Agilent 53150A Product Overview

Full-featured CW Microwave Counters for Field, Factory or Lab
Product Overview
• Agilent 53150A 20 GHz Counter
• Agilent 53151A 26.5 GHz Counter
• Agilent 53152A 46 GHz Counter
High performance microwave counters: at home in the field, bench or ATE environment
• Simultaneous power and frequency measurement with analog peaking indicator
• GPIB and RS-232 standard
• Lightweight and rugged
• Battery optional
2
Convenience, portability and outstanding performance
The innovative designs of the Agilent 53150 Series microwave counters offer an uncluttered, feature­laden front panel. These designs present no-compromise performance and quality in a surprisingly small, light, battery operated product.
The convenience of a single microwave input
The Agilent 53150 Series has an advanced sampler that integrates a separate zero bias Schottky diode for the accurate measurement of input power. This allows measurement of both frequency and power with a single connection. No compromise in frequency coverage is required for this capability. The ultrawideband microwave input covers the entire RF and microwave spectrum, from intermediate frequencies IFs) of 50 MHz to millimeter waves.
The power measurement accuracy and repeatability of these counters rivals power meters with diode sensors. Since frequency and power of the input signal are measured simultaneously, adjustment for the diode’s frequency response is done automatically. And like the latest in diode sensors, compensation is also made for deviation from square law.
Field tough but ready for benchtop or ATE applications
The Agilent 53150 Series is as comfort­able in the field as in the laboratory. The rugged case with an integrated tilting handle can tolerate the vibration and shock expected in field use. The backlit LCD display ensures visibility in all environments, from dark to full sunlight, at distances exceeding 15 feet.
If ac power is unavailable, the internal, replaceable camcorder batteries pro­vide at least 2.5 hours of continuous operation. The unit can also be powered from an external 11-18 Vdc source.
For benchtop and ATE applications, the Agilent 53150 Series delivers full functionality and high measurement speed. The fully programmable RS-232 interface and high speed GPIB interface are standard features.
No compromise performance
The Agilent 53150 Series offers exceptional sensitivity by utilizing a single board design with low phase noise PLL circuitry. Despite their simple appearance, these counters retain all the powerful functions one expects in precision instrumentation: measurement averaging, arbitrary as well as nulling offsets for both frequency and power, display of power in either dBm or Watts.
Additional capabilities include full control of resolution, sampling rate, and GPIB address plus extensive self-diagnostics, fast acquisition times and full programmability. Performance surpasses the industry standard Agilent 5350 Series, in virtually every aspect, in a package less than half the weight and size.
3
Input characteristics Agilent 53150A Agilent 53151A Agilent 53152A
Frequency range
Channel 1 (Normal mode) 10 Hz - 125 MHz 10 Hz - 125 MHz 10 Hz - 125 MHz
(Low pass filter enabled) 10 Hz - 50 kHz 10 Hz - 50 kHz 10 Hz - 50 kHz
Channel 2 50 MHz - 20 GHz 50 MHz - 28.5 GHz 50 MHz - 26.5 GHz
Sensitivity
Channel 1
10-30 Hz 40 mV 40 mV 40 mV 30 Hz-125 MHz 25 mV 25 mV 25 mV
Channel 2
50-250 MHz –20 dBm –20 dBm –20 dBm
0.25-12.4 GHz –33 dBm –33 dBm –33 dBm
12.4-18 GHz –33 dBm –33 dBm –30 dBm 18-20 GHz –29 dBm –29 dBm –27 dBm 20-26.5 GHz N/A –25 dBm –27 dBm
26.5-40 GHz N/A N/A –23 dBm 40-46 GHz N/A N/A –17 dBm
Maximum input
Channel 1 2 Vrms 2 Vrms 2 Vrms Channel 2
50 MHz - 2 GHz +5 dBm +5 dBm +5 dBm 2-46 GHz +13 dBm +13 dBmq +13 dBm
Damage level
Channel 1 120 V (dc + ac pk) linearly 120 V (dc + ac pk) linearly 120 V (dc + ac pk) linearly Channel 2 derated to 5 Vrms at 125 MHz derated to 5 Vrms at 125 MHz derated to 5 Vrms at 125 MHz
+27 dBm +27 dBm +27 dBm
Impedance
(Nominal) Channel 1 1 M/60 pF 1 M/60 pF 1 M/60 pF Channel 2 50 50 50
Connector
Channel 1 BNC female BNC female BNC female Channel 2 SMA/APC-3.5 SMA/APC-3.5 2.92 mm removable,
compatible female compatible female SMA/APC-3.5 compatible female
SWR
Channel 2
50-250 MHz 1.5:1 typical 1.5:1 typical 1.5:1 typical
0.25-10 GHz 2.0:1 typical 2.0:1 typical 2.0:1 typical 10-20 GHz 3.0:1 typical 3.0:1 typical 3.0:1 typical 20-26.5 GHz N/A 3.0:1 typical 2.5:1 typical
26.5-46 GHz N/A N/A 2.5:1 typical
Coupling
Channel 1 ac ac ac Channel 2 ac ac ac
Acquisition time
(1 MHz FM rate) Channel 1 N/A N/A N/A Channel 2 (FM Auto/FM Off) 125 ms/100 ms 125 ms/100 ms 140 ms/115 ms
Resolution
Channel 1/Channel 2 1 Hz to 1 MHz 1 Hz to 1 MHz 1 Hz to 1 MHz
Measurement specifications and characteristics
All measurement specifications are over the full signal and temperature ranges unless otherwise noted.
Loading...
+ 5 hidden pages