KYORITSU KEW 1109, KEW 1109S Instruction Manual

INSTRUCTION MANUAL
ANALOGUE MULTIMETER
KEW 1109S

Table of Contents

Page
1. Safety Warnings ............................................................................... 1


Before Using the Multimeter .......................................................... 3
2. Features ........................................................................................... 5
3. Specifications .................................................................................... 6
4. Instrument Layout ............................................................................. 8
5. How to Read Scales ......................................................................... 9
6. Operating Instructions ..................................................................... 11
6-1 Preparation ................................................................................ 11
6-2 DC Voltage Measurements ...................................................... 12
6-3 AC Voltage Measurements ....................................................... 13
6-4 Low Frequency (dB) Measurements ........................................ 14
6-5 AC Output Voltage Measurements Using Output
Terminal ................................................................................... 15
6-6 DC Current Measurements ...................................................... 16
6-7 Resistance Measurements ....................................................... 17
6-8 Terminal Current LI, Terminal Voltage LV &
Diode Test ................................................................................ 20
6-9 Measuring I
CEO
(leakage current of transistors) ......................... 22
6-10 Measuring hFE (DC Current Amplification Factor) ................... 24
7. Fuse & Battery Replacement .......................................................... 27
 1 

1. Safety Warnings

The instrument must be used by a competent, trained p erson and operated in strict accordance with the instructions. Kyoritsu Electrical Instruments Works, Ltd will not accept liability for any damage or injury caused by misuse or non-compliance with the instructions or safety procedures. It is essential to read and understand the safety rules contained in the instructions. They must be observed when using the instrument.
This instruction manual contains warnings and safety rules
which must be observed by the user to ensure safe operation of the instrument and retain it in safe condition for a long period. Therefore, read these operating instructions thoroughly and completely before using the instrument.
#
WARNING This is a warning for the user to avoid electric shock hazard. The symbol # on the instrument means that the user must read the instructions in this manual for safe operation of the instrument.
#
CAUTION This is a caution for the user to avoid damage to the instrument.
#
WARNING
● Never open the back case during measurements.
● Never use the instrument to measure voltage higher than 250V
on an industrial power line.
● When meas ur in g a high voltage greater t ha n 250V (in s ma ll
powe r suppl y circu it a nd no t on pow er tr ansmissi on li ne), connect the test leads to the circuit under test after it is once de-energiz ed. Do not touch the wiring or test leads by hand during voltage measurements. After the measurements, turn off power to the circuit under test and disconnect the test leads.
Never test a circuit voltage higher than 250V with the rang e
se l e c t or sw itch errone o u s l y se t to one of the curren t or resistance range positions. The fuse may not protect the circuit.
 2 
● Never use the instrument in an explosive atmosphere especially
when making current measurements.
#
CAUTION
● Before makin g mea surements check that the range selec tor
switch is at a proper range position. Make sure to remove the tips of the test leads from the circuit under test when changing the measuring range during measurements.
● Do not apply voltage to the current or resistance ranges. It may
result in a fuse blow or instrument damage.
● Make certain to set the range selector switch to the OFF position
after every use.
● It is recommended that the range selector switch should be set
to the 250mA DC position to protect the instrument against the
possible shock or vibration in transit.
Note: Take good care not to make voltage measurements with
the range selector switch at the 250mA DC position. The
fuse may blow or instrument get damaged.
● Do not exp o se the ins t r ument to the di r e ct sun, ext r eme
temperature and humidity or dew fall for a long period. Also, care must be taken not to give a shock to the instrument by dropping or inadequate handling.
● Since the meter cover has been given an anti-static treatment
do not rub it strongly with a dry cloth. Where anti-static charges are present on the meter window after it has been used for a long period of time, causing the meter pointer to deflect in an abnormal way, coat its surface with anti-static chemicals for plastics or wipe it with a dilute solution of anti-static fluid as a temporary measure.
● Understanding Some of the Basics in Electrical Testing Before
Using the Multimeter
 3 
● Auxiliary Units (Prefixes) There are a number of measurement units used for multimeters. Vol t (V), am p e r e (A) an d ohm (Ω ) are mo s t widely us e d as
meas ure ment un its to indi cat e electri cal p otent ial , current a nd resistance. However, it is not always straitforward to handle these units as they sometimes too large or too small for practical use or calculation. Prefixes are therefore used as auxiliary units to simplify the usage of such measurement units.
The following table shows some of the examples:
Auxiliary Unit M k m μ n P
Read mega- kilo- milli- micro- nano- pico-
Multiply 10
6
10
3
10
-3
10
-6
10-9 10
-12
Example
2MΩ
2000kΩ
2kΩ
2000Ω
250mA
0.25A
50μA
0.05mA
200nF
0.2μF
1000pF
0.001μF
● Notes on Voltage & Current Measurements It is important to understand the basic differences between current
Voltage measurements are designed to detect potential difference
between two points. Make certain that the multimeter is connected in parallel with the circuit under test.
Current measurements are intended to monitor the consumption of
current in the circuit resulting from the application of voltage. Make sure to connect the multimeter in series with the circuit under test.
Generally speaking, the internal resistance of a voltmeter should
preferably be larger, while that of an ammeter should be smaller. If the multimeter is erroneously connected in parallel with a circuit for current measurements, the likely result will be the flow of excessive current and aubsequent damage to the fuses and other electronic components. To avoid such a potential danger it is necessary to have a good understanding of voltage and current measurements.
 4 
Voltage Measurements Current Measurements
(Parallel Connection) (Series Connection)
Measuring voltage at both ends of a resistor.
M e a su r in g c u r r e n t b ei ng consumed by a resistor.
I
M
.......... Meter Current
IR .......... Load Current
E .......... Power Source Voltage
RM ........ Internal Resistance of Multimeter
Fig. 1
IR+I
M  IR
> > I
M
R
M・IR
< < E
R
M・IR
I
M
I
R
I
R
Power Source
Power Source
E E Load Load
R
M
=
DC.V Range DC.mA Range
COM
COM
RM = 0
V
mA
mA
Battery (Power Source)
Battery (Power Source)
Battery (Power Source)
Battery (Power Source)
Resistor
(Load)
Resistor
(Load)
 5 

2. Features

● Mirrored scale for easy and accurate reading.
● 19 measuring ranges for a wide scope of application.
● OUTPUT terminal to cut off the DC component of AC voltage being
measured.
● hFE scale for transistor checking.
● dB scale (−10 - +62dB).
● Safety designed input terminals and test leads.
 6 

3. Specifications

Functions Measuring Ranges Accuracy
DC Voltage (7 ranges)
0-0.1 V 0-0.5V 0-2.5V
0-10V (20kΩ/V)
0-50V 0-250V 0-1000V
±3% of full scale
0-10V
AC Voltage (4 ranges)
0-50V (9kΩ/V)
0-250V 0-1000V
±3% of full scale
DC Current (4 ranges)
0-50μA
0-2.5mA
0-25mA
0-250mA
(
terminal voltage
)
approx. 100mV
(
terminal voltage
)
approx. 100mV
(
terminal voltage
)
approx. 150mV
(
terminal voltage
)
approx. 550mV
±3% of full scale
Resistance (4 ranges)
0-2kΩ (20Ω mid-scale) 0-20kΩ (200Ω mid-scale) 0-2MΩ (20kΩ mid-scale) 0-20MΩ (200kΩ mid-scale)
±3% of scale length
Current across Terminals at Resistance Range (LI) (4 ranges)
0-150mA (x1 range) 0-15mA (x10 range) 0-150μA (x1k range)
±5% of scale length (Battery voltage at 3V)
0-60μA (x10k range)
(Battery voltage at 12V)
Voltage across Terminals at Resistance Ranges (LV) (4 ranges)
0-3V (x1 , x10, x1 k ranges)
±5% of scale length (Battery voltage at 3V)
0-12V (x10k)
(Battery voltage at 12 V)
 7 
Low Frequency Output Using OUTPUT Terminal
0-10V 0-50V 0-250V 0-1000V
Refer to frequency characteristic chart
Low Frequency Output (dB) (4 ranges)
10V AC
10 - +22dB 50V AC +4 - +36dB 250V AC +18 - +50dB 1000V AC +30 - +62dB (0 dB = 0.775V(1mW) across a 600Ω line)
Refer to frequency characteristic chart
DC Current Amplication Factor (hFE) (1 range)
hFE: 0-1000 (atΩ× 10 range)
±3% of scale length
Dimensions: 100 (W) x 150 (L) x 47 (D)mm Weight: Approx. 330g Accessories: Test Leads (Model 7066A) - 1 set
F250V 0.5A fuse - 2 pcs. (including one spare fuse
also installed into the housing case)
1.5V Battery Type R-6P, SUM-3, AA or equivalent ­2 pcs.
(Installed into the instrument) 9V Battery Type 6F22, 006P PP3, or equivallent
- 1 pce (installed into the instrument)
Instruction Manual - 1 copy
 8 

4. Instrument Layout

Meter Pointer
Meter Zero Adjust Screw
OUTPUT Terminal
+(P) Terminal
-COM (N) Terminal
Front Panel
Mirrored Scale Plate
Meter Window
Housing Case
Ohm Zero Adjust Knob
Range Selector Switch
⑦ ⑧
 9 

5. How to Read Scales

Measuring Ranges
Scales
H o w t o Read Scales
input Terminals
DC Voltage (7 ranges)
0.1V
0.5V
2.5V 10V 50V 250V 1000V
B 10 B50 B 250 B 10 B50 B250 B10
X0.01 X0.01 X0.01 X1 X1 X1 X100
+&
COM
AC Voltage (4 ranges)
10V 50V 250V 1000V
C10
B50 B250 B10
X1 X1 X1 X100
+&
COM
DC Current (4 ranges)
50μA
2.5mA 25mA 250mA
B50 B250 B250 B250
X 1 X0.01 X0.1 X1
+&
COM
Resistance (4 ranges)
X1Ω x10Ω x1kΩ X10kΩ
A 0-2k A0-2k A0-2k A0-2k
X1 X10
X1k X10k
+&
COM
Current across Terminals (LI)(4 ranges)
X1Ω X10Ω X1kΩ X10kΩ
E15 E15 E15 E15
X10(mA) X1(mA) X 10(μA) X 4(μA)
+&−COM
Voltage across Terminals (LV)(4 ranges)
x1Ω X10Ω X1kΩ X10kΩ
F3 F3 F3 F3
X1(V) X1(V) X1(V) X4(V)
+&−COM
Loading...
+ 25 hidden pages