Keysight N9021B User Manual

Keysight Technologies N9021B MXA Signal Analyzer
Option EXM, External Mixing Upgrade Kit
Installation Note
Notices
© Copyright 2021 Keysight Technologies, Inc.
Keysight Technologies makes no warranty of any kind with regard to this material, including but not limited to, the implied warranties of merchantability and fitness for a particular purpose. Keysight Technologies shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material.
Manual Part Number
N9021-90004
Edition
Edition 1, January 2021
Printed in USA/Malaysia
Published by: Keysight Technologies, Inc. 1400 Fountaingrove Parkway Santa Rosa, CA 95403

Option EXM, External Mixing Upgrade Kit

Option EXM, External Mixing Upgrade Kit
Products Affected: N9021B MXA
Serial Numbers: All
To Be Performed By: (X) Keysight Service Center
(X) Personnel Qualified by Keysight
( ) Customer
Estimated Installation Time: Estimated Adjustment Time: Estimated Verification Time:
a.To ensure that these newly installed options are functioning properly, the procedure that follows includes the requirement of performing certain adjustments and performance verification tests. However, the completion of these tests does not guarantee that the instrument meets all advertised specifications.
1.5 Hours
1.0 Hours
4.0 Hours
a
a
Software and test equipment is required for making adjustments and for performance verification testing.
Information on how to obtain this software can be found at:
www.keysight.com/find/calibrationsoftware
While Keysight does recommend that a full calibration be performed after the installation of this upgrade, the end user must ultimately determine whether they want this service or not. If a full calibration is required, arrangements regarding the level of calibration must be made between the end user and the calibration provider.

Introduction

This kit includes parts to upgrade a N9021B MXA signal analyzer to add Option EXM, External Mixing. This kit includes parts to upgrade a MXA regardless of frequency range option installed.
In addition to installing the hardware to support Option EXM, a license for the following option will also be installed:
— N9021B-EXM, External Mixing
Installation Note N9021-90004 3
Option EXM, External Mixing Upgrade Kit

Installation Kit Parts List

Quantity Description Keysight Part Number
1 Adapter- Coaxial Straight Female-SMA to Female-SMA, 50W 1250-1666
1 Washer, Lock, Internal Tooth, ¼ inch 2190-0067
1 Nut-Hex-Double-Chamfer 1/4-36-THD.125-IN-THK, Stainless Steel 2950-0223
1 50 Ohm Termination, SMA male 1810-0118
1 Opt EXM Cable Kit with Wire Markers (includes cable W30, listed below) N9021-60006
1 Cable Assembly, External Mixing, Front Panel (W29) N9020-20166
1 Cable Assembly, mmW Front End to W29 (W17) (for use in millimeter wave
analyzers)
1 Cable Assembly, Coaxial 275 mm LG (W30) (for use in millimeter wave
analyzers)
1 Label, Warning N9030-80018
14 Screw, M3 x 0.5, (6 mm long), flat head 0515-1946
5 Cable Ties 1400-0249
1 Entitlement Certificate 5964-5178
1 Entitlement Certificate Envelope 5967-7169
1 Installation Note This note
a.This cable is included in the Opt EXM Cable Kit with Wire Markers, p/n N9021-60006.
N9020-20167
8121-2025 labeled ‘903’ and ‘13’
a
with ends
4 Installation Note N9021-90004
Option EXM, External Mixing Upgrade Kit

Tools Required

— T-10 TORX Driver
— T-20 TORX Driver
— 5/16-inch torque wrench
— ¼-inch open-end wrench
—Diagonal cutters
— Keysight Calibration and Adjustment Software, N7814A (revision E.25.00 or later)
— Test equipment and computer supported by the X- Series Performance Tests and Adjustment
Software
— N9021B MXA Signal Analyzer Service Guide, N9021-90002.
— Microsoft Windows based personal computer with internet access and USB port
— USB storage device with > 4 GB free memory

Initial Instrument Functionality Check

Power on the instrument and allow the instrument to boot up. Run an alignment and display the measurement screen. (The instrument will probably display a spectrum analyzer screen and you will see the instrument sweeping.)
There should be no alignment failures. If there are failures, investigate and fix the problem before continuing.
Before you disassemble the instrument, turn the power switch to Standby. After the instrument has completely shut down, unplug the instrument. Failure to unplug the instrument can result in personal injury.
Electrostatic discharge (ESD) can damage or destroy electronic components. All work on electronic assemblies should be performed at a static-safe workstation. Refer to the documentation that pertains to your instrument for information about static-safe workstations and ordering static-safe accessories.
Installation Note N9021-90004 5
Option EXM, External Mixing Upgrade Kit

Analyzer Disassembly

If the instrument is placed on its face during any of the following procedures, be sure to use a soft surface or soft cloth to avoid damage to the front panel, keys, or input connector.
Make sure any adapters on the front panel are removed.
1. Disconnect the instrument from ac power.
2. Refer to Figure 1. Using the T-20 driver, remove the 4 screws (1) (two on each side) that
attach the handle strap (2) on each side of the instrument.
3. Remove the four key locks from the instrument four bottom feet.
4. Remove the four instrument bottom feet (6).
5. Using the T-20 driver, remove the four screws (including washers) (3) that hold the rear feet
(4) in place.
6. Pull the instrument cover (5) off towards the rear of the instrument.
Figure 1 Instrument Outer Case Removal
6 Installation Note N9021-90004
Option EXM, External Mixing Upgrade Kit

Top Brace Removal

Refer to Figure 2. To remove the top brace (1), use the T-10 driver to remove the twelve screws (3) (0515-0372) attaching the top brace to the chassis. Remove the wire hold down 1 (as shown in
Figure 3). Remove and discard the 14 screws (15 screws with Option B5X) (4) (0515-1946)
attaching the top brace to the boards.
Figure 2 Top Brace Removal
Installation Note N9021-90004 7