through the use of
constant velocity pressure measurement accomplished with a pitot tube and
pressure sensor mounted on the control board. Transport, altitude, and
temperature differences require the airflow re-calibration steps as outlined in
Chapter 5 and 6. Use the “Calibration” menu in Chapter 5 to reset the
“Pressure Sensor” values for “motor speed” and “airflow face velocity.”
Quiet, Energy-Efficient ECM Impeller
The electronically commutated motorized (ECM) impeller is 95% energy
efficient providing “green” energy savings (See Appendix D). Because of
the high efficiency, the impeller runs extremely quiet for the comparative
amount of airflow volume. All Paramounts run at or below 62 dbA at 80 fpm
and most Paramounts run at or below 59 dbA at 80 fpm. By design, the ECM
impeller is located after the filters assuring an intrinsically safe design where
all contaminated air is filtered effectively. See Chapter 2: Prerequisites for
complete airflow exhaust volume, noise pressure, and face velocities of 60,
80 and 100 fpm. The ECM impeller has a programmable blower timer
should you desire additional “green” energy savings.
Audible/Visual Alarms
The audible/visual alarms remind the operator to check the filters with
sampling tubes or replace the filter. Filter replacement alarms occur in
audible/visual modes for both filter saturation activated by the Safety-First
Vapor Sensor or the final filter elapsed time. A low airflow alarm occurs
when the face velocity drops below programmed levels.
Large Control Panel LCD Display
The large (3" x 1.5") LCD message center provides continuous status reports
for each filter, displays airflow speed, and displays alarm messages for both
filters and airflow. The touchpad can be programmed for filter types, filter
check times, and filter replacement times. Audible/visual alert time messages
appear after the specified number of hours has elapsed. The audible/visual
alarm reminds the operator to check the filters with sampling tubes or replace
the filters. Filter replacement audible/visual alarms occur for both the SafetyFirst Vapor Sensor and final filter elapsed time.
Product Service 1-800-522-7658
Loading...
+ 56 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.