The information contained in this manual and the accompanying products are copyrighted and all rights
reserved by Labconco Corporation. Labconco Corporation reserves the right to make periodic design
changes without obligation to notify any person or entity of such change.
Warranty
Labconco provides a warranty on all parts and factory workmanship. The warranty includes areas of
defective material and workmanship, provided such defect results from normal and proper use of the
equipment.
The warranty for all Labconco products will expire one year from date of installation or two years
from date of shipment from Labconco, whichever is sooner, except the following;
• Purifier
• SteamScrubber
• Blood Drawing Chairs carry a ten-year warranty.
• Carts carry a lifetime warranty.
• Glassware is not warranted from breakage when dropped or mishandled.
®
Logic®+ Biological Safety Cabinets and PuriCare® Procedure Stations carry a fiveyear warranty from date of installation or six years from date of shipment from Labconco,
whichever is sooner.
®
& FlaskScrubber® Glassware Washers carry a two-year warranty from date
of installation or three years from date of shipment from Labconco, whichever is sooner.
This limited warranty covers parts and labor, but not transportation and insurance charges. In the
event of a warranty claim, contact Labconco Corporation or the dealer who sold you the product. If
the cause is determined to be a manufacturing fault, the dealer or Labconco Corporation will repair or
replace all defective parts to restore the unit to operation. Under no circumstances shall Labconco
Corporation be liable for indirect, consequential, or special damages of any kind. This statement may
be altered by a specific published amendment. No individual has authorization to alter the provisions
of this warranty policy or its amendments. Lamps and filters are not covered by this warranty.
e due to corrosion or accidental breakage is not covered.
Dama
Returned or Damaged Goods
Do not return goods without the prior authorization from Labconco. Unauthorized returns will not be
accepted. If your shipment was damaged in transit, you must file a claim directly with the freight
carrier. Labconco Corporation and its dealers are not responsible for shipping damages.
The United States Interstate Commerce Commission rules require that claims be filed with the delivery
carrier within fifteen (15) days of delivery.
Limitation of Liability
The disposal and/or emission of substances used in connection with this equipment may be governed by
various federal, state, or local regulations. All users of this equipment are required to become familiar
with any regulations that apply in the user’s area concerning the dumping of waste materials in or upon
water, land, or air and to comply with such regulations. Labconco Corporation is held harmless with
respect to user’s compliance with such regulations.
Contacting Labconco Corporation
If you have questions that are not addressed in this manual, or if you need technical assistance, contact
Labconco’s Customer Service Department or Labconco’s Product Service Department at 1-800-8215525 or 1-816-333-8811, between the hours of 7:00 a.m. and 6:00 p.m., Central Standard Time.
Part #3848310 Rev. C
ECO H904
Original instructions
T
AABBLLEE
T
CHAPTER 1: INTRODUCTION 1
CHAPTER 2: PREREQUISITES 2
Space Requirements 2
Clearance 2
Location Requirements 3
Exhaust Requirements 4
Electrical Requirements 4
Service Line Requirements 5
CHAPTER 3: GETTING STARTED 6
Unpacking the Biosafety Cabinet 7
Preparing the Biosafety Cabinet for Operation 7
Moving and Lifting the Biosafety Cabinet 8
Installing the Biosafety Cabinet on an Existing Work Surface 8
Installing the Biosafety Cabinet on a Labconco Base Stand 8
Telescoping Base Stands 9
Manual or Electric Hydraulic Lift Base Stands 9
SoLo Electric Hydraulic Lift Base Stands 9
Connecting the Biosafety Cabinet to Utility Service Lines 10
Optional Ventus
Optional Vacu-PassTM Cord and Cable Portal Use 12
Drain Valve Installation 13
Initial Certification 14
CHAPTER 4: PERFORMANCE FEATURES AND SAFETY
HEPA Filters 15-16
ULPA Filters 16
Laminar Airflow 17
Directional Airflow 18
Motor/Blower 19
Cabinet Air Intakes (Grilles), Ductwork & Air Balance Controls 20
Ultraviolet (UV) Lamp 20
Safety Precautions 21-22
CHAPTER 5: USING THE CABINET 23
O
O
TM
PRECAUTIONS 15
C
FF
Exhaust System Connections 11
OONNTTEENNTTSS
C
Original instructions
System Reset Switch 23
Information Center 24
Alarm Screens 25
Operating the Sliding Sash 26
Starting the Biosafety Cabinet 26
The Logic+ Touchpad 27
Navigating the Logic+ Menu Screens 28
Navigating the MyLogicTM Menu Screens 29-31
Setting the Clock 29
Configuring the Logic+ 30-31
Navigating the Settings Screens 31-34
Display Options 31-32
Startup Tone 32
RS-232 Output Rate 32
UV Settings 33-34
UV Lamp Hourmeter 33
Reset UV Lamp Hourmeter 33
Change UV Lamp Life 34
Service Menu Screens 34
Timer Operation 35
Interval Timer Operation 35
Stopwatch Timer Operation 35
If An Airflow Alert Activates 35
Resetting the Airflow Alert System 35
Working in the Biosafety Cabinet 36-38
CHAPTER 6: MAINTAINING THE CABINET 39
Routine Maintenance Schedule 39-40
Service Operations 40-44
Work Surface Removal 40
Front Grille Removal 40-41
Towel Catch Removal 41-42
Front Panel Removal & Installation 42-43
Changing the Fluorescent Lamps 43
Changing the Optional UV Lamp 43-44
Resetting a Circuit Breaker 44
Storage 44
CHAPTER 7: TROUBLESHOOTING 45-47
APPENDIX A: COMPONENTS 48-49
APPENDIX B: DIMENSIONS 50
Units of Measure 31
Security Lock 32
Original instructions
APPENDIX C: SPECIFICATIONS 51
Electrical Data 51
Motor Specifications 51
Environmental Conditions 52
APPENDIX D: ACCESSORIES 53
APPENDIX E: QUICK CHART 54
DECLARATION OF CONFORMITY 55
CAUTION – See Manual. When this symbol is on the unit it
indicates a caution that is detailed in this manual.
ATTENTION - Voir manuel. Lorsque ce symbole est allumé
l'appareil, il indique une mise en garde qui est indiqué dans
ce manuel.
This biosafety cabinet offers many unique features to enhance safety,
performance and ergonomics. To take full advantage of them, please acquaint
yourself with this manual and keep it handy for future reference. If you are
unfamiliar with how biosafety cabinets operate, please review Chapter 4: Performance Features and Safety Precautions before you begin working in
the cabinet. Even if you are an experienced biosafety cabinet user, please
review Chapter 5: Using the Cabinet; it describes the biosafety cabinet’s
features so that you can use it efficiently.
This manual and other technical information is available in PDF format
at our website: www.labconco.com.
If the unit is not operated as specified in this manual it may impair the
protection provided by the unit.
Si l'unité n'est pas utilisée comme spécifié dans ce manuel il peut
diminuer la protection fournie par l'unité.
Product Service 1-800-522-7658
1
Original instructions
CChhaapptteerr 22::
PPrreerreeqquuiissiitteess
Before you install the Logic+, you need to prepare the site for installation.
Examine the location where you intend to install the cabinet. You must be
certain that the area is level and of solid construction. In addition, a dedicated
source of electrical power must be located near the installation site.
Carefully read this chapter to learn:
• Location requirements.
• Electrical power requirements.
• Exhaust requirements.
• Service utility requirements.
• Space requirements.
Refer to Appendix C: Specifications, for complete biosafety cabinet electrical
and environmental conditions, specifications and requirements.
Space Requirements
The dimensions for the Logic+ A2 biosafety cabinets are shown in Appendix B:
Dimensions.
Clearance
A minimum clearance of at least 6 inches (150 mm) is suggested on the top
and both sides of the cabinet for service.
2
Product Service 1-800-522-7658
Original instructions
Location Requirements
Note: The biosafety cabinet should be located away from traffic patterns,
doors, fans, ventilation registers, fume hoods and any other air-handling
devices that could disrupt its airflow patterns. All windows in the room
should remain closed. Figure 2-1 shows the preferred location for the
biosafety cabinet.
Preferred location
Main Disconnect Device
(Power cord plug in back
of the electronics module).
Chapter 2: Prerequisites
Figure 2-1
Air register blocked or redirected
to prevent cabinet disruption
Alternate location
Do not position the unit so that it is difficult to operate the main
disconnect device.
Ne placez pas l'appareil de sorte qu'il est difficile de faire fonctionner le
dispositif principal de déconnexion.
Product Service 1-800-522-7658
3
Original instructions
Chapter 2: Prerequisites
Exhaust Requirements
If you intend to connect the biosafety cabinet to the optional VentusTM
Canopy Connection Kit, first examine the location to ensure that it
accommodates the cabinet’s exhaust duct. The area directly above the
cabinet’s exhaust port should be clear of structural elements, water and utility
lines, or other fixed obstructions. There should be enough clearance to
accommodate a 10-inch diameter duct. Avoid cabinet locations that require
an elbow directly above the cabinet’s exhaust connection or an excessive
number of elbows in the exhaust system. For further information about the
cabinet’s exhaust system requirements, please refer to Chapter 3: Getting Started.
Electrical Requirements
The biosafety cabinet models have the following electrical requirements:
Note: A dedicated outlet with an appropriate circuit breaker should be
located as close as possible to the right rear side of the cabinet, at a
height even with, or higher than, the top of the bench or stand. Consult
your local electrical codes for properly rated circuit breakers. For safe
operation the dedicated outlet must provide the protective earthing
ground connection to the cabinet.
Note: On 100 and 115 VAC models, both electrical outlets are protected by a
ground fault interrupter circuit (GFIC). Labconco does not recommend
plugging the biosafety cabinet into a GFIC outlet.
4
Product Service 1-800-522-7658
Original instructions
Electrical outlets in the cabinet are restricted to 5 amps
Do not use any detachable power cord that is not adequately
Service Line Requirements
Chapter 2: Prerequisites
maximum current.
Prises électriques dans l'armoire sont limitées à 5 courant
maximum ampères.
rated for the unit.
Ne pas utliser un fil électrique amovible qui n’est pas du tension
nominale de l’appareil.
All utility service lines should be ¼ inch O.D., brass, copper, or stainless
steel, and equipped with an easily accessible shut-off valve. If the service line
pressure exceeds 40 PSI, it must be equipped with a pressure regulator to
reduce the line pressure.
Note: The use of flammable gases or solvents should be avoided in the
biosafety cabinet. Open flame in the cabinet will disrupt the laminar airflow
in the cabinet and may damage the HEPA filters. Flammable gases or
solvents may reach explosive concentrations in the cabinet or ductwork. If
you feel that the procedure requires the use of an open flame or flammable
materials, contact the institution’s safety office.
The use of air or gases under high pressure should be avoided as they may
seriously disrupt the airflow patterns in the cabinet.
Product Service 1-800-522-7658
5
Original instructions
CChhaapptteerr 33::
GGeettttiinngg SSttaarrtteedd
Now that the installation is properly prepared, you are ready to inspect,
install, and certify the Logic+ biosafety cabinet. This chapter covers how to:
• Unpack and move the biosafety cabinet.
• Install the cabinet.
• Connect the electrical supply source.
• Connect the service lines.
• Connect to an exhaust system (optional).
• Arrange certification of the biosafety cabinet.
Tools required for installation the biosafety cabinet include two 1/2"
wrenches, a flat-blade screwdriver, a #2 Phillips screwdriver, and a
carpenter’s level.
Note: The biosafety cabinet models weigh between 400–700 lbs. (182-318
kg). The shipping pallet allows for lifting with a mechanical lift truck or
floor jack. If you must lift the product manually, use at least six (6) persons
and follow safe-lifting guidelines.
6
Product Service 1-800-522-7658
Original instructions
Unpacking the Biosafety Cabinet
Carefully remove the outer carton and inspect the cabinet for damage that
may have occurred in transit. If the biosafety cabinet is damaged, notify the
delivery carrier immediately and retain the entire shipment intact for
inspection by the carrier.
Note: United States Interstate Commerce Commission rules require that
claims be filed with the delivery carrier within fifteen (15) days of delivery.
Do not return goods without the prior authorization of Labconco.
Unauthorized returns will not be accepted.
If the cabinet was damaged in transit, you must file a claim directly with the
freight carrier. Labconco Corporation and its dealers are not responsible for
shipping damages.
Do not discard the carton or packing material for the biosafety cabinet until
all of the components have been checked, installed and tested.
The cabinet is secured to the pallet in two places on each side. To access the
nuts and bolts, remove the side panels by removing and keeping the two
Phillips screws on both panels. Swing the front of each panel away from the
cabinet, and lift it straight up to remove the panel from the cabinet.
Note: The side panels must be removed to access the fasteners that secure the
biosafety cabinet to the pallet. DO NOT ATTEMPT TO LIFT THE
BIOSAFETY CABINET BY THE SIDE PANELS; DAMAGE WILL
OCCUR.
Chapter 3: Getting Started
Preparing the Biosafety Cabinet for Operation
Installation instructions (Labconco P/N 1056801) are attached to the sash of
the biosafety cabinet. If these instructions are missing or unclear, contact
Product Service at 800-821-5525 or 816-333-8811. The following are located
in a box underneath the work surface:
• User’s Manual CD
• Drain Valve Assembly and fasteners
• Power Cord
• Product Registration Card
• Vacu-Pass
If you did not receive one or more of the components listed for the cabinet, or
if any of the components are damaged, contact Labconco Corporation
immediately for further instructions.
Product Service 1-800-522-7658
TM
Accessories (optional)
7
Original instructions
Chapter 3: Getting Started
Moving and Lifting the Cabinet
Move the cabinet, attached to its pallet, by using a floor jack, or a furniture
dolly underneath the unit. DO NOT move the cabinet by tilting it onto a hand
truck.
When lifting the cabinet DO NOT lift the cabinet in the middle front area of
the hull. Lifting here may bend or distort the bottom of the cabinet, causing
damage to the unit.
Installing the Biosafety Cabinet on an Existing
Work Surface
Note: The biosafety cabinet is very top heavy. Use caution when lifting or
moving it.
When installing the biosafety cabinet onto an existing work surface or benchtop,
ensure that the structure can safely support the combined weight of the cabinet
and any related equipment. The work surface should be at least as wide as the
cabinet and 31 inches (787 mm) deep to properly support the unit.
A hole or notch may be cut in the supporting surface in the right front corner
to accommodate the optional drain valve.
Installing the Cabinet on a Labconco Base Stand
Labconco offers accessory Base Stands in a variety of configurations to suit
your particular needs. Stands can be ordered with adjustable telescoping legs
or with a manually or electrically adjustable hydraulic lift.
8
Product Service 1-800-522-7658
Original instructions
Telescoping Base Stands
These stands are included with some Logic+ models, or are available
separately. The base stands for each width cabinet are listed in Table 3-1
below. An optional caster wheel kit is available (part # 3730500).
Manual or Electric Hydraulic Lift Base Stands
These base stands offer infinitely adjustable height between 25.5 and 33.5
inches (648 to 851 mm), giving a cabinet work surface height of 28.0 to 36.0
inches. The height is adjusted either by a manual (hand crank) or electric
pump that drives hydraulic rams in the legs of the stands. All of the hydraulic
stands are equipped with fixed feet, but can be converted to caster wheels
with the addition of Caster Kit #3784000. The base stands for each cabinet
size is listed in Table 3-2 below.
Chapter 3: Getting Started
Table 3-1
Width Base Stand w/Feet Model #
3' 3401003
4' 3401004
5' 3401005
6' 3401006
Width Manual Lift Stand # Electric (115V) Lift Stand # Electric (230V) Lift Stand #
3' 3780200 3780100 3780103
4' 3780201 3780101 3780104
5' 3780203 3780106 3780107
6' 3780202 3780102 3780105
Table 3-2
Note: When installing the cabinet on the hydraulic lift base stand, ensure that
the hydraulic lines and the electrical cord are clear of any obstructions before
installing the cabinet on the stand or operating the lift system.
SoLoTM Electric Hydraulic Lift Base Stands
These base stands permit the Logic+ to be lowered enough to be transferred
through a standard doorway as low as 78 inches. Casters provide mobility
and lock in place. The SoLo Stands for each Purifier Logic+ series model is
listed below.
Connecting the Biosafety Cabinet to Utility
Service Lines
Note: Some models have a solenoid valve connected to the service valve on
the right side, rear position. The solenoid prevents gas from flowing to the
service valve when the unit blower is off. It is the only service valve position
that can be fitted with a solenoid valve. Connect the gas service to the
solenoid valve.
The service lines (if any) should be connected to the tube fitting(s) on the
outside of the liner wall as shown in Figure 3-1. To install the tubing, follow
these steps:
1. Ensure that the tubing is ¼ inch O.D., soft metal, and that the end has
been completely deburred.
2. Route the tubing from the rear of the cabinet, ensuring that it will line up
with the slot in the back of the side panel. The slot is located from 8 ¾ to
11 ¼ inches (222 to 288 mm) from the bottom of the cabinet.
Note: Make sure that the tube routing will not contact any electrical wires.
DO NOT loop service line tubing within the side panels of the cabinet.
3. Make sure that the nut on the tube fitting is loose, but do not remove it.
Look inside the fitting to make sure the tube ferrule is there.
4. Push the tube into the fitting until it is properly seated. The tube will go
approximately ¾ inch (19 mm) into the fitting.
5. Tighten the tube fitting nut hand tight and then, using a 7/16-inch wrench,
tighten it at least ¾ turn more.
6. Close the service valve in the biosafety cabinet and then slowly open the
shutoff valve on the service valve. Test all fittings for leakage. Tighten
the tube nut slightly if needed.
10
Figure 3-1
Tube fitting nut
Tube fitting
Product Service 1-800-522-7658
Original instructions
Optional VentusTM Canopy Exhaust Connection
Certain applications such as working with odorous products or volatile toxic
materials will require the connection of the biosafety cabinet to an exhaust
system.
Note: The canopy connection, also referred to as a thimble or air gap
connection, allows single or multiple biohazard cabinets to be connected to
an exhaust system. During operation, the exhaust system draws all of the
cabinet’s exhaust air, plus a volume of room air (through the vent in the
canopy) into the exhaust duct. Canopy connections function as a “shock
absorber” allowing the system to function properly during changes in room
air pressure.
For information on selecting a canopy connection, go to Appendix D:
Accessories.
Chapter 3: Getting Started
Note: If the research involves the use of toxic compounds or volatile
materials, contact the facility’s safety officer or Labconco to ensure that the
biosafety cabinet and its exhaust system are compatible with the materials
you will be working with.
Product Service 1-800-522-7658
11
Original instructions
Chapter 3: Getting Started
Optional Vacu-PassTM Cord & Cable Portal Use
Note: There must be enough clearance to pass the cord or cable between the
Logic+’s exterior dress panel and any obstruction.
Note: Some Vacu-Pass components and the cord or cable passing through it
may become contaminated during use of the cabinet. Ensure all potentially
contaminated components are surface decontaminated before handling or
removal from the cabinet.
1. Remove the grommet from the liner side wall. Remove the solid sealing
plug from the body of the portal by either pressing it through from the
outside, or by carefully inserting a spatula or similar device between the
sealing plug and the body of the portal, and prying the plug out.
2. Pass the cord or cable through the body of the portal, and then through
one of the plugs that has been cut for cord or cable use, then through the
grommet, as shown in Figure 3-2.
Note: select a plug with a hole that is slightly smaller than the cord or cable,
to create a proper seal. This will also help minimize movement of the cord or
cable if it is accidentally pulled during use.
Figure 3-2
3. Position the cord or cable as it will be used in the cabinet, and then push
the plug back into the body of the portal until it seats in the portal.
Reinstall the grommet.
12
Product Service 1-800-522-7658
Original instructions
Drain Valve Installation
In order to prevent damage during shipping, the drain valve assembly has not
been installed. If desired, the valve should be installed after the cabinet is in
its final location.
To install the valve assembly, follow these steps:
Note: The work surface is heavy. Use caution when handling it.
4. Lift the work surface out of the biosafety cabinet by lifting on the knobs
at the front of the work surface. Steady the work surface while pulling it
straight out the front of the cabinet.
5. Using a putty knife, remove and discard the stainless steel cover that is
sealed over the drain mounting holes. Scrape out remaining sealant that is
around the holes.
6. Apply a light coating of silicone sealant (user supplied) to the mounting
surface of the drain assembly. Attach the drain assembly under the
bottom of the cabinet as shown in Figure 3-3. Wipe off any excess sealant
from the cabinet bottom. Ensure that the center drain hole is unobstructed.
Chapter 3: Getting Started
7. Make sure the drain valve is in the closed position.
8. Reinstall the work surface.
9. Allow the silicone sealant to cure for at least eight hours before exposing
it to liquid.
NOTE: The drain valve assembly
attaches to the underside of the cabinet
bottom.
Apply a light coat of silicone sealant to
this surface of the connector, aligning the
three holes in the connector with the three
holes in the biosafety cabinet liner.
Figure 3-3
Product Service 1-800-522-7658
13
Loading...
+ 42 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.