KYORITSU KEW 3121A, KEW 3123A, KEW 3122A Instruction Manual

KEW 3121A, 3122A, 3123A
KEW3121A, 3122A KEW3123A
INSTRUCTION MANUAL
HIGH VOLTAGE INSULATION TESTER

CONTENTS

Page
1Safety warnings …………………………………………………1
2Features …………………………………………………………2
3Specifications ………………………………………………3 4
4Instrument Layout ………………………………………………5
5Operating Instructions …………………………………………6
5-2Battery Check ……………………………………………6
5-5Use of Guard Terminal ……………………………………8
6Battery Replacement ……………………………………………9
7Accessories and options ………………………………………9
7-1.Metal part for Line Probe, and replacement ……………9
9Cleaning Meter Cover …………………………………………12
1

1. Safety warnings

Please read through these operating instructions before using the instrument to avoid any dangers such as electrical shock and to ensure safe operation of the instrument.
Pay particular attention to all WARNING and CAUTION in this instruction manual. WARNING indicates warnings to avoid electrical shock and CAUTION indicates cautions to avoid damage to the instrument.
#
WARNING
Th
is instrument cannot test live conductors. Ensure that the
circuit to be tested is powered off before performing a test.
Never open the Battery cover or remove the Instrument panel
during a measurement.
Confirm that the Rotary switch is at the OFF position, and then
connect the test leads.
Do not attempt to make measurement in the presence of
flammable gasses. Otherwise, the use of the inst
rument may
cause sparking, which can lead to an explosion.
Never attempt to make any measurement if any abnormal
conditions, such as a broken cover or exposed metal parts are present on the Instrument or test leads.
Set the Function switch to OFF position after use, and ensure
the Test button is released and unlocked.
#
CAUTION
Do no t exp ose the ins tru men t to direct sunligh t, high
temperature
s, humidity or dew.
Never leave the instrument in the environment of 60˚C or
higher temperatures.
Do not use the instrument if it or test lead is wet.
Remove the batteries if the instrument is to be stored and will
not be in use for a long period.
Use a damp cloth with neutral detergent or water for cleaning
the instrument. Do not use abrasives or solvents.
2

2. Features

●  Batt ery powered, the instrum ents test ins ulation up to 100000MΩ at 2500V for KEW 3121A, 200000MΩ at 5000V for KEW 3122A and 200GΩ at 5000V and 400GΩ at 10000V for KEW 3123A.
●  Suited for heavy duty electrical maintenance and servicing of industrial installations, cables, transformers, generators and switchgear where high voltage insulation tests are required.
●  Dual sc a l e s fo r low an
d hi g h ra n ges wh i c h ch a nge automatically. Colour coded scales for easy reading plus LED's that illuminate in matching colour.
●  Drip proof construction. The case is sealed with rubber
gaskets to protect internal circuit against rain.
●  Hard carrying case furnished as standard accessory. Houses
both instrument and test leads in compact form. Made of plastic, it is highly water resistant.
●  Designe
d for low power consumption. Since the maximum
current consumption is 90mA eight pieces of 1.5V SUM-3or equivalent ) permit about 6 hours of continuous operation even when the instrument is used on maximum load or twice longer on minimum load.
●  Rated output voltage is maintained down to 100MΩ for KEW 3121A, 200MΩ for KEW 3122A and 0.2GΩ/0.4GΩ for KEW 3123A. This permits accurate measurements of
low insulation
resistance.
●  Optional adapter MODEL8324 is available for connection to recorder and enables cable insulation monitoring.