The CS-7 is a software program for installation on a PC. It performs image processing of X-ray images received from
an AeroDR SYSTEM or REGIUS system.
The CS-7 adds patient data received from RIS etc. and sends diagnosis images to PACS etc. Furthermore, the CS-7
can send requests for exposure conditions to the X-ray device and receive irradiation results.
This operation manual describes instructions on the basic functions for operation of this device. Those operating this
device for the rst time should read this manual beforehand.
Also, store this manual close to this device after reading it through, so it can be used as a guide to allow optimum
operating conditions.
*
If the pages of the operation manual are smudged and illegible, replace it with a new one. (There is a fee
for this service.)
IMPORTANT
• For United States of America, this device is not intended to use for Mammography.
• One Shot stitching exposure function can only be used in some countries.
– This function cannot be used in the United States or Canada.
– This function cannot be used in the EU, EFTA member countries, Switzerland or Turkey.
• In this manual, "this device" refers to a collective term indicating CS-7 software and PC with the software installed.
Summary of usability specications (for IEC/EN 62366)
(1) Medical purposes
• Provision of disease and injury diagnostic images.
(2) Patient groups
• No patient population exists who uses or is in contact with the device.
• Patient population for the X-ray images read is not specied.
(3) Parts of body or organizations to which the device is mounted or that interact with the device
• No appropriate points. (Device does not come in contact with the patient.)
(4) Operating principle
• This device CS-7 is a software program for installation and use in commercially available PCs.
Commercially available PCs consist of a PC main body and an LCD monitor. This device receives
X-ray image information from the reader device (REGIUS series, DR Detector) and connects it with
patient information, which will be output after processing into diagnosable images by an operator.
(5) Signicant physical characteristics
• Refer to "12.1 Specifications".
(6) Signicant performance characteristics
• Refer to "12.1 Specifications".
(7) User of this device
• No special training is required to use this device. The intended users of this device are as follows.
– A professional in good health with specialist knowledge/qualications who has fully understood the
content of this document. (Such as a doctor or radiological technologist)
10
Page 11
Introduction
Disclaimer
(1) This manual may not be reproduced in whole or in part without the permission of Konica Minolta, Inc.
(2) The contents of this manual may be subject to change without prior notice.
(3) Konica Minolta, Inc. is not responsible for any claims made for malfunction or damage caused by installation,
relocation, modification, maintenance, and repair made by anyone except Konica Minolta and contractors
designated by Konica Minolta.
(4) Konica Minolta, Inc. is not responsible for any claims made for malfunction or damage to Konica Minolta
products, caused by third-party products not installed by Konica Minolta.
(5) Konica Minolta, Inc. is not responsible for any claims made for malfunction or damage caused by maintenance
or repair using maintenance parts other than those specied by Konica Minolta.
(6) Konica Minolta, Inc. is not responsible for any claims made for malfunction or damage caused by not observing
the precautions and operation methods described in the operation manual.
(7) Konica Minolta, Inc. is not responsible for any claims made for malfunction or damage caused by the
environment, such as power source and installation environment, that is described in the Installation
Requirement or operation manual, that is not suitable for the installation requirements.
(8) Konica Minolta, Inc. is not responsible for any claims for malfunction or damage caused by acts of nature such
as res, earthquakes, oods, or lightning strikes.
(9) Konica Minolta, Inc. is not responsible for any claims for malfunction or damage caused by using this device for
any purpose other than that specied for this device.
(
10
) Diagnostic and treatment action is performed under the responsibility of the physician(s). Konica Minolta, Inc. is
not responsible for any diagnostic/treatment conditions or diagnostic/treatment results.
(11) Konica Minolta, Inc. is unable to assume any liability relating to the following events arising due to systems
infected by computer viruses, worms and other malicious software potentially harmful to computers (hereunder
referred to as "malwares"):
- The erasure, corruption, or unwanted disclosure of data stored within this device, including personal and
sensitive patient information.
- Malfunction of the device or accidents occurring as a result of the malfunction of the device
- Infection of other products upon connection with the device, or damages arising therefrom.
For certain functions of this device, separate options and licenses are needed. For details regarding options and licenses, contact Konica Minolta service representatives.
11
Page 12
Introduction
Term description
The meanings of terms used in this operation manual are as follows:
TermsDescription
• Collective term indicating DR Detector, AeroDR Interface Unit, AeroDR Interface Unit2, AeroDR
AeroDR SYSTEM
DR Detector
AeroDR Interface Unit
AeroDR Interface Unit2
AeroDR Generator Interface UnitRelays signals between the X-ray device, the DR Detector, and this device.
AeroDR Generator Interface Unit2Relays signals between the X-ray device, the DR Detector, and this device.
AeroDR Battery ChargerCharges the DR Detector. It also has the registration function for the DR Detector.
AeroDR Battery Charger2Charges the DR Detector. It also has the registration function for the DR Detector.
Access pointCollective term indicating AeroDR Access Point and a general-purpose access point.
I/F Cable
AeroDR UF Cable
AeroDR Portable RF Unit
AeroDR Portable RF Unit2
AeroDR Portable UF UnitReceives an image data from the DR Detector and transfers it to this device.
AeroDR Stitching SystemUsed when performing stitching exposures using the DR Detector.
CalibrationProcessing for calibrating the characteristics of the DR Detector for each pixel
CRAn acronym for computed radiography
DRAn acronym for digital radiography
JMAn acronym for job manager
REGIUS
REGIUS SIGMA/SIGMA2Used as the collective terms for "REGIUS SIGMA" and "REGIUS SIGMA2".
CR cassette
Generator Interface Unit, AeroDR Generator Interface Unit2, AeroDR Battery Charger, AeroDR
Battery Charger2, AeroDR Access Point, etc.
• Collective term indicating AeroDR SYSTEM, AeroDR SYSTEM 2 and SKR3000.
Supplies power to the AeroDR Generator Interface Unit, AeroDR Generator Interface Unit2,
and access point, and supplies power to and charges the battery of the DR Detector when
wired cables are used. It also has a hub function.
• Supplies power to the access point, and supplies power to and charges the battery of the
DR Detector when wired cables are used. It also has a hub function.
• Relays signals between the X-ray device, the DR Detector, and the CS-7.
Use with the DR detector wired connection. It also includes the DR detector charging and
registration functions.
In the case of capturing images with the wall stand and table, register the DR detector
direction.
Receives an image data from the DR Detector and transfers it to this device. Also, this unit
interfaces with the portable X-ray device.
• Interfaces with the mobile X-ray device.
• Charges or registers the DR Detector when using the AeroDR Battery Charging KIT and
AeroDR I/F Cable2.
Collective term indicating the REGIUS series (Excluding "REGIUS SIGMA" and "REGIUS
SIGMA2").
When indicated individually, they are shown as "REGIUS 110", "REGIUS 110HQ",
"REGIUS 210", "REGIUS 190", and "REGIUS 170".
Used as the collective terms for the cassettes in the REGIUS series and REGIUS SIGMA/
SIGMA2.
12
Page 13
TermsDescription
Abbreviation for Radiology Information System
RIS
HostA server to save image information
X-ray device
HS processing
F processing
E processing
IG
Exposure body partThis is the exposure body part, such as head, neck, and chest.
Exam. Tag
Exposure condition
Exposure setting
Wired cableCollective term indicating I/F Cable and AeroDR UF Cable.
Aero Sync
Stitching holder
A system to manage reservation and result of examinations and treatment that use radiology equipment
An X-ray exposure device constructed of an X-ray generator device, an operation panel,
and so on
Image processing to reduce noise
Used to reduce the noise while maintaining the denition of an image.
Image processing for adjusting the sharpness of an image
Used to maintain the sharpness of each image depending on the exposure body part.
Image processing for compressing the dynamic range
Depending on the body part, used to provide a suitable diagnostic image even with high
contrast images.
An acronym for Intelligent-Grid.
This is a function that corrects scattered X-ray radiation.
A button that has registered exposed body parts, size, direction, etc. It is positioned on the
body part selection screen.
Indicates conditions relating to exposure in general such as exposure device, exposure
method, exposure setting, and positioning.
Indicates the tube voltage, X-ray tube current, exposure time, etc. displayed on the console
of the X-ray device.
This is a mode in which exposure is performed without being synchronized with the X-ray device.
Collective term indicating the holder used for One Shot stitching exposure. The holder can
load two or three DR Detectors.
The positions of slots containing the stitching holder DR Detectors are labeled as upper,
middle, and lower.
The triangle mark is on top.
Introduction
Upper, Middle, Lower
Upper
Middle
Lower
13
Page 14
Introduction
Structure of pages
Exampleofpagestructure
(1)
(2)
(3)
(4)
(5)
1.1 OperationFlow
1.1.1 Displayingtheexposure
screen
Perform the following procedure when to display the
exposure screen.
Turn onthe powerfor X-ray devices,etc.
1
connectedtothisdevice.
Pressthepowerswitchonthisdevice.
2
• Startup messages are displayed sequentially.
• Internal system initialization is performed while
the system startup screen is displayed.
• Upon completion of initialization, the initial
screen (list screen) is displayed.
•••••••••••••••••••••••••••••••••••••
IMPORTANT
• When sharing a REGIUS among several units of this
device, start the unit with the JM installed first; after
conrming startup of this unit, start the other units.
The operating procedure is described in sequential numerical
steps.
Shows a screen on which operations are performed.-
Describes important items for operation. Be sure to read them.
IMPORTANT
-
HINT
14
Page 15
Introduction
Screen display
Note that explanatory screens may be omitted from this manual when not needed.
Characters and icons displayed on the screen
Characters and icons displayed on the screen are shown with [ ]. Messages displayed on such as the conrmation
screen are shown with " ".
Notation example of tab: [Image Proc] tab
[Output] tab
Notation example of message: "Are you sure you want to
suspend this study?"
Notation example of button: [Cropping]
[Masking]
[Text]
Notation of operation methods
This manual describes main operation methods as follows.
NotationDescription
Press
SelectOperation to select from several options such as from a pull-down menu or tabs.
InputOperation to enter any string using a keyboard or the input panel displayed on the screen.
DragOperation to move an operation target while touching or clicking it.
Drop
Operation to touch the screen or click with a mouse to operate operation targets such as
buttons.
Operation to release the operation target being dragged by quitting touching it or releasing
the button being clicked.
15
Page 16
Introduction
Tool classification
Tool classification
For setting the display items of patient registration
up data.
Items to be set with the user tool
Items to be set with user tool are shown in the following diagram.
More information about the user tool is described in the "CS-7 User Tool Operation Manual".
[Setting]
[Backup/Restore]
selection button
display area
[System] area
[Panel] area[Panel Setting]
[Overlay] area
[Exam. Tag] area
[Image Processing] area
[Hospital Information] area
[Backup] area
Tool selection button
[Setting]
[Screen Setting]
[Patient DB]
[Trauma Patient]
[Marker]
[Stamp]
[Exposure Condition]
[RIS Code Mapping]
[Default Exam. Tag]
[Image Processing (Exam. Tag)]
[Image Processing (List)]
[Operator]
[Department]
[Location]
[Requesting Physician]
[Backup]
Function
For confirming and editing the information
of the institution and device.
screen, list screen, etc. and the layout.
For editing the patient information
registered in the database.
For setting the automatic number issuing function
and default Exam. Tag for trauma patients.
For setting DR Detector.
For setting the marker to be
overlaid on images.
For setting the stamp to be
overlaid on images.
For adding and editing the Exam. Tag settings.[Edit Exam Tag]
For adding and editing information about
the exposure settings.
For associating RIS codes sent from
the RIS with codes for this device.
For selecting an Exam. Tag that is set for a
default Exam. Tag.
For adjusting the image processing
parameter of the Exam. Tag.
For adding and editing information about
image processing.
For registering and editing the
selection of the operator names.
For registering and editing the
selection of the hospital departments.
For registering and editing the
selection of the hospital wards.
For registering and editing the selections for
requesting physician names.
For saving the settings of this device,
patient information and logs.
[Restore] area
[Auto backup] area
[Log Management]
[Reject Log]
[Restore]
[Setting]
[Maintain DB and images]
16
For managing operation logs of this device.
For managing logs of rejected images of
this device.
For restoring saved settings and patient
information.
For setting the auto backup function.
For restoring or deleting automatically backed
Page 17
Introduction
For setting the dedicated application for aggregating
Tool classification
Tool classification
[Adjustment]
[Analysis]
selection button
display area
[System] area
[Panel] area
[CS-7] area
[LongLen Unit] area
[REGIUS Σ] area
[Image Quality] area
[Statistics Analysis] area
Tool selection button
[Date and Time]
[Touch Panel]
[Calibration]
[Update]
[Release Notes]
[Adjustment]
[Information Notification Setting]
[Error History]
[Fractional Masking]
[QA/QC]
[Statistics Analysis]
[Setting]
For setting the date and time.
For adjusting the touch panel.
For conducting the calibration of the DR
Detector.
For updating this device.
For displaying the version of this device and
updated contents.
For adjusting the AeroDR Stitching Unit.
For setting the notification procedure of the
communication control information.
For displaying the error history screen.
For setting fractional masking.
For evaluating and managing image quality.
For aggregating and taking the statistics of
exposure history.
and taking the statistics of exposure history.
Function
[AeroRemote]
[Manual]
[Quality Management] area
[Test] area
[Service] area
[User Manual] area[Operation Manual]For displaying an operation manual.
[Other Manual] area
[Raw Image Output] For outputting raw images to media.
[AFSSAPS Tool]
[Drop/Impact PanelCheck]
[Error Panel]
[Quick Repair]
[Revert to before replacement]
[User Tool Operation Manual]
[PDA Operation Manual]
[Other]For displaying other manuals.
[Other 2] - [Other 5]
For outputting raw images to the host.
For performing the DR Detector test.
For displaying DR Detectors in which an
error was detected.
For replacing the DR Detector.
For reverting the DR Detector settings to
the settings before replacement.
For displaying a user tool operation manual.
For displaying a PDA operation manual.
For displaying other manuals.
17
Page 18
18
Page 19
Chapter
1
Safety Precautions & Warnings
This chapter describes precautions and warnings
to ensure safe use of this device.
19
Page 20
0197
1.1 Symbols relating to safety
1.1.1 Safety alert symbol
This is the "safety alert symbol". This symbol alerts you
to matters and/or operations potentially hazardous to
yourself and other people. Read these messages and
follow the instructions carefully.
1.1.2 Warning notice (signal
words)
Signal words indicate a degree of potential hazards
in the use of the product.
Signal words include the following three types, which
are used according to risk of damage caused by dan-
ger and the severity of damage:
DANGER
Indicates an imminently hazardous situation which, if
not avoided, will result in death or serious injury.
WARNING
Indicates a potentially hazardous situation which, if
not avoided, could result in death or serious injury.
CAUTION
Indicates a potentially hazardous situation which, if
not avoided, could result in minor or moderate injury.
It may also be used to indicate a hazardous situation
where only physical damage is likely to occur.
1.1.3 Description of graphic
symbols
Indicates the off position for the main
power switch of this device.
Indicates the on position for the main
power switch of this device.
Indicates the power on or standby position.
Indicates that the AC power is being
used.
Indicates that hazardous voltage is being
used, and caution is required.
It means conformity with the Medical
Devices Directive 93/42/EEC.
0197 is Notied Body number.
EC Directive 93/42/EEC does not
cover animal use. So, the notied body
whose identication number is 0197 is
not responsible for animal use.
20
Page 21
1.2 Safety precautions
Read all safety precautions thoroughly before using
this device.
Be sure to observe the safety precautions described in
this section.
1.2.1
Precautions before usage
WARNING
• EMC Statement - CS-7 is designed to provide reasonable protection against harmful interference in a
typical medical installation. The personal computer
of CS-7 generates, uses and can radiate radio
frequency energy and, if not installed and used in
accordance with the instructions, may cause harmful interference to other devices in the vicinity. However, there is no guarantee that interference will not
occur in a particular installation. If CS-7 does cause
harmful interference to other devices, which can be
determined by turning the equipment o and on, the
operator is encouraged to try to correct the interfer-
ence by one or more of the following measures:
– Reorient or relocate the receiving device.
–
Increase the separation between the equipment.
– Connect the equipment into an outlet on a circuit
different from that to which the other device(s)
are connected.
– Consult the manufacturer or eld service techni-
cian for help.
CAUTION
• This device is not explosion-proof, so do not use
any ammable or explosive gas near this device.
• For the basic operation of the personal computer,
display monitor, and optional parts for this device,
refer to their operation manuals.
• Please follow the rules and regulations of your relevant authorities in the disposal of this product, accessories, options, consumables, media and their
packing materials.
This symbol means: Do not dispose of this
product together with your household waste!
Please refer to the information of your local
community or contact our dealers regarding
the proper handling of end-of-life electric and
electronic equipments.
Recycling of this product will help to conserve
natural resources and prevent potential negative
consequences for the environment and human
health caused by inappropriate waste handling
.
1.2.2 Precautions for usage
WARNING
Chapter 1
CAUTION
• If the pages of the operation manual are smudged
and illegible, replace it with a new one. (There is a
fee for this service.)
• Use this device outside of the patient environment.
• This device conforms with the METI guidelines to
reduce harmonic emissions caused by electrical and
electronic equipment for household and general use.
• The operators (hospitals and clinics) hold responsibility for the usage and maintenance of this device.
Do not use this device unless you are a physician
or certied person under law.
• Before using this device, conrm that this device is
operating normally.
• When a problem occurs with this device, turn the
power off, attach an appropriate sign, such as
"malfunction", on this device, and contact Konica
Minolta technical representatives.
• Check the following additional information for images and patients and use in operation and treatment.
– The input patient information is correct.
– The input patient information has the intended
result.
– When patient information is acquired from other
devices, it is the patient in question.
– The patient being examined matches the patient
information shown on the display monitor.
– When sending the image to the output device or
media, the patient information is correct.
Before treatment, the image on the output device
–
or on the media matches the patient information.
–
The result is correct when image replacement
and supplemental information edit (image laterality/patient orientation) is performed in the exposure screen
.
21
Page 22
1.2 Safety precautions
WARNING
• Image processing functions must be used by quali-
ed persons with specialized knowledge of image
diagnosis equipment and image processing. In addition, suciently conrm the results of image processing.
• Do not perform diagnosis with the images displayed
on this device.
• Risk of misidentification of patients may occur if
modifications to patient information for examinations or examination information acquired from a
higher system such as a Radiology Information
System (RIS) are made. When the information is
modified, notify the system administrator of the
modification without delay, and in addition, make
sure that the patient information and/or examination
information are correctly modied on output devices
such as PACS to which this device sent the image
information and examination information.
• Konica Minolta will not be liable for the loss of important data such as patient information arising
from the introduction of this system. Important data
require separate assurance of security, backup, etc.
• When lightning is likely to occur, stop work and
unplug the power plug from the wall outlet. When it
thunders before unplugging the power plug do not
touch the device, including any of the cables. Doing
so may cause a re or electric shock.
• If there is any smoke, odor, or abnormal sound, it
may cause a re if use is continued, so immediately
turn the power switch off, unplug the power plug
from the wall outlet, and contact Konica Minolta
technical representatives.
• Take the following precautions when using this device and software.
– Konica Minolta or a contractor designated by
Konica Minolta will install the software of this de-
vice.
– Do not disassemble or modify this device.
– Do not connect any devices that were not pur-
chased from Konica Minolta.
– Do not install software purchased from other
companies.
– Do not uninstall software that has already been
installed.
– Take precautions to prevent infection from com-
puter viruses.
– Do not change the setting for the OS (operating
system), or delete folders or les.
– Do not operate Windows Explorer unless speci-
ed.
– Do not use the DVD drive for other purposes.
– Confirm that the hard disk access lamp is out
before removing disks.
– Do not turn the power switch o or pull out the
power cable while the system is operating.
WARNING
• Take note of the following to reduce the risk of re,
electric shock, or electrical leakage:
– Always use the specied cables such as a pow-
er cable.
– Use wall outlet with the correct rating as a power
source.
– Connect the power plug to the wall outlet prop-
erly without any slack.
– Use a power source with a ground wire.
– If you do not plan to use this device for an ex-
tended period of time, unplug the power plug.
– The supplied power cable is dedicated for this
device, so do not use it elsewhere.
– Avoid exposure to liquids such as water.
– Make sure that foreign material, such as pieces
of metal or wires, does not get inside.
– Do not handle the power plug with wet hands.
– Do not let dirt or dust adhere to the power plug.
– Do not use extension cables.
– Do not place many loads on a single electrical
outlet.
– Do not damage the power cable. Also, do not
use damaged cables.
– Do not block the ventilating openings.
22
Page 23
1.2 Safety precautions
CAUTION
• Take the following precautions when performing ex-
posure using this device.
– Check whether or not the patient needs extra
care before performing exposure.
– Check the patient information before performing
exposure.
• Turn the power o after nishing the work to main-
tain stable operation. If the power to the system is
kept on continuously for 24 hours out of necessity,
turn the power o and on once a week.
• To use functions that share image data, integrate the
pending reason and other settings for this device.
• Use with suitable settings for image processing parameters. To use functions that share image data,
set the same values for the image processing parameters of this device.
• Do not use devices that emit electromagnetic
waves such as mobile phones, PHS, or pocket
pagers near this device.
• Do not use continuously for extended periods of
time. If you need to use the device continuously for
extended periods, take a 10 to 15 minute break every hour to give your eyes and hands a break.
• Take note of the reception status for radios and
TVs near this device, since interference may occur
in them when this device is in use.
• Check the touch positions on the touch panel when
starting this device, turning on the display monitor,
or unplugging and plugging in the USB cable, etc.
When the touch position and the cursor position
are out of sync, or the cursor does not move even
when touched, leave it for about two minutes, or
check the USB cable connections and restart this
device and the display monitor.
• Be sure to check the image orientation of images
received from medical imaging devices (CR, DR,
etc.).
• Shut down this device after confirming that the
hard disk access lamp is out. Also, when turning on
again, do so after 10 seconds have elapsed.
1.2.3
Precautions for installing,
moving, and storing
WARNING
• Take the following precautions when moving this
device.
– Do not subject to shock or vibration.
– Do not start moving until the power is turned o,
and operation has stopped completely.
– Do not move with the power cable or any other
cable connected.
CAUTION
• Because connections of the X-ray device can only
be made by Konica Minolta or its designated contractors, contact Konica Minolta or its designated
contractors.
• Contact Konica Minolta or dealers specified by
Konica Minolta to install or move this device.
• Take note of the following when installing or storing
this device.
– Do not install or store in a location where it may
be adversely aected by atmospheric pressure,
temperature, humidity, ventilation, sunlight, dust,
salt-air, or air containing sulfur.
– Do not install or store in a location where it is not
stable, ventilation is insufficient, the difference
in light-dark is great, electromagnetic waves are
generated, or where subjected to vibration or
shock.
– Do not install or store in a location where chemi-
cal agents are used or stored.
– Do not lay this device on its back, upside down
or directly on the oor.
– Do not connect or disconnect the interface cable
while the power cable is connected.
– Do not use an interface cable other than the
specied one.
– Do not damage the interface cable. Also, do not
use damaged interface cables.
– Do not leave packing materials around.
– Do not subject this device to strong shocks or
vibration by dropping it, etc.
Chapter 1
23
Page 24
1.2 Safety precautions
1.2.4
Precautions regarding care
and maintenance of this device
WARNING
• Do not remove the lithium batteries.
• Turn the power o and remove the power plug from
the wall outlet.
CAUTION
• Perform the maintenance and inspection periodically. Also, contact Konica Minolta technical representatives for cleaning of the inside.
• Immediately after turning the power o, be careful
of hot components inside this device.
• Install the power cable, interface cable, etc.
securely.
• Do not rub or strike this device with hard objects as
it will cause damage.
1.2.5 Precautions regarding
CD/DVD use
WARNING
• The optical head of the CD/DVD device discharges
a laser, so take care not to look in the object lens of
the optical head.
CAUTION
• Take the following precautions to prevent reading
errors etc.
– Do not subject to impact.
– Do not touch disk surfaces.
– Place in cases when storing.
– Do not place in locations with a lot of dust.
– Do not place in locations with high temperature
and high humidity.
– Do not place in direct sunlight.
• Attach labels to CD/DVD surfaces and cases so
that the recorded content can be identied.
• The types of media supported by this device are
CD-R and DVD-R only. Also, please note that it is
possible that when saving on media other than the
following, writing errors or reading errors may occur. Please be forewarned.
• If dust adheres to the object lens of the optical
head in the CD/DVD drive, it is possible that writing errors and reading errors may occur. In such
case, remove the dust with a commercial DVD lens
cleaner.
24
Page 25
1.2 Safety precautions
1.2.6 Precautions regarding
display monitor use
WARNING
• If the LCD monitor shows the same image for ex-
tended periods of time, it will burn onto the screen.
CAUTION
• Use with screen brightness and contrast at a suitable level. Also, do not turn brightness up too high
when using in a dark room.
• The backlight for the LCD monitor will deteriorate
over time and will gradually grow darker. Check that
the display monitor is at a suitable level of brightness before using this device.
• View more than 80 cm from the display monitor.
• Use in a bright room is recommended due to eye
fatigue being cumulative when used in a dark room.
• When small dots of light and dark appear on the image in the LCD monitor, check whether it is a defect
in the LCD monitor or the image information before
using.
• The 23-inch multi-touch monitor (RadiForce
MS235WT) cannot be operated while the user is
wearing gloves.
• If you use Windows 10, do not swipe the touch
screen from the edge to the center. Otherwise,
Windows settings screen will appear, and forcible
shutdown of this system or unintentional altering of
its settings may occur.
1.2.7 Notice relating to
information security
WARNING
• Systems infected by computer viruses, worms and
other malicious software potentially harmful to computers (hereunder referred to as "malware") may
result in the erasure, corruption or unwanted disclosure of data stored within this device, including personal and sensitive patient information. Infections
by malware may also result in malfunction of the
device, as well as causing the infection to spread to
other products.
• This device may only be used in combination with
software conrmed to be safe and compatible, and
installed according to predetermined installation
procedures.
• The introduction of the latest security updates to
the Operating System, as well as providing for anti-
virus software and updating virus denition les on
a regular basis, are considered eective in prevent-
ing the infection of the system by malware. Please
refer to a Konica Minolta service representative for
details.
• To prevent the infection of the device by malware,
notice the following things:
– Do not connect the device to equipment not pur-
chased from Konica Minolta, or equipment not
certied for connection to the device.
– Do not install software purchased from other
companies.
– Do not uninstall software that has already been
installed.
– Do not change the setting for the OS (operating
system) or files and application software associated with the OS, or delete folders or files
thereof.
– Always confirm external storage media to be
free of malware prior to use.
– Use the device in a safe network environment
incorporating adequate security measures for
the prevention of malware infection.
– Refrain from any hazardous activity leading to
potential vulnerability against malware infection.
• Take precautions not to allow others to see the device screen displaying sensitive patient information.
Privacy protection display lters should not be used
as such lters may adversely aect the appearance
of displayed images.
• Take precautions not to allow others to operate the
device.
Chapter 1
25
Page 26
1.2 Safety precautions
WARNING
• Be sure to implement the following cyber-security
measures:
– Provide for an adequate level of information se-
curity measures within the facility using this device, and other facilities connected to the facility
via networks.
– Isolate the device from the connected network
immediately upon infection by malware, concurrently prohibiting the use of the infected device.
– Upon infection of the device by malware, or
the potential infection thereof, promptly contact
Konica Minolta service representative.
• Manage the account information (user name, ID,
etc.), password and other sensitive information as
follows:
– Use a combination of upper- and lower-case al-
phabet and numeric characters to prevent easy
identication by malicious parties.
– Do not leave records of sensitive information to
avoid uncertied disclosure.
– Passwords should consist of at least eight char-
acters.
– Passwords should be assigned with expiration
dates, and changed on a regular basis.
1.2.8 Precautions for each
function
WARNING
Precautions on stitching combination processing
• Adjust the combination position by using the markers as supplementary standards in addition to body
parts of the subject. It is also recommended that
you arrange the measure in the combined image
part during exposure to simplify the conrmation of
the combined parts.
Precautions for One Shot stitching exposure
• One Shot stitching exposure outputs the combined
picture by correcting the images' overlapping part
captured by DR panel. Depending on the capturing
conditions, lateral stripes or decrease in contrast
might occur in a correction area, or decrease in
contrast might also occur if a structure horizontally
long overlaps with the correction area. Make sure
the correction area and the parts in an examination
are not overlapping.
• Make sure to observe the number of DR panels
that are loaded in the stitching holder through the
exposure screen before conducting the exposure.
Precautions for image measurements
• Sometimes the exposure orientation causes accidental errors with regard to the actual length of
the subject. Use while taking into account any error
between image and measured results.
Precautions when running HS processing
• The noise is reduced by HS processing, and you can
expect signals that are unclear due to noise to become
clearer, but signals that do not exist will not be visible.
• When reviewing whether to reduce dosage by means
of HS processing, consider lowering the signal component by means of reducing dosage and always
check whether images are suitable for diagnosis
.
26
Page 27
1.2 Safety precautions
WARNING
Precautions when applying scattered radiation
correction (IG) processing
• It is expected that scattered radiation correction pro-
cessing will remove scattered radiation components
thus enhance subject contrast; however, it does
and
not ensure that the same image quality as that which
is generated when a grid is used will be obtained.
will not have any eect if the desired image is not
contained in the original signals.
• Applying the scattered radiation correction processing
may result in unnatural images. In particular, applying
this processing to images for which grids were used
may increase the contrast too sharply. Make sure to
perform image verication to conrm if the application
of this processing is suitable for diagnosis.
• Preset a tube voltage value that is close to the ac-
tual exposure conditions as the image processing
parameter of the Exam. Tag.
Precautions on magnetic card reader
• Check that patient information read from a magnetic card is consistent with the information written
in the magnetic card.
• Do not connect multiple magnetic card readers to a
single device. It may cause malfunctions.
• Do not change the settings of the dip switch on the
bottom of the magnetic card reader.
Precautions on bar code reader
• The reading window discharges a laser, so take
care not to look in the laser or point the laser toward the eye. In addition, perform periodic inspection of the reading window.
• Laser is discharged during disassembly. Do not disassemble.
Precautions on blackening process for areas
outside the exposure eld
• In the blackening process for areas outside the exposure eld, if the distance between the edge of the
X-ray
exposure eld and the edge of the cassette is
3 cm or less, the masking function may fail to recog-
nize the masking eld. When using a collimator, set
the
edge of the X-ray exposure eld to be sucient-
ly distant from the edge of the cassette
use preview on the [Output] tab or [Overlay] tab to
check that the masking area is appropriately set
before the image is output, and modify the masking
area on the viewer screen if needed.
. In addition,
CAUTION
Notes on the chest wall blackening process
• Recognition of unexposed areas on the breast
may be lost in the chest wall blackening process.
Use preview on the [Output] tab or [Overlay] tab to
check that the chest wall blackening area is not in
the diagnosis area before the image is output, and
set the chest wall blackening process to OFF on
the viewer screen if needed.
Precautions on mammogram
• For United States of America, this device is not intended to use for Mammography.
• Use upon conrming that missing regions in images
on the reception area and exposure eld exceeding
the reception area are within the allowable range.
Also, make sure that additional information such as
stamps do not overlap with the breast area in images.
• Confirm that additional information such as markers and stamps is not superimposed on the breast
area.
• When displaying two mammograms on two frames,
ne-tune and check the horizontal positions of the
images.
Precautions for the use of the highlight func-
tion for the tube/gauze
• If the tube/gauze is dicult to identify on the image,
adjust the highlight level. When the tube/gauze is
highlighted too much, lower the highlight level.
Precautions when merging study images
• When studies are merged, their patient and study
information are updated.
Precautions on the lung lack/motion blur detection function
• In images showing lesions, foreign substances,
etc., lung lack or motion blur may be detected even
when none exist.
Precautions on multi-study function
• Be careful since, in order for patient information to
be determined automatically from the four information variables (patient ID, name, date of birth, and
sex), a patient will not be treated as the same patient if the four variables dier.
Chapter 1
27
Page 28
1.2 Safety precautions
CAUTION
Precautions for Web/Remote Desktop Linkup
• Do not use the Web/Remote Desktop Linkup for the
following purposes:
– Diagnosis of illnesses, prevention of illness,
monitoring, treatment, or pain relief.
– Diagnosis of injuries or physical disabilities,
monitoring, treatment, pain relief, or corrective
procedures.
– Physiological or anatomical examinations, sub-
stitutions or corrections.
– Control of conception.
– Diagnosis, treatment, alleviation, or prevention
of human or animal illness, abnormal physical
conditions, or diagnostic signs of such illnesses
or conditions.
– Healing, modication, or adjustment of human or
animal physical functions or anatomy.
– Diagnosis of pregnancy.
– Care for humans or animals during pregnancy,
while delivering, or after birth, including for natal
care and non-medicinal birth control.
Precautions on the login control mode
• Be careful of password management.
Precautions on REGIUS plates
• Frequency and duration of use of the REGIUS
Precautions on high precision monitors
• Since a high precision monitor has a different as-
• High precision mammographies are not displayed.
Precautions on the AeroDR Stitching Unit
• Always ensure the exposure ready LED (blue) of
*
plates
are automatically monitored when they are
scanned by the reader. The replacement schedule
may not be properly managed if the plate is not
used for an extended period of time. Therefore, use
of the REGIUS plates as evenly as possible is recommended. (*
Konica Minolta computed radiography)
pect ratio from that of the exposure and viewer
screens, when you enlarge, reduce, or move an
image on the exposure or viewer screens, an image of a dierent area may be displayed on such a
monitor.
the AeroDR Stitching Unit is lit before performing an
exposure. Additionally, do not press the rst stage
of the exposure switch until the exposure ready
LED (blue) is lit.
Photostimulable phosphor plate for
CAUTION
Precautions on a stitching cassette
• There are small dierences in magnication of im-
ages made with REGIUS plates which are adjacent
within a stitching cassette. For this reason, overlapping parts of images made with REGIUS plates
which are adjacent do not completely match. (When
the exposure distance is 1.5 m, misalignment in the
horizontal direction will be approximately 0.5 mm at
maximum.)
Precautions on the grid
• For grid during exposure, use 34-40 lp/cm. When
exposing with a grid other than specified, some-
times moire cannot be eliminated.
* For the DR Detector, the grid to be used is speci-
ed. Refer to the AeroDR Operation Manual for
details.
• The grid removal process reduces the response of
the frequency components that make up the grid
image. The principle of this process is that the grid
image uses a specic range of spatial frequencies.
Therefore, part of the signal components of images
other than the grid image may also be reduced if
these images have signal components near the
frequency range of the grid image. When you have
performed exposure using a grid, check whether
the images obtained by removing the grid are suitable for diagnosis.
Precautions when changing the exposure device
• When changing the device (for example, a wall
stand reader to a table reader), dierent exposure
conditions may have been set for the same body
part depending on the settings of the body part that
is created or edited by the user master. Therefore,
carefully check both the body part and its exposure
conditions before starting the examination.
Precautions when the image is used on a device other than the one used for output
• When using the image on another device for the
purpose of measurement etc., check the consistency between the devices before use. In particular,
care must be taken when decreasing the image
size because some correction may be applied to
the image depending on the reduction ratio.
Precautions on changing the reading area
• If annotation has been set for the image, changing
the reading area will disable the annotation setting.
Set annotation again.
28
Page 29
CAUTION
Precautions when sending exposure setting
• When sending exposure setting to an external Xray device, check the exposure conditions on the Xray device. Be especially careful of X-ray exposure
settings when performing an exposure on a small
child.
Precautions when delivering an image from
this device to other devices
• Conrm that additional information such as set annotation is correct.
Precautions for exposure with the DR Detector
• Before exposure, conrm on the screen of this device that the DR Detector is registered in the exposure room. Also, perform exposure after conrming
that the DR Detector to be used for exposure has
been selected.
• Do not leave the DR Detector in the vicinity of a
heating device such as a heated carpet etc.
1.2 Safety precautions
Chapter 1
Precaution on exposure dose
• After exposure, S values are displayed in the image display areas of this device. Check that the S
values indicate the appropriate dosage for exposed
images.
Precautions when performing exposure in Aero
Sync mode
• Confirm that the device is ready for exposure
through its display before performing exposure.
• Conrm that a conrmation is sounded from the device after the start of exposure.
• Do not let the DR Detector vibrate or receive shock
until the preview image is displayed.
• Conrm that the DR Detector’s battery level is full
before performing exposure.
• Use within the range of the pre-checked exposure
conditions.
• Be aware that the exposure procedure must be
changed in some cases when the settings are
changed for the Aero Sync button or initial detection
level. Be sure to press the Aero Sync button after
changing the Aero Sync button setting from OFF to
ON.
• If the Aero Sync button or initial detection level has
not been set appropriately, a malfunction may occur
due to the DR Detector being subjected to impact.
Make sure values are changed appropriately.
29
Page 30
30
Page 31
Chapter
2
Product Overview
This chapter describes the overview of this device.
31
Page 32
2.1 Overview of this device
(HOST)
REGIUS
This section describes the functions and system conguration example.
2.1.1 Functions
This device is a DIRECT DIGITIZER used with an image diagnosis device, medical imaging device and image storage device c
This device digitally processes patient images collected by the medical imaging device to provide image and patient
information.
2.1.2 System conguration example
The system conguration example is as shown below.
onnected via the network.
Electronic Medical
Records
X-Ray device operation panel
X-Ray Exposure
information
FTP
DICOM Print
HIS/RIS
DICOM G/W
DICOM MWM
DICOM MPPS
DICOM
DR
Detector
CS-7
(1) Order Information
(2) Image Information
(3) Patient Information
Image Data
DICOM Storage
REGIUS
SIGMA/SIGMA2
JPEG
DICOM PDI
Printer
PACS
Media
32
Page 33
2.1 Overview of this device
HINT
• This device is intended to be connected to DR Detector or to REGIUS, REGIUS SIGMA/SIGMA2 as an input device, and to
printer or PACS as an output device.
• For details on how to use DR Detector and REGIUS, REGIUS SIGMA/SIGMA2, refer to their respective operation manuals.
• 7 DR Detectors can be connected to this device.
• 15 REGIUS series can be connected to this device.
• Only a single REGIUS SIGMA/SIGMA2 unit can be connected to this device.
• A REGIUS series unit and a REGIUS SIGMA/SIGMA2 unit cannot be connected to this device at the same time.
This chapter describes the screen structure of this device,
and the function of each button on the screen, etc.
35
Page 36
3.1 Screen connections
The following describes the connection between various screens and how they transition. For details, see the corresponding page in the manual for each screen.
Meaning of the arrows
:Screen transition operation
:
zInitial screen
Patient registration screen
List screen
Operation that returns to
the previous screen
zSystem monitor screen
The initial screen to be displayed will be either the patient
registration screen or the list screen. The patient registration screen and the list screen are switched by selecting
tabs.
Press [Perform].
Press [Back].
zBody part selection screen
Press
[System].
This is the screen to select an
exposure body part.
Press [System].
Press [Close].
Press [Close].
System Monitor
Warranty Component
This screen is used to confirm
the status of this device and devices connected to this device,
as well as the time to replace the
regular replacement parts. Each
screen is switched using tabs.
Press [Perform].
zExposure screen
This is the screen for exposing
and adjusting the exposed im-
age. By switching the control
panel, various operations can be
performed.
Press [ ]/[ ].
Press [
[
] or
].
Press [Tool].
36
Press [Return].
zViewer screen
This is the screen for ne adjustment
of the exposed image. More detailed
image adjustment can be performed
than the exposure screen.
Page 37
3.2 Patient registration screen
3.2.1 Structure of patient registration screen
This is the screen for inputting the patient information. The patient registration screen is displayed when the
[PatientReg.] tab is selected from the main tabs.
(1)
(2)
(6)
(3)
(7)
(8)
(4)
(9)
Chapter 3
(5)
NumberNameFunctions
Date and time, operator name, exposure room name, icons of devices
(1) Application bar
(2) Sorting tab
(3) Input item area
Display/hide button
(4)
(Input panel)
(5)Tool bar area
connected to this device are displayed.
When the [ ] icon is displayed in the leftmost position of the application
bar, pressing the date and time display area hides the screen.
There are 2 types of tab, the main tab and the sub tab.
[PatientReg.] tab, [OrderList] tab and [Completed] tab are displayed in the
main tab. In setting, one more sub tab can be added.
When all tabs cannot be displayed within the screen, tab scroll buttons
[ ] are displayed.
This is the area to input patient information. Some items require input with
the input panel (keyboard).
Displays or hides the input panel display.
[System]: The system monitor screen is displayed.
[Web/Remote Desktop Linkup]:
The screen of the HOST or RIS linked to the Web/Remote Desktop is
displayed.
[Perform]: Runs the examination of the selected patient.
[Mammo Print ON] and [Mammo Print OFF]:
This either enables mammogram printing (normal print mode) or enables the print composer.
Setting this to [ON] enables mammogram printing of the exposure
screen (normal print mode). Setting this to [OFF] enables the print
composer. The device switches between [ON] and [OFF] every time the
button is pressed.
37
Page 38
3.2 Patient registration screen
NumberNameFunctions
(6)Scroll button
(7)Clear buttonThe input patient information disappears.
(8)Search buttonSearch begins with the input patient information.
(9)trauma button
Pressing [
displays the previous patient information.
Automatically issues a number in emergency case in order to start the
examination.
This is displayed when the automatic emergency number issuing function
is turned on with the user tool.
] displays the next patient information, and pressing [ ]
IMPORTANT
• Creating too many tabs makes it dicult to nd the target tab and lowers the operability. Do not add more tabs than neces-
The date and time, operator name, exposure room name, icons of devices connected to this device, and so on are
displayed in the application bar.
Date and time/operator name
display area
Device information
display area
Date and time/operator name display area
The current date and time is displayed in the date and time display area, and the operator name is displayed in the
operator name display area.
When the
[ ]
icon is displayed on the left side of the date and time display area, press the date and time display
area to hide it.
Date and time display areaOperator name display area
Device information display area
Chapter 3
The device information display area of the application bar displays the device icons as well as the states (battery
level and signal strength) of the DR Detectors registered in this device.
Signal strength
Battery level
Device icon
When an image display area has focus, a frame is displayed around the device information that corresponds to the
image display area.
Frame indicating focus
zDevice icon display
IconDevice name
Wall stand reader
Table reader
DR Detector
zDR Detector battery level display
Status
Not being
charged
Charging
Color
of icon
RedYellowGreen
Battery level
HighLow
AeroDR Stitching Unit
39
Page 40
3.2 Patient registration screen
DR Detector signal strength display
z
Out of range
Wireless condition
GoodBad
Wired
connection
IMPORTANT
• When the signal strength antenna indicator displays one or fewer bars, cycle time will take longer or wireless communication
may not be possible. It is recommended that this function is used when the signal strength antenna indicator displays two or
Device information display area when using the portable X-ray device
When this device is operated with the portable X-ray device, the network connection status, etc. are displayed on the
application bar.
Signal strength
Network name
Battery level
Device icon
zBattery level display of this device
(when using the portable X-ray device)
Status
Not being
charged
Charging
Color
of icon
RedYellowGreen
IMPORTANT
• When the signal strength antenna indicator displays one or fewer bars, cycle time will take longer or wireless communication
may not be possible. It is recommended that this function is used when the signal strength antenna indicator displays two or
more bars.
• If the battery charge drops to 5% or less, the system may enter a sleep state. If the battery icon turns red or the message
that the battery level is low appears, please charge this device. If this device cannot be charged right away, shut down this
device.
(13)Examination information display areaThe details of the examination are displayed.
(14)Scroll bar/scroll buttons
(15)Tool bar area
Displays or hides the examination information display area.
The tab is displayed when the examination information display area is
displayed.
Drag the scroll bar to scroll a list left and right. Pressing [
scrolls the screen by frame, and pressing [ ] scrolls by page.
The operation button is displayed for a focused examination.
[System]: The system monitor screen is displayed.
[Web/Remote Desktop Linkup]:
The screen of the HOST or RIS linked to the Web/Remote Desktop is
displayed.
[Delete]: Deletes the selected examination.
[Modify]:
Displays the information modication screen for the selected examina-
tion.
[Refer]: Displays the examination history of the selected examination.
[Append]: Adds the Exam. Tags to the completed examinations.
[Refresh]: When connected to a higher system, refreshes lists.
[Guard]:
Disables deletion of the selected examination. Press again and protec-
tion is canceled.
[Merge]: Combines the selected studies and moves images.
[Perform]: Performs the selected examination.
[Mammo Print ON] and [Mammo Print OFF]:
This either enables mammogram printing (normal print mode) or enables the print composer.
Setting this to [ON] enables mammogram printing of the exposure
screen (normal print mode). Setting this to [OFF] enables the print
composer. The device switches between [ON] and [OFF] every time
the button is pressed.
(14)
(15)
]
HINT
• The [Web/Remote Desktop Linkup] icon displays the registered RIS, HOST, or other name.
(6)Exam. Tag display areaExam. Tag for the selected body part is displayed.
(7)Exam. TagThis is the button to determine exposure conditions for the body part.
Switching button
(8)
(Exam. Tag display area)
Date and time, operator name, exposure room name, icons of devices
connected to this device are displayed.
Patient name, patient ID, date of birth, sex, etc. are displayed.
The emergency mark is displayed for emergency patients, and the medical warning mark is displayed for pregnant patients or other patients
who need extra care during X-ray exposure.
When the set tab is selected, body part is displayed in the body part
display area linked to the selected tab.
The exposure body part is displayed. With a maximum of 2 pages, each
page can display up to 8 buttons at the same time.
Switches the pages of the body part display area. Maximum is 2 pages.
When there are multiple pages for the Exam. Tag display area, move
pages by pressing [
] or [ ].
(7)
Chapter 3
(8)
43
Page 44
3.4 Body part selection screen
(9)
(10)
(11)
(12)
(13)
(14)
(15)
NumberNameFunctions
(9)Exam. Tags display areaPress the Exam. Tag to display the added Exam. Tags.
Exam. Tags and device icon are displayed.
(10)Exam. Tags
(11)Delete buttonDeletes the selected Exam. Tags (focused Exam. Tags).
(12)Scroll bar/scroll buttons
(13)Append and move buttonsMoves the focus to the added Exam. Tag.
(14)Tool bar area
(15)Tool bar area
HINT
• When exposed, the exposed image thumbnails and Exam. Tags are displayed in the Exam. Tags display area.
• [Schedule] is not displayed in the tool bar area when the body part selection screen is displayed from the exposure screen.
• When moving the focus to the added Exam. Tag, press the move-to-add button, [ ].
: Displayed when the Exam. Tags are changed on the exposure
screen.
: Displayed when the Exam. Tags are added on the exposure screen.
Drag the scroll bar to scroll the Exam. Tags display area left and right.
Pressing [
pressing [ ] scrolls pages.
During registration in the Exam. Tags display area, operation buttons are
displayed for the focused examination.
[System]: The system monitor screen is displayed.
[Schedule]:
Reserves the selected Exam. Tags (focused Exam. Tags). Pressing
[Schedule] reserves the examination and displays the list screen.
[Back]: Displays the list screen without reserving the examination.
[Perform]: The exposure screen is displayed.
Displayed when the exposure screen transitions to the Body Part Selec-
tion screen.
[OK]: Displays the exposure screen.
[Cancel]: Returns the display to the previous screen.
] scrolls the exposure body part list by frame, and
The exposure screen consists of several areas. Here, each area is described.
(1)
(2)
(3)
(4)
NumberNameFunctions
(1) Application bar
(2) Patient information display area
(3) Study display area
Currently selected order No./Number of
(4)
orders
(5)Work area
(6)Control panel
(7)Tool barDisplays various functions, and a shortcut to the control panel, etc.
Date and time, operator name, exposure room name, icons of devices
connected to this device are displayed.
Patient name, patient ID, date of birth, sex, etc. are displayed.
The emergency mark is displayed for emergency patients, and the medical warning mark is displayed for pregnant patients or other patients who
need extra care during X-ray exposure.
]:
[
The patient information editing screen is displayed ([
played when exposing with the multi-study).
The order list or body part selection is displayed. The displayed details
can be switched using tabs. Pressing [ ] can move the display on
an order by order basis. Pressing [ ] moves the display to the starting
order while pressing [ ] displays the last order.
Displays the currently specied order of studies and the total number of
studies.
Displays exposure conditions and exposed images, etc. The displayed
details can be switched with tabs.
Used when performing some operation on the image display area that
has focus. The operation panel switches automatically depending on
the content in the image display area that has focus, and it can also be
switched by selecting tabs.
] is not dis-
(5)
Chapter 3
(6)
(7)
45
Page 46
3.5 Exposure screen
HINT
HINT
(8)(10)
(9)
NumberNameFunctions
[Rotate Left]: Turns the image to the left.
• The image is rotated by 90 degree to the left from the 360 degree
position.
[Rotate Right]: Turns the image to the right.
(11)
(12)
(8)Function button
(9)Status displayDisplays the current rotation status by the angle value.
(10)Extended area display button
(11)Exam. Tag button
(12)Viewer buttonThe viewer screen is displayed.
• The image is rotated by 90 degree to the right from the 0 degree
position.
[Flip]:
Flips the image. When the image is ipped, the icon is in the selected
state.
[< Prev.]: Displays the previous image.
[Next >]: Displays the next image.
[Text]: The viewer screen is displayed. You can add an annotation.
[ROI Adjust]:
The viewer screen is displayed. You can change the ROI area.
Function buttons more than the maximum number of function buttons that can
be placed are stored. The stored function buttons pop up when the extended
area display button is pressed.
The following function buttons are set by default.
[S/G Drag]:
Switches to the operation mode used to adjust the S/G value. Pressing
it again cancels the operation mode for adjustment. You can adjust the
S and G values by dragging the image while the operation mode for
adjustment is turned on.
[Invert]: Inverts the black and white colors of the image.
Changes or adds the Exam.Tag. The Exam. Tag buttons can be switched by
pressing and holding this button.
: Changes the Exam. Tag.
: Adds the Exam. Tag.
HINT
•
The patient information display area can be small-sized by setting. Emergency and medical warning marks are not displayed in
the small-sized area.
•
The Reject Image button may be displayed on the tool bar depending on the setting.
• The tool bar may also be displayed vertically on the right of the work area.
• Both of the Exam.Tag buttons may be displayed simultaneously depending on the setting.
• For the [Rotate Left] and [Rotate Right] buttons, you can congure the setting to switch whether to display both the buttons
or display either one. If either one is displayed, hold down the button to switch to the other button.
• The types and locations of function buttons can be congured using the user tool.
In the work area, examination content (Exam. Tags and exposed images) and previews when sent to the host are
displayed. Display of the examination content and the previews is switched by selecting tabs.
(4)
(1)
(2)
(3)
NumberNameFunctions
Study/Queue Cont. tabSwitches the display between the study content and the output preview.
(1)
Displays exposed images after exposure, or exposure setting before
(2) Image display area
(3) Exam. Tags display area
(4)Full-Screen buttonDisplays captured images on the full screen.
(5)Screen display format switching buttonChanges screen display format.
(6)Output button
(7)Reject button
exposure.
When the [Queue Cont.] tab is selected, previews at the time of output
are displayed.
Displays ip mark, output format, and Exam. Tags, etc. The meaning of
the icons are listed below.
: Not ipped
: Flipped
: A format
: AB format
Outputs exposed images to the preset output destination. After output,
the icon changes to [ ].
Sets an exposed image as rejected. After setting the image rejected, the
icon changes to [ ].
(5)
Chapter 3
(6)
(7)
49
Page 50
3.5 Exposure screen
HINT
•
In the output preview, images are displayed in the order of transmission.
•
In the output preview, rejected images are not displayed.
•
The screen display format has a single-image format and a multiple-image format to display multiple screens in alignment. Sin-
gle-image format is suitable for checking details of the image. Multiple-image format is suitable for comparing multiple images.
For work areas when unexposed, the type of exposure device, status, irradiation information, exposure conditions,
etc. are displayed.
(1)
(2)
(3)
(4)
(5)
Chapter 3
NumberNameFunctions
(1) Status
(2) Device iconThe exposure device icon and the size of the DR Detector is displayed.
(3) Exposure informationThe exposure information during the exposure is displayed.
(4)Exposure condition iconExposure condition is displayed as an icon.
(5)Reading conditions
IMPORTANT
• When using AeroDR 3 1417HD and a CR casette, be sure to check the resolution displayed in the reading conditions when
On the control panel, adjustment items for exposure
and a screen to adjust exposed images are displayed.
Control panels basically switch automatically according to the operation objective and circumstances, but
sometimes you need to switch them manually. The
types of control panels are as follows.
Exp.Info: Expo-
sure conditions
Image Proc: Im-
age processing Acquisition
Output
(For host)
Overlay
(For printer)
Output
(For printer)
Overlay
(For host)
Chapter 3
AeroSync
QUICK
Exp.Info: Irra-
diation results
53
Page 54
3.5 Exposure screen
HINT
•
Up to three tabs are displayed on the control panels. If a
desired tab is not displayed, press [ ] displayed
on the tab and scroll it until the desired one is displayed.
•
Hold down
or longer to display the drop-down list of tabs. If you select
a tab name from the list, the selected tab is displayed.
3.5.7 Control panel (Image Proc: Image processing)
This is the control panel to adjust images after exposure. Basic image adjustments such as change of S/G value, etc.
can be made.
(1)
(2)
(3)
(4)
(5)
(6)
(7)
(8)
(9)
NumberNameFunctions
Adjusts the S value.
Pressing [ ] increases or decreases the S value step by step.
(1) S Value slider/button
(2) G Value slider/button
(3) ROI/ROI Adjustment buttons
(4)ZOOM+/ZOOM-/FIT/Detail buttons
Pressing [ ] increases or decreases the S value in multiples of the
preset value.
The S value can also be adjusted by dragging the slider.
Adjusts the G value.
Pressing [ ] increases or decreases the G value step by step.
Pressing [ ] increases or decreases the G value in multiples of the
preset value.
The G value can also be adjusted by dragging the slider.
Press [ROI] and ROI is displayed. Press again and it is released. Press
[ROI Adjustment] and the viewer screen is displayed.
Changes the display method of the image that has focus.
[ZOOM+]:
Switches to the zoom-in mode. Pressing it again cancels the zoom-in
mode.
Pressing the image in the zoom-in mode enlarges the display 1 step at
a time.
[ZOOM-]:
Switches to the zoom-out mode. Pressing it again cancels the zoomout mode.
Pressing the image in the zoom-out mode reduces the display 1 step at
a time.
[FIT]:
Displays the enlarged (or reduced) image to t in the frame size (size
of the image display area).
[Detail]:
The viewer screen is displayed and the exposed image is displayed
zoomed in.
(11)(10)
56
Page 57
3.5 Exposure screen
NumberNameFunctions
They change screen display format.
[Pan]:
Switches to the pan mode. Press it again to cancel the pan mode. You
can move a desired image by dragging it while the pan mode is turned
on.
(5)Pan/S/G/Rotate&Flip buttons
(6)E process/F process/Hybrid/Tool buttons
(7)Text buttonThe viewer screen is displayed. Place the annotation.
(8)Cropping button
(9)Masking buttonThe viewer screen is displayed, so set the masking area manually.
(10)Preview button
(11)Reset buttonReturns to the state when image was displayed.
[S/G]:
Switches to the operation mode used to adjust the S/G value. Pressing
it again cancels the operation mode for adjustment. You can adjust the
S and G values by dragging the image while the operation mode for
adjustment is turned on.
[Rotate&Flip]:
Press this button to display the screen to rotate and ip an image.
Performs the process when pressed.
The viewer screen is displayed, so set the eective image area manually.
Displays a preview image in which the set output conditions are reected.
Chapter 3
57
Page 58
3.5 Exposure screen
3.5.8 Control panel (Acquisition)
This is the control panel to perform setting of automatic read and reading conditions.
(1)
(2)
(3)
(4)
(5)
(6)
NumberNameFunctions
Change the automatic read settings of the DR Detector.
[Size/Pos.]:
The reading size, reading direction, and reading position are automatically set.
[Position]:
(1) Automatic Reading button
(2) Reading Size buttonSpecify the reading size.
(3) Reading Direction buttonSpecify the reading orientation (portrait/landscape).
(4)Reading Position buttonSpecify the reading position.
(5)Resolution button
(6)Maximum Exposure Time button
The reading position is automatically set. Select the reading size/direction using the Reading Size button/Reading Direction button.
[OFF]:
Turns the automatic reading settings o. Select the reading size/direc-
tion/position using the Reading Size button/Reading Direction button/
Reading Position button.
Specify the reading resolution when using the AeroDR 3 1417HD and a
CR casette.
Change the maximum exposure time.
[Maximum Exposure Time]: displays the current maximum exposure
time.
[Default]: sets the value set to the Exam. tag.
[Longer]: sets the time that is one stage longer than the value set to the
Exam. tag.
58
Page 59
3.5 Exposure screen
IMPORTANT
• In the case of CR cassettes, "Standard" (175μm) or "High" (87.5μm) can be selected. The resolution in the case of Exam.
Tag for mammograms is however 87.5μm for "Standard" and 43.75μm for "High".
• In the case of AeroDR 3 1417HD, "Standard" (200μm) or "High" (100μm) can be selected.
Conrm the operational environment of this device before setting to "High" (100μm) because its image size is 3 times larger
than those of AeroDR 1417HQ, AeroDR 1417S, AeroDR 1717HQ, AeroDR 1012HQ, AeroDR 2 1417HQ, or AeroDR 2
1417S.
• In the case of AeroDR 1417HQ, AeroDR 1417S, AeroDR 1717HQ, AeroDR 1012HQ, AeroDR 2 1417HQ, or AeroDR 2
1417S, the resolution is set to "Standard" (175μm). It is not possible to select the resolution.
• In some cases the maximum exposure time is xed by the X-ray device and is not displayed. For more details, contact Koni-
ca Minolta technical representatives.
• The maximum exposure time of the One Shot stitching exposure using the maximum exposure time button diers from that
This control panel can customize the button types and their layouts according to the function used. This control panel
is mainly used for image adjustments when operating this device using a portable X-ray device.
(1)
(2)
(3)
Chapter 3
NumberNameFunctions
Adjusts the S value.
(1) S Value slider/button
(2) G Value slider/button
(3) Custom area
HINT
• Contrast invert is a function for checking the image on the screen. It has no eect on the output.
• The S Value slider/button and G Value slider/button are always displayed.
• In the custom area, you can display the shortcut to the control panel.
This is the control panel to perform setting of host overlay.
3.5 Exposure screen
(1)
(2)
(3)
(4)
(5)
(6)
(7)
(8)
(9)
(10)
NumberNameFunctions
(1) Host/Printer switch button
(2) Same settings for host and printer
(3) Marker 1 selection buttonSet the marker to display.
(4)Marker 1 position buttonSet the position of the marker to display.
(5)Marker 2 selection buttonSet the marker to display.
(6)Marker 2 position buttonSet the position of the marker to display.
(7)Stamp selection buttonSet the stamp to display.
(8)Stamp position buttonSet the position of the stamp to display.
(9)Scale selection buttonSelect type of scale or grid.
(10)Preview buttonDisplays a preview image in which the set overlay content is reected.
Switches overlay for host and for printer.
Press this button when sending the same overlay conditions set for host
to printer. Not displayed when the printer is not registered as the output
device.
Chapter 3
65
Page 66
3.5 Exposure screen
3.5.15 Control panel (Overlay - for printer)
This is the control panel to perform setting of printer overlay.
(1)
(2)
(3)
(4)
(5)
(6)
(7)
(8)
(9)
(10)
NumberNameFunctions
(1) Host/Printer switch button
(2) Same settings for host and printer
(3) Marker 1 selection buttonSet the marker to display.
(4)Marker 1 position buttonSet the position of the marker to display.
(5)Marker 2 selection buttonSet the marker to display.
(6)Marker 2 position buttonSet the position of the marker to display.
(7)Stamp selection buttonSet the stamp to display.
(8)Stamp position buttonSet the position of the stamp to display.
(9)Scale selection buttonSelect type of scale or grid.
(10)Preview buttonDisplays a preview image in which the set overlay content is reected.
Switches overlay for host and for printer.
Press this button when sending the same overlay conditions set for printer to host. Not displayed when the host is not registered as the output
device.
66
Page 67
3.6 Viewer screen
This is the screen to perform image adjustment. More detailed image adjustment can be performed than the image
adjustment on the exposure screen.
(1)
(2)
(3)
(4)
(5)
(6)
(7)
Chapter 3
NumberNameFunctions
(1) Application bar
(2) Patient information display area Patient name, patient ID, date of birth, and sex, etc. are displayed.
(3) Tool bar (a)
(4)Image display areaImages to adjust are displayed.
(5)Tool bar tabSwitches various adjustment functions.
(6)Tool bar (b)Tools related to the selected tool bar tab are displayed.
(7)Return buttonReturns to the previous screen.
Date and time, operator name, exposure room name, icons of devices
connected to this device are displayed.
[S/G Drag]: Adjust the S and G values by dragging the image.
[S/G Value]: Adjusts S value and G value.
[ZOOM+]:
Switches to the zoom-in mode. Pressing it again cancels the zoom-in
mode. Pressing the image in the zoom-in mode enlarges the image 1
step at a time.
[ZOOM-]:
Switches to the zoom-out mode. Pressing it again cancels the zoom-out
mode. Pressing the image in the zoom-out mode reduces the image 1
step at a time.
[Fit]:
Displays the enlarged (or reduced) image to t in the frame size (size of
the image display area).
[Pan]: Moves the image.
[Rotate&Flip]: Rotates/ips the image.
[Grid Suppression]: Performs grid removal.
[Text]: Annotation is added/edited.
[ALL Reset]:
Resets all operations and restores the system to the status in which the
image was displayed.
67
Page 68
3.6 Viewer screen
HINT
• The Return button may be displayed within the tool bar (a) depending on the setting. For more details, contact Konica Mi-
This is the screen to check type and status of regular replacement parts when the time for replacement comes.
(1)
(2)
(4)
(3)
(5)
(6)
(7)
NumberNameFunctions
(1) Function switching tabSwitches between the [System Monitor] and [Warranty Component] tabs.
(2) Display selection menu
(3) Replacement unit listDisplays the reader name, replacement unit, and status.
ON/OFF switching button
(4)
(Replacement Notice)
(5)Unit usage status display
(6)Edit button
(7)System menu
Selects the type of replacement unit or replacement unit status.
Switches the replacement notication function on/o.
Displays the usage status and replacement estimation of the replacement
unit chosen from the list.
Used by service engineers to edit the information about the replacement
unit. This button is not displayed in the log-in mode.
Starts a utility or shuts down the system. This is displayed regardless of
tabs being selected.
[Media]: The media output screen is displayed.
[Utility]:
Starts utilities. This button is not displayed if the user authorization of
the logged-in operator is "Technologist".
[Update]: Refreshes information.
[Checkup]: Displays the checkup setting screen.
[Manual]:
The "CS-7 Operation Manual" is displayed. This button is not displayed
if the manual le is not deployed.
[TeamViewer]: Displays the TeamViewer startup button.
[Close]:
Exits the system monitor screen, and the previous screen is displayed.
Chapter 3
73
Page 74
74
Page 75
Chapter
4
General Operations
This chapter describes general operation methods of this device.
75
Page 76
4.1 Startup and shutdown
4.1.1 Startup of this device
Perform the following procedure when starting up this
device.
Turn on the power for X-ray devices, etc.
1
connected to this device.
Press the power switch on this device.
2
• Startup messages are displayed sequentially.
• Internal system initialization is performed while
the system startup screen is displayed.
• Upon completion of initialization, the initial
screen (patient registration screen or list screen)
is displayed, and this device is ready for operation.
Patient registration screen
IMPORTANT
• When sharing a REGIUS among several units of this
device, start the unit with the JM installed first; after
conrming startup of this unit, start the other units.
The DR Detector stops, linked to the
shutdown processing of this device.
Each device recovers from sleep or
transitions to sleep, linked to the startup/
shutdown processing of this device.
Details to link
78
Page 79
4.2 Login control
4.2.1 Login screen
It is possible to set a login screen that requires input of
user name and password when starting operations of
this device, or when resuming operations after returning from screen saver.
Login screen
HINT
• When the automatic login function is used, you are
automatically logged in when this device starts up and
the initial screen can be displayed.
• If LDAP login authentication has been enabled, the
domain input eld is displayed on the login screen.
•••••••••••••••••••••••••••••••••••••
4.2.2 Login
When using login mode, log in with the following procedure.
Logo is performed when switching users and so forth.
Press [
the bottom of the screen, then follow the procedure
described below to log o.
1
] to display the tool bar that is displayed at
Press [Logo].
4.2.4 Switch of operators
When 2 or more operator names are registered, their
names can be switched. Operator names can be
switched only when the automatic login function is enabled.
Press [Operator] in the application bar.
1
• The operator name selection screen is displayed.
Press the operator name to switch to it.
2
• Logging o.
• After logo, the login screen is displayed.
HINT
• [Logo] cannot be pressed until an image is acquired
from a reader device and the actual image is dis-
played.
• When [Logoff] is pressed during an examination, a
message stating that the examination is paused is displayed. Press [OK] on the conrmation screen to log
o, pausing the examination.
There are 2 types of input panel displayed on the screen, and you can input characters by pressing buttons. You can
input alphanumeric characters.
Display the input panel
z
Press the button for switching the input panel display.
Hide the input panelDisplay the input panel
The display switches every time the button for switching the input panel display is pressed.
Input alphanumeric
z
Press [ABC] or [QWERTY].
[ABC] input panel[QWERTY] input panel
Letters are displayed in alphabetical order, or in the order on a keyboard.
82
Page 83
4.4 Registration of the DR Detector
The DR Detector will be automatically registered when
it is inserted into an AeroDR Battery Charger/AeroDR
Battery Charger2 or connected to an Wired cable.
Insert the DR Detector into the AeroDR
1
Battery Charger/AeroDR Battery Charger2.
Alternatively, connect the DR Detector to
an Wired cable.
• The DR Detector is registered.
Check registration of the DR Detector in
2
the icon display of the application bar.
• When registered normally, the DR Detector icon
is displayed in the application bar.
Chapter 4
83
Page 84
84
Page 85
Chapter
5
Sequence of Examination
This chapter describes the operation steps
for examination procedures performed with this device.
85
Page 86
5.1 Screen transition of examinations
The following describes the screen transition from examination registration to examination end and the typical operations performed in each screen.
Meaning of the arrows
: Screen transition operation
Operation that returns to the previous
:
screen or an operation that can be omitted
(1) Initial screen
Patient registration screen
Screens are switched
over by selecting tabs.
List screen
• New registration of patient information
• Search registration
• Emergency registration
Press [Perform].
Press
[Back].
(2) Body part selection screen
• Manual body part selection
• Changing/deleting Exam.
Tags
Press [Perform].
Press
[
]/[].
• Selection of examination
(4) Viewer screen
• Change of S value/G value
• E/F/HS parameter adjustments
• Adjustment of ROI area
•
Positioning/editing of annotation
Press [Tool], etc.
Press [Return].
(3) Exposure screen
• Exposure with DR Detector/CR cassette
• Image adjustment
86
Press [
• Pressing [
] or [ ].
] will
produce an output
to the host or printer
depending on the
settings.
Page 87
5.2
Exposure procedure with DR Detector (Basic connection)
5.2.1 Exposure procedure ow
The following describes the basic ow of exposure procedure using the DR Detector. See each item for detailed procedures.
5.2.2
New registration of patient information
5.2.3 Selection of examination …… (2)
5.2.4 Manual body part selection ………… (3)
5.2.5 Exposure procedure ………………… (4)
5.2.6
Image adjustment on the exposure screen
… (1)
… (5)
Chapter 5
5.2.7 Examination completion and output … (6)
NumberTask
Perform a new patient registration in the patient registration screen.
(1)
(2) Select the examination target in the list screen.
(3) Set the exposure body part and exposure condition.
(4)Press the exposure switch.
(5)Adjust the exposed images after exposure. Change the S value/G value, move the ROI area, etc.
When all exposures are nished, the examination is completed. Image is simultaneously and automatically output upon
(6)
examination completion. (depending on the settings)
87
Page 88
5.2 Exposure procedure with DR Detector (Basic connection)
Input patient name, sex and date of birth.
5.2.2 New registration of
patient information
New registration of patient information is performed
with the patient registration screen. Select the
[PatientReg.] tab on the initial screen to display the patient registration screen.
Press the column for the patient ID.
1
4
• After each inputting item, press [Enter].
• If you press [Enter], the cursor will move to the
• If the length of an entry exceeds the dimensions of the
display pane of the item or if the entry spreads across
multiple lines, the long-sentence comment pop-up but-
ton [
•••••••••••••••••••••••••••••••••••••
] is displayed.
Press the input panel, and input the patient ID.
2
Press [Enter].
3
• Registration is performed in the patient data-
base of this device, and the exposure screen or
the body part selection screen is displayed.
HINT
• Because input of "Sex", "Pregnancy" are displayed
with selection buttons instead of the input panel, press
the selection buttons.
• When entering "Institution Residence", "Hospital
Department" and "Requesting Physician", select the
names from the pull-down menu or selection button,
or input them directly with the keyboard.
•••••••••••••••••••••••••••••••••••••
88
• With the input item as a search key, search the patient
database for this device, or a higher system.
• When there is a patient ID that matches, name, sex
and age are displayed.
When there is no patient ID that matches, the cursor
5.2 Exposure procedure with DR Detector (Basic connection)
HINT
• Linking to the insertion/removal of the DR Detector
into/from the wall stand, the icons inside the applica-
tion bar change as follows.
When the DR Detector is inserted into the wall stand
When the DR Detector is removed from the AeroDR
Battery Charger
• Movement of focus for the image display area is linked
to the state of insertion/removal of the DR Detector
into/from the wall stand.
• When interlock switch is released after receiving the
exposure timing from an X-ray apparatus, the DR
Detector is enabled to allow an operator to press the
exposure switch of the X-ray apparatus. A warning
sounds when the interlock cannot be released (cannot
press the exposure switch).
• X-ray exposure information is acquired automatically
when linked to the X-ray device.
•
The exposure eld is automatically recognized, and displayed on the screen as the eective image area. If automatic recognition of the exposure eld has failed, set it
manually by pressing [Cropping] in the [Image Proc] tab.
• The exposure eld automatic recognition function can
be set for each body part, and the xed eective im-
age area can be selected.
• Congure irradiation elds of at least 5x5 cm for the
following areas.
Infant hip/Front view of lumbar/Side view of lumbar/
Radial position of lumbar/Front view of head/Towne
view of the head/Turkey saddle/Cheekbone axis/Tem-
poromandibular joint/Front view of jawbone/Side view
of jawbone/Front view of the child chest/Side view of
the child chest/Progress observation of child chest/
Front view of infant chest/Side view of infant chest and
abdomen/Progress observation of infant chest and ab-
domen/Newborn chest/Newborn skeleton/Front view
of cervix/Side view of cervix/Radial position of cervix/
Cervix opening/Ribs/Side view of sternum/Radial posi-
tion of sternum/Clavicle/Shoulder joint/Scapula/Ster-
noclavicular joint/Hips/Limbs (excluding radial position
of heel bone)
Congure settings for other areas so that at least one
edge of the irradiation eld is at least 15 cm.
If these conditions are not satised, irradiation elds
5.2 Exposure procedure with DR Detector (Basic connection)
5.2.7 Examination completion
and output
When all exposures are nished, end the examination
and perform output.
Press [ ].
1
• When the examination is nished, exposed im-
ages are batch output to a preset output destination.
IMPORTANT
• Always check the X-ray irradiation results before nishing the examination.
• When the examination ends and no exposure images
exist, the examination itself is deleted. At this time, a
conrmation screen is displayed on whether to delete
the examination.
→ Press [OK] to delete the examination itself.
→ Press [Cancel] to return to the previous screen.
* A message "No images are captured. Cannot com-
plete study." may be displayed depending on the
settings. In such cases, press the [Pause] button
or [Termination] button. Either of the two buttons is
• If there is an unexposed image, a conrmation screen
is displayed.
•••••••••••••••••••••••••••••••••••••
•••••••••••••••••••••••••••••••••••••
Chapter 5
• The display returns to the initial screen.
→ Press [OK] to complete the examination by delet-
ing data for the unexposed portion and display the
initial screen.
→ Press [Cancel] to return to the previous screen.
• When outputting to a host, the output destination can
be switched between the predetermined host and a
backup host. Furthermore, when the output conditions
are dierent for each backup destination, recalculation
of output size is performed.
• Depending on the setting at the time of installation,
on the edge of the output lm (the output image when
output to a host) a mark is added showing whether the
image is ipped or not.
: Flipped
: Not ipped
• If the AeroDR 3 1417HD is used to perform an exposure with a reading resolution of 100μm, the resolution
can be converted to 175μm and output, depending on
• With Aero Sync mode exposure, X-rays with default intensity that reached the DR Detector are detected, and image acquisition is started. Select any one of the following X-ray detection levels during exposure according to the model or conditions of
the DR Detector to be used.
– Small
Used for general imaging such as that of the chest and limb bones.
– Med
This is eective for imaging of the abdomen or lumbar front of patients with thick bodies as well as in the case where the
direct X-ray region does not enter the DR Detector due to the narrowed exposure eld. Be careful not to subject the DR
Detector to strong shocks during exposure.
– Large
This is more eective than the medium detectionlevel for imaging of the abdomen or lumbar front of patients with thick
bodies as well as in the case where the direct X-ray region does not enter the DR Detector due to the narrowed exposure
eld. Request that the patient stands still for as long as possible during exposure.
– Extra Large
Exposures can be performed at a level higher than [Large]. This is eective for imaging of the abdomen or lumbar front of
patients with thick bodies as well as in the case where the direct X-ray region does not enter the DR Detector due to the
narrowed exposure eld. Be careful not to subject the DR Detector to strong shocks during exposure.
• [Extra Large] is displayed when using either the AeroDR 3 1417HD, AeroDR 2 1417HQ or AeroDR 2 1417S.
• The control panel display [AeroSync] and items that can be selected dier depending on the type of DR Detector.
If there are any radioactive substances remaining in the patient's body after nuclear medicine testing or some other reason, the
AeroDR 2 1417HQ and the AeroDR 2 1417S may recognize that X-rays were emitted and start image transfer. When the im-
age is transferred, change the X-ray sensitive setting of the image processing controller to "Large" before performing exposure.
If you change the X-ray sensitivity to "Large", make sure that all of the following conditions are met before performing exposure. If you perform exposure under other conditions, you may not be able to get an image.
– On the DR Detector, set up an area (2 cm x 2 cm or larger) on which the X-rays are emitted directly.
– Emit the X-rays onto the entire DR Detector.
– Emit the X-rays by using a tube current of 140 mA or higher or a tube voltage of 90 kV or higher.
•
If there are any radioactive substances remaining in the patient's body after nuclear medicine testing or some other reason, the
AeroDR 3 1417HD may recognize that X-rays were emitted and start image transfer. When the image is transferred, change
the X-ray sensitive setting of the image processing controller to "Large" or "Med" before performing exposure.
If you change the X-ray sensitivity to "Large", make sure that all of the following conditions are met before performing exposure. If you perform exposure under other conditions, you may not be able to get an image.
– Emit the X-rays onto the entire DR Detector.
– Emit the X-rays by using a tube current of 140 mA or higher or a tube voltage of 90 kV or higher.
• Once the X-ray detection level is selected, it is kept until an examination is completed if it is not changed. If you want to re-
turn it to the initial detection level for each exposure, contact Konica Minolta technical representatives.
5.3 Exposure with DR Detector (Aero Sync mode)5.3 Exposure with DR Detector (Aero Sync mode)
5.3.1 Exposure ow
The following describes the basic ow of exposure using the DR Detector. See each item for detailed procedures.
5.3.2
New registration of patient information
5.3.3 Selection of examination …… (2)
5.3.4 Manual body part selection ………… (3)
5.3.5 Exposure ……………………………… (4)
5.3.6
Image adjustment on the exposure screen
… (1)
… (5)
5.3.7 Examination completion and output … (6)
NumberTask
Perform a new patient registration in the patient registration screen.
(1)
(2) Select the examination target in the list screen.
(3) Set the exposure body part and exposure condition.
(4)Perform exposure.
(5)Adjust the exposed images after exposure. Change the S value/G value, move the ROI area, etc.
When all exposures are nished, the examination is completed. Image is simultaneously and automatically output upon
(6)
examination completion. (depending on the settings)
Chapter 5
99
Page 100
5.3 Exposure with DR Detector (Aero Sync mode)
5.3.2 New registration of
patient information
New registration of patient information is performed
with the patient registration screen. Select the
[PatientReg.] tab on the initial screen to display the patient registration screen.
Press the column for the patient ID.
1
Input patient name, sex and date of birth.
4
• After each inputting item, press [Enter].
• If you press [Enter], the cursor will move to the
next item.
HINT
• If the length of an entry exceeds the dimensions of the
display pane of the item or if the entry spreads across
multiple lines, the long-sentence comment pop-up but-
• Because input of "Sex", "Pregnancy" are displayed
with selection buttons instead of the input panel, press
the selection buttons.
• When entering "Institution Residence", "Hospital
Department" and "Requesting Physician", select the
names from the pull-down menu or selection button,
or input them directly with the keyboard.
•••••••••••••••••••••••••••••••••••••
100
• With the input item as a search key, search the patient
database for this device, or a higher system.
• When there is a patient ID that matches, name, sex
and age are displayed.
When there is no patient ID that matches, the cursor