5. PREPARATION FOR MEASUREMENT ··············································· 12
5-1 Removing the Cover ······································································· 12
5-2 Battery Voltage Check ····································································· 12
5-3 Test Probe Connection ··································································· 12
6. OPERATION ·························································································· 13
6-1 Disconnection and Check of Power Source of
the Circuit under Test ····································································· 13
6-2 Insulation Resistance Measurement ··············································· 14
6-3 Continuity Measurement (Resistance Tests) ·································· 16
6-4 Continuous Measurement ······························································· 17
7. FUNCTIONS ·························································································· 18
7-1 TRAC-LOK Mode (Model 3007A only) ············································ 18
7-2 AUTO NULL ···················································································· 18
7-3 Back Light (Model 3007A only)························································ 18
7-4 Auto Power Off ················································································ 18
○ The instrument is designed to produce in accordance with the following
standards and supplied in the best conditions after successfully clearing
the inspection.
・IEC 61010-1 Overvoltage CAT III 300V Pollution Degree 2
・IEC 61010-2-31 (Safety requirements for hand-held probe assemblies)
This instruction manual describes warnings and safety rules which must be
observed by the user to ensure safety operation of the instrument and to
retain it in safe condition. Therefore, read through these instructions before
using the instrument.
WARNING
Read through and understand instructions contained in this manual
●
before using the instrument.
Save and keep the manual handy to enable quick reference whenever
●
necessary.
The instrument must only be used by a competent trained person and
●
operated in strict accordance with the instructions. KYORITSU will not
accept any liability for any damage or injury caused by misuse or noncompliance with the instructions or safety procedures.
It is essential to understand the safety rules contained in the manual.
●
They must be observed on using the instrument.
Be sure to observe the above rules strictly. Not following the
instructions may cause injury or instrument damage.
○ The symbol on the instrument means that the user must refer to the
relevant section of this manual for safe operation of the instrument. There
are three kinds of the symbol . Read the following instructions of each
symbol carefully.
― 1 ―
DANGER is reserved for conditions and actions that are likely to cause
serious or fatal injury.
WARNING is reserved for conditions and actions that can cause
serious or fatal injury.
CAUTION is reserved for conditions and actions that can cause minor
injury or instrument damage.
DANGER
Do not use this instrument on energized (LIVE) circuits.
●
Do not take measurement in the presence of flammable gasses.
●
Otherwise, the use of the instrument may cause sparkling, which leads
to an explosion.
Always keep your fingers behind the barrier on test probe during
●
measurement.
Never use the instrument if its surface or your hand is wet.
●
Never open the battery compartment cover while taking measurement.
●
WARNING
Do not attempt to take any measurements, if any abnormal conditions
●
occur, such as broken test probe and cracked enclosure of the
instrument.
Never change function switches with test probe connected to the
●
equipment under test.
Do not install substitute parts, perform any unauthorized modification, or
●
disassemble the instrument. Return the instrument to Kyoritsu or your
distributor for service and repair so as to ensure the safety features are
maintained.
Do not replace batteries when the surface of the instrument is wet.
●
Make sure to disconnect the test probe from the instrument before
●
opening the battery compartment cover for battery replacement.
― 2 ―
CAUTION
Always make sure to set the function switch or range selector switch to
●
the appropriate position before taking measurements.
Do not expose the instrument to the direct sun, dew fall or extreme
●
temperature and/or humidity.
When the instrument will not be in use for a long period of time, place it
●
in storage after removing batteries.
Use a damp cloth soaked in water or neutral detergent for cleaning the
●
instrument. Do not use abrasives or solvents.
― 3 ―
2. FEATURES
MODEL-3005A/3007A are microprocessor controlled insulation-continuity
testers.
● Designed to safety standards:
IEC 61010-1 Overvoltage CAT III 300V Pollution Degree 2
IEC 61010-2-31 (Safety requirements for hand-held probe assemblies)
● Display with back light function to facilitate work at night or dimly lit
locations (Model 3007A only)
● Bar graph to indicate measured results
● Strap belt to make both hands' operation easier
● Live circuit warning indication and buzzer
● Auto discharge function
When insulation resistance is measured, electric charges stored in
capacitive circuits are automatically discharged after testing. Discharge
can be checked with live circuit warning.
● Auto power off function
To prevent the instrument from being left powered on and conserve
battery power, the instrument automatically turns off approx. 10 minutes
after the last switch operation.
● LOK mode (Model 3007A only)
The test current is removed once a stable reading is reached to prevent
unnecessary battery consumption.
― 4 ―
3. SPECIFICATIONS
● Measuring Range and Accuracy (at 23±5°C, relative humidity 45 -75%)
○ Insulation Resistance Ranges:
Nominal Output Voltage250V500V1000V
Measuring
Ranges
Open-Circuit250V DC500V DC1000V DC
Voltage+20%,-0%+20%,-0%+20%,-0%
Nominal Current
Short-Circuit
Current
20MΩ
Accuracy
○ Continuity Ranges:
Ranges20Ω200Ω2000Ω
Measuring Ranges0 ~ 19.99Ω0 ~ 199.9Ω0 ~ 1999Ω
Open-Circuit
Voltage
Measuring Current
at 0.2Ω ~ 2Ω
Accuracy
○ AC Voltage Indication
200MΩ
0~1GΩ
2000
MΩ
1G~2GΩ
0 ~ 600V±5%rdg ±3dgt
0 ~19.99MΩ0 ~19.99MΩ0 ~19.99MΩ
0 ~ 199.9MΩ0 ~ 199.9MΩ0 ~ 199.9MΩ
0 ~ 1999MΩ0 ~ 1999MΩ0 ~ 1999MΩ
1mA DC min. 1mA DC min.1mA DC min.
at 0.25MΩat 0.5MΩat 1MΩ
1.5mA approx
±1.5%rdg ±5dgt
±10%rdg ±3dgt
7 ~ 12V approx
200mA min
±1.5%rdg ±5dgt
±3%rdg ±3dgt
±1.5%rdg ±3dgt
― 5 ―
● Typical Number of Measurements.
(central tendency for supply voltage up to 8V)
Insulation Resistance Ranges:
Approx. 1000 times min. at load 0.5MΩ
Continuity Ranges:
Approx. 700 times min. at load 1Ω
● Operating Errors (IEC 61557-2,-4)
FunctionsRange
20MΩ0.50 ~ 19.99MΩ
1000V200MΩ1.0 ~ 199.9MΩ
2000MΩ10 ~ 1000MΩ
20MΩ0.50 ~ 19.99MΩ
500V200MΩ1.0 ~ 199.9MΩ
2000MΩ10 ~ 100MΩ
20MΩ0.25 ~ 19.99MΩ±30%
250V200MΩ1.0 ~ 199.9MΩ
2000MΩ10 ~ 100MΩ
20Ω0.20 ~ 19.99Ω
Ω200Ω1.0 ~ 199.9Ω
2000Ω10 ~ 1999Ω
The influencing variations used for calculating the operating error are
denoted as follows;
Temperature:0℃ and 35℃
Supply voltage : 8V to 13.8V
● Applicable Standards
IEC 61010-1Over voltage CAT III 300V Pollution Degree 2
IEC 61010-2-31Safety requirements for hand-held probe
IEC 61557-1/2/4Measuring equipment for low voltage
IEC 61326-1EMC
Measuring range to keep Maximum percentage
operating erroroperating error
assemblies
distribution systems
― 6 ―
Loading...
+ 16 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.