Kodak, RVG, Trophy, Ultra-Speed, and Insight are trademarks of
Eastman Kodak Company. Other product or company names
mentioned in this manual are used only for the purpose of
identification and may be trademarks of their respective owners.
SM701_K2200_2_en
Table of Contents
1 Safety and Regulatory Information
Conventions Used in This Manual............................................................................................................1-1
General Safety Guidelines.........................................................................................................................1-1
Warnings and Safety Instructions .............................................................................................................1-2
IEC Symbols Used ....................................................................................................................................1-5
Ceiling-mounted Unit (Optional) ..............................................................................................................2-3
Mounted on Mobile Stand (Optional) .......................................................................................................2-4
Floor-mounted Unit (Optional) .................................................................................................................2-5
Control Timer Unit....................................................................................................................................2-6
Additional Features ...................................................................................................................................3-5
4 Care and Maintenance
General Maintenance.................................................................................................................................4-1
The information contained in this manual is based on the experience and
knowledge relating to the subject matter gained by Eastman Kodak
Company prior to publication. No patent license is granted by this
information.
Eastman Kodak Company reserves the right to change this information
without notice, and makes no warranty, express or implied, with respect to
this information. Kodak shall not be liable for any loss or damage, including
consequential or special damages, resulting from any use of this information,
even if loss or damage is caused by Kodak's negligence or other fault.
Conventions Used in This Manual
CAUTION:
Caution points out procedures that you must follow precisely to avoid
damage to the system or any of its components, yourself or others, loss
of data, or corruption of files in software applications.
Note
Notes provide additional information, such
as expanded explanations, hints, or
reminders.
Important
Important highlights critical policy
information that affects how you use this
manual and this product.
General Safety Guidelines
•This product is designed and manufactured to ensure maximum safety
of operation. Operate and maintain it in strict compliance with the safety
precautions and operating instructions contained in this manual.
•This product meets all the safety requirements applicable to medical
equipment. However, anyone attempting to operate the system must be
fully aware of potential safety hazards.
•There are no user serviceable parts in this system. The product must be
installed, maintained, and serviced by qualified service personnel
according to procedures and preventive maintenance schedules in the
product service manual. If your product does not operate as expected,
contact your Service Representative.
•Do not modify this product in whole or in part without prior written
approval from Eastman Kodak Company.
•The assembly, extensions, adjustments, modifications, and repairs must
be performed by an authorized Service Representative. Your radiology
system must be installed in premises that comply with applicable
standards.
•Personnel operating and maintaining this system should receive training
and be familiar with all aspects of operation and maintenance.
1 August 2006SM701_K2200_2_en1-1
•To ensure safety, read all user manuals carefully before using the system
and observe all Caution, Important, and Note callouts located
throughout the manual.
•Keep this manual with the equipment.
•Reading this manual does not qualify you to operate, test, or calibrate
this system.
•Unauthorized personnel are not allowed access to the system.
•If the product does not operate properly or fails to respond to the
controls as described in this manual:
• Follow the safety precautions as specified in this manual.
• Stop using the equipment and do not make or authorize any changes
to it.
• Immediately contact your Service Representative, report the problem,
and await further instructions.
•X-ray systems manufactured by Eastman Kodak Company comply with
safety standards throughout the world for optimum protection against
radiation risks.
•Be aware of the product specifications and of system accuracy and
stability limitations. Consider these limitations before making any
decision based on quantitative values. If you have any doubts, consult
your Sales Representative.
CAUTION
X-rays can be dangerous if used incorrectly. Take precautions even when
following the instructions in this manual.
Use conventional commercially available equipment to protect yourself
and your
CAUTION
against scattered radiation risks.
•If you fail to comply with these instructions, Eastman Kodak Company
will not be responsible for the safety reliability, and characteristics of the
equipment.
:
:
Warnings and Safety Instructions
CAUTION:
Do not operate the equipment in the presence of explosive liquids,
vapors, or gases. Do not plug in or turn on the system if hazardous
substances are detected in the environment. If these substances are
detected after the system has been turned on, do not attempt to turn off
the unit or unplug it. Evacuate and ventilate the area before turning off
the system.
DANGER: THIS IS AN ELECTRICAL UNIT. DO NOT EXPOSE IT
TO WATER SPRAY. SUCH ACTION MAY CAUSE AN
ELECTRICAL SHOCK OR A MALFUNCTION OF THE
UNIT.
1-2SM701_K2200_2_en1 August 2006
WARNING
The user is responsible for the operation and maintenance of this unit.
This unit must only be operated by legally qualified persons.
The cover of the unit must not be opened by the operator.
Inspection and maintenance operations should only be carried out by an
approved Kodak Dental Systems technician.
WARNING
This unit must be installed in an x-ray room that complies with current
installation standards. From this location, visual or audio communication
must be maintained with the patient, together with access to the control
interface during exposure.
WARNING
Do not operate the unit if there is the threat of an earthquake.
Following an earthquake, ensure that the unit is operating properly before
using it again.
Failure to observe this precaution may expose patients to hazards.
WARNING
X-ray equipment can be hazardous to patients and the operator if the
exposure safety factors and operating instructions are not observed.
WARNING
Do not place objects within the field of operation of the unit.
WARNING
We recommend that the patient and the operator wear protective lead-lined
aprons, unless other Radiation Protection Protocols apply locally.
Ensure that any parts of the unit that may come into contact with the patient
and the operator have been disinfected after each patient has been exposed to
x-rays.
If the unit develops a fault, turn it off (O) and display a sign that states “Out
of Service.”
WARNING
The operator must ask the patient to refrain from moving during the entire
period of exposure.
In the European Union, this symbol indicates that when the last
user wishes to discard this product, it must be sent to an
appropriate facility for recovery and recycling.
Contact your local Kodak representative or refer to
www.kodak.com/go/recycle for additional information on the
collection and recovery programs available for this product.
1 August 2006SM701_K2200_2_en1-3
Labeling Summary
Safety Labels
CHASSIS GROUND STUD
ATTENTION: CONSULT ACCOMPANYING DOCUMENTS
CAUTION: IONIZING RADIATION
1-4SM701_K2200_2_en1 August 2006
IEC Symbols Used
The system may have labels with one or more of the following symbols.
These symbols indicate the IEC standards to which the system conforms.
Caution — consult accompanying documents
Protective earth
Power ON
Power OFF
Regulatory Information
The product conforms to the following safety standards: IEC/EN 60 601-1
Medical Electrical Equipment General Requirements for Safety, IEC/EN 60
601-2 Medical Electrical Equipment Electro-Magnetic Compatibility
Requirements and Tests.
CE Conformity
The Kodak intraoral x-ray system is a Class II b medical device, which bears
the following mark of conformity: .
U.S. Regulations
U.S. federal law restricts this device to sale by or on the order of a
dentist.
CAUTION:
1 August 2006SM701_K2200_2_en1-5
2
System Overview
Components
Figure 1.KODAK 2200 Intraoral X-ray System
High-frequency x-ray generator
•Transformer and associated electronics, and an oil-bathed x-ray tube
•Beam-limiting device
• Radiation diameter – 6 cm (2 3/8 in.)
•
Distance from x-ray tube focal spot to skin – 20 cm (7 7/8 in.
•Angle scale and handle to facilitate positioning
Wall framework
•Contains the high-frequency generator’s control electronics
)
1 August 2006SM701_K2200_2_en2-1
Control timer unit
•Anatomical selection and digital display of parameters (kV, mA,
exposure time)
•Self-test of the microprocessor each time the unit is activated
•Alarm during incorrect operation
•RVG key that automatically adjusts the exposure parameters (time
and mA) if you are using RVG (RadioVisioGraphy)
Scissor arm
•Wall-mounted with a choice of extensions
Figure 2.Side view of KODAK 2200 Intraoral X-ray System
Table 3.Types of Scissor Arms
ExtensionRSpan A
CG 64547.0 cm
(18.5 in.)
CG 64664.8 cm
(25.5 in.)
CG 64882.5 cm
(32.5 in.)
170.0 cm
(66 15/16 in.)
188.0 cm
(74 in.)
205.0 cm
(80 11/16 in.)
On/off switch
•Contains built-in light
Rectangular collimator (optional)
•Different sizes adapted to films and RVG sensors
Additional options
•Separate control timer
•An additional beam-limiting device measuring 10 cm (4 in.) that limits
radiation to a diameter of 6 cm (2 3/8 in.) on the skin
•Separate exposure switch (if the control panel is attached to the wall
framework)
•Ceiling-mounted unit
•Floor-mounted unit
•Unit mounted on mobile stand
2-2SM701_K2200_2_en1 August 2006
Ceiling-mounted Unit
Figure 4.KODAK 2200 Intraoral X-ray System ceiling-mounted unit
High-frequency x-ray generator
Ceiling-mounted unit containing the high-frequency x-ray
generator’s control electronics
Separate timer/control unit for the x-ray generator
Scissor arm
On/off switch with built-in light
Rectangular collimator
1 August 2006SM701_K2200_2_en2-3
Mounted on Mobile Stand (Optional)
Figure 5.KODAK 2200 Intraoral X-ray System mounted on mobile
stand
High-frequency x-ray generator
Mobile stand containing the high-frequency x-ray generator’s
control electronics
Timer/control unit for the x-ray generator
Scissor arm
On/off switch with built-in light
Rectangular collimator
Handle
Foot brake
2-4SM701_K2200_2_en1 August 2006
Floor-mounted Unit (Optional)
Figure 6.KODAK 2200 Intraoral X-ray System floor-mounted unit
Every dental specialist would like to produce high-quality intraoral
radiographs that reveal maximum detail with the minimum dose to the
patient, show teeth and anatomic structures accurately with a minimum of
distortion or magnification, and have optimal density and contrast to
maximize their use for the detection of dental diseases.
To obtain high-quality intraoral radiography with maximum details, take
extra care in all three steps of the radiography process: positioning the
patient, the x-ray generator, and the imaging system; exposing the film or the
sensor; and processing the film.
Seat the patient with the sagittal plane vertical.
•For radiography of the upper maxillary, the Frankfort plane (nose-ear
plane) must be horizontal
•For radiography of the lower maxillary, the occlusal plane must be
horizontal
Figure 8.Patient positioning
Positioning the x-ray generator
The scissor arm allows you to accurately position the generator for any type
of exposure. The beam-limiting device maintains a distance of at least 20 cm
(8 in.) between the focal spot and the skin, which allows you to use either the
paralleling technique or the bisecting technique.
Paralleling technique
The positioning tool used in the paralleling technique allows you to align the
beam and the receptor. An adapted collimator reduces the dosage by limiting
surface exposure.
Bisecting technique
When using the bisecting technique, do not use a rectangular collimator. This
limits the risk of misaligning the x-ray beam and the image receptor.
1 August 2006SM701_K2200_2_en3-1
Figure 9.Paralleling technique (left) and Bisecting technique (right)
Positioning the imaging receptor
Using the KODAK 2200 Intraoral X-ray System, you may create an x-ray
image on one of three different types of imaging receptors:
•Conventional silver halide films, such as KODAK INSIGHT or
KODAK ULTRA-SPEED dental films
•Digital sensors, such as KODAK RVG sensor
•Phosphor plate
Properly placing the receptor is critical. Check your dental radiography text
for information about proper placement of the imaging receptor.
Improperly positioning the film or sensor results in errors on the radiograph,
such as distorted teeth and roots, elongation, magnification, and/or
overlapping contacts. The paralleling technique generally reduces the risk of
such errors. However, if you improperly position the system, angulation
errors can occur (angulation of the receptor to the tooth itself).
If the exit pattern of the beam is not aligned with the imaging receptor, then
part of the radiograph will not be exposed to radiation and the final
radiograph will have some clear (unexposed) areas. This defect is called
“cone cuts”.
The imaging receptor is marked to indicate the tube side. If the orientation is
not correct, the resulting radiograph is lighter and may show artifacts, such as
foil pattern or sensor cable.
3-2SM701_K2200_2_en1 August 2006
Exposure
Exposure Parameters
Procedure
Because each receptor (film, digital sensor, or phosphor plate) has its own
sensitivity to x-ray radiation, the choice of receptor affects the exposure
parameters. For instance, sensitivity class for conventional dental films is
characterized with a letter D, E, or F where F is more sensitive than E, and E
more sensitive than D. Consequently, the required dose for the correct
exposure goes down with each increase in sensitivity.
Adjust for these differences in sensitivity by changing the exposure times
(f+, f- of the operating controls) as indicated in Table 22 on page 5-8.
Like film, different digital sensors have varying sensitivities, as indicated in
Table 23 on page 5-8, so you need to adjust exposure for sensors as well.
Use the tables, which are established under manufacturers’ standard
conditions, as guidelines and adjust your exposures as needed, based on the
results you get from your system’s exposure and processing.
1. Turn on the system.
The following displays light up:
• On button.
• Tooth or special exposure selector (bitewing or occlusal).
• Child or adult selector.
• Display unit showing exposure parameters: kV, mA, type of film
used, exposure time.
2. Program the generator using the control unit.
a. Select the exposure mode (RVG or film) depending on the
equipment.
The displays shows:
• 7 mA for film.
• 4 mA for RVG (use this mode to prevent overexposure of digital
sensors).
b. Select the receptor type (Table 22 on page 5-8 or Table 23 on
page 5-8)
c. Select the kV value:
• 60 kV for high-contrast radiographs.
• 70 kV for lower-contrast radiographs that provide a wider latitude
and more levels of gray (for periodontal work).
d. Select the type of patient: child or adult.
e. Adjust the receptor based on the patient’s size:
• Increase the receptor type (f+) for large patients (due to risk of
underexposure).
• Reduce the receptor type (f-) for small patients (due to risk of
overexposure).
f. Select the tooth or the exposure.
The unit displays the exposure time.
1 August 2006SM701_K2200_2_en3-3
Note
Exposure times range from 0.027 to
1.19 sec. for conventional films and from
0.014 to 0.624 sec. for digital sensors.
3. Make an exposure.
a. Press the radiology control key.
Processing
CAUTION
:
The operator must instruct the patient to refrain from moving during
the entire exposure.
The x-ray emission indicator lights up and an audible signal is
emitted.
b. Keep pressing until the x-ray emission light goes out and the audible
signal stops.
CAUTION
:
If you stop pressing the control key before the exposure ends, a
manipulator alarm is activated. It indicates that the x-ray emission was
interrupted prematurely and that there is a risk of underexposure. The
digital display alternates between display of OP ERROR and the
amount of selected time not used. To stop the alarm, select a tooth or
special exposure (Bitewing or Occlusal).
When using conventional film, process the film according to manufacturer’s
instructions. Develop the film under safelight conditions in an automatic
processor or manually.
If you use an automatic processor, refer to the processor’s manual. Be sure to
maintain the mechanically and keep the solutions replenished.
If you develop film manually, follow precisely the manufacturer’s
recommendations
for solution preparation, development time, and solution
temperature. Any deviation from the manufacturer’s recommendations (such
as a solution that is too concentrated or diluted, too hot or cold, or if film is
processed for the wrong amount of time), will adversely affect the quality of
the final radiograph.
3-4SM701_K2200_2_en1 August 2006
Additional Features
•KODAK 2200 Intraoral X-ray System uses a high-frequency
technology that has several advantages:
• Shorter exposure times, reducing the risk of blur due to movement of
the patient or film during exposure.
• Reduction in x-ray dose to patients because the KODAK 2200
System emits fewer soft rays absorbed by patients that do not
contribute to the radiological picture.
•Choice of high voltage between 60 kV and 70 kV:
• 60 kV for high-contrast radiographs giving a clear image of the
endodontal instruments or clearly displaying tooth decay.
• 70 kV for better reproduction of levels of gray (soft tissue), useful for
periodontal work.
•A thermal safety system prevents the generator from overheating in case
of intensive use. This system can prohibit any exposure as long as the
generator did not cool down. A COOLING error message appears on
the display unit and an audible signal is emitted untill the cooling period
is over.
CAUTION
Do not turn off the system. If you turn off power, the microprocessor
does not calculate the cooling time, and for safety reasons considers that
the system has not gone into the cooling cycle.
•While the exposure is taken, the exposure time counts off on the control
unit display.
If the exposure is interrupted (such as by releasing the key), the audible
and visible manipulator alarm is activated and the remaining exposure
time is displayed. This information makes it easier to decide whether to
develop the film or to start another exposure. (If the remaining time is
short, you may develop the film.)
To stop the manipulator alarm, press any tooth selection or exposure
type (bitewing or occlusal).
•The system includes a self-test function that is activated when turning
the system on while simultaneously pressing the RVG key for a few
seconds.
The self-test checks:
• All the lights on the control unit.
• The audible alarm and display unit.
At the end of the self-test, the total number of exposures taken by this
unit since it was installed is displayed. A short beep indicates the test is
over.
At the end of the test, the systems returns to normal operation.
:
1 August 2006SM701_K2200_2_en3-5
4
Care and Maintenance
General Maintenance
To make sure that the system functions correctly, you must have it serviced
annually by an authorized KODAK Dental Systems dealer. In addition,
every three months inspect the equipment and make sure of the following:
Generator
Mechanical support
Control unit and electrical installation
Functioning
•The certification label is legible.
•There are no oil leaks.
•The wall framework is securely attached to the wall.
•All the labels are legible.
•The scissor arm is stable in all positions.
•The symbols are legible.
•The control unit cable and the power supply cable are in good condition.
•The ground is correctly installed.
•The radiology control key returns to its initial position after use.
•The audible signal is audible and the x-ray emission light is visible when
you make an exposure (Occlusal, film type 9, 60 kV, 7 mA).
Important
If the result of any of these checks is not
satisfactory, discontinue using the
equipment and contact an authorized
Kodak Dental Systems technician.
•The message “OP. ERROR” is displayed when you make an exposure
(Occlusal, film type 9, 60 kV, 7 mA) and release the control button
before the exposure time has elapsed.
Timer self-test
•Turn off the system.
•Press and hold the RVG key and turn on the system. When the first light
comes on, release the RVG key.
All the control unit functions and indicator lights will be tested one by
one. At the end of the test, the total number of exposures made by the
machine is displayed.
•A short beep indicates that the test has ended.
1 August 2006SM701_K2200_2_en4-1
Cleaning
Disinfecting
Clean the outside of the system with a damp paper towel or soft cloth using
an alcohol-based, non-corrosive cleaner.
If necessary, wipe off surfaces with disinfectant.
CAUTION
• Do not allow liquids to drip into the system.
• Do not spray cleaner or disinfectant directly onto the machine.
• Protect the system from contamination using barriers available
from dental distributors.
• Follow the manufacturer’s safety recommendations when using the
cleaner or disinfectant.
:
Error Messages
Table 10.Error messages
Error messageCauseHow to cancel
COOLINGCooling cycle; this message can appear
during a period of intensive use.
CAUTION
If you turn off power to the system, the microprocessor does not calculate the cooling time, and for
safety reasons considers that the system has not gone into the cooling cycle.
OP. ERROR
plus audible alarm
:
Release of the radiography control
button before the end of the exposure.
The display shows the remaining
exposure time. (Based on this time,
decide whether to develop the film or
make another exposure.)
Do not turn off the system. The error
message will disappear when the system
returns to a satisfactory temperature.
Press any key on the arch to stop the
alarm.
KV ERRORThe generator’s high voltage value is
different from the required value.
POWER ERROROver- or under-tension of the main
power.
Power supply problem on the filament.
Cut filament.
4-2SM701_K2200_2_en1 August 2006
Turn off the system and restart. If the
problem persists, call a qualified service
technician and discontinue using the
equipment.
Turn off the system and restart. If the
problem persists, call a qualified service
technician and discontinue using the
equipment.
Troubleshooting
Table 11.Troubleshooting
ProblemCauseSolution
Nothing lights upUnit is disconnected.Connect the unit.
Fuse F1 is burned out or defective.Replace the fuse.
Circuit breaker is off.Turn on the circuit breaker.
Control unit does not
light up
No x-ray emissionGenerator is cooling.Wait for the COOLING message to
X-ray emission works,
but exposure is too light
or completely white
Control unit is disconnected.Connect the control unit.
Fuse F1 is burned out or defective.Replace the fuse.
Control unit is defective.Call a qualified service technician.
disappear.
Radiology control key is defective.Call a qualified service technician.
Wrong receptor type is being used.Change the receptor type. (Refer to
Table 22 on page 5-8 and Table 23 on
page 5-8.)
Generator is positioned incorrectly.Adjust the position of the generator.
Exposure time is too short.Increase the exposure time.
Development time is too short.Increase the development time. (Refer to
the development instructions.)
Developer is too cold.Heat the developer.
Developer is too old or diluted.Replace with fresh developer.
RVG key is incorrectly selected.Verify your exposure settings. (Refer to
the exposure procedure.)
X-ray emission works,
but exposure is too dark
Receptor is facing the wrong way.Reposition the receptor.
Unit was incorrectly installed.Call a qualified service technician.
Wrong receptor type is being used.Change the receptor type. (Refer to
Table 22 on page 5-8 and Table 23 on
page 5-8.)
Exposure time is too long.Decrease the exposure time.
Development time is too long.Decrease the development time. (Refer
to the development instructions.)
Developer is too hot.Cool the developer.
Developer is too concentrated.Adjust the concentration or replace the
developer.
RVG/film key was incorrectly selected.Verify your exposure settings. (Refer to
the exposure procedure.)
1 August 2006SM701_K2200_2_en4-3
5
Specifications
According to IEC Standard 601-2-7
Manufacturer
Kodak -Trophy
A subsidiary of Eastman Kodak Company
4, rue F. Pelloutier - Croissy-Beaubourg
77437 Marne-la Vallée Cedex 2
France
Models
•Dental X-ray diagnosis devices, class 1, type B, intermittent use
•KODAK 2200-TR: equipped with tube TRX 708 from
KODAK-TROPHY
•KODAK 2200-C: equipped with tube OCX / 65-G from CEI
Electric power supply (during exposure)
•230–240 V AC (± 10%), 50 Hz, 5 A, apparent resistance 0.5 Ω
•100–110–130 V AC (± 10%), 50/60 Hz, 12 A, apparent resistance 0.2 Ω
Electric power supply (no exposure)
•230–240 V AC (± 10%), 50 Hz, 100 mA
•100–110–130 V AC (± 10%), 50/60 Hz, 100 mA
Rated high voltage and maximum corresponding current
•Film mode: 70 kV, 7 mA
•RVG mode: 70 kV, 4 mA
Current/voltage combinations for a maximum output power of:
•490 W in film mode, 70 kV / 7 mA
•280 W in RVG mode, 70 kV / 4 mA
Rated power for exposure time of 0.1 sec.
•Film mode: 490 W
•RVG mode: 280 W
Rate of use
•At 70 kV, 7 mA and 0.1 sec. and at the maximum tank temperature: one
exposure every 8 sec.
Minimum value of the current/time product in the range of
conformity
•0.07 mAs at 7 mA
•0.04 mAs at 4 mA
1 August 2006SM701_K2200_2_en5-1
Selection of parameters
•Film mode: 70 kV / 7 mA and 60 kV / 7 mA
•RVG mode: 70 kV / 4 mA and 60 kV / 4 mA
Area of conformity to IEC standard 60601-2-7 (2002)
•Reproducibility of the emitted radiation: conform
•Linearity of the emitted radiation: conform
•Precision in radiography: conform
Measurementconditions
•kV: Indirect measurement using a kV peakmeter
•mAs: Direct measurement in the circuit using a mAs-meter
•Exposure time: Indirect measurement on the kV signal at 75% of the
peak value
Storage and transportation conditions
•Temperature: -10°C to 60°C (14°F to 140°F)
•Relative humidity: 10% to 95%
•Atmospheric pressure: 700 to 1060 hPa
Dimensions and weight
•Control unit:
•Wall framework:
•X-ray emitting unit:
•Scissor arm:
•Mobile stand
(optional):
•Floor column
(optional):
•Ceiling column:
16 x 9 x 4 cm
(6.3 x 3.5 x 1.6 in.)
51.4 x 18.9 x 10.8 cm
(20.2 x 7.4 x 4.3 in.)
43.8 x 22.6 x 12 cm
(17.2 x 8.9 x 4.7 in.)
87.3 x 13.3 x 6.3 cm
(34.4 x 5.2 x 2.5 in.)
90 x 60 x 110 cm
(35.4 x 23.6 x 43.3 in.)
24 x 23 x 90 cm
(9.4 x 9.1 x 35.4 in.)
50 x 50 x 154 cm
(19.7 x 19.7 x 60.6 in.)
0.4 kg
(0.9 lb)
4.3 kg
(9.5 lb)
4.3 kg
(9.5 lb)
9 kg
(19.8 lb)
40 kg
(88.2 lb.)
20 kg
(44.2 lb.)
12,8 kg
(28.2 lb.)
Scissor arm
•Equipped with gas jack specially designed for this application; proven to
function correctly after more than 400,000 cycles.
Electromagnetic compatibility
•KODAK 2200 Intraoral System complies with the European Directive
89/336/EEC and the IEC 60601.1.2 (2001) standard.
Classification: Group 1, Class B.
5-2SM701_K2200_2_en1 August 2006
X-ray Generator
Rated high voltage70 kV70 kV
Rated anodic power490 W490 W
Table 12.Main characteristics of the x-ray generator
KODAK-TROPHY type TRX 708 CEI type OCX/65-G
Maximum heat accumulated in the
anode
Rated value of focal spot (IEC
60336/1993)
Target materialsTungstenTungsten
Target slope19°19°
Filtration due to fixed materials0.6 mm (0.023 in.) eq. Al0.6 mm (0.023 in.) eq. Al
Figure 13.Heating and cooling curves for KODAK-TROPHY TRX 708
8,700 J10,000 J
0.7 mm (0.027 in.)0.7 mm (0.027 in.)
Joules
Cooling
mn
tube
kJ
min
Figure 14.Heating and cooling curves for CEI OCX/65-G tube
1 August 2006SM701_K2200_2_en5-3
Equipped X-ray Generator
IEC standard 60601-2-28 (1993)Conform
Type of protection against electric shocksClass I
Degree of protection against electric shocksType B
Rated value of inherent filtration1.5 mm (0.059 in.) eq. Al
Rated value of additional filtration1.0 mm (0.039 in.) eq. Al
Rated value of total filtration2.5 mm (0.098 in.) eq. Al
Beam-limiting cone, focal spot/skin distance20 cm (7 7/8 in.)
Maximum accumulated heat32,500 J
Maximum continuous thermal dissipation7 W
Table 15.Equipped x-ray generator
Amount of leaking radiation at maximum rate
during one hour of use
Maximum field of symmetrical radiation6 cm (2 3/8 in.) diameter
Position and tolerances of the focal point on the
reference axis
< 0.25 mGy
0 mm +/-0.5 mm (0.020 in.)
Figure 16.X-ray generator
Reference axis
Target angle
Focal point
5-4SM701_K2200_2_en1 August 2006
Energy in Joules
Cooling
Time in minutes
Figure 17.Heating and cooling curves of the KODAK 2200 system
tube head
1 August 2006SM701_K2200_2_en5-5
Position of Identification Labels
Figure 18.Identification labels
WARNING: MAGNETIC AREA
Machine identificationX-ray emitting unit
5-6SM701_K2200_2_en1 August 2006
Tables of Exposure Times
The following exposure time tables were established with a unit equipped
with a standard cone that corresponds to a focus-to-skin distance of 20 cm
(8 inches).
Table 19.Exposure times in seconds for type 6 films
70kV - 7mACone 20 cm (8 in.) 60kV - 7mACone 20 cm (8 in.)
MaxillaryChildAdultMaxillaryChildAdult
Incisor0.120.18Incisor0.240.37
Bicuspid0.120.18Bicuspid0.240.37
Premolar0.150.23Premolar0.310.47
Molar0.180.27Molar0.360.54
MandibularMandibular
Incisor0.100.15Incisor0.190.29
Bicuspid0.100.15Bicuspid0.190.29
Premolar0.110.16Premolar0.220.33
Molar0.120.18Molar0.240.37
BitewingBitewing
Anterior0.100.15Anterior0.190.29
Posterior0.120.18Posterior0.240.37
Occlusal0.200.31Occlusal0.400.61
Table 20.Exposure times in seconds for type 6 sensors
70kV - 4mACone 20 cm (8 in.)60kV - 4mACone 20 cm (8 in.)
MaxillaryChildAdultMaxillaryChildAdult
Incisor0.060.10Incisor0.130.19
Bicuspid0.060.10Bicuspid0.130.19
Premolar0.080.12Premolar0.160.25
Molar0.090.14Molar0.190.29
MandibularMandibular
Incisor0.050.08Incisor0.100.15
Bicuspid0.050.08Bicuspid0.100.15
Premolar0.060.09Premolar0.110.17
Molar0.060.10Molar0.130.19
BitewingBitewing
Anterior0.050.08Anterior0.100.15
Posterior0.060.10Posterior0.130.19
Occlusal0.110.16Occlusal0.210.32
1 August 2006SM701_K2200_2_en5-7
Note
If you use a long cone (30 cm or 12 in.),
increase exposure times by changing film
type to 3 steps higher.
Other filmsSpeed class F2 or 3
Other filmsSpeed class E4 or 3
Other filmsSpeed class D6 or 7
Table 23.Choice of sensor type (RVG mode 4 mA)
Note
If you use a long cone (30 cm or 12 in.),
increase exposure times by changing film
type to 3 steps higher.
ManufacturerNameSensor type setting
KodakRVG 50007
TrophyRVG Access7
KodakRVG 60006
TrophyRVG Ultimate6
TrophyRVG Reference Hi Resol mode6
TrophyRVGui Hi resol mode6
TrophyRVG THD4
TrophyRVG Reference Hi Sensi mode1
TrophyRVGui Hi Sensi mode1
Other brandsAdjust your sensor type*
5-8SM701_K2200_2_en1 August 2006
Emitted Doses
Table 24.Dose measured at extremity of 20 cm (8 in.) cone: film mode
70 kV—7 mA60 kV—7 mA
To obtain the dose in mGy.cm2, multiply values in the tables by the exposed
surface. The exposed surface depends on the collimator that is used and is
indicated in Table 28 on page 5-13.
Table 28.Exposure surface versus type of collimator used
Collimator type Format (mm)Used with
digital sensor
A19 x 24Size 0-4.6
B23 x 35Size 1Size 0 22 x 358.3
C31 x 39Size 2Size 1 24 x 40
Standard cone60 mm
-Size 3 27 x 54
diameter
Used with
film
Size 2 31 x 41
Size 4 57 x 76
Exposure
surface (cm
12.1
28.3
2
)
1 August 2006SM701_K2200_2_en5-13
EASTMAN KODAK COMPANY
343 State Street
Rochester, N.Y. 14650
Kodak, RVG, Trophy, Ultra-Speed, and Insight are trademarks of
Eastman Kodak Company. Other product or company names
mentioned in this manual are used only for the purpose of
identification and may be trademarks of their respective owners.