Keysight N9911X Configuration Guide

Keysight N9911X Economical Waveguide Calibration Components
Configuration Guide
Notices
CAUTION
WARNING
© Keysight Technologies 2013, 2014
No part of this manual may be reproduced in any form or by any means (including electronic storage and retrieval or transla­tion into a foreign language) without prior agreement and written consent from Key­sight Technologies as governed by United States and international copyright laws.
Manual Part Number
N9911-90003
Edition
August 2014
Keysight Technologies 1400 Fountaingrove Parkway Santa Rosa, CA 95403
Warranty
The material contained in this docu­ment is provided “as is,” and is sub­ject to being changed, without notice, in future editions. Further, to the max­imum extent permitted by applicable law, Keysight disclaims all warran­ties, either express or implied, with regard to this manual and any infor­mation contained herein, including but not limited to the implied warran­ties of merchantability and fitness for a particular purpose. Keysight shall not be liable for errors or for inciden­tal or consequential damages in con­nection with the furnishing, use, or performance of this document or of any information contained herein. Should Keysight and the user have a separate written agreement with warranty terms covering the material in this document that conflict with these terms, the warranty terms in the separate agreement shall control.
Technology Licenses
The hardware and/or software described in this document are furnished under a license and may be used or copied only in accordance with the terms of such license.
Restricted Rights Legend
52.227-19 (June 1987) or any equivalent agency regulation or contract clause. Use, duplication or disclosure of Software is subject to Keysight Technologies’ standard commercial license terms, and non-DOD Departments and Agencies of the U.S. Gov­ernment will receive no greater than Restricted Rights as defined in FAR
52.227-19(c)(1-2) (June 1987). U.S. Govern­ment users will receive no greater than Limited Rights as defined in FAR 52.227-14 (June 1987) or DFAR 252.227-7015 (b)(2) (November 1995), as applicable in any technical data.
Safety Notices
A CAUTION notice denotes a haz­ard. It calls attention to an operat­ing procedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met.
A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly per­formed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice until the indicated condi­tions are fully understood and met.
If software is for use in the performance of a U.S. Government prime contract or sub­contract, Software is delivered and licensed as “Commercial computer soft­ware” as defined in DFAR 252.227-7014 (June 1995), or as a “commercial item” as defined in FAR 2.101(a) or as “Restricted computer software” as defined in FAR

N9911X Economical Waveguide Calibration Components

N9911X Economical Waveguide Calibration Components are used to calibrate Keysight FieldFox Handheld Analyzers. These components are purchased separately — it is not necessary to buy a kit. This allows you to purchase only the components you need, saving the expense of non-necessary components and a storage box.
Use the following information to help determine which N9911X Economical Waveguide Calibration Components to purchase.

Types of Calibration Components

Here is a brief explanation of the types of calibration components that you can purchase. Refer to
“Calibration Component Part Numbers and Photos” on page 3 to see images of the components.
Adapter, Waveguide-to-Coax
The adapter is intended for adapting the analyzer’s coaxial connectors to waveguide flanges.
Te r mi n a ti o n
A termination is also called a load. It is connected directly to the adapter.
¼ Wavelength Offset Shim
A ¼ wavelength offset shim is also called a ¼ wavelength section. The shim is connected between the adapter and the short to create an offset short.
Short
A short is also called a flush short. It is connected directly to the adapter, or used as an offset short when combined with the ¼ wavelength offset shim.

Coax Cables

To perform a 1-port calibration, one coaxial cable is required. To perform a 2-port calibration, two coaxial cables are required. Each cable is connected between an analyzer test port and a waveguide-to-coax adapter. A list of Keysight cables available for purchase is in the “Accessories” section of the online document Keysight FieldFox Handheld Analyzers Configuration Guide (part number 5990-9836EN).

Coupling Set (Hardware) for Flanges

Hardware items such as screws, nuts, washers, and slip pins are necessary for coupling flanges together. One coupling set (english/imperial or metric) contains all of the flange hardware you need for performing both 1-port and 2-port calibrations. It is not necessary to buy more than one coupling set.
Configuration Guide N9911-90003 1
Loading...
+ 10 hidden pages