SERVICE NOTE
SUPERSEDES: None
E8285A CDMA Mobile Station Test Set
Serial Numbers: US00000000/US40021375
Firmware Upgrade (From A.02.XX And Earlier) Requires Running DI­AGLIM7 From TESTS Screen
Duplicate Service Notes:None
To Be Performed By: Agilent Technologies-Qualifie d Pe r sonnel or Custome r
Situation:
Test Sets which are being up gr aded with ne w firmwa re (from A.0 2. XX and earlie r to a newer revision) requir e runn ing DI AGLIM7. DI AGLIM7 is a n int ernal R O M-ba se d IBAS IC program which rec onfigures specific diag nostic limits for proper hardware operation and compatibil ty. Fa ilure to run DIAGLI M7 a f te r a fi rm wa re upgrade may result in incorrect di­agnostic resu lts, e rror me ssages, or mis- dia gnosis of Test Set opera tional integrity.
ADMINISTRATIVE INFORMATION
SERVICE NOTE CLASSIFICATION:
INFORMATION ONLY
AUTHOR:
JR
ENTITY:
PL13
Continued
DATE: November 2000
ADDITIONAL INFORMATION:
© 2000 AGILENT TECHNOLOGIES PRINT ED IN U.S.A.
1 of 2
Page 2 Service Note E8285A-01
Solutio n / A ction:
Run DIAGLIM7 after upgradin g Tes t Set firmwar e (from A. 02. XX and ear lier to a newer re­vision). Use the following procedure:
Step 1: Check the Test Set’s curre nt f irmwa re re vision on the CONFIG screen and upgrade the firm­ware if necessary. Firmware c an be pu rchase d by ordering Agilent P/N E8285- 61815.
Step 2: After the firmwa re has been upgrad ed and the Test Set has powe r ed up suc c essfully; go to the TESTS screen, select ’ROM’ from the ’Select Procedure Location:’ field. Select ’DI­AGLIM7’ from the ’Select Procedure Filename:’ field. Press Run Test (k1). Wait until the diagnostic li mits have been download ed (the ’ *’ in the uppe r right c or ne r of th e scr e en disa p­pears).