13.Case, Strap and Shoulder-pad assembly ...................................... 33
— 2 —
www..com
information@itm.com1.800.561.8187
1.
SAFE TESTING (READ BEFORE USING)
Electricity is dangerous and can cause injury and death. To avoid possible
electric shock, personal injury or damage of instrument, always treat it with
the greatest of respect and care. If you are not quite sure how to proceed,
stop and take advice from a qualified person.
1. This instrument must only be used by a competent and trained person and
operated in strict accordance with the instructions. KYORITSU will not
accept liability for any damage or injury caused by misuse or noncompliance with the instructions or with the safety procedures.
2. It is essential to read and to understand the safety rules contained in the
instructions. They must always be observed when using the instrument.
3. This instrument is only intended for single phase operation at 230V AC
+10%, -15% phase to earth or phase to neutral operation, and then only for
Loop, RCD and Uc testing. For use in the continuity testing and insulation
testing modes this instrument must be used ONLY on circuits
which are de-energized.
4. Verify the tester's operation by measuring a known voltage before and
after using it.
5. When conducting tests do not touch any exposed metalwork associated
with the installation. Such metalwork may become live for the duration of
the test.
6. Never open the instrument case (except for fuse and battery
replacement and in this case disconnect all leads first) because dangerous
voltages are present. Only fully trained and competent electrical engineers
should open the case. If a fault develops, return the instrument to your
distributor for inspection and repair.
7. If the overheat symbol appears in the display "
instrument from the mains supply and allow to cool down.
8. For loop impedance tests to prevent unwanted tripping during loop testing
all residual current devices(RCDs) must be taken out of the circuit and
temporarily replaced with a suitably rated MCB unit. The RCD must be
replaced after the loop test is completed.
9. If abnormal conditions of any sort are noted (such as a faulty display,
unexpected readings, broken case, cracked test leads, etc) do not use the
tester and return it to your distributor for repair.
10. For safety reasons only use accessories (test leads, probes, fuses, cases,
etc) designed to be used with this instrument and recommended by
KYORITSU. The use of other accessories is prohibited as they are unlikely to
have the correct safety features.
11. When testing, always be sure to keep your fingers behind the safety
barriers on the test leads.
12.
During testing it is possible that there may be a momentary degradation of
the reading due to the presence of excessive transients or discharges on
the electrical system under test. Should this be observed, the test must be
repeated to obtain a correct reading. If in doubt, contact your distributor.
13.
The sliding shutter on the back of the instrument is a safety device. The
instrument should not be used if it is damaged or impaired in any way, but
#
" disconnect the
— 1 —
www..com
information@itm.com1.800.561.8187
returned to your distributor for attention.
14.
Do not operate the function switch while the instrument is connected to a
circuit. If, for example, the instrument has just completed a continuity test
and an insulation test is to follow, disconnect the test leads from the circuit
before moving the function switch.
15. Do not rotate function switch when test button is depressed. If the function
switch is inadvertently moved to a new function when the test button is
depressed or in lock-down position the test in progress will be halted. To
reset, release test button and press again to restart testing on new function.
16.
THE WIRING CHECK LED (P-E, P-N) of this instrument is to protect the user
from electrical shock resulting from incorrect connection of Line and Neutral
or Line and Earth.When the Neutral and Earth conductors are incorrectly
wired, the WIRING CHECK LED function cannot identify the incorrect
connection. Other procedures and test must be conducted to check and
confirm that the wiring is correct prior to making measurement. Do not use
this instrument to check the correct wiring of the power supply. Kyoritsu will
not be held liable for any accident that may result from incorrect wiring of
the power supply line.
17.
Use a damp cloth and detergent for cleaning the instrument. Do not use
abrasives or solvents.
CAT. Ⅲ
Symbols used on the instrument
Designed to protect against transient overvoltages in a building
wiring installation (low-voltage distribution level)
protected throughout by DOUBLE INSULATION or REINFORCED
INSULATION
This instrument satisfies the marking requirement defined in the
WEEE Directive. This symbol indicates separate collection for
electrical and electronic equipment.
Caution, risk of electric
shock
Protection against wrong
connection is up to 440V
— 2 —
#
Caution (refer to accompanying
instruction manual)
Earth Ground
www..com
information@itm.com1.800.561.8187
2. INSTRUMENT LAYOUT
Fig. 1
IΔn SELECT SWITCH: Available FUNCTION NO.6, 7, 8, 9, 10
(MEMORY SELECT SWITCH)
0°/180°SELECT SWITCH: Available FUNCTION NO.4, 6, 7, 8, 9
(MEMORY RECALL SWITCH)
UL VALUE SELECT SWITCH: Available FUNCTION NO.6, 7, 8, 9
(ENTER SWITCH)
AUTO NULL SWITCH: Available FUNCTION NO.1
(MEMORY CLEAR SWITCH)
MEMORY MODE SWITCH
(MEMORY MODE EXIT SWITCH)
The switch name shown in ( ) is used in MEMORY MODE.
TestLeadwithIECConnector
LCDDISPLAY
TestLeadforContinuityandInsulationTesting
— 3 —
www..com
information@itm.com1.800.561.8187
3. FEATURES
KEW 6010B Multi-Function tester performs six functions in one instrument.
1. Continuity tester
2. Insulation resistance tester (500V/1000V)
3. Loop impedance tester
4. RCD tester
5. Uc tester
6. Mains voltage warning when operating the Loop, RCD and Uc mode.
Above test results: item1 through 5, can be saved to the internal memory; and
they can be recalled whenever necessary.
Data can be transferred from KEW6010B to PC by using MODEL8212 and
"KEW Report" (Optional accessory).
The tester is designed to Safety Standard
IEC 61010-1 CAT III (300V) Pollution degree 2, IEC 61557-1, 2, 3, 4, 6, 10.
Drip-proof construction in conformance with IP40, IEC 60529.
The instrument is supplied with:-
1. KAMP10 lead for Loop/RCD/Uc testing at socket outlets.
2. Model 7122B lead for Continuity and Insulation testing.
Continuity and insulation resistance functions have the following features: Rated current Continuity: 200mA as required in IEC 61557-4
Insulation: 1mA as required in IEC 61557-2
Live circuit warning A colour coded LED and buzzer warn if the
Continuity Null Allows automatic subtraction of test lead
Auto discharge Electric charges stored in capacitive circuits
Loop impedance, RCD and Uc testing functions have the following features: Voltage level Supply voltage is displayed when the
(Buzzer sounds when test current exceeds 200mA)
circuit under test is live.
resistance from continuity measurements.
are discharged automatically after testing by
releasing the test button.
instrument is connected to the supply until the
test button is pressed.
— 4 —
www..com
information@itm.com1.800.561.8187
Wiring check Three LEDs indicate if the wiring of the circuit
under test is correct.
Over temperature Detects overheating of the internal resistorprotection
15mA Loop Loop impedance 2000Ω range measurementmeasurement is carried out with low test current (15mA).
The current will not cause tripping out involved
DC Test Allows testing of RCDs which are sensitive to
Phase angle selector The test can selected from either the positive
UL(touch voltage limit) Select UL 25V or 50V with pressing the UL value change and Uc value select switch. Where Uc value exceeds monitoring UL, "UcH v" will be displayed without starting
And at Uc range, can monitor the Uc value.
(used for Loop tests) and of the current control
MOS-FET (used for RCD and Uc tests) displaying
a warning symbol "
further measurements.
RCD even the one with the lowest nominal
differential current (30mA).
DC fault currents.
(0°) or from the negative (180°) half-cycle
of voltage.This will prevent tripping of some
polarized RCDs when Loop testing(20Ω range
only) and may give a more accurate reading
when testing RCDs.
the RCD test.
" and automatically halting
Other features: Auto data hold Holds the displayed reading until pressed
or rotated any switches after the test is
complete and at Loop/RCD/Uc range, until the
succeeding supply is applied.
Auto power off Automatically switches the instrument off after
a period of approximately 10 minutes. The
power-off state returns to normal when the
function switch is re-set to any position.
Data memory Can store 300 measured results.
Optional Accessory
Indication Flickers while the instrument is measuring.
Model 7133B (OMA DIEC) distribution board
or
lighting circuit test lead for LOOP/RCD/
Uc testing. Data can be transferred to PC via
Optical Adapter Model 8212 (with PC software
"KEW Report")
— 5 —
www..com
information@itm.com1.800.561.8187
4. SPECIFICATION
Measurement Specification
Continuity
Open Circuit
Voltage (DC)
Greater than 6 V
Insulation Resistance
Function
500V500V+20%-0%
1000V1000V+20%-0%
Loop Impedance
Rated Voltage
(AC)
230V+10%-15%
50Hz
RCD
Function
x1/2
x110/30/100/300/500mA
FAST150mA
DC
Auto
Ramp
Short Circuit
Current
Greater than
200mA@2Ω
Open Circuit
Voltage (DC)
Rated
Voltage
(AC)
230V
+
10%
-
15%
50Hz
Rated Current
1mA or greater
1mA or greater
Nominal Test Current
at 0Ω External Loop
25A/10ms
15mA/350ms max.
Test Current
10/30/100/300/500mA
10/30/100/300mA
500mA
Goes up by 10% from 20%
to 110% of IΔn. 300ms x 10
RangeAccuracy
20/200Ω
Auto - Ranging
Up to 2Ω
Over 2Ω
RangeAccuracy
@ 500kΩ
20/200MΩ
Auto-Ranging
@ 1MΩ
RangeAccuracy
20Ω
2000Ω
@ KAMP10 Test lead
Test
Current
Duration
Test CurrentTrip Time
2000ms-8% -2%
2000ms
50ms
2000ms
200ms
+2% +8%
±10%
±4%
± (3%rdg + 4dgt)
± (3%rdg + 3dgt)
± (3%rdg + 3dgt)
± (3%rdg + 8dgt)
Accuracy
± (1%rdg
+3dgt)
Uc
Rated Voltage (AC)Test CurrentRangeAccuracy
5mA at IΔn=10mA
230V+10%-15% 50Hz
www..com
15mA at IΔn=30/100mA
150mA at IΔn=300/500mA
— 6 —
100V
+5% +15%rdg
±8dgt
information@itm.com1.800.561.8187
Voltage Measurement
Rated Voltage (AC)Measuring Range (AC)Accuracy
100-250V 50Hz100-300V3% rdg
@ Loop/RCD/Uc Range
To prevent wrong connection of test leads and to maintain safety, the dedicated
terminals used for continuity and insulation tests are automatically covered
when using the terminals for Loop impedance, RCD and Uc tests.
Typical Number of Tests (central tendency for supply voltage up to 8V at R6P)
Continuity Ranges : Approx. 700 times min. at load 1Ω
Insulation Resistance Ranges :
Approx. 1000 times min. at load 0.5MΩ(500V)
Approx. 800 times min. at load 1MΩ(1000V)
LOOP/RCD/Uc Ranges : Operational lifetime:5h (In case of continuous
duty)
Operating instrumetal uncertainty
●
Operating instrumetal uncertainty of Continuity (IEC 61557-4)/Insulation Resistance (IEC 61557-2)
Function
Continuity
lnsulation
Resistance
Range
20Ω
200Ω
500V
1000V
Measuring range to keep
operating uncertainty
0.20-19.99Ω
20.0-199.9Ω
0.50 - 199.9MΩ
1.00 - 199.9MΩ
The influencing variations used for calculating the operating instrumetal uncertainty are denoted as follows;
Temperature : 0℃ and 35℃
Supply voltage : 8V to 13.8V
Maximum percentage
operating uncertainty
±30%
● Operating instrumetal uncertainty of Loop Impedance (IEC 61557-3)
Range
20Ω
2000Ω
Measuring range to keep
operating uncertainty
0.4-19.99Ω
100-1999Ω
Maximum percentage
operating uncertainty
±30%
The influencing variations used for calculating the operating instrumetal uncertainty are denoted as follows:
Temperature : 0℃ and 35℃
Phase angle : At a phase angle 0° to 18°
System frequency : 49.5Hz to 50.5Hz
System voltage : 230V+10%-15%
Supply voltage : 8V to 13.8V
— 7 —
www..com
information@itm.com1.800.561.8187
● Operating instrumetal uncertainty of RCD (IEC 61557-6)
Function
×1/2
×1, FAST
Auto Ramp
Operating uncertainty of trip current
-10% - 0%
0% -+10%
-10% -+10%
The influencing variations used for calculating the operating error are denoted as follows:
Temperature : 0℃ and 35℃
Earth electrode Resistance (shall not exceed below) :
IΔn (mA)
10
30
100
300
500
Earth electrode resistance (Ωmax.)
UL50V
2000
600
200
130
80
UL25V
2000
600
200
65
40
System voltage : 230V+10%-15%
Supply voltage : 8V to 13.8V
Instrument dimensions:- 175 X 115 X 86mm
Instrument weight:- 840g including batteries.
Reference conditions Specifications are based on the following conditions
except where otherwise stated:-
1.Ambient temperature: 23± 5℃
2.Relative humidity 45% to 75%
3.Position: horizontal
4.AC power source 230V, 50Hz
5. DC power source: 12.0 V, ripple content 1% or
less
6. Altitude up to 2000m, Indoor use
Battery type Eight R6 or LR6 batteries.
Low battery warning "" symbol appears in the display if the battery
voltage drops below 8V.
Operating temperature 0 to +40℃ , relative humidity 80% or less, no
and humidity. condensation.
Storage temperature -20 to +60℃ , relative humidity 75% or less, no
and humidity condensation.
Surge protection transient overvoltage 4000V
Insulation resistance greater than 50MΩ at 1000V DC
(between enclosures and electrical circuits)
— 8 —
www..com
information@itm.com1.800.561.8187
LED indication of live Illuminates if there is an alternating voltage of 20V
circuit warning AC or more in the circuit under test before Continuity
or Insulation resistance tests. When DC voltage is
detected across the measuring terminal the LED
lights up.
LED indication of The P-E and P-N LEDs illuminate when the wiring of
correct polarity the circuit under test is correct. The reverse LED
" is lit when P and N are reversed.
"
Display The liquid crystal display has 3 1/2 digits with a
decimal point and units of measurement (Ω, MΩ, V,
mA and ms) relative to selected function.
Overload protectionThe continuity test circuit is protected by a 0.5 A 600
V fast acting (HRC) ceramic fuse mounted in the
battery compartment, where a spare fuse is also
stored.
The insulation resistance test circuit is protected by
a resistor against 1200 V AC for 10 seconds.
Mains Voltage On connecting test leads to the circuit under test at
Indication Loop, RCD and Uc ranges, the LCD reads VL-PE.
Indications are as follow:
Less than 100V : "Lo v"
100V~259V : voltage value and ""
260V~300V : voltage value and "Hi v"
alternately, and ""
Over than 300V : "Hi v" and ""
— 9 —
www..com
information@itm.com1.800.561.8187
Loading...
+ 24 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.