Compatible with new, next-generation light sources including PWM-controlled sources.
For simple but accurate illuminance measurements. Makes creating illuminance
measurement systems such as multi-point measurement systems easy!
<Mini receptor ><Standard receptor>
T-10A T-10MA / T-10WSA / T-10WLA
Receptor
diffuser
window:
Ø 25 mm
Main Features
Reliable, worry-free illuminance meters that
conform to JIS AA Class and DIN Class B
Illuminance Meters T-10A and T-10MA conform to Class AA of JIS
C 1609-1: 2006 "Illuminance meters Part 1: General measuring
instruments" and DIN 5032 Part 7 Class-B " Photometry;
classication of illuminance meters and luminance meters"
requirements to provide high-accuracy, high-reliability, worry-free
measurements.
Illuminance meters conforming to these standards are required for
measurements of general illumination light sources, white LED lamps
for illumination, etc. in a variety of industrial elds.
Removable receptor
The receptor and main body can be detached from each other and
then connected using a LAN cable, making it easy to install as part
of an inspection system.
Compatible with PWM-controlled lighting. Enables
measurements of next-generation light sources.
Conventional illuminance meters often cannot accurately
measure PWM-controlled light sources, but the T-10A series
of illuminance meters can be used to accurately measure
even such light sources.
Easy, inexpensive multi-point measurement
(2 to 30 points).
Illuminance distribution
of a projector etc. can
be easily measured with
a single instrument and
several receptors.
T-10A
Conforms to JIS AA Class
and DIN class B
Can be used for general
measurements of illuminance.
T-10MA
(Cord length: 1 m)
Conforms to JIS AA Class
and DIN class B
Enables illuminance
measurements of small areas.
Can be used for illuminance
measurements in narrow spaces
where the standard receptor won't t.
It can also be easily installed on
various kinds of equipment or jigs
for measuring light levels such as
illumination.
Receptor
diffuser
window:
Ø 14 mm
T-10WSA
T-10WLA
(Cord length: 5 m)
(Cord length: 10 m)
Conforms to JIS requirements
for special illuminance meters
Custom orderWaterproof
The mini receptor and cord are both
waterproof, so they can be used for
measurements in water.
They can be used for illuminance
control for shery-related applications
(such as sh farming, etc.) or for
measuring outdoor illuminance on
rainy days.
Multi-point illuminance measuring system
z
5-point example: Architectural lighting, etc.
z
9-point example: Projectors, etc.
●
[T-10A 9-point measuring system composition]
Illuminant Meter T-10A 1 unit
T-10A Receptor Head 8 units
Adapter units for Main Body T-A20 1 unit
Adapter units for Receptor Head T-A21 9 units
AC Adapter 1 unit
Data Management Software T-S10w 1 set
Main applications
z
Government testing organizations
z
Research/inspection at illumination
equipment makers
z
Maintenance at factories, ofces, hospitals,
etc.
z
25-point example: Street lighting, etc.
z
Illuminance control of security lighting,
street lighting, etc.
z
As sensor for equipment measuring lightdistribution characteristics, etc.
Illuminance Meter T-10A seriesIlluminance Meter T-10A series
23
Center of tripodsocket
Center of receptor window
7.5
Ø25
35
174
117.5
69
Ø45.2
T-10A
T-10MA/ T-10W
A/ T-10WLA
30
Reference
plane
Relative sensitivity (%)
Wavelength (nm)
Data Management Software T-S10w (Optional accessory)
Convenient, easy-to-use Excel® add-in software
Reads measurement data from T-10A series Illuminance
Meters directly into Excel
then be performed easily using the various functions of Excel
Main specifications of Data Management
⿎
®
. Further processing of data can
®
.
Software T-S10w
Typ eAdd-in for Excel
Operating
environment
Compatible
instruments
(Excel® is required to use this add-in.)
One of the following environments with Excel
* Languages in parenthesis ( ) are the OS language.
Windows
Chin ese)
Windows
Chin ese)
* For details on system requirements for above versions of
Windows
* Not compatible with 64 -bit versions of ofce 2010.
T-10A, T-10MA, T-10W
T-10W
®
®
installe d:
®
XP + Excel® 2003 (English, Japanese, or Simplied
®
7 + Excel® 2010 (English, Japanese, or Simplied
®
and/or Excel®, refer to their respective specications.
Ideally, the relative spectral responsivity of the illuminance
meter should match V (λ) of the human eye for photopic vision.
As shown in the graph above, the relative spectral responsivity
of Konica Minolta Illuminance Meters T-10A/10MA is within
6% (f1´) of the CIE spectral luminous efciency V (λ).
CIE ; Commission Internationale de I´Eclairage
f1´(CIE symbol) ; The degree to which the relative spectral
responsivity matches V (λ) is characterized by means of the error f1´.
0
%
2020
4040
6060
80 80
100
80%
60%
40%
20%
Since the brightness at the measurement plane is proportional
to the cosine of the angle at which the light is incident, the
response of the receptor must also be proportional to the
cosine of the incidence angle.
For Konica Minolta Illuminance Meters T-10A/10MA, the
cosine response f2 is within 3%.
The graph above shows the cosine correction characteristics
of Konica Minolta Illuminance Meters T-10A/10MA.
Ideal curve
Konica Minolta Illuminance
Meters
T-10A/10MA
System diagram
⿎
Data management using computer
Data Management software
T-S10w
(Including
T-A15 )
USB Cable
T-A15
AA-size batteries
2 pcs.
Strap
Personal computer
(Commercial product)
Data recording using printer
RS-232C printer
e
Cas
T-A10
Ca
p (with Strap)
T-A13
(for T-10A)
Mi
ni C
ap
T-A14
(for T-10M A)
Main Specications of T-10A
⿎
Model
Typ eMulti-function digital illuminance meter with detachable receptor head (Multi-point measurements of 2 to 30 points is possible)
Illuminance meter class
Receptor
Relative spectral response
Cosine response (f
Measuring rangeAuto range (5 manual ranges at the time of analog output)
Measuring function
Measuring
range
User calibration functionCCF (Color Correction Factor) setting function: Measurement value x 0.500 to 2.000
Linearity±2% ±1 digit of displayed value
Tem per ature /
humidity drif t
Computer interfaceUSB
Printer outputRS-232C
Analog output1 mV/digit, 3 V at maximum reading; Output impedance: 10 KΩ; 90% response time: 28 ms
Display3 or 4 Signicant-digit LCD with backlight illumination (Automatic illumination)
Power source
Battery life72 hours or longer (when alkaline batteries are used) in continuous measurement
Operating temperature
/humidity range
Storage temperature /
humidity range
Dimensions
Cord length
Weight (without battery) 200 g (7.0 oz.)205 g260 g (Receptor head only: 120 g)3 40 g (Receptor head only: 200 g)
*1 Conforms to requirements for Class AA of JIS C 1609-1: 2006 for all items except cosine response (f2).
*2 Although measurements below 1.00 lx are possible, they may not be stable due to the effects of electric al noise.
<Notes regarding mini receptors and waterproof mini receptors>
*Do not touch the cable during measur ements. Doing so may resul t in unstable measurement values.
*Secure the cable during measurements. Failure to do so may result in unstable measurement values.
Illuminance 0.01 to 299,900 lx; 0.001 to 29,990 fcd1.00 to 299,900 lx; 0.1 to 29.990 fcd *
Integrated
illuminance
(Commercial product)
Printer Cable
T-A12
External power
Hard Case
Hard Case
CL-A10
AC Adapter
Standard accessories
<
Illu mi nan c e M ete r T-10A
(Standard receptor head)
Conforms to requirements for Class AA of JIS C 1609-1: 2006 "Illuminance
meters Part 1: General measuring instruments" Conforms to DIN 5032 Part
7 Class B
Silicon photocell
Within 6% (f1´) of the CIE spectral luminous efciency V (λ)
)Within 3%Within 10%
2
Illuminance (lx). illuminance difference (lx). illuminance ratio (%). integrated illuminance (lx·h).
integration time (h). average illuminance (lx).
0.01 to 999,900 x 103 lx·h 0.001 to 99,990 x 103 fcd·h / 0.001 to 9999 h
Within ±3%
2 AA-size batteries / AC adapter AC-A308 (optional; for 1 to 10 receptors) or
AC adapter AC-A311 (optional; for 1 to 30 receptors)
-10 to 40°C, relative humidity 85% or less
(at 35°C) with no condensation
-20 to 55°C, relative humidity 85% or less
(at 35°C) with no condensation
69 x 174 x 35 mm
–1 m5 m10 m
><
CL-A10
Optional accessories
Illuminance Meter T-10MA
(Mini receptor head)
Main body: 69 x 161.5 x 30 mm
Receptor: Ø16.5 x 13.8 mm
Additional receptors (sold separately; Product
includes 1 receptor according to model purchased)
T-10MA Receptor Head
T-10W
S
A Receptor Head
T-10WLA Receptor Head
Custom order
T-10A
T-10MA
S
A
T-10W
T-10W
L
A
Custom models
Illu mi nan c e M ete r T-10W
(Waterpr oof mini receptor head)
Conforms to requirements for special Illuminance meters of JIS C 1609-1:
1
2006 *
5 to 40°C, relative humidity of 85% or less
(at 35°C) with no condensation
0 to 55°C, relative humidity of 85% or less
(at 35°C) with no condensation
>
T-10A
Receptor
Head
Mini plug
For multi-point and cable extension me asureme nt
Adapter Unit for
Receptor Head
Adapter Unit for
Main Body
T-A20
Example of cable
extension
Example of
multi-point System
Multi-point measurement requires use of
optional AC adapter.
A
S
With LAN category 5 cable; 1m
Illu mi nan c e M ete r T-10WLA
(Waterpr oof mini receptor head)
2
T-A21
LAN (10 BASE-T)
category 5 straight cable
(commercially
available)
About PWM-controlled lighting
PWM is the abbreviation of Pulse Width Modulation, and refers to
the method of controlling signal intensity by controlling the ratio
between the ON period and OFF period of a pulse signal.
A pulse signal is a signal which repeatedly alternates between ON
and OFF, and the percentage of ON period during a single cycle is
referred to as the "duty cycle".
PWM-controlled lighting is a method for controlling the brightness
Period when LED is lit
Dimensions
⿎
(Units: mm)
Center of
receptor window
174
117.5
Ø25
Reference
plane
Ø45.2
7.5
Ø16.5
Ø14
Center of
receptor
window
Reference
plane
13.6〜13.9
35
Center of
tripod
socket
161.5
Ø6.5
Ø12.7
Ø3
of a lamp by controlling the duty cycle (lit time) of light from a
pulse-emission source. As the lit time becomes longer, the light
becomes brighter, and conversely, as the lit time becomes shorter
the light becomes darker.
69
Illuminance Meter T-10A seriesIlluminance Meter T-10A series
45
T-10A
30
Cordlength:
T-10MA( 1 m)T-10WT-10W
S
SA ( 5 m)
LA (10 m)
PASS
FAIL
OK!
Cross-section A
Illuminance (lx)
Horizontal distance (m)
Height (m)
0011223
3
4
-1-2-3
300
100
50
10
30
5
y
illuminance Measurement Trio
Chroma Meter
CL-200A
De facto industry standard for measuring
color temperature!
0.90
520
530
0.80
540
510
550
0.70
0.60
500
0.50
0.40
4000
4500
D55
*
D65
*
*
D75
C
*
0.30
490
10000
0.20
480
0.10
470
460
380〜440
450
0.00
0.10 0.20 0.30 0.40 0.50 0.60 0.70
560
570
580
2500
2000
3000
590
3500
*
A
600
1500
B
*
610
680〜780
620
650
x
Measures color temperature
and Chromaticity!
Measures dominant
wavelength!
Even measures
excitation purity!
Application examples
Main Features
Compact and easy to carry
The CL-200A's compact body ts in your
palm. Battery-powered so it can be taken
along and used anywhere.
Data transfer using main body buttons
When using the CL-200A
with Data Management
Software CL-S10w
(included), measurements
can be taken and data
transferred to Excel
®
using
the main body buttons as
well as computer keys.
Detachable receptor head
The receptor head can be detached and then connected to
the main body using a normal LAN cable*, making it easy to
install the sensor in an inspection system.
Optional Adapter Units required for receptor head and main body
*
Can also measure illuminance
(JIS AA class)
Excel® add-in software included
Data Management Software CL-S10w
(Standard accessory)
Easy, convenient Excel® add-in
Measurement data from the CL-200A can be transferred
directly into Excel
managed freely within Excel
Includes LED ranking function
Color variations, the top topic in the LED industry, can be
quantied and a ranking function is also provided.
JIS correlated color temperature
Correlated color temperature is determined using the
equations dened by JIS (Japanese Industrial Standards).
Multi-point measurement and user
calibration also possible
Multi-point measurement management using up to 30
receptor heads is possible.
User calibration function enables compensation of
measurement values to match a desired standard.
Calibration can be performed by two methods: Singlepoint calibration or RGB calibration.
®
. The transferred data can then be
®
.
For lighting production
and adjustment
When using various types of light
sources in a room or open space,
it is sometimes necessary to check
the color of the lighting.
By using the CL-200A, it is
possible to adjust the lighting color
so that the food in a restaurant looks delicious.
For color-viewing
cabinet maintenance
A color-viewing cabinet like that
shown at left is used in industries
such as the printing industry to
visually evaluate nished work
under controlled conditions. This
color-viewing cabinet provides
illumination at a specic illuminance and color temperature by
using uorescent lamps, halogen lamps, etc. The CL-200A can
be used for the daily maintenance and control of these lamps as
well as to indicate when replacement is needed.
Multi-point illuminance measuring system
z
5-point example: Architectural lighting, etc.
z
9-point example: Projectors, etc.
For evaluating light
source characteristics
Evaluation of the light
distribution of LED illumination
modules or the illuminance
distribution of lighting xtures
can be evaluated.
For projector lightsource research and
color inspection
The CL-200A can be used to
measure the white balance and
uniformity of microprojectors,
etc. with internal LED light
sources. The ability to connect multiple receptors using LAN
cables enables measurement of not only a single point in
the center, but up to a maximum of 30 points over the entire
projected area.
For LED billboard
development and
maintenance
The CL-200A enables quality
control of the LED modules for
digital signage to be performed
easily. If modules with different
color tones are used together, the billboard will look mottled,
but by measuring the chromaticity and color temperature of
modules using the CL-200A and selecting modules based on
measured values, billboard uniformity can be achieved.
z
25-point example: Street lighting, etc.
Chroma Meter CL-200AChroma Meter CL-200A
67
PASS
FAIL
[CL-200A 9 point measuring system composition]
Chroma Meter CL-200A 1 unit
CL-200A Receptor Head 8 units
Adapter Unit for Main Body T-A20 1 unit
Adapter Unit for Receptor Head T-A21 9 units
AC Adapter 1 unit
Data Management Software CL-S10w 1 set