Kipp&Zonen Brewer MkIII User Manual

Brewer MkIII
Spectrophotometer
Service manual
REV
DESCRIPTION
DCN #
DATE
APPD
-
Initial Release
-----
98-11-06
TKLL
A
New Brewer Electronics
822
99-02-01
TKLL
B
Updated
868
99-08-13
C
Updated
55
05-10-21
KBo
D
Updated
9
06-02-03
KBo
E
Hamamatsu PMT
08-11-14
KBo
ii
iii
MKIII SERVICE MANUAL
1
TABLE OF CONTENTS
PREFACE ............................................................................................................................................... 3
1 PURPOSE AND APPLICATION ...................................................................................................... 3
2 EQUIPMENT REQUIRED ............................................................................................................... 3
3 MAJOR EQUIPMENT FAILURES ................................................................................................... 5
3.1 "No Operation" - No Power Indication ...................................................................................... 5
3.2 Trouble Shooting Sequence for "No Operation" Condition ........................................................ 6
4 OPERATING TEST FAILURES ..................................................................................................... 11
4.1 Mercury Lamp Circuit .............................................................................................................. 11
4.2 Trouble shooting a Mercury Lamp Test Failure ....................................................................... 11
4.3 Standard Lamp Circuit ............................................................................................................ 13
4.4 Standard Lamp Failure ........................................................................................................... 13
4.5 Trouble-shooting Standard Lamp Test Failure ........................................................................ 13
4.6 Light Detection System Failure ............................................................................................... 15
4.7 Trouble-shooting the Light Detection System ......................................................................... 15
4.8 Stepping motor control ............................................................................................................ 17
4.9 Photon Counter Circuitry ......................................................................................................... 17
4.10 High Speed Amp Board ....................................................................................................... 17
4.11 Ratemeter ............................................................................................................................ 18
4.12 Temperature monitor ........................................................................................................... 18
4.13 High voltage Circuits ........................................................................................................... 18
4.14 Clock / Calandar .................................................................................................................. 18
4.15 A/D Converter ...................................................................................................................... 18
4.16 Breather Desiccant Cartridge .............................................................................................. 19
4.17 Humidity Sensor (Optional) ................................................................................................ 19
5 TEST RESULTS OUT OF SPECIFICATION ................................................................................. 21
6 OPTICS CARE AND CLEANING .................................................................................................. 23
6.1 Optics Care ............................................................................................................................. 23
6.2 Optics Cleaning ....................................................................................................................... 23
7 AZIMUTH TRACKER .................................................................................................................... 25
7.1 Brewer Power Problems. ........................................................................................................ 25
7.2 Bewer Communications Problems .......................................................................................... 25
7.3 Tracker Electrical Problems .................................................................................................... 26
7.3.1 Sensor Problems .......................................................................................................... 26
7.3.2 No Tracker Operation. .................................................................................................. 26
7.4 Tracker Mechanical Problems ................................................................................................ 27
8 MAIN ELECTRONICS FIRMWARE / CONFIGURATION LOADING PROCEDURES .................. 29
8.1 Installation of Brewcmd Software ............................................................................................ 29
8.2 Restore Back to COSMAC Mode ............................................................................................ 29
8.3 New Configuration Upload ...................................................................................................... 30
8.4 Uploading new Firmware and configuration files .................................................................... 31
8.5 Main Board flat cable connections ............................................ Error! Bookmark not defined.
9 WARRANTY .................................................................................................................................. 33
TABLE OF CONTENTS
MKIII SERVICE MANUAL
2
Recommendations by Environment Canada
Mark III Brewer Ozone Spectrophotometers are recommended by Environment Canada (EC) as the significantly superior model of Brewer instrument with which to measure ozone in the ultraviolet (UV) region of the spectrum. EC strongly discourages the use of other models of the Brewer instrument for the measurement of ultraviolet radiation or ozone in the UV because of the much poorer stray light performance of the single monochromator versions of the instrument.
MKIII SERVICE MANUAL
3
PREFACE
This document has been developed to aid an operator when a Brewer Spectrophotometer stops operating or whose operating characteristics have changed since leaving KIPP & ZONEN B.V.. It has been developed from the experience of KIPP & ZONEN and present users of the Brewer. The most likely electronic failures and ones which can be repaired in the field have been documented.
1 PURPOSE AND APPLICATION
The purpose of this manual is to help the operator diagnose problems with a Brewer Spectrophotometer to an assembly level. Component level diagnostics will not be covered since the replacement of many components may require instrument realignment or re-calibration. Potentiometers should not be adjusted nor set screws removed without completely defining the problem area. If the information in this document is not sufficient, it is recommended that the user contact KIPP & ZONEN before dismantling of any assemblies beyond what is called for in this manual. In several cases the operator will be asked to replace components that are not included in spares kits. In these cases please contact KIPP & ZONEN for information on obtaining the component.
2 EQUIPMENT REQUIRED
1) Volt-Ohmmeter, capable of measuring from 1 mV to 2000 volts DC.
2). A clip on or in-line current meter.
3). In place of the above items, a digital Volt-Ohmmeter, such as the one contained in the KIPP & ZONEN UV Stability Kit is adequate except for Voltage readings of greater than 1000V.
4) An oscilloscope may be useful (1 millivolt; 20 megahertz bandwidth).
5) An external UVB lamp may be useful when the internal quartz-halogen lamp appears at fault.
6) If the control computer is suspect, a second computer with Brewer software loaded may be needed.
7) A terminal communications program such as Telix, Procomm, Xtalk or Windows HyperTerminal.
MKIII SERVICE MANUAL
4
MKIII SERVICE MANUAL
5
3 MAJOR EQUIPMENT FAILURES
A major equipment failure is defined as a condition where there is no communication between the Brewer Spectrophotometer (Brewer) and the external control computer (PC). Symptoms are that no commands go to the Brewer, and no information is received by the PC. This is a "no operation" condition.
3.1 "NO OPERATION" - NO POWER INDICATION
1. Reset the Brewer
If any problem is fixed at any point in this trouble shooting procedure attempt to launch the Brewer Software. When the Brewer Software is launched the Brewer should send the following message to the PC:
BREWER OZONE SPECTROPHOTOMETER
#nnnnn
AES SCI-TEC
CANADA
VERSION 1 Jan 01, 1998
If this message is received then communications has been established between the Brewer and the control computer. If the instrument does not reset (as seen by motors initializing), the problem is probably in the Brewer, but may also be in the communications cable, the PC or the software. A fresh set of software can be loaded onto the Computer from the original discs to eliminate it as a potential problem. Ensure that the Brewer configuration files are preserved as they contain important operating information such as the COM: port number. Check the configuration files to ensure that none of the files have been corrupted.
2. Check Power
If the GREEN LED on the Brewer is illuminated, then there is power to the instrument, and the Main Power Supply is producing 5V. If the LED is not illuminated, then the AC power source should be checked at the AC outlet and the end to end continuity of the Power Cable tested. The AC power should then be connected directly to the Brewer, bypassing the Tracker. If there is still a no-power indication, then the Brewer top cover should be removed and the MAIN power supply checked to see if it feels warmer than ambient.
3. Check Power Switch and Fuses
If the cover is not warm, then perhaps the Power Switch is defective, or a fuse requires replacement. With power disconnected, use an Ohmmeter to confirm that the Switch is functional, and check the two fuses next to the power supply cover as indicated in the photograph. If a fuse needs changing, ensure that AC power is disconnected, and replace the fuse with the appropriate spare. To check the fuse inside the power supply, disconnect AC power from the Brewer, remove the Main Power Supply cover (four 6-32 screws) and measure or visually inspect the fuse and replace it if necessary. When re-installing the cover, ensure that the Power Supply connectors remain connected.
3 MAJOR EQUIPMENT FAILURES
MKIII SERVICE MANUAL
6
If the power supply appears to be normal, then the LED may be defective, or there may be a faulty connection.
4. Measure Voltages The main power supply voltages can be checked at the test points on J23 of the Main Electronics board. (Refer to the chart for correct voltages).
Connector J23 Monitor points
Pin 1 Pin 2 Pin 3 Pin 4 Pin 5
Pin 6
Pin 7
Pin 8
Pin 9 Pin 10 5v
5v
-5v
12v
12v
-12v
24v
2.5v
4.5v
Gnd
3.2 TROUBLE SHOOTING SEQUENCE FOR "NO OPERATION" CONDITION
It has been established that there is Power, motors reset, but that there is absolutely no communication between the PC and the Brewer.
1. Reset A software reset should be attempted after the "Brewer Failed to Respond" message appears on the PC monitor, or the PC has tried a number of times to establish contact with the Brewer. The
software reset is done by pressing key on the PC following the error message, or launching the Brewer software from the DOS prompt. If this fails to establish communications, the PC should be given a ‘cold’ boot (power OFF/ON).
2. Check cables If there is still no response, the cables should be inspected to ensure that a connector has not been inadvertently removed or pulled out of place. The Communications cable to the Brewer from the PC, as well as the power cable to the Brewer should be checked. All cables involving the linking of peripherals should be checked to ensure that they are plugged in as well. After this has been done another software reset can be tried.
3. Test serial port Check that the computer serial port is operating correctly by running a serial communications program such as Telix, Procomm, or Windows Hyper Terminal. Make sure the serial communications program is set to 1200 baud, is in full duplex mode, and the correct serial port has been selected. Refer to the communications program documentation for correct operation. Remove the RS422 adapter from the back of the computer and connect pins 2 and 3 of the computer serial port together. While running the communications program in full Duplex mode, characters typed should be seen echoing to the monitor. Disconnecting the jumper wire should stop the echo-back. This test confirms that the computer serial port is operating correctly. If the computer did not echo the characters change the Brewer’s configuration to use another port (see section 3.6 in the Operator’s Manual) or have the defective serial port repaired before doing any further tests.
MKIII SERVICE MANUAL
7
4. Test RS422 adapter The RS422 adapter at the computer can be tested by removing the communications cable from the adapter, connecting pins 1-3, plugging the adapter into the Computer COM: port, and checking for echo-back as in step 3. Now switch connections to pins 2-4 and test again. The RS422 adapter is operating correctly if characters echo back. If a fault is indicated, then either the adapter or the power module is defective. The power module output should be approximately +13V DC with no load or +9V with a load (the centre pin is positive.) If the voltage is normal, then replace the RS422 adapter and test again.
5. Test communications cable If the tests have passed to this point, the communications cable between the Brewer and the computer can be tested. Reconnect the cable to the RS422 adapter and connect it back to the computer. Remove the communications cable from the Tracker surge suppressor box and connect a wire jumper to connect pins C-B. Test the cable by sending characters from the computer as in step 3; move the jumper to connect pins I-J and test it again.
Reconnect the communication cable to the tracker and disconnect the communications cable at the Brewer. Short pins C-B and then pins I-J and test as in step 4. If the test passes, reconnect the cable to the Brewer and open the cover of the Brewer. Turn the power off and disconnect the IDC connector (the IDC connector,J7, is the connector attached to the shielded ribbon cable shown in the picture below) Use a small wire jumper to connect pins 2-8, and then pins 4-6, using the echo­back test each time. (The IDC pins begin from 1 at the triangle stamped on the side of the connector and increment in odd numbers on the same row - 1, 3, 5, 7, 9 on the first row and 2, 4, 6, 8, 10 on the next row, in the same direction). If the test fails at any point of this cable test procedure the faulty connection should be repaired before proceeding to the next test. If all of these procedures have positive results, then the problem lies in the Brewer Main Electronics Board. Reconnect all the cabling to the instrument and continue to the next test.
As an alternative for testing cables, each line can be tested for continuity by measuring point to point with a ohmmeter, using the appropriate schematic as a guide. The resistance through the surge arrestor is 10 ohms so that a typical resistance should be 12 ohms from the RS422 adapter to the IDC connector at the Brewer Main Electronics board.
3 MAJOR EQUIPMENT FAILURES
MKIII SERVICE MANUAL
8
6. Line Driver
If the reset of the motors is heard and seen, but there is no message sent to the PC, then there may be a problem with the line driver on the Main Electronics board. Ensure that power is turned off and remove the Main Electronics cover plate and replace IC U11, (on a socket 2 inches up and 2 inches to the left of the LED). Make sure to use proper grounding precautions before touching the electronics board. Leave the cover plate off and turn on the instrument. The motors should be observed to reset and if the computer was left in the serial communications program, the Brewer reset message will be displayed on the screen. Exit the communications program and launch the Brewer operating program. Normal Brewer commands can be used for checking correct operation.
7. Check Mode of Microprocessor
If there is still no operation, there may be a firmware problem, or a problem on the Main Electronics Board, so the functionality of this board should be tested. There is a Red status LED located near the bottom left hand side of the board - it should be flashing slowly in ‘Cosmac mode’ as per the following table. If the Brewer is in some other mode change back to the Cosmac mode (step 10).
LED Mode indication
¼sec
¼sec
¼sec
¼sec
¼sec
¼sec
¼sec
¼sec
Cosmac mode normal operation
On
Off
Off
Off
On
On
On
Off
Cosmac mode with configuration missing
On
Off
Off
Off
On
Off
On
Off
Loadmode waiting for code to be downloaded
On
Off
On
Off
On
Off
On
Off
Loadmode with bad flash memory
On
On
On
On
Off
Off
Off
Off
Loadmode with good flash memory
On
On
On
On
On
On
On
Off
Opmode with configuration present
On
On
Off
Off
On
On
Off
Off
Opmode without configuration present
On
On
Off
Off
On
Off
On
Off Checking Flash memory
On
Off
On
Off
Off
Of
Off
Off
Changing from load to cosmac or load mode
On
On
On
On
On
On
On
On
MKIII SERVICE MANUAL
9
8. Check Voltages
If the LED is not flashing, then the correct voltages should be confirmed for the power supply on the Main Electronics board J23, as per section 3.1.
9. Test Main Power Supply
If the power supply values have changed, the power connections should be checked. If they appear normal, and the voltages are still not at proper levels, replacement should be considered. However, there may be a short circuit in one of the assemblies causing the power supply to be loaded. Shut the power off and disconnect the power connector J1 from the Main Electronics board and from the J4 Lamp control board. Repeat the voltage measurements at the connector and see if all the voltages are present. If the voltages are correct then determine which assembly is at fault by installing the connectors one at a time. Replace the main power supply if necessary, and ensure when first turned on that the assemblies are disconnected to allow setup of voltages to correct values. The main 5 volts monitored at the main board should be set to 5.0 volts with the single potentiometer on the main power supply.
10. Restore Brewer to Cosmac Mode
If the flashing LED on the Main Electronics board indicates that it is not in the cosmac mode, and the software has not been successful in correcting it during the Reset sequence, then the following procedure should be followed:
A. Exit the serial communications program and change to the subdirectory c:\bdata\nnn.
B.Type BREWCMD , and observe that the instrument has responded and note which mode is indicated. The program should display the same mode as the mode observed on the status LED.
C. If the mode is ‘cosmac mode’ then type loadmode , and the display should return indicating ‘loadmode’. Now type opmode and within a few seconds the Brewer will be in opmode.
3 MAJOR EQUIPMENT FAILURES
MKIII SERVICE MANUAL
10
D. Type readlog to display past error history, each line displayed will be a possible clue to the problem that caused the failure. If motor failures have occurred, the status will tell you which motor and what the failure is. Usually it is a sensor problem or a wiring problem that prevents the motor from resetting. If no problems are displayed, then the system can be placed back into cosmac mode for further testing.
E. Type loadmode and following the return of the prompt, type cosmacmode . The motors will reset and within a minute the system will display the status that it is in cosmac mode.
F. Quit the Brewcmd program by typing quit .
11. Run Brewer Software.
Run the Brewer operating program by typing BREWER to check the correct operation of the instrument. Select routines that you are familiar with and ensure that all functions and data appear normal.
12. Replace Main Electronics Board
If no communications return when running Brewcmd then there may be a problem with the Main Electronics Board. Replace the Main Electronics board with a spare board, ensuring that all cables are reconnected to the original connectors. Normally the spare Main Electronics board is preloaded with the same parameters as those that came with the instrument. Therefore no reprogramming or uploading of the configuration parameters are needed. If the configuration parameters have changed, then refer to the procedure to upload new firmware and configuration data to the board, section 8.
The main board has a number of flat cables attached to it. The following lists the connectors for the flat cables and the electronics they are connected with. “Top” (at the front) stands for pins 1-16. “Bottom” (at the back) are pins 17-32:
J2 top BA-W77/A Lamp Board J2 bottom BA-W77/B J1 on hv module J13 top BA-W76/B iris J13 bottom BA-W76/A zenith J14 top BA-W76/B fw #1 J14 bottom BA-W76/B fw#2 J15 top BA-W76/A micrometer #2 (MKIII only) J15 bottom BA-W76/A micrometer #1 J16 bottom BA-W76/C fw#3 (MK IV only} J17 top BA-W76/B slitmask (shutter) J17 bottom BA-W76/A azimuth J12 top BA-W76/A control switch’s J7 top Communication J7 bottom humidity sensor
13. Contact KIPP & ZONEN
If you have followed this procedure and not located the problem contact KIPP & ZONEN for further assistance.
Webpage: www.kippzonen.com Email: info.holland@kippzonen.com
MKIII SERVICE MANUAL
11
4 OPERATING TEST FAILURES
In the Brewer Spectrophotometer, most of the operating tests are done using either the mercury lamp or the quartz-halogen lamp. If any of the other tests fail, the user should proceed immediately to a mercury lamp test (HG) or a standard lamp test (SL). Mercury lamp test failures and standard lamp test failures are two major problems which can occur. In both cases calibration of the instrument may have been compromised. For the purpose of troubleshooting, these two sections have been separated from the main part of the instrument. The Light Detection System is made up of the zenith prism, foreoptics, spectrometer, slitmask, PMT, High Voltage Circuitry, High Speed Amplifier (part of the PMT), interconnecting harnesses, and the Main Electronics board which houses the photon counting circuits. Without this section being operational, no testing, or measuring, can be done. There are a few points to remember in the event there is a major failure somewhere in the system, whether it is the Mercury Lamp, Standard Lamp, or Light Detection System. The error message "lamp not on test terminated" is an indication of a failure, either in the lamps or in the Light Detection System. Changes in dark count may be symptomatic of a number of problems which may occur in the Brewer. If Dark Count begins to increase or becomes erratic, the slitmask may be at fault. Note that it is normal for Dark Count to increase with temperature. Dark Count changes may also indicate motor power supply problems, PMT performance problems, or High Speed Amplifier problems. If the dark count changes, there may be a changing optical condition in the Light Detection System. A higher dark count can also be the result of a poor ground somewhere in the system or high humidity inside the instrument. Grounding connections should be checked, especially between the PMT housing and the Main Electronics. The desiccant should be changed more frequently in higher humidity locations.
4.1 MERCURY LAMP CIRCUIT
The Mercury lamp circuitry (Fig 10-4.2) is a constant current source designed to keep the lamp current constant throughout the temperature range of the instrument. A voltage regulator, connected to a specially selected resistance wire, regulates the lamp current with minimal circuitry.
4.2 TROUBLE SHOOTING A MERCURY LAMP TEST FAILURE
The Mercury lamp test fails, and software responds with "lamp not on test terminated".
1. Check Motor Positions
Retry the test and look through each Viewing Port to see if the lamp turns on. Ensure that the zenith prism and the filter wheels are in the correct position.
4 OPERATING TEST FAILURES
MKIII SERVICE MANUAL
12
2. Check Temperature
Observe the temperature of instrument (TE). If the Brewer temperature is less than 0 degrees Celsius, the lamp may be too cold and may not fire. If this is the case, the Standard Lamp may be turned on for a few minutes to warm the Mercury lamp.
3. Check Lamp Voltage and Current
There is the possibility of a lamp failure. Perform an AP (A/D voltages printout) test. If the Mercury Lamp current is 0.0, and the Mercury Lamp Voltage is near 15 volts, then the lamp is probably bad. Measure the voltage across the lamp, pins 6 and 9 on P111. (Reference Figure 10.6-2.2). It should be approximately 0 volts with the lamp off (B0) and 13 volts with the lamp on (B1).
4. Mercury Lamp Replacement
To replace the bulb, loosen the two thumb screws (item 6, Figure 10.6-2.2) and carefully withdraw the lamp holder. Take care not to touch the new lamp with bare hands. Ensure the lamp is tight in its socket and cleaned with isopropyl alcohol after replacement. The quartz-halogen lamp should also be checked at this time, as blackening of the glass could reduce the amount of HG light which is passing through it to the foreoptics. Re-assemble the HG lamp by reversing the above steps and test the lamp with B1 and HG commands.
5. Check Micrometer Position
If the lamp is working fine, and light can be seen in the Viewing Ports, then the position of the micrometer should be checked. Remove the “Spectrometer” cover and check to see that the micrometers are in the correct position as indicated in the Final Test Record, and on the Log Sheet taped to the Spectrometer cover. If the Micrometers are not where the Log Sheet indicates, they should be adjusted manually and an HG test attempted again.
6. Attempt a Standard Lamp Test
If the test still fails, the problem could be in a misalignment of a component in the Spectrometer, a problem in the slit mask, or a fault in the PMT or counting circuitry. Attempt a standard lamp test. The standard lamp has a much higher intensity and does not depend on the position of the micrometer and slitmask as much as the mercury lamp test. If the standard lamp test operates correctly, the SL printout should be examined carefully to see if any characteristics have changed. If the dark count has increased significantly, this may be an indication that the alignment has changed and may require a mirror adjustment. Check to see if any of the intensities or ratios have changed, which may indicate that the optics have deteriorated, or the PMT performance or counting has degraded. If the standard lamp test results appear to be normal, then KIPP & ZONEN should be contacted before proceeding.
7. Measure High Voltage
If the standard lamp test is not operational, then it can be assumed that there is an optical failure, or a PMT failure. If the mercury lamp test is normal while the standard lamp has failed, this usually indicates a problem with the lamp or the circuitry that controls the lamp. The high voltage should be measured using a voltmeter and looking at test point E16 on the Main Electronics board. There should be a voltage between 2.5 volts and 5 volts, which has been factory set (see final test record for the setting), and should not be adjusted. This voltage is equal
MKIII SERVICE MANUAL
13
to the actual high voltage divided by 409. There may be a variation of a volt or two of the high voltage, but the voltage at test point should be very close to the final test record value. If it is significantly different, the high voltage circuit may have failed. This will require a skilled technician for repair or KIPP & ZONEN should be contacted.
8. Check Light Detection System
If the high voltage appears to be normal from the test point reading, then there is a possibility that something has failed in the Light Detection System, reference Section 4.7.
4.3 STANDARD LAMP CIRCUIT
The standard lamp current regulator is similar to the Mercury lamp circuit except the resistance wire is chosen to regulate the supply current with less variation than the mercury lamp. The circuit maintains the operating current within 3 milliamps over a range of -20 to + 40 degrees C, and is even more stable in the operating range of 10 to 30 degrees.
4.4 STANDARD LAMP FAILURE
The quartz-halogen (Standard) lamp is the lamp most frequently used in self diagnostics of the Brewer. With it, calibration can be monitored, and correct operation of many systems can be inferred. The software for the standard lamp tests has been designed to give an error message in the event of a failure. The message is "Lamp not on test terminated". The message may be somewhat misleading in that there are many conditions which can make it appear that the Lamp is not on. These conditions include PMT performance, filterwheel positioning, Zenith prism position, photon counting electronics, and the I/O electronics used to turn the lamp on and off.
4.5 TROUBLE-SHOOTING STANDARD LAMP TEST FAILURE
Error message, "Lamp not on test terminated".
1. Check if Lamp is Burned Out
Retry the test and if the lamp is on, it will be seen through the viewing ports if the Filter Wheels, iris, and the Zenith Prism have moved into the proper position. Perform an AP (A/D voltages printout) test and compare results with previous tests, paying particular attention to standard lamp current and voltage. The current, channel 14, will be approximately zero and the voltage, channel 15, will exceed 14 volts if the lamp is unplugged or burnt out.
2. Attempt HG Test
If the test continues to fail, a mercury lamp test should be attempted. If the mercury lamp test operates correctly, then the Light Detection System is probably operating correctly. If so, a mercury lamp calibration should be done and the Standard Lamp test should be repeated.
4 OPERATING TEST FAILURES
MKIII SERVICE MANUAL
14
3. Observe Slitmask Performance.
If the mercury lamp test is normal, the spectrometer cover should be removed and another standard lamp test done while observing the slitmask. If the standard lamp fails again while the slitmask is operating correctly, KIPP & ZONEN should be consulted.
4. Photon Counting Circuitry
If the mercury lamp test is not normal, there is probably a failure in the photon counting circuitry and section 4.9 of this manual should be consulted.
5. Replace Standard Lamp
Before starting disassembly, it should be confirmed that the lamp filament is open circuit by checking with an ohmmeter across pins 1 and 5 of J111 (lamp connector). Reference fig. 10.6.2-2. The Standard Lamp is located above the mercury lamp and is attached to a removable plate. Remove the two screws holding the plate. Replace the bulb with a new one, taking care not to touch it with bare fingers- use a tissue or soft cloth and clean the lamp with isopropyl alcohol after installation. If the bulb is blackened, it should be replaced regardless whether or not it is operational. Replace the connector assembly and test the lamp with the B2 command. Be sure to use B0 to turn the lamp off. Another standard lamp test should now be tried. If there is no success, there is a possibility of circuit failure on the Lamp Control board, or the Main Electronics board.
6. Check Main Electronics Board
Send the command to turn the Standard Lamp circuit on (B2), and measure the voltage at pin15 of U14 on the Main Electronics Board. The voltage should be +5 volts when the lamp is off and 0 volts when the lamp is on. If this voltage does not change, the Main Electronics board should be replaced, or KIPP & ZONEN consulted.
7. Check Cable
If there was an indication of a 0 to 5V voltage swing at pin 15 of U14, but still no lamp operation, the cable that connects the Main Electronics board to the Lamp Control board should be checked.
8. Test Lamp Control Board
If the cable it is good , the Lamp Control Board should be tested for correct operation. The lamp control for the standard lamp is very simple - Q1 pin 2 supplies power to the current regulator VR2 (both components are on the bottom edge of the Lamp Control Board). Measure Q1 pin 2, and when the standard lamp is commanded on, it should go to 17 volts. VR2 pin 2 should be 10 volts. If it this point does not go to 10 volts, then the current regulator circuitry is faulty and must be repaired or the complete board must be replaced. Refer to the lamp control board schematic Fig
10.4-2.
9. Test Lamp Cable
If the voltage at VR2 does change on command, then the lamp cable should be tested for continuity.
MKIII SERVICE MANUAL
15
4.6 LIGHT DETECTION SYSTEM FAILURE
A Light Detection System failure may show up in any or all of the diagnostic or measurement tests of the Brewer. In Lamp diagnostics it will usually be shown on the Computer Display as "Lamp Not On .. test terminated” message. In a measurement (DS for example), the counts may simply go to zero, or some other incorrect number, depending on the nature of the failure. Results of Lamp tests are predictable so an HG and/or SL should be attempted if there is any suspicion of a Light Detection problem. The Light Detection System is comprised of the foreoptics, spectrometer, micrometers, slit mask, and the PMT (including the high speed amplifier). Adjustments of any of these assemblies should be done with extreme caution, unless they are specified in this manual, as Instrument calibration may be affected.
4.7 TROUBLE-SHOOTING THE LIGHT DETECTION SYSTEM
1. Confirm Motor Positions are Correct
If Lamp tests fail and the message "Lamp not on test terminated" is displayed, the SL command should be issued. With the lamp on, the correct positions of the zenith prism, filterwheels, and iris
should be confirmed. The correct positions are printed to the screen when the SL command is issued. Light from the lamp should seen in both view tubes.
2. Test Photon Counting Circuit
If the SL test fails the photon counting circuitry should be tested. Turn off the power, remove the ribbon cable from the PMT output, and plug it into J22 of the Main Electronics Board. This connector is the output of a 1MHz Pulse Simulator (see picture). Turn on the Brewer power and from the Main Menu enter the Teletype command, TT. At the teletype prompt, enter the command R,0,7,20;O. The display should show 8 numbers, each being approximately 2,280,140 counts. Exit TT by pressing the Home key, turn off Brewer power, replace the PMT cable, and turn power back on. If the test is normal, the Counting System is operating correctly. Now check components in the Spectrometer.
3. Check Micrometer Position
In a low light level and dry room, remove the Brewer cover and the Spectrometer cover. Confirm that the micrometers are in the correct positions as indicated on the ‘Micrometer Position’ log sheet that is taped to the Spectrometer cover.
4. Check Mirror and Grating
Check that the spherical mirrors, the gratings, and the two flat mirrors appear to be in place, and that their mounting assemblies appear normal. ** TAKE CARE NOT TO TOUCH THE SURFACES OF EITHER THE MIRRORS OR THE GRATINGS AS PERMANENT DAMAGE MAY OCCUR **** .
4 OPERATING TEST FAILURES
MKIII SERVICE MANUAL
16
5. Test Slitmask
Slitmask operation can be confirmed by returning to Telytype operation (TT), sending the R,0,7,20 command, and noting that the slitmask moves back and forth pausing briefly at each of the 8 slitmask positions. Commands R,X,X,20 (where X =0 to 7) can be sent in sequence. After each command a slot in the slit mask should line up with a slot in the slit plate EXCEPT on command R,1,1,20, which is the Dark Count position. If the slitmask does not move, or appears to move incorrectly, then the I/O cable, Motor, or the motor control micro or motor driver IC on the Main Electronics board may be faulty. The Main Electronics board should be replaced and another test attempted.
6.Check Dark Count
If the slitmask appears to be functioning correctly note what the count values are when the command R,0,7,20;O is issued in Teletype. There is a very important difference between a count rate of zero and a count rate near zero. A high count rate indicates correct operation. A low count rate indicates that only dark count (a function of the PMT) is being measured and the light coming through the slit mask is not being measured. A zero count rate indicates a problem with the PMT or its electronics.
7. Measure High Voltage Test Point
Replace the Spectrometer cover, and remove the cover from the Main Electronics board. Measure the high voltage test point E16 on the main electronics board (see picture), if it is not the same as the value given in the Final Test Record, section 5.0, then adjust R4 on the high voltage control board.
8. Check High Voltage Cable
If the test point voltage cannot be achieved, there may be a short circuit in the high voltage line somewhere between the HV module and the actual PMT dynode chain. Care should be taken when inspecting this area because there can be as much as 1800 volts at the connection points. Check the high voltage cable for continuity.
9. Measure High Voltage at PMT
The rear cover of the PMT housing can be removed (turn CCW) and the actual high voltage measured on the internal connector. This measurement must be done with a voltmeter rated for 2000 volts or by using a high voltage probe. If the high voltage is still not as indicated on the Final Test Records, then contact KIPP & ZONEN.
10. Test High Speed Amp Board
If high voltage is correct, there may be a failure in the high speed amp board. TP1 on the high speed amplifier board (inside the PMT housing) should read -10 millivolts (referenced to the HV connector shield) or the value recorded in the Final Test Record, section 4.0. If the value of this test point is incorrect, an adjustment of the on-board potentiometer can be attempted. If this test fails, replace the high speed amplifier board. If this test point is normal, the photon counter cable should be inspected to ensure that all connections and wires are in tact. Finally, an oscilloscope can be used to monitor test point TP4 on the high speed amp board, and then pins 1,2,3,4, and 12,13,14, 15 of the IC, SN75114N. There should be clean square waves, the frequency of which should vary with light intensity.
MKIII SERVICE MANUAL
17
If these signals are normal, consult KIPP & ZONEN before proceeding further. If any of the signals are missing or are not clean square waves, replace this board.
11. Contact KIPP & ZONEN
If all of these tests fail, there is a probable fault in the PMT, and KIPP & ZONEN should be contacted before proceeding.
4.8 STEPPING MOTOR CONTROL
The stepping motor control circuitry provides for the control of each stepping motor through individual motor microprocessors and stepping motor driver integrated circuits. This type of motor control allows the system to move the motors simultaneously, and to monitor the sensors very efficiently. All of the motor microprocessors are controlled by the main processor through a serial "I2C" bus. Each motor is optimized for speed and taylored to each function in the Brewer. The configuration file is included with the control software to allow changes and updates to the system. The configuration file is optimized for each instrument and normally does not require operator modifications unless recommended by KIPP & ZONEN.
4.9 PHOTON COUNTER CIRCUITRY
A schematic diagram of the photon counter circuitry is given in figure 10.4-1. The photomultiplier signal which has been amplified and divided by a line driver on the pulse amplifier board is received by a line receiver on the Main Electronics board. This signal is fed to a pair of binary counters. The outputs of the 2 counters are compared and if they differ by more then a set amount the second counter is disabled and a “PMT counter failure” message is generated in the Firmware Log. The counters are each connected to output ports which are connected to the data bus of the system microprocessor. The main gate is generated by an electronic circuit that provides a very precise gating signal to enable counting for a specific length of time. The microprocessor decides when the pulse counting circuitry is initiated and the circuitry then triggers the gate to allow counting of the photons.
4.10 HIGH SPEED AMP BOARD
The photomultiplier, and pulse amplifier board are enclosed inside the photomultiplier subassembly. The sub assembly is designed to allow access to the pulse amplifying circuitry without upsetting the alignment of the PMT or the optics. The photomultiplier is an EMI 9125UVA type. It is enclosed in a magnetic shield at the high voltage potential of the photo cathode. A small area on the shield is open to allow light to fall on the cathode. A high voltage power supply provides stepped voltages to the photomultiplier through a resistor divider circuit at the base of the tube. Since differential thermal expansion between the photomultiplier glass and the PTFE base may cause tube breakage under extreme temperature conditions, it is recommended that the photomultiplier housing never be subjected to temperatures of less than -50C or greater than +60C. The pulse amplifier board is located directly behind the base of the photomultiplier tube. This circuitry accepts the photon pulse signals from the photomultiplier, amplifies them, discriminates the signal level from current leakage, divides the amplified photon pulses by four, and finally outputs the pulse on a line driver. Since the circuitry is extremely sensitive to feedback and RF noise, it is located in close proximity to the photomultiplier.
4 OPERATING TEST FAILURES
MKIII SERVICE MANUAL
18
4.11 RATEMETER
The ratemeter circuit provides a voltage output of the photon count rate. It consists of a two­transistor charge pump on the high speed amp board, and an integrator-amplifier on the Main Electronics board. A transistor is connected as a current pump and driven by one of the outputs of a J-K flip flop on the high speed amplifier board. The pulses from the flip flop are integrated by an RC network then scaled by an op-amp.
4.12 TEMPERATURE MONITOR
The temperature monitor is a temperature sensitive bridge using YSI linearized thermistors. The bridge output is amplified to produce a suitable scale for the A/D converter. There are six thermistor circuits available, one of which is located in a hole drilled in the side of the front flange of the PMT housing. The other five are used to measure temperature at points around the instrument as status information. One of the sensors monitors outside temperature.
4.13 HIGH VOLTAGE CIRCUITS
The high voltage (600-1600 volts) for the photomultiplier (PMT) is provided by a DC-DC converter / high voltage control circuit board.
4.14 CLOCK / CALANDAR
The real time clock/calendar circuit is a battery backed up clock with integral static random access memory. A lithium battery is used to keep the clock operating when there is no power applied to the Brewer. The clock is year 2000 compatible.
4.15 A/D CONVERTER
The A/D converter has a 10-bit conversion resolution and is built into the microprocessor. Analog multiplexers are utilized to select the channel to monitor. The AP routine is the main routine that accesses the data and displays the information on the PC monitor, saves it to a file, or prints it on the printer.
MKIII SERVICE MANUAL
19
4.16 BREATHER DESICCANT CARTRIDGE
The breather desiccant cartridge allows the Brewer to compensate for normal pressure changes, while preventing the influx of moist air. Note that the breather desiccant does not remove moisture from the Brewer. Extra desiccant is needed to remove moisture from inside the instrument. Breather desiccant is housed in a plastic container with a breathing tube placed at the bottom of the cartridge. The other end of the tube is connected to a port at the bottom on the underside of the Brewer and is exposed to the outside air. As the air is drawn in from the outside, the air is dried as it passes the desiccant. Reference figure 10.8.
4.17 HUMIDITY SENSOR (OPTIONAL)
The humidity sensor is a monolithic IC sensor that provides a voltage output proportional to relative humidity. The sensor is buffered by an Op Amp and the output is connected to one of the analog input channels of the A/D converter. A temperature sensor is located near the humidity sensor to provide temperature compensation for the device. The two measured values, relative humidity and temperature allow the absolute moisture to be calculated. The absolute humidity (grams/m3) value is displayed on the PC screen along with a numerical “moisture index” value. Refer to Figure 10.4-5 for the sensor specifications and Figure 10.4-6 schematic for the humidity sensor assembly.
4 OPERATING TEST FAILURES
MKIII SERVICE MANUAL
20
MKIII SERVICE MANUAL
21
5 TEST RESULTS OUT OF SPECIFICATION
Whenever any diagnostic tests are out of specification, (SL, HG,DT,RS,) the following should be done or considered.
1. If the deadtime has increased or decreased significantly, the High Speed Amplifier board may be defective, the slitmask may have become misaligned, or a ground in the instrument wiring may have become resistive. High moisture levels inside the instrument may also cause abnormal deadtime results.
2. A test lamp may have deteriorated to the point where tests are affected, and the lamps may require replacement. They can be inspected as mentioned in the previous sections, and replaced if necessary.
3. The optical surfaces within the zenith prism may have become smudged with a fingerprint or may have become dusty over a period of time. The prism should be cleaned, referring to section 6.2.
4. The micrometer may be "sticking" and not in its exact location after a measurement or a test. It should be inspected and the drive mechanism carefully cleaned with tissue and isopropyl alcohol.
5. Dark count is a convenient characteristic to monitor as it is printed out in many tests and measurements. Dark count changes may be the result of moisture, changes in the slit mask, PMT, high Voltage, or counting circuitry.
6. When the dark count of an instrument changes, or becomes erratic, a slitmask Run-Stop test should be done to see if there are any changes in the slitmask circuitry. If this does not give any indication of the problem, there may be a ground somewhere which has become resistive, and it should be checked.
7. The High voltage test can be done, and the results compared to those in the Final Test Records.
8. Generally speaking PMT problems are not field repairable.
9. Desiccants should be changed at regular intervals. The frequency of desiccant change depends on the climate. Desiccants will require changing much more frequently in humid environments. Alternately the humidity indicator can be used to indicate that desiccant needs changing – desiccant should be changed if the moisture index provided be the electronic humidity sensor is greater then 10 or if the paper indicator for 40% humidity is no longer blue. Desiccants can be removed and dried overnight at approximately 60 degrees C if necessary.
If these checks fail to locate the problem, KIPP & ZONEN should be consulted as to further investigation.
5 TEST RESULTS OUT OF SPECIFICATION
MKIII SERVICE MANUAL
22
MKIII SERVICE MANUAL
23
6 OPTICS CARE AND CLEANING
6.1 OPTICS CARE
- Never touch the polished surface of optics.
- Hold an optic only by its edges.
- Reduce the need for cleaning optics as much as possible.
6.2 OPTICS CLEANING
- Use latex gloves when handling or cleaning optics.
- Do not reuse cleaning materials.
SPECTROMETER MIRRORS, ND FILTERS, QUARTZ PRISMS AND LENSES: In the event of dust, blow off with rubber hand pump (avoid using your breath) or use bottled dry nitrogen to remove particles. In the event of grease or fingerprints, dampen a very soft cloth with methyl ethyl alcohol and wipe gently. NOTE: Wipe marks will almost certainly remain. To remove the wipe marks, dampen a soft tissue with a mixture of ether and isopropyl alcohol and wipe gently. The tissue should be dampened to the extent that while wiping, the mixture evaporates one to two millimeters behind the tissue.
QUARTZ DOME AND EXTERIOR QUARTZ WINDOW: Spray with window cleaner and wipe clean with a soft cloth.
LAMPS: Dampen a soft cloth with methyl or ethyl alcohol and wipe gently.
POLARIZING FILTER: Use only a soft, dry cloth or tissue to remove dirt and finger marks. The filter should always be wiped gently to avoid marring the finish.
DIFFRACTION GRATINGS: Do NOT touch or attempt to clean.
6 OPTICS CARE AND CLEANING
MKIII SERVICE MANUAL
24
MKIII SERVICE MANUAL
25
7 AZIMUTH TRACKER
Problems in the Azimuth Tracker affect the Brewer’s ability to track the sun, and thus any measurement that depends on direct solar radiation may be in jeopardy. Regardless of the problem, it is usually manifests itself as frequent failures of the Solar Siting test (SI), or the Steps per Revolution test (SR). The Main Electronics Board sends motion commands to the Tracker, and receives sensor indications from the Tracker via the Azimuth Tracker connector on the front of the Brewer base. Indirect problems can affect Brewer power and communications since both of these cables pass through the Tracker housing assembly. Communications is routed through a surge suppressor assembly mounted on the bottom of the Tracker, and Power is supplied to the Brewer via the Tracker enclosure wherein a second set of surge suppressors is located. If the surge suppressors in the communications line are faulty, a no-communications condition can exist. The communication cable from the Computer can be connected to bypass the Tracker as a troubleshooting aid in determining if the surge suppressor assembly is at fault. Similarly, power can be connected directly to the Power Connector on the Brewer in the event that a “no power” problem exists in the Brewer. The Power switch on the Tracker will not affect power to the Brewer, but a failure of the fuse in the Tracker will interrupt both Brewer and Tracker power. Tracker problems can be either mechanical or electrical.
7.1 BREWER POWER PROBLEMS.
Power for the Brewer passes through the Tracker Case where it is connected to a set of MOV surge suppressors. In the event of large transient voltages on the power cable, one or more of these components may become damaged, and subsequently result in AC power short circuits which may affect power to both the Brewer and the Tracker. Connecting the AC power directly to the Brewer and observing if operation is restored is a method of troubleshooting this subassembly. An ohmmeter can then be used to determine the specific component at fault. Power connections in the Tracker, including the Fuse, can also be checked in the event of power problems in the Brewer.
7.2 BEWER COMMUNICATIONS PROBLEMS
A failure in the surge suppressor in the communications link can be checked by connecting the communication cable from the Computer directly into the Brewer. If Brewer/Computer communications are restored, the surge suppressor subassembly is likely at fault.
7 AZIMUTH TRACKER
Loading...
+ 101 hidden pages