Before using this instrument,
please read this manual.
Page 2
Safety Symbols
The following symbols are used in this manual or CM-3600A to prevent accidents
which may occur as a result of incorrect use of the instrument.
Denotes an instruction regarding a safety warning or note.
Read the instruction carefully to ensure safe and correct use.
Denotes an instruction regarding the risk of electric shock.
Read the instruction carefully to ensure safe and correct use.
Denotes an instruction regarding the risk of fire.
Read the instruction carefully to ensure safe and correct use.
Denotes a prohibited action.
This action must never be performed.
Denotes an instruction.
This instruction must be strictly adhered to.
Denotes a prohibited action.
Never disassemble the instrument.
Denotes an instruction.
Be sure to disconnect the AC adapter from the AC outlet.
This symbol indicates A.C.
This symbol indicates D.C.
This symbol indicates class II protection against electric shock.
Trademarks
• Windows® is a registered trademark of Microsoft Corporation in the United States and other countries.
Notes on this Manual
• Copying or reproduction of all or part of the contents of this manual without KONICA MINOLTA’s permission is strictly prohibited.
• The contents of this manual are subject to change without prior notice.
• Every effort has been made in the preparation of this manual to ensure the accuracy of its contents.
However, should you have any questions or find any errors, please contact your retailer or a KONICA
MINOLTA-authorized service facility.
• KONICA MINOLTA will not accept any responsibility for consequences arising from the use of the
instrument.
Page 3
Safety Precautions
To ensure correct use of this instrument, read the following points carefully and adhere to them.
After you have read this manual, keep it in a safe place where it can be referred to anytime a question arises.
(Failure to adhere to the following points
WARNING
may result in death or serious injury.)
Do not use the instrument in places
where flammable or combustible
gases (gasoline, etc.) are present.
Doing so may cause a fire.
Always use the AC adapter supplied as
a standard accessory or the optional AC
adapter, and connect it to an AC outlet
of the rated voltage and frequency. If an
AC adapter other than those specified by
KONICA MINOLTA is used, it may result in
damage to the unit, fire or electric shock.
If the instrument will not be used for a
long time, disconnect the AC adapter
plug from the AC outlet. Accumulated
dirt or water on the prongs of the
AC adapter’s plug may cause a
fire and should be removed.
Do not disassemble or modify the
instrument or the AC adapter. Doing
so may cause a fire or electric shock.
Take special care not to allow liquid or
metal objects to enter the instrument.
Doing so may cause a fire or electric
shock. Should liquid or metal objects
enter the instrument, turn the power
OFF immediately, disconnect the
AC adapter plug from the AC outlet,
and contact the nearest Konica
Minolta-authorized service facility.
The instrument should not be operated if it
is damaged or the AC adapter is damaged,
or if smoke or odd smells occur. Doing
so may cause a fire. In such situations,
turn the power OFF immediately,
disconnect the AC adapter plug from the
AC outlet and contact the nearest Konica
Minolta-authorized service facility.
Do not forcibly bend, twist, or pull
the AC adapter power cable. Do not
scratch or alter the power cable or
place heavy objects on it. Doing so
may damage the power cable and
cause a fire or electric shock.
Insert the power plug fully and
securely. Incomplete inserting may
cause fire or electric shock.
(Failing to adhere to the following points may result in
CAUTION
Do not perform measurement with the
specimen measuring port
directed towards your eyes.
Doing so may damage your eyes.
Be careful not to get your hand caught in
the openable section of the instrument.
Doing so may result in injury.
injury or damage to the instrument or other property.)
Always hold the plug itself when
disconnecting the AC adapter plug
from the AC outlet. Pulling on the
power cable may damage it and
cause a fire or electric shock.
Do not insert or disconnect the AC adapter
plug from an AC outlet with wet hands.
Doing so may cause electric shock.
Do not place the instrument on an
unstable or sloping surface. Doing so
may result in its falling or overturning,
causing injury. Be careful not to drop
the instrument when carrying it.
Make sure that the AC outlet is located
near the instrument and that the AC
adapter plug can be connected to and
disconnected from the AC outlet easily.
When cleaning, disconnect the power plug.
Connecting may cause electric shock.
1
Page 4
INTRODUCTION
This spectrophotometer is designed for spectral measurement of color and color differences in
various industries. It can measure both reflected and transmitted color with high accuracy.
Packing Materials
⿎
General Packing Materials
Keep all packing materials (cardboard box, cushioning material, plastic bags, etc.) in a safe
place. The CM-3600A is a precision measuring instrument. They can be used to protect the
instrument from impact and vibration during shipment to Konica Minolta for maintenance.
Should they be lost or damaged, contact the nearest Konica Minolta-authorized service facility.
Protective Cap for Sample Holder
The CM-3600A is delivered with no target mask attached. To protect the specimen
measuring port, a protective cap is attached to the sample holder.
This protective cap must be removed before using the CM-3600A.
When you transport the CM-3600A to another place, the protective cap
must be attached. Keep the protective cap in a safe place.
Protective cap
2
Page 5
Notes on Use
⿎
Be sure to use this instrument properly. Use of this instrument in ways other than those specified in
this manual may result in risk of injury, electric shock, instrument damage, or other problems.
Operating Environment
• Use the CM-3600A at an ambient temperature between 13°C and 33°C and relative humidity 80% or less (at
33°C) with no condensation. Be sure to use the instrument within this range. Do not use it in areas of rapid
temperature changes.
• Do not leave the CM-3600A in direct sunlight or near sources of heat, such as stoves, etc. The internal
temperature of the instrument may become much higher than the ambient temperature in such cases.
• Do not use the CM-3600A in areas where dust, cigarette smoke or chemical gases are present. Doing so
may cause deterioration in performance or a breakdown.
• Do not use the CM-3600A near equipment which produces a strong magnetic field (such as speakers,
etc.).
• The CM-3600A belongs to installation category I products (equipment which is powered by an AC adapter
connected to commercially available power).
• The CM-3600A belongs to pollution degree 2 products (equipment which may cause temporary electrical
hazards due to contamination or condensation or products which are used in such an environment).
• Do not use the CM-3600A at altitudes higher than 2,000 m.
• The CM-3600A and the AC adapter supplied as a standard accessory have been designed exclusively for
indoor use. They should never be used outdoors because rain or other factors may damage the instrument.
Measurement
• Make sure no dirt or dust get into the specimen measuring port.
• When using the instrument for long periods of time, the measurement value may change depending on
changes in the environment. Therefore, in order to achieve accurate measurements, we recommend that
white calibration be done regularly using the White Calibration Plate.
White Calibration Plate
• The calibration data for the White Calibration Plate was measured at 23°C.
To achieve the highest accuracy when measuring absolute values (colorimetric values), calibration and
measurement should be performed at 23°C.
• Do not allow the White Calibration Plate to get scratched or stained with such as fingerprints.
• Do not move the White Calibration Plate while it has been caught by the sample holder. Doing so may
damage the White Calibration Plate.
• When the White Calibration Plate is not in use, be sure to close the cover so that the White Calibration
Plate is not exposed to ambient light.
Target Mask
• Do not touch the Target Mask's inner surface (black-coated surface) by hand, scratch it or make it dirty.
• When not in use, Target Masks should be stored in the accessory case (CM-A215) so that they will not be
exposed to ambient light.
Power Source
• Make sure that the power switch is set to OFF (" I ") when the CM-3600A is not in use.
• Always use the AC adapter supplied as a standard accessory (AC-A308) and connect it to an AC outlet of
the rated voltage and frequency.
• Use an AC power supply of the rated supply voltage (within ±10%).
• Do not connect the AC adapter to an overloaded electrical circuit. In addition, do not wrap or cover the AC
adapter with cloth or other material while in use. Doing so may cause an electric shock or fire.
3
Page 6
System
• Do not subject the CM-3600A to strong impact or vibration. Doing so may cause deterioration of performance or breakdown.
• The specimen measuring port and integrating sphere are extremely precise components, and great care
should be taken to prevent them getting dirty or exposing them to impacts. When the CM-3600A is not in
use, be sure to attach a target mask to the measuring port to prevent entry of foreign matter.
• The CM-3600A may cause interference if used near a television, radio, etc.
• Since the CM-3600A uses a microcomputer, external magnetic noise may cause malfunction. In this case,
turn the power OFF, and wait 30 minutes, and then turn it ON again.
About Transmittance Measurement
• Do not spill a specimen or other liquid on the instrument. If any liquid is spilled on the instrument, wipe it off
immediately with a soft, dry cloth.
4
Page 7
Notes on Storage
⿎
• The CM-3600A should be stored at temperatures between 0°C and 40°C, and at a relative humidity of 80%
or less (35°C) without condensation. Do not store the instrument in areas subject to high temperatures,
high humidity, sudden changes in temperature, or where freezing or condensation may occur, because
these circumstances may cause a breakdown. It is recommended to store the CM-3600A with a drying
agent at a temperature around 20°C.
• Do not leave the CM-3600A inside a car such as in the trunk. Otherwise, the temperature and/or humidity
may exceed the allowable range for storage during midsummer or midwinter, resulting in a breakdown.
• Keep the packing materials used for shipment and use them to transport the CM-3600A. This protects the
instrument from sudden changes in temperature, vibration, and shock.
• Do not store the CM-3600A in areas where dust, cigarette smoke or chemical gases are present. Doing so
may cause deterioration in performance or a breakdown.
• Entry of dust into the specimen measuring port will hinder accurate measurement. When the instrument
is not in use, you must close the transmittance chamber cover and cover the instrument with the supplied
Dust Cover to prevent the entry of dust into the integrating sphere.
• The White Calibration Plate may become discolored if left exposed to light. Therefore, make sure to close
the cover when it is not in use so that the White Calibration Plate is not exposed to ambient light during
storage.
• The Target Masks may discolor if they are left exposed to light. When they are not in use, keep them in a
safe place to prevent exposure to light and to protect them from scratches and dust. And store them in the
accessory case (CM-A215).
• Take care not to leave the CM-3600A for a long period of time with a target mask attached. The sample
holder may stick to the target mask.
• Be sure to keep all packing materials (cardboard box, cushioning material, plastic bags, etc.). They can be
used to protect the instrument during transportation to the service facility for maintenance (re-calibration
etc.).
Notes on Cleaning
⿎
• If the CM-3600A becomes dirty, wipe it with a soft, clean dry cloth. Never use solvents such as thinner or
benzene.
• If the White Calibration Plate becomes dirty, wipe it with a soft, clean dry cloth. If dirt is difficult to remove,
wipe it off with a cloth dampened with commercially-available lens cleaning solution.
Then remove the solution with a cloth dampened with water, and leave the plate to dry.
• If the inner surface (black-coated surface) of the Target Masks, or the inside of the integrating sphere, get
dirty, contact a Konica Minolta-authorized service facility.
• Should the CM-3600A break down, do not try to disassemble and repair it by yourself. Contact a Konica
Minolta-authorized service facility.
Disposal Method
⿎
• Make sure that the CM-3600A and its accessories and packing materials are either disposed of or recycled
correctly in accordance with local laws and regulations.
Used to perform white calibration for measurement
of reflectance and to perform measurement
of transmittance. A CD-ROM containing white
calibration data and software for writing the white
calibration data are supplied with this accessory.
Target Mask CM-A105
CM-A106
CM-A107
Used to change the illumination area (measurement
aperture) according to the specimen.
Measurement and illumination (aperture area at specimen
surface) areas for each target mask are as follows:
CM-A105(LAV): ø25.4 mm /ø30 mm
CM-A106(MAV): ø8 mm /ø11 mm
CM-A107(SAV): ø4 mm /ø7 mm
Zero Calibration Box CM-A104
Used to perform zero calibration for measurement
of reflectance and to perform haze measurement
at measurement of transmittance.
AC Adapter AC-A308
Used to supply power from an AC
outlet to the CM-3600A.
Input: 100 to 240 V 50/60 Hz
Output: 8 V 1.5 A
Plug design: Center-negative
USB Cable (3 m) IF-A21
Used to connect the instrument to
a personal computer (PC).
Accessory Case CM-A215
Dust Cover CM-A110
8
Page 11
Optional Accessories
Software
SpectraMagic™ NX CM-S100w
This software provides various functions (e.g., data
processing and file management) and allows the user
to operate the CM-3600A using a personal computer.
Transmittance Specimen Holder CM-A96
Used to hold the specimen for measurement
of transmittance. It can hold specimens
up to 22.5 mm (7/8 in.) thick.
① Viewfinder .............................Used to locate the position of the specimen for measurement of reflectance.
② Transmittance chamber .......Place the specimen in this chamber to perform
measurement of transmittance.
③ Target mask ..........................Select a suitable target mask (for ø30 mm (LAV), ø11 mm (MAV) and ø7 mm
(SAV) illumination) according to the specimen and attach it to the CM-3600A.
④ Sample holder .......................Used to hold the specimen, white calibration plate or zero calibration box.
⑤ Power switch .........................Used to turn power ON and OFF.
⑥
Transmittance chamber cover
⑦ Viewfinder cover ...................Open this cover to check the position of the
⑧
USB connection terminal
⑨
AC adapter input socket
...Covers the transmittance chamber used for transmittance measurements.
specimen for measurement of reflectance.
.........Used to connect the instrument to a PC with the supplied USB cable
(IF-A21 or IF-A22).
.........Used to connect the AC adapter supplied with the CM-3600A.
11
Page 14
Measurement Procedure
○This manual explains how to prepare the CM-3600A and how to set a specimen.
○The CM-3600A is controlled by a PC to perform measurements.
○For a description of measuring method using SpectraMagic™ NX
(optional), refer to the SpectraMagic™ NX instruction manual.
Flow of Preparation and Measurement
⿎
Connecting the PCConnect the CM-3600A to the PC with the USB cable.
(P. 13)
Connecting the AC adapterConnect the CM-3600A to the AC outlet
with the AC adapter. (P. 14)
Starting the PC Turn on the PC to be used to control
(starting Windows)the CM-3600A.
Starting the softwareStart the software and set it for control of the CM-3600A.
Turning the power ON Turn the power ON. (P. 15)
Attaching a target mask
Performing zero calibration
Performing white (100%)
calibration
<Reflectance measurement><
Attach the target mask
to be used. (P. 16)
Position the zero calibration
box and perform zero
calibration. (P. 17)
Position the white calibration
plate and perform white
calibration. (P. 18)
Transmittance measurement
Attach the LAV (Ø25.4 mm)
target mask and set the white
calibration plate in position. (P. 16)
Position the Transmittance
Zero Calibration Plate and
perform zero calibration. (P. 19)
Perform 100% transmittance
with nothing or a cell
containing distilled water in the
transmittance chamber. (P. 20)
>
Positioning a specimen Position the specimen on / in the CM-3600A. (P. 21)
Performing measurement Perform measurements.
Turning the power OFF When measurements are complete, turn the power OFF. (P. 15)
Exit out of the software and turn off the PC.
12
Page 15
Connecting a Personal Computer
Connect the instrument to a PC with the supplied USB cable IF-A21 (3 m).
Memo
• To connect the instrument with a PC, it is recommended that you use software that enables connection
and operation of the instrument (such as the optional Color Management Software SpectraMagic™ NX).
• The USB communication port of the instrument conforms to USB 1.1.
○To connect the instrument to a PC, you need to install the USB driver dedicated to the CM-3600A. Install
the USB driver supplied with the software that enables connection and operation of the instrument.
○The instrument is not designed to be powered via the USB cable. You need to connect the AC adapter.
○Make sure that the USB connector plug is oriented correctly and connected securely.
○When connecting/disconnecting the USB cable, be sure to hold the connector plug.
Do not pull on or forcibly bend the cable. Otherwise, wire breakage may result.
○Make sure that the cable has sufficient length. Putting tension on the
cable may cause connection failure or wire breakage.
○To connect the USB cable connector, check the shape of the receptacle
(connection terminal) and insert the connector fully until it is secured.
Operating Procedure
In general, a USB cable can be connected/disconnected while the instrument is turned ON; however,
you need to turn OFF the instrument in the procedure below. See P. 14 for how to connect the AC adapter
and P. 15 for how to switch the power on and off.
Turn OFF the instrument (Press the
1.
“ ” side of the Power switch.).
Connect the B connector of the USB
2.
cable to the USB connection terminal
(B type) of the instrument.
○Fully insert the connector and
ensure secure connection.
Connect the A connector of the USB
3.
cable to the USB port of the PC.
Connect the AC adapter and turn ON the
4.
instrument
○When you are prompted to install the USB
driver, specify the USB driver included
with the software or the white calibration
data CD and complete the installation.
○When using the optional Color Management
Software SpectraMagic™ NX, refer to the
SpectraMagic™ NX Installation Guide.
○After installation of the USB driver has
finished, switch the instrument off for a few
seconds and then switch it back on.
(Press the “ I ” side of the Power switch.)
.
13
Page 16
Connecting the AC Adapter
WARNING
Always use the AC adapter supplied as a standard accessory or specified replacement AC
adapter with the CM- 3600A, and connect it to an AC outlet of the rated voltage and frequency.
Failure to do so may damage the CM3600A or the AC adapter, causing a fire or electric shock.
If the CM-3600A will not be used for a long time, disconnect the AC
adapter from the AC outlet. Accumulated dirt or water on the prongs of the
AC adapter's plug may cause a fire and should be removed.
Do not insert or disconnect the AC adapter with wet hands. Doing so may cause electric shock.
Insert the power plug fully and securely. Incomplete inserting may cause fire or electric shock.
Do not disassemble or modify the AC adapter. Doing so may cause a fire or electric shock.
Do not unplug or plug in the AC adapter with the instrument’s power
switch set to ON. Doing so may cause malfunction.
Connecting Procedure
Make sure that the power switches of both
1.
CM-3600A and host PC are set to OFF (" ").
Insert the output plug of the AC
2.
adapter into the AC adapter input
socket on the rear of the CM-3600A.
Insert the input plug of the AC
3.
adapter into an AC wall outlet.
○The AC Adapter AC-A308 supplied as
the standard accessory must be used.
○Before disconnecting the AC adapter, the
power switch must be set to OFF ("
").
14
Page 17
Turning Power ON and OFF
WARNING
The CM-3600A should not be operated if the CM-3600A or the AC adapter is
damaged, or smoke or strange odors occur. Doing so may result in a fire. In such
situations, turn the power OFF immediately, disconnect the AC adapter from the
AC outlet, and contact the nearest Konica Minolta-authorized service facility.
Procedure
To turn the power ON and to light the lamp on
1.
the power switch, set the power switch to ON
(" I ") .
To turn the power OFF, set the power switch
2.
to OFF (" ").
15
Page 18
Attaching a Target Mask
WARNING
Do not place the CM-3600A on an unstable or sloping surface. Doing so may result in its
falling or overturning, causing injury. Take care not to drop the CM-3600A when carrying it.
Be careful around openings in the CM-3600A. Failure to do so
may result in fingers being trapped causing injury.
The CM-3600A allows you to select a target mask from the following three
types according to the specimen and your application.
hand, and attach it to the CM-3600A. Make
sure that it fits into the concave area
around the integrating sphere opening.
Memo
• When attaching a target mask, make sure that
the black-coated side faces the CM-3600A.
Release the sample holder to close it.
4.
Notes on Use of Target Mask
⿎
ake care not to scratch or make the inner surface (black-coated surface) of the target masks dirty with
○ T
such as fingerprints.
○The target masks may become discolored if left in areas exposed to light. Therefore, make sure that
target masks which are currently not in use are kept inside the accessory case (CM-A215) to prevent
exposure to light.
○When not using the CM-3600A, attach one of the target masks or the protective cap to prevent dust
entering the integrating sphere.
○Take care not to leave the CM-3600A for a long period of time with a target mask attached. The sample
holder may stick to the target mask.
16
Page 19
Attaching the Zero Calibration Box
WARNING
Do not perform measurement with the specimen measuring port directed
towards your face. Doing so may cause damage to your eyes .
Be careful around openings in the CM-3600A. Failure to do so
may result in fingers being trapped causing injury.
The zero calibration box is used to perform zero calibration for measurement of reflectance and to perform
haze measurement at measurement of transmittance.
○
Make sure that there is nothing in the transmittance chamber.
○
○
Before performing zero calibration, attach the target mask to be used for measurements.
○
○
Before performing zero calibration, set the same measurement area, specular component (SCI/
○
SCE), and UV light quantity as when performing zero calibration using the software. In the fluorescent
measurement that does not require rigorous accuracy (that does not perform fluorescent calibration),
perform measurement under the condition that the UV cut-off filter does not cover the xenon lamp (UV
light quantity is 99.9).
Procedure
Pull the sample holder toward
1.
you and keep it open.
○The sample holder will remain open
when opened more than 70 degrees.
Fit the projections of the zero calibration
2.
box into the grooves on the CM3600A and then close the sample
holder to hold the box in place.
Notes on Use of Zero Calibration Box
⿎
○Take care not to scratch, touch, or make the inside of the zero calibration box dirt y with such as
fingerprints.
○If the inside of the zero calibration box gets dirty, wipe it with a soft, clean, dry cloth.
○If dirt is difficult to remove, dampen a cloth with commercially available lens cleaning liquid and wipe the
zero calibration box. Then wipe off the liquid with a cloth dampened with water, and leave the box to dry.
○Should the inside of the zero calibration box get so dirty that it
cannot be cleaned, replace the box with a new one.
17
Page 20
Attaching the White Calibration Plate
CAUTION
Do not perform measurement with the specimen measuring port directed
towards your face. Doing so may cause damage to your eyes .
Be careful around openings in the CM-3600A. Failure to do so
may result in fingers being trapped causing injury.
The white calibration plate is used to perform white calibration for measurement of
reflectance and to cover the reflectance measuring port when performing transmittance
measurements (zero calibration, 100% calibration, measurement).
○
Make sure that there is nothing in the transmittance chamber.
○
○
Before performing white calibration, attach the target mask to be used for measurements.
○
○
Before performing white calibration, set the same measurement area, specular component (SCI/SCE),
○
and UV light quantity as when performing zero calibration using the software.
In the fluorescent measurement that does not require rigorous accuracy (that does not perform fluorescent
calibration), perform measurement under the condition that the UV cut-off filter does not cover the xenon lamp (UV
light quantity is 99.9).
Procedure
Pull the sample holder toward
1.
you and keep it open.
○The sample holder will remain open
when opened more than 70 degrees.
Position the white calibration plate so that it is
2.
held in place by the sample holder by fitting the
sample holder pad into the concave (rear) side
of the white calibration plate as shown at right.
Notes on Use of White Calibration Plate
⿎
Memo
• When performing white calibration, the white calibration data for the white calibration plate
being used is required. The white calibration data for the white calibration plate included
with the instrument is stored in the instrument’s memory at the time of shipment.
○
The white calibration plate may become discolored if left exposed to light. Therefore, when not in use,
○
make sure that the lid is closed to prevent exposure to light.
○
Take care not to scratch, touch, or make the white calibration plate surface dirty with such as fingerprints.
○
○
If the white calibration plate gets dirty, wipe it with a soft, clean, dry cloth.
○
○
If dirt is difficult to remove, dampen a cloth with commercially available lens cleaning liquid and wipe the white
○
calibration plate. Then wipe off the liquid with a cloth dampened with water, and leave the plate to dry.
○
Should the white calibration plate get so dirty that it cannot be cleaned, replace the plate with a new one.
○
Updating White Calibration Data
⿎
○
You may the “Data Setting Tool software” stored on the CD-ROM accompanying White Calibration Plate
○
CM-A139 or the optional Color Data Software SpectraMagic™ NX to set the white calibration data.
18
Page 21
Attaching the Transmittance Zero Calibration Plate
The Transmittance Zero Calibration Plate is used for performing zero calibration for transmittance measurements.
○
Before starting transmittance calibration and measurements, attach the LAV (ø25.4 mm) target mask and
○
position the white calibration plate at the reflectance measuring port.
○
Before starting transmittance calibration and
○
measurements, set the measurement area to LAV and
the specular component to SCI using the software.
○
Before starting transmittance calibration and
○
measurements, set the same UV light quantity as
when performing measurements using the software.
In the fluorescent measurement that does not
require rigorous accuracy (that does not perform
fluorescent calibration), perform measurement
under the condition that the UV cut-off filter does not
cover the xenon lamp (UV light quantity is 99.9).
Procedure
(Optional Accessory)
Open the cover of the transmittance chamber.
1.
Position the Transmittance Zero Calibration
2.
Plate in the transmittance chamber so that it
completely blocks the illumination window.
Memo
• When using the optional Transmittance Specimen
Holder, position the Transmittance Zero Calibration
Plate in the holder. For information on installing
the optional Transmittance Specimen Holder,
refer to the instructions included with it.
Close the cover of the transmittance chamber.
3.
Notes on Use of Transmittance Zero Calibration Plate
⿎
○
Take care not to scratch, touch, or make the surface of the transmittance zero calibration plate dirty with
○
such as fingerprints.
○
If the transmittance zero calibration plate gets dirty, wipe it with a soft, clean, dry cloth.
○
○
If dirt is difficult to remove, dampen a cloth with commercially available lens cleaning liquid and wipe the zero
○
calibration plate. Then wipe off the liquid with a cloth dampened with water, and leave the plate to dry.
○
Should the transmittance zero calibration plate get so dirty that it cannot be cleaned, replace the plate
○
with a new one.
Illuminationwindow
19
Page 22
Performing 100% calibration
Two methods for performing 100% calibration for
transmittance measurements are possible:
• 100% calibration to air:
When the specimen to be measured is in sheet or solid
form, 100% calibration should be performed with the
transmittance chamber empty.
• 100% calibration to water:
When the specimen to be measured is in liquid form
and will be measured using a cell (optional accessory),
100% calibration should be performed using distilled (or
pure) water in the same size and type cell as will be used for measurements.
○
Before starting transmittance calibration and measurements, attach the LAV (ø25.4 mm) target mask and
○
position the white calibration plate at the reflectance measuring port.
○
Before starting transmittance calibration and measurements, set the measurement area to LAV and the
○
specular component to SCI using the software.
○
Before starting transmittance calibration and measurements, set the same UV light quantity as when
○
performing measurements using the software.
In the fluorescent measurement that does not require rigorous accuracy (that does not perform fluorescent calibration),
perform measurement under the condition that the UV cut-off filter does not cover the xenon lamp (UV light quantity is 99.9).
After setting up the instrument as described below for your selected 100% calibration method, perform 100%
calibration from your software.
Setting up instrument for 100% calibration to air
⿎
Procedure
Open the cover of the transmittance chamber.
1.
Make sure that there is nothing (specimen,
2.
cell, Transmittance Zero Calibration Plate,
etc.) between the illumination window and the
receiving window in the transmittance chamber.
Memo
• It is no problem that the optional Transmittance
Specimen Holder CM-A96 is installed in the
transmittance chamber.
Close the cover of the transmittance chamber.
3.
Setting up instrument for 100% calibration to water
⿎
Procedure
Open the cover of the transmittance chamber.
1.
Fill the cell with distilled (or pure) water to
2.
approximately 2/3 (when using any of the cells
available as optional accessories) and position
the cell in the optional Transmittance Specimen
Holder installed in front of the illumination
window inside the transmittance chamber.
○
Although the Transmittance Specimen Holder can be installed in front of either the illumination
○
or receiving window, it is recommended that it is installed in front of the illumination window.
In this position, the illumination / observation geometry is diffuse illumination/0° viewing angle.
○
The cell used for calibration should be the same type and have the same optical path length as the
○
cell which will be used for measurements.
○
The cell should be positioned in the optional Transmittance Specimen Holder so that the clear sides
○
of the cell are facing the illumination and receiving windows.
○
The level of water in the cell should be above the top of the illumination and receiving windows when
○
the cell is positioned in the optional Transmittance Specimen Holder.
Close the cover of the transmittance chamber.
3.
20
Illumination window
Transmittance
chamber cover
Receiving window
Cell or Plastic Cell
Light
Page 23
Setting a Specimen
WARNING
Do not use the CM-3600A in places where flammable or combustible gases (gasoline fumes, etc.)
are present. Doing so may cause a fire.
Do not disassemble or modify the CM-3600A. Doing so may cause a fire or electric shock.
The CM-3600A should not be operated if it is damaged, or smoke or strange odors occur. Doing so
may result in a fire.
In such situations, turn the power OFF immediately, disconnect the AC adapter from
the AC outlet, and contact the nearest Konica Minolta-authorized service facility.
CAUTION
Do not perform measurement with the specimen measuring port directed
towards your face. Doing so may cause damage to your eyes.
Be careful around openings in the CM-3600A. Failure to do so
may result in fingers being trapped causing injury.
Reflectance measurement P. 22
To measure the reflectance of a film - or
plate-like specimen, the specimen needs
to be secured with the sample holder. If the
specimen cannot be secured in this way,
remove the sample holder and hold the
specimen against the measurement aperture.
Sample
The sample holder
can hold samples
with thicknesses
up to 30 mm.
When measuring specimens that are thicker than 30 mm
Removing the Sample Holder
⿎
Sample holder
Transmittance measurement P. 23
Transmittance chamber
Target Mask
(LAV) and White
Calibration Plate
Sample holder
Procedure
Use a Philips screwdriver to turn
1.
the two screws counter-clockwise
to remove the sample holder.
○
Keep the screws and sample holder in a safe place.
○
Sample holder
Two screws
21
Page 24
Reflectance measurement
⿎
○Make sure that there is nothing in the transmittance chamber.
○Before starting reflectance measurements, attach the target mask to be used for measurements.
○Before starting reflectance measurements, set the measurement area, specular component, and UV light
quantity using the software.
Procedure
Pull the sample holder toward
1.
you and keep it open.
○The sample holder will remain open when
it is opened more than 70 degrees.
Secure the specimen with the sample holder.
2.
Open the viewfinder cover and
3.
check the measuring point.
Memo
• When checking the measuring point while you are
seated, position the viewfinder cover at the desired
angle so that the image reflected in the mirror
inside the viewfinder cover can be observed .
• When the viewfinder cover is opened, the
specimen will be illuminated for 60 seconds by a
lamp to enable you to check the measuring point.
○Do not exert excessive force on the
viewfinder cover while it is open.
○It is not possible to perform measurement
if the viewfinder cover is open.
Adjust the position of the specimen
4.
so that the area to be measured is
centered in the measurement aperture
and then close the viewfinder cover.
○When adjusting the position of the specimen, the
sample holder must be pulled and kept open. This
will prevent the sample holder from scratching
the surface of the specimen during adjustment.
About opacity measurements
⿎
When using the optional Color Management Software SpectraMagic™ NX, pairs of white-backed and blackbacked measurements can be taken and opacity can be calculated.
22
Page 25
Transmittance measurements
⿎
○Before starting transmittance calibration and
measurements, attach the LAV (ø25.4 mm)
target mask and position the white calibration
plate at the reflectance measuring port.
○Before starting transmittance calibration and
measurements, set the measurement area to LAV and
the specular component to SCI using the software.
○Before starting transmittance calibration and
measurements, set the UV light quantity using the software.
Procedure
Open the cover of the transmittance chamber.
1.
Place the specimen against the illumination
2.
window tightly. (When measuring a liquid
specimen, the specimen container must be
placed against the illumination window.)
○The specimen must be placed so that
the area to be measured entirely covers
the entire illumination window.
Memo
• Although the specimen can be plaied in front of either the
illumination or receiving window, it is reccommended
that it is placed in front of the illumination window.
In this posit ion, the il lumi nati on/obse rvat ion
geometry is diffuse illumination/0° viewing angle.
In the position in front of the receiving window, the
illumination/observation geometry is approximately 0°
:0° . It does not strictly satisfy the difinition of sush as
CIE No.15. However, this presents no problem if the values are used as relative values.
• It is recommended that the optional Transmittance Specimen Holder CM-A96 be used to hold the specimen in place.
• In the case of a liquid specimen, use of an optional cell (CM-A97 to 99, A130 to 132) is recommended.
○If you are going to use a container other than the above cells, it must be transparent and constructed in such
a way that it can be positioned with opposing surfaces parallel to both illumination and receiving windows.
○Take care not to spill the liquid specimen on the CM-3600A. Should there be any spillage, immediately wipe
it with a soft, dry cloth.
○Never measure flammable liquids.
○No measurements must be performed if the specimen or inside and outside of the specimen container is
scratched or dirty. If you hold the measuring surface of the specimen with your fingers, it will be marked with
fingerprints, so always hold another part of the specimen when setting it in place.
○Air bubbles on the inner surface of the specimen's container will hinder correct measurements, so make
sure that all air bubbles are removed before measurement. (Air bubbles sometimes develop even if the
container is just left standing still.)
○Correct measurements will be hindered if water drops develop on the specimen or its container due to
condensation.
○Liquids containing particles may cause unstable measurements due to precipitation of the particles.
Illuminationwindow
Illuminationwindow
Receiving window
Receiving window
Close the cover of the transmittance chamber.
3.
About haze measurements
⿎
When the White Calibration Plate is positioned over the
reflectance measuring aperture, the measurement geometry
for transmittance measurements becomes di:0°. When the
Zero Calibration Box is positioned over the reflectance
measuring aper tu r e , the measure m e n t geometr y for
transmittance measurements becomes de:0°.
The illumination/observation system does not strictly satisfy the definition of haze (ASTM D 1003).
However, this presents no problem if the values are used as relative values.
When using the optional Color Management Software SpectraMagic™ NX, these two types of transmittance
measurements (di:0°, de:0°) with and without specimen can be taken and haze can be calculated.
23
Page 26
Cleaning the CM-3600A and Accessories
WARNING
Do not disassemble or modify the CM-3600A or AC adapter.
Doing so may cause a fire or electric shock.
The CM-3600A should not be operated if it is damaged, or if smoke or strange odors occur. Doing
so may result in a fire. In such situations, turn the power OFF immediately, disconnect the AC
adapter from the AC outlet, and contact the nearest Konica Minolta-authorized service facility.
CAUTION
When cleaning, disconnect the power plug. Connecting may cause electric shock.
Be careful around openings in the CM-3600A. Failure to do so
may result in fingers being trapped causing injury.
Zero Calibration Box and White Calibration Plate
⿎
Wipe gently with a dry soft cloth. If dirt is difficult to remove, dampen a cloth with commercially available lens
cleaning liquid and wipe. Then wipe off the liquid with a cloth dampened with water, and leave it to dry.
○When cleaning, take care not to scratch the zero calibration box or white calibration plate.
Target Mask
⿎
Use a blower to remove dirt and dust from the target masks.
○Do not touch the black-coated surface of the target masks with fingers or wipe it with a cloth. If the target
masks get so dirty that dirt cannot be removed using a blower, contact the nearest Konica Minolta-
authorized service facility.
Inside Integrating Sphere
⿎
Make sure that there is nothing placed
1.
against the illumination window.
Block the receiving window so
2.
that no dust or dirt enters.
Open the sample holder and remove
3.
dust and dirt using a blower.
○Do not touch the white-coated inner surface
of the integrating sphere, wipe it with a cloth
or place any object against it. If the whitecoated surface gets so dirty that dirt cannot be
removed using a blower, contact the nearest
Konica Minolta-authorized service facility.
Illumination window
Receiving window
Receiving Window
⿎
Set the measurement area to SAV using your
1.
software.
Memo
• This will cause the lens to move closer to the
receiving window, making it easier to clean.
Use a blower to remove dirt and dust from the
2.
receiving window.
○Do not put your fingers into the receiving
window or touch any optical lenses.
Receiving window
24
Page 27
Error Message
The following error messages may appear when you use SpectraMagic® NX CM-S100w, optional color
management software, to control the instrument. If you see an error message, follow the instructions shown
in the table below. If a problem persists, contact the nearest Konica Minolta-authorized service facility.
Error MessegeProblem / possible causeActionRefer to page
Time for Periodic
Calibration.Please contact
the closest Service Center.
File Open Error(Incorrect
File Extension).
Flash not ready.The illumination unit has
File open error.The calibration data file
Flash Error.The xenon lamp
Not Calibrated.Zero calibration and
Incorrect Calibration
Procedure.
A/D Error.A/D conversion failed.
EEPROM Error.The EEPROM may
Circuit is not operating
properly.
Not Ready.Charging for flashing the
It is time to perform
periodic calibration.
The file extension
is incorrect.
not been charged.
cannot be read properly.
failed to emit light.
• Xenon lamp dead
• Illumination circuit failure
• Sensor failure"
white calibration have
not been performed.
The calibration procedure
is not correct.
The count value is not
appropriate for zero (0%) or
while (100%) calibration.
• A/D converter failure
• Circuit failure"
be damaged.
The motor for changing
the measurement
diameter, adjusting the
UV light amount, or
switching between SCI
and SCE is not working.
xenon lamp is not complete.
Contact the nearest KONICA
MINOLTA-authorized
service facility and request
periodic calibration.
Select the file that has the
correct file extension.
Contact a KONICA
MINOLTA-authorized
service facility.
Load the original calibration
data file again. If the
problem persists, contact
the nearest Konica Minoltaauthorized service facility.
Retry measurement or
perform recalibration. If the
message persists, contact
the nearest Konica Minoltaauthorized service facility.
Perform zero recalibration
and white calibration.19, 20
Use the zero calibration
box for zero calibration
and the while calibration
plate for white calibration.
Turn the power off and then
on again. If the message
persists, contact the
nearest Konica Minoltaauthorized service facilit
Contact the nearest
KONICA MINOLTAauthorized service facility.
Turn the power off and then
on again. If the message
persists, contact the
nearest Konica Minoltaauthorized service facility
Wait for at least 3 seconds
after the last flash.
If the message persists,
contact the nearest
Konica Minolta-authorized
service facility.
SpectraMagic™
SpectraMagic™
—
NX (optional)
instruction
manual.
—
NX (optional)
instruction
manual.
—
17, 18
15
15
—
25
Page 28
Error MessegeProblem / possible causeActionRefer to page
Finder port is open.The viewfinder cover stays
open during measurement.
The measured object does
not exhibit fluorescence.
UV adjustment is not allowed
due to the UV specimen not
containing fluorescence.
The fluorescent calibration
coefficients could not
be determined.
The fluorescent
calibration coefficient
cannot be determined.
No response from instrument Communication with
the instrument failed.
• The power of the
instrument is off.
• COM port setting
is incorrect.
• Communication
settings are incorrect.
Detected instrument
The instrument is incorrect. Connect CM-3600A.
disagrees with connected
one.\nConnect again.
File is different to
current status.
The measurement
conditions in the fluorescent
calibration coefficient file
(.krd) designated to be read
are different from the actual
measurement conditions.
UV adjustment not
acceptable.
UV adjustment with a
Ganz&Griesser option has
not yet included all the
required measurements.
Close the viewfinder cover
before measurement.
Perform UV adjustment
using a specimen that
contains fluorescence.
Change Standard value
or tolerance settings,
or use a different
Fluorescent Standard;
then retry "UV adjustment".
Turn on the power of
the instrument.
Check and correct
the COM port and
communication settings.
Adapt the actual
measurement conditions
to the measurement
conditions in the .krd
file or select a file that
corresponds with the actual
measurement conditions.
Measure the fluorescent
sample four times when
selecting "Ganz&Griesser 4"
or five times when selecting
"Ganz&Griesser 5".
22
—
—
15
—
SpectraMagic™
NX (optional)
instruction
manual.
SpectraMagic™
NX (optional)
instruction
manual.
26
Page 29
TROUBLESHOOTING GUIDE
If a problem occurs with the Spectrophotometer, please check the following points before
requesting service. If the problem continues to occur even after the suggested corrective
actions have been taken, contact the nearest Konica Minolta-authorized service facility.
ConditionCheckpointRecommended action
Reflectance measurement
results seem strange.
Transmittance
measurement results
seem strange.
Data input/output between
the Spectrophotometer
and a computer cannot
be performed.
Commands cannot be input
to the Spectrophotometer
from a computer.
Was specimen
positioned correctly?
Are the white calibration
data correct?
Was white calibration
performed correctly?
Was zero calibration
performed correctly?
Is there any obstacle in the
transmittance chamber?
Was measurement mode
set to reflectance?
Was specimen
positioned correctly?
Was 100% calibration
performed correctly?
Was zero calibration
performed correctly?
Was measurement mode
set to transmittance?
Is the USB cable connected
correctly to both the
Spectrophotometer
and the computer?
Is the software
operating correctly?
Open the viewfinder cover and
check measurement point.
Set the correct white
calibration data.
Attach White Calibration
Plate correctly and perform
white calibration correctly.
Attach Zero Calibration
Plate correctly and perform
zero calibration correctly.
Make sure that there is nothing
in the transmittance chamber.
By referring to the operation
manual of the software, perform
the operation correctly.
Position specimen correctly.
Attach White Calibration Plate
correctly and perform 100%
calibration correctly using
the appropriate calibration
standard (air or distilled water).
Attach White Calibration
Plate correctly, block all light
from reaching the receptor
window, and perform zero
calibration correctly.
By referring to the operation
manual of the software, perform
the operation correctly.
Connect the USB cable correctly
bitween the Spectrophotometer
and the computer.
By referring to the operation
manual of the software, perform
the operation correctly.
Set POWER switch of
Spectrophotometer to OFF
and then set it back to ON.
Refer
to page
22
18
18
17
22
—
23
20
19
—
13
—
15
27
Page 30
28
Page 31
Explanations
29
Page 32
lIIumination/Observation System
Measuring Reflected Colors
⿎
The flow of measurement is shown below.
The geometry of the CM-3600A conforms to CIE No.15, ISO 7724/1, ASTM E 1164, DIN 5033 Teil 7, and JIS
Z 8722 condition c (diffused illumination/perpendicular viewing system) standards, and offers both di:8° (SCI:
specular component included ; Total reflectance) and de:8° (SCE: specular component excluded ; Diffuse
reflectance) measurements.
① Illumination
Light from xenon
lamps diffuse in the
integrating sphere
and illuminate the
specimen uniformly.
① Light from the pulsed xenon lamps are diffused by
reflection from the inner surface of the integrating
② Receiving
a. Light reflected
by the specimen
are received.
b. Light diffused in the
integrating sphere
are received.
③ Sensing
Light from the specimen-measuring and
illumination-monitoring optical fibers are
transmitted to sensors, where the light in the
wavelength range of 360 to 740 nm is divided
into 10 nm-pitch components and projected
onto the sensor array sections, which
convert the light intensity of each component
into proportional currents and output the
currents to the analog processing circuit.
sphere, and finally illuminate the specimen uniformly.
② a. The light reflected by the specimen surface at an angle of 8° to the normal to the surface
is received by the specimen-measuring optical system and guided to the sensor.
b. The diffused light in the integrating chamber is received by the illumination-
monitoring optical fiber and guided to the sensor.
③ The light from the specimen-measuring optical fiber and from the illumination-monitoring optical fiber
is divided into each wavelength component and projected onto the sensor array sections, which
convert the light into proportional currents and output the currents to the analog processing circuit.
Memo
• By using the outputs from the specimen-measuring sensor and the illumination-monitoring
sensor for calculations, compensation for slight differences in the spectral characteristics
and intensity of the illumination light is performed (double-beam system).
Measuring Transmitted Colors
⿎
The flow of measurement is shown below.
The geometry conforms to CIE No. 15, ASTM E1164, DIN 5033 Teil 7, and JIS Z 8722 condition g,
and offers di:0° (SCI: Total transmittance) and de:0° (SCE: Diffuse transmittance) measurements.
① Illumination
Light from xenon
lamps diffuse in the
integrating sphere
and illuminate
the specimen.
① Light from the pulsed xenon lamps are diffused by
reflection from the inner surface of the integrating sphere
② Receiving
a. Light passing through
the specimen
are received.
b. Light diffused in the
integrating sphere
are received.
③ Sensing
Light from the specimen-measuring and
illumination-monitoring optical fibers are
transmitted to sensors, where the light in the
wavelength range of 360 to 740 nm is divided
into 10 nm-pitch components and projected
onto the sensor array sections, which
convert the light intensity of each component
into proportional currents and output the
currents to the analog processing circuit.
and the surface of the white calibration plate covering the
reflectance measurement aperture, and finally illuminate the specimen in the transmittance chamber.
② a.
The light transmitted by the specimen is received by the specimen-
measuring optical system and guided to the sensor.
b. The diffused light in the integrating chamber is received by the illumination-
monitoring optical fiber and guided to the sensor.
③ The light from the specimen-measuring optical fiber and from the illumination-monitoring optical
fiber is divided into each wavelength component and projected onto the sensor array section, which
convert the light into proportional currents and output the currents to the analog processing circuit.
Memo
• By using the outputs from the specimen-measuring sensor and the illumination-monitoring
sensor for calculations, compensation for slight differences in the spectral characteristics
and intensity of the illumination light is performed (double-beam system).
30
Page 33
Illumination and Measurement Areas
The CM-3600A allows you to select a target mask from three types: LAV (for ø25.4mm measurements),
MAV (for ø8mm measurements) and SAV (for ø4mm measurements), according to the specimen and your
application. Select and attach a suitable target mask (illumination area) for each measurement area.
Target Mask
⿎
Since the CM-3600A has no target mask detection function, it is not possible for the instrument to
determine which target mask has been attached. Thus, when switching from one target mask to another,
be sure that the intended target mask is attached and the instrument is properly set from the software.
Furthermore, since the target masks are coated in black and measurement is influenced by the
condition of this coated surface, do not touch this surface with your hands, scratch it or make it dirty.
Measurement Area
⿎
When the measurement area is switched from one to another, the condensing lens of the receiving optical
system is moved by the motor according to the commands fromthe software running on a personal computer.
System Configuration
Spectral sensor
Light-receiving lens
Transmittance chamber
LAV
MAV
SAV
Illumination-monitoring
optical fiber
Viewfinder frosted glass
Viewfinder mirror
Viewfinder lens
Integrating sphere
Baffle
400 nm UV cut filter
Xenon flash lamp 3
Xenon flash lamp 1
Xenon flash lamp 4
420 nm UV cut filter
Sample holder
31
Xenon flash lamp 2
Page 34
Overview of Simultaneous Measurement of SCI/SCE
The CM-3600A offers simultaneous measurement of SCI (specular
component included)/SCE (specular component excluded).
With conventional models, SCI and SCE are switched mechanically by opening and closing an optical trap
provided on the integrating sphere. This conventional method requires mechanical switching whenever SCI
and SCE need to be switched. In addition, it is not possible to start measurement until switching is completed.
The CM-3600A has eliminated mechanical switching and enables acquisition of SCI and SCE
data by performing calculations with the measured data obtained using two light sources.
To viewing system
SCI/SCE SIMULTANEOUS
⿎
MEASUREMENT
○Light sources ① and ② are located as
illustrated at right. Light source ① flashes first.
・ Light source ① is a normal diffused
Memo
type, and enables the user to obtain SCI
measurement data when it flashes.
○Next, light source ② flashes.
・ Light source ② enables numerical control of
Memo
specularly reflected light. The data obtained when
this light source flashes (i.e., the amount of light
on the surface of the specimen) and the one
obtained when light source ① flashes can then
be used to calculate the SCE measured data.
2
Specimen
1
By performing the above measurement and calculation during each measurement, both SCI and SCE
measurement data can be obtained simultaneously without the need for mechanical switching.
32
Page 35
Fluorescent Measurement
The CM-3600A incorporates two types of xenon lamps as light sources for fluorescent measurements
(UV full light source and UV cut light source) and performs numeric calculation of the reflectance
when the specimen is illuminated by these two light sources to obtain fluorescent reflectance.
WHEN FLUORESCENT CALIBRATION IS PERFORMED:
⿎
When SpectraMagic™NX is used, the following four fluorescent calibration methods
are available to enable accurate measurement of fluorescent reflectance.
1. Profite mode
Correction coefficients for fluorescent measurement are obtained based on the calibrated reflectance profile of
the fluorescent standard plate (the reflectance profile is created by entering reflectance for each wavelength).
2. Tint mode
Correction coefficients for fluorescent measurement are obtained so that the measured
CIE Tint value for the fluorescent standard plate is within the specified range (Tint value
for the fluorescent standard plate calibrated with a D65 light source is entered) .
3. Whiteness (WI) mode
Correction coefficients for fluorescent measurement are obtained so that the measured CIE
WI (whiteness index) value for the fluorescent standard plate is within the specified range (WI
value for the fluorescent standard plate calibrated with a D65 light source is entered).
4. Tint and WI mode
Correction coefficients for fluorescent measurement are obtained so that the measured CIE Tint
and WI values for the fluorescent standard plate are both within the specified range (Tint and
WI values for the fluorescent standard plate calibrated with a D65 light source are entered).
CALCULATING FLUORESCENT REFLECTANCE
⿎
Based on the measured reflectances for the fluorescent standard plate obtained under the UV full light
source and under the UV cut light source, the amount of fluorescence is obtained. The correction coefficients
for fluorescent measurement are then determined so that the given reference values are satisfied.
(Calibration example: Profile mode)
Under UV full light
source
Under D65 light source
(entered reference value)
Under UV cut light
source
These correction coefficients are then used to obtain the fluorescent reflectance by performing numeric
calculations using the measured reflectances under the UV full light source and under the UV cut light source.
(The reflectances at cut wavelengths under UV cut light source are set to 0%.)
As a result, the CM-3600A can keep the fluorescent output quantity as close
to the standard as possible, without the need for adjusting the quantity of
ultraviolet light, which is required in the case of conventional models.
WHEN FLUORESCENT CALIBRATION IS NOT PERFORMED
⿎
Under UV
full light source
Under
D65 light
source
b a
Under UV cut light source
Correction coefficient=b/a
The UV full xenon lamp of the CM-3600A has a spectral distribution similar to that of the D65
light source. Therefore, when high-accuracy fluorescent measurements are not required,
normal measurement of reflectance can be used to measure fluorescent reflectance.
UV CUT LIGHT SOURCES
⿎
With the CM-3600A, UV400 cut light source (with radiation at wavelengths of 390 nm or lower eliminated)
and UV420 cut light source (with radiation at wavelengths of 410 nm or lower eliminated) are provided.
Light-receiving
element
Spectral separation
device
Wavelength range 360 to 740 nm
Wavelength pitch10 nm
HaIf bandwidthApprox.10 nm
Reflectance range 0 to 200%; resolution: 0.01%
Sphere sizeø152 mm
Light source4 pulsed xenon lamps
Minimum
interval between
measurements
Measurement/
illumination area
(Selectable)
Repeatability
Inter instrument
agreement
UV adjustmentInstantaneous numerical adjustment
UV cut filter400 nm cutoff and 420 nm cutoff
Transmittance
chamber
InterfaceUSB 1.1
PowerAC100 to 240 V 50/60 Hz (Using included AC adapter)
Operating
temperature/
humidity range (*1)
Storage
temperature/
humidity range
Size (WxHxD)244 x 205 x 378 mm
Weight11.5 kg
with simultaneous measurement of SCI (specular component included) / SCE
(specular component excluded) Conforms to CIE No.15, lSO 7724/1, ASTM
E 1164, DIN 5033 Teil 7 and JIS Z 8722 condition c standard. Transmittance
: di:0°, de:0° (diffused illumination, 0-degree viewing) Conforms to CIE No.15,
ASTM E 1164, DIN 5033 Teil 7 and JIS Z 8722 condition g standard.
Silicon photodiode array (dual 40 elements)
Diffraction grating
Normal SCI/ SCE measurement: 4 sec.
Transmittance measurement: 3 sec.
UV-cut/ UV-adjusted measurement: 5 sec.
LAV : ø25.4 mm/ ø30 mm
MAV : ø8 mm/ ø11 mm
SAV : ø4 mm/ ø7 mm
When white calibration plate is measured 30 times at 10-sec.
intervals after white calibration has been performed;
Spectral reflectance: Standard deviation within 0.1%
Colorimetric values: Standard deviation within E
Mean E
tiles compared to values measured with master body.
Width: 133 mm; depth: approx. 50 mm; measurement dia.: approx. 17 mm
Transmittance Specimen Holder (Optional accessory): Sample
holder for both plate-shaped and liquid samples (removable)
13 to 33°C, relative humidity 80% or less (at 33°C) with no condensation
0 to 40°C, relative humidity 80% or less (at 35°C) with no condensation
*
0.15 (LAV/SCI) Average for 12 BCRA Series II color
ab
*
ab
0.02
• The specifications given here are subject to change without prior notice.
35
Page 38
MEMO
36
Page 39
< CAUTION >
KONICA MINOLTA WILL NOT BE LIABLE FOR ANY DAMAGES RESULTING FROM THE
MISUSE, MISHANDLING, UNAUTHORIZED MODIFICATION, ETC. OF THIS PRODUCT,
OR FOR ANY INDIRECT OR INCIDENTAL DAMAGES (INCLUDING BUT NOT LIMITED
TO LOSS OF BUSINESS PROFITS, INTERRUPTION OF BUSINESS, ETC.) DUE TO THE
USE OF OR INABILITY TO USE THIS PRODUCT.