8. Non Contact Voltage (NCV): > 25V AC
SYMBOLS USED ON LCD
~
AC Measurement DC Measurement
Negative DC Value
-
Overload: Range Exceeded
O
.L.
Low Battery
Minimum Reading
MIN
Duty Cycle Mode Frequency Mode
%
Voltage Measurement
V
Resistance in Ohms Diode Test
Ω
Capacitance in Farads Continuity Test
F
Relative / Zero Mode
-6
Micro 10
µ
3
Kilo 10
k
ELECTRICAL SPECIFICATIONS
DC Voltage Measurement
Range Resolution Accuracy
600mV ˜ 600V 0.1mV ˜ 0.1V ± (0.5% + 4 digits)
1000V 1V ± (0.8% + 10 digits)
Overload Protection: 1000V
AC Voltage Measurement
Range Resolution Accuracy
600mV ˜ 750V 0.1mV ˜ 1V ± (2.0% + 5 digits)
Overload Protection: 750V RMS
Frequency Response: 45 to 400 Hz
Minimum Voltage for Frequency Measurement: 200mV
Response: True RMS
Auto Range Active
AT
Auto Power-Off Active
Apo
HOLD
Hold Active
Maximum Reading
M
AX
Current in Amps
A
Nano 10
n
Milli 10
m
Mega 10
M
AC Current Measurement
Range Resolution Accuracy
600µA 0.1µA ± (2.0% + 5 digits)
2000µA 1µA ± (1.5% + 5 digits)
60A 0.01A ± (2.9% + 15 digits)
600A ~ 1000A 0.1A ~ 1A ± (1.9% + 8 digits)
Overload Protection:
• Voltage: 600V RMS
• Current: 2000µA (leads), 1000A (clamp)
Frequency: 45 to 400Hz
Minimum Current for Frequency Measurement: 400µA (leads) or 20A (clamp)
Response: True RMS
Resistance Measurement
Range Resolution Accuracy
600Ω ˜ 6MΩ 0.1Ω ˜ 0.001MΩ
60MΩ 0.01MΩ
Overload Protection: 600V RMS
Capacitance Measurement
Range Resolution Accuracy
60nF ˜ 6000µF 0.01nF ˜ 1µF ± (3.5% + 6 digits)
-9
-3
6
Overload Protection: 600V RMS
Frequency Measurement
Range Resolution Accuracy
99.99Hz ˜ 499.9kHz 0.01Hz ˜ 100Hz ± (0.1% + 4 digits)
Overload Protection: 600V RMS
Duty Cycle Measurement
Range Resolution Accuracy
0.1 ˜ 99.9% 0.1% ± (0.2% per kHz + 0.1% + 5 digits)
Overload Protection: 600V RMS
Diode Test
Overload Protection Range Test Current Open Circuit Voltage
600V RMS 6.0V Appx. 0.25mA< 1.8V DC
Continuity Test
Overload Protection Open Circuit Voltage Threshold (Appx.)
600V RMS
<
0.44V
± (1.0% + 4 digits)
± (2.0% + 4 digits)
~40Ω
Turn instrument off and disconnect test leads. Clean the instrument by using
a damp cloth. Do not use abrasive cleaners or solvents.
Remove the batteries when instrument is not in use for a prolonged period of
time. Do not expose to high temperatures or humidity. After a period of storage
in extreme conditions exceeding the limits mentioned in the Specifications
section, allow the instrument to return to normal operating conditions before
using it.
Caution: This symbol indicates that equipment and its accessories
shall be subject to a separate collection and correct disposal.
WARRANTY
www.kleintools.com/warranty
CLEANING
STORAGE
DISPOSAL / RECYCLE
Instruction Manual
ENGLISH
• 1000A AC/DC
• TRUE RMS
• AUTO / MANUAL
RANGE
• NON CONTACT
VOLTAGE TESTER
• MAX / MIN / HOLD
• ANALOG BAR GRAPH
• 3-5/6 DIGIT
5999 COUNT LCD
• BACKLIGHT
• WORKLIGHT
CL2500
DC Current Measurement
Range Resolution Accuracy
600µA 0.1µA
2000µA 1µA
60A 0.01A ± (2.5% + 15 digits)
600A ~ 1000A 0.1A ~ 1A ± (1.5% + 8 digits)
Overload Protection:
• Voltage: 600V RMS
• Current: 2000µA (leads), 1000A (clamp)
± (1.2% + 3 digits)
Non Contact Voltage Detector
On Voltage
Appx. 25V AC and above
CUSTOMER SERVICE
KLEIN TOOLS, INC.
450 Bond Street
Lincolnshire, IL 60069
www.kleintools.com
139677 Rev. 11/13 B
ENGLISH
CL2500
Instruction Manual
GENERAL SPECIFICATIONS
The Klein Tools CL2500 is a True RMS, auto-ranging clamp meter. It measures
AC / DC voltage, AC / DC current, resistance, capacitance, frequency, and duty
cycle. It can also test non-contact voltage, diodes, and continuity.
• Operating Altitude: 2000 meters
• Storage Altitude: 10000 meters
• Humidity: 0% to 80% at 32°F to 95°F (0°C to 35°C)
• Operating Temperature: 0°C / 32°F to 40°C / 104°F < 75% R.H.
• Storage Temperature: -20°C / -4°F to 60°C / 140°F
• Accuracy Temperature: 18°C / 64°F to 28°C / 82°F
• Temperature Coefcient: 0.1*(specified accuracy) / °C
• Dimensions: 8.09" x 2.56" x 1.38" (205 mm x 65 mm x 35 mm)
• Weight: 14 oz. (400 g)
• Calibration: Accurate for one year
• Certications: UL & cUL standard UL3111-1 Listed
• Pollution Degree: 2
• Accuracy: ± (% of reading + # of least significant digits)
• Ingress Protection: IP42
• Drop Protection: 1 m (3 ft.)
• Safety Rating: CATIII 1000V, CAT IV 600V
0% to 70% at 32°F to 131°F (0°C to 55°C)
WARNINGS
To ensure safe operation and service of the tester, follow these instructions.
Failure to observe these warnings can result in severe injury or death.
• Before each use, verify meter operation by measuring
a known voltage or current.
• Never use the meter on a circuit with voltages that exceed the
category based rating of this meter.
• Do not use the meter during electrical storms, or in wet weather.
• Do not use the meter or test leads if they appear to be damaged.
• Ensure meter leads are fully seated, and keep fingers away from
the metal probe contacts when making measurements.
• Do not open meter to replace batteries while probes are connected.
• Use caution when working with voltages above 60V DC,
or 25V AC RMS. Such voltages pose a shock hazard.
• To avoid false readings that can lead to electrical shock,
replace batteries if a low battery indicator appears.
• Unless measuring voltage or current, shut off and lock out
power before measuring resistance or capacitance.
• Always adhere to local and national safety codes. Use individual
protective equipment to prevent shock and arc blast injury where
hazardous live conductors are exposed.
SYMBOLS
AC Alternating Current Warning or Caution
~
DC Direct Current Dangerous levels
DC/AC Voltage or Current Double Insulated Class II
Ground
AC Source
Safe for disconnect
from live conductors
FEATURE DETAILS
1
B
8
E
J
F
G
D
4
H
C
I
4
K
A
A. Use CAT IV 600 rated leads or higher.
Do not attempt to measure more than 1000V DC or 750V AC or 2000µA.
B. Keep hands below line when measuring high current levels.
C. Auto Power-Off (Apo)
• Device will power off after 30 minutes non-use.
• Turn the dial or press a button to wake.
• Disabled during Max / Min function.
• Holding Select button while turning on disables Auto Power-Off.
D. Select Functionality Button
• Quick press (<0.5 sec):
» Switch between AC and DC.
» Switch between
• Long press (>1 sec):
» Switch between AC, Frequency, and Duty Cycle modes.
» Zero the DC current clamp reading.
E. Hold / Backlight / Worklight
• Press to hold the current input on the display.
• Press again to return to live reading.
• Press for 2 seconds to enable / disable lights.
• Using lights decreases battery life.
F. Auto / Manual Range
• Press repeatedly to cycle through manual ranges.
• Press for 2 seconds to return to auto ranging mode.
•
AT is displayed on LCD only during auto ranging mode.
G. Max / Min Hold
• Press to enter Max / Min mode; the largest and smallest
values will be saved while in this mode.
• Press repeatedly to alternate between the maximum and minimum readings.
• Press for 2 seconds to return to live reading and clear
the stored maximum and minimum values.
and .
H. I. Battery Replacement
• When indicator is displayed on the LCD, batteries must be
replaced.
• Remove the back screw and replace 2 x AAA batteries.
J. Magnetic Back
• Attach instrument to metal for hands-free use.
K. Bar Graph
• The bar graph shows an approximate analog representation
of a measurement.
• The bar graph responds much faster than the digital display.
• The scale of the bar graph is zero to the maximum reading of
the selected range.
FUNCTION INSTRUCTIONS
1
2
6
7
5
3
1. AC / DC Current (large): < 1000A
Features:
• Center wire in guides for best accuracy.
• Opposing currents cancel. Use line-splitter (USA only) when necessary.
2. AC / DC Current (small): < 2000µA
Features:
• Select AC or DC current source.
Current above 2000µA will damage instrument.
•
3. AC / DC Voltage: < 750V AC or 1000V DC
Features:
• Select AC or DC voltage source.
MAX/MINRANGEHOLD
ZERO (DC)
MAX/MINRANGEHOLD
MAX/MINRANGEHOLD
4. Frequency (Hz) / Duty Cycle (See FEATURE DETAILS)
Features:
• Select , , or setting.
5. Resistance: < 60MΩ
Features:
•
6. Capacitance: < 6000µF
Features:
• Safely discharge capacitor before measurement.
• Reading may take up to 60 seconds for large capacitors.
7. Diode / Continuity
HOLD
HOLD
Do not measure resistance on a live circuit.
HOLD
MAX/MIN
RANGE
Diode Features:
HOLD MAX/MIN
Display shows:
• Forward voltage drop if forward biased.
• “O.L.” if reverse biased.
Continuity Features:
HOLD MAX/MIN
• Display shows resistance.
• Buzzer sounds if less than 40
Ω.