Keysight N5445A InfiniiMax III, N2848A InfiniiMax III+, N5439A InfiniiMax III, N2836A InfiniiMax III, N5441A InfiniiMax III, N5444A InfiniiMax III, N2835A InfiniiMax III, N2787A, N2803A User Manual
No part of this manual may be reproduced in
any form or by any means (including electronic storage and retrieval or translation
into a foreign language) without prior agreement and written consent from Keysight
Technologies, Inc. as governed by United
States and international copyright laws.
Manual Part Number
N2800-97020
Edition
May 2017
Published in USA
Published by:
Keysight Technologies, Inc.
1400 Fountaingrove Parkway
Santa Rosa, CA 95403 USA
Warranty
The material contained in this document is provided “as is,” and is subject
to being changed, without notice, in
future editions. Further, to the maximum extent permitted by applicable
law, Keysight disclaims all warranties,
either express or implied, with regard
to this manual and any information
contained herein, including but not
limited to the implied warranties of
merchantability and fitness for a particular purpose. Keysight shall not be
liable for errors or for incidental or
consequential damages in connection
with the furnishing, use, or performance of this document or of any information contained herein. Should
Keysight and the user have a separate
written agreement with warranty terms
covering the material in this document
that conflict with these terms, the warranty terms in the separate agreement
shall control.
U.S. Government Rights
The Software is "commercial computer software,"
as defined by Federal Acquisition Regulation
("FAR") 2.101. Pursuant to FAR 12.212 and
27.405-3 and Department of Defense FAR Supplement ("DFARS") 227.7202, the U.S. government acquires commercial computer software
under the same terms by which the software is
customarily provided to the public. Accordingly,
Keysight provides the Software to U.S. government customers under its standard commercial
license, which is embodied in its End User License
Agreement (EULA), a copy of which can be found
at
http://www.keysight.com/find/sweula. The
license set forth in the EULA represents the exclusive authority by which the U.S. government may
use, modify, distribute, or disclose the Software.
The EULA and the license set forth therein, does
not require or permit, among other things, that
Keysight: (1) Furnish technical information related
to commercial computer software or commercial
computer software documentation that is not customarily provided to the public; or (2) Relinquish
to, or otherwise provide, the government rights in
excess of these rights customarily provided to the
public to use, modify, reproduce, release, perform, display, or disclose commercial computer
software or commercial computer software documentation. No additional government requirements beyond those set forth in the EULA shall
apply, except to the extent that those terms,
rights, or licenses are explicitly required from all
providers of commercial computer software pursuant to the FAR and the DFARS and are set forth
specifically in writing elsewhere in the EULA. Keysight shall be under no obligation to update,
revise or otherwise modify the Software. With
respect to any technical data as defined by FAR
2.101, pursuant to FAR 12.211 and 27.404.2 and
DFARS 227.7102, the U.S. government acquires
no greater than Limited Rights as defined in FAR
27.401 or DFAR 227.7103-5 (c), as applicable in
any technical data. 52.227-14 (June 1987) or
DFAR 252.227-7015 (b)(2) (November 1995), as
applicable in any technical data.
Safety Notices
A CAUTION notice denotes a hazard.
It calls attention to an operating
procedure, practice, or the like that,
if not correctly performed or
adhered to, could result in damage
to the product or loss of important
data. Do not proceed beyond a CAU-TION notice until the indicated conditions are fully understood and
met.
A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like
that, if not correctly performed or
adhered to, could result in personal
injury or death. Do not proceed
beyond a WARNING notice until the
indicated conditions are fully
understood and met.
Technology Licenses
The hardware and/or software described in
this document are furnished under a license
and may be used or copied only in accordance with the terms of such license.
Page 3
Contents
1 Using InfiniiMax III Series Probes
2 InfiniiMax III Probe Heads
3 Maintaining Probe Heads
InfiniiMax III Probe Amplifiers 6
InfiniiMax III and III+ Probing System Family Diagram 9
Proper Handling Techniques 11
N5477A Sampling Scope Adapter 16
Cautions 17
N5445A InfiniiMax III Differential Browser Probe Head 20
N2848A InfiniiMax III+ QuickTip InfiniiMode Probe Head 25
N5439A InfiniiMax III ZIF Probe Head 29
N2836A InfiniiMax III Solder-in Probe Head 35
N5441A InfiniiMax III Solder-in Probe Head 39
N5444A InfiniiMax III 2.92 mm / 3.5 mm / SMA Probe Head 42
N2835A InfiniiMax III Differential Connectivity Kit and Accessories 44
Strain Relieving the Probe Heads 46
Specifications and Characteristics 58
InfiniiMax III Probe Heads Dimensions 65
5 Calibration / Deskew Procedure
Procedure 70
6 Performance Verification
Bandwidth Performance Verification 78
DC Input Resistance Performance Verification 96
Performance Test Record 99
7Performance Plots
N2836A 26 GHz Solder-In Head (with N2803A) 103
N2838A/N5439A 25 GHz PCB ZIF Tip/Head (with N2803A) 106
N5440A/N5439A 28 GHz Ceramic ZIF Tip/Head (with N2803A) 109
N5447A/N5439A 28 GHz Ceramic High-Sensitivity ZIF Tip/Head (with N2803A) 112
N5441A 16 GHz Solder-In Head (with N2803A) 115
3
Page 4
N5444A 30 GHz SMA 2.92 mm Head (with N2803A) 118
N5445A 30 GHz Browser Head, 1 mm span (with N2803A) 121
N5445A 30 GHz Browser Head, 2 mm span (with N2803A) 124
N5445A 30 GHz Browser Head, 3 mm span (with N2803A) 127
8SPICE Models
SPICE Models for N5440/1/4/5/7A Probe Heads 130
SPICE Models for N2838A and N2836A Probe Head 136
Simplified InfiniiMax III Schematic 139
4InfiniiMax III Series Probes
Page 5
Keysight InfiniiMax III Series Probes
CAUTION
User’s Guide
1Using InfiniiMax III Series
Probes
InfiniiMax III Probe Amplifiers 6
InfiniiMax III and III+ Probing System Family Diagram 9
Proper Handling Techniques 11
N5477A Sampling Scope Adapter 16
Cautions 17
This chapter contains documentation on using the InfiniiMax III probing system.
The InfiniiMax III probing system offers you the highest performance available for
measuring differential and single-ended signals, with flexible connectivity solutions for
today’s high-density ICs and circuit boards. Four different InfiniiMax III probe amplifiers
ranging from 16 GHz to 30 GHz are available for matching your probing solution to your
performance and budget requirements. A proprietary 200 GHz fT InP (indium phosphide)
IC process with backside ground vias and novel thick film technology is utilized to
accommodate your highest performance needs.
The family diagram for the InfiniiMax III probing system is shown in “InfiniiMax III and III+
Probing System Family Diagram" on page 9. This system is designed to give you the
maximum flexibility in matching your probe to your setup. The following pages will discuss
each of the adapters, probe amplifiers, and probe heads in detail.
Refer to “Proper Handling Techniques" on page 11 before connecting any probe to the
oscilloscope.
5
Page 6
1Using InfiniiMax III Series Probes
InfiniiMax III Probe Amplifiers
There are four InfiniiMax III amplifier models:
•N2800A 16 GHz Probe Amplifier
•N2801A 20 GHz Probe Amplifier
•N2802A 25 GHz Probe Amplifier
•N2803A 30 GHz Probe Amplifier
Each probe amplifier is pre-loaded with its specific measured S-parameters. When used
with the Infiniium 90000 X-series, the oscilloscope downloads these parameters and
automatically corrects the response of the unique probe system.
Figure 1InfiniiMax III Probe Amplifier
Menu Button
Press this button to bring up the Probe dialog box in the oscilloscope GUI.
Light Button
Press this button to turn the LED headlights on the browser probe head on/off. Pressing
and holding this button will ramp the intensity of the LED headlights. You may want to
adjust the brightness to accommodate different lighting or glare conditions.
6InfiniiMax III Series Probes User’s Guide
Page 7
Sensitivity Indicator
CAUTION
CAUTION
CAUTION
If you are using the ZIF probe head, the sensitivity indicator tells you which ZIF tip is
connected to the probe head. If the N5447A ceramic high sensitivity ZIF tip is connected
then the “Hi” sensitivity LED will be illuminated. If the N5440A ceramic ZIF tip is connected
then the “Normal” sensitivity LED will be illuminated. This will assist you in ensuring that
the correct ZIF tip is selected in the oscilloscope GUI’s probe dialog box.
If you select the wrong pairing of ZIF probe head and ZIF probe tip in the oscilloscope GUI’s
probe menu, your waveforms will look incorrect.
DC Calibration Indicator
Indicates if the particular combination of probe amplifier, probe head, and oscilloscope
channel input is calibrated. Any time you use a new probe head type, a new probe
amplifier, or a different channel on the oscilloscope, the DC Cal indicator turns amber,
indicating that a DC calibration is required. After performing a calibration as described in
Chapter 5, “Calibration / Deskew Procedure”, the DC Cal indicator turns green.
Using InfiniiMax III Series Probes1
Channel Identification Rings
Use the channel-identification rings to match each probe to the color of the oscilloscope’s
input channel to which it is connected. This enables you to quickly know which probe is
connected to each channel without having to trace cables back to the oscilloscope inputs.
Channel Connection
Refer to “Proper Handling Techniques" on page 11 before connecting any probe to the
oscilloscope.
The connector labeled A in Figure 1 on page 6 plugs into one of the oscilloscope channel
inputs. If connecting it directly to an Infiniium 90000 X-series oscilloscope, plug it into the
channel and then turn the gray clutch/dial around the input (Figure 2 on page 8) until it
clicks. To disconnect the probe amplifier, loosen the clutch on the oscilloscope input and
disconnect the probe amplifier. If you are using an InfiniiMax III probe amplifier with
Keysight’s 86100D sampling oscilloscope, you will need to use the N5477A sampling
scope adapter.
The connector labeled B in Figure 1 on page 6 connects to one of the InfiniiMax III probe
heads.
InfiniiMax I and II probe heads are not compatible with the InfiniiMax III probe amplifiers and
vice versa.
InfiniiMax III Series Probes User’s Guide7
Page 8
1Using InfiniiMax III Series Probes
NOTE
Figure 2Clutches on an Infiniium 90000 X-series oscilloscope
The clutches on the Infiniium 90000 X-series do not require a mechanical calibration. They will maintain
the same tolerance as the Keysight torque wrench (8 in lbs ±1 in lbs).
Probe Amplifier Bandwidth Upgrades
The InfiniiMax III probe amplifiers are upgradeable. When you order a probe bandwidth
upgrade, you receive a new probe with a new model number and new serial number. The
bandwidth upgrades are:
•N5446A-001 16 GHz to 20 GHz bandwidth upgrade
•N5446A-002 20 GHz to 25 GHz bandwidth upgrade
•N5446A-003 25 GHz to 30 GHz bandwidth upgrade
•N5446A-004 16 GHz to 25 GHz bandwidth upgrade
•N5446A-005 16 GHz to 30 GHz bandwidth upgrade
•N5446A-006 20 GHz to 30 GHz bandwidth upgrade
8InfiniiMax III Series Probes User’s Guide
Page 9
Using InfiniiMax III Series Probes1
InfiniiMax III and III+ Probing System Family Diagram
The following diagram shows the InfiniiMax III and InfiniiMax III+ probes and probe heads.
The diagram is not drawn to scale.
Figure 3InfiniiMax III / III+ Probing System
InfiniiMax III Series Probes User’s Guide9
Page 10
1Using InfiniiMax III Series Probes
CAUTION
CAUTION
CAUTION
NOTE
For extreme temperature testing, use the N5450B InfiniiMax extreme temperature
extension cable with the N5441A solder-in head. The N5441A can withstand temperatures
from –55°C to +150°C for up to 250 test cycles.
None of the N2800/1/2/3A probe amplifiers can withstand the extreme temperatures (–55°C
to +150°C) that the N5450B can withstand. When using the N5450B extension cable, do not
subject the InfiniiMax III probe amplifier to extreme temperatures.
None of the other probe heads are designed for extreme temperature testing.
The N5449A includes one N2873A probe. The adapter is specifically tuned for the N2873A
probe. Similar probes (1 ΜΩ input) can be used. Other probes may not meet the
bandwidth specification.
When using the N5450B extension cable, do not subject the InfiniiMax III probe amplifier or
probe head (other than the N5441A solder-in probe head) to extreme temperatures.
N2849A QuickTips are also compatible with N2851A QuickTip Probe Heads, which are designed for
1130/1/2/4A InfiniiMax I and 1168/9A InfiniiMax II probes.
Use the N5443A Performance Verification (PV) accessory fixture to properly
position the probe for PV testing.
10InfiniiMax III Series Probes User’s Guide
Page 11
Proper Handling Techniques
CAUTION
CAUTION
CAUTION
CAUTION
CAUTION
CAUTION
This guide will assist you in properly handling your InfiniiMax III N2800A-series probes to
prevent damage and maintain high performance. For more safe-handling information, go
to www.keysight.com/find/scope-demo, click on See Keysight's probes in action, and then
InfiniiMax III ESD Best Practices demo video listed under the Document Library tab.
Electrostatic discharge (ESD) can quickly and imperceptibly damage or destroy
high-performance probes, resulting in costly repairs. Always wear a wrist strap when handling
probe components.
Probes are sensitive devices and should be treated with care. Do not bend or kink the probe
amplifier cable. Do not drop heavy objects on the probe, drop the probe from large heights,
spill liquids on the probe, etc. Any of these examples can significantly degrade the
performance of the probe.
When storing the probe, it is best to coil the cable in a large radius and avoid a net twist in the
cable during the process. This can be done in a similar manner to how garden hoses or
extension cords are typically coiled.
Using InfiniiMax III Series Probes1
InfiniiMax I and II probe heads cannot be used with InfiniiMax III probe amplifiers and
InfiniiMax III probe heads cannot be used with InfiniiMax I and II amplifiers.
Never allow the probe head to be connected to the probe amplifier, if the probe amplifier is not
connected to the oscilloscope channel.
Always disconnect an N5441A solder-in probe head from the probe amplifier before
unsoldering, moving to a new position, and re-soldering the head.
Before Connecting a Probe
InfiniiMax probes and accessories are ESD sensitive devices and should be treated with
care. Before using or handling the probe or accessories, always wear a grounded ESD wrist
strap and ensure that cables and probe heads are discharged before being connected.
All work, including connecting probe amplifiers to the oscilloscope, should be performed
at a static-safe work station as shown in Figure 4 on page 12.
InfiniiMax III Series Probes User’s Guide11
Page 12
1Using InfiniiMax III Series Probes
Figure 4Static-Safe Work Station
Many scopes including Keysight's 90000X series have a front-panel ground socket. You
can plug the wrist strap into the ground socket as seen in the following picture.
Figure 5Wrist Strap Connected to Oscilloscope Ground Socket
The static-safe work station shown in Figure 4 uses two types of ESD protection:
•Conductive table-mat and wrist-strap combination.
•Conductive floor-mat and heel-strap combination.
Both types, when used together, provide a significant level of ESD protection. Of the two,
only the table-mat and wrist-strap combination provides adequate ESD protection when
used alone. To ensure user safety, the static-safe accessories must provide at least 1 MΩ
of isolation from ground. Purchase acceptable ESD accessories from your local supplier.
12InfiniiMax III Series Probes User’s Guide
Page 13
WARNING
These techniques for a static-safe work station should not be used when working on
CAUTION
circuitry with a voltage potential greater than 500 volts.
Connecting a Probe to an Oscilloscope Channel
To protect against ESD damage, always use the following steps when connecting your
probe to the oscilloscope.
1If the Device Under Test (DUT) is not grounded to the oscilloscope via the AC mains
ground, connect the DUT ground to the oscilloscope ground. An example of a device
having a floating ground would be a battery-powered DUT.
2Attach the probe head to the DUT.
At this point, the N2800A-series probe amplifier must not be connected to the oscilloscope or
the probe head.
Using InfiniiMax III Series Probes1
Figure 6Probe Head Attached to DUT
3Connect the probe amplifier to the oscilloscope while hand tightening the dark gray clutch
ring. The clutch ring limits the applied torque to 8 in-lb (±1 in-lb).
Figure 7Probe Amplifier Connected to Oscilloscope
4Connect the probe head to the probe amplifier.
InfiniiMax III Series Probes User’s Guide13
Page 14
1Using InfiniiMax III Series Probes
CAUTION
CAUTION
When connecting a probe head to a probe amplifier, push straight in. When disconnecting a
probe head from an amplifier, pull the probe head connectors straight out of the sockets.
Never bend the probe head in order to pry it loose from the amplifier. Also, do not wiggle the
probe head up and down or twist it to remove the connectors from the sockets. This can
damage the pins in the amplifier or the probe head itself.
Figure 8Probe Head Connected to Probe Amplifier
Disconnecting a Probe from an Oscilloscope Channel
Always disconnect the probe head from the probe amplifier before:
•disconnecting the probe amplifier from the oscilloscope.
•switching the probe amplifier from one oscilloscope channel to another.
Never allow the probe head to be connected to the probe amplifier, if the probe amplifier is not
connected to the oscilloscope channel.
14InfiniiMax III Series Probes User’s Guide
Page 15
Using InfiniiMax III Series Probes1
CAUTION
Figure 9Probe Improperly Disconnected from Oscilloscope while Probe Head is Connected
to the Probe Amplifier
Probing the DUT
When making your measurements, you’ll often need to probe different locations on the
DUT. You can safely move any of the following three probe heads without having to first
break the amplifier-to-head connection:
•N5445A differential browser head
•N5439A ZIF head
•N5444A 2.92 mm/3.5 mm/SMA head
The only exception is when the DUT is not grounded to the oscilloscope via the AC mains
ground. In this case, connect the DUT ground to the oscilloscope ground before moving
the probe. An example of a device having a floating ground would be a battery-powered
DUT.
When probing with an N5441A solder-in probe head, always disconnect the probe head
from the amplifier before moving the head to a new probing point. This is because some
soldering-iron tips can hold a charge which can damage the probe amplifier.
Always disconnect an N5441A solder-in probe head from the probe amplifier before
un-soldering, moving to a new position, and re-soldering the head.
InfiniiMax III Series Probes User’s Guide15
Page 16
1Using InfiniiMax III Series Probes
N5477A Sampling Scope Adapter
The N5477A Sampling Scope Adapter allows you to connect the InfiniiMax III probing
system to the Infiniium 86100D DCA-X sampling oscilloscope or other RF instruments.
Figure 10N5477A InfiniiMax III Sampling Scope Adapter
Connect the NMD/3.5 mm side of the adapter to one of the channel inputs on the
sampling oscilloscope. Then connect the power cord to either the probe power output on
the DCA-X 86100D wideband oscilloscope (if the module being used has this receptacle)
or to the 1143A power module. Finally, connect the other side of the adapter to the
InfiniiMax III probe amplifier.
Previously, the DCA-X wideband oscilloscope was limited to 13 GHz of probing, but with
the N5477A sampling oscilloscope adapter, the DCA-X can now probe up to 30 GHz.
The N5477A sampling oscilloscope adapter can also be used to connect the InfiniiMax III
probing system to generic 50 Ohm test equipment as long as the 1143A power supply and
5062-1247 NMD male-to-3.5mm female adapter are used. Both of these parts can be
ordered through Keysight Technologies.
16InfiniiMax III Series Probes User’s Guide
Page 17
Cautions
CAUTION
CAUTION
CAUTION
CAUTION
CAUTION
CAUTION
Using InfiniiMax III Series Probes1
In addition to the CAUTION notices below, also refer to the CAUTION notices listed in each
section of this book.
Electrostatic discharge (ESD) can quickly and imperceptibly damage or destroy
high-performance probes, resulting in costly repairs. Always wear a wrist strap when handling
probe components.
Probes are sensitive devices and should be treated with care. Do not bend or kink the probe
amplifier cable. Do not drop heavy objects on the probe, drop the probe from large heights,
spill liquids on the probe, etc. Any of these examples can significantly degrade the
performance of the probe.
Whenever you disconnect a probe head from an InfiniiMax amplifier, pull the probe head
connectors straight out of the sockets. When connecting a probe head to an amplifier, push
straight in also. Never bend the probe head in order to “pop” it loose from the amplifier. Also,
do not wiggle the probe head up and down or twist it to remove the connectors from the
sockets. This can damage the pins in the amplifier or the probe head itself.
When storing the probe, it is best to coil the cable in a large radius and avoid a net twist in the
cable during the process. This can be done in a similar manner to how garden hoses or
extension cords are typically coiled.
InfiniiMax I and II probe heads cannot be used with InfiniiMax III probe amplifiers and
InfiniiMax III probe heads cannot be used with InfiniiMax I and II amplifiers.
Always remove the probe head from the device under test (DUT) before disconnecting the
probe amp from the oscilloscope.
InfiniiMax III Series Probes User’s Guide17
Page 18
1Using InfiniiMax III Series Probes
18InfiniiMax III Series Probes User’s Guide
Page 19
Keysight InfiniiMax III Series Probes
CAUTION
User’s Guide
2InfiniiMax III Probe Heads
N5445A InfiniiMax III Differential Browser Probe Head 20
N2848A InfiniiMax III+ QuickTip InfiniiMode Probe Head 25
N5439A InfiniiMax III ZIF Probe Head 29
N2836A InfiniiMax III Solder-in Probe Head 35
N5441A InfiniiMax III Solder-in Probe Head 39
N5444A InfiniiMax III 2.92 mm / 3.5 mm / SMA Probe Head 42
N2835A InfiniiMax III Differential Connectivity Kit and Accessories 44
Strain Relieving the Probe Heads 46
This chapter documents the InfiniiMax III probe heads.
Refer to “Proper Handling Techniques" on page 11 before connecting any probe head to a
probe amplifier.
19
Page 20
2InfiniiMax III Probe Heads
NOTE
N5445A InfiniiMax III Differential Browser Probe Head
The N5445A browser head (30 GHz) is the best choice for the
general-purpose trouble shooting of differential signals with
spring-loaded tips and variable spacing from 20 mil to 125 mil (or
0.5 mm to 3.1 mm).
The span between the signal tips is easily adjusted with a thumb
wheel on the browser (see Figure 11).
For performance plots, refer to Chapter 7, “Performance Plots.
Figure 11N5445A InfiniiMax III Differential Browser Head
Using the LED Headlights
A pair of LED headlights are integrated into the tip of the browser to illuminate the probing
area for better visibility. The headlights are controlled via the Light button on the InfiniiMax
III probe amplifier (see Figure 1 on page 6). Pressing this button turns the headlights on /
off while pressing the button and holding it down ramps the intensity of the headlights.
Adjusting the Tip Span
Turn the browser’s thumb wheel (see Figure 12) to adjust the tip spacing from 20 to
125 mil (0.5 mm to 3.1 mm). Do not force the adjustment near the end of its range.
20InfiniiMax III Series Probes User’s Guide
Page 21
Figure 12Adjusting the Tip Span
Tip-Span Entry for Probe Calibration
InfiniiMax III Probe Heads2
When calibrating the probe, you will be prompted to enter the span setting on the
oscilloscope’s Probe Calibration dialog box. To determine the tip span, do the following
steps:
1Adjust the browser’s tip span for your measurement.
2Locate the tip-span gauge on the browser’s protective cap as shown in Figure 13.
3Determine which of the three possible tip-span settings most closely matches the
browser’s tip span.
Figure 13Tip Span Gauge on Browser Protective Cap
InfiniiMax III Series Probes User’s Guide21
Page 22
2InfiniiMax III Probe Heads
CAUTION
Mounting the Browser
There are two holes on the back side of the browser. Use these holes to mount the browser
to a customer designed holder. Figure 14 below shows the dimensions of these mounting
holes.
Figure 14Dimensions of Mounting Holes
Probe Along the Browser’s Axis
To prevent tip damage, probe along the browser’s axis as shown in Figure 15. Hold the
probe vertical and perpendicular to the circuit board.
Figure 15Proper Probe Handling
To avoid damaging the browser’s tips, do not apply a side load to the browser.
22InfiniiMax III Series Probes User’s Guide
Page 23
InfiniiMax III Probe Heads2
CAUTION
CAUTION
CAUTION
Figure 16Improper Probe Handling
Do not apply too much force when browsing. The weight of the probe in your hand should be
sufficient. The axial travel of the probe is about 15 mils (0.4 mm).
The browser’s protective cap should be kept on the browser at all times except when probing.
Always remove the browser from the device under test (DUT) before disconnecting the probe
amp from the oscilloscope.
Replaceable Parts
Figure 17 shows the replaceable parts for th N5445A.
Figure 17Browser Tips and Ground Blade
InfiniiMax III Series Probes User’s Guide23
Page 24
2InfiniiMax III Probe Heads
Table 1N5445A Replaceable Parts
DescriptionQtyPart Number
Browser tips4N5476A
Ground Blade1N4855A
Ground Blade Screws1N4856A
N2787A 3D Probe Positioner
Using the N2787A 3D probe positioner with the N5445A browser probe head reduces the
chance of breaking the browser tips and ensures that the tips maintain solid contact. Use
the following steps to position the probe using the N2787A:
1Lock the vertical compliance of the probe positioner.
2Clamp the browser into the positioner, aligning the browser’s slot with the positioner’s
gripping pad.
3While holding the browser, loosen the main knob and position the probe.
4Use the browser’s own weight to depress the tips, and tighten the main knob to lock the
probe’s position.
Figure 18Using the Browser with the N2787A 3D Probe Positioner
24InfiniiMax III Series Probes User’s Guide
Page 25
InfiniiMax III Probe Heads2
NOTE
N2848A InfiniiMax III+ QuickTip InfiniiMode Probe Head
The N2848A QuickTip probe head is used with an
N2849A QuickTip and together they provide the
following advantages:
•Easy-to-make secure magnetic mechanical connection between probe head and
QuickTip as shown in Figure 19. No latch lever is used!
•Extreme temperature environments such as temperature chambers.
Figure 19Magnet Connections in Probe Head
When used with N2830A-series probe amplifiers, the N2848A is InfiniiMode compatible.
InfiniiMode allows you to make differential, common mode, and single ended measurements
without having to re-solder the tip leads. InfiniiMode is not available when using with
N2800A-series probe amplifiers.
Table 2N2848A with N2804A Probe Amplifier
ItemCharacteristic
Bandwidth16 GHz
Input C (Differential)340 fF
Input C (Single Ended200 fF
Rise Time27.1 ps (10 - 90%)
Fall Time19.3 ps (20 - 80%)
Permanently solder any number of N2849A QuickTips to your DUT as shown in Figure 20
on page 26. Because the probe head is magnetically connected (instead of mechanically
connected) to the QuickTip, you can effortlessly connect and disconnect to each QuickTip.
For best performance, position the QuickTip vertically on the DUT. The N2849A QuickTip
has two signal leads and two ground leads. The ground leads have minimal effect on your
differential measurements. However, if you are making only differential measurements you
InfiniiMax III Series Probes User’s Guide25
Page 26
2InfiniiMax III Probe Heads
CAUTION
NOTE
can optionally cut off the ground leads or fold them out of the way. Be aware that without
the ground leads, the mechanical stability of the QuickTip will be reduced and you will
need to stabilize the probe head.
Figure 20Probing with the N2848A Probe Head and N2849A QuickTip
Do not replace or repair the N2849A QuickTip’s resistor or ground leads. Attempting to do so
will damage the ability of the tip to mate with the N2848A probe head.
The N2848A does not include any N2849A QuickTips. The N2849A must be ordered separately.
26InfiniiMax III Series Probes User’s Guide
Page 27
N2849A QuickTips are also compatible with N2851A QuickTip InfiniiMode Probe Heads,
CAUTION
CAUTION
CAUTION
NOTE
NOTE
which are designed for 1130/1/2/4A InfiniiMax I and 1168/9A InfiniiMax II probes.
Connecting a QuickTip to the DUT
Use the following tips when soldering the QuickTips to your DUT:
•Orient the QuickTip vertically as shown in Figure 20 on page 26.
•Solder the four leads to vias or surfaces.
•When soldering to a via, always trim the lead close to the via’s underside.
•If a lead is to be soldered to a surface and not a via, make a stronger solder joint by
bending the end of the lead 90°. For signal leads, bend the wire approximately half way
between the resistor and the end of the wire. Bend the ground leads at about the same
distance.
Always mechanically strain-relieve the QuickTip head before using to protect both your probe
accessories and DUT from damage.
Be careful not to damage the tip wires when handling the QuickTips. Wires can be carefully
reshaped with tweezers or fingers if necessary.
InfiniiMax III Probe Heads2
The QuickTips are very fragile. They must be manufactured in this way in order to meet the
high-performance, high bandwidth applications they are intended for. Be extremely careful
when handling.
Resistor and wire leads on the QuickTip are factory trimmed to the proper length for use. Adding wire
length to the tip of the mini-axial lead resistors or to the ground leads will degrade the performance of
the probe.
Soldering the ground wires is not required when making differential or single-ended (+ or – leads)
measurements.
InfiniiMax III Series Probes User’s Guide27
Page 28
2InfiniiMax III Probe Heads
Cleaning the Magnetic Connections
If the three magnetic connections in the head become dirty, clean the connections using
the following steps:
1Use compressed air or a cloth to remove any loose dirt.
2Gently rub a small piece of tack putty (supplied with the probe) against the magnetic
connections to clean off any remaining surface grime.
Figure 21N2848A Head Before and After Cleaning
28InfiniiMax III Series Probes User’s Guide
Page 29
InfiniiMax III Probe Heads2
NOTE
NOTE
NOTE
N5439A InfiniiMax III ZIF Probe Head
The N5439A Zero Insertion Force (ZIF) probe head
provides 28 GHz bandwidth. The head supports three
types of economical replaceable tips which are shown in
Table 3. Solder as many tips onto your DUT as needed.
Because of the ZIF tip’s extremely low loading, the tips
can remain on the DUT as you easily move the probe
head from one probing site to the next.
For information on mechanically securing the probes to
protect your equipment and designs from damage, refer to “Strain Relieving the Probe
Heads" on page 46.
Table 3Available ZIF Tips
ZIF TipDescriptionVariable Tip SpacingQty
N2838A25 GHz 450W PC board tip provides robust design with
high bandwidth.
N5440A28 GHz 450W ceramic
(normal sensitivity) for
extremely low probe
loading.
N5447A28 GHz 450W ceramic
high-sensitivity for
extremely low probe
loading.
The N5439A does not include any ZIF tips. You must order either the N2838A, N5440A, or
N5447A in addition to N5439A.
5 mil to 250 mil
(0.127 mm to 6.35 mm).
5 mil to 80 mil
(0.127 mm to 2 mm).
5 mil to 80 mil
(0.127 mm to 2 mm).
5
5
5
To install or repair the N2838A resistor leads, refer to “Replacing Axial Resistor
Tips" on page 50.
For performance plots, refer to Chapter 7, “Performance Plots.
InfiniiMax III Series Probes User’s Guide29
Page 30
2InfiniiMax III Probe Heads
CAUTION
Ensuring Maximum N2838A Tip Performance
The specifications and performance plots of the N2838A ZIF tip were measured
with a nominal spacing of 40 mil (1 mm). In order to achieve the proper response
as shown in the performance plots, keep the mini-axial lead resistors roughly
parallel as shown in Figure 22, and use the tip wires on the mini-axial leads to get
the desired span.
If you need to position the resistors different than shown in this figure (that is,
resistor bodies close together or spread apart), use Keysight’s N2807A and
N2808A PrecisionProbe products to perform an AC calibration of the probe, which
properly captures the response. Increasing the spacing to 250 mil degrades the
performance some, but PrecisionProbe can be used to compensate or qualify the
effect.
Figure 22Proper Position of Resistors
Soldering a ZIF Tip to the DUT
The ZIF tips are very fragile. They must be manufactured in this way in order to meet the
high-performance, high bandwidth applications they are intended for. Be extremely careful
when handling.
30InfiniiMax III Series Probes User’s Guide
Page 31
InfiniiMax III Probe Heads2
NOTE
CAUTION
1Break off a ZIF tip/handle combination from the packaging holder at the point shown in the
figure.
Figure 23Five ZIF Tips on Packaging
2Flux and tin the leads on the target DUT.
Figure 24Preparing the DUT Leads
3While holding the plastic form, form the ZIF tip wires to match the DUT’s pitch and angle.
4Flux the ZIF tip wires and DUT leads.
5Position the ZIF tip with the gold traces facing up as shown in Figure 25 and carefully
re-flow the solder. This orients the tip so that it will properly mate with the probe head.
When soldering the tip to your DUT, use the tip handle to hold the tip. This allows you to position these
small tips without damaging them.
Do not dwell on this solder joint.
InfiniiMax III Series Probes User’s Guide31
Page 32
2InfiniiMax III Probe Heads
CAUTION
CAUTION
CAUTION
Figure 25Soldering the ZIF Tip
6Pull the ZIF handle from the ZIF tip in the direction shown in Figure 26. ZIF tips can be
carefully handled with your fingertips and reinserted into a plastic handle if necessary.
Figure 26Removing the Handle from the ZIF Tip
7You can connect ZIF tips to any of the locations on a DUT that you need to probe. The
probe head can be quickly moved between the tips.
Always mechanically strain-relieve the ZIF head before using to protect both your probe
accessories and DUT from damage. Refer to “Strain Relieving the Probe Heads" on page 46.
Be careful not to damage the tip wires when handling the ZIF tips. Wires can be carefully
reshaped with tweezers or fingers if necessary.
Connecting the Probe Head to the ZIF Tip
1Add strain relieve for the ZIF probe head as described in “Strain Relieving the Probe
Heads" on page 46.
2Form the coaxial cables to bring the probe head near the tip. Press the lever down on the
ZIF probe head (see Figure 27) and slide the probe head onto the tip. Pressing on this lever
removes the clamping force of the connector and enables you to insert or remove ZIF tips.
Stop if you encounter any resistance at all when sliding the probe head over the ZIF tip. Check
your alignment, make sure the lever is pressed, and try again. Inserting the ZIF tip should
require “zero” insertion force.
32InfiniiMax III Series Probes User’s Guide
Page 33
CAUTION
Always use the lever when inserting or removing ZIF tips.
NOTE
Figure 27Probe Head with Location of the Lever
InfiniiMax III Probe Heads2
Figure 28Probe Head Connected to a Soldered ZIF Tip
For more repeatable results, orient the probe connection perpendicular to the device as shown in
Figure 29.
InfiniiMax III Series Probes User’s Guide33
Page 34
2InfiniiMax III Probe Heads
NOTE
Figure 29Probe Oriented Perpendicular to Device
3To move the probe head to a different tip, press the lever and remove the probe head from
the ZIF tip. The ZIF tip remains soldered to the DUT. Then, simply connect the ZIF head to
another ZIF tip at a different location on the DUT.
You can use tweezers to actuate the lever in tight places.
34InfiniiMax III Series Probes User’s Guide
Page 35
N2836A InfiniiMax III Solder-in Probe Head
NOTE
The N2836A InfiniiMax III solder-in probe head is an economical semi-permanent
connection that provides up to 26 GHz of system bandwidth. The span of the leads can be
adjusted from 5 mil - 250 mil (0.127 mm - 6.35 mm).
To use the head’s InfiniiMode capability, use this head with N2830A-series probe
amplifiers. InfiniiMode allows you to make differential, common mode, and single ended
measurements without having to re-solder the tip leads.
Figure 30N2836A InfiniiMax III solder-in probe head
InfiniiMax III Probe Heads2
For information on mechanically securing the probes to protect your equipment and
designs from damage, refer to “Strain Relieving the Probe Heads" on page 46.
To install or repair the N2836A resistor leads, refer to “Replacing Axial Resistor
Tips" on page 50.
Figure 31Probe Head Connection to DUT
InfiniiMax III Series Probes User’s Guide35
Page 36
2InfiniiMax III Probe Heads
CAUTION
Ensuring Maximum Performance
The specifications and performance plots of the N2836A probe head were
measured with a nominal spacing of 40 mil (1 mm). In order to achieve the proper
response as shown in the performance plots, keep the mini-axial lead resisto rs
roughly parallel as shown in Figure 32, and use the tip wires on the mini-axial
leads to get the desired span.
Figure 32N2836A With 1 mm Tip Spacing
If you need to position the mini-axial lead resistors different than shown in this figure (that
is, resistor bodies close together or spread way apart), use Keysight’s N2807A and N2808A
PrecisionProbe products to perform an AC calibration of the probe. The AC calibration will
capture the response properly. Increasing the spacing to 250 mil will degrade the
performance some, but Precision Probe can be used to compensate or qualify the effect.
The axial resistors on the N2836A solder-in probe head are fragile. They must be
manufactured in this way in order to meet the high-performance, high bandwidth applications
they are intended for. Be careful when handling.
Soldering the Probe Head to the DUT
To solder the probe head to your DUT, complete the following steps. The procedure is very
similar to that for the ZIF probe tips used with the N5439A probe head. This procedure
does not show soldering the ground leads, but the same techniques are used.
1Apply flux to your target leads as shown in Figure 33.
Figure 33Applying Flux
2Tin the leads with a small amount of solder.
36InfiniiMax III Series Probes User’s Guide
Page 37
InfiniiMax III Probe Heads2
CAUTION
Figure 34Tin the Leads
3Use tweezers to form the probe head wires to fit your DUT’s geometry.
4Flux the DUT leads and your probe head wires.
Figure 35Applying Flux to Leads and Wires
5Position the probe head wires on the DUT leads and quickly re-flow the solder as shown in
Figure 36.
Do not leave the iron in contact with the probe head for more than a few seconds at a time.
Figure 36Positioning Wires
InfiniiMax III Series Probes User’s Guide37
Page 38
2InfiniiMax III Probe Heads
38InfiniiMax III Series Probes User’s Guide
Page 39
N5441A InfiniiMax III Solder-in Probe Head
CAUTION
CAUTION
NOTE
The N5441A InfiniiMax III solder-in probe head is an economical semi-permanent
connection that provides up to 16 GHz of system bandwidth. The span of the leads can be
adjusted from 5 mil - 80 mil (0.127 mm - 2 mm).
The N5441A probe head can be used with the N5450B InfiniiMax extreme temperature
extension cable. This is the only InfiniiMax III probe head that can withstand the –55°C to
+150°C extreme temperature range (for up to 250 test cycles).
When using the N5450B extension cable, do not subject the InfiniiMax III probe amplifier or
probe head (other than the N5441A solder-in probe head) to extreme temperatures.
InfiniiMax III Probe Heads2
Figure 37N5441A InfiniiMax III solder-in probe head
The wires on the N5441A are fragile. They must be manufactured in this way in order to meet
the high-performance, high bandwidth applications they are intended for. Be careful when
handling.
For information on mechanically securing the probes to protect your equipment and
designs from damage, refer to “Strain Relieving the Probe Heads" on page 46.
To install or repair the tip wires on the N5441A, refer to “Replacing N5441A Probe
Head Wires" on page 53.
Soldering the Probe Head to the DUT
To solder the probe head to your DUT, complete the following steps. The procedure is very
similar to that for the ZIF probe tips used with the N5439A probe head.
1Position the probe head near the location on the DUT where you want to solder the probe.
2Add strain relieve for the probe head as described in “Strain Relieving the Probe
Heads" on page 46.
3Apply flux to your target leads as shown in Figure 38.
InfiniiMax III Series Probes User’s Guide39
Page 40
2InfiniiMax III Probe Heads
CAUTION
Figure 38Applying Flux
4Tin the leads with a small amount of solder.
Figure 39Tin the Leads
5Use tweezers to form the probe head wires to fit your DUT’s geometry.
6Flux the DUT leads and your probe head wires.
Figure 40Applying Flux to Leads and Wires
7Position the probe head wires on the DUT leads and quickly re-flow the solder as shown in
Figure 41.
Do not leave the iron in contact with the probe head for more than a few seconds at a time.
40InfiniiMax III Series Probes User’s Guide
Page 41
Figure 41Positioning Wires
InfiniiMax III Probe Heads2
InfiniiMax III Series Probes User’s Guide41
Page 42
2InfiniiMax III Probe Heads
N5444A InfiniiMax III 2.92 mm / 3.5 mm / SMA Probe Head
The N5444A InfiniiMax III 2.92 mm/3.5 mm/SMA probe head provides 30 GHz bandwidth
and allows you to connect two 2.92 mm, 3.5 mm, or SMA cables to make a differential
measurement on a single oscilloscope channel.
The N5444A provides for a termination to a common DC voltage rather than to ground,
which is required for many signal standards. It is implemented such that from DC to
approximately 1 kHz, the termination is 55 Ohms to the termination voltage, and above
approximately 10 kHz, the termination is 50 Ohms to 0.9 times the termination voltage.
The termination voltage range is ±4V with a minimum step of 5 mV and a maximum
current of 80 mA. The termination voltage can be controlled internally by the oscilloscope
or applied externally using the supplied DC jack.
Figure 42N5444A InfiniiMax III 2.92mm/3.5mm/SMA Probe Head
N5448B Coaxial Phase Matched Cable Pair
For extending the cable length of the N5444A InfiniiMax III probe head and add flexibility
and convenience to the probing setup, you can order N5448B 10 inches (25 cm) long
coaxial phase matched cable pair. This cable pair supports 2.92 mm male-to-2.92 mm
male connection.
Figure 43N5448B Coaxial Phase Matched Cable Pair
Before connecting these cables to the N5444A probe head, you must first remove the
supplied rigid cables of the N5444A probe head. Figure 44 on page 43 shows the N5448B
cables attached to the N5444A probe head.
42InfiniiMax III Series Probes User’s Guide
Page 43
Figure 44N5444A Probe Head With N5448B Cables Attached
CAUTION
InfiniiMax III Probe Heads2
The maximum bend radius for the N5448B coaxial cable pair is 30 mm. Bending these
cables at too tight a radius or twisting the cables can cause damage, reduce
performance, and impact the precision of these cables.
Also, ensure that the plastic caps that are provided with these cables are installed
when the cables are not in use.
For details on the N5448B cable, refer to its guide available in the Document Library tab of
this product’s page on www.keysight.com.
Probe Offset Calibration with the N5444A
To achieve more accurate probe offset calibration with the N5444A, an SMA
shorting adapter (Keysight part number: 1250-3999) is required. This adapter is
provided with the N5444A (with unit serial number US50072545 or higher) and
should be connected to the N5444A's "—" input connector during the calibration.
The latest Infiniium firmware includes instructions on the Probe Calibration dialog box to
remind you to install the adapter. If the Infiniium oscilloscope's firmware is older than
version 5.50.33, you can still attach the adapter and the resulting calibration will be valid.
InfiniiMax III Series Probes User’s Guide43
Page 44
2InfiniiMax III Probe Heads
N2835A InfiniiMax III Differential Connectivity Kit and Accessories
In addition to the individual probe heads described in the previous topics, the
N2835A differential connectivity kit is also available for the InfiniiMax III probes.
This kit provides multiple quantities of the four InfiniiMax III probe heads as shown
in Figure 45. You can order this kit either at the same time as InfiniiMax III probe
amplifiers or separately later.
Figure 45Probe Heads Included in the N2835A Differential Connectivity Kit (not to scale)
Table 4Supplied Accessories (Sheet 1 of 2)
Qty
Description
N2836A InfiniiMax III 26 GHz Differential Solder-In Probe
Head
Replacement Axial Resistors Kit10N2836-68701
N5439A InfiniiMax III 28 GHz Differential ZIF Solder-in Probe
Head
N2838A InfiniiMax III 25 GHz ZIF Tip Kit2 kits (5 tips in
N2848A InfiniiMax III QuickTip Probe Head2N2848A
N2849A InfiniiMax QuickTip Tips Kit 2 kits (4 tips in
N5445A InfiniiMax III Browser Probe Head1N5445A
Replacement Tips
Tweezer for replacing tips
Screw Driver
Protective End Cap
Supplied
2N2836A
2N5439A
each kit)
each kit)
4
1
1
1
Part
Number
N2838A
N2849A
N5476A
N5445-23801
N5445-23802
N5445-44101
44InfiniiMax III Series Probes User’s Guide
Page 45
Table 4Supplied Accessories (Sheet 2 of 2)
InfiniiMax III Probe Heads2
Description
Qty
Supplied
Ground Blades4N5445-68700
Part
Number
InfiniiMax III Series Probes User’s Guide45
Page 46
2InfiniiMax III Probe Heads
Strain Relieving the Probe Heads
High-performance probes have small physical geometries to ensure the lowest possible
loading and best electrical response. Because of their small size, probing accessories are
often delicate. It is important to mechanically secure your probes to protect both your
equipment and designs from damage. There are several methods that Keysight
recommends:
Tack-putty
Keysight recommends the use of tack-putty for securing both probe heads and amplifiers.
Wrap a small amount of tack-putty around your probe head cables, taking care to not
pinch them. The mass can then be secured to a rigid body neat your DUT.
A similar techniques can be used to secure probe amplifiers where you apply some
tack-putty to the underside of the probe amplifier body and attach it to a rigid body near
your DUT.
Figure 46Using tack-putty to strain relieve a probe head and amplifier
You can also use putty with a positioner, such as the Keysight N2787A. The same
positioner can also be used to support your probe amplifier as shown in Figure 18 on
page 24.
Figure 47Using putty with the N2787A 3D Probe Positioner
46InfiniiMax III Series Probes User’s Guide
Page 47
Low-temperature Hot Glue
CAUTION
CAUTION
You can also use low-temperature hot glue to secure cables.
Only use LOW temperature hot glue. To remove the hot glue, warm it with a heat gun set on
low. Only heat the hot glue enough to remove it.
Figure 48Using low-temperature hot glue to secure a probe head
InfiniiMax III Probe Heads2
General Tips on Strain Relief
Keep in mind that different accessories have different cable stiffness. You should choose a
strain relief method appropriate for the cable stiffness. For instance, it is best to secure the
stiffer N5439A near the SMP connectors and form the cable to the optimal location.
Other strain-relief options like tape or hook-and-loop work fine as well, but keep the
following guidelines in mind to protect your probing investment.
Do not kink cables. Do not crush cables. Do not use aggressive adhesives or high
temperatures.
Table 5Replaceable Parts
PartPart Number
Extra wire (for solder-in probe head only)01169-81301 (7 mil)
Many probe heads come equipped with replaceable resistor or wire tips which can
be replaced or repaired. Using the procedures in this chapter, you can extend the
life of the following items:
•N2836A probe heads
•N5441A probe heads
•N2838A ZIF tips
Do not replace or repair the N2849A QuickTip’s resistor or ground leads. Attempting to do so
will damage the ability of the tip to mate with the N2848A probe head.
49
Page 50
3Maintaining Probe Heads
NOTE
CAUTION
CAUTION
Replacing Axial Resistor Tips
The procedure in this sections shows you how to replace the 130 ohm axial resistors that
are located at the tip of the
•N2836A solder-in probe head and
•N2838A ZIF tip.
•These resistors can become worn or damaged with use. Order the replacement axial
resistor kit (N2836-68701) which provides 10 resistors.
The pictures in the following procedure show the N2836A solder-in probe head, but the same
procedure applies to the N2838A ZIF tip.
Recommended Equipment
•Vise or clamp for holding tip.
•Metcal STTC-022 (600 °C) or STTC-122(700 °C) tip soldering iron or equivalent. The
600 °C tip will help limit burning of the FR4 tip PC board.
•0.381 mm (0.015 in) diameter RMA flux standard tin/lead solder wire.
•Fine stainless steel tweezers.
•Rosin flux pencil, RMA type (Kester #186 or equivalent).
•Diagonal cutters.
•Magnifier or low power microscope.
•Ruler.
Procedure
1As shown in Figure 49, clamp the probe head or ZIF tip in a vise. Tweezers can be used to
As the probe heads and tips are easily damaged, only experienced soldering technicians
should attempt this repair.
hold the probe head or ZIF tip away from the vise. When using tweezers, grip the tip either
on the sides or top and bottom.
When tightening the vise, use light force to avoid damaging the solder-in probe head.
50InfiniiMax III Series Probes User’s Guide
Page 51
Figure 49Clamping the Part
NOTE
NOTE
CAUTION
CAUTION
Maintaining Probe Heads3
2Grab each resistor lead or wire with tweezers and pull very gently up. Touch the soldering
iron to solder joint just long enough for the resistor to come free of the probe head tip.
Figure 50Removing the Resistor
Clean the soldering iron tip of any excess solder before using.
The solder joint has very low thermal mass so the joint will quickly melt and release.
To limit burning and damage to the PC board, do not keep the soldering iron in contact with
the tip any longer than is necessary.
Excessive dwell time with the iron will permanently damage the flip-chip resistor.
InfiniiMax III Series Probes User’s Guide51
Page 52
3Maintaining Probe Heads
CAUTION
CAUTION
NOTE
3Use the soldering iron and solder to fill the holes in preparation for mounting the new
resistors (or wires).
Do not leave the iron in contact with the tip any longer than necessary.
4Use the flux pencil to coat the solder joint area with flux.
5Locate the trim gauge which is supplied with the N2836-68701 replacement axial resistor
kit.
a
Place a resistor over the lead length gauge shown in Figure 51. Trim the
leads to match the drawing. The orientation of the lead is not important.
b Place a resistor over the bend gauge and bend the leads to match the
drawing. This bend fits in the hole in the tip’s PC board.
Figure 51Trim Guage
6Holding the resistor lead or wire in one hand and soldering iron in the other, position the
end of the resistor lead (after the 90 degree bend) over the solder filled hole.
Figure 52Soldering in a new axial resistor
7Touch the soldering iron to the side of the hole. When the solder in the hole melts, the
resistor lead will fall into the hole.
Remove the soldering iron as soon as the lead falls into the hole.
Because the thermal mass of the joint is very small, extra dwell time with the soldering iron is not
needed to ensure a good joint.
8Using a digital volt meter, measure the resistance from the coax center conductor to the
resistor tip. The DC resistance should measure 450 ohms.
52InfiniiMax III Series Probes User’s Guide
Page 53
Replacing N5441A Probe Head Wires
CAUTION
Use the following procedure to install or replace the wire leads on the N5441A solder-in
probe head. Depending on your probing application, you can order either 5 mil or 7 mil
wire as listed in the following table. For example, use the 5 mil wire for attaching to small
vias.
Table 6Required Wire Type
Wire DiameterPart Number
0.007 inch (tin-plated nickel wires)01169-81301
0.005 inch (tin-plated nickel wires)01169-21306
Table 7Recommended Equipment
Equipment
Maintaining Probe Heads3
Vise or clamp for holding tip
Metcal STTC-022 (600 °C) or STTC-122 (700 °C) tip soldering iron or equivalent. The 600 °C tip will help limit burning of the
FR4 tip PC board.
0.381 mm (0.015 in) diameter RMA flux standard tin/lead solder wire
Fine stainless steel tweezers
Rosin flux pencil, RMA type (Kester #186 or equivalent)
Flush cutting wire cutters
Magnifier or low power microscope
Keysight supplied trim gauge (01169-23801)
Procedure
1Use the vise or clamp to position the tip an inch or so off the work surface for easy access.
If using a vise, grip the tip on the sides with light force. When tightening the vise, use light
force to avoid damaging the solder-in probe head If using a tweezers clamp, grip the tip either
on the sides or at the top and bottom.
InfiniiMax III Series Probes User’s Guide53
Page 54
3Maintaining Probe Heads
CAUTION
Figure 53Clamping the Probe Head
2As shown in Figure 54, remove the old wires with tweezers while re-flowing the solder from
the underside of the probe.
Apply heat quickly to avoid damaging your probe.
Figure 54Removing the Old Wire
3If necessary, add a small amount of solder to the holes and apply flux.
54InfiniiMax III Series Probes User’s Guide
Page 55
Maintaining Probe Heads3
CAUTION
Figure 55Adding Solder and Flux
4Re-flow the solder from the underside and insert a new piece of wire. It is best to shape the
wire into an “L” before attempting to insert.
Do not dwell with the iron in contact with the probe head.
Figure 56Adding a New Wire
Figure 57New Wires Properly Attached
5Trim any wire stubs on the probe head underside.
InfiniiMax III Series Probes User’s Guide55
Page 56
3Maintaining Probe Heads
Figure 58Trim Wire Stubs
6Use the included trim gauge to cut the wire lengths. Doing so ensures the best
performance from your probe head.
Figure 59Trim Gauge Placed on Wires
Figure 60Removing Excess Wire
7Check the DC resistance of each probe leg when you have replaced the wires. The correct
resistance should be 450 ohms.
56InfiniiMax III Series Probes User’s Guide
Page 57
Keysight InfiniiMax III Series Probes
User’s Guide
4Performance Specifications /
Characteristics
The InfiniiMax III series probe amplifiers and probe heads are designed to be used within
the temperature range of the 90000-X series oscilloscope. The only exception is the
N5441A solder-in probe head. The N5441A solder-in probe head temperature range is
from –55°C to +150°C for up to 250 test cycles. The N5441A is compatible with the
N5450B InfiniiMax extreme temperature extension cable.
57
Page 58
4Performance Specifications / Characteristics
Specifications and Characteristics
Table 8Probe Amplifiers
Probe AmplifierBWtr
N2800A16 GHz27.1 ps
N2801A20 GHz21.7 ps
N2802A25 GHz17.4 ps
N2803A30 GHz14.3 ps
58InfiniiMax III Series Probes User’s Guide
Page 59
Performance Specifications / Characteristics4
Table 9N2803A Amplifier with N2836A, N5441A, and N2848A Probe Heads (Sheet 1 of 2)
Specifications / CharacteristicsN2836A
Solder-In
Probe Bandwidth
(–3 dB), Probe Only
Rise and Fall Time, Probe Only
(Single-Ended)
Rise and Fall Time, Probe Only
(Differential Mode)
Input Resistance (> 10 kHz)Rse = 500Ω each input to
ground
Rdiff = 1 kΩ
Rcm = 250Ω
Input Voltage Range
(Differential or Single-Ended)
Input Common Mode Range±12V DC to 250 Hz
DC attenuation Ratio6:16:15:1 or 10:1
Offset Range (for probing a
single-ended signal)
Input Referred Noise Spectral
Density
Input Referred Noise Example4 mVrms4 mVrms
*Denotes warranted characteristic - all others are typical
**Harmonic distortion < –38 dB is standard; < –34 dB wider input range with slightly increased distortion
1.6Vpp, ±0.8V
(HD2&3< –38 dBc typ.),
2.5Vpp, ±1.25V
(HD2&3< –34 dBc typ.) **
±1.25V > 250 Hz
±16V±16V±16V
23.9 nV/rt (Hz)23.9 nV/rt (Hz)23.9 nV/rt (Hz)
Rse = 500Ω each input to
ground
Rdiff = 1 kΩ
Rcm = 250Ω
1.6Vpp, ±0.8V
(HD2&3< –38 dBc typ.),
2.5Vpp, ±1.25V
(HD2&3< –34 dBc typ.) **
±12V DC to 250 Hz
±1.25V > 250 Hz
Rse = 500Ω
ground
Rdiff = 1 kΩ
Rcm = 250Ω
1.6Vpp, ±0.8V
(HD2&3< –38 dBc typ.),
2.5Vpp, ±1.25V
(HD2&3< –34 dBc typ.)**
±12V DC to 250 Hz
±1.25V > 250 Hz
each input to
InfiniiMax III Series Probes User’s Guide59
Page 60
4Performance Specifications / Characteristics
Table 9N2803A Amplifier with N2836A, N5441A, and N2848A Probe Heads (Sheet 2 of 2)
Specifications / CharacteristicsN2836A
Solder-In
Maximum Input Voltage18V peak CAT I18V peak CAT I18V peak CAT I
*Denotes warranted characteristic - all others are typical
**Harmonic distortion < –38 dB is standard; < –34 dB wider input range with slightly increased distortion
N5441A
Solder-In)
N2848A QuickTip Head
with N2849A Tip
60InfiniiMax III Series Probes User’s Guide
Page 61
Performance Specifications / Characteristics4
Table 10N2803A Amplifier with N2838A, N5447A, and N5440A ZIF Tips (Sheet 1 of 2)
Specifications / CharacteristicsN5439A ZIF Probe Head
N2838A
PC Board ZIF Tip
Probe Bandwidth
(–3 dB), Probe Only
Rise and Fall Time, Probe Only
(Single-Ended)
Rise and Fall Time, Probe Only
(Differential Mode)
System Bandwidth
(–3 dB) with DSO/DSAX93204A
Rise and fall time with
DSO/DSAX93204A
Input CapacitanceCdiff = 95 fF
DC Input Resistance*Rse = 50 kΩ ±2% each
Input Resistance (> 10 kHz)Rse = 500Ω each input to
Rse = 50 kΩ ±2% each
input to ground
Rdiff = 100 kΩ ±2%
Rcm = 25 kΩ ±2%
Rse = 500Ω each input to
ground
Rdiff = 1 kΩ
Rcm = 250Ω
N5447A
High Sensitivity Ceramic
ZIF Tip
28 GHz (typical)
20.9 ps (10-90%)
13.8 ps (20-80%)
15.5 ps (10-90%)
11.0 ps (20-80%)
15.5 ps (10-90%)
11.0 ps (20-80%)
Cdiff = 32 fF
Cse = 44 fF
Rse = 25 kΩ ±2% each
input to ground
Rdiff = 50 kΩ ±2%
Rcm = 12.5 kΩ ±2%
Rse = 250Ω each input to
ground
Rdiff = 500Ω
Rcm = 125Ω
Input Voltage Range
(Differential or Single-Ended)
Input Common Mode Range±12V DC to 250 Hz
DC attenuation Ratio6:16:13:1
Offset Range (for probing a
single-ended signal)
Input Referred Noise Spectral
Density
Input Referred Noise Example4 mVrms4 mVrms2 mVrms
*Denotes warranted characteristic - all others are typical
**Harmonic distortion < –38 dB is standard; < –34 dB wider input range with slightly increased distortion
InfiniiMax III Series Probes User’s Guide61
1.6Vpp, ±0.8V
(HD2&3< –38 dBc typ.),
2.5Vpp, ±1.25V
(HD2&3< –34 dBc typ.) **
±1.25V > 250 Hz
±16V±16V±8V
23.9 nV/rt (Hz)23.9 nV/rt (Hz)12.0 nV/rt (Hz)
1.6Vpp, ±0.8V
(HD2&3< –38 dBc typ.),
2.5Vpp, ±1.25V
(HD2&3< –34 dBc typ.)**
±12V DC to 250 Hz
±1.25V > 250 Hz
0.8Vpp, ±0.4V
(HD2&3< –38 dBc typ.),
1.6Vpp ±0.8V
(HD2&3< –34 dBc typ.)**
±6V DC to 250 Hz
±0.65V > 250 Hz
Page 62
4Performance Specifications / Characteristics
Table 10N2803A Amplifier with N2838A, N5447A, and N5440A ZIF Tips (Sheet 2 of 2)
Specifications / CharacteristicsN5439A ZIF Probe Head
N2838A
PC Board ZIF Tip
Maximum Input Voltage18V peak CAT I18V peak CAT I18V peak CAT I
*Denotes warranted characteristic - all others are typical
**Harmonic distortion < –38 dB is standard; < –34 dB wider input range with slightly increased distortion
N5440A
Ceramic ZIF Tip
N5447A
High Sensitivity Ceramic
ZIF Tip
62InfiniiMax III Series Probes User’s Guide
Page 63
Performance Specifications / Characteristics4
Table 11N2803A Amplifier with N5444A and N5445A Probe Heads (Sheet 1 of 2)
Specifications / CharacteristicsN5444A
SMA, 3.5 mm, 2.92 mm Adapter
Probe Bandwidth
(–3 dB), Probe Only
Rise and Fall Time, Probe Only
(Single-Ended)
Rise and Fall Time, Probe Only
(Differential Mode)
System Bandwidth
(–3 dB) with DSO/DSAX93204A
Rise and fall time with DSO/DSAX93204A15.5 ps (10-90%)
Input CapacitanceN/ACdiff = 35 fF
DC Input Resistance*55Ω to VtermRse = 50 kΩ ±2% each input to gnd
Input Resistance (> 10 kHz)50Ω to (0.909)(Vterm)Rse = 500Ω each input to ground
30 GHz (typical)30 GHz (typical),
15.5 ps (10-90%)
11.0 ps (20-80%)
15.5 ps (10-90%)
11.0 ps (20-80%)
30 GHz30 GHz
11.0 ps (20-80%)
N5445A
Browser
28 GHz warranted
16.2 ps (10-90%)
10.9 ps (20-80%)
14.5 ps (10-90%)
10.3 ps (20-80%)
14.3 ps (10-90%)
10.2 ps (20-80%)
Cse = 50 fF
Rdiff = 100 kΩ ±2%
Rcm = 25 kΩ ±2%
Rdiff = 1 kΩ
Rcm = 250Ω
Input Voltage Range
(Differential or Single-Ended)
Input Common Mode Range±6V DC to 250 Hz
DC attenuation Ratio6:16:1
Offset Range
(for probing a single-ended signal)
Input Referred Noise Spectral Density23.9 nV/rt (Hz)23.9 nV/rt (Hz)
Input Referred Noise Example4 mVrms4 mVrms
Maximum Input Voltage8V peak without violating maximum
*Denotes warranted characteristic - all others are typical
**Harmonic distortion < –38 dB is standard; < –34 dB wider input range with slightly increased distortion
1.6Vpp, ±0.8V (HD2&3< –34 dBc),
2.5Vpp, ±1.25V (HD2&3< –38 dB)**
±1.25V > 250 Hz without violating
maximum input power
±6V without violating maximum
input power
input power
1.6Vpp, ±0.8V (HD2&3< –34 dBc),
2.5Vpp, ±1.25V (HD2&3< –38 dB)**
±12V DC to 250 Hz
±1.25V > 250 Hz
±16V
18V peak CAT I
InfiniiMax III Series Probes User’s Guide63
Page 64
4Performance Specifications / Characteristics
P
max
rms
VinV
term
–()
()
2
55
---------------------------------------
=
Table 11N2803A Amplifier with N5444A and N5445A Probe Heads (Sheet 2 of 2)
Specifications / CharacteristicsN5444A
SMA, 3.5 mm, 2.92 mm Adapter
Maximum Input Power125 mW calculated with the
N5445A
Browser
N/A
following equation for each input:
*Denotes warranted characteristic - all others are typical
**Harmonic distortion < –38 dB is standard; < –34 dB wider input range with slightly increased distortion
64InfiniiMax III Series Probes User’s Guide
Page 65
InfiniiMax III Probe Heads Dimensions
All dimensions included in this section are in millimeters [inches].
N2836A InfiniiMax III Solder-in Head Dimensions
Performance Specifications / Characteristics4
N2838A InfiniiMax III PCB ZIF Tip with N5439A ZIF Head Dimensions
InfiniiMax III Series Probes User’s Guide65
Page 66
4Performance Specifications / Characteristics
N2848A InfiniiMax III QuickTIp Tip with N2849A QuickTip Head
Dimensions
N5440A/N5447A InfiniiMax III Ceramic ZIF Tip with N5439A ZIF Head
Dimensions
66InfiniiMax III Series Probes User’s Guide
Page 67
Performance Specifications / Characteristics4
N5441A InfiniiMax III Solder-in Head Dimensions
InfiniiMax III Series Probes User’s Guide67
Page 68
4Performance Specifications / Characteristics
68InfiniiMax III Series Probes User’s Guide
Page 69
Keysight InfiniiMax III Series Probes
User’s Guide
5Calibration / Deskew
Procedure
To perform a calibration/deskew for the InfiniiMax III Probing system, you will be using the
N5443A Calibration/Deskew Fixture.
This fixture enables you to not only calibrate / deskew your InfiniiMax III probe, but it also
ensures you properly position the probe amplifier during the procedure via the plastic
holder. You can remove the plastic holder from the fixture by removing the four screws on
the bottom side of the holder if you prefer.
69
Page 70
5Calibration / Deskew Procedure
Procedure
1To perform a DC calibration / calibrated skew, make sure the 50 ohm terminator (included
with N5443A) is connected to the fixture as shown below.
2Hold the N5443A upright and connect the fixture to the Cal Out on the oscilloscope. Turn
the nut on the Cal Out counter-clockwise to tighten.
3While holding the fixture upright with one hand, use an 8 in. lbs. torque wrench to fully
tighten the connector as shown on the next page.
70InfiniiMax III Series Probes User’s Guide
Page 71
Calibration / Deskew Procedure5
CAUTION
4Connect the probe amplifier to one of the channel inputs on the oscilloscope.
5Insert the amplifier into the top of the fixture holder. The amplifier can slide up and down in
the holder to adjust the probe head position.
Always wear an ESD wrist strap when working with active probes. Not doing so can result in
the probe becoming permanently damaged.
6Before connecting to the N5443A fixture, form the N5439A probe head ZIF tip wires as
shown below (if using the ZIF tip for the calibration / deskew). If you are using the browser
probe head then you do not need to adjust the shape of the tips.
InfiniiMax III Series Probes User’s Guide71
Page 72
5Calibration / Deskew Procedure
Form the wires as shown below if using the N5441A solder-in probe head.
7Connect your probe head to the amplifier.
8Position the probe head wires (if you are using the ZIF or Solder-in probe heads) so they
curl towards the scope as shown below.
72InfiniiMax III Series Probes User’s Guide
Page 73
Calibration / Deskew Procedure5
9Press the Menu button on the probe amplifier to bring up the Probe Setup dialog box on
the oscilloscope.
10In the Probe Setup dialog box, click on the Calibrate Probe ... button. This will open the
Probe Calibration dialog box.
11In the Probe Calibration dialog box, select the probe head you are using in the Head Label
(Type) field, select the Calibrated Atten/Offset radio button, and press the Start
Atten/Offset Calibration... button.
InfiniiMax III Series Probes User’s Guide73
Page 74
5Calibration / Deskew Procedure
NOTE
12A dialog box will then open that tells you how to connect the probe head to the
calibration/deskew fixture.
The fixture has three spring-loaded fingers (shown below) that clamp probe head wires to
the fixture if you are using the ZIF probe head or the Solder-in probe head.
13On the fixture, the center gold trace is signal and the large plates on either side are both
ground as shown below.
74InfiniiMax III Series Probes User’s Guide
Page 75
Calibration / Deskew Procedure5
CAUTION
NOTE
CAUTION
Never solder a probe tip to the thickfilm gold. The gold will immediately dissolve into the
solder and disappear.
14Follow the instructions on the oscilloscope to clamp the wires to either the signal or
ground. When connecting the probe wires to the fixture, press down on the spring-loaded
fingers described in Step 12 and insert the probe wires. You can check that the wires are
connected correctly by pressing the autoscale button on the front panel and checking that
you have a stable step on screen (note that pressing autoscale will close the Probe
Calibration dialog box).
People tend to have their own preference for this step in terms of the angle to view and insert the probe
wires under the spring-loaded fingers. Try different angles to determine your optimum method.
When connecting the probe head to the fixture, do not press down with much force or you
could snap off the fixture from the Cal Out connection. Light contact is all that is needed for
the calibration.
15If you are using the browser probe head, it is recommended that you use the N2787A 3D
Probe Positioner to hold the browser in place (as shown below). The browser uses
spring-loaded tips so you do not need much force to get a solid contact. You can check
that the tips are connected correctly by pressing the autoscale button on the front panel
InfiniiMax III Series Probes User’s Guide75
Page 76
5Calibration / Deskew Procedure
CAUTION
NOTE
and checking that you have a stable step on screen (note that pressing autoscale will close
the Probe Calibration dialog box).
When connecting the probe head to the fixture, do not press down with much force or you
could snap off the fixture from the Cal Out connection. Light contact is all that is needed for
the calibration.
16With the probe head correctly connected to the fixture, press OK in the dialog box on the
oscilloscope and wait for the calibration to finish. It will report whether the calibration was
successful or not.
17Once the calibration has successfully completed, the DC Cal LED on the probe amplifier
will turn green indicating that the particular combination of probe amplifier, probe head,
and oscilloscope channel input has been calibrated.
Any time you use a new probe head type, a new probe amplifier, or a different channel on the
oscilloscope, the DC Cal LED will be amber, indicating that a DC calibration is required.
18Once the DC vertical calibration has successfully completed, select the Calibrated Skew
radio button from the Probe Calibration dialog box on the oscilloscope. Then press the
Start Skew Calibration... button and follow the on-screen instructions for the skew
calibration.
76InfiniiMax III Series Probes User’s Guide
Page 77
Keysight InfiniiMax III Series Probes
CAUTION
User’s Guide
6Performance Verification
Bandwidth Performance Verification 78
DC Input Resistance Performance Verification 96
Performance Test Record 99
This chapter describes the equipment and procedures needed to verify the performance of
InfiniiMax III probes. Due to the very high frequency of the InfiniiMax III probing system, it is
important to carefully adhere to the techniques and procedures described in this chapter
to accurately measure the performance.
It is also important to note that the performance measured here is of the probe by itself.
Keysight high performance real-time scopes (and sampling scopes under certain
conditions) will apply probe correction that will further enhance the performance of the
probes.
Electrostatic discharge (ESD) can quickly and imperceptibly damage or destroy high
performance probes, resulting in costly repairs. Always wear a wrist strap when handling
probe components and insure that cables are discharged before being connected.
77
Page 78
6Performance Verification
Bandwidth Performance Verification
This section documents the bandwidth performance of the N2803A InfiniiMax III Probe
Amplifier with the N5439A ZIF probe head and N5440A ceramic ZIF tip or with the N5445A
browser probe head.
Keysight recommends a test interval of one year or 2000 hours of operation.
Equipment Needed
•Keysight 2 port E8361A/C Vector Network Analyzer:
•Or equivalent VNA that covers at least a 50 MHz to 34 GHz range. This procedure is
written assuming the E8361A/C PNA. If a different VNA is used, references that are
specific to the PNA will need to be modified.
•Needs proper test port cables and/or adaptors to provide male 2.92 mm connectors
at reference planes. If test port cables are 2.4 mm or 1.85 mm cables, then Keysight
11904A and 11904D adaptors can be used to convert to 2.92 mm male connectors.
•Needs to be capable of using a Touchstone file to de-embed at a port.
•Needs to have bias port for port 1 of the VNA (i.e. has internal bias T’s and a BNC
•Maury Microwave 8775B2 2.92 mm male broadband load, or other 2.92 mm male load
with similar or better return loss. A high quality 2.92 mm adaptor to a 2.4 mm or
1.85 mm VNA calibration load with required return loss could be used.
•Keysight N5477A Autoprobe II/3.5 mm Adaptor.
•Keysight 1143A Power supply.
•Keysight 5062-1247 outside thread 3.5 mm (male) to 3.5 mm (female) adaptor.
•Keysight N4692A-00F 2.92 mm (female/female) ECal module. Or, other 2.92 mm
calibration kit that can calibrate to the 2.92 mm male connectors at the test ports.
•BNC 50 ohm male terminator or equivalent; not a critical part. For example, Pomona
number 3840-50 or 4119-50.
•InfiniiMax III Probe Head. Either the N5439A ZIF probe head with the N5440A ceramic
ZIF probe tip or the N5445A browser probe head.
78InfiniiMax III Series Probes User’s Guide
Page 79
VNA Setup
CAUTION
•Power level: –6 dBm.
•Sweep: Log Frequency 50 MHz to 34 GHz, 284 points (100 pts/decade).
•IF BW: 1 kHz.
•Test port cables and adaptors (if needed) to provide male 2.92 mm connectors at
•Install the BNC 50 ohm terminator to the bias input for port 1 of the VNA (on the rear
•Clear all traces from display, then select S12 to display.
•Set scale for S12 to 3 dB/div, with reference level to 0 dB and reference position to 5
Procedure
1Calibrate the PNA to the two male 2.92 mm connectors (connectors can be seen in
Figure 62) using the N4692A-00F ECal module (or equivalent 2.92 mm cal kit).
Performance Verification6
measurement planes.
panel of E8361 PNA). This provides a DC 50 ohm termination for the probe amp output.
divisions.
As with all precision connector interfaces, make sure to torque all connections using the
proper torque wrench!
2Prepare the N5439A ZIF probe head for connection to the PV fixture as shown in Figure 61.
a
Install a N5440A ceramic ZIF tip into the N5439A ZIF probe head. Make sure
it is fully inserted.
b Bend the tip wires down at their halfway point using fine tweezers.
c Slightly spread the tips wires to better match the spacing needed for the PV
fixture.
Figure 61N5439A ZIF Probe Head
3Connect the 1250-1749 adapter and N5443A PV fixture assembly to the calibrated ports of
the PNA as shown in Figure 62. A small bench vise is useful to hold PV fixture steady.
InfiniiMax III Series Probes User’s Guide79
Page 80
6Performance Verification
Figure 62Setup for measuring “Vin” of probe
4Connect the N5477A AutoProbe II adaptor to the 1143A power supply and turn on the
power supply.
a
Make sure the probe offset control button on the 1143A is set to “Zero” so no
probe offset is applied.
b The 5062_1247 adapter should be attached to the N5477A and properly
torqued.
5Connect the probe amp pod end to the N5477A and torque connector.
6Connect the probe to the N5443A PV fixture:
a
Probe amp with ZIF probe head is inserted into the PV holder far enough
that the tip wires can easily reach the pinchers on the PV fixture.
b Form the coax cables so that the tip wires are close to the pincher points
before trying to connect the tip wires. The connectors between the probe
head and the probe amp can be rotated to align the probe tip properly to the
pinchers. Since the center trace of the PV fixture is above the ground plane,
the probe head should be tipped slightly so the tip wires touch the center
trace and ground plane at the same time.
c Depress the actuators on the pinchers and carefully insert one wire under
the center pincher and the other wire under one of the side pinchers. Either
polarity of the probe can be tested and will yield the same results (but
80InfiniiMax III Series Probes User’s Guide
Page 81
Performance Verification6
opposite phase) if the probe is working properly. Figure 63 shows a close up
of the tip wires positioned under the pinchers.
d Ideally the probe head should not be angled toward the port 2 side of the PV
fixture, but a slight angle of 5 degrees is acceptable. If angled too much, the
measured BW of the probe will be degraded due to coupling from the trace
to the probe tip.
Figure 63Close-up of tip wires positioned under pinchers
7Install the proper file to de-embed the 1250-1749 adapter and the output side of the
N5443A (i.e. the path from the male 2.92 mm connector to the probe point of the N5443A)
from port 1 of the VNA.
a
Create a Touchstone file for the 1250-1749 and N5443A PV fixture by
cutting and pasting the text in “Touchstone File (1250-1749 & N5443A)" on
page 88. Name the file Adaptor_1250_1749___OutputSideOfFixture_N5443A.s2p.
b On the VNA, go to menu “Calibration/Fixturing Selections/2 Port
De-embedding” and select Port 1.
c Set S2P file selection to the file saved in step a.
d Check the “Enable De-embedding”.
e Under “Calibration” menu, select “Fixturing ON/off” to turn on
de-embedding.
8Trigger VNA to perform a single sweep. a. Press “Trigger” under Channel Setup, and then
the green soft-key for “Single”. Display should look like Figure 64. If it looks noticeably
different, the probe tip wires may not be making contact under the pinchers.
9Under “Trace/Math/Memory” select “Data->Memory”. This will save the de-embedded
input voltage trace into the memory.
InfiniiMax III Series Probes User’s Guide81
Page 82
6Performance Verification
Figure 64De-embedded “Vin” trace
10Now move the 2.92 mm male test port 1 connector to the 5062_1247 adapter and N5477A
Autoprobe II adapter assembly. Connect the Maury Microwave 8775B2 2.92 mm male
broadband load to the 1250_1749 adaptor and N5443A PV fixture assembly. This new
setup is shown in Figure 65. Torque all connections.
11Install the proper file to de-embed the 5062_1247 adapter and N5477A adapter from port
1 of the VNA.
a
Create a Touchstone file for the 5062-1247 and N5477A by cutting and
pasting the text in “Touchstone File (5062-1247 & N5477A)" on page 92.
Name the file Adapter_5062_1247___Adapter_N5477A.s2p.
b Go to menu “Calibration/Fixturing Selections/2 Port De-embedding” and
select Port 1.
c Set S2P file selection to the file saved in step a.
d Make sure the “Enable De-embedding” box is still checked.
e Under “Calibration” menu, make sure “Fixturing ON/off” is still checked so
file is being used for de-embedding.
82InfiniiMax III Series Probes User’s Guide
Page 83
12Trigger VNA to perform a single sweep.
a
Press “Trigger” under Channel Setup, and then the green soft-key for
“Single”.
b Under “Scale” menu, adjust the reference level until the 50 MHz point (left
side of the screen) is at center screen. Reference level should be
approximately -15.3 dB, but can vary a few tenths of a dB either way.
c Display should look like Figure 66. If it looks noticeably different, the probe
tip wires may not be making contact under the pinchers.
Performance Verification6
Figure 65Setup to measure “Vout” of probe
InfiniiMax III Series Probes User’s Guide83
Page 84
6Performance Verification
Figure 66De-embedded “Vout” trace
13Under menu “Trace/Math/Memory” select “Data/Memory” in the “Data Math” box.
This will divide the current trace (de-embedded vout trace) by the memory
a
trace (de-embedded vin trace) and therefore show the voltage transfer
function of the probe or “vout/vin”.
b Again, adjust the “Reference Level” in the scale menu so the 50MHz point is
at center screen. The display should look like Figure 67.
c Turn on a marker and adjust it to where the trace crosses 3 dB below the
50 MHz point (which is 1 division below center screen since screen is set to
3 dB/div).
d Verify that the BW is ≥ 26 GHz for the N5440A ceramic ZIF tip, N5439A ZIF
probe head, and N2803A 30 GHz probe amp combination.
84InfiniiMax III Series Probes User’s Guide
Page 85
Performance Verification6
NOTE
Figure 67De-embedded “Vout/Vin” response of the probe
You only need to perform the Bandwidth Performance Verification test with the ZIF tip as described
above. If the probe passes with the ZIF probe head then it will pass with the browser probe head as well.
You are not required to test the bandwidth specification with both probe heads. If you have the choice
between using the ZIF probe head or the browser, Keysight suggests using the ZIF probe head because
measurements are more repeatable with it and it is easier to make a proper connection.
To measure the performance of the N5445A Browser with the N2803A probe amp, the
same procedure as described above is used except the browser is held in position with a
probe positioner. This setup is shown in Figure 69 and Figure 69 using the Keysight
N2787A 3D Probe Positioner.
1Tilt the browser slightly so the two tips touch the center trace and the ground plane on the
PV fixture at about the same time.
2Probe the halfway point between the where the pinchers are and where the microstrip
transitions into the rectangular coax line.
3Set the span of the browser to the minimum possible as should in this picture:
InfiniiMax III Series Probes User’s Guide85
Page 86
6Performance Verification
Figure 68Probe Tips on Microstrip
86InfiniiMax III Series Probes User’s Guide
Page 87
Performance Verification6
Figure 69N2787A 3D Probe Positioner used to measure the N5445A browser with the
N2803A amp
InfiniiMax III Series Probes User’s Guide87
Page 88
6Performance Verification
NOTE
Touchstone File (1250-1749 & N5443A)
Cut and paste the following text and save in a text file named
Adapter_1250_1749___OutputSideOfFixture_N5443A.s2p. This file is used in Step 7 on page 81.
You’ll also find the following data on the Adobe AIR version of the Probe Resource Center (PRC).
Copying this data from the PRC is the simplest most reliable method to get the data. To download the
PRC, visit http://www.Keysight.com/find/PRC.
! freq S11 S21 S12 S22
!Port 1=female 3.5mm connector of 1250-1749 adaptor, Port 2=probe point on N5443A fixture
# Hz S DB R 50
Cut and paste the following text and save in an text file named
Adapter_5062_1247___Adapter_N5477A.s2p. This file is used in Step 11 on page 82.
You’ll also find the following data on the Adobe AIR version of the Probe Resource Center (PRC).
Copying this data from the PRC is the simplest most reliable method to get the data. To download the
PRC, visit http://www.keysight.com/find/PRC.
! freq S11 S21 S12 S22
!Port 1=female side of 5062_1247 adaptor, Port 2=male side of N5477A
# Hz S DB R 50
InfiniiMax III Probe HeadEither the N5439A ZIF probe head with the N5440A ceramic ZIF tip, the N5441A
solder-in probe head, or the N5445A browser probe head. (You only need to perform
the Performance Verification test on one of these probe heads, not all of them. If it
passes for one of them, then it will pass for all of them.)
96InfiniiMax III Series Probes User’s Guide
Page 97
Performance Verification6
Measuring Input Resistance for ZIF Probe Heads or Solder-in Probe Heads
1Connect the BNC to SMA adapter and BNC to Banana Plug adapters as shown in
Figure 70.
2Position the PV fixture on a table top and clamp it with a small bench vise to steady it.
Ensure that the PV fixture is flush with the table top so that when the banana plugs are
probed, it does not rock the PV fixture.
Figure 70Measuring the differential input resistance of a ZIF or solder-in probe head
3Connect the probe amplifier to the oscilloscope or power supply so it is powered.
4Connect the ZIF or solder-in probe head to the probe amp and insert it into the PV fixture
as shown Figure 70.
5Depress the pincher fingers on the PV fixture so they open and carefully insert the tip wires
under the pinchers. Release the pichers once the tips are inserted.
6As shown in Figure 70, measure the DC input resistance between the banana plugs. Since
one tip wire is connected to the signal line and the other tip is connected to the PV fixture
ground, this is a measurement of the differential input resistance. It should be 100 kOhms
+/- 2% (98 to 102 kOhms).
7To measure the single-ended input resistance, measure the resistance between the signal
plug of the banana adapter and the probe amplifier ground, which can be accessed as
shown in Figure 71 (through the vent window of the probe amplifier).
Figure 71Measuring the single-ended input resistance of a ZIF or solder-in probe head
InfiniiMax III Series Probes User’s Guide97
Page 98
6Performance Verification
Measuring Input Resistance for the Browser Probe Head
The following image shows the correct setup for measuring the differential input resistance
of the N5445A Browser probe head (Figure 72).
Figure 72Setup for Measuring Input Resistance for Browser Probe Head
1Set the browser for a fairly wide span (span does not impact this DC measurement).
2Plug the probe amplifier into the browser.
3Clamp the browser with the N2787A probe positioner.
4Hold the browser with one hand and loosen the arm locking knob on the probe positioner
with the other hand.
5Carefully position one tip over the signal trace on the PV fixture and the other over the
ground plane. Then let the weight of the browser slightly compress the tips so good
contact is made.
6Now the differential and single-ended input resistance can be measured in the same
manner as for the ZIF and solder-in probe head procedure above.
98InfiniiMax III Series Probes User’s Guide
Page 99
Performance Test Record
NOTE
The recommended test interval is 1 Year/2000 hours.
Keysight TechnologiesKeysight InfiniiMax III Series Probe
Model Number:Tested by:
Serial Number:Work Order Number:
Recommended next test date:Date:
Performance Verification6
Probe Head
(only required to test one)
Bandwidth Performance Test
N5439A with N5440A≥ 26 GHz
N5445A≥ 28 GHz
DC Input Resistance Performance Test
N5439A with N5440A98 to 102 kΩ (differential mode)
N5445A98 to 102 kΩ (differential mode)
N5441A98 to 102 kΩ (differential mode)
Test LimitsResultPass/Fail
49 to 51 kΩ (single-ended mode)
49 to 51 kΩ (single-ended mode)
49 to 51 kΩ (single-ended mode)
InfiniiMax III Series Probes User’s Guide99
Page 100
6Performance Verification
100InfiniiMax III Series Probes User’s Guide
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.