KYORITSU 2002R User Manual

取扱説明書 INSTRUCTIONMANUAL
KEWSNAP
SERIES
MODEL 2002PA 2002R
デジタルクランプメータ DIGITALCLAMPMETER
―25―
Contents
ENGLISH
1.Safety Warnings ……………………………………………………… 26 2.Features ……………………………………………………………… 28 3.Specifications
3−1 MODEL 2002PA…………………………………………………29
3−2 MODEL 2002R ………………………………………………… 31 4.Instrument Layout …………………………………………………… 34 5.Preparation for Measurement
5−1 Battery Voltage Check ……………………………………… 35
5−2 Checking Switch Setting and Operation ……………………35 6.Measurement
6−1 Current Measurement…………………………………………36
6−2 Voltage Measurement ……………………………………… 38
6−3 Resistance Measurement …………………………………… 39 7.Notes on Functions
7−1 Data Hold ………………………………………………………41
7−2 Sleep Function…………………………………………………41
7−3 Output for Recorder …………………………………………42
7−4 Mode Switching Function ……………………………………43 8.Battery Replacement ………………………………………………… 44 9.Optional Accessories …………………………………………………45
―26―
1.SafetyWarnings
This instrument has been designed and tested according to IEC
Publication 61010Safety Requirements for Electronic Measuring
Apparatus. This instruction manual contains warnings and safety
rules which must be observed by the user to ensure safe operation of
the instrument and to retain it in safe condition. Therefore, read
through these operating instructions before starting using the
instrument.
WARNING
Read through and understand instructions contained in this manual before starting to use the instrument.
Save and keep the manual handy to enable quick reference whenever necessary.
Be sure to use the instrument only in its intended applications and to follow measurement procedures described in the manual.
Be sure to understand and follow all safety instructions contained in the manual.
Be sure to observe the above instructions. Failure to follow the above instructions may cause injury, instrument damage and/or damage to equipment under test.
The symbol indicated on the instrument means that the user
must refer to related parts in the manual for safe operation of the instrument. Be sure to carefully read the instructions following each symbol in this manual.
DANGER is reserved for conditions and actions that are likely to
cause serious or fatal injury.
WARNING is reserved for conditions and actions that can cause
serious or fatal injury.
CAUTION is reserved for conditions and actions that can cause
minor injury or instrument damage.
―27―
DANGER
Never make measurement on the circuit above 750VAC or 1000VDC.
Do not attempt to make measurement in the presence of flammable
gasses, fumes, vapor or dust. Otherwise, the use of the instrument may cause sparking, which can lead to an explosion.
Transformer jaw tips are designed not to short the circuit under test. If equipment under test has exposed conductive parts, however, extra precaution should be taken to minimize the possibility of shorting.
Never attempt to use the instrument if its surface or your hand is wet.
Do not exceed the maximum allowable input of any measurement
range.
Never open the battery compartment cover and the instrument case when making measurement.
Verify proper operation on a known source before use or taking action as a result of the indication of the instrument.
WARNING
Never attempt to make any measurement if any abnormal conditions are noted, such as broken case, cracked test leads and exposed metal parts.
Do not turn the function selector switch with test leads connected to the instrument.
Do not install substitute parts or make any modification to the instrument. Return the instrument to your distributor for repair or re­calibration.
Do not try to replace the batteries if the surface of the instrument is wet.
Always switch off the instrument and make sure to disconnect test leads
before opening the battery compartment cover for battery replacement.
CAUTION
Always make sure to check the function selector switch is set to an appropriate position before starting measurement.
Always make sure to insert the plug of each lead fully into the appropriate terminal on the instrument.
Be sure to set the function selector switch to the "OFF" position after use. When the instrument will not be in use for a long period, place it in storage after removing the batteries.
Do not expose the instrument to the direct sun, high temperature or dew fall.
Use a cloth dipped in water or neutral detergent for cleaning the instrument. Do not use abrasives or solvents.
―28―
Tear-drop-shaped jaws for ease of use in crowded cable areas and other tight places.
Accurate true-RMS reading of AC current or voltage with distorted waveform (MODEL2002R)
Provides a wide measuring range from 0 up to 2000A.
Terminal cover to avoid the use of an incorrect terminal.
Measures current variation as short as 10 msec with peak-hold
feature.
Provides output to a chart recorder for current variation recording.
Designed to international safety standards.
IEC61010-1 (CAT.600V /CAT.1000V Pollution degree 2) IEC61010-031, IEC61010-2-032
Data hold function to allow easy readings in dimly lit or hard-to-read locations.
Sleep feature to conserve battery power.
Permits easy continuity check with a beeper.
Provides a dynamic range of 4,000 counts full scale.
Provides wide measuring range of voltage and resistance in auto-
ranging.
Wide frequency range from 40Hz to 1kHz (Current measuring range: 0-1500A).
Transformer jows fitted with guard to further improve safety.
Protected throughout by double or reinforced insulationF.
2.Features
―29―
3−1 MODEL2002PA
Measuring Ranges and Accuracy(at 23±5, relative humidity 45-75%)
AC Current 〜400A,〜2000
3.Specifications
Range
400
2000
Measuring Range
0〜400.0A 0〜1000
10001500 15002000
Measuring Range 0〜40.00
15.0400.0 150750
Measuring Range
0〜±40.00V ±15.0〜±400.0V ±150〜±1000
Measuring Range 0〜400.0Ω
0.1504.000kΩ
1.5040.00kΩ
15.0400.0kΩ
Output Voltage/Measuring Range
0〜400.0mV /0〜400A 0〜150.0mV /0〜1500
150.0200.0mV/15002000
Resolution
0.1
1A
Resolution
0.01
0.1
1V
Resolution
0.01
0.1
1V
Resolution
0.1Ω
1Ω
10Ω 100Ω
AccuracyFrequency Range
±1.0rdg±3dgt(50/60Hz) ±2.0rdg±3dgt(40〜1kHz) ±1.0rdg±3dgt(50/60Hz) ±3.0rdg±3dgt(40〜1kHz) ±3.0rdg (50/60Hz
AccuracyFrequency Range) ±1.0rdg±2dgt(50/60Hz) ±1.5rdg±3dgt(40〜1kHz)
Accuracy
±1.0rdg±2dgt
Accuracy
±1.5rdg±2dgt
AccuracyFrequency Range
±1.5rdg±0.5mV(50/60Hz) ±2.5rdg±0.5mV(40〜1kHz) ±1.5rdg±0.5mV(50/60Hz) ±3.5rdg±0.5mV(40〜1kHz) ±3.5rdg (50/60Hz
Maximum Measurement Time
Continuous
15min min
Range 40 400 750
Range 40 400 1000
Range 400Ω 4kΩ 40kΩ 400kΩ
Range 400
2000
AC VoltageV)Auto-ranging
Initially set to the 40V range. Input impedance is about 1MΩ.
DC Voltage(VAuto-ranging
Initially set to the 40V range. Input impedance is about 1MΩ.
Resistance(Auto-ranging)
Initially set to the 400Ω range. In the continuity check mode, fixed to the 400Ω range and when the reading is not more than 50±35Ω, the buzzer beeps.
OUTPUT(AC Current Ranges
DC Output: 100.0mV per 1000 counts (Output impedance: about 10kΩ)
Electromagnetic compatibility (EMC) EN61000-4-3 Radiated RF electromagnetic field immunity RF field strength
=1V/m
, total accuracy:specified accuracy
RF field strength
=3V/m
, total accuracy:specified accuracy
+2%
of range
―30―
Operating System Dual Integration
Display Liquid crystal display with a maximum
count of 4000
Low Battery Warning "BATT" symbol is displayed on the digital display.
Overrange Indication "OL" is displayed where input exceeds the upper limit of a range
Response Time Approx. 2 seconds
Sleep function Automatically powered down in about 10
minutes after the last switch operation
Data Hold Available in all ranges provided the peak measurement mode is deactivated.
Operating Environmental indoor use
conditions altitude up to 2000m
Storage Temperature 2060, relative humidity up to 85%
and Humidity without condensation
Operating Temperature 040, relative humidity up to 85%
and Humidity without condensation
Conductor Size Approx. 54.5diameter max.
Overload Protection 2400A AC for 10sec
1200V AC/DC for 10sec 600V AC for 10sec
Withstand Voltage 5320V AC for 5 seconds between electrical circuit and housing cases or metal parts of jaws
Insulation Resistance 10MΩ or greater at 1000V between electrical circuit and housing cases or metal parts of jaws
Safety Standard IEC 61010-1: Measurement CAT. 600V/CAT.1000V, pollution degree 2.
Dimensions 247(L)×105(W)×49(Dmm
Weight Approx. 470g(battery included)
Power Source Two R6P(DC1.5V) batteries or equivalent
Current Consumption Approx. 5mA max. (Approx. 20μA in the
sleep mode)
Accessories Test leads M-7107 Two R6P batteries Instruction manual Recorder Output Plug M-8201 Carrying case M-9094
Optional Accessories Multi-Tran M-8008 Output Probe M-7014, etc.
.
―31―
3−2 MODEL2002R
Measuring Ranges and Accuracy(at 23±5, relative humidity 45-75%)
AC Current 〜400A,〜2000 (9 counts or less is corrected to 0)
Range
400
2000
Measuring Range
0〜400.0A 0〜1000
10001500 15002000
Measuring Range 0〜40.00
15.0400.0 150750
Measuring Range
0〜±40.00V ±15.0〜±400.0V ±150〜±1000
Measuring Range 0〜400.0Ω
0.1504.000kΩ
1.5040.00kΩ
15.0400.0kΩ
Output Voltage/Measuring Range
0〜400.0mV /0〜400A 0〜150.0mV /0〜1500
150.0200.0mV/15002000
Resolution
0.1
1A
Resolution
0.01
0.1
1V
Resolution
0.01
0.1
1V
Resolution
0.1Ω
1Ω
10Ω 100Ω
AccuracyFrequency Range
±1.5%rdg±3dgt4565Hz) ±2.5%rdg±3dgt401kHz) ±2.0%rdg±5dgt4565Hz) ±3.0%rdg±5dgt401kHz) ±4.0%rdg 50/60Hz
AccuracyFrequency Range) ±1.0rdg±2dgt(4565Hz) ±1.5rdg±3dgt(40〜1kHz)
Accuracy
±1.0rdg±2dgt
Accuracy
±1.5rdg±2dgt
AccuracyFrequency Range
±2.0%rdg±0.5mV4565Hz) ±3.0%rdg±0.5mV401kHz) ±2.5%rdg±0.5mV4565Hz) ±3.5%rdg±0.5mV401kHz) ±4.5%rdg 50/60Hz
Maximum Measurement Time
Continuous
15min min
Range 40 400 750
Range 40 400 1000
Range 400Ω 4kΩ 40kΩ 400kΩ
Range 400
2000
AC Voltage〜V)Auto-ranging (9 counts or less is corrected to 0)
Initially set to the 40V range. Input impedance is about 1MΩ. When approx.300V or more is applied to the instrument instantaneously, the measured value is indicated on 750V range.
DC Voltage(VAuto-ranging
Initially set to the 40V range. Input impedance is about 1MΩ.
Resistance(Auto-ranging)
Initially set to the 400Ω range. In the continuity check mode, fixed to the 400Ω range and when the reading is not more than 50±35Ω, the buzzer beeps.
OUTPUT(AC Current Ranges
DC Output: 100.0mV per 1000 counts (Output impedance: about 10kΩ)
CF (Crest Factor) CF=3 or less accuracy+1% (45〜65Hz)、less than AC3000A/AC1200V Peak
Electromagnetic compatibility (EMC) EN61000-4-2 Electrostatic discharge immunity(ESD) Performance criteria B
―32―
Operating System Dual Integration
Display Liquid crystal display with a maximum
count of 4000
Low Battery Warning "BATT" symbol is displayed on the digital display.
Overrange Indication "OL" is displayed where input exceeds the upper limit of a range
Response Time Approx. 2 seconds (at full scale)
Sleep function Automatically powered down in about 10
minutes after the last switch operation
Data Hold Available in all ranges provided the peak measurement mode is deactivated.
Operating Environmental indoor use
conditions altitude up to 2000m
Storage Temperature 2060, relative humidity up to 85%
and Humidity without condensation
Operating Temperature 040, relative humidity up to 85%
and Humidity without condensation
Conductor Size Approx. 54.5diameter max.
Overload Protection 2400A AC for 10sec
1200V AC/DC for 10sec 600V AC for 10sec
Withstand Voltage 5320V AC for 5 seconds between electrical circuit and housing cases or metal parts of jaws
Insulation Resistance 50MΩ or greater at 1000V between electrical circuit and housing cases or metal parts of jaws
Safety Standard IEC 61010-1: Measurement CAT. 600V/CAT.1000V, pollution degree 2.
Dimensions 247(L)×105(W)×49(Dmm
Weight Approx. 470g(battery included)
Power Source Two R6P(DC1.5V) batteries or equivalent
Current Consumption Approx. 10mA max. (Approx. 20μA in the
sleep mode)
Accessories Test leads M-7107 Two R6P batteries Instruction manual Recorder Output Plug M-8201 Carrying case M-9094
Optional Accessories Multi-Tran M-8008 Output Probe M-7014, etc.
―33―
―34―
4.InstrumentLayout
TransformerJaw
JawTrigger
Display
ModeSwitch
ResetSwitch
DataHoldSwitch
TerminalCover
DropPreventionStrap
V/Ω Terminal
COMTerminal
OUTPUTTerminal
TestLeads
Mー7107
RecorderOutputPlug
Mー8102
DataHoldMode
Resistance
Voltage
Current
PeakModeMeasurementBatteryVoltageAlertContinuityCheckMode
NegativePotential
ACCurrent
DCCurrent
● LCD INDICATOR
FunctionSwitch
―35―
5.PreparationforMeasurement
5−1 CheckingBatteryVoltage
Set the function selector switch to any position other than "OFF". When the display is clear without "BATT" showing, proceed to
measurement.
When the display blanks or "BATT" is indicated, replace the
batteries according to section 8: battery replacement.
NOTE
It is possible that display is kept in blank while the function selector switch is set to a position other than "OFF". This is due to sleep function which automatically powers the instrument down in a certain period of time after the last switch operation. To operate the instrument in this case, set the switch back to the "OFF" position, then to the desired position, or press any button.
5−2 CheckingSwitchSettingandOperation
Make sure that the function selector switch is set to the correct position, the instrument is set to the correct mode and the data hold function is deactivated. Otherwise, desired measurement cannot be made. (See section 6 for measurement instructions and section 7 for notes on functions.)
―36―
6−1 CurrentMeasurement
WARNING
Do not make measurement on a circuit above 750V AC. This may cause shock hazard or damage to the instrument or equipment under test.
Transformer jaw tips are designed not to short the circuit under test. If equipment under test has exposed conductive parts, however, extra precaution should be taken to minimize the possibility of shorting.
Do not make measurement with the battery compartment cover removed from the instrument.
Do not make current measurement with the test leads connected to the V/Ω and COM terminals.
When measuring current is not less than 1000A, make sure to stop measurement within the maximum measuring time shown below. Otherwise, transformer jaws may heat to cause a fire or deformation of molded parts, which will degrade insulation.
10001500A:15min. 15002000A:5min.
6−1−1 ACCurrentMeasurement(NormalMode)
Set the function selector switch to the "400A" or "2000A"
position and make sure that the current under test does not exceed the upper limit of the measuring range you are selecting.
Press the trigger to open the transformer jaws and clamp them onto
the conductor under test.
6.Measurement
Correct
Incorrect
―37―
Take the reading on the display.
NOTE
During current measurement, keep the transformer jaws fully
closed. Otherwise, accurate measurement cannot be made. The maximum measurable conductor size is 54.5mm in diameter.
When measuring a larger current, the transformer jaws may buzz. This is not a fault and does not affect the accuracy at all.
6−1−2 PeakCurrentMeasurement
Set the function selector switch to the "400A" or "2000A" position.Press the mode switch to select the peak mode. "PEAK" will be
shown on the display.
Press the trigger to open the transformer jaws and clamp them onto
the conductor under test. Then, press the reset switch.
The display shows the current's crest value divided by the square
root of two. Therefore, when the current is sinusoidal, the reading equals RMS value.
To reset the display, press the reset switch. Note: When this is done, the reading goes off for about one second.
After the measurement is over, press the mode switch to return to
the normal mode.
NOTE
In the peak measurement mode, the data hold feature is disabled.
When a measured value is 9 counts or less, it is corrected to 0
(MODEL2002R).
Vp
INPUTCurrent
Vp/
󰵫2
PeakHold
―38―
6−2 VoltageMeasurement
WARNING
Do not make measurement on a circuit above 750V AC or 1000V DC. This may cause shock hazard or damage to the instrument or equipment under test.
Do not make measurement with the battery compartment cover removed.
6−2−1 DCVoltageMeasurement
Set the function selector switch to the " V" position.Slide the terminal cover to the left. Plug the red test lead into the V/Ω
terminal and the black test lead into the COM terminal.
Connect the other end of the test leads to the circuit under test.
Take the reading on the display. When the red lead is the negative potential, the "" sign is shown on the display.
6−2−2 ACVoltageMeasurement
Set the function selector switch to the "V" position.Slide the terminal cover to the left. Plug the red test lead into the V/Ω
terminal and the black test lead into the COM terminal.
Connect the other end of the test leads to the circuit under test. Take
the reading on the display.
NOTE
For high sensitivity, there are parts which do not indicate "0".
Blacktestlead
Redtestlead
―39―
6−3 ResistanceMeasurement
WARNING
Before attempting to make measurement, make sure that the circuit under test is not live. The instrument is protected against a voltage up to 600V.
Do not make measurement with the battery compartment cover removed.
6−3−1 ResistanceMeasurement(NormalMode)
Set the function selector switch to the "Ω/" position.Slide the terminal cover to the left. Plug the red test lead into the V/Ω
terminal and the black test lead into the COM terminal.
Short the tip of the test leads and check whether the display reads "0".Connect the tip of the test leads to the circuit under test. Take the
reading on the display.
NOTE
When shorting the tip of the test leads, the display may read a very
small resistance instead of "0." This is the resistance of the test leads, not a fault.
If one of the test leads is open, the display reads "OL".
Black
testlead
Redtestlead
―40―
6−3−2 ContinuityCheck
Set the function selector switch to the "Ω/ " position.Slide the terminal cover to the left. Plug the red test lead into the V/Ω
terminal and the black test lead into the COM terminal.
Press the mode switch to set the instrument to the continuity check
mode. The measuring range is fixed to 400Ω and the " " symbol is shown on the display.
Short the tip of the test leads and make sure that the display reads
"0" and the buzzer beeps.
Connect the tip of the test leads to the circuit under test. The
display reads the resistance and the buzzer beeps when the reading is not more than about 50Ω.
NOTE
When shorting the tip of the test leads, the display may read a very
small resistance instead of "0". This is the resistance of the test leads, not a fault.
If one of the test leads is open, the display reads "OL".
―41―
7−1 DataHold
This is a function used to freeze the measured value on the display. Press the data hold switch. The reading becomes frozen and the
"H" symbol is shown on the display, indicating the instrument in the data hold mode.
To exit the data hold mode, press the data hold switch again to
release it.
NOTE
When the function selector switch is turned while the instrument is
in the data hold mode, the data hold function remains activated. To make measurement in this case, release the data hold switch by pressing it and exit the data hold mode.
The data hold function is disabled in the peak measurement mode
on the AC current range.
When the sleep function is activated, the data hold mode turns to
the normal mode.
7−2 SleepFunction
This is a function to prevent the instrument from being left powered on in order to conserve battery life. The instrument automatically enters the sleep (powered-down)
mode about 10 minutes after the last switch operation. To exit the sleep mode, press the data hold, reset or mode switch
or turn the function selector switch back to "OFF", then to any other
position.
How to Exit the Sleep Mode
Turn the function selector switch from "OFF" to another position with
the data hold switch pressed. Then, "P.OFF" is shown on the display. This disables the sleep function and enables continuous use of the instrument.
7.NotesonFunctions
―42―
To enable the sleep function, turn the function selector switch back
to "OFF", then to any other position.
NOTE
The instrument consumes small amount of current in the sleep
mode. When the instrument is not in use, make sure to set the function selector switch to "OFF".
7−3 RecorderOutput
Only on the "400A" or "2000A" range, DC voltage proportional to the input current is output from the OUTPUT terminal. Set the function selector switch to the "400A" or "2000A"
position.
Slide the terminal cover to the right and insert the recorder output
plug into the OUTPUT terminal for connection with a recorder or other recording device.
NOTE
Output voltage is 1mV/A on the "400A" range and 0.1mV/A on
the "2000A" range. Set an appropriate input sensitivity on the recorder.
The peak hold function does not apply to the recorder output even
if the instrument is in the peak hold mode.
For long term measurement, disable the sleep function. (See
section 7-2 for sleep function.)
―43―
[Howtoattachapairofleadstotherecorderoutputplug]
Attach one end of the leads to the recorder output plug observing
correct polarity, then connect the other end to the recorder.
DANGER
Never apply voltage to the output terminal.
7−4 ModeSwitchingFunction
On a AC current ("400A" or "2000A") or the resistance(Ω/) range, press the mode switch to cycle through the measurement modes. The instrument is initially set to the normal mode and can be switched to the peak or continuity check mode by means of the mode switch. (See section 6-1-2 for peak current measurement and section 6-3-2 for continuity check.)
Normal
Peak
Peakcurrentvalueisdisplayed.
(Responsetime:10ms)
"PEAK"isshownonthedisplay.
Pressthemodeswitch
Normal
ContinuityCheck
Thebuzzerbeepsiftheresistance isnotmorethanabout50Ω. Thebuzzersymbol isshown on thedisplay.
≪ACCurrentRange(400Aor2000A)≫
≪ResistanceRange≫
Pressthemodeswitch
Pressthemodeswitch
Pressthemodeswitch
―44―
WARNING
To avoid electric shock hazard, make sure to set the function selector switch to "OFF" and remove the test leads from the instrument before trying to replace batteries.
CAUTION
Do not mix new and old batteries.
Make sure to install batteries in correct polarity as indicated in the
battery compartment.
NOTE
If the instrument is powered on, but the display blanks or "BATT" is
shown on the display, replace the batteries.
Set the function selector switch to the "OFF" position.Unscrew and remove the battery compartment cover on the bottom
of the instrument.
Replace the batteries observing correct polarity. Use two new R6P
batteries.
Replace and screw the battery compartment cover.
8.BatteryReplacement
Screw
Battery
Battery compartment cover
―45―
9.OptionalAccessories
Multi-Tran MODEL 8008 extends the capability of MODEL 2002PA or MODEL 2002R, allowing measurement up to 3000A or on a large bus-bar or conductor.
Set the function selector switch to "400A".As shown in the figure below, clamp MODEL 2002PA or MODEL
2002R onto the pickup coil of MODEL 8008.
Clamp MODEL 8008 onto the bus-bar or conductor under test.Take the reading on MODEL 2002PA or MODEL 2002R and
multiply it by 10.
MAX 150㎜
MAX 100㎜
―46―
MEMO
―47―
MEMO
―48―
MEMO
―49―
MEMO
DISTRIBUTOR
92―1579A
03―08
Kyoritsu reserves the rights to change specifications or designs described in this manual without notice and without obligations.
Loading...