Agilent N8972A Quick Reference Guide

Noise Figure Analyzers
NFA Series

Quick Reference Guide

Manufacturing Part Number: N8972-90003

May 2000

© Copyright 2000 Agilent Technologies

Safety Notices

This instrument has been designed and tested in accordance with IEC Publication 61010-1+A1+A2:1991 Safety Requirements for Electrical Equipment for Measurement, Control and Laboratory Use and has been supplied in a safe condition. The instruction documentation contains information and warnings which must be followed by the user to ensure safe operation and to maintain the instrument in a safe condition.

The information contained in this documentissubject to change without notice.

Agilent Technologies makes no warranty of any kind with regard to this material, including but not limited to, the implied warranties of merchantability and fitness for a particular purpose. Agilent Technologies shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material.
The following safety symbols are used throughout this manual. Familiarize yourself with the symbols and their meaning before operating this instrument.
WARNING Warning denotes a hazard. It calls attention to a procedure
which, if not correctly performed or adhered to, could result in injury or loss of life. Do not proceed beyond a warning noteuntil the indicated conditions are fully understood and met.
CAUTION Caution denotes a hazard. It calls attention to a procedure that, if not
correctly performed or adhered to, could result in damage to or destruction of the instrument. Do not proceed beyond a caution sign until the indicated conditions are fully understood and met.
ii
NOTE Note calls out special information for the user’s attention. It provides
operational information or additional instructions of which the user should be aware.
The instruction documentation symbol. The product is marked with this symbol when it is necessary for the user to refer to the instructions in the documentation.
This symbol is used to mark the on position of the power line switch.
This symbol is used to mark the standby position of the power line switch.
This symbol indicates that the input power required is AC.
WARNING This is a Safety Class 1 Product (provided with a protective
earthing ground incorporated in the power cord). The mains plug shall only be inserted in a socket outlet provided with a protected earth contact. Any interruption of the protective conductor inside or outside of the product is likely to make the product dangerous. Intentional interruption is prohibited.
WARNING If this product is not used as specified, the protection provided
by the equipment could be impaired. This product must be used in a normal condition (in which all m eans for protection are intact) only.
WARNING No operator serviceable parts inside. Refer servicing to qualified
personnel. To prevent electrical shock do not remove covers.
iii
WARNING For continued protection against fire hazard, replace line fuses
only with the same type and ratings (115V range; type F 5A 125V; 239V range F 5A 250V). The use of other fuses or materials is prohibited.
CAUTION To prevent electrical shock, disconnect the instrument from the mains
(line) before cleaning. Use a dry cloth or one slightly dampened with water to clean the external case parts. Do not attempt to clean internally.
Environmental requirements: This product is designed for indoor use only and to meet the following environmental conditions:

• Operating temperature: 0° Cto+55° C

• Operating humidity: <95% relative
• Altitude: up to 4500 m

Warranty

This Agilent Technologies instrument product is warranted against defects in material and workmanship for a period of three years from date of shipment. During the warranty period, Agilent Technologies Company will, at its option, either repair or replace products which prove to be defective.
Forwarranty service or repair,this product must be returned to a service facility designated by Agilent Technologies. Buyer shall prepay shipping charges to Agilent Technologies and Agilent Technologies shall pay shipping charges to return the product to Buyer. However, Buyer shall pay all shipping charges, duties, and taxes for products returned to Agilent Technologies from another country.
Agilent Technologieswarrantsthat its software and firmware designated
iv
by Agilent Technologies for use with an instrument will execute its programming instructions when properly installed on that instrument. Agilent Technologies does not warrant that the operation of the instrument, or software, or firmware will be uninterrupted or error-free.

LIMITATION OF WARRANTY

The foregoing warranty shall not apply to defects resulting from improper or inadequate maintenance by Buyer, Buyer-supplied software or interfacing, unauthorized modification or misuse, operation outside of the environmental specifications for the product, or improper site preparation or maintenance.
NO OTHER WARRANTY IS EXPRESSED OR IMPLIED. AGILENT TECHNOLOGIES SPECIFICALLY DISCLAIMS THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE.

EXCLUSIVE REMEDIES

THE REMEDIES PROVIDED HEREIN ARE BUYER’S SOLE AND EXCLUSIVE REMEDIES. Agilent Technologies SHALL NOT BE LIABLE FOR ANY DIRECT, INDIRECT, SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES, WHETHER BASED ON CONTRACT, TORT, OR ANY OTHER LEGAL THEORY.
Where to Find the Latest Information Documentation is updated periodically. For the latest information about
Agilent NFA Noise Figure Analyzers, including firmware upgrades and application information, please visit the following Internet URL:

http://www.agilent.com/find/nf/

v

Manufacturer's Declaration

This statement is provided to comply with the requirements of the German Sound Emission Directive, from 18 January 1991.
This product has a sound pressure emission (at the operator position) < 70 dB(A).
• Sound Pressure Lp < 70 dB(A).
• At Operator Position.
•NormalOperation.
• According to ISO 7779:1988/EN 27779:1991 (Type Test).
Herstellerbeschein igung
Diese Information steht im Zusammenhang mit den Anforderungen der Maschinenlärminformationsverordnung vom 18 Januar 1991.
• Schalldruckpegel Lp < 70 dB(A).
•AmArbeitsplatz.
• Normaler Betrieb.
• Nach ISO 7779:1988/EN 27779:1991 (Typprüfung).
vi
Contents
1. Getting Started
WhatYouwillFindinthisChapter..............................2
OverviewoftheFront-Panel....................................3
OverviewoftheRear-Panel.....................................6
DisplayAnnotation...........................................8
OverviewoftheFrontPanelKeys..............................11
HowtheFrontPanelKeysareOrganized ......................11
NavigatingThroughtheMenuSystem.........................11
PerformingCommonFileOperations............................13
FormattingaDiskette......................................13
SavingaFile..............................................15
LoadingaFile.............................................16
RenamingaFile...........................................16
CopyingaFile.............................................17
DeletingaFile ............................................18
WorkingwithTables.........................................19
2. Making Basic Measurements
WhatYouwillFindinthisChapter.............................22
EnteringENRData..........................................23
Selecting a Common ENR Table . . . . . . . . . .....................23
EnteringENRTableData...................................24
SavinganENRTable.......................................27
EnteringaSpotENRValue..................................28
ChangingtheDefaultTcoldvalue.............................28
SettingtheMeasurementFrequencies ..........................29
SelectingSweepFrequencyMode.............................29
SelectingListFrequencyMode...............................30
vii
Contents
SelectingFixedFrequencyMode............................. 32
SettingtheBandwidthandAveraging.......................... 33
SelectingaBandwidthValue................................33
SettingAveraging.........................................33
CalibratingtheAnalyzer.....................................34
Toperformacalibration....................................34
SelectingtheInputAttenuationRange........................ 35
DisplayingtheMeasurementResults...........................36
SelectingtheDisplayFormat................................ 36
SettingwhichResultTypesareDisplayed.....................38
Graphicalfeatures ........................................ 39
SettingtheScaling........................................41
WorkingwithMarkers.....................................43
IndicatinganInvalidResult..................................50
3. Advanced Features
WhatYouwillFindinthisChapter............................52
SettingupLimitLines.......................................53
CreatingaLimitLine......................................54
SettingLossCompensation...................................55
ConfiguringLossCompensation .............................55
4. Performing System Operations
WhatYouwillFindinthisChapter............................58
SettingtheGPIBAddresses..................................59
ToSettheGPIBAddresses..................................59
ConfiguringtheSerialPort................................... 60
ConfiguringtheLOGPIB .................................... 61
viii
Contents
ConfiguringtheCharacteristicsofanExternalLO ................62
CustomCommandSet......................................62
SettlingTime.............................................64
MinimumandMaximumFrequencies .........................65
ConfiguringtheInternalAlignment.............................66
TurningAlignmentOffandOn...............................66
ChangingAlignmentMode..................................66
DisplayingError,SystemandHardwareInformation..............67
DisplayingtheErrorHistory.................................67
DisplayingSystemInformation...............................67
DisplayingHardwareInformation ............................67
PresettingtheNoiseFigureAnalyzer...........................68
DefiningthePower-On/PresetConditions........................69
SettingthePowerOnConditions .............................69
SettingthePresetConditions................................69
RestoringSystemDefaults....................................70
SettingtheTimeandDate....................................71
Toturnthetimeanddateonandoff...........................71
Tosetthetimeanddate.....................................71
ConfiguringaPrinter ........................................72
ToConfigureaPrinter......................................72
TestingCorrectPrinterOperation ............................72
ix
Contents
x

1 Getting Started

This chapter introduces you to basic features of the Noise Figure Analyzer, including front panel and rear panel descriptions, and an overview of the display annotation.
1
Getting Started

WhatYouwillFindinthisChapter

What You will Find in this Chapter

This chapter covers the following:

• Overview of the Front-Panel

• Overview of the Rear-Panel
• Display Annotation
• Overview of the Front Panel Keys
• Performing Common File Operations
• Working with Tables
2 Chapter1
Overview of the Front-Panel
Figure 1-1 Front panel parts overview
Getting Started
Overview of the Front-Panel
Table 1-1 Front panel item descriptions
Item Description
1 Vie wing Angle keys allow you to adjust the display. 2The 3 Menu keys are the unlabeled keys next to the screen. The menu
Chapter 1 3
Esc (escape) key cancels any entry in progress.
key labels are shown on the display next to these unlabeled keys.
Getting Started
Overview of the Front-Panel
Table 1-1 Front panel item descriptions
Item Description
4TheMEASUREMENT functions allow you to configure the
measurement mode and set the NFA parameters needed for making measurements.
Frequency/Points and Averaging/Bandwidthkeysactivate
The the primary set up function keys and access menus of related functions.
Calibrate key r emoves any second stage noise contribution
The from the measurement. The from this menu, you can enter the ENR data.
Meas Mode and Mode Setup keys are used to configure the
The NFA to measure mixers and frequencies greater than the basic frequency of the NFA using a Local Oscillator.
ENR key accesses the ENR menu,
5The
DISPLAY functions allow you to configure the display
results.
6The
CONTROL functions control the NFA’s setup of Loss
Compensation
and input calibration ranges. The this group, as is full screen display. The
and Limit Line.TheCorr key sets correction
Sweep mode is controlled in
Full Screen functions in
all display formats.
7
SYSTEM functions affect the state of the Noise Figure Analyzer.
Various setup and alignment routines are accessed with the System key.
The green
Preset key resets the Noise Figure Analyzer to a
known state.
File key menu allows you to save and load traces, ENR
The tables, limit-line tables, and frequency lists to or from the NFA memory or the floppy disk drive. The
Save function defined under File.
the
Print Setup menu keys allow you to configure hardcopy
The output. The
Print key sends hardcopy data to the printer.
Save Trace key executes
8 The Media Door on the right side of the front panel accesses the
3.5 inch disk drive.
4 Chapter1
Table 1-1 Front panel item descriptions
Item Description
9 The Data Entry Keys, which include the Up/Down arrow keys,
RPG (rotatable knob), and numeric keys, allow you to enter or change the numeric value of an active function.
The RPG allows continuous change of functions such as, center frequency, averages, and marker position.
The Up/Down arrow keys allow discrete increases or decreases of the active function value.
Getting Started
Overview of the Front-Panel
10
Prev key accesses the previously selected menu.
The 11 Not currently supported. 12 13
PROBE POWER provides power for other accessories.
NOISE SOURCE DRIVE OUTPUT +28V PULSED this
connector provides a 28 Vdc level to switch the noise source on.
The noise source is off when no voltage is applied. 14
Tab Keys are used to move between table input fields, and to
move within the fields of the dialog box accessed by the
menu keys. 15
INPUT 50This is the signal input connector for the Noise
Figure Analyzer. 16
The Next Window key selects which graph or result
parameter is active.
Pressing Zoom key while in graph mode allows you to
switch between the dual-graph and single-graph to display the
active graph. 17 Press the
Help key and then any front panel or menu key to get a
short description of the key function and the associated remote
command.
File
18 The
(On) key turns the Noise Figure Analyzer (NFA) on, while
O (Standby) key switches the NFA to standby.
the
Chapter 1 5
Getting Started
Overview of the Rear-Panel
Overview of the Rear-Panel
Figure 1-2 Rear panel parts overview
Table 1-2 Rear panel item d escriptions
Item Description
1
6 Chapter1
Power input is the input for the AC line-power source.
Table 1-2 Rear panel item d escriptions
Item Description
2 Line Fuse.Thefuseisremovedbytwistingcounterclockwise
1/4 turn. Replace only with a fuse of the same rating. See the
label on the rear panel. 3 Service Connector. The service connector is for service use only.
Getting Started
Overview of the Rear-Panel
4
VGA OUTPUT drives an external VGA compatible monitor with
a signal that has 31.5 kHz horizontal, 60 Hz vertical
synchronizing rate, non-interlaced. 5 6 7
PARALLEL interface parallel port is for printing only.
RS-232 interface supports remote instrument operation.
10 MHz REF IN accepts an external frequency source to provide
the 10 MHz, −15 to +10 dBm frequency reference used by the
Noise Figure Analyzer. 8
10 MHz REF OUT provides a 10 MHz, 0 dBm minimum,
timebase reference signal. 9 10 11 12
LO GBIB port is for the control of an external LO by the NFA.
MAIN GPIB interface port supports remote instrument operation.
AUX OUT (TTL) it is not currently supported.
AUX IN (TTL) it is not currently supported.
13 Power On Selection selects an instrument power preference.
Chapter 1 7
Getting Started

Display Annotation

Display Annotation

The graph display annotation, shown in Figure 1-3, is referenced by numbers, which are listed with a description and a function key indicating which key activates the function related to the annotation.
Figure 1-3 Display Ann otation
32
31
30
29 28
27 26
25 24
23
1
2 3 4 65 7 8
9
10
11
12
22
21
13
14151617181920

Each item is given a description and where applicable a function key associated with it.

8 Chapter1
Table 1-3 Display annotation item descriptions
Item Description
1 The active function area displays the label and value of the
currently active key.
Getting Started
Display Annotation
2 The time and date display, controlled by the
key, under the
System key menus.
3 The marker 1 frequency, controlled by the
State menu keys.
Time/Date menu
Marker(1⇑ ) and
4 The marker 1 amplitude. 5 The marker 2 frequency, controlled by the
State menu keys.
Marker(2⇑ ) and
6 The marker 2 amplitude. 7 The GPIB annunciators RLTS. 8 The data invalid indicator appears when a measurement starts. It
disappears after a complete sweep. 9 The key menu title, this is dependent on which key is selected. 10 The key menu. 11 A non-active menu key. 12 The marker 4 frequency,controlled by the
Marker(4⇓)and State
menu keys. 13 The marker 4 amplitude. 14 The frequency span or stop frequency, controlled by the
Span
or Stop Freq key.
Freq
15 Displays whether the measurement is corrected or uncorrected,
controlled by the calibration state and the
Corr key.
16 Displays whether Loss Compensation is On or Off, controlled
Loss Comp key.
by the 17 The number of points, controlled by the
Points menu key.
Chapter 1 9
Getting Started
Display Annotation
Table 1-3 Display annotation item descriptions
Item Description
18 The marker 3 frequency, controlled by the Marker(3)and State
menu keys. 19 The marker 3 amplitude. 20 The number of averages, controlled by the
Averages menu key.
21 The display status line, displays instrument status and error
messages. 22 The bandwidth, controlled by the
Bandwidth menu key.
This is fixed at 4 MHz on the N8972A model. 23 The center frequency or start frequency, controlled by the
Center Freq or Start Freq menu keys.
24 The lower trace scale, controlled by the
Scale/Div menu key.
(This is auto-coupled to 25 and 26.) 25 The lower trace lower limit, controlled by the
Lower Limit menu
key. (This is auto-coupled to 24 and 26.) 26 The lower trace upper limit, controlled by the
Upper Limit menu
key. (This is auto-coupled to 24 and 25.) 27 The lower trace result type, controlled by the 28 The upper trace scale, controlled by the
Result menu key.
Scale/Div menu key.
(This is auto-coupled to 29 and 30.) 29 The upper trace lower limit, controlled by the
Lower Limit menu
key. (This is auto-coupled to 28 and 30.) 30 The upper trace upper limit, controlledbythe
Upper Limit menu
key. (This is auto-coupled to 28 and 29.) 31 The upper trace result type, controlled by the
Result key.
32 The limit line failure indicator.
10 Chapter1
Getting Started

Overview of the Front Panel Keys

Overview of the Front Panel Keys

How the Front Panel Keys are Organized

The front panel keys are divided into four main groups:
MEASURE keys, w hich are used to configure the measurement
• parameters
CONTROL keys, which are used to configure advanced measurement
• parameters
SYSTEM keys, which perform system-level operations
DISPLAY keys, which adjust the display characteristics of the
• measurement

Navigating Through the Menu System

Menu keys Pressing any of the grey front panel keys in the MEASURE, DISPLAY,
RESULT or SYSTEM key groupings accesses menus of functions that are
displayed along the right-hand side of the display. These keys are called menu keys. See Figure 1-4.
Chapter 1 11
Getting Started
Overview of the Front Panel Keys
Figure 1-4 Menu Keys
Action keys Pressing any of the white keys (
and Print) invokes an action and these keys are called action keys.
To activate a menu key function
To activate a menu key function, press the key immediately to the right of the screen menu key. The menu keys that are displayed depend on whichfrontpanelkeyispressedandwhichmenulevelorpageis selected.
Selecting a function within a menu key
Some menu keys have functions contained within them, for example, and Off. To turn the function on, press the menu key so that On is underlined. To turn the function off, press the menu key so that Off is underlined.
For a summary of all front panel keys and their related menu keys, see theUser’sGuideortheanalyzeronlinehelp.
12 Chapter1
Calibrate, Full Screen, Restart, Save Trace
On
Getting Started

Performing Common File Operations

Performing Common File Operations

This section covers:

• Formatting a diskette

• Saving a file
•Loadingafile
• Renaming a file
• Coping a file
• Deleting a file

FormattingaDiskette

The format is MS-DOS. It is not necessary to format your diskette with the Noise Figure Analyzer; pre-formatted disks can be used with the NoiseFigureAnalyzer.
Step 1. Place the diskette you wish to format into the diskette drive (A:\) of the
NoiseFigureAnalyzer.
Step 2. Access the file manager menu by pressing
Figure 1-5.
Chapter 1 13
File key, File Manager.See
Getting Started
Performing Common File Operations
Figure 1-5 File Manager Menu
Step 3. Start the format process by pressing Step 4. Press
Enter, a second time to format the disk.
The format process takes approximately three minutes. You are now ready to save files to the disk.
14 Chapter1
Format,thenEnter.
Getting Started
Performing Common File Operations

SavingaFile

You can save files (ENR tables, states, traces, limits, frequency lists, or screens) to a floppy disk (A:\), or the internal drive (C:\)oftheNoise Figure Analyzer.
Step 1. To access the Save menu press
File, Save.
Step 2. Selectthetypeoffileyouwanttosave.
For example, if you have a limit line table data present and want to save it, press
Step 3. Select the limit tables file you wish to save (
Forexample,tosavefile2,press
Limits.
1, 2, 3 or 4).
2.
Step 4. Enter a filename using the Alpha Editor menu keys. See“Using the
Alpha Editor” on page HIDDEN. File names are limited to eight (8) characters.
Step 5. Selectthedriveyouwishtosavetobypressing
directory and file list, press
Select.
Tab ,tomoveto
NOTE If the correct drive is not listed in the Path: field, highlight “..” at the
top of the directory list. This enables you to move up a directory. Press
Select. To highlight the desired drive,[-A-] or [-C-])usethearrowkeys
or the RPG, press Select when highlighted.
Step 6. Press
Enter,tosavethefiletothedrive.
Chapter 1 15
Getting Started
Performing Common File Operations

Loading a File

You can load files (ENR tables, states, limits or frequency lists) from a floppy disk (A:\), or the internal drive (C:\).
NOTE Not all the file types you save can be loaded back into the Noise Figure
Analyzer. For example, screen files and trace files. The trace file is in a CSV (comma separated value) format, designed for use with a PC.
Step 1. To access the Load menu press
File, Load.
Step 2. Select the type of file you want to load (ENR tables, states, limits or
frequency lists).
Step 3. Selectthedrivewhereyourfileislocatedbypressing
to highlight [-C-] or [-A-],thenpress
Select.
Tab .UsetheRPG
Step 4. Select the file you want to load into the Noise Figure Analyzer by
changing the highlighted file with the up or down arrow keys to highlight the file name.
Step 5. Press
Enter to load the specified file.

Renaming a File

You can rename a file in the [-C-] or [-A-] drive as follows:
Step 1. Press Step 2. Select the type of file you want to rename (ENR tables, states, traces,
Step 3. Select the drive where you file is located, by pressing the
File, File Manager, Rename toaccesstheRenamemenuitems.
limits, frequency lists or screens). For example, if you are renaming a ENR table file, press
Select.Tochangedrive,usethearrowkeystohighlight[-C-] or [-A-],
then press
Select.
ENR. Tab key, press
Step 4. Select the file you want to rename by moving the cursor with the RPG or
arrow keys to highlight the file name.
Step 5. Press
Tab →to enter the Alpha Editor menu. File names are limited to
eight (8) characters.
16 Chapter1
Loading...
+ 58 hidden pages