Agilent 37717C Remote Control Manual

HP 37717C Communications Performance Analyzer
Remote Control Manual
Copyright Hewlett­Packard Ltd.1998
All rights reserved. Reproduction, adaption, or translation without prior written permission is prohibited, except as allowed under the copyright laws.
HP Part No. 37717-90430
First edition, 11/98
Printed in U.K.
Warranty
The information contained in this document is subject to change without notice.
Hewlett-Packard makes no warranty of any kind with regard to this material, including, but not limited to, the implied warranties or merchantability and fitness for a particular purpose.
Hewlett-Packard shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material.
WARNING
Warning Symbols Used on the Product
!
The product is marked with this symbol when the user should refer to the instruction manual in order to protect the apparatus against damage.
The product is marked with this symbol to indicate that hazardous voltages are present
The product is marked with this symbol to indicate that a laser is fitted. The user should refer to the laser safety information in the Calibration Manual.
Hewlett-Packard Limited Telecommunications Networks Test Division South Queensferry West Lothian, Scotland EH30 9TG
Remote Control Manual
HP Part Number 37717-90430
Printed in U.K. November 1998
HP 37717C Communications Performance Analyzer
In This Book
This manual contains all the information necessary for remote control of the HP 37717C Communications Performance Analyzer.
4
Contents
1 Methods of Remote Control
Methods of Remote Control 1-2 Connecting the HP 37717C to HP-IB 1-3
Operating Distances 1-3 Communication with the System Controller 1-3
Connecting the HP 37717C to RS-232-C 1-4
Terminal Mode 1-4 Computer Mode 1-6 Communication with the Terminal/Computer 1-6
Connecting the HP 37717C to a LAN 1-10
Current Settings 1-11 New Settings (BOOTP) 1-11 New Settings (Keyboard Entry) 1-12
To Initialize the HP 37717C 1-14 Remote Control Hints & Tips 1-15
Default Settings 1-15 Instrument Coupling 1-15 Error Checking 1-15 Command Completion 1-16 Setup Delays 1-16 Status Registers 1-16 Gating Control 1-17 Frequency Offsets 1-18 User Locks 1-18
2 Common Commands
Common Commands 2-3 SCPI Command Format 2-3
Command Syntax 2-3
v
Contents
SCPI Long Form Command Headers 2-4 Linking Command Statements 2-4 Parameters 2-4
Remote Control Commands 2-6 Instrument Options 2-7 INSTrument subsystem 2-8 SOURce subsystem 2-9 SOURce subsystem - Transmitter Common Commands 2-11 INPut subsystem 2-14 SENSe subsystem 2-15 SENSe subsystem - Receiver Common Commands 2-17 SENSe subsystem - Test Timing 2-20 SENSe subsystem - Analysis Control 2-22 SENSe subsystem - Result Returning Commands 2-22
Common Results 2-22
SENSe subsystem - Trouble Scan Results 2-23 Common System Commands 2-26 SENSe subsystem - Configuring Graphics 2-27 SENSe subsystem - Managing Graphics Stores 2-29 SENSe subsystem - Retrieving Graphics Store Data 2-31 SENSe subsystem - Retrieving Data for a Single Graph 2-33 SENSe subsystem -Obtaining Graphics End of Measurement Results 2-35 STATus subsystem 2-44 SYSTem Subsystem 2-47 IEEE common capabilities 2-62
3 PDH Command Reference
PDH Command Reference 3-3 OUTPut subsystem 3-4 SOURce subsystem- Transmitter Unstructured PDH Commands 3-9 SOURce subsystem -Transmitter Structured PDH Commands 3-17
vi
Contents
SOURce subsystem - Transmitter SPDH Test Signal Settings Commands 3­28 SOURce subsystem - Transmitter SPDH Test Function Commands 3-36 INPut subsystem 3-51 SENSe subsystem - Receiver Unstructured PDH Settings 3-57 SENSe subsystem - Receiver Structured PDH Settings 3-62 SENSe subsystem - Receiver SPDH Test Signal Settings Commands 3-68 SENSe subsystem - Receiver SPDH Test Function Commands 3-74 SENSe subsystem - Analysis Control 3-75 SENSe subsystem - Alarm Scan Control 3-80 SENSe subsystem - SPDH Round Trip Delay Control 3-81 SENSe subsystem - Result Returning Commands 3-82
Unstructured PDH Short Term Results 3-82 Unstructured PDH Cumulative Results 3-82 Unstructured PDH and Payload G.821 Analysis Results 3-83 Unstructured PDH M.2100 Analysis Results 3-83 Unstructured PDH FAS Error Analysis Results 3-84 Unstructured PDH CRC Error Analysis Results 3-84 Unstructured PDH REBE Error Analysis Results 3-85 Unstructured PDH Alarm Seconds Results 3-85 Frequency Results (PDH and SPDH) 3-86
3-86 Structured PDH Short Term Results 3-86 Structured PDH Cumulative Results 3-88 Structured PDH and Payload G.821 Analysis Results 3-89 Structured PDH 140 Mb/s G.826 Analysis Results 3-89 Structured PDH 34 Mb/s G.826 Analysis Results 3-90 Structured PDH 8 Mb/s G.826 Analysis Results 3-91 Structured PDH 2 Mb/s G.826 Analysis Results 3-92 Structured PDH DS1 G.826 Analysis Results 3-93 Structured PDH DS3 G.826 Analysis Results 3-94 Structured PDH BIT G.826 Analysis Results 3-95
vii
Contents
Structured PDH M.2100 Analysis Results 3-95 Structured PDH M.2110 Analysis Results 3-97 Structured PDH M.2120 ANALysis Results 3-98 Structured PDH 2 Mb/S FAS Error Analysis Results 3-98 Structured PDH 8 Mb/S FAS Error Analysis Results 3-99 Structured PDH 34 Mb/S FAS Error Analysis Results 3-99 Structured PDH 140 Mb/S FAS Error Analysis Results 3-100 Structured PDH CRC Error Analysis Results 3-100 Structured PDH REBE Error Analysis Results 3-101 Structured PDH Alarm Seconds Results 3-102 Structured PDH Round Trip Delay Results 3-102
FETCh subsystem 3-103
4 SDH Command Reference
SDH Command Reference 4-3 OUTPut subsystem 4-4 SOURce subsystem - Transmitter SDH Settings Commands 4-7
SDH Clock settings 4-7 Thru mode settings 4-9 SDH Mapping Settings 4-11 AU Layer Selection 4-11 TU Layer Selection 4-12 Payload Layer Selection 4-13 TUG Channel 4-14 TU Payload and Test Pattern 4-15 Background Settings 4-18 Old Commands 4-22
SOURce subsystem - Transmitter SDH OVERHEAD SETUP 4-23 SOURce subsystem - Transmitter SDH Test Function Commands 4-36
viii
Contents
INPut subsystem 4-63 SENSe subsystem - Receiver SDH Settings 4-67
SDH Mapping settings 4-67 AU Layer Selection 4-68 TU Layer Selection 4-68 Payload Layer Selection 4-69 TUG Channel 4-70 TU Payload and Test Pattern 4-71 Old Commands 4-75
SENSe subsystem - Receiver SDH Test Function Commands 4-76 SENSe subsystem - Alarm Scan Control 4-83 SENSe subsystem - SDH Tributary Scan Control 4-85 SENSe subsystem - Optical Power Measurement Settings 4-87 SENSe subsystem - Result Returning Commands 4-88
Frequency Results (SDH) 4-88 SDH Short Term Results 4-88 SDH Cumulative Results 4-89 SDH RS B1 BIP Analysis Results 4-90 SDH MS B2 BIP Analysis Results 4-90 SDH Path MS FEBE Analysis Results 4-91 SDH Path B3 BIP Analysis Results 4-92 SDH Path FEBE Analysis Results 4-92 SDH Path IEC Analysis Results 4-93 SDH Tributary Path BIP Analysis Results 4-93 SDH Tributary Path FEBE Analysis Results 4-94 SDH Block Based Bit Analysis Results 4-94 SDH M.2101 Analysis Results 4-95 SDH Service Disruption Results 4-96 SDH Optical Power Result 4-97 SDH Pointer Activity Results 4-97 SDH Alarm Seconds Results 4-98
FETCh subsystem 4-99
ix
Contents
5 SONET Command Reference
SONET Command Reference 5-3 OUTPut subsystem 5-4 SOURce subsystem - Transmitter SONET Settings Commands 5-7
SONET Clock settings 5-7 Thru mode settings 5-9 SONET Mapping Settings 5-11 SPE Layer Selection 5-11 VT Layer Selection 5-12 Payload Layer Selection 5-13 VT Group 5-13 VT Payload and Test Pattern 5-14 Background Settings 5-18
SOURce subsystem - Transmitter SONET OVERHEAD SETUP 5-21 :SOURce subsystem - Transmitter SONET Test Function Commands 5-32 INPut subsystem 5-59 SENSe subsystem - Receiver SONET Settings 5-63
SONET Mapping settings 5-63 SPE Layer Selection 5-64 VT Layer Selection 5-64 Payload Layer Selection 5-65 VT Group 5-66 VT Payload and Test Pattern 5-67
SENSe subsystem - Receiver SONET Test Function Commands 5-71 SENSe subsystem - Alarm Scan Control 5-77 SENSe subsystem - SONET Tributary Scan Control 5-79 SENSe subsystem - Optical Power Measurement Settings 5-81 SENSe subsystem - Result Returning Commands 5-82
Frequency Results (SONET) 5-82 SONET Short Term Results 5-82 SONET Cumulative Results 5-83 SONET Section B1 BIP (CV-S) Analysis Results 5-84
x
Contents
SONET Line B2 BIP (CV-L) Analysis Results 5-84 SONET Line FEBE (REI-L) Analysis Results 5-85 SONET Path B3 BIP (CV-P) Analysis Results 5-85 SONET Path FEBE (REI-P) Analysis Results 5-86 SONET CV-IEC Analysis Results 5-86 SONET Virtual Tributary Path BIP (CV-V) Analysis Results 5-87 SONET Virtual Tributary Path FEBE (REI-V) Analysis Results 5-87 SONET Block Based Bit Analysis Results 5-88 SONET M.2101 Analysis Results 5-88 SONET Service Disruption Results 5-90 SONET Optical Power Result 5-90 SONET Pointer Activity Results 5-90 SONET Alarm Seconds Results 5-91
FETCh subsystem 5-93
6 ATM Command Reference
ATM Command Reference 6-3 OUTPut subsystem 6-4 SOURce subsystem- Transmitter ATM Commands 6-6 INPut subsystem 6-20 SENSe subsystem - Receiver ATM Settings 6-22
ATM Layer 6-22
SENSe subsystem - Result Returning Commands 6-30
ATM Short Term Results 6-30 ATM Cumulative Results 6-31 ATM EM BIP Analysis Results 6-33 ATM FEBE/REI Analysis Results 6-33 ATM REBE Analysis Results 6-34 ATM CRC4 Analysis Results 6-34
xi
Contents
ATM Alarm Seconds Results 6-35 ATM Mean Cell Transfer Delay Results 6-35
FETCh subsystem 6-36
7 Jitter Command Reference
Jitter Command Reference 7-3 OUTPut subsystem 7-4 SOURce subsystem - Transmitter Jitter Commands 7-6 INPut subsystem 7-18 SENSe subsystem - Receiver Jitter Settings 7-21 SENSe subsystem - Result Returning Commands 7-24
Jitter/Wander Results 7-24
FETCh subsystem 7-26
8 Status Reporting
Status Reporting 8-2
Status Byte 8-5 Standard Event Status Register 8-6 QUEStionable Status Register 8-7 OPERation Status Register 8-8 INSTrument Status Register 8-9 DATA Status Register 8-10 PDH Status Register 8-11 FAS Status Register 8-12 BINary Status Register 8-13 SDH Status Register 8-14 SDH2 Status Register 8-15
xii
Contents
SONet Status Register 8-16 SONet2 Status Register 8-17 JITTer Status Register 8-18 SPDH Status Register 8-19 M140 Status Register 8-20 M34 Status Register 8-21 M8 Status Register 8-22 M2 Status Register 8-23 DS3 Status Register 8-24 DS2 Status Register 8-25 DS1 Status Register 8-26 ATM Status Register 8-27 DISK Status Register 8-28 Variations in BINary Input to DATA Status Register 8-31 Programmed Status Reporting 8-32
9 General Information
General Information 9-2 SCPI Overview 9-2 HP-IB Universal Commands 9-3
Device Clear (CLEAR) 9-3 Serial Poll (SPOLL) 9-3 Remote Enable (REMOTE) 9-3 Local Lockout (LOCAL LOCKOUT) 9-4 Local (LOCAL) 9-4
xiii
Contents
10 Application Program Examples
Initializing the HP 37717C 10-2 Setup the HP 37717C PDH Tx (Main Settings) 10-3 Setup the HP 37717C PDH Rx (Main Settings) 10-4 Setup the HP 37717C PDH Tx (Main & Structured Settings) 10-5 Setup the HP 37717C PDH Rx (Main & Structured Settings) 10-7 Setup the HP 37717C PDH Tx to add Errors & Alarms 10-8 Setup the HP 37717C PDH Rx to perform Bit Error Measurements 10-9 Setup the HP 37717C SDH Tx 10-10 Setup the HP 37717C SDH Rx 10-12 Setup the HP 37717C SDH Tx to add Errors & Alarms 10-13 Setup the HP 37717C SDH Tx Overhead Bytes 10-14 Setup the HP 37717C SDH Tx for Overhead Byte Sequencing 10-15 Perform HP 37717C SDH Rx Measurements 10-16 Setup the HP 37717C SDH Rx to retrieve Overhead Monitor Bytes 10-17 Setup the HP 37717C SDH Rx to perform Overhead Byte Capture 10-18 Setup the HP 37717C SONET Tx 10-19 Setup the HP 37717C SONET Rx 10-21 Setup the HP 37717C SONET Tx to add Errors & Alarms 10-22 Setup the HP 37717C SONET Tx Overhead Bytes 10-23 Setup the HP 37717C SONET Tx for Overhead Byte Sequencing 10-24 Perform HP 37717C SONET Rx Measurements 10-25 Setup the HP 37717C SONET Rx to retrieve Overhead Monitor Bytes 10­26 Setup the HP 37717C SONET Rx to perform Overhead Byte Capture 10­27
11 Backwards Compatibility
Backwards Compatibility 11-2
xiv
Contents
12 SCPI Error Messages
SCPI Error Messages 12-2
General Index
xv
Contents
xvi
1

1 Methods of Remote Control

Methods of Remote Control
Methods of Remote Control
The HP 37717C can be remotely controlled in one of four ways :
HP-IB Options A3B, A3D (1A8) Provides a parallel interface that allows the connection of other
devices to the system for example : Frequency Counter; Printer; Signal Generator. Allows great flexibility in communicating data and controlling data and provides one of the easiest methods of constructing automatic systems. If long distance communication is required, suitable HP-IB Extenders must be connected within the test system at both ends of the communication link.
RS-232-C Options A3B, A3D (1CW) Provides a serial interface that can be connected directly to the
RS-232-C port of a terminal or computer. Only the controller and the HP 37717C can be connected within the system. If long distance communication is required, the HP 37717C can be connected directly to the RS-232-C port of a Modem and carried over a telephone line.
LAN Option A3B (1F7) Provides a parallel interface that allows the connection of other
instruments and controllers to the system for example : workstations; other HP 37717C’s; other test sets that have a LAN interface.
DNA Option USS
Provides control via a Front Panel displayed on a PC connected to a LAN. Other test sets that have the Virtual Remote capability can be connected to the LAN and controlled via the same PC. See HP E4540A Manual, HP part number E4540-
90004.
1-2
Methods of Remote Control

Connecting the HP 37717C to HP-IB

The following points should be considered when connecting the HP 37717C to HP-IB :
Operating distances
Communication with the system controller

Operating Distances

1. The total length of HP-IB cable must not exceed 2 meters (6 feet) × the number of devices in the system.
2. The total length of HP-IB cable, used to interconnect all devices in the system, must not exceed 20 meters (65 feet).
Operating distances can be increased by using HP-IB Extenders, HP 37204A or HP 37201A :
Up to 1250 meters use HP 37204A. Over 1250 meters use two HP 37201A and two suitable Modems.

Communication with the System Controller

Each device in the system must have a unique address to allow the controller to access each one individually. On the HP 37717C the address is set on the OTHER display under REMOTE CONTROL :
T o set the HP-IB Address pressOTHER; REMOTE CONTROL and set the display as shown opposite. Address 5 has been selected but any Address in the range 0 to 30 can be selected.
1-3
Methods of Remote Control

Connecting the HP 37717C to RS-232-C

Two methods of controlling the HP 37717C via RS-232-C are available :
Terminal Simple method of control on a command by command basis. Computer More extensive method of control allows programs to be run.

Terminal Mode

Terminal mode is provided to allow the HP 37717C to be controlled by a "dumb" terminal and provides the following features :
When connection is established via RS-232-C the HP 37717C responds with a "prompt".
Any command sent to the HP 37717C is echoed back to the terminal to provide a local record.
If a mistake is made in the command, the appropriate SCPI Error Message and the "entered" command is returned to the terminal to provide a local record.
A command history buffer is provided with a capacity of 20 commands. This allows frequently used commands to be retransmitted without having to retype the command :
To list the history buffer contents type : !l (lower case L) To retransmit the last Command sent type : To retransmit a previously sent command type : !n where n = the commands
listed number.
When REMOTE CONTROL PORT [RS232] is selected on the OTHER REMOTE CONTROL display the HP 37717C acts as DCE (Data Communications Equipment). Dumb terminals act as DTE (Data Terminal Equipment). This makes connection of the HP 37717C to a dumb terminal straightforward as shown in Table 1-1.
RS-232-C cable HP Part Number 5060-4461 will provide this connection.
1-4
Methods of Remote Control
Table 1-1 HP 37717C (Options A3B & A3D) to RS-232-C Terminal
37717C Pin # Signal Terminal Pin # Signal 2 Transmit Data (TX) 2 Transmit Data (TX) 3 Receive Data (RX) 3 Receive Data (RX) 5 Signal Ground 7 Signal Ground
RS-232-C cable HP Part Number 24542U will provide this connection.
Connection via a Modem
Full Modem control is not possible and the Modem should be configured to force DSR, CD, CTS and DTR true.
Since Modems are always configured as DCE (Data Communications Equipment), the same as the HP 37717C, a crossover is required in the three wire connection as shown in Table 1-2.
Table 1-2 HP 37717C (Options A3B & A3D ) to Modem
37717C Pin # Signal Modem Pin # Signal 2 Transmit Data (TX) 3 Receive Data (RX) 3 Receive Data (RX) 2 Transmit Data (TX) 5 Signal Ground 7 Signal Ground
RS-232-C cable HP Part Number 24542M will provide this connection.
1-5
Methods of Remote Control

Computer Mode

Computer mode is provided to allow the HP 37717C to be controlled by a RS-232-C computer allowing programs to be run. Unlike Terminal mode commands are not echoed back and SCPI Error Messages are not returned unless the appropriate query command is issued.
When REMOTE CONTROL PORT [RS232] is selected on the OTHER REMOTE CONTROL display the HP 37717C acts as DCE (Data Communications Equipment). The Computer should be configured to act as DTE (Data Terminal Equipment). This makes connection of the HP 37717C to a computer straightforward as shown in Table 1-3.
Table 1-3 HP 37717C (Option A3B & A3D) to RS-232-C Computer
37717C Pin # Signal Computer Pin # Signal 2 Transmit Data (TX) 2 Transmit Data (TX) 3 Receive Data (RX) 3 Receive Data (RX) 5 Signal Ground 7 Signal Ground
RS-232-C cable HP Part Number 24542U will provide this connection.
Connection via a Modem
Full Modem control is not possible and the Modem should be configured to force DSR, CD, CTS and DTR true.
Since Modems are always configured as DCE (Data Communications Equipment), the same as the HP 37717C, a crossover is required in the three wire connection as shown in Table 1-2 .

Communication with the Terminal/Computer

Remote control via RS-232-C requires that the HP 37717C RS-232-C interface settings match those of the terminal/computer. These settings are made on the OTHER display under REMOTE CONTROL.
1-6
Methods of Remote Control
Using , and the display softkeys set REMOTE CONTR OL PORT [RS232]. Set CONTROLLER TYPE, XON/XOFF, SPEED, PARITY and STOP BITS to match the settings on your terminal/ computer.
.
1-7
Methods of Remote Control
Controlling the HP 37717C
The simplest way to verify the connection and become familiar with remote RS­232-C operation is to send a few commands and observe their effect. In this e xample using a terminal we will :
Take the HP 37717C under remote control
Initialize the HP 37717C
Read the contents of the error register
Start a test period
Add errors
Obtain the result (Option dependent)
Return the HP 37717C to local operation
NOTE Select CONTROLLER TYPE [TERMINAL] on the OTHER REMOTE CONTR OL
display and connect the PDH OUT port to the PDH IN port for the duration of this exercise.
Table 1-4 HP 37717C to RS-232-C Computer
Terminal Input Comment :SYST:REM Takes the HP 37717C under remote control.
On the instrument the indicator above LOCAL is lit.
*RST Default settings, registers cleared. :SYST:ERR? The SCPI Error Message and number can be read,
even in local operation mode. (+0, "No error")
:SENS:DATA:TEL:TEST
Selects a manual test period.
:TYPE MAN :SENS:DATA:TEL:TEST
ON
Starts the manual test period. On the instrument the indicator above is lit.
1-8
RUN/STOP
Methods of Remote Control
Table 1-4 HP 37717C to RS-232-C Computer
Terminal Input Comment
:SOUR:DATA:TEL:ERR
Adds a single bit error.
:BIT ONCE :SOUR:DATA:TEL:ERR
Adds a single bit error.
:BIT ONCE :SOUR:DATA:TEL:ERR
Adds a single bit error.
:BIT ONCE :SENS:DATA:TEL:TEST
OFF :SENS:DATA "ECO:PDH
:BIT"
Halts the manual test period. On the instrument the indicator above is extinguished.
RUN/STOP
Option UKK, Unstructured PDH -Returns the bit error count in numeric form, in this example 3.
or
:SENS:DATA "ECO:SPDH :BIT" RETURN
Option UKJ, Structured PDH - Returns the bit error count in numeric form, in this example 3.
:SYST:LOC Returns the HP 37717C to local control. On the
instrument the indicator above LOCAL is extinguished.
1-9
Methods of Remote Control

Connecting the HP 37717C to a LAN

Two methods of controlling the HP 37717C via a LAN are available :
Terminal Simple method of control on a command by command basis. Computer More extensive method of control allows programs to be run.
The following information should be obtained from the LAN System Administrator before connecting to the LAN:
Internet address (Example 015.144.180.205)
Subnet Mask (Example 255.255.248.0).
Default Gateway IP Address (Example 15.144.176.1).
The PORT NUMBER and MAC ADDRESS are assigned to the HP 37717C at the time of shipment.
Connect the HP 37717C LAN interface to the LAN network. The HP 37717C must now be configured to accept the IP address, Subnet Mask and
Default Gateway IP Address.
OTHER
Press , and set up the display as shown opposite.
The HP 37717C can be configured to accept the IP address, Subnet Mask and Default Gateway IP Address in one of two ways:
1-10
REMOTE CONTROL
Methods of Remote Control

Current Settings

The IP Address, Subnet Mask and Gateway IP Address will adopt the current settings.
The Port Number and MAC Address are displayed for information.

New Settings (BOOTP)

To use BOOTP your system administrator must set up a BOOTP server on the network.
When YES is selected the power must be cycled on the HP 37717C to allow for a change of settings.
When power is restored the HP 37717C will issue BOOTP requests until it receiv es a reply from a suitably configured BOOTP server .
The Settings of IP Address, Subnet Mask and Gateway IP Address will be adopted as the Current Settings.
1-11
Methods of Remote Control

New Settings (Keyboard Entry)

Use and
DECREASE DIGIT
INCREASE DIGIT
Address, Subnet Mask and Gateway IP Address.
Use and to move between the IP Address, Subnet Mask and Gateway IP Address.
The New Settings will not become active until the power is cycled (power off/ power on) on the HP 37717C.
The entered New settings are checked to ensure they conform to the relationships that should exist between the IP Address, Subnet Mask and Gateway IP Address. If the entered data contravenes these relationships, the HP 37717C will change the data to more sensible values.
Configuration
The HP 37717C can only accept one tcp connection at a time and it cannot establish an outgoing call. The connection has to be directed to port 5001.
to set the IP
For Example (from a UNIX system): telnet <IP address> 5001 If ping <IP address> is issued (from a UNIX system), commands issued will be
echoed at the local end of the connection.
Controlling the HP 37717C
The simplest way to verify the connection and become familiar with remote LAN operation is to send a few commands and observe their effect. In this example using a terminal we will:
Take the HP 37717C under remote control
Initialize the HP 37717C
Read the contents of the error register
Start a test period
Add errors
Obtain the result (Option dependent)
Return the HP 37717C to local operation
1-12
Methods of Remote Control
NOTE Connect the PDH OUT port to the PDH IN port for the duration of this exercise.
Table 1-5 HP 37717C to RS-232-C Computer
Terminal Input Comment :SYST:REM Takes the HP 37717C under remote control. On the
instrument the indicator above LOCAL is lit. *RST Default settings, registers cleared. :SYST:ERR? The SCPI Error Message and number can be read,
even in local operation mode. (+0, "No error") :SENS:DATA:TEL:TEST :TYPE MAN Selects a manual test period. :SENS:DAT A:TEL:TEST ON Starts the manual test period. On the instrument the
indicator above is lit.
RUN/STOP
:SOUR:DATA:TEL:ERR :BIT ONCE Adds a single bit error. :SOUR:DATA:TEL:ERR :BIT ONCE Adds a single bit error. :SOUR:DATA:TEL:ERR :BIT ONCE Adds a single bit error. :SENS:DATA:TEL:TEST OFF Halts the manual test period. On the instrument the
indicator above is extinguished.
RUN/STOP
:SENS:DATA "ECO:PDH :BIT" Option UKK, Unstructured PDH -Returns the bit
error count in numeric form, in this example 3.
or
:SENS:DATA? "ECO:SPDH :BIT" Option UKJ, Structured PDH -Returns the bit error
count in numeric form, in this example 3. :SYST:LOC Returns the HP 37717C to local control. On the
instrument the indicator above LOCAL is
extinguished.
1-13
Methods of Remote Control

To Initialize the HP 37717C

Regardless of the current setup the following command will initialize the HP 37717C. It sets the HP 37717C to the factory default settings and clears all registers.
*RST
It is recommended that you do not rely on default settings but program each setting to what you require.
1-14
Loading...
+ 588 hidden pages