Kodak II Plus User manual

Page 1
Publication Part No. 6F6595
15 MAR 04
Supercedes 927 0532
10 NOV 03
INSTALLATION INSTRUCTIONS
for the
PIERCER IMPROVEMENT KIT
for the
Kodak AUTOMIXER II
Service Code: 3112
Kodak AUTOMIXER II Plus
Service Code: 3439
Important
Qualified service personnel must install this equipment.
HEALTH IMAGING
Confidential, Restricted Information
Page 2
Installation Instructions
PLEASE NOTE: The information contained herein is based on the experience and knowledge relating to the subject matter gained by Eastman Kodak
Company prior to publication.
No patent license is granted by this information.
Eastman Kodak Company reserves the right to change this information with out notice, and makes no warranty, express or implied, with respect to this information. Kodak shall not be liable for any loss or damage, including consequential or special damages, resulting from any use of this information, even if loss or damage is caused by Kodak’s negligence or other fault

Installation

Purpose
If chemistry drains slowly from the bottles, because the PIERCER in Kodak AUTOMIXER II or AUTOMIXER II Plus does not pierce the cap and seal completely, you will need to install the following parts:
MODEL of Automixer
AUTOMIXER II 9171279 2 (one for each PIERCER
AUTOMIXER II Plus 9270531 2 (one for each PIERCER
Installation Instructions 6F6595 1
These parts are available from Service Parts Management.
Part Numbers Required
Parts Needed for each AUTOMIXER
for BOTTLE “A”)
for BOTTLE “A”)
2
15 MAR 04 – 6F6595
Page 3
Installation Instructions
Installing the new PIERCER Parts to an AUTOMIXER II Plus
Important These parts improve piercing of the cap and seal of the chemistry bottles, do NOT remove the existing PIERCERS.
Warning
Dangerous Voltage
1. De-energize and disconnect the main power from the AUTOMIXER.
TOP COVERS
BOTTLES
2. Open the TOP COVERS and remove any empty BOTTLES, that may be inserted into the MIXER.
6F6595 - 15 MAR 04
3
Page 4
Installation Instructions
INSERT
INSERT – installed on PIERCER, top view
tabs
opening
PIERCER
INSERT – installed on PIERCER, bottom view
PIERCER BOTTLE “A”
PIERCER
tabs
BASIN
2. Install INSERT onto the PIERCER used for BOTTLE “A” for DEVELOPER and FIXER, making sure to align the opening over the moveable blade. Make sure tabs on INSERT lock underneath the PIERCER
2. Close the TOP COVERS.
3. Energize the AUTOMIXER and check for correct operation.
4
15 MAR 04 – 6F6595
Page 5
Installation Instructions
Installing the new PIERCER Parts to an AUTOMIXER II
Important These parts improve piercing of the cap and seal of the chemistry bottles, do NOT remove the existing PIERCERS.
Warning Dangerous Voltage
1. De-energize and disconnect the main power from the AUTOMIXER.
2. Open the TOP COVERS and remove any empty BOTTLES, that may be inserted into the MIXER.
Caution
New blades have sharp points, use extreme caution when handling and installing!
3. Install the 2 interlocking new blades over the existing PIERCERS used for BOTTLE “A” for
DEVELOPER and FIXER.
New interlocking blades (2 pieces)
PIERCER BOTTLE “A”
Original PIERCER
6F6595 - 15 MAR 04
5
Page 6
Installation Instructions
4. Close the TOP COVERS.
5. Energize the AUTOMIXER and check for correct operation.
Publication Date Publication No. ECO No. Changed Pages File Name Notes
15 OCT 03 927 0532 ------- -------- AM2 Piercer.doc New Publication
15 MAR 04 6F6595 ------- All AM2 Piercer.doc Publication part number
changed only, content remains the same.
Kodak, Min-R, Industrex, and X-Omat
Printed in the U.S.A. HEALTH IMAGING
EASTMAN KODAK COMPANY Rochester NY 14650
are trademarks.
6
15 MAR 04 – 6F6595