KYORITSU 5020 Instruction Manual

INSTRUCTION MANUAL
KEW LOGGER
SERIES
Current Logger KEW 5010
Current&Voltage Logger
KEW 5020
Introduction
Thank you for purchasing KEW LOGGER 5010/5020.
󱾀
Follow the procedure below and set time on the instrument before use.
1) Install the KEW LOG SOFT 2 and the USB Driver according to the instructions written in the Install Manual for KEW LOG Soft 2.
2) Run KEW LOG SOFT 2 once install is completed. (Refer to 4. Start
KEW LOG SOFT 2 in the Install Manual.)
3) Confirm that the LOGGER and a PC is firmly connected with a USB
Cable and click Time Synchronizing.
4) Confirm that the Logger connected with the PC is listed on the
List of detected loggers. (The USB Driver might not be installed correctly when a message No Logger is detected. is displayed. Refer to the backside of the Cautions for installing USB Driver install or 6. Trouble shooting (P.8) of the Install Manual and re-install the USB Driver.)
5) Click Time Synchronizing Button while the LOGGER is powered on and checks are marked in the box for each connected LOGGER to set time.
󱾀
Read this manual and do your desirable settings. (Setting should be done by using KEW LOG Soft 2. Settings for some Recording modes can be done on the LOGGER.
󱾀
Read through 1. Safety Warnings in this manual before using the LOGGER in order to ensure safe operation of the instrument.
Contents
1. Safety warnings
…………………………………………………………
1
2. Features
…………………………………………………………………
4
3. Instrument layout
………………………………………………………
5
3-1) Panel
………………………………………………………………
5
3-2) LCD
…………………………………………………………………
5
3-3) Displayed message
………………………………………………
7
3-4) Function of Button
…………………………………………………
8
3-5) Range/ Filter function
……………………………………………
9
4. Recording procedures
…………………………………………………
10
Step1: Start-up
………………………………………………………
11
Step2: Confirmation of set value
……………………………………
12
Step3: Preparation before a recording
………………………………
13
Step4: Start of recording
……………………………………………
15
Step5: Stop of recording
………………………………………………
17
5. Recording modes and conditions
………………………………………
18
Normal recording mode
………………………………………………
19
Trigger recording mode
………………………………………………
22
Capture recording mode
………………………………………………
24
Power Quality Analysis mode (only 5020)
……………………………
24
6. Recording modes
………………………………………………………
25
Flashing Detection LED
…………………………………………………
28
7. Simplified Power Integration Function
…………………………………
31
(available on KEW LOG Soft 2)
8. Other settings (Setting2)
………………………………………………
34
9. Confirmation of recorded data (CALL)
…………………………………
39
10. Data transfer to PC
……………………………………………………
43
11. Battery replacement
……………………………………………………
45
12. Auto-power off Function & External power supply
……………………
46
13.Trouble-shooting
…………………………………………………………
47
14.Specification
……………………………………………………………
49
1
1. Safety Warnings
This in st rument has been des igned, manufac tu red and te st ed according to IEC 61010: Safety requirements for Electronic Measuring appar atus, and del ivered in the best condition after passed the inspection. This instruction manual contains warnings and safety rules which must be observed by the user to ensure safe operation of the instrument and retain it in safe condition. Therefore, read through these operating instructions before using the instrument.
WARNING
󱾀
Read through and understand the instructions contained in this manual before using the instrument.
󱾀
Keep the manual at hand to enable quick reference whenever necessary.
󱾀
The instrument is to be used only in its intended applications.
󱾀
Understand and follow all the safety instructions contained in the
manual. It is essential that the above instructions are adhered to. Failure to follow the above instructions may cause injury, instrument damage and / or damage to equipment under test.
The symbol
indicated on the instrument means that the user must
refer to the related par ts in the manual for safe operation of the instrument. It is essential to read the instructions wherever the
symbol appears in the manual.
DANGER is reserved for conditions and actions that are likely
to cause serious or fatal injury.
WARNING is reserved for conditions and actions that can cause
serious or fatal injury.
CAUTION is reserved for conditions and actions that can cause
injury or instrument damage.
2
DANGER
󱾀
Never make measurement on a circuit in which voltage over AC300V exists.
󱾀
Do n ot m ake mea sur em ent w hi l e thun der i s rumb lin g. Stop measurement immediately and disconnect the instrument from the object under test.
󱾀
Do not attempt to make measurement in the presence of flammable gasses. Otherwise, the use of the instrument may cause sparking, which can lead to an explosion.
󱾀
Transformer jaw tips are designed not to short the circuit under test. If equipment under test has exposed conductive parts, however, extra precaution should be taken to minimize the possibility of shorting.
󱾀
Never attempt to use the instrument if its surface or your hand is wet.
󱾀
Do not exceed the maximum allowable input of any measuring range.
󱾀
Never open the Battery Cover during a measurement.
󱾀
Verify proper operation on a known source before use or taking action as a result of the indication of the device.
WARNING
󱾀
Never attempt to make measurement if any abnormal conditions, such as broken case and exposed metal parts are found on the instrument.
󱾀
Do not install substitut e pa rt s or mak e any modi ficat ion to the instrument. For repair or re-calibration, return the instrument to your local KEW distributor from where it was purchased.
󱾀
Do not try to replace the batteries if the surface of the instrument is wet.
󱾀
Disconnect all the cords and cables from the object under test and power off the instrument before opening the Battery Cover for Battery replacement.
CAUTION
󱾀
Put the instrument on a stable place where is free from vibrations or shocks.
󱾀
Keep away Floppy Disks, Mag Cards, PCs and Displays from the magnet, which is attached to the backside of the instrument.
󱾀
Do not expose the instrument to the direct sun, high temperature and humidity or dewfall.
󱾀
Be sure to power off the instrument after use. When the instrument will not be in use for a long period, place it in storage after removing the batteries.
󱾀
Use a cloth dipped in water or neutral detergent for cleaning the instrument. Do not use abrasives or solvents.
3
Safety symbols
Refer to the instructions in the manual.
󱽩
Indicates a Instrument with double or reinforced insulation Indicates that this instrument can clamp on live bare conductors
when the voltage to be tested is below Circuit – Ground-to-Earth voltage against the indicated Measurement Category.
Indicates AC
Indicates DC
󱾀
Measurement categories(Over-voltage categories)
To ensure safe operation of measuring instruments, IEC 61010 establishes safety standards for various electrical environments, categorized as CAT.I to CAT.IV, and called measurement categories. Higher-numbered categories correspond to electrical environments with greater momentary energy, so a measuring instrument designed for CAT.III environments can endure greater momentary energy than one designed for CAT.II. CAT.I : Secondary electrical circuits connected to an AC electrical
outlet through a transformer or similar device.
CAT.II : Primary electrical circuits of equipment connected to an AC
electrical outlet by a power cord.
CAT.III : Primary electrical circuits of the equipment connected directly to
the distribution panel, and feeders from the distribution panel to outlets.
CAT.IV : The circuit from the service drop to the service entrance, and
to the power meter and primary over-current protection device (distribution panel).
4
2. Features
󱾀
KEW 5020 is a Data Logger capable of measuring Leakage Current, Load Current and Voltage. (KEW5010 is for measuring Leakage Current and Load Current).
󱾀
Following sensors are used for measuring current & voltage. * Leak/Load Current Clamp Sensor KEW 8146/8147/8148 * Load Current Clamp Sensor KEW 8121/8122/8123 * Voltage Sensor KEW 8309 (for KEW5020)
󱾀
TRMS measurement & recording of AC Current (50/60Hz) & AC Voltage (50/60Hz, only on KEW5020)
󱾀
LED to indicate pre-set current/ voltage value is exceeded. (Trigger/ Capture Recording, Power Quality Analysis modes)
󱾀
Can store 60,000 data when using 1ch, and when using all 3ch, can store 20,000 data at each channel. (Normal recording mode)
󱾀
Data will not be lost at battery replacement or at low battery voltage as it is stored in nonvolatile memory.
󱾀
Long time recording is available by a use of external AC Adaptor (optional parts) Installed ba tte rie s enables backup at a te mpo ral ly electricity failure. Size AA Alkaline Battery: about 10-day measurement.
󱾀
Can transfer the recorded data to PC via USB cable.
󱾀
Protected throughout by double (reinforced) insulation
󱽩
󱾀
This instrument provides 3 Recording modes and Power Quality Analysis mode (only KEW5020). Various insulation monitoring are available with 4 recording modes. Understand characteristic of each recording mode and select an appropriate mode.
5
    ◆
Detection LEDs P.28
    ◆
USB P.43
    ◆
Port for External power supply P.46
3. Instrument layout
3-1) Panel
3-2) LCD
6
Mark Details
Channel number (CH No.) : Selected Channel number
is displayed.
Memory Block Number : Memory block (No. 1 to
3) in use is indicated. (P.
38)
Sensor Mark : Displayed on the CH No . to Indicate the
connected Sensor.
Clock Mark : Indicates time
Timer Mark : Timer function is activated. (Stand-by till the preset time.)
Recording Mark : Recording is being performed.
Auto-power-off is disabled. (Instrument won't be off automatically.)
Battery Mark : Indicates battery voltage in 4-level
Recording mode
Sub-indication : Indicates Menu items
Date : Month, Day
Main indication : Di splays m ea su re d, pres et and
recording values
Guide mark : Items can be switched with /
button.
One-time : Recording stops when memory becomes full
Endless : Overwrite the old data, and store the latest data. Range Hold/ Auto Range mark :
Status of the selected Range
Filter mark : Indicates Filter function is activated.
7
3-3) Displayed message
Message Meaning
Sensor is not connected.
Over-range
Menu: Setting1 (SET.1) P.18
View or change the recording mode/ condition.
Menu: Setting2 (SET.2) P.34
View or change the Location information and
auto-power-off function. Menu: Recall(CALL) P.39 Recorded qty., Max value reference, Recall
Normal recording mode
Trigger recording mode
Capture recording mode
Power Quality Analysis mode [KEW5020 only]
PC data in transit
Warning of memory clear
Error (improper Sensor is connected under Power Quality Analysis mode/ KEW5020 only) (P.47) All the Memory block are occupied. (P.16) Move the data and clear the memory. (P.34)
Sub-indication Sensor Type
Load current clamp sensor
Voltage Sensor
Leakage & Load current clamp sensor
Leakage current clamp sensor
8
3-4) Function of Button
Power ON / OFF
Menu mode
Recording/ Measurement mode
Button Power ON Power OFF
Pressing at least 1 sec (while the LOGGER is off)
Pressing at least 1 sec (except for recording status)
Button Function
Enter into Menu mode
Start and stop recording
Switch channels
Switch Range and Filter (P.9
Button Menu Setting Change (flashing)
Select Menu Setting change, Enter
Back Cancel
Switch Menu item Increase number
Switch Menu item Decrease number
9
3-5) Range/ Filter Function
Range Configuration Range configuration varies depending on each connected Sensor. Refer to "12. Specification" (P.45) in this manual.
Auto-ranging Auto-ran ging is available only in Normal Recording mo de. It isn't available on a Sensor providing one single range. (e.g. Voltage Sensor "KEW8309" for KEW5020)
Range Hold Select the ranges with
mark to fix measurement range under
Normal Recording mode.
Measurement range in Trigger/ Capture Recording mode is decided
depending on a detection level. Switching of measurement ranges
with the button isn't possible.
Filter Function Lowpass Filter works when Filter Function is ON
mark lights up) and filter out the frequency in higher harmonics band. (Cutoff Frequency: approx 160Hz, attenuation rate -24dB)
Range/ Filter Function for each Recording mode
Recording mode Function
Normal recording
* Switching ranges, On/Off Filter * Auto-ranging is available.
Trigger/ Capture
recording
* On/Off Filter * Range corresponds to detection level * Auto-ranging isn't available.
Power Quality Analysis
[KEW5020 only]
* On/Off Filter * Auto-ranging isn't available.
10
Following explains the flow of operation: through preparation to the stop of recording.
P.11
P.12
P.13
P.15
P.17
* The recorded data can be viewed either by the following two methods. (1) On a PC: Follow the instruction described at "10. Data transfer to
PC" in this manual (P.43) for data transfer, and Help for PC Software "KEW LOG Soft 2" for operation method.
(2) On the instrument: Refer to Confirmation of recorded data (CALL)
(P.39)
* Press the
button at least 1 sec. to power off the instrument.
(This is because to prevent operational error.)
4. Recording procedures
Step1: Start-up
Step2: Confirmation of the set value
Step3: Preparation before recording
Step4: Start of recoding
Step5: Stop of recording
Se lect the approp ri ate sensor, and connect it to the instrument.
Confirm the recording mode.
Install the instrument and do setups for each channel.
Start recording.
Stop recording.
11
1. Press the button and power off th e instrument. Connect a Clamp/ Voltage Sensor to CH1 (for multiple connection, starting from CH1) while the instrument is being off. Connect them firmly with careful attention to the orientation of the connector.
2. Press the button at least 1 sec. to power on the instrument. Release the button when all indications are displayed on the LCD. First , ti me is disp la yed , and then sensors being connected are displayed.
If incorrect time is displayed whenever
powering on the instrument, internal battery for clock may be exhausted. In this case, send back the instrument to your local KEW distributor for repair.
3. Check the connected Sensors.
Sensor (P.7) and the full-scale value are displayed.
(beginning from CH1)
Sensor mark is displayed on the CH No,. to which a Sensor is connected .
Only CH1 is used in Capture REC mode.
Voltage sensor is connected to CH1 in PQA REC mode (only on KEW5020).
Voltage sensor cannot be used with KEW5010.
Power off the instrument and repeat steps from 1. to re-connect a Sensor.
Step1: Start-up
12
4. Can make measure men t right after powering on the instrument.
* Each time pressing the button,
CH1 to CH3 switches. When (non-connect) is displayed on the LC D, a se ns or is not connected to the sele ct ed chann el ; or the connection is incorrect.
Check the mark indicating the selected recording mode. Refer to 5. Recording
modes and conditions in this manual
to ch a nge th e re c o rd i ng mo d e or condit ion (Recording interval / Preset current)
Recording mode Details Refer to
Normal
Perform s continuous me asuremen t/ recording at the pre-set intervals. (15-kind: 1 sec. to 60min.)
P.19
Trigger
Records 8 data (0.8 sec.) in total with time information when pre-set detection level is exceeded.
P.22
Capture
Record s 10 (50Hz) to 12 waveforms (60Hz) when preset detection level is exceeded. (Graph can be viewed on PC)
P.24
(PC setting)
Power Quality Analysis
(only 5020)
Records date and time information when preset Swell, Dip and Short interruption value are exceeded.
P.24
(PC setting)
Step2: Confirmation of preset value
13
1. Clamp on the measured object and fix
the Sensor. Firmly fix it so as not to
easily come off.
Care should be taken, when connecting
the Voltage Sensor designed only for
KE W5020, n ot to shor t the object
under test.
2. Fix the instrument.
1) Fix the instrument with magnet :
Can fix the instrument to metallic plate
with the magnet on its backside.
2) Hang the instrument on hook :
Can fix the instrument with a hook or
screw by using the hooking hole on the
top of the instrument.
(Firmly fix it so as not to easily come
off.)
3. Press the button to switch the
di sp lay of mea su re d val ue am on g
CH1 to 3. Confirm a proper Sensor is
connected to an appropriate CH.
Step3: Preparation before recording
14
4. Check a Range/ Filter Press the button to check their on/off
status.
Press the button to change them.
Independ ent setting of Range & Filter is available on each CH.
Point
At Normal Rec ording mode, Range and Filte r sw itches each time pressing the
button.
Auto-ranging is also available.
At Trigger/Capture/Power Quality Analysis mo des, only Fi lt er on/ of f is available . Range is selected automa tically based on the detection level (preset value for Trigger/ Capture Recording). See Setting1 "SEt.1".
Se tting ch ange isn 't availab le durin g recording; setting should be done prior to a recording.
mark m ean s Ra nge H old Fun c ti o n is
activated.
mark means Auto-ran ging Function is
activated.
5. Wh en only the le ft mo st segment of the
Ba tt er y m ar k is flashing on th e LCD , it means battery voltage is low.
Rep lac e th e ba tte rie s wi th ne w on es.
Wh en no th in g i s displ ay ed on the LCD, the batteries are exhausted. Replace the batteries with new ones.
Loading...
+ 39 hidden pages