Before using this instrument,
please read this manual.
Page 2
Safety Symbols
The following symbols are used in this manual or CM-3700A to prevent accidents
which may occur as a result of incorrect use of the instrument.
Denotes an instruction regarding a safety warning or note.
Read the instruction carefully to ensure safe and correct use.
Denotes an instruction regarding the risk of electric shock.
Read the instruction carefully to ensure safe and correct use.
Denotes an instruction regarding the risk of fire.
Read the instruction carefully to ensure safe and correct use.
Denotes a prohibited action.
This action must never be performed.
Denotes an instruction.
This instruction must be strictly adhered to.
Denotes a prohibited action.
Never disassemble the instrument.
Denotes an instruction.
Be sure to disconnect the AC adapter from the AC outlet.
This symbol indicates A.C.
This symbol indicates D.C.
This symbol indicates class II protection against electric shock.
Trademarks
• Windows® is a registered trademark of Microsoft Corporation in the United States and other countries.
Notes on this Manual
• Copying or reproduction of all or part of the contents of this manual without KONICA MINOLTA’s permission is strictly prohibited.
• The contents of this manual are subject to change without prior notice.
• Every effort has been made in the preparation of this manual to ensure the accuracy of its contents.
However, should you have any questions or find any errors, please contact your retailer or a KONICA
MINOLTA-authorized service facility.
• KONICA MINOLTA will not accept any responsibility for consequences arising from the use of the instrument.
Page 3
Safety Precautions
To ensure correct use of this instrument, read the following points carefully and adhere to them.
After you have read this manual, keep it in a safe place where it can be referred to anytime a question arises.
(Failure to adhere to the following points
WARNING
may result in death or serious injury.)
Do not use the instrument in places where
flammable or combustible gases (gasoline,
etc.) are present.Doing so may cause a fire.
Always use the AC adapter supplied as
a standard accessory or the optional AC
adapter, and connect it to an AC outlet
of the rated voltage and frequency. If an
AC adapter other than those specified by
KONICA MINOLTA is used, it may result in
damage to the unit, fire or electric shock.
If the instrument will not be used for a
long time, disconnect the AC adapter
plug from the AC outlet. Accumulated
dirt or water on the prongs of the
AC adapter’s plug may cause a
fire and should be removed.
Do not forcibly bend, twist, or pull the AC
adapter power cable. Do not scratch or alter
the power cable or place heavy objects
on it. Doing so may damage the power
cable and cause a fire or electric shock.
Do not disassemble or modify the
instrument or the AC adapter. Doing
so may cause a fire or electric shock.
Take special care not to allow liquid or
metal objects to enter the instrument.
Doing so may cause a fire or electric
shock. Should liquid or metal objects
enter the instrument, turn the power
OFF immediately, disconnect the
AC adapter plug from the AC outlet,
and contact the nearest Konica
Minolta-authorized service facility.
The instrument should not be operated if it
is damaged or the AC adapter is damaged,
or if smoke or odd smells occur. Doing
so may cause a fire. In such situations,
turn the power OFF immediately,
disconnect the AC adapter plug from the
AC outlet and contact the nearest Konica
Minolta-authorized service facility.
Always hold the plug itself when
disconnecting the AC adapter plug
from the AC outlet. Pulling on the
power cable may damage it and
cause a fire or electric shock.
Insert the power plug fully and
securely. Incomplete inserting may
cause fire or electric shock.
(Failing to adhere to the following points may result in
CAUTION
Do not perform measurement with the
specimen measuring port
directed towards your eyes.
Doing so may damage your eyes.
Be careful not to get your hand caught in
the openable section of the instrument.
Doing so may result in injury.
When carrying the instrument, fix it with tape,
etc. to prevent the transmittance specimen
chamber cover from opening and closing.
Doing so may result in your hand caught by
the cover and result in injury.
injury or damage to the instrument or other property.)
Do not insert or disconnect the AC adapter
plug from an AC outlet with wet hands.
Doing so may cause electric shock.
Do not place the instrument on an
unstable or sloping surface. Doing so
may result in its falling or overturning,
causing injury. Be careful not to drop
the instrument when carrying it.
Make sure that the AC outlet is located
near the instrument and that the AC
adapter plug can be connected to and
disconnected from the AC outlet easily.
When cleaning, disconnect the power plug.
Connecting may cause electric shock.
1
Page 4
INTRODUCTION
This spectrophotometer is designed for spectral measurement of color and color differences in
various industries. It can measure both reflected and transmitted color with high accuracy.
Packing Materials
General Packing Materials
Keep all packing materials (cardboard box, cushioning material, plastic bags, etc.) in a safe
place. The CM-3700A is a precision measuring instrument. They can be used to protect the
instrument from impact and vibration during shipment to Konica Minolta for maintenance.
Should they be lost or damaged, contact the nearest Konica Minolta-authorized service facility.
Protective cap for the specimen measuring port (integrating sphere opening)
The CM-3700A is delivered with no target mask attached. To protect the specimen
measuring port, a protective cap is attached to the specimen measuring port.
This protective cap must be removed before using the CM-3700A.
When you transport the CM-3700A to another place, the protective cap
must be attached. Keep the protective cap in a safe place after putting it in the accessory case (CM-A156).
Protective cap
2
Page 5
Notes on Use
Be sure to use this instrument properly. Use of this instrument in ways other than those specified in
this manual may result in risk of injury, electric shock, instrument damage, or other problems.
Operating Environment
• Use the CM-3700A at an ambient temperature between 13°C and 33°C and relative humidity 80% or less (at
33°C) with no condensation. Be sure to use the instrument within this range. Do not use it in areas of rapid
temperature changes.
• Do not leave the CM-3700A in direct sunlight or near sources of heat, such as stoves, etc. The internal
temperature of the instrument may become much higher than the ambient temperature in such cases.
• Do not use the CM-3700A in areas where dust, cigarette smoke or chemical gases are present. Doing so
may cause deterioration in performance or a breakdown.
• Do not use the CM-3700A near equipment which produces a strong magnetic field (such as speakers, etc.).
• The CM-3700A belongs to installation category I products (equipment which is powered by an AC adapterconnected to commercially available power).
• The CM-3700A belongs to pollution degree 2 products (equipment which may cause temporary electrical
hazards due to contamination or condensation or products which are used in such an environment).
• Do not use the CM-3700A at altitudes higher than 2,000 m.
• The CM-3700A and the AC adapter supplied as a standard accessory have been designed exclusively for
indoor use. They should never be used outdoors because rain or other factors may damage the instrument.
Measurement
• Make sure no dirt or dust get into the specimen measuring port.
• When using the instrument for long periods of time, the measurement value may change depending on
changes in the environment. Therefore, in order to achieve accurate measurements, we recommend that
white calibration be done regularly using the White Calibration Plate.
White Calibration Plate
• The calibration data for the White Calibration Plate was measured at 23°C.
To achieve the highest accuracy when measuring absolute values (colorimetric values), calibration and
measurement should be performed at 23°C.
• Do not allow the White Calibration Plate to get scratched or stained with such as fingerprints.
• Do not move the White Calibration Plate while it has been caught by the sample holder. Doing so may
damage the White Calibration Plate.
• When the White Calibration Plate is not in use, be sure to close the cover so that the White Calibration
Plate is not exposed to ambient light.
Target Mask
• Do not touch the Target Mask’s inner surface (white-coated surface) by hand, scratch it or make it dirty.
• When not in use, Target Masks should be stored in the accessory case (CM-A156) so that they will not be
exposed to ambient light.
Power Source
• Make sure that the power switch is set to OFF (“ I “) when the CM-3700A is not in use.
• Always use the AC adapter supplied as a standard accessory (AC-A308) and connect it to an AC outlet of
the rated voltage and frequency.
• Use an AC power supply of the rated supply voltage (within ±10%).
• Do not connect the AC adapter to an overloaded electrical circuit. In addition, do not wrap or cover the AC
adapter with cloth or other material while in use. Doing so may cause an electric shock or fire.
3
Page 6
System
• Do not subject the CM-3700A to strong impact or vibration. Doing so may cause deterioration of performance or breakdown.
• The specimen measuring port and integrating sphere are extremely precise components, and great care
should be taken to prevent them getting dirty or exposing them to impacts. When the CM-3700A is not in
use, be sure to attach a target mask to the measuring port to prevent entry of foreign matter.
• The CM-3700A may cause interference if used near a television, radio, etc.
• Since the CM-3700A uses a microcomputer, external magnetic noise may cause malfunction. In this case,
turn the power OFF, and wait 30 minutes, and then turn it ON again.
About Transmittance Measurement
• Do not spill a specimen or other liquid on the instrument. If any liquid is spilled on the instrument, wipe it off
immediately with a soft, dry cloth.
4
Page 7
Notes on Storage
• The CM-3700A should be stored at temperatures between 0°C and 40°C, and at a relative humidity of 80%
or less (35°C) without condensation. Do not store the instrument in areas subject to high temperatures,
high humidity, sudden changes in temperature, or where freezing or condensation may occur, because
these circumstances may cause a breakdown. It is recommended to store the CM-3700A with a drying
agent at a temperature around 20°C.
• Do not leave the CM-3700A inside a car such as in the trunk. Otherwise, the temperature and/or humidity
may exceed the allowable range for storage during midsummer or midwinter, resulting in a breakdown.
• Keep the packing materials used for shipment and use them to transport the CM-3700A. This protects the
instrument from sudden changes in temperature, vibration, and shock.
• Do not store the CM-3700A in areas where dust, cigarette smoke or chemical gases are present. Doing so
may cause deterioration in performance or a breakdown.
• Entry of dust into the specimen measuring port will hinder accurate measurement. When the instrument
is not in use, you must close the transmittance chamber cover and cover the instrument with the supplied
Dust Cover to prevent the entry of dust into the integrating sphere.
• The White Calibration Plate may become discolored if left exposed to light. Therefore, make sure to close
the cover when it is not in use so that the White Calibration Plate is not exposed to ambient light during
storage.
• The Target Masks may discolor if they are left exposed to light. When they are not in use, keep them in a
safe place to prevent exposure to light and to protect them from scratches and dust. And store them in the
accessory case (CM-A156).
• Take care not to leave the CM-3700A for a long period of time with a target mask attached. The sample
holder may stick to the target mask.
• Be sure to keep all packing materials (cardboard box, cushioning material, plastic bags, etc.). They can be
used to protect the instrument during transportation to the service facility for maintenance (re-calibration
etc.).
Notes on Cleaning
• If the CM-3700A becomes dirty, wipe it with a soft, clean dry cloth. Never use solvents such as thinner or
benzene.
• If the White Calibration Plate becomes dirty, wipe it with a soft, clean dry cloth. If dirt is difficult
to remove, wipe it off with a cloth dampened with commercially-available lens cleaning solution.
Then remove the solution with a cloth dampened with water, and leave the plate to dry.
• If the inner surface (white-coated surface) of the Target Masks, or the inside of the integrating sphere, getdirty, contact a Konica Minolta-authorized service facility.
• Should the CM-3700A break down, do not try to disassemble and repair it by yourself. Contact a Konica
Minolta-authorized service facility.
Disposal Method
• Make sure that the CM-3700A and its accessories and packing materials are either disposed of or recycled
correctly in accordance with local laws and regulations.
Used to perform white calibration for measurement
of reflectance and to perform measurement
of transmittance. A CD-ROM containing white
calibration data and software for writing the white
calibration data is supplied with this accessory.
Target Mask CM-A91
CM-A92
CM-A93
Used to change the illumination area (measurement
aperture) according to the specimen.
Measurement and illumination (aperture area at specimen
surface) areas for each target mask are as follows:
CM-A91 (SAV) : 3 × 5 mm / 5 × 7 mm
CM-A92 (MAV) : ø8 mm / ø11 mm
CM-A93 (LAV) : ø25.4 mm / ø30 mm
Zero Calibration Box CM-A155
Used to perform zero calibration for measurement
of reflectance and to perform haze measurement
at measurement of transmittance.
AC Adapter AC-A308
Used to supply power from an AC
outlet to the CM-3700A.
Input: 100 to 240 V 50/60 Hz
Output: 8 V 1.5 A
Plug design: Center-negative
USB Cable (3 m) IF-A21
Used to connect the instrument to
a personal computer (PC).
Accessory Case CM-A156
Dust Cover CM-A69
8
Page 11
Optional Accessories
Software
SpectraMagic™ NX CM-S100w
This software provides various functions (e.g., data
processing and file management) and allows the user
to operate the CM-3700A using a personal computer.
Transmittance Specimen Holder CM-A96
Used to hold the specimen for measurement
of transmittance. It can hold specimens
up to 22.5 mm (7/8 in.) thick.
(MAV) and ø30mm (LAV) measurements) according
to the specimen and attach it to the CM-3700A.
⑦ Transmittance specimen chamber cover ....Covers the transmittance specimen chamber when slid closed.
⑧ USB connection terminal ............................ Used to connect the instrument to a PC with the supplied USB cable
(IF-A21 or IF-A22).
⑨ AC adapter input socket..............................Used to connect the AC adapter supplied with the CM-3700A.
11
Page 14
Measurement Procedure
¥ This manual explains how to prepare the CM-3700A and how to set a specimen.
¥ The CM-3700A is controlled by a PC to perform measurements.
¥ For a description of measuring method using SpectraMagic™ NX
(optional), refer to the SpectraMagic™ NX instruction manual.
Flow of Preparation and Measurement
Connecting the PCConnect the CM-3700A to the PC with the USB cable.
(P.13)
Connecting the AC adapterConnect the CM-3700A to the AC
outlet with the AC adapter. (P.14)
Starting the PC Turn on the PC to be used to control
(starting Windows)the CM-3700A.
Starting the softwareStart the software and set it for control of the CM-3700A.
Turning the power ON Turn the power ON. (P.15)
Attaching a target mask
Performing zero calibration
Performing white (100%)
calibration
<Reflectance measurement><
Attach the target mask
to be used. (P.16)
Position the zero calibration
box and perform zero
calibration. (P.19)
Position the white
calibration plate and perform
white calibration. (P.20)
Transmittance measurement
Attach the LAV (Ø25.4 mm)
target mask and set the white
calibration plate in position. (
Position the Transmittance
Zero Calibration Plate and
perform zero calibration. (P.21)
Perform 100% transmittance
with nothing or a cell
containing distilled water in the
transmittance chamber. (P.22)
P.16
>
)
Positioning a specimen Position the specimen on / in the CM-3700A. (P.23)
Performing measurement Perform measurements.
Turning the power OFF When measurements are complete, turn the power OFF. (P.15)
Exit out of the software and turn off the PC.
12
Page 15
Connecting a Personal Computer
Connect the instrument to a PC with the supplied USB cable IF-A21 (3 m).
Memo
• To connect the instrument with a PC, it is recommended that you use software that enables connection
and operation of the instrument (such as the optional Color Management Software SpectraMagic™ NX).
• The USB communication port of the instrument conforms to USB 1.1.
{
To connect the instrument to a PC, you need to install the USB driver dedicated to the CM-3700A. Install
the USB driver supplied with the software that enables connection and operation of the instrument.
{
The instrument is not designed to be powered via the USB cable. You need to connect the AC adapter.
{
Make sure that the USB connector plug is oriented correctly and connected securely.
{
When connecting/disconnecting the USB cable, be sure to hold the connector plug. Do not pull on or
forcibly bend the cable. Otherwise, wire breakage may result.
{
Make sure that the cable has sufficient length. Putting tension on the cable may cause connection failure
or wire breakage.
{
To connect the USB cable connector, check the shape of the receptacle (connection terminal) and insert
the connector fully until it is secured.
Operating Procedure
In general, a USB cable can be connected/disconnected while the instrument is turned ON; however,
you need to turn OFF the instrument in the procedure below. See P.14 for how to connect the AC adapter
and P.15 for how to switch the power on and off.
Turn OFF the instrument (Press the “ ” side
1.
of the Power switch.).
Connect the B connector of the USB cable to
2.
the USB connection terminal (B type) of the
instrument.
{
Fully insert the connector and ensure secure
connection.
Connect the A connector of the USB cable to
3.
the USB port of the PC.
Connect the AC adapter and turn ON the
4.
instrument
{
When you are prompted to install the USB driver,
specify the USB driver included with the software
or the white calibration data CD and complete
the installation.
{
When using the optional Color Management
Software SpectraMagic™ NX, refer to the
SpectraMagic™ NX Installation Guide.
{
After installation of the USB driver has finished,
switch the instrument off for a few seconds and
then switch it back on.
(Press the “ I ” side of the Power switch.)
.
13
Page 16
Connecting the AC Adapter
WARNING
Always use the AC adapter supplied as a standard accessory or specified replacement AC
adapter with the CM- 3700A, and connect it to an AC outlet of the rated voltage and frequency.
Failure to do so may damage the CM- 3700A or the AC adapter, causing a fire or electric shock.
If the CM-3700A will not be used for a long time, disconnect the AC adapter from the AC outlet.
Accumulated dirt or water on the prongs of the AC adapter's plug may cause a fire and should be removed.
Do not insert or disconnect the AC adapter with wet hands. Doing so may cause electric shock.
Insert the power plug fully and securely. Incomplete inserting may cause fire or electric shock.
Do not disassemble or modify the AC adapter. Doing so may cause a fire or electric shock.
Do not unplug or plug in the AC adapter with the instrument’s power switch set to ON. Doing so may
cause malfunction.
Connecting Procedure
Make sure that the power switches of both
1.
CM-3700A and host PC are set to OFF (" ").
Insert the output plug of the AC adapter into
2.
the AC adapter input socket on the rear of the
CM-3700A.
Insert the input plug of the AC adapter into an
3.
AC wall outlet.
{
The AC Adapter AC-A308 supplied as the
standard accessory must be used.
{
Before disconnecting the AC adapter, the power
switch must be set to OFF (“ “).
14
Page 17
Turning Power ON and OFF
WARNING
The CM-3700A should not be operated if the CM-3700A or the AC adapter is damaged, or smoke
or strange odors occur. Doing so may result in a fire. In such situations, turn the power OFF
immediately, disconnect the AC adapter from the AC outlet, and contact the nearest Konica Minoltaauthorized service facility.
Procedure
To turn the power ON and to light the lamp on
1.
the power switch, set the power switch to ON
(“ I “) .
To turn the power OFF, set the power switch
2.
to OFF (“ “).
15
Page 18
Attaching a Target Mask
WARNING
Do not place the CM-3700A on an unstable or sloping surface. Doing so may result in its
falling or overturning, causing injury. Take care not to drop the CM-3700A when carrying it.
Be careful around openings in the CM-3700A. Failure to do so
may result in fingers being trapped causing injury.
The CM-3700A allows you to select a target mask from the following three types according to the specimen
and your application.
Target mask
CM-A91 (for LAV 3×5 mm measurements: illumination area: 5×7 mm)
CM-A92 (for MAV ø8 mm measurements: illumination area: ø11 mm)
CM-A93 (for SAV ø25.4 mm measurements: illumination area: ø30 mm)
Procedure
Pull the sample holder toward you and keep it
1.
open.
{
The sample holder will remain open when
opened more than 70 degrees.
Remove the currently attached target mask
2.
by pulling it toward you.
Memo
• The target mask is held on by magnets.
Align the target mask positioning holes of the
3.
target mask to be attached with the target mask
positioning pins of the Spectrophotometer's
specimen port baseplate and slide the target
mask in against the specimen port baseplate.
Memo
• The target mask should be attached in the direction
that mak es the white painted s urface face inside
(toward the CM-3700A side) and position the letters
“MADE IN JAPAN” at the top.
Release the sample holder to close it.
4.
Notes on Use of Target Mask
{
Take care not to scratch or make the inner surface (white-coated surface) of the target masks dirty with
such as fingerprints.
{
The target masks may become discolored if left in areas exposed to light. Therefore, make sure that
target masks which are currently not in use are kept inside the accessory case (CM-A156) to prevent
exposure to light.
{
When not using the CM-3700A, attach one of the target masks or the protective cap to prevent dust
entering the integrating sphere.
{
Take care not to leave the CM-3700A for a long period of time with a target mask attached. The sample
holder may stick to the target mask.
16
Page 19
Calibration of Reflectance and transmittance measurements
Calibration for Reflectance Measurements
After the Spectrophotometer has been switched on, zero calibration and white calibration must
be performed before spectral reflectance measurements are taken. Zero calibration is performed
using the Zero Calibration Box (which has a reflectance of 0%) to set the zero reference point for
reflectance measurements; white calibration is performed using a white calibration plate (which
has a reflectance of approximately 100% and for which spectral reflectance data are provided as
white calibration data) to set a high-reflectance reference point for reflectance measurements.
White calibration data
The white calibration data of the White Calibration Plate are provided under the following six kinds of
measurement conditions. When the CM-3700A is purchased, these data are already set within the instrument.
• Measurement area : SAV (3 x 5 mm) / Specular component : SCI
After the Spectrophotometer has been switched on, 0% calibration and 100% calibration must be performed
before spectral transmittance measurements are taken. 0% calibration is performed by blocking the light
path with an opaque material (a material which has a transmittance of 0%) to set the 0% reference point
for transmittance measurements. 100% calibration is performed using either air (when measuring solid
specimens) or distilled water (when measuring liquid specimens) to set a high-transmittance reference point
for transmittance measurements; in both cases (air and water) the transmittance is assumed to be 100% for
all wavelengths.
{
Since using the LAV target mask and setting the measurement area to LAV and the SCI/SCE setting
to SCI provides the most stable data for transmittance measurements, it is recommended that
transmittance measurements and calibration (both 0% calibration and 100% calibration) for transmittance
measurements be performed under these conditions.
{
The EEPROM of the Spectrophotometer has space for one set of calibration data for transmittance
measurements.
100% calibration data
Because it is necessary to set the 100% transmittance data to the CM-3700A for all wavelengths, 100%
calibration data are normally set to all wavelengths at shipment.
17
Page 20
Performing calibration
Calibration (both zero calibration and white calibration or both 0% calibration and 100% calibration) is effective only
while the Spectrophotometer remains switched on. If it is switched off and then switched on again, both zero calibration
and white calibration (or both 0% calibration and 100% calibration) must be performed before measurements are taken.
c
The calibration becomes ef fective only when being per formed under six kinds of measurement
conditions. Therefore, for example, when performing the measurement of SCI and SCE alternately,
performing the measurement by switching the measurement area in a sequential order, or by switching
the UV light quantity, if you perform zero calibration and white calibration (0% calibration and 100% of
calibration) in advance as to up to six kinds of necessary measurement conditions, you do not have to
perform calibration every time the measurement conditions are changed. However, to more accurately
perform measurement, it is recommended to perform zero calibration and white calibration (0%
calibration and 100% calibration) every time measurement conditions are changed.
c
When the calibration has been performed under the seventh kind of measurement conditions, the oldest
calibration becomes ineffective and the calibration performed under the seventh kind of measurement
conditions becomes effective as the latest. Moreover, when the calibration is re-performed under
the same measurement conditions as one calibration that was already performed, not that the oldest
calibration becomes ineffective, but the calibration already performed under the same measurement
conditions becomes ineffective and the re-performed calibration becomes effective as the latest.
Example) First, zero calibration and white calibration are performed under the following six kinds of
measurement conditions in numerical order.
① Measurement area
② Measurement area
③ Measurement area
④ Measurement area
⑤ Measurement area
⑥ Measurement area
:
SAV / Specular component
:
MAV / Specular component
:
LAV / Specular component
:
SAV / Specular component
:
MAV / Specular component
:
LAV / Specular component
:
SCI / Target mask
:
SCI / Target mask
:
SCI / Target mask
:
SCE
/ Target mask
:
SCE
/ Target mask
:
SCE
/ Target mask
:
SAV / UV light quantity 99.9
:
MAV / UV light quantity 99.9
:
LAV / UV light quantity 99.9
:
SAV / UV light quantity 99.9
:
MAV / UV light quantity 99.9
:
LAV / UV light quantity 99.9
In this case, the calibration can be performed under all the six kinds of measurement conditions. Moreover, the
calibration can be performed when UV light quantity is set except 99.9. However, in this case, the measured
value is calculated and output based on the nearest UV light quantity (in this example 99.9).
Next, if the calibration under the conditions
⑦ Measurement area : SAV / Specular component : SCE / Target mask : SAV / UV light quantity 99.9
is re-performed, not that the oldest calibration under the conditions ① becomes ineffective, but the
calibration under the conditions ④ , which was performed under the same measurement conditions,
becomes ineffective and the calibration under the conditions ⑦ becomes effective as the latest.
Next, if the calibration under the seventh kind of conditions
⑧ Measurement area : MAV / Specular component : SCI / Target mask : MAV / UV light quantity 90.0
is performed, the oldest calibration under the conditions ① becomes ineffective and the calibration under
the conditions ⑧ becomes effective as the latest. As a result, effective calibrations can be arranged in the
following chronological order.
② Measurement area : MAV / Specular component : SCI / Target mask : MAV / UV light quantity 99.9
③ Measurement area : LAV / Specular component : SCI / Target mask : LAV / UV light quantity 99.9
⑤ Measurement area : MAV / Specular component :
⑥ Measurement area : LAV / Specular component :
⑦ Measurement area : SAV / Specular component :
⑧ Measurement area : MAV / Specular component : SCI / Target mask : MAV / UV light quantity 90.0
{
The measured values are affected by UV light quantity to some extent. If you want to perform rigorous
measurement, after changing the UV light quantity, zero calibration and white calibration (0% calibration
and 100% calibration) should be performed. When performing the fluorescent measurement that does
not require rigorous measured values, the measurement should be performed with the UV light quantity
of 99.9.
SCE
/ Target mask : MAV / UV light quantity 99.9
SCE
/ Target mask : LAV / UV light quantity 99.9
SCE
/ Target mask : SAV / UV light quantity 99.9
{
Due to temperature changes in the surrounding environment and heat generation caused by repeated
applications, the indicated values may subtly deviate, but if the calibration is regularly performed, the
measurement accuracy will improve.
18
Page 21
Attaching the Zero Calibration Box
WARNING
Do not perform measurement with the specimen measuring port directed
towards your face. Doing so may cause damage to your eyes .
Be careful around openings in the CM-3700A. Failure to do so
may result in fingers being trapped causing injury.
The zero calibration box is used to perform zero calibration for measurement of reflectance and to perform
haze measurement at measurement of transmittance.
{
Make sure that there is nothing in the transmittance chamber.
{
Before performing zero calibration, attach the target mask to be used for measurements.
{
Before performing zero calibration, set the same measurement area, specular component (SCI/
SCE), and UV light quantity as when performing zero calibration using the software. In the fluorescent
measurement that does not require rigorous accuracy (that does not perform fluorescent calibration),
perform measurement under the condition that the UV cut-off filter does not cover the xenon lamp (UV
light quantity is 99.9).
Procedure
Pull the sample holder toward you and keep it
1.
open.
{
The sample holder will remain open when
opened more than 70 degrees.
Fit the projections of the zero calibration
2.
box into the grooves on the CM3700A and then close the sample
holder to hold the box in place.
Notes on Use of Zero Calibration Box
{
Take care not to scratch, touch, or make the inside of the zero calibration box dirty with such as fingerprints.
{
If the inside of the zero calibration box gets dirty, wipe it with a soft, clean, dry cloth.
{
If dirt is difficult to remove, dampen a cloth with commercially available lens cleaning liquid and wipe the
zero calibration box. Then wipe off the liquid with a cloth dampened with water, and leave the box to dry.
{
Should the inside of the zero calibration box get so dirty that it cannot be cleaned, replace the box with a
new one.
19
Page 22
Attaching the White Calibration Plate
CAUTION
Do not perform measurement with the specimen measuring port directed
towards your face. Doing so may cause damage to your eyes .
Be careful around openings in the CM-3700A. Failure to do so
may result in fingers being trapped causing injury.
The white calibration plate is used to perform white calibration for measurement of reflectance and to cover
the reflectance measuring port when performing transmittance measurements (zero calibration, 100%
calibration, measurement).
{
Make sure that there is nothing in the transmittance chamber.
{
Before performing white calibration, attach the target mask to be used for measurements.
{
Before performing zero calibration, set the same measurement area, specular component (SCI/SCE),
and UV light quantity as when performing zero calibration using the software.
In the fluorescent measurement that does not require rigorous accuracy (that does not perform fluorescent
calibration), perform measurement under the condition that the UV cut-off filter does not cover the xenon lamp (UV
light quantity is 99.9).
Procedure
Pull the sample holder toward you and keep it
1.
open.
c
The sample holder will remain open
when opened more than 70 degrees.
Position the white calibration plate so that it is
2.
held in place by the sample holder by fitting the
sample holder pad into the concave (rear) side
of the white calibration plate as shown at right.
Notes on Use of White Calibration Plate
Memo
• When performing white calibration, the white calibration data for the white calibration plate being used is
required. The white calibration data for the white calibration plate included with the instrument is stored in the
instrument’s memory at the time of shipment.
{
The white calibration plate may become discolored if left exposed to light. Therefore, when not in use,
make sure that the lid is closed to prevent exposure to light.
{
Take care not to scratch, touch, or make the white calibration plate surface dirty with such as fingerprints.
{
If the white calibration plate gets dirty, wipe it with a soft, clean, dry cloth.
{
If dirt is difficult to remove, dampen a cloth with commercially available lens cleaning liquid and wipe the
white calibration plate. Then wipe off the liquid with a cloth dampened with water, and leave the plate to dry.
Should the white calibration plate get so dirty that it cannot be cleaned, replace the plate with a new one.
If the White Calibration Plate is changed, reset the white calibration data to the new plate’s.
Updating White Calibration Data
c
You may the "Data Setting Tool software" stored on the CD-ROM accompanying White Calibration Plate
CM-A139 or the optional Color Data Software SpectraMagic™ NX to set the white calibration data.
20
Page 23
Attaching the Transmittance Zero Calibration Plate
The Transmittance Zero Calibration Plate is used for performing zero calibration for transmittance measurements.
c
Before starting transmittance calibration and measurements, attach the LAV (ø25.4 mm)
target mask and position the white calibration plate at the reflectance measuring port.
c
Before starting transmittance calibration and
measurements, set the measurement area to LAV and
the specular component to SCI using the software.
c
Before starting transmittance calibration and
measurements, set the same UV light quantity as
when performing measurements using the software.
In the fluorescent measurement that does not
require rigorous accuracy (that does not perform
fluorescent calibration), perform measurement
under the condition that the UV cut-off filter does not
cover the xenon lamp (UV light quantity is 99.9).
Procedure
(Optional Accessory)
Open the cover of the transmittance chamber.
1.
Position the Transmittance Zero Calibration
2.
Illumination
window
Plate in the transmittance chamber so that it
completely blocks the illumination window.
Memo
• When using the optional Transmittance Specimen
Holder, position the Transmittance Zero Calibration
Plate in the holder. For information on installing
the optional Transmittance Specimen Holder,
refer to the instructions included with it.
Close the cover of the transmittance chamber.
3.
Notes on Use of Transmittance Zero Calibration Plate
c
Take care not to scratch, touch, or make the surface of the transmittance zero calibration plate dirty with
such as fingerprints.
c
If the transmittance zero calibration plate gets dirty, wipe it with a soft, clean, dry cloth.
c
If dirt is difficult to remove, dampen a cloth with commercially available lens cleaning liquid and wipe the
zero calibration plate. Then wipe off the liquid with a cloth dampened with water, and leave the plate to dry.
c
Should the transmittance zero calibration plate get so dirty that it
cannot be cleaned, replace the plate with a new one.
21
Page 24
Performing 100% calibration
Two methods for performing 100% calibration for
transmittance measurements are possible:
• 100% calibration to air:
When the specimen to be measured is in sheet or solid
form, 100% calibration should be performed with the
transmittance chamber empty.
• 100% calibration to water:
When the specimen to be measured is in liquid form and will be
measured using a cell (optional accessory), 100% calibration
should be performed using distilled (or pure) water in the same
size and type cell as will be used for measurements.
{
Before starting transmittance calibration and measurements, attach the LAV (ø25.4 mm) target mask and
position the white calibration plate at the reflectance measuring port.
{
Before starting transmittance calibration and measurements, set the measurement area to LAV and the
specular component to SCI using the software.
{
Before starting transmittance calibration and measurements, set the same UV light quantity as when
performing measurements using the software.
In the fluorescent measurement that does not require rigorous accuracy (that does not perform fluorescent calibration),
perform measurement under the condition that the UV cut-off filter does not cover the xenon lamp (UV light quantity is 99.9).
After setting up the instrument as described below for your selected 100%
calibration method, perform 100% calibration from your software.
Setting up instrument for 100% calibration to air
Procedure
Open the cover of the transmittance chamber.
1.
Make sure that there is nothing
2.
(specimen, cell, Transmittance Zero
Calibration Plate, etc.) between the
illumination window and the receiving
window in the transmittance chamber.
Memo
• It is no problem that the optional Transmittance Specimen Holder CM-A96 is installed in the transmittance chamber.
Close the cover of the transmittance chamber.
3.
Setting up instrument for 100% calibration to water
Procedure
Open the cover of the transmittance chamber.
1.
Fill the cell with distilled (or pure) water to
2.
Illumination
window
approximately 2/3 (when using any of the cells
available as optional accessories) and position
the cell in the optional Transmittance Specimen
Holder installed in front of the illumination
window inside the transmittance chamber.
c
Although the Transmittance Specimen Holder can be installed in front of either the illumination
or receiving window, it is recommended that it is installed in front of the illumination window.
In this position, the illumination / observation geometry is diffuse illumination/0° viewing angle.
c
The cell used for calibration should be the same type and have the same
optical path length as the cell which will be used for measurements.
c
The cell should be positioned in the optional Transmittance Specimen Holder so that
the clear sides of the cell are facing the illumination and receiving windows.
Close the cover of the transmittance chamber.
3.
22
Transmittance
chamber cover
Re c eivi ng
window
Light
Cell or Plastic Cell
Page 25
Setting a Specimen
WARNING
Do not use the CM-3700A in places where flammable or combustible gases (gasoline fumes, etc.)
are present. Doing so may cause a fire.
Do not disassemble or modify the CM-3700A. Doing so may cause a fire or electric shock.
The CM-3700A should not be operated if it is damaged, or smoke or strange odors occur. Doing so
may result in a fire.
In such situations, turn the power OFF immediately, disconnect the AC adapter from
the AC outlet, and contact the nearest Konica Minolta-authorized service facility.
CAUTION
Do not perform measurement with the specimen measuring port directed
towards your face. Doing so may cause damage to your eyes.
Be careful around openings in the CM-3700A. Failure to do so
may result in fingers being trapped causing injury.
Reflectance measurement
{
Make sure that there is nothing in the transmittance chamber.
{
Before starting reflectance measurements, attach the target mask to be used for measurements.
{
Before starting reflectance measurements, set the measurement area, specular component, and UV light
quantity using the software.
Procedure
Pull the sample holder toward you and keep it
1.
open.
c
The sample holder will remain open when
it is opened more than 70 degrees.
Secure the specimen with the sample holder.
2.
Slide the measurement port baseplate
3.
lock in the direction of the arrow, open
the measurement port baseplate, and
check the position of the specimen.
{
Hold the measurement port baseplate when
sliding the measurement port baseplate lock to
keep it from falling open.
{
Do not apply strong pressure to the measurement
port baseplate when the measurement port
baseplate is open.
{
The measurement cannot be performed when
the measurement port baseplate is open.
Check the
position of the
specimen
23
Page 26
When widely opening the measurement port baseplate
{
If you remove the stopper below the measurement
port baseplate, the measurement port baseplate can
be widely opened.
{
Remove the two stopper attaching screws by turning
them counterclockwise using a Phillips-head screwdriver.
{
Keep the removed screws and stopper in a safe place.
{
When widely opening the measurement port baseplate,
install the CM -3700A at the edge of desk so that the
sample holder does not touch the desk below.
Repeat the procedure 2 and 3, and after deciding
4.
the position of the specimen so that the part that
you want to measure is located at the measurement
point, close the measurement port baseplate.
{
When shifting the position of the specimen, in
order to protect the surface of the specimen, you
should pull the sample holder toward the front
and keep it open.
{
When closing the measurement port baseplate, be sure that the measurement port baseplate lock
is at the open position (slid in the direction of the arrow). If the lock is in the closed position, the
measurement port baseplate cannot be closed.
{
Do not open the measurement port baseplate and the transmittance specimen chamber cover while
performing measurement. Doing so may result in inaccurate measurement.
two screws
Stopper
Open
position
The sample holder for measuring the reflectance
To measure the reflectance of a film - or plate-like
specimen, the specimen needs to be secured with
the sample holder. If the specimen cannot be secured
in this way, remove the sample holder and hold the
specimen against the measurement aperture.
The sample holder can hold
samples with thicknesses
up to 30 mm.
Removing Sample Holder
Procedure
To remove the specimen holder when
1.
taking reflectance measurements of large
spesimens, use a crosspoint screwdriver
to turn the four screws of the specimen
holder counterclockwise, remove the screws,
and then remove the specimen holder.
Sample
Sample holder
Sample holder
{
Keep the screws and sample holder in a safe
place.
About opacity measurements
When using the optional Color Management Software SpectraMagic™ NX, pairs of white-backed and blackbacked measurements can be taken and opacity can be calculated.
24
four
screws
Page 27
Transmittance measurements
{
Before starting transmittance calibration and
measurements, attach the LAV (ø25.4 mm) target
mask and position the white calibration plate at the
reflectance measuring port.
{
Before starting transmittance calibration and
measurements, set the measurement area to LAV
and the specular component to SCI using the
software.
{
Before starting transmittance calibration and
measurements, set the UV light quantity using the
software.
Procedure
Open the cover of the transmittance chamber.
1.
Place the specimen against the illumination
2.
window tightly. (When measuring a liquid
specimen, the specimen container must be
placed against the illumination window.)
{
The specimen must be placed so that the area
to be measured entirely covers the entire
illumination window.
Memo
• Although the specimen can be plaied in front of either the illumination or receiving window, it is reccommended
that it is placed in front of the illumination window. In this position, the illumination/observation geometry is
diffuse illumination/0° vowing angle. In the position in front of the receiving window, the illumination/observation
geometry is approximately 0° :0° . It does not strictly satisfy the difinition of sush as CIE No.15. However, this
presents no problem if the values are used as relative values.
• It is recommended that the optional Transmittance Specimen Holder CM-A96 be used to hold the specimen in place.
• In the case of a liquid specimen, use of an optional cell (CM-A97 to 99, A130 to 132) is recommended.
c
If you are going to use a container other than the above cells, it must be transparent and constructed in such
a way that it can be positioned with opposing surfaces parallel to both illumination and receiving windows.
{
Take care not to spill the liquid specimen on the CM-3700A. Should there be any spillage, immediately
wipe it with a soft, dry cloth.
{
Never measure flammable liquids.
{
No measurements must be performed if the specimen or its container is scratched or dirty. If you hold the
measuring surface of the specimen with your fingers, it will be marked with fingerprints, so always hold
another part of the specimen when setting it in place.
{
Air bubbles on the inner surface of the specimen’s container will hinder correct measurements, so make
sure that all air bubbles are removed before measurement. (Air bubbles sometimes develop even if the
container is just left standing still.)
{
Correct measurements will be hindered if water drops
develop on the specimen or its container due to
condensation.
{
Liquids containing particles may cause unstable
measurements due to precipitation of the particles.
Illumination
window
Re c eivi ng
window
Close the cover of the transmittance chamber.
3.
25
Page 28
The transmittance chamber for transmittance measurement
Set the liquid and transmittance specimen to the transmittance
chamber. If you cannot set the specimen because it touches the
Light Trap in the transmittance chamber, remove the Light Trap
before setting the specimen.
Transmittance chamber
Target Mask
(LAV) and White
Calibration Plate
Sample holder
How to remove / attach the Light Trap
{
Before removing or attaching the Light Trap, be sure to set the specular component to SCI using the
software.
Removing Light Trap
If the specimen for transmittance
1.
measurements cannot be positioned because
the light trap is in the way, the light trap
can be removed by turning the screw of the
light trap counterclockwise, removing the
screw, and then removing the light trap.
Replacing Light Trap
To replace the light trap, hook the flange of
1.
the light trap into the Spectrophotometer as
shown, inset the screw, and turn the screw
clockwise until snug.
About haze measurements
When the White Calibration Plate is positioned over the reflectance measuring aperture, the measurement geometry
for transmittance measurements becomes di:0°. When the Zero Calibration Box is positioned over the reflectance
measuring aperture, the measurement geometry for transmittance measurements becomes de:0°.
The illumination/observation system does not strictly satisfy the definition of haze (ASTM D 1003).
However, this presents no problem if the values are used as relative values.
When using the optional Color Management Software SpectraMagic™ NX, these two types of transmittance
measurements (di:0°, de:0°) with and without specimen can be taken and haze can be calculated.
26
Page 29
Cleaning the CM-3700A and Accessories
WARNING
Do not disassemble or modify the CM-3700A or AC adapter.
Doing so may cause a fire or electric shock.
The CM-3700A should not be operated if it is damaged, or if smoke or strange odors occur. Doing
so may result in a fire. In such situations, turn the power OFF immediately, disconnect the AC
adapter from the AC outlet, and contact the nearest Konica Minolta-authorized service facility.
CAUTION
When cleaning, disconnect the power plug. Connecting may cause electric shock.
Be careful around openings in the CM-3700A. Failure to do so
may result in fingers being trapped causing injury.
Zero Calibration Box and White Calibration Plate
Wipe gently with a dry soft cloth. If dirt is difficult to remove, dampen a cloth with commercially available lens
cleaning liquid and wipe. Then wipe off the liquid with a cloth dampened with water, and leave it to dry.
{
When cleaning, take care not to scratch the zero calibration box or white calibration plate.
Target Mask
Use a blower to remove dirt and dust from the target masks.
{
Do not touch the black-coated surface of the target masks with fingers or wipe it with a cloth. If the target masks
get so dirty that dirt cannot be removed using a blower, contact the nearest Konica Minolta-authorized service
facility.
Inside Integrating Sphere
Remove the light trap cover. (See P.26)Remove the light trap cover. (See P.26)
1.
{
Be sure to set the SCI/SCE setting to SCI before
removing the light trap.
{
Store the light trap and screw carefully.
Make sure that there is nothing placed
2.
against the illumination window.
Block the receiving window so that no dust
3.
or dirt enters.
Receiving windowIllumination window
Set the specular component to SCE using the
4.
software.
Open the Measurement port baseplate and
5.
remove dust and dirt using a blower.
{
Do not touch the white-coated inner surface of the
integrating sphere, wipe it with a cloth or place any
object against it. If the white-coated surface gets
so dirty that dirt cannot be removed using a blower,
contact the nearest Konica Minolta-authorized
service facility.
27
Page 30
Receiving Window
Set the measurement area to SAV using your
1.
software.
Memo
• This will cause the lens to move closer to the
receiving window, making it easier to clean.
Use a blower to remove dirt and dust from the
2.
receiving window.
{
Do not put your fingers into the receiving window or
touch any optical lenses.
Receiving window
28
Page 31
Error message
The following error messages may appear when you use SpectraMagic® NX CM-S100w, optional color
management software, to control the instrument.
If you see an error message, follow the instructions shown in the table below. If a problem persists, contact
the nearest Konica Minolta-authorized service facility.
Error MessegeProblem / possible causeActionRefer to page
Flash ErrorThe light amount of the
xenon lamp has been
lowered to 50% or less.
Degradation of the
•
xenon lamp
Contamination of the
•
integrating sphere
Flash not ready.The illumination unit has
not been charged.
Flash Error.The xenon lamp
failed to emit light.
Xenon lamp dead
•
Illumination circuit failure
•
Sensor failure"
•
Incorrect Calibration
Procedure.
A/D Error.A/D conversion failed.
Attached target mask is
for smaller measurement
area than the selected
measurement area.
Circuit is not
operating properly.
Not Ready.Charging for flashing
The calibration procedure
is not correct.
The count value is not
appropriate for zero (0%)
or while (100%) calibration.
A/D converter failure
•
Circuit failure"
•
No target mask is installed
or the target mask
installed is too small.
The motor for changing
the measurement
diameter, adjusting the
UV light amount, or
switching between SCI
and SCE is not working.
the xenon lamp is
not complete.
Clean the integrating
sphere according to the
instructions in "Cleaning
the CM-3700A and
Accessories" on page
27. If the problem
persists, contact the
nearest Konica Minoltaauthorized service facility.
Contact a KONICA
MINOLTA-authorized
service facility.
Retry measurement or
perform recalibration. If the
message persists, contact
the nearest Konica Minoltaauthorized service facility.
Use the zero calibration
box for zero calibration
and the while calibration
plate for white calibration.
Turn the power off and
then on again. If the
message persists, contact
the nearest Konica Minoltaauthorized service facility.
I n s t a l l a t a r g e t m a sk
suitable for or larger than
the measurement area or
diameter.
Turn the power off and
then on again. If the
message persists, contact
the nearest Konica Minoltaauthorized service facility.
Wait for at least 3 seconds
after the last flash.
If the message persists,
contact the nearest
Konica Minolta-authorized
service facility.
27
-
-
19,20
15
16
15
-
29
Page 32
No response from
instrument
Communication with
the instrument failed.
The power of the
•
instrument is off.
COM port setting
•
is incorrect.
Communication
•
settings are incorrect.
Hardware initialization
at power on (of lens,
filter and trap positions)
is in progress.
Time for Periodic
Calibration. Please
It is time to perform
periodic calibration.
contact the closest
Service Center.
Not Calibration data.
Set Calibration data.
The calibr a t ion data for
while calibration is not found.
Are you sure?
Check UV Settings.UV Settings has been
set to "100% Full".
UV adjustment not
possible.Change
Standard value or
Current settings did
not allow adjustment
of the filter position.
tolerance settings,
or use a different
Fluorescent Standard.
UV adjustment not
possible. UV is set
to XX and cannot be
Current settings did
not allow adjustment
of the filter position.
adjusted further in
the required direction.
Change Standard value
or tolerance settings,
or use a different
Fluorescent Standard.
Turn on the power of the
instrument.
Check and correct
the COM port and
communication settings.
15
After turning the power on,
wa it for t en and several
seconds until the lens, trap
and UV filter positions are
initialized.
Contact a KONICA
MINOLTA-authorized
service facility and request
periodic calibration.
Load the calibration data
into the instrument.20
Set the UV Settings to
"UV adjustment" .
Refer to the
SpectraMagic™
NX (optional)
instruction manual.
Change Standard value
or tolerance settings,
or use a different
Fluorescent Standard;
34
then retry "UV adjustment".
Change Standard value
or tolerance settings,
or use a different
Fluorescent Standard;
then retry "UV
34
adjustment".
-
-
30
Page 33
TROUBLESHOOTING GUIDE
If a problem occurs with the Spectrophotometer, please check the following points before
requesting service. If the problem continues to occur even after the suggested corrective
actions have been taken, contact the nearest Konica Minolta-authorized service facility.
ConditionCheckpointRecommended action
Reflectance measurement
results seem strange.
Transmittance
measurement results
seem strange.
Data input/output between
the Spectrophotometer
and a computer cannot
be performed.
Commands cannot be input
to the Spectrophotometer
from a computer.
Was specimen
positioned correctly?
Are the white calibration
data correct?
Was white calibration
performed correctly?
Was zero calibration
performed correctly?
Is t here a ny obstacle in the
transmittance chamber?
Was measurement mode
set to reflectance?
Was specimen
positioned correctly?
Was 100% calibration
performed correctly?
Was zero calibration
performed correctly?
Was measurement mode
set to transmittance?
Is the USB cable connected
correctly to both the
Spectrophotometer
and the computer?
I s th e so f t w a r e op e r a t i n g
correctly?
Open measurement port
baseplate and check
measurement point.
Set the correct white
calibration data.
Attach White Calibration
Plate correctly and perform
white calibration correctly.
Attach Zero Calibration
Plate correctly and perform
zero calibration correctly.
Make sure that there is nothing
in the transmittance chamber.
By referring to the operation
manual of the software, perform
the operation correctly.
Position specimen correctly.
Attach White Calibration Plate
correctly and perform 100%
calibration correctly using
the appropriate calibration
standard (air or distilled water).
Attach White Calibration
Plate correctly, block all light
from reaching the receptor
window, and perform zero
calibration correctly.
By referring to the operation
manual of the software, perform
the operation correctly.
Connect the USB cable correctly
bitween the Spectrophotometer
and the computer.
By referring to the operation
manual of the software, perform
the operation correctly.
Set POWER switch of
Spectrophotometer to OFF
and then set it back to ON.
Refer
to page
23
20
20
19
23
—
23
22
21
—
13
—
15
31
Page 34
32
Page 35
Explanations
33
Page 36
lIIumination/Observation System
Measuring Reflected Colors
The flow of measurement is shown below.
The geometry of the CM-3700A conforms to CIE No.15, ISO 7724/1, ASTM E 1164, DIN 5033 Teil 7, and JIS
Z 8722 condition c (diffused illumination/perpendicular viewing system) standards, and offers both di:8° (SCI:
specular component included ; Total reflectance) and de:8° (SCE: specular component excluded ; Diffuse
reflectance) measurements. (SCI/SCE switchable)
① Illumination
Light from xenon
lamps diffuse in the
integrating sphere
and illuminate
the specimen
uniformly.
① Light from the pulsed xenon lamps are diffused by
② Receiving
a. Light reflected
by the specimen
are received.
b. Light diffused in the
integrating sphere
are received.
③ Sensing
Light from the specimen-measuring and
illumination-monitoring optical fibers are
transmitted to sensors, where the light in
the wavelength range of 360 to 740 nm is
divided into 10 nm-pitch components and
projected onto the sensor array sections, which
convert the light intensity of each component
into proportional currents and output the
currents to the analog processing circuit.
reflection from the inner surface of the integrating
sphere, and finally illuminate the specimen uniformly.
② a. The light reflected by the specimen surface at an angle of 8° to the normal to the surface is received by
the specimen-measuring optical system and guided to the sensor.
b. The diffused light in the integrating chamber is received by the illumination-monitoring optical fiber and
guided to the sensor.
③ The light from the specimen-measuring optical fiber and from the illumination-monitoring optical fiber
is divided into each wavelength component and projected onto the sensor array sections, which
convert the light into proportional currents and output the currents to the analog processing circuit.
Memo
• By using the outputs from the specimen-measuring sensor and the illumination-monitoring
sensor for calculations, compensation for slight differences in the spectral characteristics
and intensity of the illumination light is performed (double-beam system).
Measuring Transmitted Colors
The flow of measurement is shown below.
The geometry conforms to CIE No. 15, ASTM E 1164, DIN 5033 Teil 7, and JIS Z 8722 condition g,
and offers di:0° (SCI: Total transmittance) and de:0° (SCE: Diffuse transmittance) measurements.
① Illumination
Light from xenon
lamps diffuse in the
integrating sphere
and illuminate
the specimen.
① Light from the pulsed xenon lamps are diffused by
reflection from the inner surface of the integrating sphere
② Receiving
a. Light passing through
the specimen
are received.
b. Light diffused in the
integrating sphere
are received.
③ Sensing
Light from the specimen-measuring and
illumination-monitoring optical fibers are
transmitted to sensors, where the light in
the wavelength range of 360 to 740 nm is
divided into 10 nm-pitch components and
projected onto the sensor array sections, which
convert the light intensity of each component
into proportional currents and output the
currents to the analog processing circuit.
and the surface of the white calibration plate covering
the reflectance measurement aperture, and finally illuminate the specimen in the transmittance chamber.
② a.
The light transmitted by the specimen is received by the specimen-measuring optical system and guided to the sensor.
b. The diffused light in the integrating chamber is received by the illumination-monitoring optical fiber and
guided to the sensor.
③ The light from the specimen-measuring optical fiber and from the illumination-monitoring optical
fiber is divided into each wavelength component and projected onto the sensor array section, which
convert the light into proportional currents and output the currents to the analog processing circuit.
Memo
• By using the outputs from the specimen-measuring sensor and the illumination-monitoring
sensor for calculations, compensation for slight differences in the spectral characteristics
and intensity of the illumination light is performed (double-beam system).
34
Page 37
Illumination and Measurement Areas
The CM-3700A allows you to select a target mask from three types: LAV (for ø25.4mm measurements),
MAV (for ø8mm measurements) and SAV (for 3×5mm measurements), according to the specimen and your
application. Select and attach a suitable target mask (illumination area) for each measurement area.
Target Mask
Each target mask is equipped with pins at the bottom of the target mask plate; the number and position of
these pins indicate to the Spectrophotometer which target mask is attached (and thus which illumination
area is being used). However, changing the target mask does not automatically change the measurement
area of the Spectrophotometer. Therefore, when changing the target mask, follow the direction of the color
conditioning software CM-S100w.
The target mask for SAV measurements is essentially the same as the one for MAV measurements, but
with a 0.3 mm-thick diffusion plate attached to increase the illumination area and reduce the edge-loss error
which occurs with semitransparent specimens.
The target mask surface which faces into the integrating sphere has a high-reflectance white coating. Since
the reflectance of this coating affects measurements, be careful to protect the white surface from scratches
and never touch the white surface.
Measurement Area
When the measurement area is switched from one to another, the condensing lens of the receiving optical
system is moved by the motor according to the commands fromthe software running on a personal computer.
System Configuration
Spectral sensor 2
(for monitoring illumination)
Xenon arc lamp
35
Lens system
for measurements
Spectral sensor 1 (for measurements)
Page 38
Fluorescent Measurement
The Spectrophotometer is equipped with a UV cutoff filter positioned in front of the pulsed xenon arc lamp.
The position of this filter is controlled by a motor and allows the amount of UV included in the illumination
to be controlled. When the pulsed xenon arc lamp is completely blocked by the UV cutoff filter, all light at
wavelengths below 390 nm is eliminated from the illumination. The position of the filter can also be adjusted
to provide illumination which closely matches CIE illuminant D65.
UV Adjustment
The position of the filter (and thus the amount of UV included in the illumination)
can be adjusted in 1000 steps, from 0.0 to 99.9% by software.
For the highest possible accuracy when measuring fluorescent specimens, fluorescent
calibration using a standard specimen for which standard values should be performed by
measuring the standard specimen repeatedly and the position of the UV filter adjusted so that
the measured values closely match the standard values. In addition, since the xenon lamp will
age and the light output by the lamp will change over time, it is recommended that fluorescent
calibration be performed periodically if fluorescent specimens will be measured.
WHEN FLUORESCENT CALIBRATION IS PERFORMED
When SpectraMagic™NX is used, the following four fluorescent calibration methods
are available to enable accurate measurement of fluorescent reflectance.
1. Tint mode
Correction coefficients for fluorescent measurement are obtained so that the measured
CIE Tint value for the fluorescent standard plate is within the specified range (Tint value
for the fluorescent standard plate calibrated with a D65 light source is entered).
2. Whiteness (WI) mode
Correction coefficients for fluorescent measurement are obtained so that the measured CIE Tint
and WI values for the fluorescent standard plate are both within the specified range (Tint and
WI values for the fluorescent standard plate calibrated with a D65 light source are entered).
WHEN FLUORESCENT CALIBRATION IS NOT PERFORMED
If extremely high accuracy is not required and fluorescent calibration will not be performed, it is
recommended that fluorescent specimens be measured with the UV filter completely removed
from in front of the xenon lamp, allowing 99.9% of UV to be included in the illumination.
Detector
Wavelength range
Wavelength pitch
Half bandwidth
Measuring range
Light source
Minimum
measurement
interval
Measurement/
illumination area
Repeatability
Inter-instrument
agreement
UV adjustment
Transmittance
chamber
Interface
Power
Operation
temperature/
humidity range (*1)
Storage
temperature/
humidity range
Size (W x H x D)
Weight
Reflectance
Transmittance
Silicon photodiode array with flat holographic grating
360 to 740 nm
10 nm
Approx. 14 nm average
0 to 200%; Resolution: 0.001%
Pulsed xenon arc lamp
3 seconds
Reflectance
Transmittance ø25 mm / Approx. ø20 mm
1. When white calibration plate is measured 30 times at 10-sec. intervals after white
calibration has been performed:
Spectral reflectance: Standard deviation within 0.05%
Chromaticity: Standard deviation within E*
2. When black tile (BCRA Series II; reflectance: 1%) is measured 30 times at 10-second
intervals after white calibration has been performed:
Spectral reflectance: 380 to 740 nm: Standard deviation within 0.02%
360 and 370 nm: Standard deviation within 0.04%
Chromaticity: Standard deviation within E*
Mean E*
Max E*
ab
color tiles compared to values measured with Konica Minolta master body.
Computer controlled: continuously variable
Maximum sample thickness: Approx. 50 mm
Maximum sample length: Unlimited (no sides when transmittance chamber cover is open)
Sample holder (optional) for holding sheet samples or containers
of liquid samples can be installed/removed
USB 1.1
AC 100 to 240 V 50/60 Hz 25 VA (using included AC adapter)
13 to 33°C, relative humidity 80% or less (at 33°C) with no condensation
0 to 40°C, relative humidity 80% or less (at 35°C) with no condensation
271 x 274 x 500 mm (10-11/16 x 10-3/4 x 19-11/16 in.)
18 kg (39.7 lb.)
switchable
Conforms to CIE No.15, ISO 7724/1, ASTM E 1164, DIN 5033 Teil 7
and JIS Z 8722 condition c standard.
di:0°, de:0° (diffuse illumination/0° viewing angle)
Conforms to CIE No.15, ASTM E 1164, DIN 5033 Teil 7 and
JIS Z 8722 condition g standard.
Changeable between SAV, MAV, and LAV
SAV : 3x5 mm measurement / 5x7 mm illumination
MAV : ø8 mm measurement / ø11 mm illumination
LAV : ø25.4 mm measurement / ø28 mm illumination
0.08 (typical) Average for 12 BCRA Series II color tiles.
ab
0.3 (corresponds to approx. E*
0.005
ab
0.05
ab
0.2) for any of 12 BCRA Series II
CMC
*1 Operating temperature/humidity range of products for North America : 13 to 33°C, relative humidity 80% or less (at 31°C)
with no condensation
The specifications and drawings given here are subject to change without prior notice.
38
Page 41
< CAUTION >
KONICA MINOLTA WILL NOT BE LIABLE FOR ANY DAMAGES RESULTING FROM THE
MISUSE, MISHANDLING, UNAUTHORIZED MODIFICATION, ETC. OF THIS PRODUCT,
OR FOR ANY INDIRECT OR INCIDENTAL DAMAGES (INCLUDING BUT NOT LIMITED
TO LOSS OF BUSINESS PROFITS, INTERRUPTION OF BUSINESS, ETC.) DUE TO THE
USE OF OR INABILITY TO USE THIS PRODUCT.