Kipp&Zonen COMBILOG User Manual

COMBILOG 1022
Data Logger
Hardware Manual Version 1.04
Issue: 22.01.2010
Technical data are subject to change!
Hardware Manual COMBILOG 1022
2
Copyright 1995-2010, Theodor Friedrichs & CO
Copyrights: Operating instructions, manuals and software are
subject of copyright. Copying, duplication, translation, conver­sion into any electronic medium or any machine readable form, as a whole or in parts, is not permitted, with the exception of making a back-up copy of the software for saving purposes, in­sofar as this is technically feasible and is recommended by our company. Contraventions will lead to compensation.
Limitation of Liability: No liability is assumed by Theodor Friedrichs for damages and/or injury resulting from use of equipment supplied by this company. In no event will Theodor Friedrichs be liable for indirect or consequential damages what­soever resulting from loss of use, data or profits arising out of connection with the use of performance of Theodor Friedrichs products. Theodor Friedrichs products are not designed, inten­ded, or authorized for use as components and medical sys­tems, or other applications indeed to support or sustain life, or for any other application in which the failure of the Theodor Friedrichs product(s) could create a situation where personal injury or death may occur. Any claims against Theodor Friedrichs in connection with the hardware and software products described in this manual can exclusively be based on the guarantee regulations. Any further claims are excluded, in particular Theodor Friedrichs does not give any guarantee as to the correctness of the contents of this manual. Changes are subject to alteration and can be executed any time without advanced notice.
Trade Marks: Without going into details, we want to point out the usage of indications and entered trade marks, in particular the indications and trade marks of Microsoft Corporation, In-
ternational Business Machines Corporation and Intel Corpora­tion.
________________________________________
Hardware Manual COMBILOG 1022
3
Hardware Manual COMBILOG 1022
4
CHAPT ER S UR VE Y
Page
1 GENERAL PRELIMINARY REMARKS...............10
2 SYSTEM DESCRIPTION........................................12
3 INSTALLATION.......................................................19
4 SIGNAL PROCESSING..........................................28
5 FUNCTIONAL DESCRIPTION..............................34
6 DISPLAY / MENU OPERATION...........................60
7 DATA STORAGE.....................................................74
8 MASTER FUNCTION..............................................81
9 INITIATION AND TEST..........................................84
10 STRUCTURE OF THE BUS TOPOLOGY...........86
11 COMMUNICATION..................................................99
12 WEBSERVER.........................................................144
13 SPECIFICATIONS.................................................147
14 SIMPLIFIED DRAWINGS....................................154
Hardware Manual COMBILOG 1022
5
TABLE O F CO NT EN TS
Page
1 GENERAL PRELIMINARY REMARKS...............10
1.1 About this Manual....................................................................10
1.2 Important Notice......................................................................10
1.3 Contact for Inquiries................................................................11
2 SYSTEM DESCRIPTION........................................12
2.1 System Overview.....................................................................12
2.2 Range of Application...............................................................13
2.3 Features..................................................................................15
2.4 Configuration Software............................................................17
3 INSTALLATION.......................................................19
3.1 Mounting / Fixing.....................................................................19
3.2 Protective System....................................................................19
3.3 Ambient Temperature..............................................................20
3.4 Front panel / Pin Assignment..................................................20
3.5 Connection .............................................................................22
3.6 Power Supply..........................................................................23
3.7 Bus Connection.......................................................................25
3.8 Sensor Connection..................................................................26
3.9 Several Sensors at one Data Logger......................................27
3.10 Module Jack............................................................................27
4 SIGNAL PROCESSING..........................................28
4.1 Analogue Inputs.......................................................................28
4.2 Digital Inputs/Outputs..............................................................28
4.3 Power Switch...........................................................................29
4.4 Internal Reference Voltage, Offset- and Drift Correction.........30
4.5 Internal Processing..................................................................31
4.6 Scan Rate and Power Consumption.......................................32
4.7 Signal Processing....................................................................33
Hardware Manual COMBILOG 1022
6
5 FUNCTIONAL DESCRIPTION..............................34
5.1 Analogue Input Channel..........................................................35
5.1.1 Voltage Measurement.............................................................36
5.1.2 Current Measurement..............................................................38
5.1.3 Resistance Measurement........................................................40
5.2 Digital Input Channel...............................................................45
5.2.1 Digital Status Recording..........................................................46
5.2.2 Frequency Measurement.........................................................47
5.2.3 Progressive Counter................................................................49
5.3 Digital Output Channel............................................................50
5.3.1 Digital Status Output:...............................................................50
5.4 Arithmetic Channel..................................................................52
5.5 Setpoint Channel.....................................................................58
5.6 Alarm Channel.........................................................................58
5.7 Serial Channel.........................................................................58
5.8 Threshold Values.....................................................................58
5.9 Error Handling.........................................................................59
6 DISPLAY / MENU OPERATION...........................60
6.1 Display and Operation.............................................................60
6.2 Menu Items..............................................................................60
6.3 SD-Card...................................................................................72
6.3.1 Remove Card..........................................................................72
6.3.2 Firmware Update.....................................................................73
7 DATA STORAGE.....................................................74
7.1 General Remarks to Data Storage..........................................74
7.2 Modes of Data Storage............................................................74
7.3 Storage Medium......................................................................75
7.4 Internal Data Storage..............................................................76
7.5 External Data Storage with SD Card.......................................78
8 MASTER FUNCTION..............................................81
8.1 Master Function.......................................................................81
Hardware Manual COMBILOG 1022
7
9 INITIATION AND TEST..........................................84
9.1 Before Connecting the Device.................................................84
9.2 After Connecting the Device....................................................84
9.3 Configuration of the Data Logger............................................84
10 STRUCTURE OF THE BUS TOPOLOGY...........86
10.1 Bus Interface...........................................................................87
10.2 Bus Structure...........................................................................88
10.3 Transmission Speed and Line Length.....................................89
10.4 Bus Cable................................................................................90
10.5 Bus Plug..................................................................................91
10.6 Bus Termination......................................................................92
10.7 Shielding..................................................................................94
10.8 PC Bus Connection.................................................................95
10.9 Potential Equalization..............................................................96
10.10 Adjustment of Address and Baud Rate...................................96
11 COMMUNICATION..................................................99
11.1 Bus Interface...........................................................................99
11.2 Bus Protocol............................................................................99
11.3 Data Format...........................................................................100
11.4 Output Format.......................................................................101
11.5 Transmission Sequence........................................................104
11.6 ASCII protocol.......................................................................106
11.6.1 Telegram Format for the ASCII Protocol...............................106
11.6.2 Instruction Set in the ASCII-Protocol.....................................109
11.7 PROFIBUS protocol..............................................................116
11.7.1 Telegram Format for the Profibus Protocol...........................116
11.7.2 Instruction Set in the PROFIBUS-Protocol............................120
11.8 MODBUS protocol.................................................................128
11.8.1 Telegram Format for the MODBUS-RTU Protocol................128
11.8.2 Instruction Set in MODBUS-RTU Protocol............................129
11.9 Sample Program....................................................................140
11.10 Autocall Function...................................................................141
Hardware Manual COMBILOG 1022
8
11.11 Modem Connection...............................................................143
12 WEBSERVER.........................................................144
12.1 General..................................................................................144
12.2 System Info............................................................................144
12.3 System Log............................................................................144
12.4 Data View..............................................................................144
12.5 Data Logger...........................................................................145
12.6 Configuration / Login.............................................................145
12.7 Single Channel Configuration................................................145
12.8 Power Switch.........................................................................145
12.9 Serial Channel.......................................................................145
12.10 Logger Configuration.............................................................146
12.11 Password...............................................................................146
12.12 Protocol.................................................................................146
12.13 Logout....................................................................................146
13 SPECIFICATIONS.................................................147
13.1 Power Supply........................................................................147
13.2 Signal Inputs/Outputs............................................................147
13.3 Signal Processing..................................................................147
13.4 Analogue Inputs (8 per Module)............................................148
13.5 Digital Inputs/Outputs (6 per Module)....................................149
13.6 Interfaces...............................................................................149
13.7 Operating Conditions.............................................................151
13.8 Electromagnetic Compatibility...............................................151
13.9 Shell.......................................................................................152
13.10 Circuit....................................................................................152
13.11 Accessories / Notice for Orders.............................................152
14 SIMPLIFIED DRAWINGS....................................154
14.1 Front View.............................................................................154
Hardware Manual COMBILOG 1022
9

1 GENERAL PRELIMINARY REMARKS

1.1 About this Manual

The manual contains all important information concerning the function, installation and initiation of the data logger
COMBILOG 1022.
The description of the configuration software for the COMBILOG-System is available as Online-Help within the con­figuration software COMBILOG.EXE.

1.2 Important Notice

Make sure to use the data logger COMBILOG 1022 exclusively in accordance with the notices, technical data and operating conditions mentioned in this manual. In case of inexpert hand­ling or wrong application possible disturbances, measuring er­rors, effects on or from other appliances and facilities as well as possible endangering of human lives or tangible assets can­not be excluded!
Therefore if you have not yet operated the data logger COMBILOG 1022, you should first of all study this manual thoroughly. While initiating or operating the appliance or in case service is required always observe the notices given in this manual.
Please note further that there are other special regulations to be observed in case of application in potentially explosive sur­roundings (EExe, EExi, ...). These, however are not subject of
Hardware Manual COMBILOG 1022
10
this manual, which only explains the general use of the data logger COMBILOG 1022.

1.3 Contact for Inquiries

In case of inquiries concerning the data logger COMBILOG 1022 please contact your local distributor or directly Theodor Friedrichs & Co. GmbH.
Hardware Manual COMBILOG 1022
11

2 SYSTEM DESCRIPTION

2.1 System Overview

The COMBILOG 1022 is a data logger with compact design, combined with integrated LC-display and memory slot suitable for SD cards. This data logger was developed for meteorological, hydrologic­al and environmental measuring systems, but it is equally suit­able for countless further applications in industrial production. The COMBILOG 1022 features high performance, compact design (SMD components), low power consumption and mod­erate price.
The data logger is equipped with 8 analogue and 6 digital measuring channels; further channels for numeric calculation may be configured. Four serial interfaces RS232, RS485, USB and Ethernet are built-in, featuring communication via ASCII, PROFIBUS or MODBUS. Data storage is achieved by internal Flash or SD memory card, optionally. „SELECT“ switch and 4-line LCD on the front panel allow to enter or modify a number of different modes and functions, such as scan rate and averaging time, as well as offset or gain. The COMBILOG 1022 can easily be mounted on a 35 mm standard rail using its „snap-in“ clamp and is therefore suitable for control cabinet installation or similar. Thanks to its low power consumption, battery supplied systems are possible, whereby the use of a solar panel enables any ex­tension of measuring period.
Hardware Manual COMBILOG 1022
12
For applications like outdoor use there is a version with stain­less steel housing available, as well as various other accessor­ies. Configuring of the data logger is accomplished by means of an easy to handle WINDOWSTM 98 / ME / NT / 2000 / XP software.

2.2 Range of Application

As described under (2.1), the most varying measurement tasks can easily be accomplished by means of the COMBILOG 1022. Some typical applications are e.g. measurement of temperat­ure via resistance thermometers (Pt100), operation with com­bined sensors with current- or voltage output (e.g. wind speed measurement with DC generator) or position measurement and weight measurement by displacement transducers and force transducers. With these applications the data logger COMBILOG 1022 supports measuring methods with 2-, 3- and 4-wire technique. The signal processing required in accordance with the sensors used, such as gain, linearisation, offset cor­rection etc. can be adjusted individually by software. An extern­al amplifier is not required. The digital signal inputs can be used, for example, to connect switches, initiators, digit emitters and oscillators. Thus status indications can be collected and tasks like e.g. position meas­urements, displacement measurements, angular measure­ments, frequency measurements and timings can be carried out. Furthermore special 8-bit-Graycode-transmitters can be connected. Special calculations of measured values are possible by arith­metic channels. In case the 8 analogue and 6 digital inputs are not sufficient, other modules can be connected to the data
Hardware Manual COMBILOG 1022
13
logger via the RS485 bus. In this case the COMBILOG is used as a bus master to read the measured values from the slave modules.
All data can be transmitted via the integrated RS485 commu­nication interface to a subsequent control (PLC) or to a com­puter (PC). Up to 127 modules can be connected with the two­wire line over distances of several km. At the same time the communication interface features programming and configuring the individual application from a PC. If the data logger COMBILOG 1022 is not integrated in a bus, an additional RS232-interface is available for the user. This in­terface allows only a point-to-point connection up to max. 20 m (65 feet), but all functions of the RS485-interface remain avail­able. Furthermore the data logger can send messages in case of user definable conditions automatically via modem or SMS. A configuration program is included (requires Microsoft WINDOWS 98,ME,NT,2000, XP).
Hardware Manual COMBILOG 1022
14

2.3 Features

Function:
Measurement inputs for all common types of sensors for I,
U and R.
Several different sensors can be connected simultan-
eously
Measured values monitored by programmable thresholds Detection of sensor errors Detection of communication errors Programmable error handling Calculation of average values, minima, maxima, standard
deviation and other arithmetic functions
7 MByte Flash internal data storage, extendible with ex-
ternal SD card
Inputs and outputs:
8 analogue inputs (for 2-, 3- and 4-wire connection) 6 digital inputs/outputs (I/O ports), configurable
Power supply:
Power supply: +10 ... +30 VDC All connections protected against excess voltage, excess
current and reverse polarity
Battery operation is possible due to low power consump-
tion
Display and operation:
LED-status indication for digital inputs/outputs LED-status indication for malfunction and operation (ERR /
RUN)
Hardware Manual COMBILOG 1022
15
LED-status indication for the ethernet interface LC-display (4 x 16 characters) and push-/turn knob for op-
eration
Measured value processing:
Linearisation, scaling and conversion into physical units Option to adjust, modify or reset the processing paramet-
ers individually
Master function to retrieve data from external modules Programmable averaging Automatic message transmission via modem or SMS Non-volatile storage for program, parameters and data
Configuration:
Configurable with PC-software under
WINDOWS™ 98 / ME / NT / 2000 / XP
Menu-guided sensor selection in plain text Free configuration of up to 32 channels Data base for the most common sensors Definition of user-specific sensors Setting of type and principle of measurement Display of pin assignment Input of linearisation Alarm settings Programmable error handling Arithmetic combination of sensor channels Configurable measuring rate and averaging interval Configuration on file (offline-operation) Configuration via bus (online-operation)
Programming:
Allocation of address and baud rate via bus Password to save the configuration and the data memory Synchronizing of date and time with the host PC
Hardware Manual COMBILOG 1022
16
Communication:
Integrated RS 485, RS 232, USB and Ethernet communic-
ation interface
Autonomous function independent of subsequent systems Definition of the transmission protocol (ASCII and
PROFIBUS or MODBUS)
Definition of the telegram format (baudrate and parity) Definition of the output format
(field length / decimals / unit)
Simple instruction set
Shell:
Compact structural shape Attractive design Fast mounting Snap-on mounting on DIN rail 35 mm / 1.4 inch Protection IP20 Plug-in screwed terminals Module jack, ground connection

2.4 Configuration Software

The COMBILOG 1022 is delivered with a configuration soft­ware for MS WINDOWS 98/ME/NT/2000/XP. This software al­lows the individual configuration of the data logger. Measuring channels are defined as variables in a variable table. Predefined sensors can be selected from the integrated data­base. The linearisation of the sensor signals will be performed automatically. Additional sensors can be defined. Additional parameters like scan rate, averaging interval, data recording, error handling, automatic message generation, mas­ter function etc. are configurable. A password enables protec­tion of the configuration and the stored data. Instantaneous measured values can be watched directly.
Hardware Manual COMBILOG 1022
17
In case a software update for the data logger is necessary, the configuration program provides a download function, that sends the new program to the logger. Communication is supported via standard interface (RS232), telephone or GSM modem or TCP/IP protocol.
Example for a configuration:
Hardware Manual COMBILOG 1022
18

3 INSTALLATION

3.1 Mounting / Fixing

The data logger COMBILOG 1022 has a snap-on mounting for installation on standard profile rails 35 mm (1.4 inch) according to DIN EN 50022.
Installation on the DIN rail is performed by means of the four straps on the rear side of the data logger. First push the two straps on the bottom behind the notch of the DIN rail and then press the data logger on the DIN rail until the two straps on the top snap in.
In order to take the data logger off the DIN rail slide the module lateral off the rail or in case it is not possible lift the data logger slightly so that the straps on the top get off the notch and the data logger can be taken off easily by pulling.
Attention: Refer to protection earth hints in chapter 3.5!

3.2 Protective System

The data logger has an IP20 protective system. For outdoor in­stallations data logger COMBILOG 1022 can be installed in a stainless housing type 9920, thus featuring IP66 standard.
Hardware Manual COMBILOG 1022
19

3.3 Ambient Temperature

The admissible ambient temperature for the data logger COMBILOG 1022 is -45 °C to +85 °C. The admissible ambient temperature for the LC-Display is between -20°C and +60°C and becoming very slow below 0°C. Attention: For certain memory card types, differing temperat-
ure ranges have to be considered.

3.4 Front panel / Pin Assignment

The front panel of the data logger COMBILOG 1022 shows fol­lowing elements:
Figure 3.1 Front panel
Hardware Manual COMBILOG 1022
1
14 12 11 10 9
87
6
5432
1516 139
20
Description of the parts:
Number Description Number Description
1 USB
9
Module jack connection
2 Ethernet 10 LED ERR (red)
3 Voltage supply 11 LED RUN (green)
4 Power switch 12 Ethernet Link / Traffic
5 RS485 13 Press/rotary knob
6 RS232
14
8 analogue inputs
7 6 digital I/Os
15
LC-display
8 Status-LEDs for digital I/Os
16
Interface for memory card
Table 3.1 Description of the parts on the front of the device
Pin assignment:
Terminal Assignment Terminal Assignment
+10..30V Voltage supply + I/O + Digital I/O 1…6
0V Voltage supply - 0 V Ground for digital I/Os
A RS485-bus interface A SOURCE Source output
B RS485-bus interface B In+ Analogue input +
RX RS232 receive In- Analogue input -
TX RS232 transmit
AGND
Ground for analogue input
COM RS232 ground
UOUT 1 / 2
Power switch output
Table 3.2 Pin assignment
Hardware Manual COMBILOG 1022
21

3.5 Connection

Connection: plug-in screw terminals Nominal cross section: 1.5 mm² (0.02 square inch)
unifilar/fine-strand (AWG 16)
Length of wire stripping : 6 mm (0.2 inch) Alternatively available: LP-terminals (spring loaded)
(upon request)
Protection earth: 6.3mm series tabs at rear
side of housing
The best way to pull off the screw terminals is to use a small screwdriver, placed as a lever between terminal and the front of the data logger.
Not more than 2 leads should be connected with one clamp. In this case the leads should have the same conductor cross sec­tion.
Note: Wire connection is only allowed during power off. Note: In order to avoid influences from noise on the sensor sig-
nals shielded wires should be used for the power supply, the bus connection and the signal lines.
ATTENTION: Before final installation, a suitable protection earth cable with terminal has to be connected to the ground connector at the back of the data logger. Assure that the con­nection has a low impedance after mounting.
Hardware Manual COMBILOG 1022
22

3.6 Power Supply

Figure 3.2 Connection of the distribution voltage
Voltage range
+10 ... +30 VDC
Power input
0.1 W typical (up to 1.1W maximum, depending on config-
uration)
Internal protector (reversible)
excess current 0.5 A M excess voltage
Hardware Manual COMBILOG 1022
U+
U-
Versorgungsspannung
Power Supply
23
Non-regulated DC voltage between +10 and +30 VDC is suffi­cient for the power supply of the data logger COMBILOG 1022. The input is protected against excess voltage and current and against reverse polarity. The power consumption remains ap­proximately constant over the total voltage range, due to the in­tegrated switching controller.
Due to its low current consumption (max. 110 mA at 12 VDC) the data logger can also be remote-fed via longer lines. Several data loggers can be supplied parallel within the admissible voltage range, considering the voltage drop in the lines. The supply lines can also be installed in one common cable, togeth­er with the bus line, if required.
In order not to charge the data logger’s supply voltage unne­cessarily, a separate power supply for sensors with a large cur­rent requirement is recommended.
Hardware Manual COMBILOG 1022
24

3.7 Bus Connection

In general the data logger is connected to the bus by applying the signal leads A and B of the incoming bus cable and A' and B' of the outgoing bus cable together to one terminal on the module (figure 3.3).
Alternatively the bus can also be connected by a "stub cable" as shown in figure 3.4.
Owing to the removable terminal, the bus connection to other data loggers remains valid, even if one data logger is replaced by another.
Figure 3.3 Connection of the data logger to the bus
Note: When connecting the logger to the bus, the two bus in­terfaces A and B must not be interchanged.
Note: The stub cable should be as short as possible, not longer than 30 cm (12 inch).
Hardware Manual COMBILOG 1022
25
Figure 3.4 Connection of the data logger to the bus by a stub cable

3.8 Sensor Connection

The analogue and digital signal inputs and outputs are wired according to measurement task, to the transducer (sensor) that is used, and to the number of connected sensors. The pinout arrangements for the various types of measurement are de­scribed in chapter 5. The respectively valid pin assignment is determined by means of the configuration software. Since the digital outputs are "passive" the processing of extern­al elements always requires an external current supply. In case of larger loads this should be independent of the data logger supply. At the connection of inductive loads a connection with a diode is required in order to prevent possible damages by induced voltage.
Hardware Manual COMBILOG 1022
26
Following devices can be connected directly to the digital out­puts: signal lamps, small relays, switching relays for larger loads, acoustic signal installations, buzzer respectively beeper etc., as long as the connected loads are not exceeding the val­ues described in the technical data chapter 12.

3.9 Several Sensors at one Data Logger

The data logger COMBILOG 1022 can simultaneously receive and process sensor signals from several different sensors. As many sensors can be connected as there are analogue and di­gital signal inputs and outputs available (14 sensors max.; 8 analogue and 6 digital sensors).

3.10 Module Jack

The data logger COMBILOG 1022 has a bus connection facility on the left and on the right side of the housing, featuring inter­connection of the +10…30 VDC supply and the bus signal, for several COMBILOG. This kind of bus connection and of power supply is particularly advantageous if several data loggers are mounted on one com­mon profile rail side by side. In this case the connection via the terminals can be dropped, except for one module.
Note: If some COMBILOG 1022 are connected using the mod- ule jacks, the supply power has to be provided using the screw terminals. The power supply via the module jacks has to be used only for expansion modules.
Hardware Manual COMBILOG 1022
27
Note: It is necessary to take care that the current at the Module
Jack is not higher than permitted. Thus, the power supply preferably should be led to the centre of the module line. For the same reason a 100 mA self resettable fuse protects the module connector of the COMBILOG 1022 against short cir­cuits. The permanent current drawn from the module connector shall not be higher than 100 mA.
Hardware Manual COMBILOG 1022
28

4 SIGNAL PROCESSING

The data logger COMBILOG 1022 has eight analogue inputs and six digital inputs/outputs. Several different sensor signals as well as digital inputs and digital output signals can be con­nected and processed simultaneously.

4.1 Analogue Inputs

The analogue inputs serve to collect sensor signals, or to ac­quire control values respectively. They are particularly de­signed to measure voltages, currents and resistances.
There are 8 equal analogue inputs, each input can be con­figured individually.
Note: Overloads of more than ± 10 VDC will lead to false measuring results in the according analogue input channel. Overloads of more than ± 15 VDC will also have influence on the measuring accuracy of the other input channels!

4.2 Digital Inputs/Outputs

The six digital inputs/outputs of the data logger can be con­figured - independent of each other - as inputs or as outputs. The current status (in/out) is signalized by one LED each.
Hardware Manual COMBILOG 1022
29
As inputs the I/Os can be used for collecting feed-back signals, for measuring frequencies, as counters or for receiving special serial 8-bit-Graycode signals. Status information can be issued by the outputs. Thereby host-controlled or process-controlled status outputs are possible.
The digital inputs have an excess voltage protection (transil di­odes), with nominal threshold 30 V. Input voltages between
2.0 VDC and 30 DC are interpreted as logic LOW ("0"), input voltages lower than 0.9 V as logic HIGH ("1"). The maximum input current is 1.5 mA.
Figure 4.1 Definition of signal levels and logic levels
The outputs are open-collector type with a maximum voltage of 30 VDC and a maximum current of 100 mA.

4.3 Power Switch

The data logger Combilog 1022 has two power switches which act as voltage supplies for external sensors. A power switch can be assigned to an analogue channel and configured with a lead time (see chapter Power Switch ). The supply is switch on with a configurable time before the se­lected analogue input channel will be measured.
Hardware Manual COMBILOG 1022
Signal level Logical level
30

4.4 Internal Reference Voltage, Offset- and Drift Correction

An internal reference voltage serves to adjust the entire ana­logue signal processing automatically. Especially for measurement of extreme low voltages, currents and resistances, the configuration software features an addi­tional compensation of temperature drift. With current- and voltage measurement, this is realized by an internal offset measurement. The measured offset is subsequently applied to correct the measured values. For Maximum accuracy with cur­rent measurement (temperature drift less than 25 ppm/K) it is recommended to accomplish the measurement via an external shunt with a correspondingly low temperature coefficient (< 5 ppm/K). For this purpose, the input channel has to be con­figured as a voltage input. For resistance measurement, a drift correction requires an ad­ditional input channel which has to be equipped with a suitable reference resistance. This resistance should have a low tem­perature coefficient (< 5 ppm/K). In the configuration table, this channel has to be defined as a reference channel with resist­ance input for drift correction, whereby the nominal value of this resistance (at 20°C) has to be indicated. Using the above described methods, the analogue inputs can almost completely be kept free from temperature drift.
Hardware Manual COMBILOG 1022
31

4.5 Internal Processing

Next to collecting the analogue input signals, the analogue mul­tiplexer at the input of the circuit collects the internal reference voltage. All these values are then transmitted to the program­mable amplifier PGA, where the signals are amplified accord­ing to the kind and type of the connected sensors and then supplied to the A/D converter.
The A/D converter digitalises all incoming signals with a defini­tion of 16 bit and at a rate that can be preset for the module by the user. The Sigma-Delta-procedure used for the A/D-conver­sion guarantees a high accuracy and a high linearisation. The A/D-converter processes an integrated amplifier with the ampli­fier stages of 1, 2, 4, 8, 16, 32 and 64. For very small signals, the module switches to an additional amplifier with amplifier stages of 100, 200, 800 and 3200. The amplification in align­ment with the accuracy and resolution of the calculated meas­uring values results from the selection of the measuring range which will be configured by assistance of the configuration soft­ware.
This software also configures the ADC internal low pass filter, depending on the mains frequency and desired ADC frequency (selectable between 10 and 400 Hz). Standard configuration for the low pass filter is 50 Hz.
The microprocessor µP now edits the measuring signal in digit­al form. First the processor linearises and scales the signal and holds it ready for transmission via bus in programmable units. Further the processor monitors the measured values for excess of freely programmable threshold values. Thus a monitoring of
Hardware Manual COMBILOG 1022
32
failure or breaking of the sensing element or short-circuit can also be realised. The data logger can be activated - by means of appropriate configuration - to provide a corresponding signal at the digital I/O in case of alarm. The digital I/Os are directly addressed and monitored respectively by the microprocessor µP. Hereafter an arithmetical averaging of the values is carried out. The average interval is the same for each channel and is ad­justable in steps of 0.5, 1, 2, 3, 4, 5, 10, 15, 20 and 30 seconds, 1, 2, 3, 4, 5,10, 15, 20, 30 minutes, resp. 1, 2, 3, 4, 6, 8 and 12 hours. The calculated average values are finally stored in the memory. The special user program, the data for configuration, linearisa­tion and scaling etc. that are required by the processor µP for all tasks are retentively stored in a EEPROM. The timing control of the data processing is realised by an in­ternal real time clock buffered by Gold cap capacitor.

4.6 Scan Rate and Power Consumption

The A/D-converter digitalizes every signal at a rate that can be preset by the user. The scan rate can be selected between
0.25 sec. and 1 hour. In this selected time period all configured channels are scanned and processed correspondingly. The power consumption of the data logger COMBILOG 1022 de­pends on this scan rate. Between the measuring cycles the data logger will be set into a so-called “Sleep-Mode“. During this mode the data logger needs only about 20 mA with supply for the serial lines being active.
Note: A scan rate of 0.25 s causes an unsignificant higher vari­ation of instantaneous values but does not affect the averaging.
Hardware Manual COMBILOG 1022
33

4.7 Signal Processing

Arithmetical averaging is carried out using several measuring values. The averaging interval, which is the same for all chan­nels, can be set to one of the values 0.5, 1, 2, 3, 4, 5, 10, 15, 20, 30 seconds or 1, 2, 3, 4, 5, 10, 15, 20, 30 minutes or 1, 2, 3, 4, 6, 8, 12 hours respectively. The calculated values are fi­nally stored in the memory. With the configuration software the kind of averaging is select­able: Normal averaging or averaging of wind direction (con­sideres the discontinuity at NORTH). For counter variables not the average, but the number of pulses is calculated. A change of the average interval can be initiated by certain pro­gram conditions, thus featuring temporary higher time resolu­tion of measured signals.
Hardware Manual COMBILOG 1022
34

5 FUNCTIONAL DESCRIPTION

The data logger COMBILOG 1022 has a total of 32 logical channels for the collection, processing and output of various kinds of sensor information. These 32 channels can be con­figured as:
Analogue Input Channel Digital Input Channel Digital Output Channel Arithmetic Channel Setpoint Channel Alarm Channel
For each channel various kinds of channel information and pro­cessing functions can be determined. The table in appendix C gives a survey of the channel set-ups with the data logger COMBILOG 1022. The channel set-ups are carried out by means of the configuration software.
Hardware Manual COMBILOG 1022
35

5.1 Analogue Input Channel

The Analogue Input Channel collects and processes the sig­nals of the most common types of sensors. A large number of standardised sensors is already stored in the COMBILOG in­ternal sensor data base. Further sensors can be added by the user.
Following measuring principles are provided:
Voltage measurement Current measurement Resistance measurement Temperature measurement with thermocouples
For each of these principles the data logger COMBILOG 1022 offers several types of measurement. For voltage measure­ment the types of measurement single-ended and differential can be used. Currents up to 25 mA are directly measured by the data logger. Current measurements of more than 25 mA can be carried out by measuring voltage drop at an external shunt. Resistance measuring can be carried out in 2-, 3- and 4­wire technique. Each analogue input channel can be assigned an individual sample rate (see chapter Single Channel Configuration ).
Hardware Manual COMBILOG 1022
36

5.1.1 Voltage Measurement

Two methods are available for voltage measurement: single­ended and differential measurement.
With the single-ended type the voltage to be measured is con­nected between an analogue input (In+) and analogue ground (AGND). Differential measurements are realized by using two analogue inputs (In+ and In-). Measuring range is between 0 and ±10 V.
Note: With differential measurements both voltages have to be within 10 V referred to AGND (Common-Mode-Range).
It is recommended to connect the In- to A
GND
with a high ohmic
resistance.
Hardware Manual COMBILOG 1022
37
Voltage Measurement
Figure 5.1 Voltage measurement - single-ended
Figure 5.2 Voltage measurement - differential
Hardware Manual COMBILOG 1022
Source
IN+
IN-
AGND
=
+
-
U1
U
U = U1
=
+
-
Anschlussschema Schaltung
Source
IN+
IN-
AGND
=
U
U = U2-U1
Anschlussschema Schaltung
= =U1U2
=
U1
U2
CircuitConnection scheme
Connection scheme Circuit
38

5.1.2 Current Measurement

For current measurement the current source is connected between an analogue input (In+) and analogue ground (A
GND
). The load required for measurement is controlled by an internal resistor R
int
to 100 . The power capacity of this shunt is lim-
ited to 125 mW. This results in a measuring range of 25 mA maximum.
Higher currents can be measured by means of an external res­istor which is connected parallel to the current source to the analogue signal input and analogue ground (A
GND
). The power capacity of this external shunt has to be adapted to the current source to be measured, so that the voltage occurring at the analogue input does not exceed +10 V. The analogue input is configured as voltage input. The voltage has to be divided by R
ext
.
Note: The precision of the current measurement with external shunt depends on the precision of the resistor being used.
Note: The input resistance of the current measurement chan­nel depends on the current to measure!
Hardware Manual COMBILOG 1022
39
Current Measurement
Figure 5.3 Current measurement with internal shunt
Figure 5.4 Current measurement with external shunt
Hardware Manual COMBILOG 1022
Source
IN+
IN-
AGND
I
U
I = U / Rint
Anschlussschema Schaltung
Rint
Source
IN+
IN-
AGND
I
U
I = U / Rext
Anschlussschema Schaltung
Rext
Rext
Connection scheme Circuit
Connection scheme Circuit
40

5.1.3 Resistance Measurement

Resistance measurement is carried out by means of voltage measurement at a resistor, measuring the resulting voltage drop. The constant current required for the resistance meas­urements is provided by the internal supply of the data logger.
For this purpose the sensor module connects a supply point in­ternally with the analogue measurement input via a reference resistor Ro. The voltage drop Uo via resistor Ro is required as a reference for further signal processing by the module. The res­istance value of the sensor can be calculated from the input signals Ui as a multiple of the reference resistor Ro. Measuring
range is between 0 and 20 k.
Note: The data logger COMBILOG 1022 supports resistance measurement in 2-, 3- and 4-wire technique. With resistance measurement in 2-wire technique the supply lines cause an ad­ditional voltage drop, thus distorting the measuring result and influencing the measuring accuracy. Therefore it is necessary to pay attention especially with resistance measurement in 2­wire-technique. Wires with impedance as low as possible should be used. Make sure that the leads are well connected to the data logger and the sensor. With resistance measurement using 3-wire technique the potential on the supply lines will be subtracted by software. Therefore 2 measurements are neces­sary, resulting in double measuring time. With resistance measurement in 4-wire technique the drop of potential is picked up directly at the sensor, so that the measuring results are not influenced by the supply lines.
Hardware Manual COMBILOG 1022
41
Resistance Measurement
Figure 5.5 Resistance measurement in 2-wire technique
Figure 5.6 Resistance measurement in 3-wire technique
Figure 5.7 Resistance measurement in 4-wire technique
Hardware Manual COMBILOG 1022
Source
IN+
IN-
AGND
U
Rx = U/U0 * R0,
Anschlussschema Schaltung
Rx
Rx
Messwiderstand
RL
RL R0
=
U0
Rx = 2 * RL
Source
IN+
IN-
AGND
U1
Rx = (U1/U0-2*U2/U0) * R0,
Anschlussschema Schaltung
Rx
Rx
Messwiderstand
RL
RL R0
=
U0
U2
Rx = 0
Source
IN+
IN-
AGND
U2 - U3
Rx = (U2 - U3)/U0 * R0,
Anschlussschema Schaltung
Rx
Rx
Messwiderstand
RL
RL R0
=
U0
Rx = 0
Connection scheme Circuit
Connection scheme Circuit
Connection scheme Circuit
42
Temperature Measurement with Thermocouple
Thermocouples consist of two “thermoelectric wires” made of different materials (e.g. platinum and platinum rhodium) that are welded to each other at one end. If the contact and the oth­er ends of the thermoelectric wires have different temperatures, a “thermoelectric voltage” Uth appears at the contact of both thermoelectric wires. This voltage is largely proportional to the temperature difference. It can be measured and used for temperature measurement purposes. With data logger COMBILOG 1022 the thermoelectric voltage is measured differentially.
Since thermocouples can only measure a temperature differ­ence (difference between temperature to be measured and temperature at the connecting terminals on the sensor module), a terminal temperature (internal cold junction com­pensation, TC
int
) or a known temperature reference (external
cold junction compensation, TC
ext
) also have to be determined.
With measurement of temperature with internal cold junction compensation an additional temperature sensor is necessary to
measure the temperature ϑ
k
at the “cold” terminal. A special cold junction terminal is available, where a Pt100 temperature sensor is integrated directly in the terminal block. The temper­ature of the test point is determined on basis of linearization
trace to ϑ
x
= Lin(Ux+Lin
-1
ϑ
k
).
Hardware Manual COMBILOG 1022
43
If the temperature is measured by external cold junction com­pensation, a second thermocouple of the same type is re­quired, which is connected in series with the first one. The po­larity is selected so that the thermoelectric voltages subtract each other. The second thermocouple is set to a fixed refer-
ence temperature ϑr (mostly ϑ
r
= 0°C). The data logger then
calculates the temperature at the measuring position by means
of the linearisation curve as ϑ
x
= Lin(Ux+Lin
-1
ϑ
r
). The data log-
ger will be informed about the reference temperature ϑ
r
via the
configuration software (“cold junction temperature”).
Hardware Manual COMBILOG 1022
44
Temperature Measurement with Thermocouple
Figure 5.8 Temperature measurement with internal cold junction
compensation by special terminal clamp ICJ 104
Figure 5.9 Temperature measurement with external cold junction
Hardware Manual COMBILOG 1022
Source
IN+
IN-
AGND
U1
ϑ
x = Lin (U2 + Lin-1 ϑk), ϑk = Lin (U1)
Anschlussschema Schaltung
1M
ϑ
k
R0
=
U0
U2
ϑ
x
ϑ
x
zusätzliches
Kaltstellenmodul
Source
IN+
IN-
AGND
ϑ
= Lin (U1 + Lin-1 ϑp)
Anschlussschema Schaltung
1M
ϑpϑ
1M
ϑ
p
ϑ
U1
Connection scheme Circuit
Connection scheme Circuit
45

5.2 Digital Input Channel

The following functions can be realized by means of the Digital Input Channel:
Digital status recording Frequency measurement Counter 8 bit Graycode transducers, Type 4122 / 4123 8 bit status input, with additional (external) module
The above mentioned functions are based on incremental measurements except the digital status and Graycode record­ing. Incremental measuring means to count while measuring. Pulses are counted e.g. from wind speed sensors.
Furthermore it is possible to connect up to 6 sensors with a serial 8 bit Graycode output to the COMBILOG 1022, e.g. wind direction sensor type 4122 / 4123.
By means of an external module type 1025 8 bit status signals can be measured at each input. This module converts the 8 bit into a serial signal, and the COMBILOG will compose it to 1 byte again.
Hardware Manual COMBILOG 1022
46

5.2.1 Digital Status Recording

For the acquisition of digital status information (on/off, closed/open, left/right, etc.) the signal fed to the digital input is collected and is held ready for further processing in the data logger COMBILOG 1022 or for transmission via bus.
The digital input is set (switch closed) as long as the signal voltage remains under the threshold value of 1.0 V. The digital information can be scanned as 1/0 information via bus.
Figure 5.10 Digital status recording
I/O 1
status
"0" "1" "0" "1"
signal diagram:
Hardware Manual COMBILOG 1022
Anschlussschema Schaltung
0V
I/O
+ 10V .. 30V
Connection scheme
Signal diagram:
Circuit
47

5.2.2 Frequency Measurement

With frequency measurements the data logger counts the pul­ses within a certain time interval at the digital input. The user can preset this time interval by setting the time base (TB) in the range between 0.1 sec and 10 sec. The frequency is calculated by the sensor module from the number of pulses and the time base TB as:
frequency f =
number of impulses per time interval TB
length of time intervall TB
Hz
With frequency measurements always the negative signal edge (1 -> 0) is counted.
The lower the frequency f, the larger the interval between two pulses, and the larger the time base TB has to be. On the other hand the updating of the measured value decreases with an in­creasing time base. Thus the time base should be selected so as to make TB ≈ 1/fu, fu being the lowest frequency respectively the smallest change in frequency to be determined by the data logger. The error with frequency measurements thus amounts to ∆f = fu = 1/TB.
Note: The high-end frequency for the frequency measurement, i.e. the highest frequency to be measured. It is 4000 Hz.
Hardware Manual COMBILOG 1022
48
Frequency Measurement
Figure 5.11 Frequency measurement
I/O 1
time base
TB TB TB
counting
1 2 3 1 2 3 1 2 3 1
measurand
(TB = 5 sec)
signal diagram:
- high level:
TB TB TB
1 2 1 1 2
- low level:
0,6 Hz 0,6 Hz 0,6 Hz
0,4 Hz 0,2 Hz 0,4 Hz
pulse
I/O 1
time base
counting
measurand
(TB = 5 sec)
pulse
Hardware Manual COMBILOG 1022
Anschlussschema Schaltung
0V
I/O
+ 10V .. 30V
f
f
Connection scheme
signal diagram:
Circuit
49

5.2.3 Progressive Counter

When configuring a digital input as a progressive counter the data logger COMBILOG 1022 constantly monitors the digital input for a signal variation. If a negative signal edge (1 -> 0) oc­curs at the input, the current result is increased by 1.
The values may range from -2
31
to +(231-1) (about -2.1 to +2.1 billion). Above +231-1 the counting continues with -231. The val­ues can be reset to zero via the bus interface or internally after the procedure of the averaging interval.
Note: The maximum counting rate is 2400 Hz.
Note: After a voltage cut-off the counter is reset to zero.
Figure 5.12 Progressive Counter
+1 +1 +1 +1 +1
n+1 n+2 n+3 n+4 n+5
signal diagram:
I/O 1
counting
counting
measurand
pulse
Hardware Manual COMBILOG 1022
Anschlussschema Schaltung
0V
I/O
+ 10V .. 30V
Connection scheme Circuit
50

5.3 Digital Output Channel

5.3.1 Digital Status Output

The digital output channel supports:
digital status output, host-controlled digital status output, process-controlled
Via the digital inputs/outputs I/O 1 to I/O 6 on the data logger COMBILOG 1022 digital status can be output in digital form, according to the configuration. A typical case of application would be e.g. the local output of an acoustic or optical signal in case a limiting value is exceeded or undershot by a measured value. All outputs are open-collector.
The supply voltage can range from 10 up to 30 VDC. It has to be either supplied externally or taken from the power supply of the data logger.
The status of the digital output can be scanned as 1/0 informa­tion via bus.
With the host-controlled digital status output, the digital output is set according to the status information received by the data logger via bus.
With the process-controlled output of status information the data logger monitors measured values, resp. sensor channels from excess of default threshold values. The digital output is set if one or several threshold conditions are fulfilled.
Hardware Manual COMBILOG 1022
51
The thresholds can be freely defined by the user. The user can also preset the logical signal level (see also the configuration software COMBILOG.EXE).
Thus it is possible to activate a digital output depending on a specified time or periodically. This can be realized in connec­tion with an arithmetic channel, that can calculate the time or a time interval from the internal real time clock. A typical applica­tion is to switch off a modem after a specified time to reduce the power consumption of battery powered systems.
Figure 5.13 Digital status output
I/O 1
status
"0" "1" "0" "1"
signal diagram:
Hardware Manual COMBILOG 1022
Anschlussschema Schaltung
I/O
U+
Connection scheme Circuit
52

5.4 Arithmetic Channel

By means of the arithmetic channel sensor channels and con­stants can be connected with each other via arithmetic opera­tions. The result is allocated to the arithmetic channel. The for­mula can contain up to 20 operands. The calculation is per­formed with a stack depth of 20. The value is handled as a 4-byte floating point format with 24 significant bits according to IEEE, standard 754. The full scale is –1037 to +1037.
A typical application for the arithmetic channel is e.g. the de­termination of a value that cannot be measured directly, but calculated from other values (e.g. power as a product of voltage and current). Or the arithmetic channel is used for fur­ther mathematical preparation of a measuring signal, in order to obtain a particular desired format, linearisation or similar.
Special functions for which the existing commands are not suf­ficient, can be carried out by a user specific download program for the COMBILOG 1022. For this case, some implemented special functions can be used for assistance.
To integrate a serial channel the special function 1 is to be used. The serial channel converts an ASCII telegram into a float value. Two variants are possible to chose, constant length or variable length. With constant length the conversion starts at startposition and ends at startposition + length. With variable length the conversion starts at startposition and ends at the first delimiter character. Should the conversion be unsuccessful the default value will be taken. The call to special function 1 is defined as:
Hardware Manual COMBILOG 1022
53
spez1(INTERFACE; TYPE; START; VAR; DEFAULT)
INTERFACE
1: RS232
2: RS485
TYPE
1: constant length
2: variable length
START: start position (starting with '0')
VAR
for TYPE = 1: value length
for TYPE = 2: delimiter character (ASCII value)
DEFAULT: on error this is value will be returned
In defining more than one arithmetic channel with this function you can extract more than one value from one ASCII telegram.
To convert a variable value with start position 10, delimiter #, interface RS485 and default value –1 from “Value:_0023,6697#END” the call to spez1() should be spez1(2; 2; 10; 35; -1) and the result is 3.6697.
With special function spez2() some special meteorological val­ues can be calculated. The following functions are defined:
spez2(air pressure; height; 1);
Calculates the air pressure on sea level out of air pressure at given height.
spez2(app_wind; fair_wind; wind_angle; 2);
Calculates the true wind speed out of apparent wind, fair wind and the enclosed angle.
spez2(temperature; radiance_balance; saturation_deficit ;
wind_speed; 3); Calculates the evaporation acc. to Penman-Monteith.
Hardware Manual COMBILOG 1022
54
Note: The calculation time of an Arithmetic Channel is
min 0.2 ms. The overall calculation time is the sum of the times of all operands in the formula plus 0.2 ms. This has to be taken into account when calculating the sample rate.
Hardware Manual COMBILOG 1022
55
Arithmetic Operators
operations abbreviation time
Addition + 0.26 ms
Subtraction - 0.26 ms
Multiplication * 0.26 ms
Division / 0.26 ms
Modulo % 0.24 ms
Truncate value trunc 0.16 ms
Minimum value min 0.16 ms
Maximum value max 0.16 ms
Absolute value abs 0.16 ms
Square root sqrt 0.26 ms
Exponential function to base e exp 0.56 ms
Logarithm to base e In 0.26 ms
Logarithm to base 10 log 0.26 ms
Sine sin 0.44 ms
Cosine cos 0.44 ms
Tangent tan 0.66 ms
Inverse sine arcsin 0.70 ms
Inverse cosine arccos 0.70 ms
Inverse tangent arctan 0.62 ms
Lowest value from a selection Low 0.30 ms
Highest value from a selection high 0.30 ms
X
Y
power 1.64 ms
Integrator integ 0.18 ms
Differentiator deriv 0.18 ms
Read from external module read
Write to external module write
Time /seconds of the day) SecondsOfDay 0.18 ms
Sample rate SampleTime 0.18 ms
Free space on SD-Card SDSpace 0.20 ms
Hardware Manual COMBILOG 1022
56
Free space in RAM RAMSpace 0,1 ms
Application specific function 1 spec 1
1)
Application specific function 2 spec 2
1)
Application specific function 3 spec 3
1)
Application specific function 4 spec 4
1)
1)No specification available as the time depends to the specific function and the program.
Table 5.1 Arithmetic operators and processing times
The times given in the above table are based on an operating fre­quency of 48 MHz.
Remarks:
Division (/)
When dividing by zero, the positive full scale (+1037) will be assigned to the Arithmetic Channel if the numerator is positive and the negative full scale (-1037) will be assigned if the numerator is negative.
Square root (sqrt)
The square root of a negative number is zero.
Logarithm to Base e (In)
For a value 0 the negative full scale will be assigned to
the Arithmetic Channel.
Logarithm to Base10 (log)
For a value 0 the negative full scale will be assigned to
the Arithmetic Channel.
Arc functions (sin, cos, tan)
The arc values must be taken in radians (2π = 360°). If
calculating the tangent, the positive full scale will be as-
signed to the Arithmetic Channel for the arc value
2
π
and
the negative full scale for the arc value -
2
π
.
Inverse functions for sin (arcsin), cos (arccos), tan
(arctan)
Hardware Manual COMBILOG 1022
57
The results of the inverse functions are given in radians
(2π = 360°). At the function arcsin the value +
2
π
will be
assigned to the Arithmetic Channel for a value >1 and the
value –
2
π
will be assigned for a value <-1. At the function
arccos the value 0 will be assigned to the Arithmetic
Channel for a value >1 and the value will be assigned for a value <-1.
Minimum and maximum of a channel value (min, max)
With this function the minimum and maximum value of a channel appeared since the last reset has been triggered off can be determined (“pull-pointer-function”). The result value can be reset to the actual value of the measured channel via the bus or at the end of the average interval.
The functions read and write enable the data logger to re-
ceive measured values from other modules connected to the same bus, resp. to send values to them (master func­tion)
Note: Logic combinations, e.g. if-then relations, are not yet possible respectively would require a user-specific software (upon request).
Hardware Manual COMBILOG 1022
58

5.5 Setpoint Channel

This channel features transmission of values via bus to the data logger COMBILOG 1022. The values are allocated to the setpoint-channel and are thus at the disposal of the data logger for further processing.
A typical application for the setpoint-channel is e.g. the dynam­ic variation of control thresholds.

5.6 Alarm Channel

The Alarm Channel has the same features as the process-con­trolled digital output channel, the only difference is that the status information is not output locally at the digital output, but can only be scanned via the bus.

5.7 Serial Channel

The serial channel output transmits the ASCII formatted strings of the logging channels onto the configured interface. The seri­al channel function can be configured using the build in web page (see chapter 12.9)

5.8 Threshold Values

The user can preset the conditions for process-controlled digit­al status output on the data logger and for the output of an alarm signal via bus. This is carried out by means of the confi­guration software COMBILOG.EXE.
Hardware Manual COMBILOG 1022
59

5.9 Error Handling

The data logger COMBILOG 1022 can detect independently certain defects, which are result of a line break, short-circuit or communication interrupt, for example. For these defects the user can preset a certain behaviour for the data logger via the configuration software.
In case of a sensor failure the last valid value can be main­tained, set to the corresponding limits or set to a default value.
Furthermore the COMBILOG 1022 can send messages via mo­dem or SMS automatically to report errors or other conditions, e.g. if the data memory capacity becomes zero.
Hardware Manual COMBILOG 1022
60

6 DISPLAY / MENU OPERATION

6.1 Display and Operation

The data logger COMBILOG 1022 has a display with 4 lines of 16 characters each, in order to allow the indication of the measured values of each channel. Furthermore the settings of the data logger can be recalled and changed if desired (there­fore the input by the press/ rotary knob must be unlocked; refer to the corresponding section 6.2 “Menu Items“) The operation is performed via the combined press/rotary knob at the right side of the display. By turning the knob the menu items or informations can sequentially be indicated. A confirma­tion or a call of a function is performed by pressing the knob.

6.2 Menu Items

In the following diagrams all display pictures with the corres­ponding operation steps are indicated. Following symbols are used for the operation steps:
Symbols:
turn the knob clockwise
turn the knob counter clockwise
press the knob briefly (confirmation)
press the knob for approx. 1 second minimum
(abortion)
Hardware Manual COMBILOG 1022
61
Main menu
Note: The number, designation and indication of the measured value of the channels depend on the configuration.
Hardware Manual COMBILOG 1022
62
Configuration menu
Note: If the push/turn selection knob is pressed at any point in the configuration menu for about 1 second you will return to the initial position in the main menu.
Hardware Manual COMBILOG 1022
63
Setting of the display backlight
Note: The value of the display backlight can be set in steps from 0 % to 100 %. With low values the display will be set dark and with the value 100 % it will be set to maximum brightness.
Hardware Manual COMBILOG 1022
64
Setting the scan rate and the averaging interval
Configuration menu
Note: The scan rate determines after which time interval the
measured values of the channels will be measured again. Scan rate is selectable between 0.5 s and 1 h. At “Averaging“ the averaging interval can be set. It determines the time interval for the averaging of the measured values. Av­eraging interval is selectable between 1s and 12h. In the example the measured values will be measured again every 10 seconds and after 60 seconds the average value will be calculated (here by means of 6 measured values).
Hardware Manual COMBILOG 1022
65
Setting of the Automatic Switch Off and the LED Display
Configuration menu
Note: With the automatic switch off function the data logger
can be set to the saving mode if no operation is made by the press/ rotary knob in a certain time interval (30 seconds). In this case the display will be set off until a further operation takes place. If “LEDs“ is set to ON the LED-display is switched on and the two LEDs RUN and ERR on the front of the data logger show the actual operating state (mode) of the data logger. With the selection OFF the LEDs will be switched off.
Hardware Manual COMBILOG 1022
66
Setting of the Baud Rate and the Address
Configuration menu
Note: Possible values for the baud rate are 2400, 4800, 9600,
19200 and 38400 bps. Additionally to the baud rate the parity can be set . Possible values are N (no parity), E (even parity) and O (odd parity). For the address a value between 1 and 127 can be set.
Hardware Manual COMBILOG 1022
67
Lock or unlock the press/rotary knob input
Configuration menu
Normally the changing of configuration parameters by the press/rotary knob is locked (disabled). To change the paramet­er the press/rotary knob must be unlocked (enabled).
Note: If no operation is performed at the data logger for ap­prox. 30 seconds it will return to the main menu and the press/rotary knob input will automatically be locked.
Hardware Manual COMBILOG 1022
68
Network settings
If the powersave feature for the ethernet interface is enabled no connection can be established using this interface. Due to the relatively high power consumption it is essential for low power applications like mobile battery powered systems to have the ethernet powersave function enabled. The IP, gateway and netmask selection modifies the selected address. The flashing number is actually selected for modifica­tion. Rotating the knob counts the number up or down while pressing the knob selects the next number.
Hardware Manual COMBILOG 1022
69
Channels settings
In order to change the values the press/rotary knob must be pressed. Thereby a cursor will be set on the first character of the value. Pressing the knob again moves the cursor one char­acter to the right. The value at the place of the cursor can be altered by turning the knob. Clockwise (=upwards) or counter­clockwise (=downwards).
Note: Depending on the type of channel different settings can be made.
(1) the definition of offset and factor is possible for the ana-
logue input channel, the digital input channel and the digit­al output channel. These settings are used to convert the measurement value from the unit of the measured value to the unit of the measurement display.
(2) For the setpoint channel a setpoint value can be defined
if this is allowed by the configuration software COMBILOG.EXE. This value can be used by the data log-
ger for further processing (e.g. for the arithmetic channel). For the arithmetic channel and the alarm channel no settings can be made.
Hardware Manual COMBILOG 1022
70
Display and delete the data memory
Main menu
The display shows the number of stored datasets and the max­imum capacity of the data memory. The fourth line shows the approximate time until the data memory is filled. This time is displayed in days (d) or hours (h) and is calculated by the data logger assumed that the average interval is constant.
Hardware Manual COMBILOG 1022
71
Setting of the date and time
In order to change the date and the time the press/rotary knob must be pressed when the time
(1)
respectively date
(2)
is dis­played. For this purpose a cursor will be set on the first charac­ter of the date respectively time indication. Pressing the knob again will move the cursor one character to the right. The value at the place of the cursor can be altered by turning the knob.
Note: The time will be stopped if the press/rotary knob is pressed in the main menu, date and time display. The time will continue running if the date or time setting is confirmed by pressing the knob.
Hardware Manual COMBILOG 1022
Configuration Date / Time
Time change?
Combilog 1022 Date / Time
02.01.2009 00:05:10
Configuration Date / Time
Date change?
Change+
Change-
Change+
Change-
Configuration Date / Time
02.01.2009
Configuration Date / Time
00:05:10
 
 
(2)
(1)
72

6.3 SD-Card

If a SD-card is inserted into the card slot an additional menu item is displayed in the main menu. Choosing this menu item one can update the firmware of the data logger or eject the SD­card.
Hardware Manual COMBILOG 1022
SD-Card >remove card update files
Main menue >SD-Card Data recorder Auto Powerdown
FIRMWARE
UPDATE
AND
RESTART
Change +
Change -
Change+
Change-
Do you want to update with: COMBL.BIN [Yes/No]
Updatefiles
COMBL.BIN
   
 
73

6.3.1 Remove Card

If a SD-card is configured as logging destination the card has to be logged off of the system before removing it. Otherwise written data can be lost. For card removal you choose this menu item. The card will be logged out of the system without any further question.

6.3.2 Firmware Update

Using the menu item 'update files' one can start the firmware update procedure. Therefore a special firmware update file must reside in the directory 'updates' on the SD-card. After choosing ' update files' all files in the directory 'updates' are presented for selection. Selecting one file starts the update procedure after a security question. If the file contents for some reason seems to be cor­rupted, the firmware update stops. If there was no error during file checking the data logger starts the reprogramming and starts a system reboot during which the new image is written into the program flash.
Warning:
As a firmware update is a critical process a backup of all logged data and of the configuration is highly recommended!
Hardware Manual COMBILOG 1022
74

7 DATA STORAGE

7.1 General Remarks to Data Storage

The COMBILOG 1022 is able to store the calculated mean val­ues. Data storage with the COMBILOG 1022 can be accom­plished in three different manners:
Internal flash: 7 MByte are available as circulated buffer for data recording.
External SD card: The records are continuously stored on this card.
Internal flash and external SD-card: The records are stored in parallel in the internal flash and on the external SD-Card.

7.2 Modes of Data Storage

No data storage
In this mode the COMBILOG 1022 operates as measuring module, processes measuring values and sends the results to a PC upon command.
Continuous data storage
This is the normal operation of the data logger. The measured values, selected for storage, are continuously written to internal or external memory, with the preset averaging interval.
Hardware Manual COMBILOG 1022
75
Conditional data storage
Measured values are stored in the memory as long as a condi­tion defined by the configuration program is valid (e.g. a threshold is exceeded).
Conditional data storage with zoom function
In this mode two time bases for data recording are available. The selection between these time bases depends on a condi­tion selected by the configuration program.

7.3 Storage Medium

The COMBILOG 1022 has 3 storage modes.
Flash: storage only in internal Flash
SD-Card: storage only on external SD-Card
Flash & SD-Card: data will be stored on both mediums
NOTE: The storage medium can be configured on the website. Have a look at section webserver.
Hardware Manual COMBILOG 1022
76

7.4 Internal Data Storage

The COMBILOG 1022 is delivered with 8 MByte flash memory. For data recording 7 MByte memory are available as circular memory (first in, first out). If the memory is read out, at first the oldest record will be output and the corresponding space is en­abled. The data memory can be read out via one of the serial inter­faces. Communication commands are described in chapter
11.8. Every record consists of a length information, date, time and measured values.
L Time M1 M2 ... Mx S1 S2 Sx
L Length of record (2 Byte) Time Date and time of record (8 Byte) M1 First measured value (4 Byte) M2 Second measured value (4 Byte) Mx Last measured value (4 Byte) S1 First external channel (only with master function) S2 Second external channel Sx Last external channel
Hardware Manual COMBILOG 1022
77
Memory demand for one record:
Number of bytes = 8 + 4 * number of measured values
The duration of data recording until the data memory is filled can be calculated by the following formula:
86400*)*48(
*1024² * 7
n
M
d
+
=
d = duration of data recording in days n = number of values to be stored (without date and time) M = averaging interval in seconds
Example:
Storage of eight measured values per minute.
Number of Bytes = 8 + (4 * 8) = 40 bytes per record Memory demand per day = 24 * 60 * 40 = 57600 Bytes
At 7 MBytes internal flash 7 * 1024² * 60 / (40 * 86400) = 127,4 days can be recorded.
Hardware Manual COMBILOG 1022
78

7.5 External Data Storage w ith SD Card

The data will be stored in a file on the SD Card. The name of the file depends of store mode. Three different modes of file storage:
Day file
Every day a new file will be created with the name consists of day, month and year.
Example: 02062008.csv This file contained data from 2. June 2008
Month file
Every month a new file will be created with the name consists of month and year. Example: 062008.csv This file contained data from June 2008
Continues file
New data will be append to the file. The name of the file: cl_log.csv
The file contain also additional information about channel con­figuration. If the memory card is filled, no further data will be recorded. Data recording is performed in ASCII format. A separation mark (";" or tabulator) and the decimal character ("." or ",") is selectable by the user.
Note: The data logger can write data to pre-formatted memory cards only. The data logger supports the FAT 32 file system.
Hardware Manual COMBILOG 1022
79
Theodor Friedrichs & Co. delivers pre-formatted memory cards, that can directly used for the COMBILOG.
Structure of the data file:
Ident Friedrichs
V2.01 V4.01 Location COMBILOG Serial No 123456 Sample Rate 1 Slowest 1 Fastes 1 Store Rate 60 Date Time Variable 1 Variable 2 ... Variable n
02.06.2008 12:01:48 3.45 1.28 ... 3.44
02.06.2008 12:01:48 3.45 1.28 ... 3.44
02.06.2008 12:01:48 3.45 1.28 ... 3.44
02.06.2008 12:01:48 3.45 1.28 ... 3.44
Structure of a record:
T Time T M1 T M2 T … T Mx T S1 T S2 T … Sx CR LF
T Separation mark (Tabulator or “;“) Time Date and time of the data record
(DD.MM.YY HH:MM:SS) M1 First measured value M2 Second measured value Mx Last measured value S1 First external channel S2 Second external channel Sx Last external channel CR Carriage Return LF Line Feed
Hardware Manual COMBILOG 1022
80
Note: With the Configuration program a ”;” or a “TAB” (ASCII
09hex) can be selected as delimiter and a “:” or a “;” is select­able as decimal point.
Memory demand for one record:
NumBytes = 20 + field length (M1) + 1 + field length (M2) + 1…
+ field length (Sx) +1
NumBytes = 20 + number of measured values * (8 +1)
(fixed field length of 8 characters)
Example:
Storage of eight measured values per hour: NumBytes = 20 + 8 *(8 + 1) = 92 bytes per record Memory demand per day = 24 * 92 = 2208 bytes
Following equation can be used:
86400**l
Mk
d
=
d = period of data recording in days k = capacity of the memory card in byte l = length of one data set in byte M = average interval in seconds
Hardware Manual COMBILOG 1022
81

8 MASTER FUNCTION

8.1 Master Function

The two interfaces and the master function of the COMBILOG 1022 allows a configuration of a complex measurement system. In such a system with activated master function the data logger as master is able to read out the other bus users (slaves)
This feature is used to extend the number of inputs and outputs of the COMBILOG 1022 in case the 8 analogue inputs and 6 digital inputs/outputs of the data logger are not sufficient. Other COMBILOG 1022 can be connected to the RS-485 inter­face as slaves. The master data logger reads out the measured values (actual instantaneous values only!) of the slave modules automatically via the bus and stores them in its memory . By this method complex systems to record up to 92 channels can be realised easily.
The advantage of this master function is a flexible distribution of a number of inputs and outputs and the sensors can be loc­ated in an area over some kilometres. The data storage is cent­ral in the master data logger.
These data can be read out directly via the RS232 interface or via telephone or GSM modem.
Arrangement of such a measurement system with COMBILOG 1022 and slave modules is described by following steps:
Hardware Manual COMBILOG 1022
82
Configure all slave modules with the same bus parameters
as the master data logger (same protocol type, same baud rate and parity). All modules must have different modules addresses.
Connect the master data logger via RS232 with the host
PC and start the configuration program.
Set the bus parameter for the master function with module
settings. Baud rate is selectable between 2400 and 38400
bps and independent of the settings of the RS232 inter­face.
Select all values to be measured by the master data log-
ger by defining the module address and the channel num­ber of the corresponding slave module. Up to 60 external channels can be selected.
Connect all slave modules with the master via the RS485
bus. After downloading the configuration to the master data logger the data transmission between master and slave modules is started.
Note, that the configuration of slave modules is not possible after activating the master function.
All collected slave values will be added to the normal dataset, that is defined by the logging function of the master. No further calculation (averaging etc.) is executed. If averaging is neces­sary, use the read function of the arithmetic channel!
The scan rate corresponds to the “Logging Interval”
Within one RS485 bus system, only one COMBILOG 1022 can be defined as master.
Hardware Manual COMBILOG 1022
83
Figure 8.1 Example Master Slave System
Hardware Manual COMBILOG 1022
Host-Rechner
RS232
Busmaster Adresse 01
Slave Adresse 02
Slave Adresse 03
Sensoren Sensoren
Sensoren
84

9 INITIATION AND TEST

9.1 Before Connecting the Device

Before connecting the supply voltage to the data logger COMBILOG 1022, once again check all connections. Watch that the supply voltage never exceeds 30 VDC.

9.2 After Connecting the Device

After connecting the supply voltage the data logger displays the current operating state on the two LEDs at the front of the device (if the LED-display has not been switched off). The meanings of the LEDs are given in table 9.1 on the following page. If the module was switched off for a longer time, the settings of date and time should be checked and corrected if necessary (see chapter 6.2).

9.3 Configuration of the Data Logger

Before entering into operation the data logger has to be pro­grammed and configured as to its specific application. In most cases the programming has already been carried out on deliv­ery (see status of RUN-LED and ERR-LED, table 9.1). The configuration has to be carried out by the user by means of the configuration software COMBILOG.EXE on a PC. The installa­tion procedure is described in APPENDIX E.
Hardware Manual COMBILOG 1022
85
RUN
(green LED)
ERR
(red LED)
meaning
off
The supply voltage has been selected too low or
the power supply cannot supply the required
power.
off flash
The data logger is in the monitor mode. A valid
program has not yet been loaded; the appliance is
not yet ready for operation.
on
There is a sensor error detected. Possible causes
may be: 1. wrong configuration, 2. line break or
short circuit, 3. measured value too high or too
low.
flash flash
The data logger is in the download mode. Cur-
rently a program or a configuration is transmitted
to the data logger.
on
off
The supply voltage has been connected correctly.
There is no error. Data transmission to the module
via bus is not active.
on
There is a sensor error detected. Possible causes
may be: 1. wrong configuration, 2. line break or
short circuit, 3. measured value too high or too
low.
short off X
A telegram has just been dispatched from the
data logger via the bus to a control system or to a
PC.
Table 9.1 Assignment of LED functions (flash frequency approx. 1Hz)
Hardware Manual COMBILOG 1022
86

10 STRUCTURE OF THE BUS TOPOLOGY

The coupling of the data logger COMBILOG 1022 to a commu­nication bus is performed via an integrated RS485 interface. The second interface, the RS232 computer interface, is only usablein order to build point-to-point connections for a distance of max. 20 m (65 feet). At the COMBILOG 1022 the same data will permanently be given out. Only the physical characteristics of the two interfaces are different whereas only those of the RS485 are appropriate in order to build a bus topology. The RS485 bus topology is characterized by the following features:
Bus interface:
RS485, half duplex
Bus topology:
line pattern, closed at both ends by the characteristic im­pedance, stub cable to the party max. 30 cm (12 inch).
Bus medium:
shielded, twisted pair cable
Transmission speed:
ASCII-protocol:
2400 Bit/s
4800 Bit/s
9600 Bit/s
19200 Bit/s
38400 Bit/s
Hardware Manual COMBILOG 1022
87
Line length:
depends on the transmission speed, max. 1.2 km (0.75 miles) per bus segment, max. 4.8 km (3 miles) via a phys­ical bus string with 3 repeaters.
Number of bus users:
max. 32 bus users per bus segment, max. 127 bus users via a physical bus string.

10.1 Bus Interface

The bus interface in the data logger is an RS485 interface. Its advantages compared with RS232 connections are a larger number of users, its higher transmission speed, its higher im­munity from interferences and the extended line length.
transmission speed
1 K 10 K 100 K
187,5 K
1 M 10 M
[bps]
10 m (32.5 ft)
100 m (325 ft)
1000 m (3.250 ft)
1200 m (3.900 ft)
transmission
600 m (1.950 ft)
RS 232
RS 422 RS 485
route
Figure 10.1 Interrelation between transmission speed and line length
Hardware Manual COMBILOG 1022
88

10.2 Bus Structure

The bus structure is a line structure where each bus segment will be blanked off with characteristic impedance on both ends. Branches can be build up over a bi-directional signal amplifier, so called repeater. Other than that branches are not permitted (no tree topology). The max. stub to a user is not allowed to ex­ceed 30 cm (12 inches).
The following figures show a few examples for a possible set­up of bus topologies. The meaning of the symbols is: : bus user, : repeater and : bus termination.
.....
Figure 10.2 Simple line structure
.....
.....
.....
Figure 10.3 Extended line structure
Hardware Manual COMBILOG 1022
89
: :: :
...
Figure 10.4 Line structure with branches
The RS485 interface permits the simultaneous connection and operation of a maximum of 32 bus users per bus segment. Fur­ther bus segments can be constituted via bi-directional repeat­ers, and thus the number of bus users can be extended to max.
127.

10.3 Transmission Speed and Line Length

The transmission speed with the data logger COMBILOG 1022 can be adjusted between 2,400 baud and 38.4 kbps. The per­missible line lengths are reduced with increasing transmission speed. At the given transmission speeds these line lengths are about 1,200 m (3,900 feet) per bus segment. With 3 repeaters topologies with a dimension of max. 4.8 km (3 miles) can be set up.
Hardware Manual COMBILOG 1022
90
Note: These specifications refer to bus cables with a conductor
cross section of 0.22 mm² and a permissible signal attenuation of max. 6 dB referred to the overall length. According to previ­ous experience the line length can be twice as long if a two­wire circuit with a conductor cross section of at least 0.5 mm² is used.

10.4 Bus Cable

For setting up a bus topology a shielded twisted pair with at least two leads and the following electric characteristic values should be used:
characteristic impedance : 100 ... 130 at f > 100 kHz
operating capacity : max. 60 pF/m conductor cross section : min. 0.22 mm² (AWG 24) attenuation : max. 6dB referred to the
overall length
Hardware Manual COMBILOG 1022
91

10.5 Bus Plug

For installing the bus cable and the bus interface, 9-channel D-sub miniature plugs and sockets are used. The pin assign­ment for the RS485 connection according to PROFIBUS/MOD­BUS specification is given in table 10.2.
plug Pin RS485-
notation
meaning
Profibus
meaning
Modbus
1
5
6
9
DB 9
1 Shield, protective
Ground
2 RP, Reserved for
Power
VP, positive 5...24 V
D.C. Power Supply
3 RxD/TxD-P, Receive /
Transmit Data-P
PMC, Port Mode Con-
trol
4 CNTR-P, Control-P
5 B / B´ DGND, Data Ground B/B´, transceiver ter-
minal 1, V1 voltage
( V1 > V0 für binary 1
[OFF state] )
6 VP, Voltage Plus
7 RP, Reserved for
Power
8 C / C´ RxD/TxD-P, Receive /
Transmit Data-N
C/C´, signal and power
supply common
9 A / A´ CNTR-N, Control-N A/A´ transceiver termin-
al 0, V0 voltage
( V0 > V1 for binary 0
[ON state] )
Table 10.2 Pin assignment D-sub miniature plug according to PROFIBUS /
MODBUS
Hardware Manual COMBILOG 1022
92
Only signal leads A and B (and Shield) are absolutely obligat­ory for a (shielded) connection. All others can be installed to­gether with these signal leads if required.

10.6 Bus Termination

In order to avoid signal reflections on the bus, each bus seg­ment has to be blanked off at its physical beginning and at its end with the characteristic impedance. For this purpose, a ter­minating resistor Rt is installed between the bus leads A and B. In addition to that the bus lead A is connected via a pull-down resistor Rd to ground (Data Ground) and the bus lead B is con­nected via a pull-up resistor Ru to potential (VP). These resist­ors provide a defined quiescent potential in case there is no data transmission on the bus. This quiescent potential is level
high.
Figure 10.5 Bus termination
Note: The numbers in brackets in figure 10.5 indicate the pin number for the connection via the 9-channel D-sub miniature plug.
Hardware Manual COMBILOG 1022
VP
DGND
Ru
Rt
Rd
VP = +5V
Ru = 390
± 2%, min ¼ Watt
Rd = 390
± 2%, min ¼ Watt
Rt = 390
± 2%, min ¼ Watt
B
A
B (3)
A (8)
Profibus-
Bezeichnung
Combilog­Klemmen-
bezeichnung
watt
watt
watt
reference
reference
93
The bus termination can be carried out in various ways. It can either be carried out via external resistors and a separate power supply, independent of the module, according to figure
10.5. In this case we recommend to use the indicated resistors for the bus termination.
Or the bus termination is connected with the bus users at the beginning and at the end of a bus line. Most of the RS485 con­nections for controls, computers, repeaters, interface convert­ers, etc. offer this option.
Also with data logger COMBILOG 1022 this option is given. Via the bus termination plug which is available as accessory and installed at the right or left port on the front side of the device, the bus termination at this module can be additionally connec­ted. Two jumpers which connect the bus with the bus termina­tion in the data logger are integrated in the bus termination plug.
Figure 10.6 Bus termination on the COMBILOG 1022
Note: Instead of the bus termination module separate jumpers can also be used for the bus termination. In this case, please
Hardware Manual COMBILOG 1022
VP
DGND
Ru
Rt
Rd
A
B
94
make absolutely sure that the jumper clips are installed as in­dicated, and that the bus leads or the bus termination are not short-circuited by mistake!

10.7 Shielding

In case of increased interference we recommend to use shiel­ded bus cables. In this case, a shielding should also be carried out for the cables from power supply and for the signal cables.
There are varying experiences and recommendations concern­ing the kind of shield connection. In general the shield should be connected with the protector ground (not data ground!) at each bus connection. If necessary the shield should be earthed additionally several times along the course of the cable. With smaller distances, e.g. with stub cables, the immunity from noise often is improved if the shield is only applied to the stub cable exit.
Figure 10.7 Earthing of the bus line shield on the COMBILOG 1022
Hardware Manual COMBILOG 1022
Isolierung
Schirmgeflechtzentrale Erdung
Central earth in point
braided shield
isolation
95
Bus parties such as controls (PLCs), computers (PCs), repeat­ers and interface converters, a.s.o., mostly offer the possibility of applying the shield directly to the appliance or to separate shield rails. The shield rails offer the advantage of preventing possible interfering signals from being led to the appliance via the shield. These are already branched off before via the pro­tector ground.
The COMBILOG housing has no direct shield terminals. The shield of the bus cable can be earthed e.g. by so-called shield clamps.
Note: The shield must not be connected to bus interfaces A or B!
Note: The shield should always be connected to earth in a large surface, low-inductive manner.

10.8 PC Bus Connection

The bus interface of the data logger is based on the RS485 standard. Since most of the hosts are "only" equipped with RS232 interfaces, an interface converter or a plug-in board with RS485 drivers is required for conversion purposes.
Theodor Friedrichs offers a compact interface converter.
Furthermore a repeater module is available from Theodor Friedrichs. This module can be used as a repeater or as a converter. It also enables to connect the necessary bus termination with a switch. The repeater/converter has a snap­on mounting for installation on standard profile rails (DIN rail) 35 mm (1.4 inch) according to DIN EN 50022.
Hardware Manual COMBILOG 1022
96

10.9 Potential Equalization

The difference between the actual physical voltage potentials DGND of all connections with the bus must not exceed ±7 Volt. If this cannot be guaranteed, an equalization has to be provided. For most of the connections this means that the minus connection of the power supply has to be fed-through as a compensating line from connection to connection.

10.10 Adjustment of Address and Baud Rate

Before a control unit (PLC) or a computer (PC) can interchange data with a data logger via the bus, address and baud rate for the data logger have to be defined. Following hints have to be considered:
All devices have to be adjusted to the same baud rate. Within the bus topology the same address must not ap-
pear twice.
bus parameter ASCII-protocol /
MODBUS-Protocol
PROFIBUS-protocol
address 1 ..... 127 1 ..... 126
2,400 bps -
baud rate 4,800 bps -
9,600 bps 9,600 bps
19,200 bps 19,200 bps
38,400 bps -
Table 10.3 Setting variants for address and baud rate
for the data logger COMBILOG 1022
Hardware Manual COMBILOG 1022
97
If no other specifications are made on delivery, the data logger has address 1 and baud rate 19,200 bps, no parity as default. The adjustment can be changed via bus by means of the con­figuration software COMBILOG.EXE:
Adjustment via bus by means of the configuration soft-
ware:
The condition for adjusting address and baud rate via bus is that there are no different data loggers with the same address on the bus. Otherwise the bus connection has to be disconnec­ted or the supply voltage has to be interrupted for the duration of the adjustment with those data loggers having the same ad­dress as the data logger that is to be newly adjusted. The data logger to be newly adjusted does not necessarily have to be set to the same baud rate as the PC, as the adjustment pro­cedure is accomplished as described below:
The adjustment or modification of address and baud rate via bus is always carried out together with the loading of a new configuration by means of the configuration software COMBILOG.EXE. The download is carried out as described in the online help for the configuration software. The LEDs at the front indicate which data logger is being newly adjusted at the moment. If the LED ERR changes from "off" to "flash", a new program is just being transmitted to the data logger. The values are taken over as soon as the data transmission via bus has been successfully completed.
Note: The address “0” is reserved for transmission via PROFIBUS. This address can therefore not be assigned to a logger.
Hardware Manual COMBILOG 1022
98
Note: The address “127” is reserved for broadcast transmis-
sion in the PROFIBUS-protocol (Level 2) and may only be as­signed for these cases.
Adjustment via Interface, RS232 by means of the configur-
ation software:
In addition to the bus interface RS485, the data logger COMBILOG 1022 has an RS232 computer interface. By means of the configuration software COMBILOG.EXE addresses can be assigned and bus parameters can be adjusted via the RS232 interface similarly as when using the RS485 interface.
Hardware Manual COMBILOG 1022
99

11 COMMUNICATION

11.1 Bus Interface

The bus interface of the data logger is an RS485 interface ac­cording to the specifications of the EIA-RS485 USA standard. The host interface is an RS232 interface according to the spe­cifications of the EIA-RS232.

11.2 Bus Protocol

The following protocols are available for the data logger
COMBILOG 1022:
ASCII-protocol PROFIBUS-protocol (Level 2) according to
DIN 19245, part 1
MODBUS-RTU-protocol according to
PI-MBUS-300 Rev. D
The ASCII protocol can be operated simultaneously with the PROFIBUS or the MODBUS protocol. The binary protocols PROFIBUS and MODBUS can only be used exclusively. You can select the desired binary protocol through a web page. The ASCII protocol and the PROFIBUS protocol will be de­livered as standard.
Hardware Manual COMBILOG 1022
100
Loading...