KYORITSU 1052 User Manual

INSTRUCTION MANUAL
DIGITAL MULTIMETER
KEW 1051
1052
Thank you for purchasing our Model KEW1051, KEW1052 Digital Multimeter.
This instruction manual describes the specications and handling precaution for this Digital
Always observe the following instructions. Failure to do so may impair the protection provided by the instrument and probes, and may result in electrical shock or other dangers that may lead to serious injury or the loss of life. KYORITSU is in no way liable for any damage resulting from the user’s mishandling of the product. For safe use of this product, the following safety symbols are used on the product:
About This Manual
n
• Every effort has been made to ensure accuracy in the preparation of this manual.
However, should any errors or omissions come to the attention of the user, please contact
KYORITSU.
• T he conten ts of this ma nual are su bject to ch ange wi th out pri or notice bec au se of
improvement in performance or function.
• A ll rights reserved. No par t of this manual may be reproduced in any from without
KYORITSU's written permission.
1
Regarding Safe Use of This Product
W ARNIN G
CAUTION
For safe use of this product, the following safety symbols are used on the product and manual:
This indicates that the operator must refer to an explanation in the
instruction manual in order to avoid the risk of serious injury or the loss of life.
This indicates that the operator must refer to an explanation in the
instruction manual in order to avoid the risk of injury or damage to the product.
This indicates information that is essential for handling the instrument
Note
or should be noted in order to familiarize yourself with the instrument’s operating procedures and/or functions.
Danger! Handle with Care
Double Insulation
Direct Current
Alternating Current
DC/AC
Fuse
Battery
Ground
This symbol indicates that the operator must refer to an explanation in the instruction manual in order to avoid risk of injury or death of personnel or damage to the instrument.
This symbol indicates double insulation or reinforced insulation.
This symbol indicates DC voltage/current.
This symbol indicates AC voltage/current.
This symbol indicates AC and DC.
This symbol indicates a fuse.
This symbol indicates a battery.
This symbol indicates ground (earth).
2
W ARNIN G
Always observe the following instructions. Failure to do so may result in
n
electrical shock or other dangers that may lead to serious injury or the loss of life.
Test leads / Test leads with crocodile clip (optional accessory)
• Use the probes supplied by KYORITSU with this instrument.
• Do not use test leads/test leads with crocodile clip that have deteriorated or are defective.
Check test leads/test leads with crocodile clip continuity.
• Disconnect test leads/test leads with crocodile clip from the circuit under test before opening
the casing to replace the batteries or for any other reason.
• Disconnect test leads/test leads with crocodile clip from the c ircu it under test before
attaching/detaching the test leads/test leads with crocodile clip to/from the instrument.
• Disconnect test leads/test leads with crocodile clip from the instrument before opening the
casing to replace the batteries or for any other reason.
• A cap is provided on the tip of a test lead. Use a test lead with the cap on for safety (safety standards: IEC 61010-031).
• Do not use the crocodile clip of test leads being loosen or removed conditions.
Casing
• Do not use the instrument if there is any damage t o the casing or when the casing is
removed.
Fuses
• Use fuses of the specied rating when the fuse is replaced.
Operating Environment
• Do not operate the instrument in an atmosphere where any ammable or explosive gas is
present.
• Avoid using the instrument if it has been exposed to rain or moisture or if your hands are
wet.
Disassembly
• No person, except personnel from KYORITSU, is authorized to disassemble this instrument.
3
Contents
1. Overview .............................................................................................................5
2. Measurement Category .......................................................................................6
3. Specications ...................................................................................................... 7
3.1 General Specications ..............................................................................7
3.2 Accuracy ...................................................................................................9
4. Operation ..........................................................................................................12
4.1 Precautions Before Measurement ...........................................................12
4.2 Components ............................................................................................13
4.3 Measuring Instructions............................................................................ 17
4.3.1 AC Voltage Measurement ................................................................ 17
4.3.2 DC Voltage Measurement ................................................................ 17
4.3.3 Measurements with SENSOR .......................................................... 18
4.3.4 Resistance Measurement ..................................................................18
4.3.5 Continuity Check ............................................................................. 19
4.3.6 Diode Test ........................................................................................19
4.3.7 Temperature Measurement...............................................................20
4.3.8 Current Measurement ......................................................................21
4.3.9 Capacitor Measurement ................................................................... 22
4.3.10 Frequency Measurement ................................................................. 22
4.3.11
Function to change RMS detection to/from MEAN detection
mode (KEW1052 only) .........................................................................23
4.3.12 Function to turn the lter on/off ...................................................... 23
4.3.13 AUTO HOLD Function ................................................................... 24
4.3.14 Relative and percentage calculation ................................................ 24
4.3.15 MIN/MAX/AVG Function (KEW1052 only) .................................25
4.4 Memory Function (KEW1052 only) ....................................................... 26
4.5 AUTO POWER OFF Function ............................................................... 27
4.6 Set-up Function ....................................................................................... 28
4.7 Additional functions simply set when POWER ON ............................... 32
4.8 LCD Check .............................................................................................32
5. User Calibration Function .................................................................................33
6. Battery and Fuse Replacement .........................................................................35
6.1 Battery Replacement ............................................................................... 35
6.2 Fuse Replacement ...................................................................................36
7. Calibration and Maintenance ............................................................................37
8 Disposing the Product .......................................................................................37
4
1. Overview
Display
l
4-digit (LCD) Maximum Reading: 6000 Bar graph indicator
Supports a variety of measurement function
l
Measurement function
DC Voltage, AC Voltage, DC Current, AC Current, Resistance, Frequency, Temperature, Capacitor, Continuity Check, Diode Test
Other functions
Data Hold (D•H), Auto Hold (A•H), Range Hold (R•H), Maximum value* (MAX), Minimum value* (MIN), Average value* (AVG),
Zero Adjustment (Capacitor, Resistance), Relative values, Save to Memory*, LCD backlight. *: For model KEW1052 only
Switching detection modes
l
Effective value (root mean square value) detection (RMS) and mean value detection (MEAN)
can be switched during AC voltage measurement (KEW1052 only).
 Low-passlter
l
The low-pass lter can be switched on/off during AC voltage or AC current measurement.
Communication: optional communication set is required
l
• Measurement data can be transferred to a PC using an optional USB communication
package.
The data can be read by certain applications to make trend graphs or can be converted
into Excel les.
• The data can also be output from an optional printer via an optional printer communication
set.
(KEW1052 only).
Safety design
l
Complied standards: CE standards Uses a current-input terminal shutter for preventing wrong input. Uses high-performance UL-standard fuses.
5
2. Measurement Category
W ARNIN G
Internal Wiring
Outlet
Entrance Cable
Fixed Equipment, etc.
Distribution Board
CAT . II
Equipment
CAT.I
CAT .IV
CAT . III
Measurement Category (CAT.)
n
The restrictions on the maximum voltage level for which the KEW1051, KEW1052 can
be used, depend on the measurement categories specied by the safety standards.
Do not apply any input level higher than maximum allowable input.
AC/DC 1000V CAT.III AC/DC 600V CAT.IV
Measurement
category
CAT. I For measurements performed on circuits not
Description Remarks
directly connected to MAINS.
CAT. II For measurements performed on circuits directly
connected to the low voltage installation.
CAT.III For measurements performed in the building
installation.
CAT.Ⅳ For measurements performed all the source of the
low-voltage installation.
Appliances, portable equipment, ect.
Distribution board, circuit breaker, ect.
Overhead wire, cable systems, ect.
Note
Radiation immunity affects the accuracy of the KEW1051, KEW1052 under the
conditions specied in IEC61326-1.
Use of this instrument is limited to domestic, commercial, and light industry
applications. If equipment generating strong electromagnetic interference is located nearly, the instrument may malfunction.
6
3. Specications
3.1 GeneralSpecications
Measurement function: DC Voltage, AC Voltage, DC Current, AC Current, Resistance,
Frequency, Temperature, Capacitor, Continuity Check, Diode Test
Other functions: Data Hold (D•H), Auto Hold (A•H), Range Hold (R•H), Maximum
value* (MAX), Minimum value* (MIN), Average value* (AVG),
Zero Adjustment (Capacitor, Resistance), Relative values, Save to
Memory*, LCD backlight. *: For model KEW1052 only Measuring method: Σ modulation Display: 4-digit (LCD)/7-segment Maximum Reading: 6000 Polarity Indicator: “–” Appears automatically when the polarity is
negative Overrange Indicator: “ OL ” Low-battery Indicator: “
low Measurement cycle: 5 times per second (except frequency measurement : one time per second, Resistance
me as urement (6M Ω/ 60MΩ) : 2 .5 time s pe r se co nd, cap ac it or
measurement (1000μF) : max.0.14 time per second) Bar graph display approx 25 times per second (at AC, Ω)
Operating temperature and humidity ranges:
-10 to 55ºC, 80%RH or less (no condensation) 70%RH or less at 40 to 55ºC. Storage temperature and humidity ranges:
-30 to 70ºC, 70%RH or less (no condensation) Temperature coefcient:
(Accuracy at 23±5ºC× 0.1)/ºC should be added. (Temperature ranges: -10 to 18ºC and 28 to 55ºC) Power supply: AA-size (R6/LR6) 1.5V batteries: 4 Battery life: Approximately 300 hours (Operating hours of alkaline batteries when in DC voltage-mode.)
Note: The battery life varies depending on the operating conditions. Insulation resistance : 1000V DC, 100MΩ or more
” Appears when the batteries become
7
Withstand voltage: 6.88kVrms AC for ve seconds
(across input terminals and casing) External dimensions: Approximately 192(L)×90(W)×49(D)mm Weight: Approximately 560g (including batteries) Complied standards: Safety standards IEC61010-1, IEC61010-031 CAT.III (Max. input voltage: AC/DC1000V) CAT.IV (Max. input voltage: AC/DC600V) Pollution degree 2, indoor use, 2000m max. above sea level EMC standards IEC61326-1 Class B Effect of radiation immunity:
In the radio-frequency electromagnetic eld of 3 V/m, accuracy is
within ve times the rated accuracy.
Standard accessories: Batteries : 4 Test leads: 1set (M-7220) Fuse (included): 440mA/1000V (M-8926), 10A/1000V (M-8927) Instruction manual: 1 Blank cover: 1
Optional accessories: Carrying case M-9150
(for the main unit with test leads and
communication cable) Test leads (1set) M-7220
Fuse 440mA/1000V M-8926 10A/1000V M-8927 Temperature probes M-8405, 8406, 8407, 8408
Test leads with crocodile clip (1 set)
M-7234
Followings are for KEW1052 only. USB Communication set M-8241 (Software, USB adapter and
cable) Printer communication set M-8243 (Printer Adapter and Cable) Printer M-8246
AC adapter (for printer, Europe) Thermal paper for printer (10 rolls)
M-8248
M-8247
8
3.2 Accuracy
Test conditions: Temperature and humidity: 23±5ºC at 80%RH or less Accuracy: ±(% of reading + digits)
Note: Each response time is a value to rated accuracy within selected range.
DC Voltage Measurement V
Range Resolution Accuracy Input Resistance
600mV 0.1mV
6V 0.001V 11MΩ
60V 0.01V
0.09+2
10MΩ
10MΩ600V 0.1V
1000V 1V 0.15+2
NMRR: 60dB or more 50/60Hz±0.1% CMRR: 120dB or more 50/60Hz (Rs=1kΩ)
Response time: 1 sec max.
AC Voltage Measurement V
AC Coupling: RMS value detection, sine wave
MEAN value detection and RMS value calibration (KEW1052 only)
Range
Resolution
50/60Hz 40Hz to 500Hz 500Hz to 1kHz
600mV 0.1mV
6V 0.001V 11MΩ, <50pF
60V 0.01V
0.5+5 1+5
Accuracy
Impedance
10MΩ, <200pF
1.5+5 10MΩ, <50pF600V 0.1V
1000V 1V -
Accuracy: At 5 to 100% of range and 1000V range is 200 to 1000V. less than 1500V peak
For non-sinusoidal waveforms, add ±(2% + 2% of full scale), for Crest factor<3. CMRR: 60dB or more DC to 60Hz (Rs=1kΩ)
4 counts or less is corrected to 0, Response time: 2 sec max.
DC Current Measurement A
Range Resolution Accuracy Voltage Drop
600μA 0.1μA
6000μA 1μA
60mA 0.01mA
440mA 0.1mA
10A 0.01A
0.2+2
0.5+56A 0.001A
< 0.12mV/μA
< 3.3mV/mA
< 0.1V/A
Response time: 1 sec max.
Maximum
Input Voltage
1000V
DC
1000V
rms AC
Input
Maximum
Input
Voltage
1000V
rms AC
1000V
DC
Maximum
Input Current
440mA Protected by
a 440mA/1000V fuse.
10A Protected by
a 10A/1000V fuse.
9
AC Current Measurement [RMS] A
RMS value detection, sine wave
Range Resolution
600μA 0.1μA
6000μA 1μA
60mA 0.01mA
440mA 0.1mA
6A 0.001A
10A 0.01A
Accuracy
50/60Hz 40Hz to 1kHz
0.75+5 1.5+5
Voltage
Drop
< 0.12mV/μA
< 3.3mV/mA
< 0.1V/A
Maximum
Input Current
440mA Protected by a
440mA/1000V fuse.
10A Protected by a
10A/1000V fuse.
Accuracy: At 5 to 100% of range, 10A range is 2 to 10A and 440mA range is 30 to 440mA.
For non-sinusoidal waveforms, add ±(2% + 2% of full scale), for Crest factor<3.
4 counts or less is corrected to 0, Response time: 3 sec max.
ResistanceMeasurementΩ
Range Resolution Accuracy
Measuring
Current
Maximum
600Ω 0.1Ω
6kΩ 0.001kΩ <110μA
60kΩ 0.01kΩ <13μA
0.4+1
< 1.2mA < 3.5V
600kΩ 0.1kΩ <1.3μA
6MΩ 0.001MΩ 0.5+1
60MΩ 0.01MΩ
1+2 (~40MΩ)
2+2 (40~60MΩ)
<130nA
Open-loop
Voltage
< 1.3V
Input
Protective
Voltage
1000V
rms
Accuracy is specied after zero adjustment at 600Ω to 6kΩ (Resistance) Response time: 2 sec max. at 600Ω to 600kΩ, 10 sec max. at 6M to 60MΩ
Continuity Check
Range Resolution Range of Operation
The buzzer turns on
600Ω 0.1Ω
for resistances lower
than 50±30Ω.
Measuring
Current
Approx.
<1.2mA
Open-loop
Voltage
Input Protective
Voltage
<3.5V 1000V rms
Diode Test
Range Resolution Accuracy
Measuring Current
(Vf=0.6V)
Open-loop
Voltage
Input Protective
2V 0.001V 1+2 Approx. 0.5mA <3.5V 1000V rms
10
Voltage
Temperature Measurement TEMP
Range Resolution Accuracy
Input Protective
Voltage
-50 ~ 600°C 0.1°C 2+2°C
-58~ 999.9°F
-58~ 1112°F
0.1°F 1°F
2+3.6°F
2+3°F
1000V rms
Use optional Temperature Probe: Thermocouple Type K
Capacitor Measurement
Range Resolution Accuracy
Input Protective
Voltage
10nF 0.01nF 2+10
100nF 0.1nF
10μF 0.01μF
100μF 0.1μF
1000μF 1μF
2+51μF 0.001μF
3+5
1000V rms
Accuracy is specied after zero adjustment at 10n to 1μF (Capacitance).
Frequency Measurement Hz
AC Coupling, Maximum Reading 9999
Range Resolution Accuracy Input Voltage
10.00 to 99.99Hz 0.01Hz
90.0 to 999.9Hz 0.1Hz
0.900 to 9.999kHz 0.001kHz 0.4 to 600Vrms
0.02+1
0.2 to 600Vrms
9.00 to 99.99kHz 0.01kHz 0.8 to 100Vrms
11