Kodak Min R Mammography Processor User manual

{OperatorsManual}{Product io n}{HealthImaging}
Publication No. 3E0815
29SEP99
HEALTH IMAGING
OPERATOR MANUAL
for the
H176_0001AC
© Eastman Kodak Company, 1999
PLEASE NOTE The information contained herein is based on the experience and knowledge relating to the
subject matter gained by Eastman Kodak Company prior to publication. No patent license is granted by this information. Eastman Kodak Company reserves the right to change this information without notice, and
makes no warranty, express or implied, with respect to this information. Kodak shall not be liable for any loss or damage, including consequential or special damages, resulting from any use of this information, even if loss or damage is caused by Kodak’s negligence or other fault.
Warning
To avoid hazardous conditions, keep floors and floor coverings around your Processor and associated drains clean and dry at all times. Any accumulation of fluids from mixing tanks, drain lines, etc., should be cleaned up immediately. In the event of an accumulation of liquid due to backup, overflow, or other malfunctions of the drain associated with your Processor, call a plumber or other contractor to correct any problem with the drain. Kodak accepts no responsibility or liability whatsoever for the serviceability of any drain connected to or associated with a Processor. Such drains are the sole responsibility of the customer.
Table of Contents
Description Page
Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
Overview Processing Film
Processing Solutions
Correcting Difficulties Cleaning and Periodic Maintenance
New Equipment Warranty
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Start Up Feeding Sheet Film Shutdown
Preparing the Solutions Draining the Tanks Filling the Tanks
Daily Procedure Every Two Week s
Warranty Coverage How to Obtain Service Limitations
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
2 29SEP99 – 3E0815
Safety
Section 1: Safety
Chemical Hazard
Warning
Gloves and protective eyewear must be worn when working with chemistry, especially fixer solutions. If chemicals get into the eyes, flush them thoroughly with large volumes of water. If irritation persists, see a doctor.
Besides being a potential cause of falls, the chemistry can contaminate waste water and irritate the skin and eyes. Chemistry spills must be cleaned up immediately. You must:
• prevent the spilled chemistry from entering a waste water drain.
• clean up the spill with a moist mop or rag.
• dispose of cleaning material s and was te water col lecte d duri ng the clean up acc ordi ng to envi r onm ent regulations.
• ensure that the air exhaust pipe is always connected to a ventilation system, because inhalation of chemicals is dangerous to health and should be avoided.
• observe all environmental regulation for storage and disposal of waste chemicals.
• use these instructions in conjunction with the chemical instructions.
Operator Precaution
Warning
The operator must:
only be allowed to remove the Top Cover to avoid electrical and mechanical hazard.
• not exchange any internal parts.
• call service personnel to perform any repairs.
• not operate equipment with a damaged power cord.
• not use an extension cord to power this equipment.
• position the power cord so it will not be tripped over or pulled.
Equipment Damage
Caution
The equipment is not contained in a sealed cabinet. Therefore, it must not be used in locations where it can come in contact with liquids, including bodily fluids.
This equipment generates, uses, and can radiate radio frequency energy. If not installed by qualified service personnel and used in accordance with the User Guide, may cause interference to radio communications and other electronic devices. Operation of this equipment in a residential area may cause interference, in which case the user, at their own expense, will be required to take whatever measures may be required to correct the interference.
3E0815 – 29SEP99 3
OPERATORS MANUAL
Nonpatient Contact Equipment
1.83 m
2.5 m
1.83 m1.83 m
H176_0034HC
Warning
This device is not medical equipment according to EN 60 601-1 and must not enter the patient environment as defined by EN 60 601-1. The following requirements have to be met:
• Distance from the device to Nonpatient Contact Equipment: Horizontal = 1.83 meters Vertical = 2.5 meters above the floor under the patient
• Contact of the patient and the device simultaneously by the caregiver is not allowed.
• No direct electrical connection is allowed between the device and the Nonpatient Contact Equipment.
4 29SEP99 – 3E0815
Overview
Section 2: Overview
The Processor is designed to process medical x-ray sheet film that can be processed in the Kodak RP X-Omat Developer and Replenisher, the Kodak RP X-Omat Fixer and Replenisher, and the Kodak RP X-Omat EX Developer and Replenisher.
The self-threading roller transport system has:
• a film-length detection system
• a replenishment system
• a developer solution filter
• a recirculation system for the developer and fixer solutions
• an automatic standby feature
The Processor can operate at 2 different transport speed cycles:
• Standard Cycle has a transport speed of 76.2 dry to dry, for a 43 cm (17 in.) sheet of film. The developer temperature is 33.3
• Rapid Cycle has a transport speed of 101.6 ± dry to dry, for a 43 cm (17 in.) sheet of film. The developer temperature is 34.4 ±
Note
The transport speed and the developer solution filter can only be changed by qualified service personnel. A film feed signal sounds 3 seconds after the film being processed has cleared the Rollers of the Detector Crossover
Assembly. This indicates that the Processor is ready to accept another piece of film.
± 1.5 cm (30.4 ± 0.6 in.) per minute or 157 ± 5 seconds drop time,
± 0.3°C (92.0 ± 0.5°F).
± 2.0 cm (40.0 ± 0.8 in.) per minute or 118 ± 4 seconds drop time,
± 0.3°C (94.0 ± 0.5°F).
3E0815 – 29SEP99 5
OPERATORS MANUAL
Section 3: Processing Film

Start Up

H176_0001ACB H176_0001AC
TOP COVER
MAIN POWER SWITCH
[1] Check that the Processor is off. [2] Remove the Top Cover.
6 29SEP99 – 3E0815
Processing Film
FIXER/WASH CROSSOVER ASSEMBLY
WASH RACK
DEVELOPER/FIXER CROSSOVER ASSEMBLY
DRYER ASSEMBLY
[3] Check that the following are seated correctly:
• Developer Rack
•Fixer Rack
• Wash Rack
• Fixer/Wash Crossover Assembly
• Developer/Fixer Crossover Assembly
• Detector Crossover Assembly
•Dryer Assembly
FIXER RACK
DEVELOPER RACK
OVERFLOW LEVEL
WEIRS
DETECTOR
DETECTOR
CROSSOVER
CROSSOVER
ASSEMBLY
ASSEMBLY
H172_1001CCD
H172_1001CC
[4] Check that the solutions in the Developer, Fixer,
and Wash Tanks are at the Overflow Level.
[5] Turn on the water supply.
H172_1004ACE H172_1004AC
3E0815 – 29SEP99 7
WASH TANK
FIXER TANK
DEVELOPER TANK
Loading...
+ 15 hidden pages