E8285A-01
SERVICE NOTE
SUPERSEDES: None
E8285A CDMA Mobile Station Test Set
Serial Numbers: US00000000/US40021375
Firmware Upgrade (From A.02.XX And Earlier) Requires Running DIAGLIM7 From TESTS Screen
Duplicate Service Notes:None
To Be Performed By: Agilent Technologies-Qualifie d Pe r sonnel or Custome r
Parts Required: None
Situation:
Test Sets which are being up gr aded with ne w firmwa re (from A.0 2. XX and earlie r to a
newer revision) requir e runn ing DI AGLIM7. DI AGLIM7 is a n int ernal R O M-ba se d IBAS IC
program which rec onfigures specific diag nostic limits for proper hardware operation and
compatibil ty. Fa ilure to run DIAGLI M7 a f te r a fi rm wa re upgrade may result in incorrect diagnostic resu lts, e rror me ssages, or mis- dia gnosis of Test Set opera tional integrity.
ADMINISTRATIVE INFORMATION
SERVICE NOTE CLASSIFICATION:
INFORMATION ONLY
AUTHOR:
JR
ENTITY:
PL13
Continued
DATE: November 2000
ADDITIONAL INFORMATION:
© 2000 AGILENT TECHNOLOGIES
PRINT ED IN U.S.A.
1 of 2

Page 2 Service Note E8285A-01
Solutio n / A ction:
Run DIAGLIM7 after upgradin g Tes t Set firmwar e (from A. 02. XX and ear lier to a newer revision). Use the following procedure:
Step 1:
Check the Test Set’s curre nt f irmwa re re vision on the CONFIG screen and upgrade the firmware if necessary. Firmware c an be pu rchase d by ordering Agilent P/N E8285- 61815.
Step 2:
After the firmwa re has been upgrad ed and the Test Set has powe r ed up suc c essfully; go to
the TESTS screen, select ’ROM’ from the ’Select Procedure Location:’ field. Select ’DIAGLIM7’ from the ’Select Procedure Filename:’ field. Press Run Test (k1). Wait until the
diagnostic li mits have been download ed (the ’ *’ in the uppe r right c or ne r of th e scr e en disa ppears).