CAT.II Primary electrical circuits of equipment connected to an AC
electrical outlet by a power cord
CAT.III Primary electrical circuits of the equipment connected
directly to the distribution panel, and feeders from the
.
distribution panel to outlets.
Protected throughout by DOUBLE INSULATION or
REINFORCED INSULATION
User must refer to the explanations in the instruction manual.
Earth Ground
1. Safety Warnings
This instrument has been designed, manufactured and tested according
to following standards, and delivered in the best condition after passing
quality control tests.
This instruction manual contains warnings and safety rules which
have to be observed by the user to ensure safe operation of the
instrument and to maintain it in safe condition. Therefore, read
through these operating instructions before using the instrument.
DANGER
● Read through and understand the instructions contained in this
manual before using the instrument.
● Keep the manual at hand to enable quick reference whenever
necessary.
● The instrument is to be used only in its intended applications.
● Understand and follow all the safety instructions contained in the
manual.
It is essential that the above instructions are adhered to. Failure to
follow the above instructions may cause injury, instrument damage
and/or damage to equipment under test.
● The symbol indicated on the instrument means that the user must
refer to the related parts in the manual for safe operation of the
instrument. It is essential to read the instructions wherever the symbol
appears in the manual.
DANGER is reserved for conditions and actions that are likely to
cause serious or fatal injury.
WARNING is reserved for conditions and actions that can cause
serious or fatal injury.
CAUTION is reserved for conditions and actions that can cause
injury or instrument damage.
1
DANGER
● This instrument is designed to measure the earth-to-line voltage
90 ~ 264V and the line-to-line voltage up to 440V (50/60Hz). Do not
exceed the maximum allowable input of any measuring range.
● Do not attempt to make measurement in the presence of
flammable gasses. Otherwise, the use of the instrument may cause
sparking, which can lead to an explosion.
● Keep your fingers behind the safety barrier on the test leads.
● Set the Function Switch to any desirable Range before making
a measurement. Do not power on the instrument with it being
connected to the live circuit.
● Never attempt to use the instrument if its surface or your hand are
wet.
● Never open the Battery Cover during a measurement.
● Verify proper operation on a known source before use or taking
action as a result of the indication.
WARNING
● Never attempt to make any measurement if any abnormal
conditions, such as a broken cover or exposed metal parts are
present on the Instrument and test leads.
● Do not install substitute parts or make any modification to the
instrument. Return the instrument to your local KYORITSU
distributor for repair or re-calibration in case of suspected faulty
operation.
● Set the Function Switch to the OFF position when removing the
Battery Cover for battery replacement.
● If the overheat symbol “ ” appears on the display, disconnect the
instrument from the measuring point and allow to cool down.
CAUTION
● Firmly insert the plugs of test leads to the appropriate terminals.
● Set the Function switch to the OFF position after use, and remove
the batteries if the instrument is to be stored and will not be in use
for a long period.
● Use a damp cloth with neutral detergent for cleaning the instrument.
Do not use abrasives or solvents.
● Do not store the instrument if it is wet.
2
2. Procedure of removing Cover
KEW5410 has a dedicated Cover to protect against impacts from the
outside and prevent the operation part, the LCD and the Connector Block
from becoming dirty. The Cover can be detached and put on the
backside of the main body during measurement.
2-1 Method of removing the Cover
Slide and pull the Cover
in the direction of an arrow.
Fig.1
2-2 Method of storing the Cover
Turn the Cover, slide and
push it in the direction of
arrow.
Fig.2
3
3. Feature
This instrument is a digital RCD Tester to measure the trip time and trip
out current of RCDs. It also equips the function to measure the voltage.
● Measurement of RCD trip time
Conducting testing of rated residual non-operating currents at x 1/2
Range, measuring RCD trip time at x1 and x5 Ranges.
● Measurement of trip out current
Measuring trip out current by varying current automatically.
● Remote Test
Enabling a user to hold the Test Leads with his both hands by
locking the Test Button.
● Voltage Measurement
Carrying out a constant measurement of voltage in the stand-by mode
at each Range.
● Auto-detection of Contact voltage
Detecting the voltage to earth of Earth electrodes or Protective
conductors during RCD test – when applying test currents – at
measurement using EARTH in order to prevent electrical shocks
caused by the damaged earth.
Measurement will be ceased at AC50V or more.
● Dust- and Water-proof
Dust- and Water-proof construction (designed to IEC60529 IP54)
● Backlight
Facilitating working at dimly illuminated locations.
4
4. Specification
Measuring range and accuracy
●
(23ºC±5ºC, relative humidity 75% or less)
Range
x 5
x 1
x 1 / 2
AUTO
RAMP
(mA)
Rated
Voltage
100V±10%
200V+32%
/-10%
400V±10%
50 / 60Hz
Test current
IΔn
10 / 30 / 50
/ 100mA
10 / 30 / 50
/ 100 / 200
/ 500mA
10 / 30 / 50
/ 100 / 200
/ 500mA
Measuring
range
Testing time
0ms ~ 200ms
Testing time
0ms ~ 2000ms
Testing time
0ms ~ 2000ms
40%~ 110% of IΔn
(goes up by 5%)
Testing time
300ms x 15steps
Accuracy
Trip Time
±(1%rdg+3dgt)
Test Current
+2% ~ +8%
Trip Time
±(1%rdg+3dgt)
Test Current
+2% ~ +8%
Trip Time
±(1%rdg+3dgt)
Test Current
-8% ~ -2%
Test Current
at each step
-4% ~ +4%
* Only the RCD type G (without trip out time-delay) can be tested at
Auto Ramp Test; type S (time-delay) cannot be tested.
● Used location : Altitude up to 2000m, indoor use
Operating temperature : 0ºC ~ 40ºC, relative humidity 85%
●
& humidity (no condensation)
●Storage temperature : -20ºC ~ 60ºC, relative humidity 85%
& humidity (no condensation)
●Withstand Voltage : AC3700V / 1 min
(between electrical circuit and enclosure)
●Insulation resistance : 50MΩ or more / 1000V
(between electrical circuit and enclosure)
●Sleep Function : 1. Automatically enters Sleep mode in 3 min
after the last switch operation (current
consumption 75uA). This function doesn’t
work at voltage measurements. To exit
from the Sleep mode, set the Function
switch to OFF position once, and re-set it
to the Range at which a measurement to
be conducted.
2. Backlight turns off in 1 min after it lights
up.
● Dimension : 186mm x 167mm x 89mm
● Weight : 965g
● Power Source : DC12V / Size AA battery R6P(SUM-3)
x 8pcs
●Possible number of : 1200 times or more
Measurements measure every 30sec at x1/2 Range,
IΔn =100mA)
●Accessories : Instruction manual x 1pce
Strap belt x 1pce
Test lead M7128 x 1set (red & black cords)
Test lead with alligator clip M7129 x 1 set
Cord case x 1pce
Long pin M8017 x 2pcs
Size AA battery R6P(SUM-3) x 8pcs
6
Loading...
+ 16 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.