7.11 User Interface Cover113..................................................
Appendix A Abbreviations115................................................................
Appendix B Screen Summaries117...........................................................
TP-6883 4/21Table of Contents6
Safety Precautions and Instructions
IMPORT ANT SAFETY INSTRUCTIONS.
Electromechanicalequipment,
including generator sets, transfer
switches, switchgear, and accessories,
can cause bodily harm and pose
life-threateningdangerwhen
improperly installed,operated, or
maintained. To prevent accidents be
aware of potential dangers and act
safely.Read and follow all safety
precautions and instructions. SAVE
THESE INSTRUCTIONS.
This manual has several types of safety
precautions and instructions: Danger,
Warning, Caution, and Notice.
DANGER
DANGERindicatesahazardous
situation which, if not avoided, will
result in death or serious injury.
WARNING
WARNING indicates a hazardous
situation which, if not avoided, could
result in death or serious injury.
CAUTION
CAUTION indicatesa hazardous
situation which, if not avoided, could
result in minor or moderate injury.
NOTICE
NOTICE is used to address practices
not related to physical injury.
Safety decals affixed to the equipment
in prominent places alert the operator
or service technician to potential
hazards and explain how to act safely.
The decals are shown throughout this
publicationtoimproveoperator
recognition.Replace missing or
damaged decals.
Accidental Starting
WARNING
Accidental starting.
Can cause severe injury or death.
Disconnect the battery cables before
working onthe generatorset.
Remove the negative (- ) lead first
when disconnecting the battery.
Reconnect the negative (- ) lead last
when reconnecting the battery.
Disablingthegeneratorset.
Accidentalstartingcancause
severe injury or death.Before
working on the generator set or
equipment connected to the set,
disable the generator set as follows:
(1) Move the generator set master
switchtotheOFFposition.
(2) Disconnect the power to the battery
charger.(3) Remove the battery
cables,negative(- )leadfirst.
Reconnect the negative (- ) lead last
when reconnecting the battery. Follow
these precautions to prevent starting of
the generator set by an automatic
transfer switch, remotestart/stop
switch, or engine start command from a
remote computer.
Disablingthegeneratorset.
Accidentalstartingcancause
severe injury or death.Before
working on the generator set or
equipment connected to the set,
disable the generator set as follows:
(1) Press the generator set off/reset
button to shut down the generator set.
(2) Disconnect the power to the battery
charger, if equipped. (3) Remove the
battery cables, negative (- ) lead first.
Reconnect the negative (- ) lead last
when reconnecting the battery. Follow
these precautions to prevent the
starting of the generator set by the
remote start/stop switch.
Hazardous Voltage/
Moving Parts
DANGER
Hazardous voltage.
Will cause severe injury or death.
Disconnect all power sources before
opening the enclosure.
DANGER
Hazardous voltage.
Will cause severe injury or death.
Only authorized personnel should
open the enclosure.
Grounding electrical equipment.
Hazardous voltage will cause severe
injury or death.Electrocution is
possiblewheneverelectricityis
present. Ensure you comply with all
applicablecodesandstandards.
Electrically ground the generator set,
transfer switch, and related equipment
and electrical circuits. Turn off the main
circuit breakers of all power sources
before servicing the equipment. Never
contact electrical leads or appliances
when standing in water or on wet
ground because these conditions
increase the risk of electrocution.
Shortcircuits.Hazardous
voltage/current will cause severe
injury or death. Short circuits can
cause bodily injury and/or equipment
damage.Do not contact electrical
connections with tools or jewelry while
makingadjustmentsorrepairs.
Remove all jewelry before servicing the
equipment.
TP-6883 4/217Safety Precautions and Instructions
Servicingthetransferswitch.
Hazardous voltage will cause severe
injury or death. Deenergize all power
sources before servicing. Turn off the
main circuit breakers of all transfer
switch power sources and disable all
generator sets as follows: (1) Move all
generatorsetmastercontroller
switchesto the OFF position. (2)
Disconnectpower toall battery
chargers. (3) Disconnect all battery
cables,negative(- )leadsfirst.
Reconnect negative (- ) leads last when
reconnecting the battery cables after
servicing. Follow these precautions to
prevent the starting of generator sets
by an automatic transfer switch, remote
start/stop switch, or engine start
command from a remote computer.
Before servicing any components
inside the enclosure: (1) Remove all
jewelry. (2) Stand on a dry, approved
electrically insulated mat.(3) Test
circuits with a voltmeter to verify that
they are deenergized.
Makinglineorauxiliary
connections.Hazardous voltage
will cause severe injury or death. To
prevent electrical shock deenergize the
normal power source before making
any line or auxiliary connections.
deenergizedbyinlineconnector
separation.Test circuits with a
voltmeter to verify that they are
deenergized before servicing.
Testingliveelectricalcircuits.
Hazardous voltage or current will
cause severe injury or death. Have
trained and qualified personnel take
diagnosticmeasurementsoflive
circuits.Use adequately rated test
equipment with electrically insulated
probes and follow the instructions of the
test equipment manufacturer when
performing voltage tests. Observe the
following precautions when performing
voltage tests: (1) Remove all jewelry.
(2) Stand on a dry,approved electrically
insulated mat. (3) Do not touch the
enclosure or components inside the
enclosure.(4) Be prepared for the
system to operate automatically.
(600 volts and under)
Heavy Equipment
WARNING
Notice
NOTICE
Improper operator handle usage.
Use the manual operator handle on the
transferswitchformaintenance
purposes only .Return the transfer
switch to the normal position. Remove
the manual operator handle, if used,
and store it in the place provided on the
transfer switchwhenservice is
completed.
NOTICE
Electrostatic discharge damage.
Electrostaticdischarge(ESD)
damages electronic circuit boards.
Preventelectrostaticdischarge
damage by wearing an approved
grounding wrist strap when handling
electronic circuit boards or integrated
circuits. An approved grounding wrist
strap provides a high resistance (about
1 megohm), not a direct short,to
ground.
Servicingthetransferswitch
controls and accessories within the
enclosure. Hazardous voltage will
cause severe injury or death.
Disconnect the transfer switch controls
at the inline connector to deenergize
the circuit boards and logic circuitry but
allow the transfer switch to continue to
supply power to the load. Disconnect
all power sources to accessories that
are mounted within the enclosure but
are not wired through the controls and
Unbalanced weight.
Improper lifting can cause severe
injury or death and equipment
damage.
Use adequate lifting capacity.
Never leave the transfer switch
standing upright unless it is securely
bolted in place or stabilized.
TP-6883 4/218Safety Precautions and Instructions
Introduction
This manual provides operation instructions for the
Kohlerr Decision-Makerr MPAC 1500 automatic
transfer switch controller and related accessories.
The Decision-Makerr MPAC 1500 controller is available
for the transfer switch models shown below.
ModelDescription
KCSStandard-Transition Any Breaker ATS *
KCPProgrammed-Transition Any Breaker ATS *
KCC
KBS
KBP
KBC
KASStandard-Transition Electrically Operated
KAPProgrammed-Transition Electrically Operated
KGS
KGP
KEP
* Available with automatic or non-automatic controller
[ Available with automatic controller only
Closed-Transition Any Breaker ATS [
Standard-Transition Bypass/Isolation ATS [
Programmed-Transition Bypass/Isolation ATS [
Closed-Transition Bypass/Isolation ATS [
Bypass/Isolation ATS [
Bypass/Isolation ATS [
Standard-Transition Bypass/Isolation ATS [
Programmed-Transition Bypass/Isolation ATS [
Service Entrance ATS [
For Bypass/Isolation models, refer to the Installation
Manual for instructions to bypass and isolate the
transfer switch.See List of Related Materials for
document part numbers.
Information in this publication represents data available
at the time of print. Kohler Co. reserves the right to
change this literature and the products represented
without n otice and without any obligation or liability
whatsoever.
The equipment service requirements are very important
to safe and efficient operation. Inspect parts often and
perform required service at the prescribed intervals.
Obtain service from an authorized service distributor/
dealer to keep equipment in top condition.
Read this manual and carefully follow all procedures
and safety precautions to ensure proper equipment
operation and to avoid bodily injury. Read and follow the
Safety Precautions and Instructions section at the
beginning of this manual. Keep this manual with the
equipment for future reference.
List of Related Materials
A separate transfer switch installation manual provided
with the unit contains instructions for transfer switch
installation instructions, manual operation procedures,
and bypass/isolation instructions, if applicable.
Literature ItemPart Number
Specification Sheet, MPAC 1500 ControllerG11-128
Installation Manual, Model KCS/KCP/KCCTP-6833
Installation Manual, Model KBS/KBP/KBCTP-6835
Installation Manual, Model KGS/KGPTP-6836
Installation Manual, Model KEPTP-6946
Installation Manual, Model KAS/KAPTP-7190
Operation Manual, Modbus ProtocolTP-6113
TP-6883 4/219Introduction
Service Assistance
For professional advice on generator set power
requirements and conscientious service, please c ontact
your nearest Kohler distributor or dealer.
D Visit the Kohler Co. website at KOHLERPower.com.
D Look at the labels and decals on your Kohler product
or review the appropriate literature or documents
included with the product.
D Call toll free in the US and Canada 1-800-544-2444.
D Outside the US and Canada, call the nearest regional
Kohler Asia Pacific Headquarters
Singapore, Republic of Singapore
Phone: (65) 6264-6422
Fax:(65) 6264-6455
China
North China Regional Office, Beijing
Phone: (86) 10 6518 7950
(86) 10 6518 7951
(86) 10 6518 7952
Fax:(86) 10 6518 7955
East China Regional Office, Shanghai
Phone: (86) 21 6288 0500
Fax:(86) 21 6288 0550
India, Bangladesh, Sri Lanka
India Regional Office
Bangalore, India
Phone: (91) 80 3366208
(91) 80 3366231
Fax:(91) 80 3315972
Japan, Korea
North Asia Regional Office
Tokyo, Japan
Phone: (813) 3440-4515
Fax:(813) 3440-2727
TP-6883 4/2110 Service Assistance
Section 1 Operation
1.1Introduction
This section contains operation instructions, including:
D User interface panel, with display, pushbuttons, and
LED indicators
D Main menu
D System status, warnings, and faults
D Passwords
D Te s ts
D Warnings and Faults
D Reset Data
1.2Source Names
Throughout this manual, the sources are referred to as
follows. Source 1 (S1) is connected to the Normal side
of the transfer switch and is also referred to as Source N.
Source 2 (S2) is connected to the Emergency side of the
transfer switch and is also called Source E. The engine
start contacts are associated with Source 2.
1
2
ThePreferredSourceisthesourcethatwillbeusedif
both sources are available. Typically, this is the normal
utility source 1. If the transfer switch is equipped with the
optional alarm module, the Set Preferred Source menu
allows the operator to select either source as the
preferred source.Source 2 (connected to the
emergency side of the contactor) can be set as the
preferred source using this menu. See Section 4.8.3 for
more information about preferred source selection.
Other applications may use different configurations,
such as the gen-gen configuration which uses two
generator set sources and no utility.
1.3User Interface Panel
The user interface panel is located on the transfer switch
door. Figure 1-1 shows the user interface pushbuttons
and LED indicators.
34
5
9
1. Source N Available LED
2. Source N Position LED
3. Source E Position LED
4. Source E Available LED
5. USB port for SiteTecht
connection
Figure 1-1User Interface Panel
8
6. System Alert LED
7.NotinAutoLED
6
7
GM85884
8. Pushbuttons (4)
9. Display
TP-6883 4/2111Section 1 Operation
1.3.1Display
The four-line display indicates transfer switch status and
setup, including the following:
D System status
D Faults and warnings
D Active time delays
D Source voltages
D Source frequency (Hz)
D Current (amps)
D Source setup information
D Time and date
D Time and date of next scheduled exercise
The display also identifies the pushbutton functions,
which can change from menu-to-menu.
B
Down arrow (closed). Step down to the next
menu or scroll through a list.
Y
Up arrow (closed). Step back to the previous
menu.
"
Right arrow (closed). Move to the next
submenu.
Up arrow (open). Increases the selected
numerical value.
Down arrow (open). Decreases the selected
numerical value.
Right arrow (open). Steps to the next digit in a
selected numerical value.
BackSteps back to the previous menu or submenu.
1.3.2Display Contrast
To adjust the display contrast, press and hold the
second button until two rows of asterisks (*) appear.
Then press the up arrow button to increase the contrast
or the down arrow button to decrease the contrast. The
display will return to the main menu after a few seconds
if no buttons are pressed.
1.3.3Pushbuttons
The user interface panel has four pushbuttons below the
display. Pushbutton functions are shown above each
button in the last line of the display and can change from
menu-to-menu. The pushbutton functions are defined
in Figure 1-2.
Note: The current can be set to zero (0) by pressing the
first and last pushbuttons together. See section
4.16 for instructions.
End
Delay
End
OKEnters the displayed numerical value
Reset Reset the fault condition shown on the display,
SaveSaves settings shown on the display.
StartFrom the Test menu, starts the test sequence.
TestFrom the main menu, moves to the test
Ends the current time delay.
Ends an active test sequence. See
Tes t
Section 2.2.3.
(password or setting).
MainReturns to the main menu.
NextSteps to the next parameter in an item with
multiple settings (for example, in Exerciser
Setup).
or reset an accessory module after connection.
SetFrom the main menu, moves to the first
setup menu.
sequence menus. See Section 1.7.
ViewFrom the main menu, moves to the first
view menu.
Figure 1-2Pushbutton Functions
TP-6883 4/2112 Section 1 Operation
1.3.4LED Indicators
1.3.5Lamp Test
LEDs on the user interface indicate contactor position,
source availability, faults, and other conditions. The
table in Figure 1-3 describes the functions of the LED
indicators.
See Section 1.10 for more information about warnings
and faults.
Some programmable inputs will trigger the LEDs to light
or flash. See Section 4.9.
LED IndicatorCondition
Source N Available,
Green
Source E Available, RedSource E is available.
Position A, GreenContactor is in Normal position.
Position B, RedContactor is in Emergency position.
System Alert, RedFault. Identify and correct the cause
Not in Auto, RedATS is not set for automatic
Source N is available.
of the fault condition, then reset faults
at the controller. See Section 1.10.
Input active: Low Battery Voltage or
Remote Common Alarm. See
Section 4.9.
operation or a load shed (forced
transfer to OFF) sequence is active.
Flashes for manual transfer waiting.
Input active: Inhibit Transfer, Forced
Transfer to OFF. See Section 4.9.
Figure 1-3User Interface LED Indicators
To test the LEDs on the controller’s user interface, go to
the Main menu. Press the down arrow button once, then
press the Lamp Test button and verify that the LCD
menu and all 6 LEDs on the user interface illuminate.
SeeFigure1-4.
System Ready
LD Exer 12/14 @ 16:00
Norm 480V Emer 480V
Press the down arrow
button.
BView SetTest
Norm ABBCAC
##Hz ###V ###V ###V
Lamp
Press and hold the
Lamp Test button.
BYTe s tMa in
Figure 1-4Lamp Test
1.3.6Examples
Figure 1-5 illustrates navigation through the menus.
Figure 1-6 illustrates how to use the pushbuttons to step
through menus and change settings. This example
shows setting the time.
Set Common Alarms
BY"Main
Press the right arrow
button to enter the
Set Common Alarms
menu.
Set Common Alarms
Alarm Group 1
BY "Back
Press the up and down
arrow buttons to step
between alarm groups 1
and 2.
Press the right arrow to
modify settings in the
displayed group.
Figure 1-5Example: Set Common Alarms
Modify Alarm
BY "Back
Press the up and
down arrow buttons
to select Modify
Alarm or Remove
All Alarms
Remove All Alarms
BY "Back
Alarm Description
CommonAudible
BY "Back
Press the up and down
arrow buttons to step
through the alarm
descriptions. Then press
the right arrow to change
the common and audible
assignments for that
function.
Remove All Alarms
Yes/N o
BackSave
Press the up arrow button to toggle yes or no.
If Yes is displayed, pressing Save will remove
all alarms from the selected group (1 or 2).
Pressing Back exits without executing the
Remove command.
Alarm Description
Common Y Audible N
BackSave
Press the up arrow
button to step through
all combinations of yes
and no for common
and audible.
Press Save to save the
common and audible
alarm settings.
TP-6883 4/2113Section 1 Operation
1
System OK
LD Exer ##/## @ ##:##
Norm ###V Emer ###V
BViewSetTest
6
Set Time
BY "Back
Press the SET button.
2
Enter Password
Time to Enter #:##
???
?
OK
Press the open UP arrow button to
increase the first digit of the password
from 0 through 9.
Note: The default password is 0000.
3
Press the open RIGHT arrow
button to move to the next
digit. Repeat Steps 2 and 3
until the correct password is
displayed.
4
Enter Password
Time to Enter #:##
Time to Enter #:##
??
0?
Enter Password
0
000
OK
OK
Press the right arrow
button to enter the Set
Time submenu.
7
Set Time #
BackSave
Press the open up arrow
button to increase the first
digit. (Use 00:00 to 23:59 for
time.)
8
Set Time 1#:##
BackSave
Press the open right arrow
button to step to the next digit.
Repeat steps 7 and 8 until the
correct time is displayed.
9
Set Time 14:37
#:##
Press OK to enter the password.
5
Set Time/Date
BY "Main
Press the right arrow button
to enter the Set Time/Date
submenu.
Figure 1-6Example: Setting the Time
BackSave
Press Save.
TP-6883 4/2114 Section 1 Operation
1.4Main Menu
The main menu appears at system startup and displays
the following information(seeFigure1-7):
D System Status (see F igure 1-8)
Press the Test button to enter the Test mode.A
password is required. See Section 1.7.
The display returns to the main menu after 10 minutes of
no activity (no buttons pressed).
D Date and time of the next scheduled exercise run (if
programmed)
D Measured source voltages
D Pushbutton functions
Pressing the down arrow button steps to the normal
operation menus shown in Section 1.6. Step through
these menus to check the measured frequency, line
voltages, current (if the current sensing accessory is
installed), and other system information.
Some parameters will appear only under certain
conditions. For example:
D If no exercise runs are scheduled, the second line of
themainmenuisblank.
D The Daylight Saving Time settings are displayed only
if DST is enabled.
D Phaserotationandin-phasemonitoringare
displayed only for three-phase systems.
D Some parameters and time delays appear only for
programmed-transition models.
D The menus displayed during a test or exercise
sequence will vary depending on the time delay
settings.
Pressing the View button steps to the view menus
shown in Section 3.3.
Passwords are required to enter the setup and test
modes. See Section 1.5 for more information about
passwords.
System Ready
LD Exer 12/14 @ 16:00
Norm 480V Emer 480V
BViewSetTest
1. System status message
2. Next scheduled exercise, if programmed
3. Source voltages detected
Figure 1-7Main Menu
System Status Messages
Aux Switch Fault
Aux Switch Open
Exerciser Active
External Battery Low
Fail to Acquire Pref
Fail to Acquire Stby
Fail to Transfer
In Phase Waiting
Inhibit Transfer
Low Battery Voltage
Maint DIP Switch
Module Lost Comm
New Module
Peak Shave Active
Phase Rotation Error
Remote Common Alarm
System Ready
Test Mode Active
Figure 1-8System Status Messages
1
2
3
6446
Press the Set button to enter the setup mode.A
password is required. See Section 4.2 for system setup
menus.
TP-6883 4/2115Section 1 Operation
1.5Passwords
Passwords are required to enter the Test and Setup
menus.Passwords are 4-digit numbers.See
Figure 1-9 for instructions to enter the password using
the pushbuttons on the controller’s user interface.
There a re two passwords:
Setup Password. The setup password controls access
to the system setup menus, which allow changes to
system settings, time delays, etc.
For closed-transition models, the setup password is
required to initiate a transfer when the programmed
transition override function is set to manual.See
Section 1.7.3.
Note: The factory default password is 0000.
Enter Password
Time to Enter #:##
???<
>?
OK
Press the open up arrow button to increase the first digit
of the password from 0 to 9.
Enter Password
Time to Enter #:##
>0???<
OK
Test Password. The test password controls access to
the test sequence menus.The test password is
required to initiate a loaded, unloaded, or auto-loaded
test, and also to initiate a sync check test on
closed-transition models.
If the correct password is not entered within 30 seconds,
the display returns to the main menu.
The factory default password is 0000. Change the
password to allow only authorized personnel to start and
end tests or change settings.
1.5.1Changing Passwords
Use the Passwords Setup Menu to change passwords.
See Section 4.14.
1.5.2Test Password Reset and Disable
The test password can be reset to the default value or
disabled. Use the Setup Menu- Reset Data menu. See
Figure 1 -26.
Note: Disable the test password only during service
unless the transfer switch is located in a secure
location.
Press the open right arrow button to step to the next digit.
Repeat for all four digits.
Enter Password
Time to Enter #:##
>0000
Press the OK button to enter the password.
Incorrect Password
If the wrong password is entered, the Incorrect Password
message appears. Check the password and try again.
<
OK
Figure 1-9EnteringaPassword
Disabling the test password allows any user to initiate a
test sequence from the controller’s user interface
without entering a password. Initiating a t est starts the
generator set and, if a loaded test is selected, transfers
the load.
TP-6883 4/2116 Section 1 Operation
1.6Normal Operation Menus
During normal transfer switch operation, the screens
shown in Figure 1-10 or Figure 1-11 are displayed. Use
the up a nd down arrow buttons to view the sy stem status
information as shown. Press Main to return to the main
menu from any screen.
The Sequence of Operation descriptions in Sections 2.2
through 2.4 describe the transfer switch normal
operation for standard, programmed, and closed
transition models.
System Ready
LD Exer 12/14 @ 16:00
Norm 480V Emer 480V
BViewSetTest
System Ready
LD Exer 12/14 @ 16:00
Norm 480V Emer 480V
BViewSetTest
Norm ABBCAC
##Hz ###V ###V ###V
Lamp
BYTe stM ai n
Norm ANBNCN
##Hz ###V ###V ###V
BYMain
Emer ABBCAC
##Hz ###V ###V ###V
NormL1L2
##Hz###V###V
Lamp
BYTe s tMa i n
EmerL1L2
##Hz###V###V
BYMain
AmpsL1L2
########
BYMain
Time/Date
09:10 DST 01/13/06
BYMain
Daylight Saving Time
Clock Ahead 1 Hour
Sun03/12/06
BYMain
Daylight Saving Time
ClockBack1Hour
Sun11/05/06
BYMain
Normal Preferred
Util-Gen Operation
No Commit Transfer
BYMain
Standard Transition
BYMain
Emer ANBNCN
##Hz ###V ###V ###V
BYMain
Amps LALBLC
#### ########
BYMain
Time/Date
09:10 DST 01/13/06
BYMain
Daylight Saving Time
Clock Ahead 1 Hour
Sun03/12/06
BYMain
Daylight Saving Time
Clock Back 1 Hour
Sun11/05/06
BYMain
Normal Preferred
Util-Gen Operation
No Commit Transfer
BYMain
Standard Transition
Phase Rotation ABC
In Phase Enabled
BYMain
Figure 1-11Three-Phase Operation
BYMain
Figure 1-10Single-Phase Operation
TP-6883 4/2117Section 1 Operation
1.7System Test
Use the system test feature to:
D Start and run the generator set.
D Simulate a preferred source failure, resulting in a
transfer to the standby source.
D Check source synchronization (closed-transition
models only).
time delays. See Figure 1-16 for Sync Check menus for
closed-transition models.
Press the End Test designated pushbutton to end the
test. Time delays will execute as programmed when the
test is ended. Press the End Delay button to end the
currently displayed time delay, if desired.
To check the source voltage and frequency while a test
is running, press the Main button. Press the Test button
to return to the test sequence menus.
See Figure 1-12 for the test sequence menus. From the
main menu, press the Test button and then enter the
password. The password ensures that only authorized
personnel can start a test.
Press the down arrow button to navigate to the desired
test sequence. Press the Start button to start the test.
Figure 1-13 shows the menus displayed during the test
run. Menus are dependent on the system settings and
System Ready
LD Exer ##/## @ ##:##
Norm ###V Emer ###V
BViewSetTest
Enter Password
Time to Enter #:##
>?
???<
OK
Press the Test button to enter the
Test mode. A password is required.
If the emergency source is lost during a system test, the
fail to acquire standby signal is indicated immediately,
and the test is terminated. If the contactor is in the
standby position, it transfers immediately to the
preferred position.
See Section 4.9.3 for additional information about the
remote test input.
Type of Test
Auto Load
Time :## min
BMain"Start
Type of Test
Loaded
BMainStart
Type of Test
Unloaded
BMainStart
Auto Load Test
Enter Time
Time ?? min
BackOK
Figure 1-12Test Selection Menus
Type of Test
Sync Check
BMainStart
TP-6883 4/2118 Section 1 Operation
Engine Start in ##:##
Norm ###V Emer ###V
EndEnd
MainDelay Test
LD# Disc in ##:##
Norm ###V Emer ###V
EndEnd
MainDelay Test
Xfr to Off in ##:##
Norm ###V Emer ###V
EndEnd
MainDelay Test
XfrtoEmerin##:##
Norm ###V Emer ###V
EndEnd
MainDelay Test
Add LD# in ##:##
Norm ###V Emer ###V
EndEnd
MainDelay Test
System on Test
Norm ###V Emer ###V
EndEnd
MainDelay Test
LD# Disc in ##:##
Norm ###V Emer ###V
EndEnd
MainDelay Test
Xfr to Off in ##:##
Norm ###V Emer ###V
EndEnd
MainDelay Test
XfrtoNormin##:##
Norm ###V Emer ###V
EndEnd
MainDelay Test
Add LD# in ##:##
Norm ###V Emer ###V
EndEnd
MainDelay Test
Appears if load control
time delays are set
Programmed-transition
models only
Appears if load control
time delays are set
Appears if load control
time delays are set
Programmed-transition
models only
Appears if load control
time delays are set
1.7.1Unloaded System Test
When an unloaded test is initiated, the controller
immediately signals the generator to start, without
waiting for the engine start time delay to expire. The
contactor does not change position during an unloaded
test, but if the normal source should fail, the contactor
will transfer to the emergency source.
The unloaded test feature is available only with the
Util-Genset and Genset-Genset modes of operation.
The load bank control output is active during an
unloaded exercise or unloaded system test.If the
contactor transfers to the standby position during the
test, the load bank control is deactivated. (The standby
source supplies power to the load.)
1.7.2Loaded System Test
A loaded test simulates a preferred source failure,
except that the engine start time delay is bypassed. The
generator set is signaled to start immediately upon test
activation. Load control signals are issued prior to
transfer with their associated time delays. Since the
loaded test transfer will be between two live sources, the
in-phase monitor or closed transition feature will be
activated if it is enabled. If the preferred source is lost
during a loaded test with the contactor in the standby
position, the test will continue to be active, even on
restoration of preferred. If the standby source is lost and
the preferred source is available, the test will be
terminated, and the transfer switch will immediately
transfer to the preferred source position, bypassing all
time delays except the off-position requirements in a
programmed-transition system.
When a loaded test is terminated normally, the
retransfer sequence operates as though the preferred
source has been restored after a failure. All time delays
are executed and an in-phase transfer will occur if
enabled. The loaded test feature is available with the
Util-Genset, Util-Util and Genset-Genset modes of
operation.
1.7.3Closed-Transition Loaded Test
Eng Cooldown ##:##
Norm ###V Emer ###V
EndEnd
MainDelay Test
When a loaded test is initiated on a closed-transition
model, the generator set is signaled to start and the
controller monitors the sources for synchronization.
The load is transferred when the sources are
synchronized.
Note: See Figure 1-16 for Sync Check menus.
Figure 1-13Test Sequence Menus
TP-6883 4/2119Section 1 Operation
If the sources do not sync before the Fail to Sync time
delay expires, the programmed-transition override
function operates.
D If the override function is set to Automatic, a
programmed-transition transfer will occur when the
Fail to Sync time delay expires. The contactor stops
in the OFF position for the length of the off-to standby
time delay before proceeding to transfer to the
standby source.
D If the override function is set to manual, the user can
either initiate a programmed-transition type transfer
(setup password required) or cancel the test
sequence. See Figure 1-14. If neither action is taken,
thecontrollerwillcontinuetocheckfor
synchronizationandtransferifthesources
synchronize.
See Section 4.11.3 for instructions to set the
programmed-transition override function.
Manual Transfer
Password:
CancelOK
2
1. Use arrow buttons to enter the setup password and click OK
to initiate a manual programmed-transition transfer. OR
2. Press the Cancel button to cancel the test.
1
Figure 1-14Manual Transfer Menu for
Programmed-Transition Override
1.7.5Sync Check (closed-transition)
The Sync Check allows a test of the synchronization of
two available sources without initiating a transfer.
Navigate to the Type of Test, Sync Check menu and
press the Start button to begin the test. The controller
displays Syncing during the test, and the phase angle
difference is shown between two arrows. For example,
> 10 < indicates that the sources are 10 degrees out of
phase. The arrows move closer together as the sources
approachsynchronization.Whenthesources
synchronize, the controller indicates Synced and
continues to monitor the source synchronization. The
load is not transferred. See Figure 1-16. Press the End
Te st button to end the test.
Type of Test
Sync Check
BMainStart
System on Test
Norm ###V Emer ###V
Syncing
End
Main>102< Test
System on Test
Norm ###V Emer ###V
Synced
End
Main><Test
Figure 1-16Sync Check Menus
1.7.4Auto-Loaded System Test
The auto-loaded test feature is a timed, loaded test. The
auto-loaded time delay determines how long after the
transfer to standby to terminate the test and transfer
back to the preferred source. The time is defaulted to
30 minutes and can be adjusted from 1 minute to
60 minutes. See Figure 1-15.
Type of Test
Auto Loaded
Time :## min
BMain"Start
1
1. Press right arrow button to go to the Enter Time menu.
2. Use the open arrow buttons to enter the duration time for the
test. Then press OK.
3. Press Start to start the test.
32
Figure 1-15Auto Loaded Test Menus
Auto Loaded Test
Enter Time
Time: ?? min
Back OK
1.8Automatic Operation Test
Note: Close and lock the enclosure door before starting
the test procedure.
Preferred Source Selection.The test procedure
assumes that Source N is the preferred source. If the
ATS is equipped with the alarm board accessory, check
the preferred source selection before proceeding with
the automatic operation test. To check the preferred
source selection, use the down arrow button to step
down from the main s creen until Normal Preferred or
Emergency Preferred is displayed. See Figure 1-10 or
Figure 1-11.
Supervised Transfer Switch. If the transfer switch is
equipped with a supervised transfer switch, verify that it
is set to the Auto position.
Follow the procedure below to start a loaded test. Verify
that the ATS starts the generator set and transfers the
load to the emergency source, executing all time delays
that are set up to operate during a loss of the normal
source. End the test and verify that the transfer switch
transfers the load back to the normal source and
removes the enginestart signal, executing all
TP-6883 4/2120 Section 1 Operation
appropriate programmed time delays. Refer to Section
2.2.3 for a more detailed description of the test
sequence of operation.
Load control time delay settings may affect the
operation sequences.
Note: If the standby source fails during a loaded test,
1. Check the controller LED indicators to verify that
the Position N and Source N Available indicators
are lit.
If the sources do not synchronize before the fail
to sync time delay expires, operation depends
on the programmed transition override setting.
If automatic override is enabled, the ATS will
transfertheloadusingaprogrammedtransition transfer. If automatic override is not
enabled, the ATS will continue to monitor the
source synchronization and transfer when/if
the sources synchronize. The operator can
initiateaprogrammed-transitiontransfer
(password required) or cancel the transfer.
8. Press the End Test button.
9. Verify that the switch transfers the load back to
Source N.
2. Verify that the generator set m aster switch is in the
AUTO position.
3. Refer to Figure 1-17. From the main screen, press
the Test button. Enter the test password when
prompted and press OK.
4. Press the down arrow button to display Type of Test
Loaded.
5. Press the Start button.
6. Verify that the generator set starts and the Source
E Available LED lights.
7. Verifythat the switch transfers the load to Source E.
Observe the controller LEDs and display as the
time delays execute and the load is transferred.
a. Standard-TransitionModels:Afterthe
preferred-to-standby transfer time delay, verify
that the Position N LED turns off and the
Position E LED lights, indicating that the switch
has transferred the load to Source E.
b. Programmed-Transition Models:After the
preferred-to-off time delay, verify that the
Position NLED turnsoff.After the
off-to-standby time delay, check that the
Position E LED lights, indicating that the switch
has transferred the load to Source E.
c. Closed-Transition Models: See Section 1.7.3.
After the preferred-to-standby time delay, the
controllermonitorsthesourcesfor
synchronization.When the sources are in
sync,theATStransferstheloadtoSourceE
and the Position E LED lights. Both sources will
be connected for less than 100 milliseconds
before Source N is disconnected and the
Position N LED turns off.
a. Standard-TransitionModels:Afterthe
standby-to-preferred time delay, verify that the
Position E LED goes out and the Position N
LED lights, indicating that the switch has
transferred the load to Source N.
b. Programmed-Transition Models:After the
standby-to-off time delay, verify that the
Position E LED goes out.After the off-topreferred time delay, check that the Position N
LED lights, indicating that the switch has
transferred the load to Source N.
c. Closed-Transition Models: See Section 1.7.3.
After the standby-to-preferred time delay, the
controllermonitorsthesourcesfor
synchronization.When the sources are in
sync, the ATS transfers the load to Source N
and the Position N LED lights. Both sources will
be connected for less than 100 milliseconds
before Source E is disconnected and the
Position E LED turns off.
If the sources do not synchronize before the fail
to sync time delay expires, operation depends
on the programmed transition override setting.
If automatic override is enabled, the ATS will
transfertheloadusingaprogrammedtransition transfer. If automatic override is not
enabled, the ATS will continue to monitor the
source synchronization and transfer when/if
the sources synchronize. The operator can
initiateaprogrammed-transitiontransfer
(password required).
10. After the engine cooldown time delay expires, the
engine start s ignal is removed. Verify that the
generator set stops.
Note: The generator set may have an engine
cooldown time delay that causes the
generator set engine to run after the transfer
switch engine start signal is removed.
TP-6883 4/2121Section 1 Operation
System OK
LD Exer ##/## @ ##:##
Norm ###V Emer ###V
BViewSetTest
On the main screen, press the Test button.
Enter Password
Time to Enter #:##
???<
>?
Typ e of Test
Auto Load
Time :## min
BMain"Start
Typ e of Test
Loaded
BMainStart
Engine Start in ##:##
Norm ###V Emer ###V
EndEnd
Delay Test
Use the open arrow buttons to enter the Test
password. See Section 1.5 for instructions. Then
press t he OK button.
OK
Press the down arrow button to step to the
loaded test screen.
Press the Start button to initiate the test
sequence.
Additional test sequence screens may appear,
depending on the system settings.
System on Test
Norm ###V Emer ###V
End
MainTest
PresstheMainbuttontoreturntothemainscreen
during the test, if desired. From the main screen,
press Test to return to the test screen.
PresstheEndTestbuttontoendthetest.
Figure 1-17Starting and Stopping the Automatic Operation Test
TP-6883 4/2122 Section 1 Operation
1.9Exercise
Schedule exercise runs through the Set Exercise
menus. See Section 4.4. To run the generator set at a
time other than a scheduled exercise sequence, use the
Test function. See Section 1.7 for instructions.
When a scheduled exercise is running, the menus
shown in Figure 1-18 appear. Press Main to return to
the main menu, if desired. Press the End button to end
the exercise sequence before the scheduled stop time,
if necessary.
If a system test or peak shave is active when the
exercise is scheduled to occur, the exercise is skipped.
A preferred-source failure during an exerciser period
causes the exercise to be terminated and normal ATS
operation to resume.
An exercise event can be temporarily disabled to
prevent its execution and then re-enabled later using the
enable/disable setting in the Set Exercise menus. See
Section 4.4.
1.9.1Unloaded Exercise
An unloaded exercise starts and runs the generator set
without transferring the load.
contactor transfers to the standby position, the load
bank control will be deactivated. (The standby source
supplies power to the load.)
1.9.3Loaded Exercise
A loaded exercise starts the generator set and transfers
the load from the normal source to the standby source.
On closed-transition models, transfer will occur when
the sources are synchronized. If the sources do not
sync, press Cancel to end the exercise.
Exerciser Active
Fail to Acquire 00:59
Norm ###V Emer ###V
MainEnd
Exerciser Active
Time Remaining 00:29
Norm ###V Emer ###V
MainEnd
Figure 1-18Exercise Sequence Menus
Display during
generator set engine
start or if engine does
not start immediately.
Display during
exercise run.
1.9.2Load Bank Control
The load bank control output is active during an
unloaded exercise or unloaded system test.If the
TP-6883 4/2123Section 1 Operation
1.10Warnings and Faults
When a fault exists, the System Alert indicator flashes, a
designated output and the common fault output are
turned on, and an appropriate message is displayed to
indicate the fault. See Figure 1-19 for the location of the
System Alert indicator.
ATS warnings and faults are shown in Figure 1-20.
There are three types of warning/fault conditions:
Warning. Warnings automatically reset with a source
availability change or a transfer request.
Fault Requiring Manual Reset.Under these
conditions, normal ATS operation is halted.Active
modes are turned off. If the contactor is in the preferred
source position, the engine cooldown time delay
executes and the engine start contacts open, allowing
the generator set to shut down. See Section 1.10.1 for
instructions to reset faults.
1
Self Resetting Faults. Under these conditions, active
modes are turned off. If the contactor is in the preferred
source position, the engine cooldown time delay
executes and the engine start contacts open, allowing
the generator set to shut down. When the fault condition
is corrected, the fault is automatically cleared from the
controller and normal ATS operation continues.
1. System Alert LED
GM85888
Figure 1-19Fault Indication
ConditionTy p eDescription
Failure to Acquire Standby SourceWarningThe source voltage did not reach the acceptable range within a set
Failure to Acquire Preferred SourceWarning
IPM Synching
(In-Phase Monitor Synching)
External Battery LowWarningThe voltage of the battery connected to the external battery supply
Failure to TransferWarningThe signal to transfer is sent to the contactor and the main shaft
Src N (or Src E) Rotation ErrSelf-Resetting FaultThe detected phase rotation of one or both sources does not match
I/O Module Lost CommSelf-Resetting FaultAn I/O device has stopped communicating or does not have a
Auxiliary Switch FaultManual Reset FaultThe main shaft auxiliary switches indicate that the ATS is in more
Auxiliary Switch OpenManual Reset FaultThe main shaft auxiliary switches indicate that the ATS is in neither
Manual Reset FaultThe Source1 or Source2 circuit breaker in the service entrance
time (see Time Delays). For example, the standby source generator
set did not start.
Synchronize time delay.
Note: If the sources do come into phase after the time delay
expires, the warning is automatically cleared and normal ATS
operation continues.
module (EBSM) is low.
auxiliary switch fails to indicate a complete ATS position change.
The controller will attempt to transfer the unit three times before the
fault is indicated.
the preselected setting.
correct address specified. Fault resets if communication is
reestablished.
than one position, or the position changed when no signal was sent
to initiate the change.
position (all inputs are open).
transfer switch has tripped due to an overcurrent condition. Identify
and correct the cause of the fault before resetting the controller.
detected. See Section 1.11.1 to reset.
module with the same address. See Section 1.11.2 to reset.
Figure 1-20Warnings and Faults
TP-6883 4/2124 Section 1 Operation
1.10.1 Fault Reset
To clear a fault or warning condition and reset the
System Alert LED, go to the Main menu and press the
down arrow button to open the Reset menu. See
Figure 1 -19 and Figure 1-21. Then press the button
labeled Reset.A fault reset does not change the
controller settings.
Module Status Change
Norm ###V Emer ###V
Reset View SetTest
Reset
New Module
Press Reset.
Press Reset.
See Section 1.11, Accessory Module Faults, for
instructions to correct and reset faults related to the I/O
modules and other accessory modules.
When a fault message
is displayed, press the
down arrow
on the Main menu to
step to the Reset Fault
menu.
Then press the Reset
button.
(B) button
Fault Message
LD Exer ##/## @ ##:##
Norm ###V Emer ###V
BView SetTest
Reset
Fault Description
BYReset Main
Figure 1-21Fault Reset
1.11Accessory Module Faults
Accessory modules are optional equipment.
1.11.1Module Status Change
Connecting or disconnecting one or more accessory
modules can cause the Module Status Change
message to be displayed.
Module Connection (new or reconnected module)
Installing or reconnecting one or more accessory
modules triggers the Module Status Change message.
See Figure 1-22. Press the Reset button to display
Reset New Module. Press the Reset button from that
menu. The controller recognizes the module type(s).
See Figure 1-23.
BYReset Main
Figure 1-22Menus after Module Connection
Module
Type
AOB
SOB
POB
Description
Switch/Alarm Module (alarm option board)
Standard I/O Module (standard option board)
High-Power I/O Module (power option board)
Figure 1-23Module Types
Disconnected Module
If one or more accessory modules are disconnected
from the controller, the message Module Status Change
appears. See Figure 1-24. Pressing the Reset button
displays the m essage Check Module Setup to Clear
Fault. Use the following Module Uninstall Procedure to
uninstall modules after disconnection.
Module Status Change
Norm ###V Emer ###V
Reset View SetTest
Check Module Setup
to Clear Fault
BYReset Main
Press Reset.
Press Main and follow
Module Uninstall
Procedure.
Figure 1-24Menus after Module Disconnection
Navigate to the Set Input/Outputs>Set Aux I/O menu to
check that the controller has recognized the connected
modules.
See Section 4.9 for instructions to assign programmable
inputs and outputs to I/O modules. Go to Section 4.10
for instructions to assign functions to the audible alarm
for an Alarm Module.
TP-6883 4/2125Section 1 Operation
Module Uninstall Procedure
1.11.2Module Status Conflict
1. Press Main t o return to the main menu.
2. Press Set to enter setup mode.
3. Enter the setup password.
4. Press the down arrow to step to the Set Inputs/
Outputs menu.
5. Navigate to the Set Auxiliary I/O menu.See
Figure 1-25. Press the right arrow button to see the
status of module 1. Press the down arrow to step to
the next module, if necessary, until the menu
shows Status: Lost.
6. Press the right arrow button to move to the
Uninstall Module menu. Verify that the menu says
Uninstall Module Yes.(Press the open arrow
button to toggle no/yes, if necessary.)
7. When Yes is displayed, press Save to uninstall the
module.
8. Repeat the uninstall procedure for additional
modules, if necessary.
Other Module Status Change Conditions
A Module Status Change message that cannot be
cleared as described in this section may indicate a
failure of the controller’s real-time clock. Carefully follow
the Module Connection or Module Uninstall procedures
to attempt to reset the fault. If the fault cannot be reset,
the controller’s logic board may need to be replaced.
Contact an authorized distributor/dealer for service.
The message Module Status Conflict appears if one
type of module is replaced with another type of module
that has the same address. Follow the procedure below
to resolve the conflict.
Procedure to Clear a Module Status Conflict
1. Disconnect power to the transfer switch.
2. Disconnect the module.
3. Close the enclosure door and reconnect power to
the ATS. The display will show Module Status
Change.
4. Press the button labeled Reset. The display will
show Check Module Setup to Clear Fault.
5. Follow the procedure in Section 1.11.1 to uninstall
the module through the ATS controller keypad.
6. Disconnect power to the ATS.
7. Connect the new module.
8. Close the enclosure door and reconnect power to
the ATS. The display will show Module StatusChange. SeeFigure1-22.
9. Press the button labeled Reset to display ResetNew Module. Press the reset button from that
menu. The controller will now recognize the new
module type.
10. Navigate to the Set Auxiliary I/O menu to check the
status and settings for the new module.See
Figure 1-25. Press the right arrow button to see the
status of module 1. Press the down arrow t o step to
the next module, if necessary,
Set Inputs/Outputs
BY"Main
Main Board I/O
BY"Back
Auxiliary I/O
BY"Back
Figure 1-25Uninstall Module
Set
Set
Module 1
Type POBAddr 1
Status: Lost
BY"Back
Pressthedownarrowbuttontosteptothe
next module, if necessary.
Module 1
Uninstall Module
Yes
Save Back
TP-6883 4/2126 Section 1 Operation
1.12Reset Data
1.12.1 Reset Maintenance Records
Be sure to read and understand the information in this
section before resetting records or parameters.
Note: Resetting to the d efault parameters will reset all
parameters to a factory default setting.
Use the Reset Data menus to set records or parameters
back to factory default settings. See Figure 1-26.
1. Use the black arrow buttons to step to the desired
menu.
2. Press the open up arrow button to toggle Yes or No
until Yes is displayed.
3. Press Save to reset the records or parameters to
the factory defaults. Pressing Back exits the menu
without resetting.
Reset Data
BY "Main
BSave Back
BSave Back
Reset
Maintenance Records
Yes /No
Reset
Event History
Yes /No
Reset the maintenance records after transfer switch
service to update the last maintenance date and totals
since reset that are displayed in the maintenance
records menu.
1.12.2 Reset Event History
Resetting the event history clears the events from the
event history log. The history lists the 100 most recent
transfer switch events, including transfers and DIP
switch setting changes as well as faults and alarms.
Figure 1-26Reset Data
Reset
Default Parameters
Yes /No
BSave Back
Reset
Exerciser Setup
Yes /No
BSave Back
Reset
Test Password
Yes /No
BSave Back
Disable
Test Password
Yes /No
BSave Back
Note: Resetting to the default parameters will reset all
parameters, including the system voltageand frequency, to a factory default setting. The
transfer switch will not operate correctly if the
system voltage and frequency do not matc h the
sources.
Note: Disable the test password only during service
unless the transfer switch is installed in a secure
location.
TP-6883 4/2127Section 1 Operation
1.12.3 Reset Default Parameters
1.12.4 Reset and Disable Test Password
Resetting to the default parameters will reset all
parameters, including the system voltage andfrequency, to a factory default setting. The default
system voltage and frequency settings may not match
the settings for your application.
The transfer switch will not operate correctly if the
system voltage and frequency do not match the
sources. Use the Set Sources menu to set the system
voltage and frequency after resetting to the default
parameters. See Section 6 for instructions.
Check the system operation to verify the settings after
resetting.
Reset the Test password to r eturn the test password to
the default, 0000.
Note: Disable the test password only during service
unless the transfer switch is installed in a secure
location.
Disabling the test password allows any user to initiate a
test sequence from the controller’s user interface
without entering a password. Initiating a test starts the
generator set and, if a loaded test is selected, transfers
the load.
TP-6883 4/2128 Section 1 Operation
Section 2 Sequence of Operation
This section explains the transfer switch sequence of
operation during the following events:
D Controller power-up or reset
D Preferred source loss and return
D Te s t
D Exercise
D Emergency source loss and return
The Sequence of Operation descriptions in Sections 2.2
through 2.4 describe the transfer switch normal
operation for standard, programmed, and closed
transition models. Operation can be affected by faults
such as the normal or emergency contacts failing to
open or close when signaled to do so.
2.1Controller Power-up/Reset
Following is an explanation of the sequence of operation
for the Decision-Makerr MPAC 1500 ATS Controller
when power is initially applied to the controller or a
controller reset occurs.
3. Contactor position and source availability are
determined.
4. If neither source is acceptable, the contactor does
not change p osition.
5. If both sources are available, the controller
immediately transfersthe contactorto the
preferred source.
6. If only one source is available, the controller
immediately transfers the contactor to that source,
executing only the off-position and load control
time delays.
If the available source is the preferred source, and
the contactor is in the standby position, the
contactor transfers to preferred, the engine
cooldown time delay runs, and then the engine
start contacts open.
If the available source is the preferred source and
the contactor is already in the preferred position,
the engine start contacts open immediately,
bypassing the engine cooldown time delay.
TP-6883 4/2129Section 2 Sequence of Operation
2.2Sequence of Operation,
Standard Transition Models
2.2.2Exerciser Operation, Standard
Transition
Operation can be affected by faults such as the normal
or emergency contacts failing to open or close when
signaled to do so.
2.2.1Preferred Source Loss and
Return, Standard Transition
Following is an explanation of the transfer switch
sequence of operation when Preferred Source failure is
detected.
Preferred Source Fails
1. Load control contacts open.
2. Engine start time delay expires.
3. The generator is signaled to start.
4. The generator starts and the standby source
becomes available.
5. Preferred-to-standby time delay expires.
6. Contactor transfers to standby.
7. Post-transfer load control sequences run.
Unloaded Exercise Sequence Starts
1. Exerciser timer begins.
2. The generator is signaled to start.
3. The generator starts and the standby source
becomes available.
4. The load bank control is activated.
Unloaded Exercise Sequence Ends
1. The load bank control is deactivated.
2. Engine cooldown time delay expires.
3. The engine start contacts open, signaling the
generator to stop.
Loaded Exercise Sequence Starts
1. Exerciser timer begins.
2. The generator is signaled to start.
3. The generator starts and the standby source
becomes available.
8. Load control contacts close.
Preferred Source Returns
1. Standby-to-preferred and pre-transfer load control
time delays expire.
2. Load control contacts open.
3. Contactor transfers to preferred source.
4. Post-transfer load control sequences and engine
cooldown time delay expire.
5. Load control contacts close.
6. The engine start contacts open, signaling the
generator to stop.
4. Preferred-to-standby time delay and pre-transfer
load control sequences run.
5. Load control contacts open.
6. Contactor transfers to standby.
7. Post-transfer load control sequences run.
8. Load control contacts close.
Emergency Source Fails (Normal Source is
available)
1. Exerciser is deactivated.
2. Load control contacts open.
3. Contactor immediately transfers to preferred.
4. Immediate failure to acquire standby alarm.
5. Post-transfer load control sequences and engine
cooldown time delay expire.
6. Load control contacts close.
7. Engine start contacts open.
TP-6883 4/2130Section 2 Sequence of Operation
Loading...
+ 94 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.