KYORITSU 2009A User Manual

KEWSNAP
SERIES
KEWSNAP2009A
DIGITALCLAMPMETER
KYORITSUELECTRICALINSTRUMENTS WORKS,LTD.
INSTRUCTIONMANUAL
Contents
5−1 Checking Battery Voltage …………………………………… 12 5−2 Checking Switch Setting and Operation ……………………12
6.Measurement ………………………………………………………… 13
6−1 DC Current Measurement ……………………………………13 6−2 AC Current Measurement ……………………………………14 6−3 DC Voltage Measurement ……………………………………15 6−4 AC Voltage Measurement ……………………………………16 6−5 Resistance Measurement …………………………………… 17 6−6 Continuity Check ………………………………………………18 6−7 Frequency Measurement …………………………………… 18 6−8 Peak Measurement ………………………………………… 19 6−9 Average Measurement ……………………………………… 21
7.Other Functions ……………………………………………………… 22
7−1 Sleep Function ……………………………………………… 22 7−2 Data Hold Function ……………………………………………22 7−3 LoHz Function …………………………………………………23
7−4 OUTPUT Terminal …………………………………………… 23 8.Battery Replacement ………………………………………………… 26 9.Optional Accessories …………………………………………………27
―1―
1.SafetyWarnings
This instrument has been designed and tested according to IEC Publication 61010: Safety Requirements for Electronic Measuring Apparatus. This instruction manual contains warnings and safety rules which must be observed by the user to ensure safe operation of the instrument and retain it in safe condition. Therefore, read through these operating instructions before starting using the instrument.
WARNING
Read through and understand instructions contained in this manual before starting using the instrument.
Save and keep the manual handy to enable quick reference whenever necessary.
Be sure to use the instrument only in its intended applications and to follow measurement procedures described in the manual.
Be sure to understand and follow all safety instructions contained
in the manual. Failure to follow the above instructions may cause injury, instrument damage and/or damage to equipment under test.
The symbol indicated on the instrument means that the user
must refer to related parts in the manual for safe operation of the
instrument. Be sure to carefully read the instructions following
each symbol in this manual.
DANGER
is reserved for conditions and actions that are likely to
cause serious or fatal injury.
WARNING
is reserved for conditions and actions that can cause
serious or fatal injury.
CAUTION
is reserved for conditions and actions that can cause
minor injury or instrument damage.
―2―
Followingsymbols are used on the instrumentand in theinstruction manual.Attentionshouldbepaidtoeachsymboltoensureyoursafety.
Refertotheinstructionsinthemanual. Thissymbolis marked where the user must refer tothe instructionmanualso as not to cause personal injury or instrumentdamage.
Indicatesaninstrumentwithdoubleorreinforcedinsulation. Indicates that this instrument canclamp onbare conductors
whenmeasuring a voltage corresponding to theapplicable Measurementcategory,whichismarkednexttothissymbol.
IndicatesAC(AlternatingCurrent). IndicatesDC(DirectCurrent). IndicatesACandDC.
DANGER
Never make measurement on a circuit above 750VAC/1000VDC.
Do not attempt to make measurement in the presence of flammable gasses, fumes, vapor or dust. Otherwise, the use of the instrument may cause sparking, which can lead to an explosion.
Never attempt to use the instrument if its surface or your hand is wet.
Do not exceed the maximum allowable input of any measurement range.
Never open the battery compartment cover when making measurement.
Never try to make measurement if any abnormal conditions, such as broken Transformer jaws or case is noted.
The instrument is to be used only in its intended applications or conditions. Otherwise, safety functions equipped with the instrument doesn't work, and instrument damage or serious personal injury may be caused.
WARNING
N
ever attempt to make any measurement, if the instrument has any
structural abnormality such as cracked case and exposed metal part.
Do not turn the function selector switch with plugged in test leads connected to the circuit under test.
Do not install substitute parts or make any modification to the instrument. Return the instrument to Kyoritsu or your distributor for repair or re-calibration.
Do not try to replace the battery if the surface of the instrument is wet.
Always switch off the instrument before opening the battery compartment cover for battery replacement.
―3―
CAUTION
Make sure that the function selector switch is set to an appropriate position before making measurement.
Always make sure to insert each plug of the test leads fully into the appropriate terminal on the instrument.
Make sure to remove the test leads from the instrument before
making current measurement.
Do not expose the instrument to the direct sun, extreme temperatures or dew fall.
Be sure to set the function selector switch to the OFF" position after use. When the instrument will not be in use for a long period of time, place it in storage after removing the battery.
Use a damp cloth and detergent for cleaning the instrument. Do not use abrasives or solvents.
Measurement categories (Over-voltage categories)
To ensure safe operation of measuring instruments, IEC61010
establishes safety standards for various electrical environments,
categorized as CATto CAT, and called measurement
categories. Higher-numbered categories correspond to electrical
environments with greater momentary energy, so a measuring
instrument designed for CATenvironments can endure greater
momentary energy than one designed for CATⅡ.
CAT : Secondary electrical circuits connected to an AC electrical
outlet through a transformer or similar device.
CAT Ⅱ : Primary electrical circuits of equipment connected to an
AC electrical outlet by a power cord.
CAT Ⅲ : Primary electrical circuits of the equipment connected
directly to the distribution panel, and feeders from the distribution panel to outlets.
CAT :The circuit from the service drop to the service entrance,
and to the power meter and primary over-current protection device (distribution panel).
S
Incoming wire
CAT.IV
ocket
Interior wiring
CAT.III
Transformer
CAT.I
CAT.II
―4―
Tear-drop-shaped jaws for ease of use in crowded cable areas and
other tight places.
Accurate true-RMS reading of AC current or voltage with distorted
waveform.
LoHz mode automatically adjusts sample rate for easy reading of low
frequency AC current or voltage.
Average function for easy reading of input with large variation.
Auto-null function for easy zero adjustment.
Provides frequency reading in AC current or voltage measurement.
Auto-ranging feature on current, voltage and resistance ranges.
Wide measuring range from 0 up to 2000A.
Terminal cover to avoid the use of an incorrect terminal.
PEAK function for measuring a peak of input
Provides recorder output for long hour monitoring
Data Hold function for easy reading in dimly light or hard-to-read
locations
Sleep feature to extend battery life.
Permits easy continuity check with a beeper
Provides a dynamic range of 4,000 counts full scale
Wide frequency range from 30Hz to 1kHz
Uses shrouded transformer jaws to further improve safety
Designed to CAT.600VAC,DC / CAT.750VAC,1000VDC pollution
degree2 specified by the international safety standard:IEC61010.
2.Features
―5―
3.Specifications
Measuring Ranges and Accuracy (at 23±5℃, 45-75% relative humidity) DC Current A
Auto-ranging (Defaults to Lo): Lo:0-399.9A (Shifts to Hi at 400.0A) Hi:150-2000A
(Shifts to Lo at 149A. At 2020A or above,"OL" is shown.)
AC Current A (Crest factor(CF): 3.0 or less, peak current: 3000A or less)
Auto-ranging (Defaults to Lo) : Lo:0-399.9A (Shifts to Hi at 400.0A) Hi:150-2000A (Shifts to Lo at 149A. At 2020A or above, "OL" is shown.)
DC Voltage V (Input impedance: 2MΩ)
Auto-ranging (Defaults to Lo): Lo:0-39.99V (Shifts to Mid at 40.00 V) Mid:15.0-399.9V (Shifts to Lo at 14.9V and to Hi at 400.0V) Hi:150-1000A
(Shifts to Mid at 149V. At 1020V or above, OL" is shown.)
Measuring Range (Auto-ranging)
0〜±399.9A ±150〜±2000
Accuracy
±1.5rdg±2dgt
Measuring Range (Auto-ranging)
0399.9 1501700 17012000
Accuracy
±1.5rdg±3dgt(50/60Hz) ±3.0rdg±4dgt(30〜1kHz) ±3.5rdg±3dgt(50/60Hz)
Measuring Range (Auto-ranging)
0〜±39.99
±15.0〜±399.9V ±150〜±1000
Accuracy
±1.0rdg±2dgt
―6―
AC Voltage 〜V
(Crest factor(CF): 3.0 or less, peak voltage: 1200V or less)
Auto-ranging (Defaults to Lo): Lo:0-39.99V (Shifts to Mid at 40.00V) Mid:15.0-399.9V (Shifts to Lo at 14.9 and to Hi at 400.0V) Hi:150-750V
(Shifts to Mid at 149V. At 770V or above,OL" is shown.)
Resistance Ω
Auto-ranging (Defaults to Lo): Lo:0-399.9Ω (Shifts to Hi at 400.0Ω) Hi:150-3999Ω
(Shifts to Lo at 149Ω. At 4000Ω or above, OL" is shown.)
Continuity Check Ω (Range fixed)
Frequency
OUTPUT(Output impedance: about 10kΩ)
Measuring Range (Auto-ranging)
039.99
15.0399.9 150750
Accuracy
±1.5rdg±3dgt(50/60Hz)
±2.0rdg±4dgt(30〜1kHz)
Measuring Range (Auto-ranging)
0〜399.9Ω 1503999Ω
Accuracy
±1.5rdg±2dgt
Measuring Range
10399.9Ω
Accuracy
±1.5rdg±2dgt
Measuring Range
0〜±399.9A ±150〜±2000A 0〜399.9
1501700 17012000
Output Voltage(mVDC)
0〜±399.9mV ±15.0〜±200.0mV 0〜399.9mV
15.0170.0mV
170.1200.0mV
DC
AC
Accuracy
±1.5rdg±3mV
±1.5rdg±3mV(50/60Hz) ±3.0rdg±3mV(40〜1kHz) ±3.5rdg±3mV(50/60Hz)
Measuring Range
103999Hz
Accuracy
±1.5rdg±5dgt
―7―
Operating System Dual Integration
Display Liquid crystal display with a maximum
count of 4000 plus annunciators
Overrange Indication "OL" is shown on the display
Response Time Approx. 2 seconds.
Sample Rate About 3 times per second.
Location for use Indoor use, Altitude up to 2000m
Temperature and Humidity
23 +/-5, relative humidity up to 75%
for Guranteed Accuracy without condensation
Operating Temperature 0-40, relative humidity up to 85%
and Humidity without condensation
Storage Temperature -20-60, relative humidity up to 85%
and Humidity without condensation
Power Source 6F22(DC9V) or equivalent battery
Current Consumption Approx. 15mA max.
Sleep function Automatically powered down in about 30
minutes after the last switch operation (power consumption: about 200μA)
Overload Protection
DC/AC current ranges: 2400A AC for 10sec DC/AC voltage ranges: 1200V AC/DC for 10sec Resistance range: 600V AC/DC for 10sec
Withstand Voltage 5500V AC, 50/60Hz for 1 minute
between electrical circuit and housing case or metal part of the jaws
Insulation Resistance 10 MΩ or greater at 1000V between
electrical circuit and housing case or metal part of the jaws
Conductor Size Approx. 55diameter max.
Dimensions 250(L)x105(W)x49(D)mm
Weight About 530g
Accessories Test leads M-7017
6F22 battery Carrying case M-9094 Instruction manual Recorder output Plug M-8201
Optional Accessories Multi-Tran M-8008
Recorder M-5100A, etc. Output Lead M-7014.
―8―
*EffectiveValue(RMS)
Most alternating currents and voltages are expressed in effective values, which are also referred to as RMS (Root-Mean-Square) values. The effective value is the square root of the average of square of alternating current or voltage values.
Many clamp meters using a conventional rectifying circuit have "RMS" scales for AC measurement. The scales are, however, actually calibrated in terms of the effective value of a sine wave though the clamp meter is responding to the average value. The calibration is done with a conversion factor of 1.111 for sine wave, which is found by dividing the effective value by the average value. These instruments are therefore in error if the input voltage or current has some other shape than sine wave.
*
CF (Crest Factor) is found by dividing the peak value by the effective value.
Examples:
DC: CF =1 Sine wave: CF=1.414 Square wave with a 1: 10 duty ratio: CF=3
―9―
4.InstrumentLayout
Transformer Jaws
Pick up current flowing through the conductor.
Jaw Trigger
Operates the transformer jaws. Press to open the Transformer Jaws.
Function Selector Switch
Selects function. Also used to power the instrument on.
Data Hold Button
Freezes the display reading. "H" is shown on the display when Data Hold is enabled.
―10―
DataHold
Resistance
Voltage
Ampere
Lowfrequencyinput
DCAautozero
Peakmode
Lowbatterywarning
ContinuityCheck
Negativesign
AC
DC
Averagemode
When the plug is inserted into the output terminal, Data Hold Switch operates as range selection switch.See 7-4 output terminal
Mode Selector Button
Selects measuring mode. The instrument defaults to the normal mode (NOR). Then, press this switch to cycle through measuring modes. In any mode, pressing this switch for more than one second returns the instrument to the normal mode.
Zero Adjust/Reset Button
Used for zero adjustment on DCA and resistance ranges. Also used to reset the display reading in the PEAK mode. On DCA range, "AUTO" is shown on the display when auto-zeroing is completed. (Auto-zeroing is available on 400A range only.)
Digital Display
Field effect digital display with maximum reading of 3999. Function symbols and decimal point are controlled by the microprocessor based on the selected function and measuring mode.
―11―
Terminal Cover
Slides over V/Ω and COM Terminals to prevent access to them when OUTPUT terminal is in use.
OUTPUT Terminal(For AC or DC current range only)
Provides DC voltage output in proportion to the AC or DC current reading. The output is connected to a recording device such as a chart recorder for long hour monitoring. No output is available on voltage and resistance ranges.
COM Terminal
Accepts the black test lead for voltage or resistance measurement.
V/Ω Terminal
Accepts the red test lead for voltage or resistance measurement.
Safety Hand Strap
Prevents the instrument from slipping off the hand during use.
Test Leads (Model 7107)
Connected to COM and V/Ω terminals for voltage or resistance measurement.
Output Plug(Model 8201)
Plugs into the OUTPUT terminal for connection to a recording device. (See section 74, OUTPUT Terminal.)
Barrier
It is a part providing protection against electrical shock and ensuring the minimum required air and creepage distances.
―12―
5.PreparationforMeasurement
5−1 CheckingBatteryVoltage
Set the function selector switch to any position other than "OFF". When the display is clear without "BATT" showing, proceed to
measurement.
When the display blanks or "BATT" is indicated, replace the battery
according to section 8. Battery Replacement.
NOTE
The Sleep feature automatically turns the instrument off in about
30 minutes after the last switch or button operation. Therefore, the
display may be blank even with the function selector switch set to
a position other than "OFF". To operate the instrument in this
case, turn the switch back to the "OFF" position, then to any other
position, or press any button.
5−2 CheckingSwitchSettingandOperation
Make sure that the function selector switch is set to the correct
position, the instrument is set to the correct measuring mode and the
Data Hold function is disabled. Otherwise, desired measurement
cannot be made. (See section 6 for measurement instructions and
section 7 for notes on functions.)
―13―
6.Measurement
6−1 DCCurrentMeasurement
DANGER
Do not make measurement on a circuit above 1000VDC. This may
cause shock hazard or damage to the instrument or equipment under
test.
Do not make measurement with the battery compartment cover
removed from the instrument.
Do not make current measurement with the test leads connected to the
V/Ω and COM terminals.
Keep your fingers and hands behind the barrier during measurement.
Set the function selector switch to the " A" position. DC" should
be shown on the upper left corner of the display.
With the transformer jaws closed and without clamping them onto
the conductor, press the Zero Adjust/Reset button for about one second to zero adjust the display. (Zero adjust feature is for 400A range only.) When zero adjustment is completed, "AUTO" appears on the display.
Press the trigger to open the transformer jaws and clamp them onto
the conductor under test, then take the reading on the display. The most accurate reading will be obtained by keeping the conductor at the center of the transformer jaws.
Correct
Wrong
―14―
NOTE
During current measurement, keep the transformer jaws fully
closed. Otherwise, accurate measurement cannot be made. The maximum measurable conductor size is approx. 55mm in diameter.
When the current flows from the upside (the display side) to the
underside of the instrument, the reading is indicated positive.
The Zero Adjust/Reset button may not completely zero adjust the
output voltage from the OUTPUT terminal. In this case, make zero adjustment on the recording device.
Turing the function selector switch to a position other than DCA
cancels the zero adjustment .
6−2 ACCurrentMeasurement
DANGER
Never use the instrument on a circuit above 750VAC. This may
cause electrical shock hazard and damage to the instrument or the
circuit under test.
Do not make measurement with the test leads plugged into the
instrument.
Do not make measurement with the battery compartment cover
removed.
Keep your fingers and hands behind the barrier during measurement.
Correct
Wrong
―15―
Set the function selector switch to the "A" position. AC" should
be shown on the upper left corner of the display.
Press the trigger to open the transformer jaws and clamp them onto
a single conductor and take the reading on the display. The most accurate reading will be obtained by keeping the conductor at the center of the transformer jaws.
NOTE
During current measurement, keep the transformer jaws fully
closed. Otherwise, accurate measurements cannot be taken. Maximum conductor size is 55in diameter.
Zero adjustment is not necessary in AC current measurement.
When the current under test measures 3% of the full scale or less,
or the frequency of the current is low, "LoHz" is indicated on the display.
6−3 DCVoltageMeasurement
DANGER
Never use the instrument on a circuit above 1000VDC. This may
cause electrical shock hazard and damage to the instrument or the
circuit under test.
Do not make measurement with the battery compartment cover
removed.
Keep your fingers and hands behind the barrier during measurement.
―16―
Set the function selector switch to the " V" position. DC" should
be shown on the upper left corner of the display.
Slide the terminal cover to the left to disclose the V/Ω and COM
terminals. Plug the red test lead into the V/Ω terminal and the black test lead into COM terminal.
Connect the tip of the red and black test leads to the positive (+) and
negative (-) sides of the circuit under test respectively. Take the reading on the display. If the connection is reversed, the display indicates the "-" sign.
6−4 ACVoltageMeasurements
DANGER
Never use the instrument on a circuit above 750VAC. This may
cause electrical shock hazard and damage to the instrument or the
circuit under test.
Do not make measurement with the battery compartment cover
removed.
Keep your fingers and hands behind the barrier during measurement.
Set the function selector switch to the "V" position. "AC" should
be shown on the upper left corner of the display.
Slide the terminal cover to the left to disclose the V/Ω and COM
terminals. Plug the red test lead into V/Ω terminal and the black test lead into the COM terminal.
Black
testlead
Redtestlead
―17―
Connect the tip of the red and black test leads to the circuit under test
and take the reading on the display.
NOTE
When the voltage under test measures 3% of the full scale or less,
or the frequency of the voltage is low, "LoHz" is indicated on the display.
6−5 ResistanceMeasurement
DANGER
Never use the instrument on an energized circuit.
Do not make measurement with the battery compartment cover
removed.
Keep your fingers and hands behind the barrier during measurement.
Set the function selector switch to the " Ω" position.Slide the terminal cover to the left to disclose the V/Ω and COM
terminals. Plug the red test lead into the V/Ω terminal and the black test lead into the COM terminal.
With the tip of the test leads shorted together, press the Zero
Adjust/Reset button to offset the resistance of the test leads.
Connect the tip of the test leads to the circuit under test and take the
reading on the display.
―18―
6−6 ContinuityCheck(400Ω rangefixed)
The continuity check mode is enabled by pressing the mode selector switch on resistance range. " " is indicated on the display to show the instrument in the continuity check mode. The buzzer beeps, if the resistance under test is 20.0Ωor less.
DANGER
Never use the instrument on an energized circuit.
Do not make measurement with the battery compartment cover
removed.
Keep your fingers and hands behind the barrier during measurement.
Set the function selector switch to the " Ω" position.Slide the terminal cover to the left to disclose the V/Ω and COM
terminals. Plug the red test lead into the V/Ω terminal and the black test lead into the COM terminal.
With the tip of the test leads shorted together, press the Zero
Adjust/Reset button to offset the resistance of the test leads.
Press the mode selector button once to enter from the normal mode
to the continuity check mode. " " should be indicated on the display.
Connect the tip of the test leads to the circuit under test. If the
resistance is 20.0Ω or less, the buzzer beeps.
6−7 FrequencyMeasurement
On ACA or ACV range, the frequency of the current or voltage under test can be counted and shown on the display.
In the frequency measurement mode, "Hz" is indicated on the display.
Trigger threshold is approx. 10V for AC voltage and approx. 10A for AC current. At frequency measurement, in case of low input signal, it often happens that measurement cannot be made. Because range is fixed at 400V for AC voltage and at 400A for AC current.
―19―
DANGER
Never use the instrument on a high voltage circuit above 750VAC.
This may cause electrical shock hazard and damage to the
instrument or the circuit under test.
Do not make measurement with the battery compartment cover
removed.
Do not make current measurement with the test leads plugged into the
instrument.
Keep your fingers and hands behind the barrier during measurement.
Set the function selector switch to the "A" or "V" position.Press the mode selector button three times to enter from the normal
mode to the frequency measurement mode. "Hz" should be indicated on the display.
Follow instructions for ACA or ACV measurement and take the
frequency reading.
NOTE
When the voltage under test measures 3% of the full scale or less,
or the frequency of the current or voltage is 40Hz or less, "LoHz" is indicated on the display.
6−8 PeakMeasurement
In the PEAK mode, the display shows current or voltage's crest in effective value. (For example, when the current or voltage is sinusoidal, the reading equals the crest value divided by the square root of two.) The display reading is constantly updated with a maximum crest.
In this mode, "PEAK" is indicated on the display.
Response time is 300ms in DC measurement and 10ms in AC measurement.
―20―
DANGER
Never use the instrument on a circuit above 750VAC/1000VDC.
This may cause electrical shock hazard and damage to the
instrument or the circuit under test.
Do not make measurement with the battery compartment cover
removed.
Do not make measurement with the test leads plugged into the
instrument.
Keep your fingers and hands behind the barrier during measurement.
The PEAK mode is available on DCA, ACA, DCV and DCA ranges.
Set the function selector switch to the desired position. NoteOnly on DCA range, press the Zero Adjust/Reset button for
about one second to zero adjust the reading with the transformer jaws closed.
Press the mode selector button twice to enter from the normal mode
to the PEAK mode. "PEAK" should be shown on the display.
Follow instructions for DCA, ACA, DCV or ACV measurement.
NoteFor accurate reading, press the Zero Adjust/Reset button to reset
the reading after clamping onto the conductor or making test lead connections to the circuit under test. Then, proceed to measurement.
Vp
INPUTCurrent
Vp/
󰵫2
PeakHold
―21―
NOTE
In the PEAK mode, the auto-ranging feature is disabled and
measuring ranges are fixed as follows. DC/ACA: 0-400.0A DC/ACV: 0-400.0A
The Sleep function is disabled in the PEAK mode as well.
6−9 AverageMeasurement
In the Average mode, "AVG" is indicated on the display.
The display reads a running average of six readings over an interval of about 2 seconds.
This mode is available on ACV, DCV, ACA and DCA ranges.
Set the function selector switch to the desired position.Press the mode selector button once to enter from the normal mode
to the Average mode. "AVG" should be indicated on the display.
Follow instructions for ACV, DCV, ACA or DCA measurement.The display shows a running average of six readings over an interval
of about 2 seconds.
―22―
7−1 SleepFunction
CAUTION
The instrument consumes small amount of battery power in the
sleep mode. Make sure to set the function selector switch to the
OFF position after use.
This is a function to prevent the instrument from being left powered on in order to conserve battery life. This function causes the instrument to enter the Sleep (powered-down) mode about 30 minutes after the last switch or button operation.
To exit the Sleep mode, turn the function selector switch back to "OFF", then to any other position, or press any button.
NOTE
Connecting the plug to the OUTPUT terminal disables the Sleep
function. The function is enabled on removing the plug from the terminal.
The Sleep function is disabled in the PEAK measurement mode.
7−2 DataHoldFunction
This is a function used to freeze the measured value on the display. Press the Data Hold button to freeze the reading. The reading will be held regardless of subsequent variation in input. "H" is shown on the upper right corner of the display while the instrument is in the Data Hold mode.
To exit the Data Hold mode, press the Data Hold button again.
7.OtherFunctions
―23―
NOTE
If the instrument in the Data Hold mode goes into "sleep," it will
return to the normal mode.
7−3 LoHzFunction
In ACV or ACA range, if frequency of the voltage or current under test is 40Hz or lower, the display indicates "LoHz" and sample rate is automatically switched from the normal 3 times/sec to 2 times/sec to reduce fluctuation of the reading. "LoHz" is also indicated where input is 3% of full scale or less.
7−4 OUTPUTTerminal(Forcurrentrangesonly)
DANGER
Never use the instrument on a circuit above 750VAC/1000VDC.
This may cause electrical shock hazard and damage to the
instrument or the circuit under test.
Do not make measurement with the battery compartment cover
removed.
Never apply voltage to the OUTPUT terminal.
Attach the output plug to a connection lead so that the output
voltage can be connected to a recording device such as a chart recorder.
―24―
Slide the terminal cover to the right to disclose the OUTPUT terminal
and insert the output plug into the terminal. Make connection to the recording device.
When the plug is inserted into the output terminal, auto-range function
is cleared. Set the range depending on the state of Data Hold Switch.
Data Hold Switch OFF 400A range Data Hold Switch ON 2000A range
NoteAfter measurement, be sure to return Data Hold Switch to OFF
position.
Set the function selector switch to the desired position (ACA or DCA)
and follow appropriate measurement instructions.
NOTE
During current measurement, keep the transformer jaws fully
closed. Otherwise, accurate measurement cannot be made. The maximum measurable conductor size is approx. 55mm in diameter.
Zero adjustment is not necessary on AC current range.
On DC current range, the Zero Adjust/Reset button may not
completely zero adjust the output voltage from the OUTPUT terminal. In this case, make zero adjustment on the recording device.
Connecting the plug to the OUTPUT terminal disables the Sleep
function. The function is enabled on removing the plug from the terminal.
―25―
Consult the output voltage specifications shown in section 3 and
adjust the sensitivity of the recording device.
For long hours of use of the OUTPUT terminal, use an Alkaline
battery, which will extend continuous recording time up to about 24 hours.
―26―
WARNING
To avoid electric shock hazard, make sure to set the function selector
switch to "OFF" and remove the test leads from the instrument before
trying to replace battery.
CAUTION
Do not mix new and old battery.
Make sure to install battery in correct polarity as indicated in the
battery compartment.
If the battery voltage becomes too low for the instrument to operate normally, "BATT" is shown on the display. Then, replace the battery. Note that when the battery is completely exhausted, the display blanks without "BATT" shown.
Set the function selector switch to the "OFF" position.
Unscrew and remove the battery compartment on the bottom of the instrument.
Replace the battery observing correct polarity. Use a new 6F22 or
equivalent battery.
Re-place and screw the battery compartment cover.
8.BatteryReplacement
Screw
Battery
Batterycompartmentcover
―27―
9.OptionalAccessories
MODEL 8008 (For AC current measurement only)
Multi-Tran Model 8008 is designed to increase the measuring capability of a clamp meter. With the use of the Multi-tran, you can not only extend current range over 3000A, but also clamp on a large bus-bar or conductor.
Set the function selector switch to the "A" position .As shown in the figure below, clamp KEW SNAP 2009A onto the
pickup coil of MODEL 8008.
Clamp MODEL 8008 onto the bus-bar or conductor under test.Take the reading on KEW SNAP 2009A and multiply it by 10.
―28―
MEMO
―29―
MEMO
DISTRIBUTOR
92―1394C
06-05
Kyoritsu reserves the rights to change specifications or designs described in this manual without notice and without obligations.
Loading...