a.To ensure that these newly installed options are functioning properly, the procedure that follows includes the
requirement of performing certain adjustments and performance verification tests. However, the completion of
these tests does not guarantee that the instrument meets all advertised specifications.
1.5 Hours
1.0 Hours
4.0 Hours
a
a
Software and test equipment is required for making adjustments and for performance verification
testing.
Information on how to obtain this software can be found at:
www.keysight.com/find/calibrationsoftware
While Keysight does recommend that a full calibration be performed after the installation of this
upgrade, the end user must ultimately determine whether they want this service or not. If a full
calibration is required, arrangements regarding the level of calibration must be made between the
end user and the calibration provider.
Introduction
This kit includes parts to upgrade a N9021B MXA signal analyzer to add Option EXM, External
Mixing. This kit includes parts to upgrade a MXA regardless of frequency range option installed.
In addition to installing the hardware to support Option EXM, a license for the following option will
also be installed:
— N9021B-EXM, External Mixing
Installation Note N9021-900043
Option EXM, External Mixing Upgrade Kit
Installation Kit Parts List
QuantityDescriptionKeysight Part Number
1Adapter- Coaxial Straight Female-SMA to Female-SMA, 50W1250-1666
1Cable Assembly, External Mixing, Front Panel (W29)N9020-20166
1Cable Assembly, mmW Front End to W29 (W17) (for use in millimeter wave
analyzers)
1Cable Assembly, Coaxial 275 mm LG (W30) (for use in millimeter wave
analyzers)
1Label, WarningN9030-80018
14Screw, M3 x 0.5, (6 mm long), flat head 0515-1946
5Cable Ties1400-0249
1Entitlement Certificate5964-5178
1Entitlement Certificate Envelope5967-7169
1Installation NoteThis note
a.This cable is included in the Opt EXM Cable Kit with Wire Markers, p/n N9021-60006.
N9020-20167
8121-2025
labeled ‘903’ and ‘13’
a
with ends
4Installation Note N9021-90004
Option EXM, External Mixing Upgrade Kit
Tools Required
— T-10 TORX Driver
— T-20 TORX Driver
— 5/16-inch torque wrench
— ¼-inch open-end wrench
—Diagonal cutters
— Keysight Calibration and Adjustment Software, N7814A (revision E.25.00 or later)
— Test equipment and computer supported by the X- Series Performance Tests and Adjustment
Software
— N9021B MXA Signal Analyzer Service Guide, N9021-90002.
— Microsoft Windows based personal computer with internet access and USB port
— USB storage device with > 4 GB free memory
Initial Instrument Functionality Check
Power on the instrument and allow the instrument to boot up. Run an alignment and display the
measurement screen. (The instrument will probably display a spectrum analyzer screen and you will
see the instrument sweeping.)
There should be no alignment failures. If there are failures, investigate and fix the problem before
continuing.
Before you disassemble the instrument, turn the power switch to Standby. After the instrument
has completely shut down, unplug the instrument. Failure to unplug the instrument can result in
personal injury.
Electrostatic discharge (ESD) can damage or destroy electronic components. All work on
electronic assemblies should be performed at a static-safe workstation. Refer to the
documentation that pertains to your instrument for information about static-safe workstations
and ordering static-safe accessories.
Installation Note N9021-900045
Option EXM, External Mixing Upgrade Kit
Analyzer Disassembly
If the instrument is placed on its face during any of the following procedures, be sure to use a soft
surface or soft cloth to avoid damage to the front panel, keys, or input connector.
Make sure any adapters on the front panel are removed.
1. Disconnect the instrument from ac power.
2. Refer to Figure 1. Using the T-20 driver, remove the 4 screws (1) (two on each side) that
attach the handle strap (2) on each side of the instrument.
3. Remove the four key locks from the instrument four bottom feet.
4. Remove the four instrument bottom feet (6).
5. Using the T-20 driver, remove the four screws (including washers) (3) that hold the rear feet
(4) in place.
6. Pull the instrument cover (5) off towards the rear of the instrument.
Figure 1 Instrument Outer Case Removal
6Installation Note N9021-90004
Option EXM, External Mixing Upgrade Kit
Top Brace Removal
Refer to Figure 2. To remove the top brace (1), use the T-10 driver to remove the twelve screws (3)
(0515-0372) attaching the top brace to the chassis. Remove the wire hold down 1 (as shown in
Figure 3). Remove and discard the 14 screws (15 screws with Option B5X) (4) (0515-1946)
attaching the top brace to the boards.
Figure 2 Top Brace Removal
Installation Note N9021-900047