Agilent 8719ES Installation and Quick Start Guide

Installation and Quick Start Guide
HP 8719ET/ES HP 8720ET/ES HP 8722ET/ES
Network Analyzers
HP Part Number: 08720-90391
Printed in USA
Print Date: August 1999
Notice
but not limited to, the implied warranties of merchantability and fitness for a particular purpose. Hewlett-Packardshallnot be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material.
© Copyright 1999 Hewlett-Packard Company
ii
Certification
Hewlett-Packard Company certifies that this product met its published specifications at the time of shipment from the factory. Hewlett-Packard further certifies that its calibration measurements are traceable to the United States National Institute of Standards and Technology, to the extent allowed by the Institute's calibration facility, and to the calibration facilities of other International Standards Organization members.
Regulatory and Warranty Information
The regulatory and warranty information is in the User's Guide.
Assistance
Product maintenance agreements and other customer assistance agreements are available for Hewlett-Packard products. For any assistance, contact your nearest Hewlett-Packard sales or service office. See Table 2-1 on page 2-26 for the nearest office.
Safety Notes
The following safety notes are used throughout this manual. Familiarize yourself with each of the notes and its meaning before operating this instrument.
WARNING Warning denotes a hazard. It calls attention to a procedure which, if
not correctly performed or adhered to, could result in injury or loss of life. Do not proceed beyond a warning note until the indicated conditions are fully understood and met.
CAUTION Caution denotes a hazard. It calls attention to a procedure that, if not
correctly performed or adhered to, would result in damage to or destruction of the instrument. Do not proceed beyond a caution sign until the indicated conditions are fully understood and met.
iii
General Safety Considerations
SOFTKEY
WARNING For continued protection against fire hazard replace line fuse only
with same type and rating (3A/250V). The use of other fuses or material is prohibited.
WARNING This is a Safety Class I product (provided with a protective earthing
ground incorporated in the power cord). The mains plug shall only be inserted in a socket outlet provided with a protective earth contact. Any interruption of the protective conductor, inside or outside the instrument, is likely to make the instrument dangerous. Intentional interruption is prohibited.
CAUTION Ventilation Requirements: When installing the instrument in a cabinet,
the convection into and out of the instrument must not be restricted. The ambient temperature (outside the cabinet) must be less than the maximum operating temperature of the instrument by 4 °C for every 100 watts dissipated in the cabinet. If the total power dissipated in the cabinet is greater than 800 watts, then forced convection must be used.
How to Use This Guide
This guide uses the following conventions:
Front-Panel Key
Screen Text This represents text displayed on the instrument’s screen.
iv
This represents a key physically located on the instrument.
This represents a “softkey,” a key whose label is determined by the instrument’s firmware.

Documentation Map

The Installation and Quick Start Guide provides procedures for installing, configuring, and verifying the operation of the analyzer. It also will help you familiarize yourself with the basic operation of the analyzer.
The User’s Guide shows how to make measurements, explains commonly-used features, and tells you how to get the most performance from your analyzer.
The Reference Guide provides reference information, such as specifications, menu maps, and key definitions.
The Programmer’s Guide provides general HP-IB programming information, a command reference, and example programs. The Programmer’s Guide contains a CD-ROM with example programs.
The CD-ROM provides the Installation and Quick Start Guide, the User’s Guide, the Reference Guide, and the Programmer’s Guide in PDF format for viewing or printing from a PC.
The Service Guide provides information on calibrating, troubleshooting,and servicing your analyzer. The Service Guide is not part of a standard shipment and is available only as Option 0BW, or by ordering HP part number 08720-90397. A CD-ROM with the Service Guide in PDF format is included for viewing or printing from a PC.
v
vi
Contents
1. Installing Your Analyzer
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-2
STEP 1. Verify the Shipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3
STEP 2. Familiarize Yourself with the Analyzer Front and Rear Panels . . . . . . . . . . . . . . .1-5
Analyzer Front Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-5
Analyzer Rear Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-6
STEP 3. Meet Electrical and Environmental Requirements . . . . . . . . . . . . . . . . . . . . . . . . .1-7
STEP 4. Configure the Analyzer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-9
To Configure the Standard Analyzer. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-10
To Configure an Analyzer with a High Stability Frequency Reference (Option 1D5) . . .1-10
To Configure the Analyzer with Printers or Plotters . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-11
To Configure the Analyzer for Bench Top or Rack Mount Use . . . . . . . . . . . . . . . . . . . . .1-16
STEP 5. Verify the Analyzer Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-20
To View the Installed Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-21
To Initiate the Analyzer Self-Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-22
To Run the Operator's Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-23
To Test the Transmission Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-24
To Test the Reflection Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-25
STEP 6. Back Up the EEPROM Disk . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-26
Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-26
Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-26
EEPROM Backup Disk Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-26
2. Quick Start: Learning How to Make Measurements
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-2
Analyzer Front Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-3
Measurement Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-5
Step 1. Choose measurement parameters with your test device connected . . . . . . . . . . . .2-5
Step 2. Make a measurement calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-5
Step 3. Measure the device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-5
Step 4. Output measurement results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-5
Learning to Make Transmission Measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-6
Step 1. Choose the measurement parameters with your test device connected . . . . . . . . .2-6
Step 2. Perform a measurement calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-7
Step 3. Measure the device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-8
Step 4. Output measurement results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-8
Measuring Other Transmission Characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-10
Learning to Make Reflection Measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-13
Step 1. Choose measurement parameters with your test device connected . . . . . . . . . . .2-14
Step 2. Make a measurement calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-15
Step 3. Measure the device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-16
Step 4. Output measurement results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-17
Measuring Other Reflection Characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-18
If You Encounter a Problem . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-23
Power-up Problems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-23
Data Entry Problems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-24
No RF Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-24
vii
Contents
viii

1 Installing Your Analyzer

1-1
Installing Your Analyzer

Introduction

Introduction
This chapter shows you how to install your analyzer and confirm the correct operation, by following the steps below:
1. Verify the shipment.
2. Familiarize yourself with the analyzer front and rear panels.
3. Meet electrical and environmental requirements.
4. Configure the analyzer.
5. Verify the analyzer operation.
6. Back up the EEPROM disk.
1-2 Chapter1
Installing Your Analyzer

STEP 1. Verify the Shipment

STEP 1. Verify the Shipment
1. Unpack the contents of all the shipping containers. WARNING: The analyzer weighs
approximately 54 pounds (25 kilograms). Use correct lifting techniques.
2. Carefully inspect the analyzer to ensure that it was not damaged during shipment.
NOTE If your analyzer was damaged during shipment, contact your nearest
Hewlett-Packardoffice or sales representative. A list of HP sales and service offices is provided in Table 2-1 on page 2-26.
Chapter 1 1-3
Installing Your Analyzer
STEP 1. Verify the Shipment
3. Verify that all the accessories have been included with the analyzer.
Part Number Item
Number
08720-90391 1 Installation and Quick Start Guide 08720-90392 2 User's Guide 08720-90393 3 Reference Guide 08753-90475 4 Programmer’s Guide 08720-90418 5 CD-ROM
unique to
instrument
unique to
country 5062-3978 8 Rack Flange Kit (Option 1CM only) 5062-4072 8 Rack Flange Kit with Handles (Option 1CP only) 5062-3990 8 Front Handle Kit (standard)
6 EEPROM Backup Disk
7 AC power cable
Description
1-4 Chapter1
Installing Your Analyzer

STEP 2. Familiarize Yourself with the Analyzer Front and Rear Panels

STEP 2. Familiarize Yourself with the Analyzer Front and Rear Panels

Analyzer Front Panel

CAUTION Do not mistake the line switch for the disk eject button. See the figure below. If the
line switch is mistakenly pushed, the instrument will be turned off, losing all settings and data that have not been saved.
1 LINE (power on/off) switch 8 RESPONSE function block 2 Display 9 ACTIVE CHANNEL keys 3 Disk drive 10 ENTRY block 4 Disk eject button 11 INSTRUMENT STATE function block 5 Softkeys
6 key
Return
7 STIMULUS function block 14 ES models only:
Chapter 1 1-5
12 key
Preset
13 ES models only: R CHANNEL connectors
ET models only:
PORT 1
and
PORT 2
REFLECTION
and
TRANSMISSION
Installing Your Analyzer
STEP 2. Familiarize Yourself with the Analyzer Front and Rear Panels

Analyzer Rear Panel

1 HP-IB connector 11 AUXILIARY INPUT connector 2 PARALLEL interface 12 EXTERNAL AM connector 3 RS-232 interface 13 EXTERNAL TRIGGER connector 4 KEYBOARD input
(mini-DIN) 14 TEST SEQUENCE 5 Power cord receptacle,
with fuse 15 LIMIT TEST 6 Line voltage selector switch 16 MEASURE RESTART 7 Fan 17 TEST SET I/O INTERCONNECT 8 10 MHZ PRECISION
REFERENCE OUTPUT
a
18 BIAS INPUTS AND FUSES
9 10 MHZ REFERENCE
ADJUST
a
19 Serial number plate
10 EXTERNAL REFERENCE
INPUT connector 20 RF IN/OUT
b
21 EXTERNAL MONITOR: VGA
a. Option 1D5 only. b. Option 085 only.
1-6 Chapter1
Installing Your Analyzer

STEP 3. Meet Electrical and Environmental Requirements

STEP 3. Meet Electrical and Environmental Requirements
1. Set the line-voltage selector to the
position that corresponds to the AC power source.
3. Ensure the operating environment meets the following requirements:
2. Ensure the available AC power source meets the following requirements:
• 90–132 VAC
• 47–66 Hz / 400 Hz (single phase) OR
• 198–265 VAC
• 47–66 Hz (single phase)
The analyzer power consumption is 350 VA maximum.
4. Verify that the power cable is not
damaged, and that the power-source outlet provides a protective earth contact.
• 0 to 55 °C
• < 95% relative humidity at 40 °C (non-condensing)
• < 15,000 feet ( 4,500 meters) altitude
Some analyzer performance parameters are specified for 25 °C ±5 °C. Refer to the Reference Guide for information on the environmental compatibility of warranted performance.
WARNING: Any interruption of the protective (grounding) conductor or disconnection of the protective earth terminal, can result in personal injury, or may damage the analyzer.
Chapter 1 1-7
Installing Your Analyzer
STEP 3. Meet Electrical and Environmental Requirements
5. Ensure there are at least six inches of clearance between the sides and back of either the stand-alone analyzer or the system cabinet.
CAUTION: The environmental temperature must be 4 °C less than the maximum operating temperature of the analyzer for every 100 watts dissipated in the cabinet. If the total power dissipated in the cabinet is >800 watts, then you must provide forced convection.
6. Set up a static-safe workstation. Electrostatic discharge (ESD) can damage or destroy electronic components.
• static-control table mat and earth ground wire: HP part number 9300-0797
• wrist-strap cord: HP part number 9300-0980
• wrist-strap: HP part number 9300-1367
• heel-straps: HP part number 9300-1308
• floor mat: not available through Hewlett-Packard
1-8 Chapter1
STEP 4. Configure the Analyzer
STEP 4. Configure the Analyzer
This step shows you how to set up your particular analyzer configuration.
• standard configuration
• Option 1D5 configuration high stability frequency reference
• printer or plotter configuration
• rack-mount configuration
Installing Your Analyzer
Chapter 1 1-9
Installing Your Analyzer
STEP 4. Configure the Analyzer
To Configure the Standard Analyzer
Connect test port cables and optional adapters if you are using other connector types.
To Configure an Analyzer with a High Stability Frequency Reference (Option 1D5)
Connect the jumper cable on the analyzer rear panel as shown.
1-10 Chapter1
STEP 4. Configure the Analyzer
PARALLEL
COPY
GPIO
COPY
GPIO
To Configure the Analyzer with Printers or Plotters
1. Connect your printer or plotter to the corresponding interface.
Installing Your Analyzer
2. If you are using the parallel interface, press and toggle until your choice of or appears.
Local
If you choose , the parallel port is dedicated for normal copy device use (printers or plotters).
If you choose , the parallel port is dedicated for general purpose I/O. The analyzer controls the data input or output through the sequencing capability of the analyzer.
Chapter 1 1-11
Installing Your Analyzer
SET ADDRESSES
PRINTER PORT
PLOTTER PORT
SET ADDRESSES
PLOTTER PORT
DISK
HPIB
PARALLEL
SERIAL
PLOTTER PORT
PRINTER PORT
HPIB
PARALLEL [COPY]
PARALLEL
STEP 4. Configure the Analyzer
3. Press and then choose either or , depending on your hardcopy device. Or, if you are plotting your files
to disk, press .
4. Press the key that corresponds to your printer or plotter interface: ,
(parallel port), or (serial port). NOTE: The plotter menu is shown as an example. It will only appear if you select . Similar interface choices will appear if you select .
• If you select , the HP-IB address selection is active. Enter the HP-IB address of your printer or plotter,
followed by .
• If you have already selected the
parallel-port configuration, you must also select inthis menu in
order to generate a hardcopy.
x1
choice for the
1-12 Chapter1
Loading...
+ 44 hidden pages