Single-Output dc Supplies
Application info
The HP 6610C Series and
HP 6630B Series of dc power
supplies are ideal for testing
porta bl e battery-powered
products, electronic assemblies, and components.
Ranging current measurement capability verifies
ampere levels down to
off-sta te microamp levels,
assuring quality of devices
being tested.
• High-volume ATE
• High throughput
• Precision measurement
For ordering
information,
see chart on
inside back cover.
5
40 W and 50 W
HP 6611C HP 6612C HP 6613C HP 6614C
Output ratings
Voltage 0 to 8 V 0 to 20 V 0 to 50 V 0 to 100 V
Current 0 to 5 A 0 to 2 A 0 to 1 A 0 to 0.5 A
Programming accuracy at 25° C ± 5° C (percent of setting plus fixed)
Voltage 0.05% + 5 mV 10 mV 20 mV 50 mV
+Current 0.05% + 2 mA 1 mA 0.75 mA 0.5 mA
Ripple and noise in the range of 20 Hz to 20 MHz, with outputs ungrounded or with either terminal grounded
Voltage (rms/p-p)
Normal mode 0.5 mV/3 mV 0.5 mV/3 mV 0.5 mV/4 mV 0.5 mV/5 mV
Current (rms) 2 mA 1 mA 1 mA 1 mA
dc Measurement accuracy via HP-IB or front panel meters with respect to actual output at 25°C ± 5° C
Voltage 0.03% + 2 mV 3 mV 6 mV 12 mV
Low current range
-20 mA to +20 mA 0.1% + 2.5 µA 2.5 µA 2.5 µA 2.5 µA
High current range
+20 mA to +rated 1 0.2% + 0.5 mA 0.25 mA 0.2 mA 0.1 mA
-20 mA to -rated 1 0.2% + 1.1 mA 0.85 mA 0.8 mA 0.7 mA
Average programming resolution (supplemental characteristic)
Voltage 2 mV 5 mV 12.5 mV 25 mV
Current 1.25 mA 0.5 mA 0.25 mA 0.125 mA
OVP 40 mV 100 mV 250 mV 500 mV
Dimensions 212.8 mm W x 368.3 mm D x 88.1 mm H
80 W and 100 W
HP 6631B HP 6632B HP 6633B HP 6634B
Output ratings
Voltage 0 to 8 V 0 to 20 V 0 to 50 V 0 to 100 V
Current 0 to 10 A 0 to 5 A 0 to 2 A 0 to 1 A
Programming accuracy at 25° C ± 5° C (percent of setting plus fixed)
Voltage 0.05% + 5 mV 10 mV 20 mV 50 mV
+Current 0.05% + 4 mA 2 mA 1 mA 0.5 mA
Ripple and noise in the range of 20 Hz to 20 MHz, with outputs ungrounded or with either terminal grounded
Voltage (rms/p-p)
Normal mode 0.3 mV/3 mV 0.3 mV/3 mV 0.5 mV/3 mV 0.5 mV/3 mV
Fast mode 1 mV/10 mV 1 mV/10 mV 1 mV/15 mV 2 mV/25 mV
Current (rms) 2 mA 2 mA 2 mA 2 mA
dc Measurement accuracy via HP-IB or front panel meters with respect to actual output at 25°C ± 5° C
Voltage 0.03% + 2 mV 3 mV 6 mV 12 mV
Low current range
-20 mA to +20 mA 0.1% + 2.5 µA 2.5 µA 2.5 µA 2.5 µA
High current range
+20 mA to +rated 1 0.2% + 1 mA 0.5 mA 0.25 mA 0.25 mA
-20 mA to -rated 1 0.2% + 1.6 mA 1.1 mA 0.85 mA 0.85 mA
Average programming resolution (supplemental characteristic)
Voltage 2 mV 5 mV 12.5 mV 25 mV
Current 2.5 mA 1.25 mA 0.5 mA 0.25 mA
OVP 40 mV 100 mV 250 mV 500 mV
Dimensions 425.5 mm W x 364.4 mm D x 88.1 mm H