KYORITSU 1110 User Manual

INSTRUCTIONMANUAL
MODEL1110
KYORITSUMULTIMETER
KYORITSUELECTRICALINSTRUMENTS WORKS,LTD.
1. Safety Warnings
Safety Requirements for Electronic Measuring Apparatus. This instruction manual
contains warnings and safety rules which must be observed by the user to ensure
safe operation of the instrument and retain it in safe condition. Therefore, read
through these operating instructions before using the instrument.
WARNING
Read through and understand instructions contained in this manual before
starting using the instrument.
Save and keep the manual handy to enable quick reference whenever
necessary.
Be sure to use the instrument only in its intended applications and to follow
measurement procedures described in the manual.
Be sure to understand and follow all safety instructions contained in the
manual.
Failure to follow the above instructions may cause injury, instrument damage
and/or damage to equipment under test.
The symbol indicated on the instrument means that the user must refer to related
parts in the manual for safe operation of the instrument. Be sure to carefully read
the instructions following each symbol in this manual.
DANGER is reserved for conditions and actions that are likely to cause
serious or fatal injury. WARNING is reserved for conditions and actions that can cause serious or
fatal injury. CAUTION is reserved for conditions and actions that can cause injury or
instrument damage.
DANGER
Make sure to set the function selector switch to the appropriate position
before making measurement. Use extreme caution not to avoid applying
voltage to the instrument with the range selector switch set to a current or
resistance range.
Do not attempt to make measurement in the presence of flammable gasses,
fumes, vapor or dust. Otherwise, the use of the instrument may cause
sparking, which can lead to an explosion.
Never attempt to use the instrument if its surface or your hand is wet.
Do not exceed the maximum allowable input of the selected measuring
range.
Never open the instrument when making measurement.
WARNING
Never attempt to make any measurement if any abnormal conditions are
noted, such as broken case, cracked test leads and exposed metal parts.
Do not turn the function selector switch with test leads connected to the
instrument.
Do not install substitute parts or make any modification to the instrument.
Return the instrument to Kyoritsu or your distributor for repair or re-
calibration.
Do not try to replace the batteries or fuse if the surface of the instrument is
wet.
Always set the function selector switch to the OFF position before opening
the instrument for battery replacement.
CAUTION
Always make sure to insert each plug of the test leads fully into the
appropriate terminal on the instrument.
Be sure to set the function selector switch to the OFF position after use.
When the instrument will not be in use for a long period of time, place it in
storage after removing the batteries.
Do not expose the instrument to the direct sun, extreme temperatures or
dewfall.
Use a damp cloth and detergent for cleaning the instrument. Do not use
abrasives or solvents.
Ranges
DC.V
AC.V
DC.A
LED
10mA approx. at 0Ω(at 3V of battery voltage) Buzzer beeps below about 100
TEMP.※2
BATT.TEST 1.5V
0.7 2.0V(Load resistance:10
approx.)
0 100
:
Ranges other than the above(-20 150:
0.3V 3V 12V 30V 120V 300V 600V 12V 30V 120V 300V 600V 60μA 30mA 300mA
×1 ×10 ×100
0 0.3V (16.7k/V) 0 3V 0 12V 0 30V (20k/V) 0 120V 0 300V 0 600V 0 12V 9k/V 0 30V 0 120V 9k/V 0 300V 0 600V 0 60μA (Terminal voltage:0.3V approx.) 0 30mA
(Terminal voltage:0.4V approx.)
0 300mA
(Terminal voltage:1V approx.)
1
0 3k(30at mid-scale) 0 30k(300at mid-scale) 0 300k(3kat mid-scale)
Measuring Range Accuracy
±3% of full scale value
±4% of full scale value
±3% of full scale value
±3% of full scale value
±3% of full scale length
±3% of scale length ±3% of scale length
±4% of scale length
2. Specifications
Measuring ranges and accuracy (23 ±5:, 75%RH or less)
1 : Small differences may result depending on the resistance of the fuse.2 : With optional temperature probe MODEL 7060.
Standards: :IEC 61010-1
Over-voltage category CAT.e 300V, pollution degree 2 Over-voltage category CAT.w 600V, pollution degree 2 IEC 61010-2-031 IEC 61326 (EMC)
Meter Movement: Self-shielding core magnet, taut-band type
Fuse: Fast acting type(F500mA/600V),φ6.3x32mm
Overload Protection: AC/DC600V: fused (Current/Resistance/0.3V/BATT. TEST
1.5V ranges) AC/DC720V: 10 seconds (600/300/120V ranges) AC/DC120V: 10 seconds (30/12V ranges) AC/DC30V:10 seconds(3V range)
Withstand Voltage: AC3700V for one minute between internal circuit and
housing case
Drop Protection: From a height of 1m onto concrete floor
Operating Temperature & Humidity:
0-40:, relative humidity up to 85% without condensation
Storage Temperature
& Humidity:
-10-50:, relative humidity up to 85% without condensation
Dimensions: 94(L)×140(W)×39(D)mm
Weight: Approx. 280g(including batteries)
Power Source: Two R6P(1.5V) or equivalent batteries
Accessories: Test leads MODEL 7066 ………………… 1
R6P battery…………………………………2 Fuse (F500mA/600V) ……………………2 Carrying case ………………………………1 Instruction manual …………………………1
Optional Accessories: Temperature probe MODEL 7060
3. Instrument Layout
Meter Cover
4. How to Read Scales
1 The thick portion of the scale indicates the allowable range of
voltage of a battery. (The lower limit of voltage for a 1.5V dry battery specified by IEC 60086 is 0.9V.) Note that satisfactory indication on the meter may not mean the battery has enough power for high load (high current consumption) application.
Range
DC.V
AC.V
DC.A
BATT.TEST 1.5V 1
TEMP.
LED
0.3V
3V
12V
30V
120V
300V
600V
12V
30V
120V
300V
600V 60μA
30mA
300mA
×1 ×10 ×100
B
B
C
B
C
B
C
C
B
C
B
C
C
B
B
A
A
A
E
D
30
30
12
30
12
30
6
12
30
12
30
6
6
30
30
2.0
-
20 150
×0.01 ×0.1 ×1 ×1 ×10 ×10 ×100 ×1 ×1 ×10 ×10 ×100 ×10 ×1 ×10 ×1 ×10 ×100 ×1 ×1
Scale Used
How to Read Scale
Regardless of indicated value
A
B
C
D
E
Pointer
Scale Plate
Meter Zero Adjust Screw
Test Lead Tips
Test Leads
Ohm Zero Adjust Knob
Terminal
Terminal
5. Preparation
Checking Test Leads and Fuse
Plug the red test lead into the + terminal and the black test lead into the -terminal, and set the function selector switch to a position in the area. Then, short the test lead tips. When the meter pointer deflects to the right, proceed to measurement.
When the meter pointer does not deflect, replace the fuse with the spare fuse, which is supplied with the instrument. If there is still no deflection, the test leads may have an open. Replace the test leads.
Adjusting Meter Pointer Zero
Turn the zero adjust screw to set the meter pointer to the "0" mark on the extreme left of the scale for accurate measurement.
Checking Function Selector Switch Setting
Make sure to set the function selector switch to the appropriate position. Otherwise, intended measurement cannot be made, or injury or instrument damage may result.
When the order of voltage or current under test is unknown, first make measurement on the highest range, and then select the appropriate range.
6. How to Make Measurements
Voltage Measurements (ACV, DCV)
Do not make measurement on a circuit above 600V AC or DC.
Do not apply voltage that exceeds the rated voltage of the selected range.
Do not turn the function selector switch during measurement.
Do not touch the metal parts of the test leads during measurement.
When the order of the voltage under test is unknown, make measurement
on the highest range.
(1)
Plug the red test lead into the + terminal and the black test lead into the -terminal. (2)Set the function selector switch to the appropriate DCV or ACV position. (3)Connect the test leads to the circuit under test so the instrument is in parallel
with the circuit. In DCV measurement, the meter pointer deflects to the right
(normal direction) when the red test lead is connected to the positive side of the
circuit under test and the black test lead to the negative side of the circuit.
Connecting the test leads the other way will reverse the pointer deflection. (4)Take reading on the appropriate scale.
Current Measurements (DCA)
Exercise caution not to apply voltage to the instrument set to a current range.
Do not turn the function selector switch knob during measurement.
Make sure to firmly connect the test leads to the circuit under test so that the
connections will not become loose. The test leads must be connected to or removed from the circuit under test with the circuit powered off.
When the order of the current under test is unknown, make measurement on
the highest range.
(1)
Plug the red test lead into the + terminal and the black test lead into the -terminal. (2)Set the function selector switch to the appropriate DCA position. (3)Power off the circuit under test. (4)Connect the test leads to the circuit under test so the instrument is in series with
the circuit. In DCA measurement, the meter pointer deflects to the right (normal
DANGER
DANGER
direction) when the red test lead is connected to the positive side of the circuit under test and the black test lead to the negative side of the circuit. Connecting
the test leads the other way will reverse the pointer deflection. (5)Power on the circuit under test. (6)Take reading on the appropriate scale.
Resistance/Continuity Check
Exercise caution not to apply voltage to the instrument set to a resistance range
Make sure to power off the circuit under test.
Resistance Measurement
(1)
Plug the red test lead into the + terminal and the black test lead into the -terminal. (2)Set the function selector switch to the appropriate resistance position. (3)Short the test lead tips. Turn the Ohm Zero Adjust Knob to set the meter
pointer to the "0" mark on the extreme right of the scale for accurate
measurement. Make this adjustment whenever the function selector switch is
turned to a different resistance position. When the zero adjustment cannot be
made, replace the batteries. (4)Connect the test leads to the circuit under test. (5)Take reading using the appropriate multiplying factor.
Note: Note that keeping the test lead tips shorted will exhaust the internal batteries.
Continuity Check
(1)
Plug the red test lead into the + terminal and the black test lead into the -terminal. (2)Set the function selector switch to the ' ' position. (3)Connect the test leads to the circuit under test. (4)Check if there is a sound of the buzzer. The buzzer beeps below about 100Ω.
Note: The meter does not read on this range.
LED Check
(1)
Plug the red test lead into the +terminal and the black test lead into the - terminal. (2)Set the function selector switch to the x10 position. (3)Connect the test leads to the LED under the light-up test. (4)When the LED does not light up, connect the test leads the other way.
Note: Connect the red test lead to the anode of the LED and the black test lead
to the cathode. The meter deflection has no meaning on this range.
Temperature (TEMP.) Measurements
(1)
Plug the red test lead into the +terminal and the black test lead into the - terminal. (2)Set the function selector switch to the x10 position. (3)
Short the test lead tips. Turn the Ohm Zero Adjust Knob to set the meter pointer
to the "0" mark on the extreme right of the scale for accurate measurement. (4)Remove both red and black test reads from the instrument. (5)Plug the red lead of temperature probe Model 7060 to the + terminal and the
black lead to the-terminal. (6)Touch the object under test with the tip of the temperature probe and take
reading on the TEMP scale.
Battery Test (BATT. TEST 1.5V)
This range measures the voltage of a battery, applying a load similar to that used in common applications (load resistance: 10).
Do not apply voltage above the rated voltage for the battery test.
Do not turn the function selector switch during a test.
(1)
Plug the red test lead into the + terminal and the black test lead into the - terminal.
DANGER
DANGER
.
(2)Set the function selector switch to the 'BATT. TEST 1.5V' position. (3)Connect the red test leads to the positive side of the battery and the black test
lead to the negative side of the battery.
(4)Take reading on the BATT TEST scale. Note: The less power a battery has, the lower the reading on this range becomes
compared to the reading on the 3V DC range.
7. Battery & Battery Replacement
To avoid electric shock hazard, be sure to set the function selector switch to
the OFF position and remove the test leads from the instrument.
Replacement fuse must have the following rating. Fast acting type, F500mA/600V,φ6.3 ×32mm
Do not mix new and old batteries.
Install the batteries observing correct polarity shown inside the instrument.
(1)Remove the test leads from the terminals on the instrument and set the function
selector switch to the OFF position.
(2)Loosen the screw on the bottom of the instrument and remove the bottom case
from the instrument.
(3)Replace the batteries or fuse with new one(s). Use;
Two R6P dry batteries A fast acting type fuse: F500mA/600V,φ6.3 ×32mm
(4)Mount the bottom case and tighten the screw.
Note: Use the spare fuse stored inside the instrument. After using the spare fuse,
store a new one as a new spare.
Fuse
WARNING
CAUTION
−
+
+
Spare Fuse
Bottom Case
−
Front Panel
Function Selector Switch
Carrying Case
DISTRIBUTOR
Kyoritsu reserves the rights to change specifications or designs described in this manual without notice and without obligations.
10―07 92―1430B