Keysight Technologies U2741A User Manual

Page 1
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176 - TestEquipmentDepot.com
Keysight U2741A USB Modular 5.5 Digits Digital Multimeter
User’s Guide
Page 2
Notices
CAUTION
WARNING
Copyright Notice
reproduced in any form or by any means (including electronic storage and retrieval or translation into a foreign language) without prior agreement and written consent from Keysight Technologies as governed by United States and international copyright laws.
Manual Part Number
U2741-90001
Edition
Edition 10, March 17, 2017
Printed in:
Printed in Malaysia
Published by:
Keysight Technologies Bayan Lepas Free Industrial Zone, 11900 Penang, Malaysia
Technology Licenses
The hardware and/or software described in this document are furnished under a license and may be used or copied only in accordance with the terms of such license.
Declaration of Conformity
Declarations of Conformity for this product and for other Keysight products Web.
may be downloaded from the
U.S. Government Rights
The Software is “commercial computer software,” as defined by Federal Acquisition Regulation (“FAR”) 2.101. Pursuant to FAR 12.212 and 27.405-3 and Department of Defense FAR Supplement (“DFARS”) 227.7202, the U.S. government acquires commercial computer software under the same terms by which the software is customarily provided to the public. Accordingly, Keysight provides Software to U.S. government customers under its standard commercial license, which is embodied
its End User License Agreement
in (EULA). The license set forth in the EULA represents the exclusive authority by which the U.S. government may use, modify, distribute, or disclose the Software. The EULA and the license set forth therein, does not require or permit, among other things, that Keysight: (1) Furnish technical
information related to commercial computer software or commercial computer software documentation that is not customarily provided to the public; or (2) Relinquish to, or otherwise provide, the government rights in excess of these rights customarily provided to the public to use, modify, reproduce, release, perform, display, or disclose commercial computer software or commercial computer software documentation. No additional government requirements beyond those set forth in the EULA shall apply, except to the extent that those terms, rights, or licenses are explicitly required from all providers of commercial computer software pursuant to the FAR and the DFARS and are set forth specifically in writing elsewhere in the EULA. Keysight shall be under no obligation to update, revise or otherwise modify the Software. With respect to any technical data as defined by FAR 2.101, pursuant to FAR
12.211 and 27.404.2 and DFARS
227.7102, the U.S. government acquires no greater than Limited Rights as defined in FAR 27.401 or DFAR
227.7103-5 (c), as applicable in any technical data.
the
Warranty
THE MATERIAL CONTAINED IN THIS DOCUMENT IS PROVIDED “AS IS,” AND IS SUBJECT TO BEING CHANGED, WITHOUT NOTICE, IN FUTURE EDITIONS. FURTHER, TO THE MAXIMUM EXTENT PERMITTED BY APPLICABLE LAW, KEYSIGHT DISCLAIMS ALL WARRANTIES, EITHER EXPRESS OR IMPLIED, WITH REGARD TO THIS MANUAL AND ANY INFORMATION CONTAINED HEREIN, INCLUDING BUT NOT LIMITED TO THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE. KEYSIGHT SHALL NOT BE LIABLE FOR ERRORS OR FOR INCIDENTAL OR CONSEQUENTIAL DAMAGES IN CONNECTION WITH THE FURNISHING, USE, OR PERFORMANCE OF THIS DOCUMENT OR OF ANY INFORMATION CONTAINED HEREIN. SHOULD KEYSIGHT AND THE USER HAVE A SEPARATE WRITTEN AGREEMENT WITH WARRANTY TERMS COVERING THE MATERIAL IN THIS DOCUMENT THAT CONFLICT WITH THESE TERMS, THE WARRANTY TERMS IN THE SEPARATE AGREEMENT SHALL CONTROL.
Safety Information
A CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met.
A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice until the indicated conditions are fully understood and met.
2 Keysight U2741A User’s Guide
Page 3

Safety Symbols

CAT II
300 V
The following symbols on the instrument and in the documentation indicate precautions which must be taken to maintain safe operation of the instrument.
Direct current (DC) Off (mains supply)
Alternating current (AC) On (mains supply)
Both direct and alternating current Caution, risk of electric shock
Three-phase alternating current
Earth (ground) terminal Caution, hot surface
Protective earth (ground) terminal Out position of a bi-stable push control
Caution, risk of danger (refer to this manual for specific Warning or Caution information)
Frame or chassis (ground) terminal In position of a bi-stable push control
Equipotentiality
Equipment protected throughout by double insulation or reinforced insulation
Category II 300 V overvoltage protection
Keysight U2741A User’s Guide 3
Page 4

General Safety Information

WARNING
CAUTION
– Do not use the device if it is damaged. Before you use the device, inspect
the casing. Look for cracks or missing plastic. Do not operate the device around explosive gas, vapor, or dust.
– Always use the device with the cables provided.
– Observe all markings on the device before establishing any connection.
– Turn off the device and application system power before connecting to
the I/O terminals.
– When servicing the device, use only the specified replacement parts.
– Do not operate the device with the cover removed or loosened.
– Use only the power adapter provided by the manufacturer to avoid any
unexpected hazards.
– If the device is used in a manner not specified by the manufacturer, the
device protection may be impaired.
– Always use dry cloth to clean the device. Do not use ethyl alcohol or any
other volatile liquid to clean the device.
– Do not permit any blockage of the ventilation holes of the device.
4 Keysight U2741A User’s Guide
Page 5

Environmental Conditions

CAUTION
This instrument is designed for indoor use and in an area with low condensation. The table below shows the general environmental requirements for this instrument.
Environmental condition Requirement
Operating temperature 0 °C to 50 °C
Operating humidity 20% to 85% RH non-condensing
Storage temperature –20 °C to 70 °C
Storage humidity 5% to 90% RH non-condensing
The U2741A USB modular digital multimeter complies with the following safety and EMC requirements.
– IEC 61010-1:2001/EN61010-1:2001 (2nd Edition)
– Canada: CAN/CSA-C22.2 No. 61010-1-04
– USA: ANSI/UL 61010-1:2004
– IEC 61326-2002/EN 61326:1997+A1:1998+A2:2001+A3:2003
– Canada: ICES-001:2004
– Australia/New Zealand: AS/NZS CISPR11:2004
Keysight U2741A User’s Guide 5
Page 6

Regulatory Markings

The CE mark is a registered trademark of the European Community. This CE mark shows that the product complies with all the relevant European Legal Directives.
ICES/NMB-001 indicates that this ISM device complies with the Canadian ICES-001. Cet appareil ISM est conforme a la norme NMB-001 du Canada.
The CSA mark is a registered trademark of the Canadian Standards Association.
The RCM mark is a registered trademark of the Spectrum Management Agency of Australia. This signifies compliance with the Australia EMC Framework regulations under the terms of the Radio Communication Act of 1992.
This instrument complies with the WEEE Directive (2002/96/EC) marking requirement. This affixed product label indicates that you must not discard this electrical or electronic product in domestic household waste.
6 Keysight U2741A User’s Guide
Page 7

Waste Electrical and Electronic Equipment (WEEE) Directive 2002/96/EC

This instrument complies with the WEEE Directive (2002/96/EC) marking requirement. This affixed product label indicates that you must not discard this electrical or electronic product in domestic household waste.

Product category:

With reference to the equipment types in the WEEE directive Annex 1, this instrument is classified as a “Monitoring and Control Instrument” product.
The affixed product label is as shown below.
Do not dispose in domestic household waste.
Keysight U2741A User’s Guide 7
Page 8
THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.
8 Keysight U2741A User’s Guide
Page 9

Table of Contents

Safety Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
General Safety Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4
Environmental Conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
Regulatory Markings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6
Waste Electrical and Electronic Equipment (WEEE) Directive
2002/96/EC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7
Product category: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7
Sales and Technical Support . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7
1 Getting Started
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16
Product at a Glance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .17
Product outlook . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .17
Product Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .19
Dimensions without bumpers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .19
Dimensions with bumpers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .20
Standard Shipped Items . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .21
Inspection and Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .22
Initial inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .22
Electrical check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .22
General maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .23
Installation and Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .24
Instrument Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .25
55-Pin backplane connector pin configuration . . . . . . . . . . . . . . . . .25
Chassis installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26
2Operation and Features
Powering Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .28
Making Measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .29
Measuring DC voltage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .29
Keysight U2741A User’s Guide 9
Page 10
Measuring AC voltage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Measuring DC current . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Measuring AC current . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Measuring resistance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Measuring frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Testing continuity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Testing diodes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Measuring temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Restoring Instrument State . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Autozero . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Ranging . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Default Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
Triggering the U2741A . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
System-Related Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
Error conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
3Measurement Tutorial
DC Measurement Considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
Noise Rejection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
Resistance Measurement Considerations . . . . . . . . . . . . . . . . . . . . . . . 52
AC Measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
Other Primary Measurement Functions . . . . . . . . . . . . . . . . . . . . . . . . . 57
Frequency measurement errors . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
DC current measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
Other Sources of Measurement Error . . . . . . . . . . . . . . . . . . . . . . . . . . 59
4Characteristics and Specifications
10 Keysight U2741A User’s Guide
Page 11

List of Figures

Figure 1-1 55-pin backplane connector pin configuration . . . . . .25
Figure 3-1 Common mode source error . . . . . . . . . . . . . . . . . . . .49
Figure 3-2 Ground loop induced error . . . . . . . . . . . . . . . . . . . . . .51
Figure 3-3 Burden voltage in current measurement . . . . . . . . . . .58
Keysight U2741A User’s Guide 11
Page 12
THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.
12 Keysight U2741A User’s Guide
Page 13

List of Tables

Table 1-1 SSI connector pin description . . . . . . . . . . . . . . . . . . .25
Table 2-1 Default settings summary . . . . . . . . . . . . . . . . . . . . . .42
Table 3-1 Thermoelectric voltage for dissimilar metals
connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .48
Table 3-2 Power dissipation for various resistance ranges . . . . .54
Table 3-3 Waveform shapes and their parameters . . . . . . . . . . .55
Keysight U2741A User’s Guide 13
Page 14
THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.
14 Keysight U2741A User’s Guide
Page 15
Keysight U2741A USB Modular 5.5 Digits Digital Multimeter User’s Guide

1 Getting Started

Introduction 16 Product at a Glance 17 Product Dimensions 19 Standard Shipped Items 21 Inspection and Maintenance 22 Installation and Configuration 24 Instrument Configuration 25
15
Page 16
1 Getting Started

Introduction

The Keysight U2741A is a 5.5 digits digital multimeter that can operate as standalone or as a modular unit when used with the U2781A USB modular instrument chassis.
U2741A can perform the following measurements:
–DC Voltage
–AC Voltage
– DC Current
– AC Current
– Resistance
–Diode Test
– Continuity Test
–Temperature
– Frequency
The U2741A is controlled remotely over a USB interface via the Keysight Measurement Manager (KMM) software. The U2741A can also be programmed using the provided drivers or via SCPI commands in your customized application.
16 Keysight U2741A User’s Guide
Page 17

Product at a Glance

Bumpers

Product outlook

Top view
Getting Started 1
Keysight U2741A User’s Guide 17
Page 18
1 Getting Started
USB Indicator
Power Indicator
Fuse holder
4-wire Ω Lo Sense
4-wire Ω Hi Sense
VOLT HI
VOLT LO
55-pin backplane connector
USB inlet
Power inlet
Fastening hole for USB cable with locking mechanism
Front view
18 Keysight U2741A User’s Guide
Rear view
Page 19

Product Dimensions

105.00 mm
175.00 mm
11.50 mm
25.00 mm

Dimensions without bumpers

Top view
Getting Started 1
Front view
Keysight U2741A User’s Guide 19
Page 20
1 Getting Started
180.00 mm
8.60 mm
117.00 mm
41.00 mm

Dimensions with bumpers

Top view
Front view
20 Keysight U2741A User’s Guide
Page 21

Standard Shipped Items

Verify that you have received the following items with your unit. If anything is missing or damaged, please contact the nearest Keysight Sales Office.
✔ 12 V, 2 A AC/DC adapter
✔ Power cord
✔ Standard Test Lead
✔ USB Standard-A to Mini-B interface cable
✔ L-Mount kit (used with the modular instrument chassis)
✔ Keysight Automation-Ready CD-ROM (contains the Keysight IO Libraries
Suite)
✔ Keysight USB Modular Products and Systems Quick Start Guide
✔ Keysight USB Modular Products and Systems Product Reference DVD-ROM
✔ Keysight Measurement Manager Quick Reference Card
✔ Certificate of Calibration
Getting Started 1
Keysight U2741A User’s Guide 21
Page 22
1 Getting Started

Inspection and Maintenance

Initial inspection

When you receive your U2741A, inspect the unit for any obvious damage such as broken terminals or cracks, dents, and scratches on the casing that may occur during shipment.
If any damage is found, notify the nearest Keysight Sales Office immediately. The front of this manual contains the warranty information.
Keep the original packaging in case the U2741A has to be returned to Keysight in the future. If you return the U2741A for service, attach a tag identifying the owner and model number. Also include a brief description of the problem.

Electrical check

For verification procedures, refer to Keysight U2741A USB Modular 5.5 Digits Digital Multimeter Service Guide. The procedures will verify to a high level of confidence that the U2741A is operating in accordance to its specifications.
22 Keysight U2741A User’s Guide
Page 23

General maintenance

NOTE
General maintenance
Any repair that is not covered in your modular product manuals should only be performed by qualified personnel.
1 Power off your module and remove the power cord and I/O cable from your
device.
2 Remove your module from the bumper casing.
3 Shake out any dirt that may have accumulated on the module.
4 Wipe your module with a dry cloth and install the bumper back in place.
Getting Started 1
Keysight U2741A User’s Guide 23
Page 24
1 Getting Started
NOTE

Installation and Configuration

Follow the step-by-step instructions shown in the Keysight USB Modular Products and Systems Quick Start Guide to get started with the preparations and installations of your U2741A.
You need to install the IVI-COM driver if you are going to use the U2741A with Keysight VEE Pro, LabVIEW, or Microsoft
®
Visual Studio®.
24 Keysight U2741A User’s Guide
Page 25

Instrument Configuration

GND
NC
GND
TRIG4
nBPUB
NC
11
GND
NC
TRIG3
GND
CLK10M
NC
10
GND
NC
GND
TRIG5
GND
NC
9
GND
NC
STAR_TRIG
NC
8
TRIG2
GND
GND
NC
ND
TRIG6
GA2
NC
7
GND
NC
TRIG1
GND
GA1
NC
6
GND
NC
GND
TRIG7
GA0
NC
5
GND
NC
TRIG0
GND
NC
NC
4
GND
VBUS
GND
+12 V
+12 V
+12 V
3
GND
GND
GND
+12 V
+12 V
+12 V
2
GND
USB_D–
USB_D+
GND
+12 V
+12 V
1
F
E
D
C
B
A

55-Pin backplane connector pin configuration

The 55-pin backplane connector is used when the U2741A module is inserted into the U2781A USB modular instrument chassis. For more details, refer to the Keysight U2781A USB Modular Instrument Chassis User's Guide.
Getting Started 1
Figure 1-1 55-pin backplane connector pin configuration
Table 1-1 SSI connector pin description
SSI timing signal Functionality
GND Ground
NC Not connected
VBUS USB bus power sensing input
USB_D+, USB_D– USB differential pair
TRIG0~TRIG7 Trigger bus
+12 V +12 V power with 4 A current
nBPUB USB backplane input detect
Keysight U2741A User’s Guide 25
Page 26
1 Getting Started
Tab le 1-1 SSI connector pin description (continued)
SSI timing signal Functional ity
CLK10M 10 MHz clock source
STAR_TRIG Star trigger
GA0,GA1,GA2 Geographical address pin

Chassis installation

The L-Mount kit is to be installed to your U2741A module.
The following instructions describe the simple procedure of installing the L-Mount kit and your module in the chassis.
1 Unpack the L-Mount kit from the packaging.
2 Remove your U2741A module from the bumper casing.
3 Use a Phillips screwdriver to fasten the L-Mount kit to your U2741A module.
4 Insert your U2741A module into the U2781A chassis with the 55-pin
backplane connector positioned at the bottom of the module.
5 Once you have slotted in the module, tighten the screws of the L-Mount kit to
secure the connection.
26 Keysight U2741A User’s Guide
Page 27
Keysight U2741A USB Modular 5.5 Digits Digital Multimeter User’s Guide

2 Operation and Features

Powering Up 28 Making Measurements 29 Restoring Instrument State 40 Default Settings 42 Triggering the U2741A 43 System-Related Operation 45
This chapter contains the details on how to configure your U2741A USB modular digital multimeter to perform the various measurement functions either through the software front panel or by sending SCPI commands remotely via the USB interface.
27
Page 28
2 Operation and Features

Powering Up

Take note of the following when you power up the U2741A.
– The U2741A can only be operated via the USB interface.
– Before you can control the U2741A, you need to install the hardware driver
and the IO Libraries Suite 14.2 or higher. Both of these are included when you purchase the U2741A. Refer to the Keysight USB Modular Products and Systems Quick Start Guide for the installation procedure.
– On the front panel of the U2741A, there are two LED indicators. Refer to
Chapter 1, “Product outlook” on page 17.
– Power indicator lights up once the U2741A is powered up. It will blink if there
is a system error.
– USB indicator will only blink when there is data exchange activity between the
U2741A and the PC.
You can control your U2741A via the KMM for U2741A or via SCPI commands sent through the USB interface from your own application programs.
The Keysight U2741A complies with the syntax rules and conventions of SCPI commands.
You can determine the U2741A's SCPI language version by sending the SYSTem:VERSion? command from the remote interface.
For a complete discussion of U2741A's SCPI syntax, refer to the Keysight U2741A Programmer's Reference.
28 Keysight U2741A User’s Guide
Page 29

Making Measurements

DC Voltage Source
+–
The following pages show how to make measurement connections and how to select measurement functions from the front panel for each of the measurement functions.

Measuring DC voltage

The DC voltage measurement function has the following features:
– Five ranges to select: 100 mV, 1 V, 10 V, 100 V, and 300 V, or auto-range.
– Input impedance is 10 MΩ for all ranges (typical).
– Input protection is 300 V on all ranges (HI terminal).
Make the connection as shown below.
Operation and Features 2
Keysight U2741A User’s Guide 29
Page 30
2 Operation and Features
Keysight Measurement Manager operation
Select the DCV function and desired range. A suitable range should be selected to give the best measurement resolution. The reading is displayed and updated continuously.
SCPI commands
The following example shows how to make a DC voltage measurement using SCPI commands.
MEASure[:VOLTage]:DC?
30 Keysight U2741A User’s Guide
Page 31

Measuring AC voltage

AC Voltage Source
+–
The AC voltage measurement function has the following features:
– Five ranges to select: 100 mVrms, 1 Vrms, 10 Vrms, 100 Vrms, and 250 Vrms,
or auto-range.
– Measures the AC-coupled true RMS.
– Measures within the stipulated accuracy at a crest factor of a maximum of 5:1
(full scale).
– Input impedance is 1 MΩ ± 2% in parallel with less than 120 pF capacitance on
all ranges.
Make the connection as shown below.
Operation and Features 2
Keysight Measurement Manager operation
Select the ACV function and desired range. A suitable range should be selected to give the best measurement resolution. The reading is displayed and updated continuously.
SCPI commands
The following example shows how to make an AC voltage measurement using SCPI commands.
MEASure[:VOLTage]:AC?
Keysight U2741A User’s Guide 31
Page 32
2 Operation and Features
DC Current Source
+
–

Measuring DC current

The DC current measurement function has the following features:
– Three ranges to select: 10 mA, 100 mA, 1 A and 2 A, or auto-range.
– Input protection fuse is 2 A, voltage rating 250 V on all ranges.
Make the connection as shown below.
Keysight Measurement Manager operation
Select the DCI function and desired range. A suitable range should be selected to give the best measurement resolution. The reading is displayed and updated continuously.
SCPI commands
The following example shows how to make a DC current measurement using SCPI commands.
MEASure:CURRent[:DC]?
32 Keysight U2741A User’s Guide
Page 33

Measuring AC current

AC Current Source
+
–
The AC current measurement function has the following features:
– Three ranges to select: 10 mA, 100 mA, 1 A and 2 A or, auto-range.
– Measures true rms value.
Make the connection as shown below.
Operation and Features 2
Keysight Measurement Manager operation
Select the ACI function and desired range. A suitable range should be selected to give the best measurement resolution. The reading is displayed and updated continuously.
SCPI commands
The following example shows how to make an AC current measurement using SCPI commands.
MEASure:CURRent:AC?
Keysight U2741A User’s Guide 33
Page 34
2 Operation and Features
Resistance
Test
Current

Measuring resistance

The resistance measurement function has the following features:
– Seven ranges to select: 100 Ω, 1 kΩ, 10 kΩ, 100 kΩ, 1 MΩ, 10 MΩ, and
100 MΩ or, auto-range.
– Supports two-wire and four-wire resistance measurement.
– Open circuit voltage is limited to less than 4.5 V on all ranges.
The figure below shows a two-wire resistance measurement connection.
34 Keysight U2741A User’s Guide
Page 35
Operation and Features 2
Test Current
The following figure shows a four-wire resistance measurement connection.
Keysight Measurement Manager operation
Select the 2-Wired Ω function and desired range for two-wire resistance measurement. Select the 4-Wired Ω function and desired range for four-wire resistance measurement. A suitable range should be selected to give the best measurement resolution. The reading is displayed and updated continuously.
SCPI commands
The following example shows how to make a resistance measurement using SCPI commands.
Two-wire: MEASure:RESistance?
Four-wire: MEASure:FRESistance?
Keysight U2741A User’s Guide 35
Page 36
2 Operation and Features
Frequency Source
~

Measuring frequency

The frequency measurement function has the following features:
– Range is based on the amplitude of the signal.
– Employs a reciprocal counting technique as the measurement method.
– Gate time can be set at 0.1 second or 1 second of the input signal.
Make the connection as shown below.
Keysight Measurement Manager operation
Select the Freq function and desired range. The reading is displayed and updated continuously.
SCPI commands
The following example shows how to make a frequency measurement using SCPI commands.
MEASure:FREQuency?
36 Keysight U2741A User’s Guide
Page 37

Testing continuity

Open or closed circuit
Current
Test
The continuity testing function has the following features:
– Uses a 1 mA ± 0.2% constant current source.
– Open circuit voltage is limited to less than 4.5 V for all ranges.
– Continuity threshold is fixed at 10 Ω.
– Response time is 60 samples/second.
Make the connection as shown below.
Operation and Features 2
Keysight Measurement Manager operation
)
Select the Cont-
)
) function. The reading is displayed and updated continuously.
SCPI commands
The following example shows how to make a Continuity Test measurement using SCPI commands.
MEASure:CONTinuity?
Keysight U2741A User’s Guide 37
Page 38
2 Operation and Features
Forward bias

Testing diodes

The diode testing function has the following features:
– Uses a 1 mA ± 0.2% constant current source.
– Open circuit voltage is limited to less than 4.5 V for all ranges.
– Response time is 60 samples/second.
Make the connection as shown below.
Keysight Measurement Manager operation
Select the Diode function. The reading is displayed and updated continuously.
SCPI commands
The following example shows how to make a Diode Test measurement using SCPI commands.
MEASure:DIODe?
38 Keysight U2741A User’s Guide
Page 39

Measuring temperature

Thermistor
The temperature measurement function has the following features:
– The measurement range is dependant on the type of temperature sensor used.
For the specification details of the temperature sensors, please refer to
Chapter 4, “Characteristics and Specifications”.
– Measurements are in auto-range for 5 kΩ thermistor probe.
– Supports the thermistor.
Make the connection as shown below.
Operation and Features 2
Keysight Measurement Manager operation
Select the Temp function and the type of thermistor used. The reading is displayed and updated continuously.
SCPI commands
The following example shows how to make a temperature measurement using SCPI commands.
MEASure:TEMPerature? THER //Used for thermistor measurement
Keysight U2741A User’s Guide 39
Page 40
2 Operation and Features

Restoring Instrument State

The U2741A automatically saves the last configuration whenever a power-down event occurs and restores to the last power-down state when you turn on the instrument.

Autozero

When autozero is enabled, the digital multimeter internally disconnects the input signal following each measurement, and takes a zero reading. It then subtracts the zero reading from the preceding reading. This prevents offset voltages present on the digital multimeter’s input circuitry from affecting measurement accuracy.
This is applicable to DC voltage, DC current, two-wire ohms and temperature measurements only.
SCPI Commands
The following commands are used to set the autozero:
VOLTage:ZERO:AUTO {OFF|ON}
CURRent:ZERO:AUTO {OFF|ON}
RESistance:ZERO:AUTO {OFF|ON}
TEMPerature:ZERO:AUTO {OFF|ON}
40 Keysight U2741A User’s Guide
Page 41

Ranging

Operation and Features 2
You can let the digital multimeter automatically select the range using auto range or you can select a fixed range using manual range. Autoranging is convenient because the digital multimeter automatically selects the appropriate range for each measurement. However, you can use manual range for faster measurements since the digital multimeter does not have to determine which range to use for each measurement.
– Autorange thresholds:
Down range at: <10% of range
Up range at: >120% of range
– For manual range, if the input signal is greater than the present range can
measure, the digital multimeter gives an overload indication: “9.9E+37” from the remote interface. For auto range, the digital multimeter will give an overload indication “9.9E+37” if the input signal is greater than the highest measurement range.
– The range is fixed for continuity tests (1 kΩ range) and diode tests (1 VDC
range with 1 mA current source output).
SCPI Commands
You can set the range using any of the following commands.
CONFigure:<function> {<range>|MIN|MAX|AUTO},
{<resolution>|MIN|MAX|DEF}
MEASure:<function>? {<range>|MIN|MAX|AUTO},
{<resolution>|MIN|MAX|DEF}
<function>:RANGe {<range>|MINimum|MAXimum|AUTO}
<function>:RANGe:AUTO {OFF|ON}
Keysight U2741A User’s Guide 41
Page 42
2 Operation and Features

Default Settings

The table below summarizes the U2741A's settings as received from the factory, at power-on or following the *RST command received over the USB remote interface.
Tab le 2-1 Default settings summary
Parameter Factory Setting Power-on/Reset State
Measurement Configuration
Function DCV DCV
Range AUTO AUTO
Resolution
Temperature Unit °C User setting
Trigger Operation
Trigger Source Auto Trigger Auto Trigger
System-Related Operation
Power-Down Recall Disabled User Setting
Stored States 0-10 Cleared No Change
Reading Output Buffer Cleared Cleared
Error Queue Cleared Cleared
Power-on Status Clear Enabled User Setting
Status Registers, Masks & Transition Filter Cleared Cleared in power-on status clear
Calibration
Calibration State Secured User Setting
Calibration Value 0 No Change
Calibration String Cleared No Change
5.5 digits 5.5 digits
enabled
42 Keysight U2741A User’s Guide
Page 43

Triggering the U2741A

At power-up, the default trigger source is immediate. To make a measurement, the following steps are to be followed:
– Configure the U2741A by selecting the function, range, resolution, and so on.
– Specify the digital multimeter's trigger source. You can select software bus
trigger or an immediate internal trigger (default trigger source).
– Ensure that the U2741A is in wait-for-trigger state to accept a trigger from the
specified source.
Immediate triggering
In the immediate trigger mode, the trigger signal is always present. When you place the U2741A in the wait-for-trigger state, the trigger is issued immediately. This is the default trigger source.
SCPI command
The following command sets the trigger source to immediate:
TRIGger:SOURce IMMediate
Operation and Features 2
Software bus triggering
The bus trigger mode is initiated by sending a bus trigger command, after selecting BUS as the trigger source.
SCPI command
The following command sets the trigger source to bus:
TRIGger:SOURce BUS
The MEASure? and READ? command overwrites the BUS trigger and triggers the digital multimeter and returns a measurement.
The INITiate command changes the state of the triggering to wait-for-trigger state. Measurements will begin when the specified trigger conditions are satisfied.
Keysight U2741A User’s Guide 43
Page 44
2 Operation and Features
Star Triggering
The star trigger can only be applied when the U2741A is connected into the U2781A modular instrument chassis. It is used to trigger multiple modular units in the chassis.
SCPI command
TRIGger:SOURce STRG
Synchronization Status
This is to configure the synchronization of multiple units of U2741A (slave only) when used in the U2781A modular instrument chassis. Only one master can be assigned at a time.
SCPI command
CONFigure:SSI {NONE|SLAVe}
Refer to the Keysight U2741A Programmer's Reference for complete description and syntax for these commands.
44 Keysight U2741A User’s Guide
Page 45

System-Related Operation

This section provides information on system-related topics such as performing a self-calibration routine and reading error conditions.

Error conditions

A record of up to 20 errors can be stored in the U2741A's error queue. Refer to the programming guide for more information on the error messages.
Keysight Measurement Manager Operation
A message box will appear once an error occurs while operating the U2741A using the KMM.
Operation and Features 2
Keysight U2741A User’s Guide 45
Page 46
2 Operation and Features
THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.
46 Keysight U2741A User’s Guide
Page 47
Keysight U2741A USB Modular 5.5 Digits Digital Multimeter User’s Guide

3 Measurement Tutorial

DC Measurement Considerations 48 Noise Rejection 49 Resistance Measurement Considerations 52 AC Measurements 55 Other Primary Measurement Functions 57 Other Sources of Measurement Error 59
The Keysight U2741A is capable of making accurate measurements but in order to achieve the best accuracy, you must take the necessary steps to eliminate potential measurement errors. This chapter describes common errors found in measurements and gives suggestions to help you avoid or minimize these errors.
47
Page 48
3Measurement Tutorial

DC Measurement Considerations

Thermal EMF errors
Thermoelectric voltages are the most common source of error in low-level DC voltage measurements. Thermoelectric voltages are generated when you make circuit connections using dissimilar metals at different temperatures.
Each metal-to-metal junction forms a thermocouple, which generates a voltage proportional to the junction temperature. You should take the necessary precautions to minimize thermocouple voltages and temperature variations in low-level voltage measurements. The best connections are formed using copper-to-copper crimped connections, as the digital multimeter's input terminals are copper alloy. The table below shows common thermoelectric voltages for connections between dissimilar metals.
Tab le 3-1 Thermoelectric voltage for dissimilar metals connections
Copper to - Approx. mV / °C Copper to - Approx. mV / °C
Cadmium-Tin Solder 0.2 Aluminum 5
Copper <0.3 Tin-Lead Solder 5
Gold 0.5 Kovar or Alloy 42 40
Silver 0.5 Silicon 500
Brass 3 Copper-Oxide 1000
Beryllium 5
48 Keysight U2741A User’s Guide
Page 49

Noise Rejection

Vf = Float Voltage R
s
= DUT Source Resistance Imbalance
R
i
= digital mul timeter Isolation Resistance
C
i
= digital mul timeter Input Capacitance
Error (v) =
R
s
+ R
i
V
f
R
i
Rs
C
i
LO
HI
Ideal
Meter
V
test
Vf x R
s
(LO-Earth)
Rejecting power-line noise voltages
An important property of integrating analog-to-digital (A/D) converters is their ability to reject power-line related noise that is present on the DC input signals. This is called normal mode noise rejection, or NMR. The digital multimeter achieves NMR by measuring the average DC input by “integrating” it over an integral number of power line cycles.
Common mode rejection (CMR)
Ideally, a digital multimeter is completely isolated from earth-referenced circuits. However, there is a finite resistance between the digital multimeter's input LO terminal and earth ground, as shown below. This can cause errors when measuring low voltages that are floating in relative to earth ground.
Measurement Tutorial 3
Figure 3-1 Common mode source error
Keysight U2741A User’s Guide 49
Page 50
3Measurement Tutorial
Noise caused by magnetic loops
If you are making measurements near magnetic fields, it is a good practice to avoid inducing voltages in the measurement connections. Conductor carrying high current is one common source of magnetic field. You can use twisted-pair cable connections to the digital multimeter to reduce the noise pickup loop area, or place the test leads as close together as possible. Loose or vibrating test leads will also induce error voltages. Tie test leads securely when operating near magnetic fields. Whenever possible, use magnetic shielding materials or increase the distance from magnetic sources to minimize this error.
Noise caused by ground loops
When measuring voltages in circuits where the digital multimeter and the device under test are both referenced to a common earth ground, a ground loop is formed. As shown in Figure 3-2, any voltage difference between the two ground reference points (V
leads. This gives rise to noise and offset voltage (usually power-line related), which are added to the measured voltage.
) will cause a current to flow through the measurement
ground
50 Keysight U2741A User’s Guide
Page 51
Measurement Tutorial 3
RL = Lead Resistance R
i
= digital multimeter Isolation
V
ground
= Voltage Drop or Ground Bus
R
L
Ri > 10 GΩ
V
ground
R
L
V
test
LO
HI
Meter
Ideal
Resistance
Figure 3-2 Ground loop induced error
The best way to eliminate ground loops is to isolate the digital multimeter from earth by not grounding the input terminals. If the digital multimeter must be earth-referenced, connect it and the device under test to the same common ground point. Also connect the digital multimeter and device under test to the same electrical outlet whenever possible.
Keysight U2741A User’s Guide 51
Page 52
3Measurement Tutorial

Resistance Measurement Considerations

When measuring resistance, the test current flows from the input HI terminal through the resistor being measured. The voltage drop across the resistor being measured is sensed internally in the digital multimeter. Therefore, the test lead resistance is also measured.
The errors mentioned earlier in this chapter for DC voltage measurements also apply to resistance measurements. Additional error sources unique to resistance measurements are discussed here.
Resistance measurements
The Keysight U2741A offers two methods for measuring resistance:
2-wire and 4-wire ohms. For both methods, the test current flows from the input HI terminal and then through the resistor being measured. For 2-wire ohms, the voltage drop across the resistor being measured is sensed internal to the digital multimeter. Therefore, test lead resistance is also measured. For 4-wire ohms, separate “sense” connections are required. Since no current flows in the sense leads, the resistance in these leads does not give a measurement error.
The errors mentioned earlier in this chapter for DC voltage measurements also apply to resistance measurements. Additional error sources unique to resistance measurements are discussed on the following pages.
4-wire ohms measurements
The 4-wire ohms method provides the most accurate way to measure small resistances. Test lead resistances and contact resistances are automatically reduced using this method. Four-wire ohms is often used in automated test applications where long cable lengths, numerous connections, or switches exist between the digital multimeter and the device-under-test. The recommended connections for 4-wire ohms measurements are shown below.
52 Keysight U2741A User’s Guide
Page 53
Measurement Tutorial 3
Removing test lead resistance errors
To eliminate offset errors associated with the test lead resistance in 2-wire ohms measurements, follow the steps below.
1 Short the tips of the test leads. The reading is the test lead resistance.
2 Click Null. The digital multimeter stores the test lead resistance as the 2-wire
resistance null value and the digital multimeter will subtract this value from the value of subsequent measurements.
Minimizing power dissipation effects
When measuring resistors designed for temperature measurements (or other resistive devices with large temperature coefficients), be aware that the digital multimeter will dissipate some power in the device under test.
If power dissipation is a problem, you should select the digital multimeter's next higher measurement range to reduce the errors to acceptable levels. The following table shows several examples:
Keysight U2741A User’s Guide 53
Page 54
3Measurement Tutorial
Tab le 3-2 Power dissipation for various resistance ranges
Range Test current DUT power at full scale
100 Ω 1 mA 100 mW
1 kΩ 0.83 mA 689 mW
10 kΩ 100 mA 100 mW
100 kΩ 10 mA 10 mW
1 MΩ 900 nA 810 nW
10 MΩ 205 nA 420 nW
100 MΩ 205 nA || 10 MΩ 35 nW
Errors in high resistance measurements
When you are measuring large resistances, significant errors can occur due to insulation resistance and surface cleanliness. You should take the necessary precautions to maintain a “clean” high-resistance system. Test leads and fixtures are susceptible to leakage due to moisture absorption in insulating materials and
“dirty” surface films. Nylon and PVC are relatively poor insulators (10
13
compared to Polytetrafluoroethylene (PTFE) insulators (10 nylon or PVC insulators can easily contribute a 0.1% error when measuring a
Ω resistance in humid conditions.
1 M
Ω). Leakage from
9
Ω) when
54 Keysight U2741A User’s Guide
Page 55
Measurement Tutorial 3
NOTE
2
V
2
-----
V
2
----­3
V
3
-----
V
3
-----
T
t
p
-------
V
CF
-----
1
1
CF
-----
⎝⎠
⎛⎞
2
–×
V
CF
-----
1VV

AC Measurements

True RMS AC measurements
True RMS responding digital multimeter, like the U2741A, measure the “heating” potential of an applied voltage. Power dissipated in a resistor is proportional to the square of the voltage waveform. This digital multimeter accurately measures true RMS voltage or current, as long as the wave shape contains negligible energy above the instrument's effective bandwidth.
The U2741A uses the same techniques to measure true RMS voltage and true RMS current.
Table 3-3 Waveform shapes and their parameters
Waveform shape Crest factor (CF) AC RMS AC+DC RMS
Keysight U2741A User’s Guide 55
Page 56
3Measurement Tutorial
AC DC+()
TrueRMS
AC2DC
2
+=
The digital multimeter's AC voltage and AC current functions measure the AC-coupled true RMS value, where only the “heating value” of only the AC components of the input waveform is measured while the DC component is rejected. As seen in the figure above; for sine waves, triangular waves, and square waves, the AC-coupled and AC+DC values are equal, since these waveforms do not contain a DC offset. However, for non-symmetrical waveforms, such as pulse trains, there is a DC voltage content, which is rejected by Keysight's AC-coupled true RMS measurements. This can provide a significant benefit. An AC-coupled true RMS measurement is desirable when you are measuring small AC signals in the presence of large DC offsets.
A good example is measuring AC ripples that are present in DC power supplies. There are situations, however, where you might want to know the AC+DC true RMS value. You can determine this value by combining results from DC and AC measurements, as shown below (Equation 1):
For the best AC noise rejection, you should perform the DC measurement at 20 NPLC.
(1)
56 Keysight U2741A User’s Guide
True RMS accuracy and high-frequency signal content
A common misconception is that “since an AC digital multimeter is true RMS, its sine wave accuracy specifications apply to all waveforms.” Actually, the shape of the input signal can dramatically affect the measurement accuracy especially when that input signal contains high-frequency components which exceed the instrument's bandwidth.
Page 57

Other Primary Measurement Functions

Frequency measurement errors

The U2741A uses a reciprocal counting technique to measure frequency. This method generates a constant measurement resolution for any input frequency. All frequency counters are susceptible to errors when measuring low-voltage, low-frequency signals. The effects of both internal noise and external noise pickup are critical when measuring “slow” signals. The error is inversely proportional to frequency. Measurement errors also occur if you attempt to measure the frequency of an input following a DC offset voltage change and therefore, you must allow the digital multimeter’s input to fully settle before making frequency measurements.

DC current measurements

When you connect the digital multimeter in series with a test circuit to measure current, a measurement error is introduced. The error is caused by the digital multimeter's series burden voltage. A voltage is developed across the wiring resistance and current shunt resistance of the digital multimeter, as shown below.
Measurement Tutorial 3
Keysight U2741A User’s Guide 57
Page 58
3Measurement Tutorial
V
b
R
s
LO
HI
R
Ideal
M
e
t
e
r
V
s
Error(%) =
–100%
× V
b
V
s
Vs = Source Voltage R
s
= DUT Source Resistance
V
b
= digital mul timeter Burden
R = digital multimeter Current Shunt
Voltage
Figure 3-3 Burden voltage in current measurement
58 Keysight U2741A User’s Guide
Page 59

Other Sources of Measurement Error

Error %()
100– R
s
×
R
s
1 MΩ+
--------------------=
Error %() 100
1
12π FR
s
Cin×××()+
----------------------------------------------1–×=
Rs = Source resistance F = Input frequency C
in
= Input capacitance (100 pF) plus cable capacitance
Loading errors (AC volts)
In the AC voltage function, the input of the digital multimeter appears as a 1 MΩ resistance in parallel with 100 pF of capacitance. The cabling that you use to connect signals to the digital multimeter also adds capacitance and loading.
For low frequencies, the loading error is (Equation 2):
At high frequencies, the additional loading error is (Equation 3):
Measurement Tutorial 3
(2)
(3)
Measurements below full scale
You can make the most accurate AC measurements when the digital multimeter is at or near the full scale of the selected range. Autoranging occurs at 10% (down-range) and 120% (up-range) of full scale. This enables you to measure some inputs at full scale on one range and 10% of full scale on the next higher range. In general, the accuracy is better on the lower range; for the highest accuracy, select the lowest manual range possible for the measurement.
High-voltage self-heating errors
If you apply more than 300 Vrms, self-heating occurs in the digital multimeter's internal signal-conditioning components. These errors are included in the digital multimeter's specifications.
Temperature changes inside the digital multimeter due to self-heating may cause additional error on other AC voltage ranges.
Keysight U2741A User’s Guide 59
Page 60
3Measurement Tutorial
Voltage Measured V
in
2
Noise
2
+=
AC current measurement errors (burden voltage)
Burden voltage errors, which apply to DC current, also apply to AC current measurements. However, the burden voltage for AC current is larger due to the digital multimeter's series inductance and your measurement connections. The burden voltage increases as the input frequency increases. Some circuits may oscillate when performing current measurements due to the digital multimeter's series inductance and your measurement connections.
Low-level measurement errors
When measuring AC voltages less than 100 mV, be aware that these measurements are especially susceptible to errors introduced by extraneous noise sources. An exposed test lead acts as an antenna and a properly functioning digital multimeter will measure the signals received. The entire measurement path, including the power line, acts as a loop antenna. Circulating currents in the loop create error voltages across any impedance in series with the digital multimeter's input. For this reason, you should apply low-level AC voltages to the digital multimeter through shielded cables. You should connect the shield to the input LO terminal.
Make sure the digital multimeter and the AC source are connected to the same electrical outlet whenever possible. You should also minimize the area of any ground loops that cannot be avoided. A high-impedance source is more susceptible to noise pickup than a low-impedance source. You can reduce the high-frequency impedance of a source by placing a capacitor in parallel with the digital multimeter's input terminals. You may have to experiment to determine the correct capacitor value for your application.
Most extraneous noise is not correlated with the input signal. You can determine the error as shown below (Equation 4):
60 Keysight U2741A User’s Guide
Correlated noise, while rare, is especially detrimental. Correlated noise always adds directly to the input signal. Measuring a low-level signal with the same frequency as the local power line is a common situation that is prone to this error.
(4)
Page 61
Keysight U2741A USB Modular 5.5 Digits Digital Multimeter User’s Guide
4 Characteristics and
Specifications
For the characteristics and specifications of the U2741A USB Modular 5.5 Digits Digital Multimeter, refer to the datasheet at
http://literature.cdn.keysight.com/litweb/pdf/5991-0042EN.pdf.
61
Page 62
4 Characteristics and Specifications
THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.
62 Keysight U2741A User’s Guide
Page 63
This information is subject to change without notice. Always refer to the Keysight website for the latest revision.
© Keysight Technologies 2008 - 2017 Edition 10, March 17, 2017
Printed in Malaysia
*U2741-90001*
Loading...