SERVICE NOTE
SUPERSEDES: NONE
E8285A CDMA Mobile Station Test Set
Serial Numbers: US00000000/US40272 633
Power Supply Failure May Occur Using 220/240Vac Line Voltage
To Be Performed By: Agilent Technologies-Qualified Personnel
Parts Required: 0950-4059 Power Suppl y
Situation:
Solution / Action:
Replace the defe ct ive power supply with replacement part 0950-4059. Record the part number of the power su ppl y be ing replaced on the CSO. If no part number is found on the power supply, ente r 0950-2665 on the CSO.
DATE: Jun e 2001
ADMINISTRATIVE INFORMATION
SERVICE NOTE CLASSIFICATION:
MODIFICATION RECOMMENDED
ACTION CATEGORY:
LOCATION CATEGORY:
AVAILABILITY: PRODUCT’S SUPPORT LIFE AGILENT RESPONSIBLE UNTIL: June 2004
IMMEDIATELY ON SPECIFIED FAILURE AGREEABLE TIME
CUSTOMER INSTALLABLE ON-SITE SERVICE CENTER
STANDARDS: LABOR 0.5 Hours
SERVICE INVENTORY:
RETURN SCRAP SEE TEXT
USED PARTS:
RETURN SCRAP SEE TEXT
AUTHOR: EK ENTITY: PL-13 ADDITIONAL INFORMATION:
© 2001 AGILENT TECHNOLOGIES PRINTED IN U.S.A.
1 of 1