KYORITSU 1109S User Manual

INSTRUCTION MANUAL
ANALOGUE MULTIMETER
KEW 1109S
Table of Contents
Page
1. Safety Warnings ............................................................................... 1
• Understanding Some of the Basics in Electrical Testing

Before Using the Multimeter .......................................................... 3
2. Features ........................................................................................... 5
3. Specifications ....................................................................................6
4. Instrument Layout ............................................................................. 8
5. How to Read Scales ......................................................................... 9
6. Operating Instructions ..................................................................... 11
6-1 Preparation ................................................................................ 11
6-2 DC Voltage Measurements ...................................................... 12
6-3 AC Voltage Measurements ....................................................... 13
6-4 Low Frequency (dB) Measurements ........................................ 14
6-5 AC Output Voltage Measurements Using Output
Terminal ................................................................................... 15
6-6 DC Current Measurements ...................................................... 16
6-7 Resistance Measurements ....................................................... 17
6-8 Terminal Current LI, Terminal Voltage LV &
Diode Test ................................................................................ 20
6-9 Measuring I
6-10 Measuring h
7. Fuse & Battery Replacement .......................................................... 27
(leakage current of transistors) ......................... 22
CEO
(DC Current Amplification Factor) ................... 24
FE
1. Safety Warnings
The instrument must be used by a competent, trained p erson and operated in strict accordance with the instructions. Kyoritsu Electrical Instruments Works, Ltd will not accept liability for any damage or injury caused by misuse or non-compliance with the instructions or safety procedures. It is essential to read and understand the safety rules contained in the instructions. They must be observed when using the instrument.
This instruction manual contains warnings and safety rules
which must be observed by the user to ensure safe operation of the instrument and retain it in safe condition for a long period. Therefore, read these operating instructions thoroughly and completely before using the instrument.
WARNING
#
This is a warning for the user to avoid electric shock hazard. The symbol # on the instrument means that the user must read the instructions in this manual for safe operation of the instrument.
CAUTION
#
This is a caution for the user to avoid damage to the instrument.
WARNING
#
● Never open the back case during measurements.
● Never use the instrument to measure voltage higher than 250V
on an industrial power line.
● When meas ur in g a high voltage greater t ha n 250V (in s ma ll
power supply circuit and not on power transmission line), connect the test leads to the circuit under test after it is once de-energized. Do not touch the wiring or test leads by hand during voltage measurements. After the measurements, turn off power to the circuit under test and disconnect the test leads.
Never test a circuit voltage higher than 250V with the rang e
selector switch erroneously set to one of the current or resistance range positions. The fuse may not protect the circuit.
 1 
● Never use the instrument in an explosive atmosphere especially
when making current measurements.
CAUTION
#
● Before makin g mea surements check that the range selec tor
switch is at a proper range position. Make sure to remove the tips of the test leads from the circuit under test when changing the measuring range during measurements.
● Do not apply voltage to the current or resistance ranges. It may
result in a fuse blow or instrument damage.
● Make certain to set the range selector switch to the OFF position
after every use.
● It is recommended that the range selector switch should be set to the 250mA DC position to protect the instrument against the
possible shock or vibration in transit.
Note: Take good care not to make voltage measurements with
the range selector switch at the 250mA DC position. The
fuse may blow or instrument get damaged.
● Do not exp o se the ins t r ument to the di r e ct sun, ext r eme
temperature and humidity or dew fall for a long period. Also, care must be taken not to give a shock to the instrument by dropping or inadequate handling.
● Since the meter cover has been given an anti-static treatment
do not rub it strongly with a dry cloth. Where anti-static charges are present on the meter window after it has been used for a long period of time, causing the meter pointer to deflect in an abnormal way, coat its surface with anti-static chemicals for plastics or wipe it with a dilute solution of anti-static fluid as a temporary measure.
● Understanding Some of the Basics in Electrical Testing Before
Using the Multimeter
 2 
● Auxiliary Units (Prefixes)
There are a number of measurement units used for multimeters. Vol t (V), am p e r e (A) an d ohm (Ω ) are mo s t widely us e d as
meas ure ment un its to indi cat e electri cal p otent ial , current a nd resistance. However, it is not always straitforward to handle these units as they sometimes too large or too small for practical use or calculation. Prefixes are therefore used as auxiliary units to simplify the usage of such measurement units.
The following table shows some of the examples:
Auxiliary Unit M k m μ n P
Read mega- kilo- milli- micro- nano- pico-
Multiply 10
Example
6
2MΩ
2000kΩ
3
10
2kΩ
2000Ω
-3
10
250mA
0.25A
-6
10
50μA
0.05mA
10-9 10
200nF
0.2μF
-12
1000pF
0.001μF
● Notes on Voltage & Current Measurements
It is important to understand the basic differences between current
and voltage measurements for proper use of multimeters.
Voltage measurements are designed to detect potential difference
between two points. Make certain that the multimeter is connected in parallel with the circuit under test.
Current measurements are intended to monitor the consumption of
current in the circuit resulting from the application of voltage. Make sure to connect the multimeter in series with the circuit under test.
Generally speaking, the internal resistance of a voltmeter should
preferably be larger, while that of an ammeter should be smaller. If the multimeter is erroneously connected in parallel with a circuit for current measurements, the likely result will be the flow of excessive current and aubsequent damage to the fuses and other electronic components. To avoid such a potential danger it is necessary to have a good understanding of voltage and current measurements.
 3 
Voltage Measurements Current Measurements
IR+I
M  IR
> > I
M
R
M・IR
< < E
R
M・IR
I
M
I
R
I
R
Power Source
Power Source
E E Load Load
R
M
=
DC.V Range DC.mA Range
COM
COM
RM = 0
V
mA
mA
Battery (Power Source)
Battery (Power Source)
Battery (Power Source)
Battery (Power Source)
Resistor
(Load)
Resistor
(Load)
(Parallel Connection) (Series Connection)
Measuring voltage at both ends of a resistor.
.......... Meter Current
I
M
I
.......... Load Current
R
E .......... Power Source Voltage
RM ........ Internal Resistance of Multimeter
M e a su r in g c u r r e n t b ei ng consumed by a resistor.
Fig. 1
 4 
2. Features
● Mirrored scale for easy and accurate reading.
● 19 measuring ranges for a wide scope of application.
● OUTPUT terminal to cut off the DC component of AC voltage being measured.
● hFE scale for transistor checking.
● dB scale (−10 - +62dB).
● Safety designed input terminals and test leads.
 5 
3. Specifications
Functions Measuring Ranges Accuracy
0-0.1 V 0-0.5V 0-2.5V
DC Voltage (7 ranges)
AC Voltage (4 ranges)
DC Current (4 ranges)
Resistance (4 ranges)
Current across Terminals at Resistance Range (LI) (4 ranges)
Voltage across Terminals at Resistance Ranges (LV) (4 ranges)
0-10V (20kΩ/V)
0-50V 0-250V 0-1000V
0-10V
0-50V (9kΩ/V)
0-250V 0-1000V
0-50μA
0-2.5mA
0-25mA
0-250mA
0-2kΩ (20Ω mid-scale) 0-20kΩ (200Ω mid-scale) 0-2MΩ (20kΩ mid-scale) 0-20MΩ (200kΩ mid-scale)
0-150mA (x1 range) 0-15mA (x10 range) 0-150μA (x1k range)
0-60μA (x10k range)
0-3V (x1 , x10, x1 k ranges)
0-12V (x10k)
terminal voltage
(
approx. 100mV
terminal voltage
(
approx. 100mV
terminal voltage
(
approx. 150mV
terminal voltage
(
approx. 550mV
) ) ) )
±3% of full scale
±3% of full scale
±3%
of full scale
±3% of scale length
±5% of scale length (Battery voltage at 3V)
(Battery voltage at 12V)
±5% of scale length (Battery voltage at 3V)
(Battery voltage at 12 V)
 6 
Low Frequency Output Using OUTPUT Terminal
Low Frequency Output (dB) (4 ranges)
0-10V 0-50V 0-250V 0-1000V
10V AC
10 - +22dB
50V AC +4 - +36dB 250V AC +18 - +50dB 1000V AC +30 - +62dB (0 dB = 0.775V(1mW)
Refer to frequency characteristic chart
Refer to frequency characteristic chart
across a 600Ω line)
DC Current Amplication Factor (hFE) (1 range)
hFE: 0-1000 (atΩ× 10 range)
±3% of scale length
Dimensions: 100 (W) x 150 (L) x 47 (D)mm Weight: Approx. 330g Accessories: Test Leads (Model 7210) - 1 set
F250V 0.5A fuse - 2 pcs. (including one spare fuse
also installed into the housing case)
1.5V Battery Type R-6P, SUM-3, AA or equivalent -
2 pcs. (Installed into the instrument) 9V Battery Type 6F22, 006P PP3, or equivallent
- 1 pce (installed into the instrument) Instruction Manual - 1 copy
 7 
4. Instrument Layout
⑦ ⑧
OUTPUT Terminal
-COM (N) Terminal
Mirrored Scale Plate
Housing Case
Range Selector Switch
Meter Pointer
Meter Zero Adjust Screw
+(P) Terminal
Front Panel
Meter Window
Ohm Zero Adjust Knob
 8 
5. How to Read Scales
DC Voltage (7 ranges)
AC Voltage (4 ranges)
DC Current (4 ranges)
Resistance (4 ranges)
Current across Terminals (LI)(4 ranges) Voltage across Terminals (LV)(4 ranges)
Measuring Ranges
0.1V
0.5V
2.5V 10V 50V 250V 1000V 10V 50V 250V 1000V
50μA
2.5mA 25mA 250mA X1Ω x10Ω x1kΩ X10kΩ X1Ω X10Ω X1kΩ X10kΩ x1Ω X10Ω X1kΩ X10kΩ
Scales
B 10 B50 B 250 B 10 B50 B250 B10
C10
B50 B250 B10 B50 B250 B250 B250 A 0-2k A0-2k A0-2k A0-2k E15 E15 E15 E15
F3 F3 F3 F3
H o w t o Read Scales
X0.01 X0.01 X0.01 X1 X1 X1 X100 X1 X1 X1 X100 X 1 X0.01 X0.1 X1 X1 X10
X1k X10k X10(mA) X1(mA) X 10(μA) X 4(μA) X1(V) X1(V) X1(V) X4(V)
input Terminals
+&
COM
+&
COM
+&
COM
+&
COM
+&−COM
+&−COM
 9 
Loading...
+ 25 hidden pages