Keysight 1200 X, EDUX1052A/G User Manual

Page 1
Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes
User's Guide
Page 2
Notices
CAUTION
WARNING
© Keysight Technologies, Inc. 2005-2020
Manual Part Number
N2137-97015
Edition
Third edition, January 2020
Printed in Malaysia Published by:
Keysight Technologies, Inc. 1900 Garden of the Gods Road Colorado Springs, CO 80907 USA
Print History
N2137-97000, September 2018
N2137-97012, January 2019
N2137-97015, January 2020
Warranty
The material contained in this document is provided "as is," and is subject to being changed, without notice, in future editions. Further, to the maximum extent permitted by applicable law, Keysight disclaims all warranties, either express or implied, with regard to this manual and any information contained herein, including but not limited to the implied warranties of merchantability and fitness for a particular purpose. Keysight shall not be liable for errors or for incidental or consequential damages in connection with the furnishing, use, or performance of this document or of any information contained herein. Should Keysight and the user have a separate written agreement with warranty terms covering the material in this document that
conflict with these terms, the warranty terms in the separate agreement shall control.
Technology License
The hardware and/or software described in this document are furnished under a license and may be used or copied only in accordance with the terms of such license.
U.S. Government Rights
The Software is "commercial computer software," as defined by Federal Acquisition Regulation ("FAR") 2.101. Pursuant to FAR
12.212 and 27.405-3 and Department of Defense FAR Supplement ("DFARS")
227.7202, the U.S. government acquires commercial computer software under the same terms by which the software is customarily provided to the public. Accordingly, Keysight provides the Software to U.S. government customers under its standard commercial license, which is embodied in its End User License Agreement (EULA), a copy of which can be found at
www.keysight.com/find/sweula. The
license set forth in the EULA represents the exclusive authority by which the U.S. government may use, modify, distribute, or disclose the Software. The EULA and the license set forth therein, does not require or permit, among other things, that Keysight: (1) Furnish technical information related to commercial computer software or commercial computer software documentation that is not customarily provided to the public; or (2) Relinquish to, or otherwise provide, the government rights in excess of these rights customarily provided to the public to use, modify, reproduce, release, perform, display, or disclose commercial computer software or commercial computer software documentation. No additional government requirements beyond those set forth in the EULA shall apply, except to the extent that those terms, rights, or licenses are explicitly required from all providers of commercial
computer software pursuant to the FAR and the DFARS and are set forth specifically in writing elsewhere in the EULA. Keysight shall be under no obligation to update, revise or otherwise modify the Software. With respect to any technical data as defined by FAR
2.101, pursuant to FAR 12.211 and 27.404.2 and DFARS 227.7102, the U.S. government acquires no greater than Limited Rights as defined in FAR 27.401 or DFAR 227.7103-5 (c), as applicable in any technical data.
Safety Notices
A CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met.
A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice until the indicated conditions are fully understood and met.
2 Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide
Page 3

InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes— At a Glance

Figure 1 InfiniiVision 1200 X-Series 4-Channel Oscilloscope

Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide 3
Page 4

Figure 2 InfiniiVision 1200 X-Series 2-Channel Oscilloscope

Mixed Signal Oscilloscope
MSOX1204G
Digital Storage Oscilloscope
DSOX1202G
200 MHz 2 GSa/s70 MHz 2 GSa/s
InfiniiVision
Push Fine
Push Zero
Push to
Select
Force
Entry
Horizontal
Leve-l
Measure
Ver ti ca l
Intensity
Push
for Fine
Push
to Zero
Tri g g e r
Push for 50%
Push to Select
Tri g g e r
FFT
External
Math
Default
Setup
Run
Stop
Single
Wave Gen
Acquire
2
1
CursorsMeas
Utility
Analyze
Display
Save
to USB
Help
Save
Recall
Auto
Scale
Back
Quick
Action
Demo
Probe Comp
Ext Tri g
Gen Out
Inputs
CH1 and CH2
150V RMS
Input
Ext Trig
30V RMS
~
1M
~
16pF
Demo
Probe Comp
1
2
Table 1 InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes Model Numbers, Bandwidths
Model: EDUX1052A EDUX1052G DSOX1202A DSOX1202G DSOX1204A DSOX1204G
Channels: 2 4
Bandwidth: 50 MHz 70 MHz
Bandwidth upgrades: 70 MHz to 100 MHz with
D1202BW1A upgrade
70 MHz to 200 MHz with D1202BW2A upgrade
100 MHz to 200 MHz with D1202BW3A upgrade
Sampling rate: 1 GSa/s 2 GSa/s (interleaved), 1 GSa/s (non-interleaved)
Memory: 200 kpts 2 Mpts (1 Mpts when the
External Trig channel is on)
Segmented memory: No Yes
4 Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide
70 MHz to 100 MHz with D1200BW1A upgrade
70 MHz to 200 MHz with D1200BW2A upgrade
100 MHz to 200 MHz with D1200BW3A upgrade
2 Mpts (interleaved), 1 Mpts (non-interleaved)
Page 5
Table 1 InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes Model Numbers, Bandwidths (continued)
Model: EDUX1052A EDUX1052G DSOX1202A DSOX1202G DSOX1204A DSOX1204G
Waveform generator: No Yes (20 MHz) No Yes (20 MHz) No Yes (20 MHz)
Mask/limit test: No Yes
The Keysight InfiniiVision 1200 X-Series and EDUX1052A/G oscilloscopes deliver these features:
7 inch WVGA display.
200,000 waveforms/second update rate.
All knobs are pushable for making quick selections.
Trigger types: edge, pulse width, and video on EDUX1000-Series models. DSOX1200-Series models add: pattern, rise/fall time, and setup and hold.
Serial decode/trigger options for: I models. DSOX1200-Series models add: CAN, LIN, and SPI.
Math waveforms: add, subtract, multiply, divide, FFT (magnitude and phase), and low-pass filter.
Reference waveforms (2) for comparing with other channel or math waveforms.
Many built-in measurements.
G-suffix models have built-in waveform generator with: sine, square, ramp, pulse, DC, noise.
USB/LAN ports makes printing, saving, and sharing data easy.
A Quick Help system is built into the oscilloscope. Press and hold any key to display Quick Help.
2
C and UART/RS232 on EDUX1000-Series
For more information about InfiniiVision oscilloscopes, see:
www.keysight.com/find/scope
Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide 5
Page 6

In This Guide

This guide shows how to use the InfiniiVision 1200 X-Series and EDUX1052A/G oscilloscopes.
When unpacking and using the oscilloscope for the first time, see:
When displaying waveforms and acquired data, see:
When setting up triggers or changing how data is acquired, see:
Making measurements and analyzing data:
Chapter 1, “Getting Started,” starting on page 21
Chapter 2, “Horizontal Controls,” starting on page 37
Chapter 3, “Vertical Controls,” starting on page 47
Chapter 4, “Analog Bus Display,” starting on page 55
Chapter 5, “FFT Spectral Analysis,” starting on page 57
Chapter 6, “Math Waveforms,” starting on page 65
Chapter 7, “Reference Waveforms,” starting on page
77
Chapter 8, “Serial Bus Decode/Trigger,” starting on
page 81
Chapter 9, “Display Settings,” starting on page 83
Chapter 10, “Labels,” starting on page 91
Chapter 11, “Triggers,” starting on page 97
Chapter 12, “Trigger Mode/Coupling,” starting on
page 121
Chapter 13, “Acquisition Control,” starting on page
129
Chapter 14, “Cursors,” starting on page 145
Chapter 15, “Measurements,” starting on page 155
Chapter 16, “Mask Testing,” starting on page 179
Chapter 17, “Digital Voltmeter,” starting on page 193
Chapter 18, “Frequency Response Analysis,” starting
on page 197
When using the built-in waveform generator, see:
When saving, recalling, or printing, see:
Chapter 19, “Waveform Generator,” starting on page
203
Chapter 20, “Save/Recall (Setups, Screens, Data),”
starting on page 215
Chapter 21, “Print (Screens),” starting on page 227
6 Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide
Page 7
When using the oscilloscope's utility
NOTE
functions, see:
For reference information, see: Chapter 24, “Reference,” starting on page 261
Chapter 22, “Utility Settings,” starting on page 231
Chapter 23, “Web Interface,” starting on page 249
When using serial bus triggering and decode features, see:
Chapter 25, “CAN Triggering and Serial Decode,”
starting on page 275
Chapter 26, “I2C Triggering and Serial Decode,”
starting on page 285
Chapter 27, “LIN Triggering and Serial Decode,”
starting on page 293
Chapter 28, “SPI Triggering and Serial Decode,”
starting on page 301
Chapter 29, “UART/RS232 Triggering and Serial
Decode,” starting on page 313
Abbreviated instructions for pressing a series of keys and softkeys
Instructions for pressing a series of keys are written in an abbreviated manner. Instructions for pressing [Key1], then pressing Softkey2, then pressing Softkey3 are abbreviated as follows:
Press [Key1]> Softkey2 > Softkey3.
The keys may be a front panel [Key] or a Softkey. Softkeys are the six keys located directly below the oscilloscope display.
Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide 7
Page 8
8 Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide
Page 9

Contents

InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes—At a
Glance / 3
In This Guide / 6
1 Getting Started
Inspect the Package Contents / 21
Power-On the Oscilloscope / 22
Connect Probes to the Oscilloscope / 23
Maximum input voltage at analog inputs / 23
Do not float the oscilloscope chassis / 23
Input a Waveform / 23
Recall the Default Oscilloscope Setup / 24
Use Autoscale / 24
Compensate Passive Probes / 25
Learn the Front Panel Controls and Connectors / 26
Front Panel Overlays for Different Languages / 33
Learn the Rear Panel Connectors / 33
Learn the Oscilloscope Display / 34
Access the Built-In Quick Help / 36
Select the User Interface Language / 36
Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide 9
Page 10
2 Horizontal Controls
To adjust the horizontal (time/div) scale / 38
To adjust the horizontal delay (position) / 39
Panning and Zooming Single or Stopped Acquisitions / 40
To change the horizontal time mode (Normal, XY, or Roll) / 40
To display the zoomed time base / 44
To change the horizontal scale knob's coarse/fine adjustment
To position the time reference (left, center, right) / 46
3 Vertical Controls
To turn waveforms on or off (channel or math) / 49
To adjust the vertical scale / 49
To adjust the vertical position / 49
To specify channel coupling / 50
XY Time Mode / 41
setting / 45
To specify bandwidth limiting / 50
To change the vertical scale knob's coarse/fine adjustment
setting / 51
To invert a waveform / 51
Setting Analog Channel Probe Options / 51
To specify the channel units / 52 To specify the probe attenuation / 52 To specify the probe skew / 53
10 Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide
Page 11
4 Analog Bus Display
5 FFT Spectral Analysis
FFT Measurement Hints / 60
FFT Units / 61
FFT DC Value / 61
FFT Aliasing / 62
FFT Spectral Leakage / 63
6Math Waveforms
To display math waveforms / 66
To perform a math function on an arithmetic operation / 67
To adjust the math waveform scale and offset / 67
Units for Math Waveforms / 67
Math Operators / 68
Add or Subtract / 68 Multiply or Divide / 69
Math Transforms / 70
FFT Magnitude, FFT Phase / 70
Math Filters / 74
Low Pass Filter / 74
7 Reference Waveforms
To save a waveform to a reference waveform location / 77
To display a reference waveform / 78
To scale and position reference waveforms / 79
To adjust reference waveform skew / 79
To display reference waveform information / 80
Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide 11
Page 12
To save/recall reference waveform files to/from a USB storage
device / 80
8 Serial Bus Decode/Trigger
9 Display Settings
To adjust waveform intensity / 83
To set or clear persistence / 85
To clear the display / 86
To select the grid type / 86
To adjust the grid intensity / 87
To add an annotation / 87
To freeze the display / 89
10 Labels
To turn the label display on or off / 91
To assign a predefined label to a channel / 92
To define a new label / 93
To load a list of labels from a text file you create / 94
To reset the label library to the factory default / 95
11 Triggers
Adjusting the Trigger Level / 99
Forcing a Trigger / 99
Edge Trigger / 99
Pulse Width Trigger / 102
Pattern Trigger / 104
Rise/Fall Time Trigger / 106
12 Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide
Page 13
Setup and Hold Trigger / 107
Video Trigger / 108
To trigger on a specific line of video / 112 To trigger on all sync pulses / 113 To trigger on a specific field of the video signal / 114 To trigger on all fields of the video signal / 115 To trigger on odd or even fields / 116
Serial Trigger / 118
12 Trigger Mode/Coupling
To select the Auto or Normal trigger mode / 122
To select the trigger coupling / 124
To enable or disable trigger noise rejection / 125
To enable or disable trigger HF Reject / 125
To set the trigger holdoff / 126
External Trigger Input / 127
13 Acquisition Control
Running, Stopping, and Making Single Acquisitions (Run
Control) / 129
Overview of Sampling / 130
Sampling Theory / 131 Aliasing / 131 Oscilloscope Bandwidth and Sample Rate / 131 Oscilloscope Rise Time / 133 Oscilloscope Bandwidth Required / 134 Memory Depth and Sample Rate / 134
Selecting the Acquisition Mode / 135
Normal Acquisition Mode / 136 Peak Detect Acquisition Mode / 136
Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide 13
Page 14
14 Cursors
15 Measurements
Averaging Acquisition Mode / 138 High Resolution Acquisition Mode / 140
Acquiring to Segmented Memory / 141
Navigating Segments / 142 Infinite Persistence with Segmented Memory / 142 Segmented Memory Re-Arm Time / 142 Saving Data from Segmented Memory / 143
To make cursor measurements / 146
Cursor Examples / 150
To make automatic measurements / 156
Measurements Summary / 157
Snapshot All / 159
Voltage Measurements / 160
Peak-Peak / 160 Maximum / 160 Minimum / 161 Amplitude / 161 Top / 161 Base / 162 Overshoot / 162 Preshoot / 163 Average / 164 DC RMS / 164 AC RMS / 165
Time Measurements / 167
Period / 167 Frequency / 168
14 Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide
Page 15
Counter / 169 + Width / 169 – Width / 169 Bit Rate / 169 Duty Cycle / 170 Rise Time / 170 Fall Time / 170 Delay / 170 Phase / 171 X at Min Y / 173 X at Max Y / 173
Count Measurements / 174
Positive Pulse Count / 174 Negative Pulse Count / 174 Rising Edge Count / 175 Falling Edges Count / 175
Measurement Thresholds / 175
Measurement Window with Zoom Display / 177
16 Mask Testing
To create a mask from a "golden" waveform (Automask) / 179
Mask Test Setup Options / 182
Mask Statistics / 184
To manually modify a mask file / 185
Building a Mask File / 188
How is mask testing done? / 191
17 Digital Voltmeter
18 Frequency Response Analysis
To make connections / 197
Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide 15
Page 16
To set up and run the analysis / 198
To view and save the analysis results / 200
19 Waveform Generator
To select generated waveform types and settings / 203
To specify the expected output load / 206
To use waveform generator logic presets / 207
To add noise to the waveform generator output / 207
To add modulation to the waveform generator output / 208
To set up Amplitude Modulation (AM) / 209 To set up Frequency Modulation (FM) / 210 To set up Frequency-Shift Keying Modulation (FSK) / 212
To restore waveform generator defaults / 213
20 Save/Recall (Setups, Screens, Data)
Saving Setups, Screen Images, or Data / 215
To save setup files / 217 To save BMP or PNG image files / 217 To save CSV, ASCII XY, or BIN data files / 218 Length Control / 219 To save Lister data files / 220 To save reference waveform files to a USB storage device / 220 To save masks / 221 To navigate storage locations / 221 To enter file names / 222
Recalling Setups, Masks, or Reference Waveforms / 222
To recall setup files / 223 To recall mask files / 223 To recall reference waveform files from a USB storage
device / 224
16 Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide
Page 17
21 Print (Screens)
22 Utility Settings
Recalling Default Setups / 224
Performing a Secure Erase / 224
To print the oscilloscope's display / 227
To set up network printer connections / 229
To specify the print options / 230
To specify the palette option / 230
I/O Interface Settings / 231
Setting up the Oscilloscope's LAN Connection / 232
To establish a LAN connection / 233 Stand-alone (Point-to-Point) Connection to a PC / 234
File Explorer / 234
Setting Oscilloscope Preferences / 236
To choose "expand about" center or ground / 237 To disable/enable transparent backgrounds / 237 To load the default label library / 237 To set up the screen saver / 237 To set Autoscale preferences / 239
Setting the Oscilloscope's Clock / 240
Setting the Gen Out Source / 240
Enabling Remote Command Logging / 241
Performing Service Tasks / 243
To perform user calibration / 243 To perform hardware self test / 244 To perform front panel self test / 244 To export crash log files / 244
Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide 17
Page 18
23 Web Interface
To display oscilloscope information / 245 To display the user calibration status / 245 To clean the oscilloscope / 245 To check warranty and extended services status / 245 To contact Keysight / 246 To return the instrument / 246
Configuring the [Quick Action] Key / 246
Accessing the Web Interface / 250
Control Instrument / 251
Remote Front Panel / 252 Remote Programming via the Web Interface / 252 Remote Programming with Keysight IO Libraries / 254
Get Image / 254
Save/Recall / 255
Saving Files via the Web Interface / 255 Recalling Files via the Web Interface / 256
Identification Function / 257
Instrument Utilities / 258
Setting a Password / 259
24 Reference
Specifications and Characteristics / 261
Measurement Category / 261
Oscilloscope Measurement Category / 261 Measurement Category Definitions / 262 Maximum Input Voltages / 262
Maximum input voltage at analog inputs / 262
18 Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide
Page 19
Environmental Conditions / 263
Declaration of Conformity / 263
Probes and Accessories / 263
Software and Firmware Updates / 264
Binary Data (.bin) Format / 264
Binary Data in MATLAB / 265 Binary Header Format / 265 Example Program for Reading Binary Data / 268 Examples of Binary Files / 268
CSV and ASCII XY files / 271
CSV and ASCII XY file structure / 272 Minimum and Maximum Values in CSV Files / 272
Acknowledgements / 273
Product Markings and Regulatory Information / 273
25 CAN Triggering and Serial Decode
Setup for CAN Signals / 275
CAN Triggering / 277
CAN Serial Decode / 279
Interpreting CAN Decode / 280 CAN Totalizer / 281 Interpreting CAN Lister Data / 282
26 I2C Triggering and Serial Decode
Setup for I2C Signals / 285
I2C Triggering / 286
I2C Serial Decode / 290
Interpreting I2C Decode / 291 Interpreting I2C Lister Data / 292
Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide 19
Page 20
27 LIN Triggering and Serial Decode
Setup for LIN Signals / 293
LIN Triggering / 295
LIN Serial Decode / 297
Interpreting LIN Decode / 299 Interpreting LIN Lister Data / 300
28 SPI Triggering and Serial Decode
Setup for SPI Signals / 301
SPI Triggering / 307
SPI Serial Decode / 309
Interpreting SPI Decode / 311 Interpreting SPI Lister Data / 312
29 UART/RS232 Triggering and Serial Decode
Setup for UART/RS232 Signals / 313
UART/RS232 Triggering / 315
UART/RS232 Serial Decode / 318
Interpreting UART/RS232 Decode / 320 UART/RS232 Totalizer / 321 Interpreting UART/RS232 Lister Data / 322
Index
20 Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide
Page 21
Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes
User's Guide

1 Getting Started

Inspect the Package Contents / 21 Power-On the Oscilloscope / 22 Connect Probes to the Oscilloscope / 23 Input a Waveform / 23 Recall the Default Oscilloscope Setup / 24 Use Autoscale / 24 Compensate Passive Probes / 25 Learn the Front Panel Controls and Connectors / 26 Learn the Rear Panel Connectors / 33 Learn the Oscilloscope Display / 34 Access the Built-In Quick Help / 36 Select the User Interface Language / 36
This chapter describes the steps you take when using the oscilloscope for the first time.

Inspect the Package Contents

• Inspect the shipping container for damage.
If your shipping container appears to be damaged, keep the shipping container or cushioning material until you have inspected the contents of the shipment for completeness and have checked the oscilloscope mechanically and electrically.
• Verify that you received the following items and any optional accessories you
may have ordered:
21
Page 22
1 Getting Started
WARNING
InfiniiVision 1200 X-Series or EDUX1052A/G oscilloscope.
Power cord (country of origin determines specific type).
Oscilloscope probes (one for each analog input channel).

Power-On the Oscilloscope

Requirements
Requirements
To po wer-on the
oscilloscope
Power
Ventilation
Line voltage, frequency, and power:
• ~Line 100-120 Vac, 50/60/400 Hz
• 100-240 Vac, 50/60 Hz
•50W max
The air intake and exhaust areas must be free from obstructions. Unrestricted air flow is required for proper cooling. Always ensure that the air intake and exhaust areas are free from obstructions.
The fan draws air in from the left side and bottom of the oscilloscope and pushes it out behind the oscilloscope.
When using the oscilloscope in a bench-top setting, provide at least 2" clearance at the sides and 4" (100 mm) clearance above and behind the oscilloscope for proper cooling.
1 Connect the power cord to the rear of the oscilloscope, then to a suitable AC
voltage source. Route the power cord so the oscilloscope's feet and legs do not pinch the cord.
2 The oscilloscope automatically adjusts for input line voltages in the range 100
to 240 VAC. The provided line cord is matched to the country of origin.
Always use a grounded power cord. Do not defeat the power cord ground.
22 Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide
3 Press the power switch.
The power switch is located on the lower left corner of the front panel. The oscilloscope will perform a self-test and will be operational in a few seconds.
Page 23

Connect Probes to the Oscilloscope

CAUTION
CAUTION
WARNING
1 Connect the oscilloscope probe to an oscilloscope input channel BNC
connector.
2 Connect the probe's retractable hook tip to the point of interest on the circuit or
device under test. Be sure to connect the probe ground lead to a ground point on the circuit.
Maximum input voltage at analog inputs
150 Vrms, 200 Vpk
Do not float the oscilloscope chassis
Defeating the ground connection and "floating" the oscilloscope chassis will probably result in inaccurate measurements and may also cause equipment damage. The probe ground lead is connected to the oscilloscope chassis and the ground wire in the power cord. If you need to measure between two live points, use a differential probe with sufficient dynamic range.
Getting Started 1

Input a Waveform

Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide 23
Do not negate the protective action of the ground connection to the oscilloscope. The oscilloscope must remain grounded through its power cord. Defeating the ground creates an electric shock hazard.
The Probe Comp signal is used for compensating probes. 1 Connect an oscilloscope probe from channel 1 to the Demo, Probe Comp terminal
on the front panel.
2 Connect the probe's ground lead to the ground terminal (next to the Demo
terminal).
Page 24
1 Getting Started
WARNING
A voltage source should never be connected to the ground terminal of this instrument. If, for any reason, the Protective Conductor Terminal is disconnected or not functioning properly and a voltage source is connected to the equipment's ground terminals, the entire chassis will be at the voltage potential of the voltage source, and the operator or bystanders could receive an electric shock.

Recall the Default Oscilloscope Setup

To recall the default oscilloscope setup: 1 Press [Default Setup].
The default setup restores the oscilloscope's default settings. This places the oscilloscope in a known operating condition.
In the Save/Recall menu, there are also options for restoring the complete factory settings or performing a secure erase (see Chapter 20, “Save/Recall (Setups, Screens, Data),” starting on page 215).

Use Autoscale

Use [Auto Scale] to automatically configure the oscilloscope to best display the input signals.
1 Press [Auto Scale].
You should see a waveform on the oscilloscope's display similar to this:
24 Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide
Page 25
Getting Started 1
2 If you want to return to the oscilloscope settings that existed before, press Undo
Autoscale.
3 If you want to enable "fast debug" autoscaling, change the channels
autoscaled, or preserve the acquisition mode during autoscale, press Fast Debug, Channels, or Acq Mode.
These are the same softkeys that appear in the Autoscale Preferences menu. See Chapter 22, “Utility Settings,” starting on page 231.
If you see the waveform, but the square wave is not shaped correctly as shown above, perform the procedure "Compensate Passive Probes" on page 25.
If you do not see the waveform, make sure the probe is connected securely to the front panel channel input BNC and to the Demo/Probe Comp terminal.

Compensate Passive Probes

Each oscilloscope passive probe must be compensated to match the input characteristics of the oscilloscope channel to which it is connected. A poorly compensated probe can introduce significant measurement errors.
Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide 25
Page 26
1 Getting Started
NOTE
Perfectly compensated
Over compensated
Under compensated
If your probe has a configurable attenuation setting (like the N2140/42A probes do), the 10:1 setting must be used for probe compensation.
1 Input the Probe Comp signal (see "Input a Waveform" on page 23). 2 Press [Default Setup] to recall the default oscilloscope setup (see "Recall the
Default Oscilloscope Setup" on page 24).
3 Press [Auto Scale] to automatically configure the oscilloscope for the Probe
Comp signal (see "Use Autoscale" on page 24).
4 Press the channel key to which the probe is connected ([1], [2], etc.). 5 In the Channel menu, press Probe. 6 In the Channel Probe menu, press Probe Check; then, follow the instructions
on-screen.
If necessary, use a nonmetallic tool (supplied with the probe) to adjust the trimmer capacitor on the probe for the flattest pulse possible.
On some probes (like the N2140/42A probes), the trimmer capacitor is located on the probe BNC connector. On other probes (like the N2862/63/90 probes), the trimmer capacitor is a yellow adjustment on the probe tip.

Learn the Front Panel Controls and Connectors

26 Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide
7 Connect probes to all other oscilloscope channels. 8 Repeat the procedure for each channel.
On the front panel, key refers to any key (button) you can press.
Page 27
Getting Started 1
6. [Auto Scale] key5. [Default Setup] key
10. Tools keys
1. Power switch
2. Softkeys
3. [Intensity] key
4. Entry knob
11. Trigger controls
12. Vertical controls
19. Demo/Probe Comp and Ground
terminals
15. [Help] key20. USB
Host
port
13. [Wave Gen] key
8. Run Control keys
9. Measure controls
7. Horizontal and Acquisition controls
16. Math function controls
18. Waveform generator
output
17. Analog channel
inputs
14. [FFT] key
Softkey specifically refers to the six keys next to the display. Menus and softkey labels appear on the display when other front panel keys are pressed. Softkey functions change as you navigate through the oscilloscope's menus.
For the following figures, refer to the numbered descriptions in the table that follows.

Figure 3 InfiniiVision 1200 X-Series 4-Channel Oscilloscopes

Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide 27
Page 28
1 Getting Started
Mixed Signal Oscilloscope
MSOX1204G
Digital Storage Oscilloscope
DSOX1202G
200 MHz 2 GSa/s70 MHz 2 GSa/s
InfiniiVision
Push Fine
Push Zero
Push to Select
Force
Entry
Horizontal
Leve-l
Measure
Vert ic al
Intensity
Push
for Fine
Push
to Zero
Trigger
Push for 50%
Push to Select
Trig g e r
FFT
External
Math
Default Setup
Run
Stop
Single
Wave Gen
Acquire
2
1
CursorsMeas
Utility
Analyze
Display
Save
to USB
Help
Save Recall
Auto Scale
Back
Quick Action
Demo
Probe Comp
Ext Tri g
Gen Out
Inputs
CH1 and CH2
150V RMS
Input
Ext Trig
30V RMS
~
1M
~
16pF
Demo
Probe Comp
1
2
22. Ext Trig
external trigger
input
21. [External] key
Back
Back

Figure 4 InfiniiVision 1200 X-Series 2-Channel Oscilloscopes

1. Power switch Press once to switch power on; press again to switch power off. See "Power-On the
Oscilloscope" on page 22.
2. Softkeys The functions of these keys change based upon the menus shown on the display next to the keys.
The Back key moves back in the softkey menu hierarchy. At the top of the hierarchy, the
Back key turns the menus off, and oscilloscope information is shown instead.
3. [Intensity] key Press the key to illuminate it. When illuminated, turn the Entry knob to adjust waveform intensity.
You can vary the intensity control to bring out signal detail, much like an analog oscilloscope.
4. Entry knob The Entry knob is used to select items from menus and to change values. The function of the Entry knob changes based upon the current menu and softkey selections.
Note that when the Entry knob symbol appears on a softkey, you can use the Entry knob, to select values.
Often, rotating the Entry knob is enough to make a selection. Sometimes, you can push the Entry knob to enable or disable a selection. Also, pushing the Entry knob can also make popup menus disappear.
28 Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide
Page 29
Getting Started 1
5. [Default Setup]
key
6. [Auto Scale]
key
7. Horizontal and
Acquisition controls
Press this key to restore the oscilloscope's default settings (details on "Recall the Default
Oscilloscope Setup" on page 24).
When you press the [Auto Scale] key, the oscilloscope will quickly determine which channels have activity, and it will turn these channels on and scale them to display the input signals. See "Use
Autoscale" on page 24.
The Horizontal and Acquisition controls consist of:
Horizontal scale knob — Turn the knob in the Horizontal section that is marked to
adjust the time/div setting. The symbols under the knob indicate that this control has the effect of spreading out or zooming in on the waveform using the horizontal scale.
Push the horizontal scale knob to toggle between fine and coarse adjustment.
Horizontal position knob — Turn the knob marked to pan through the waveform data
horizontally. You can see the captured waveform before the trigger (turn the knob clockwise) or after the trigger (turn the knob counterclockwise). If you pan through the waveform when the oscilloscope is stopped (not in Run mode) then you are looking at the waveform data from the last acquisition taken.
•[Acquire] key — Press this key to open the Acquire menu where you can select the Normal, XY, and
Roll time modes, enable or disable Zoom, and select the trigger time reference point.
Also you can select the Normal, Peak Detect, Averaging, or High Resolution acquisition modes and, on DSOX1200-Series models, use segmented memory (see "Selecting the Acquisition
Mode" on page 135).
Zoom key — Press the zoom key to split the oscilloscope display into Normal and Zoom
sections without opening the Acquire menu.
For more information see Chapter 2, “Horizontal Controls,” starting on page 37.
8. Run Control
keys
When the [Run/Stop] key is green, the oscilloscope is running, that is, acquiring data when trigger conditions are met. To stop acquiring data, press [Run/Stop].
When the [Run/Stop] key is red, data acquisition is stopped. To start acquiring data, press [Run/Stop].
To capture and display a single acquisition (whether the oscilloscope is running or stopped), press [Single]. The [Single] key is yellow until the oscilloscope triggers.
For more information, see "Running, Stopping, and Making Single Acquisitions (Run
Control)" on page 129.
Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide 29
Page 30
1 Getting Started
9. Measure
controls
The measure controls consist of:
•[Analyze] key — Press this key to access analysis features like:
Trigger level setting.
Measurement threshold setting.
Video trigger automatic set up and display.
Display a bus made up of the analog channel inputs where channel 1 is the least significant bit
and channel 4 is the most significant bit. See also Chapter 4, “Analog Bus Display,” starting on page 55.
Enable serial bus decodes. See also Chapter 8, “Serial Bus Decode/Trigger,” starting on page
81.
Reference waveforms (see Chapter 7, “Reference Waveforms,” starting on page 77).
Mask testing (see Chapter 16, “Mask Testing,” starting on page 179).
Digital voltmeter (see Chapter 17, “Digital Voltmeter,” starting on page 193).
Frequency response analysis on models with a built-in waveform generator (see Chapter 18,
“Frequency Response Analysis,” starting on page 197).
•[Meas] key — Press this key to access a set of predefined measurements. See Chapter 15,
“Measurements,” starting on page 155.
•[Cursors] key — Press this key to open a menu that lets you select the cursors mode and source.
Cursors knob — Push this knob select cursors from a popup menu. Then, after the popup menu
closes (either by timeout or by pushing the knob again), rotate the knob to adjust the selected cursor position.
10. Tools keys The Tools keys consist of:
•[Save/Recall] key — Press this key to save oscilloscope setups, screen images, waveform data, or
mask files or to recall setups, mask files or reference waveforms. See Chapter 20, “Save/Recall (Setups, Screens, Data),” starting on page 215.
• [Utility] key — Press this key to access the Utility menu, which lets you configure the
oscilloscope's I/O settings, use the file explorer, set preferences, access the service menu, and choose other options. See Chapter 22, “Utility Settings,” starting on page 231.
• [Display] key — Press this key to access the menu where you can enable persistence, adjust the
display grid (graticule) intensity, label waveforms, add an annotation, and clear the display (see
Chapter 9, “Display Settings,” starting on page 83).
• [Quick Action] key — Press this key to perform the selected quick action: measure all snapshot,
print, save, recall, freeze display. and more. See "Configuring the [Quick Action] Key" on page 246.
•[Save to USB] key — Press this key to perform a quick save to a USB storage device.
30 Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide
Page 31
Getting Started 1
T
11. Trigger controls The Trigger controls determine how the oscilloscope triggers to capture data. These controls consist of:
Level knob — Turn the Level knob to adjust the trigger level for a selected analog channel.
Push the knob to set the level to the waveform's 50% value. If AC coupling is used, pushing the Level knob sets the trigger level to about 0 V.
The position of the trigger level for the analog channel is indicated by the trigger level icon (if the analog channel is on) at the far left side of the display. The value of the analog channel trigger level is displayed in the upper-right corner of the display.
• [Trigger] key — Press this key to select the trigger type (edge, pulse width, video, etc.). See
Chapter 11, “Triggers,” starting on page 97. You can also set options that affect all trigger types.
See Chapter 12, “Trigger Mode/Coupling,” starting on page 121.
•[Force] key — Causes a trigger (on anything) and displays the acquisition.
This key is useful in the Normal trigger mode where acquisitions are made only when the trigger condition is met. In this mode, if no triggers are occurring (that is, the "Trig'd?" indicator is displayed), you can press [Force] to force a trigger and see what the input signals look like.
12. Vertical
controls
13. [Wave Gen] key On G-suffix models that have a built-in waveform generator, press this key to access waveform
14. [FFT] key Provides access to FFT spectrum analysis function. See Chapter 5, “FFT Spectral Analysis,” starting
The Vertical controls consist of:
Analog channel on/off keys — Use these keys to switch a channel on or off, or to access a
channel's menu in the softkeys. There is one channel on/off key for each analog channel.
Vertical scale knob — Use the knob marked to change the vertical sensitivity (gain) for
the selected analog input channel.
Push the vertical scale knob to toggle between fine and coarse adjustment.
The default mode for expanding the signal is about the ground level of the channel; however, you can change this to expand about the center of the display.
Vertical position knob — Use this knob to change the selected analog input channel waveform
vertical position on the display.
The voltage value momentarily displayed in the upper right portion of the display represents the
voltage difference between the vertical center of the display and the ground level ( ) icon. It also represents the voltage at the vertical center of the display if vertical expansion is set to expand about ground.
For more information, see Chapter 3, “Vertical Controls,” starting on page 47.
generator functions. See Chapter 19, “Waveform Generator,” starting on page 203.
on page 57.
Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide 31
Page 32
1 Getting Started
15. [Help] key Opens the Help menu where you can display overview help topics, select the Language (see also
"Access the Built-In Quick Help" on page 36), and select training signals that can be output on
the Demo terminal.
16. Math function
controls
17. Analog channel
inputs
18. Waveform
generator output
19. Demo/Probe
Comp, Ground terminals
The Math function controls consist of:
•[Math] key — provides access to math (add, subtract, etc.) waveform functions. See Chapter 6,
“Math Waveforms,” starting on page 65.
Vertical scale knob — Use the knob marked to change the vertical sensitivity (as with
the analog channel vertical controls).
Vertical position knob — Use the knob to change a math function waveform's vertical position on
the display (as with the analog channel vertical controls).
Attach oscilloscope probes or BNC cables to these BNC connectors.
In the InfiniiVision 1200 X-Series oscilloscopes, the analog channel inputs have 1 M impedance.
Also, there is no automatic probe detection, so you must properly set the probe attenuation for accurate measurement results. See "Setting Analog Channel Probe Options" on page 51.
On G-suffix models, the built-in waveform generator can output sine, square, ramp, pulse, DC, or noise on the Gen Out BNC. Press the [Wave Gen] key to set up the waveform generator. See
Chapter 19, “Waveform Generator,” starting on page 203.
You can also send the trigger output signal or the mask test failure signal to the Gen Out BNC connector. See "Setting the Gen Out Source" on page 240.
Demo terminal — This terminal outputs the Probe Comp signal which helps you match a probe's
input capacitance to the oscilloscope channel to which it is connected. See "Compensate
Passive Probes" on page 25. The oscilloscope can also output demo or training signals on this
terminal.
Ground terminal — Use the ground terminal for oscilloscope probes connected to the Demo/Probe
Comp terminal. See the warning in "Input a Waveform" on page 23.
32 Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide
Page 33
Getting Started 1
20. USB Host port This port is for connecting USB mass storage devices or printers to the oscilloscope.
Connect a USB compliant mass storage device (flash drive, disk drive, etc.) to save or recall oscilloscope setup files and reference waveforms or to save data and screen images. See
Chapter 20, “Save/Recall (Setups, Screens, Data),” starting on page 215.
To print, connect a USB compliant printer. For more information about printing see Chapter 21, “Print (Screens),” starting on page 227.
You can also use the USB port to update the oscilloscope's system software when updates are available.
NOTE: You must Eject a USB mass storage device before unplugging it; otherwise, the device will be marked as needing repair when connecting to a computer with the Windows operating system (even though there are no harmful effects to the device).
CAUTION:Do not connect a host computer to the oscilloscope's USB host port. A host computer sees the oscilloscope as a device, so connect the host computer to the oscilloscope's device port (on the rear panel). See "Learn the Rear Panel Connectors" on page 33.
21. [External] key Press this key to set external trigger input options. See "External Trigger Input" on page 127.
22. Ext Trig input External trigger input BNC connector. See "External Trigger Input" on page 127 for an explanation of this feature.

Front Panel Overlays for Different Languages

Front panel overlays, which have translations for the English front panel keys and label text, are available in many languages. The appropriate overlay is included when the localization option is chosen at time of purchase.
To install a front panel overlay:
1 Gently pull on the front panel knobs to remove them. 2 Insert the overlay's side tabs into the slots on the front panel. 3 Reinstall the front panel knobs.

Learn the Rear Panel Connectors

For the following figure, refer to the numbered descriptions in the table that follows.
Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide 33
Page 34
1 Getting Started
2. Kensington lock hole
3. USB Device port
4. LAN port
5. EXT TRIG input (on 4-channel oscilloscopes)
1. Power cord connector
1. Power cord
connector
2. Kensington lock
hole
3. USB Device
port
4. LAN port When the oscilloscope is connected to a network (see "Setting up the Oscilloscope's LAN
15. EXT TRIG input On 4-channel models, the external trigger input BNC connector is on the rear panel.
Attach the power cord here.
This is where you can attach a Kensington lock for securing the instrument.
This port is for connecting the oscilloscope to a host PC. You can issue remote commands from a host PC to the oscilloscope via the USB device port (see the Programmer's Guide).
Connection" on page 232), the LAN port lets you print to network printers (see Chapter 21, “Print
(Screens),” starting on page 227), access the oscilloscope's built-in web server (see Chapter 23, “Web Interface,” starting on page 249), and issue remote commands (see the Programmer's Guide).
On 2-channel models, the external trigger input BNC connector is on the front panel.
See "External Trigger Input" on page 127 for an explanation of this feature.

Learn the Oscilloscope Display

The oscilloscope display contains acquired waveforms, setup information, measurement results, and the softkey definitions.
34 Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide
Page 35
Getting Started 1
Analog channel sensitivity
Status line
Analog channels and ground levels
Trigger level
Trigger point, time reference
Delay time
Time/ div
Run/Stop status
Trigger type
Trigger source
Measurements
Trigger level
Softkey labels and information area
Cursors defining measurement
Other waveforms

Figure 5 Interpreting the oscilloscope display

Status line The top line of the display contains vertical, horizontal, and trigger setup information.
Display area The display area contains the waveform acquisitions, channel identifiers, and analog trigger, and
ground level indicators. Each analog channel's information appears in a different color.
Signal detail is displayed using 256 levels of intensity.
For more information about display modes see Chapter 9, “Display Settings,” starting on page 83.
Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide 35
Page 36
1 Getting Started
Back
Back
Softkey labels and information area
Measurements area When measurements or cursors are turned on, this area contains automatic measurement and cursor
When most front panel keys are pressed, short menu names and softkey labels appear in this area. The labels describe the softkey functions. Typically, softkeys let you set up additional parameters for the selected mode or menu.
Pressing the Back key returns through the menu hierarchy until softkey labels are off and the information area is displayed. The information area contains acquisition, analog channel, math function, and reference waveform information.
You can also specify that softkey menus turn off automatically after a specified timeout period ([Utility] > Options > Menu Timeout).
Pressing the Back key when the information area is displayed returns to the most recent menu displayed.
results.
When measurements are turned off, this area displays additional status information describing channel offset and other configuration parameters.

Access the Built-In Quick Help

To view Quick Help: 1 Press and hold the key, softkey, or knob for which you would like to view help.
Quick Help remains on the screen until another key is pressed or a knob is turned.

Select the User Interface Language

To select the user interface and Quick Help language:
1 Press [Help], then press the Language softkey. 2 Turn the Entry knob until the desired language is selected.
36 Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide
Page 37
Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes
User's Guide

2 Horizontal Controls

To adjust the horizontal (time/div) scale / 38 To adjust the horizontal delay (position) / 39 Panning and Zooming Single or Stopped Acquisitions / 40 To change the horizontal time mode (Normal, XY, or Roll) / 40 To display the zoomed time base / 44 To change the horizontal scale knob's coarse/fine adjustment setting / 45 To position the time reference (left, center, right) / 46
The horizontal controls include:
• The horizontal scale and position knobs.
•The [Acquire] key for accessing the Acquire menu.
• The zoom key for quickly enabling/disabling the split-screen zoom display.
The following figure shows the Acquire menu which appears after pressing the [Acquire] key.
37
Page 38
2 Horizontal Controls
Trigger point
Time reference
Delay time
Time/ div
Trigger source
Trigger level or threshold
XY or Roll mode
Normal time mode
Zoomed time base
Time reference

Figure 6 Acquire Menu

The Acquire menu lets you select the time mode (Normal, XY, or Roll), enable Zoom, and specify the time reference.

To adjust the horizontal (time/div) scale

38 Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide
The current sample rate is displayed in the right side information area when softkey menu labels are off.
1 Turn the large horizontal scale (sweep speed) knob marked to
change the horizontal time/div setting.
Notice how the time/div information in the status line changes. The symbol at the top of the display indicates the time reference point.
Page 39
The horizontal scale knob works (in the Normal time mode) while acquisitions are running or when they are stopped. When running, adjusting the horizontal scale knob changes the sample rate. When stopped, adjusting the horizontal scale knob lets you zoom into acquired data. See "Panning and Zooming Single or Stopped
Acquisitions" on page 40.
Note that the horizontal scale knob has a different purpose in the Zoom display. See "To display the zoomed time base" on page 44.

To adjust the horizontal delay (position)

1 Turn the horizontal delay (position) knob ( ).
The trigger point moves horizontally, pausing at 0.00 s (mimicking a mechanical detent), and the delay value is displayed in the status line.
Changing the delay time moves the trigger point (solid inverted triangle) horizontally and indicates how far it is from the time reference point (hollow inverted triangle ). These reference points are indicated along the top of the display grid.
Figure 6 shows the trigger point with the delay time set to 200 µs. The delay time
number tells you how far the time reference point is located from the trigger point. When delay time is set to zero, the delay time indicator overlays the time reference indicator.
Horizontal Controls 2
All events displayed left of the trigger point happened before the trigger occurred. These events are called pre-trigger information, and they show events that led up to the trigger point.
Everything to the right of the trigger point is called post-trigger information. The amount of delay range (pre-trigger and post-trigger information) available depends on the time/div selected and memory depth.
The horizontal position knob works (in the Normal time mode) while acquisitions are running or when they are stopped. When running, adjusting the horizontal scale knob changes the sample rate. When stopped, adjusting the horizontal scale knob lets you zoom into acquired data. See "Panning and Zooming Single or
Stopped Acquisitions" on page 40.
Note that the horizontal position knob has a different purpose in the Zoom display. See "To display the zoomed time base" on page 44.
Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide 39
Page 40
2 Horizontal Controls
NOTE

Panning and Zooming Single or Stopped Acquisitions

When the oscilloscope is stopped, use the horizontal scale and position knobs to pan and zoom your waveform. The stopped display may contain several acquisitions worth of information, but only the last acquisition is available for pan and zoom.
The ability to pan (move horizontally) and scale (expand or compress horizontally) an acquired waveform is important because of the additional insight it can reveal about the captured waveform. This additional insight is often gained from seeing the waveform at different levels of abstraction. You may want to view both the big picture and the specific little picture details.
The ability to examine waveform detail after the waveform has been acquired is a benefit generally associated with digital oscilloscopes. Often this is simply the ability to freeze the display for the purpose of measuring with cursors or printing the screen. Some digital oscilloscopes go one step further by including the ability to further examine the signal details after acquiring them by panning through the waveform and changing the horizontal scale.
There is no limit imposed on the scaling ratio between the time/div used to acquire the data and the time/div used to view the data. There is, however, a useful limit. This useful limit is somewhat a function of the signal you are analyzing.
Zooming into stopped acquisitions
The screen will still contain a relatively good display if you zoom-in horizontally by a factor of 1000 and zoom-in vertically by a factor of 10 to display the information from where it was acquired. Remember that you can make automatic measurements on displayed data only.

To change the horizontal time mode (Normal, XY, or Roll)

1 Press [Acquire]. 2 In the Acquire menu, press Time Mode; then, select:
Normal — the normal viewing mode for the oscilloscope.
40 Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide
Page 41
Horizontal Controls 2
In the Normal time mode, signal events occurring before the trigger are plotted to the left of the trigger point () and signal events after the trigger plotted to the right of the trigger point.
XY — XY mode changes the display from a volts-versus-time display to a
volts-versus-volts display. The time base is turned off. Channel 1 amplitude is plotted on the X-axis and Channel 2 amplitude is plotted on the Y-axis.
You can use XY mode to compare frequency and phase relationships between two signals. XY mode can also be used with transducers to display strain versus displacement, flow versus pressure, volts versus current, or voltage versus frequency.
Use the cursors to make measurements on XY mode waveforms.
For more information about using XY mode for measurements, refer to "XY
Time Mode" on page 41.
Roll — causes the waveform to move slowly across the screen from right to
left. It only operates on time base settings of 50 ms/div and slower. If the current time base setting is faster than the 50 ms/div limit, it will be set to 50 ms/div when Roll mode is entered.
In Roll mode there is no trigger. The fixed reference point on the screen is the right edge of the screen and refers to the current moment in time. Events that have occurred are scrolled to the left of the reference point. Since there is no trigger, no pre-trigger information is available.
If you would like to pause the display in Roll mode press the [Single] key. To clear the display and restart an acquisition in Roll mode, press the [Single] key again.
Use Roll mode on low-frequency waveforms to yield a display much like a strip chart recorder. It allows the waveform to roll across the display.

XY Time Mode

The XY time mode converts the oscilloscope from a volts-versus-time display to a volts-versus-volts display using two input channels. Channel 1 is the X-axis input, channel 2 is the Y-axis input. You can use various transducers so the display could show strain versus displacement, flow versus pressure, volts versus current, or voltage versus frequency.
Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide 41
Page 42
2 Horizontal Controls
sinθ =Aor
C
A
B
D
C
Signal must be cent ered i n “X”
Measuring
phase difference
Signals 90 degrees
out of phase
Signals
in phase
Example This exercise shows a common use of the XY display mode by measuring the
phase difference between two signals of the same frequency with the Lissajous method.
1 Connect a sine wave signal to channel 1, and a sine wave signal of the same
frequency but out of phase to channel 2.
2 Press the [Auto Scale] key, press the [Acquire] key; then, press Time Mode and
select "XY".
3 Center the signal on the display with the channel 1 and 2 position ( ) knobs.
Use the channel 1 and 2 volts/div knobs and the channel 1 and 2 Fine softkeys to expand the signal for convenient viewing.
The phase difference angle (θ) can be calculated using the following formula (assuming the amplitude is the same on both channels):
Figure 7 XY time mode signals, centered on display
4 Press the [Cursors] key.
42 Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide
5 Set the Y2 cursor to the top of the signal, and set Y1 to the bottom of the
signal. Note the ΔY value at the bottom of the display. In this example, we are using
the Y cursors, but you could have used the X cursors instead.
Page 43
Horizontal Controls 2
sinθ =
second Δ Y
first Δ Y
=
1.031
1.688
; θ = 37.65 degrees of phase shift
NOTE
6 Move the Y1 and Y2 cursors to the intersection of the signal and the Y axis.
Again, note the ΔY value.
Figure 8 Phase difference measurements, automatic and using cursors
7 Calculate the phase difference using the formula below.
For example, if the first ΔY value is 1.688 and the second ΔY value is 1.031:
Z-Axis Input in XY Display Mode (Blanking)
When you select the XY display mode, the time base is turned off. Channel 1 is the X-axis input, channel 2 is the Y-axis input, and the EXT TRIG input is the Z-axis input. If you only want to see portions of the Y versus X display, use the Z-axis input. Z-axis turns the trace on and off (analog oscilloscopes called this Z-axis blanking because it turned the beam on and off). When Z is low (<1.4 V), Y versus X is displayed; when Z is high (>1.4 V), the trace is turned off.
Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide 43
Page 44
2 Horizontal Controls
These markers show the beginning and end of the Zoom window
Normal window
Time/div for zoomed window
Time/div for normal window
Delay time momentarily displays when the Horizontal position knob is turned
Zoom window
Signal anomaly expanded in zoom window
Select Zoom

To display the zoomed time base

Zoom, formerly called Delayed sweep mode, is a horizontally expanded version of the normal display. When Zoom is selected, the display divides in half. The top half of the display shows the normal time/div window and the bottom half displays a faster Zoom time/div window.
The Zoom window is a magnified portion of the normal time/div window. You can use Zoom to locate and horizontally expand part of the normal window for a more detailed (higher-resolution) analysis of signals.
To turn on (or off) Zoom:
1 Press the zoom key (or press the [Acquire] key and then the Zoom softkey).
44 Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide
Page 45
Horizontal Controls 2
The area of the normal display that is expanded is outlined with a box and the rest of the normal display is ghosted. The box shows the portion of the normal sweep that is expanded in the lower half.
To change the time/div for the Zoom window, turn the horizontal scale (sweep speed) knob. As you turn the knob, the zoomed window time/div is highlighted in the status line above the waveform display area. The Horizontal scale (sweep speed) knob controls the size of the box.
The Horizontal position (delay time) knob sets the left-to-right position of the zoom window. The delay value, which is the time displayed relative to the trigger point) is momentarily displayed in the upper-right portion of the display when the
delay time ( ) knob is turned.
Negative delay values indicate you're looking at a portion of the waveform before the trigger event, and positive values indicate you're looking at the waveform after the trigger event.
To change the time/div of the normal window, turn off Zoom; then, turn the horizontal scale (sweep speed) knob.
For information about using zoom mode for measurements, refer to "To isolate a
pulse for Top measurement" on page 161 and "To isolate an event for frequency measurement" on page 168.

To change the horizontal scale knob's coarse/fine adjustment setting

1 Push the horizontal scale knob to toggle between fine and coarse adjustment
of the horizontal scale.
When Fine is enabled, turning the horizontal scale knob changes the time/div (displayed in the status line at the top of the display) in smaller increments. The time/div remains fully calibrated when Fine is on.
When Fine is turned off, the Horizontal scale knob changes the time/div setting in a 1-2-5 step sequence.
Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide 45
Page 46
2 Horizontal Controls

To position the time reference (left, center, right)

Time reference is the reference point on the display for delay time (horizontal position).
1 Press [Acquire]. 2 In the Acquire menu, press Time Ref; then, select:
Left — the time reference is set to one major division from the left edge of the
display.
Center — the time reference is set to the center of the display.
Right — the time reference is set to one major division from the right edge of
the display.
A small hollow triangle () at the top of the display grid marks the position of the time reference. When delay time is set to zero, the trigger point indicator (▼) overlays the time reference indicator.
The time reference position sets the initial position of the trigger event within acquisition memory and on the display, with delay set to 0.
46 Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide
Turning the Horizontal scale (sweep speed) knob expands or contracts the waveform about the time reference point (). See "To adjust the horizontal
(time/div) scale" on page 38.
Turning the Horizontal position ( ) knob in Normal mode (not Zoom) moves the trigger point indicator () to the left or right of the time reference point (). See
"To adjust the horizontal delay (position)" on page 39.
Page 47
Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes
User's Guide

3 Vertical Controls

To turn waveforms on or off (channel or math) / 49 To adjust the vertical scale / 49 To adjust the vertical position / 49 To specify channel coupling / 50 To specify bandwidth limiting / 50 To change the vertical scale knob's coarse/fine adjustment setting / 51 To invert a waveform / 51 Setting Analog Channel Probe Options / 51
The vertical controls include:
• Multiplexed vertical scale and position knobs for the selected analog channel.
• The channel keys for turning a channel on or off and accessing the channel's softkey menu.
47
Page 48
3 Vertical Controls
NOTE
NOTE
Channel, Volts/div
Channel 1 ground level
Trigger source
Trigger level or threshold
Channel 2 ground level
Keysight recommends always scaling the signal so that the entire waveform is contained between the top and bottom of the display.
For proper operation of the 1200 X-Series oscilloscope, the channel inputs must not be overdriven more than ±8 divisions. Exceeding this limit may result in signals that appear incorrect and may increase crosstalk between the input channels.
To minimize crosstalk between input channels, make sure the channel is not overdriven. Also, connecting a probe or cable to a channel will reduce crosstalk.
The following figure shows the Channel 1 menu that appears after pressing the [1] channel key.
The ground level of the signal for each displayed analog channel is identified by the position of the icon at the far-left side of the display.
48 Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide
Page 49

To turn waveforms on or off (channel or math)

NOTE
1 Press an analog channel key turn the channel on or off (and to display the
channel's menu).
When a channel is on, its key is illuminated.
Turning channels off
You must be viewing the menu for a channel before you can turn it off. For example, if channel 1 and channel 2 are turned on and the menu for channel 2 is being displayed, to turn channel 1 off, press [1] to display the channel 1 menu; then, press [1] again to turn channel 1 off.

To adjust the vertical scale

1 Turn the large knob above the channel key marked to set the vertical
scale (volts/division) for the channel.
The vertical scale knob changes the analog channel scale in a 1-2-5 step sequence (with a 1:1 probe attached) unless fine adjustment is enabled (see "To
change the vertical scale knob's coarse/fine adjustment setting" on page 51).
Vertical Controls 3
The analog channel Volts/Div value is displayed in the status line.
The default mode for expanding the signal when you turn the volts/division knob is vertical expansion about the ground level of the channel; however, you can change this to expand about the center of the display. See "To choose "expand
about" center or ground" on page 237.

To adjust the vertical position

1 Turn the small vertical position knob ( ) to move the channel's waveform up or
down on the display.
Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide 49
Page 50
3 Vertical Controls
TIP
The voltage value momentarily displayed in the upper right portion of the display represents the voltage difference between the vertical center of the display and
the ground level ( ) icon. It also represents the voltage at the vertical center of the display if vertical expansion is set to expand about ground (see "To choose
"expand about" center or ground" on page 237).

To specify channel coupling

Coupling changes the channel's input coupling to either AC (alternating current) or DC (direct current).
If the channel is DC coupled, you can quickly measure the DC component of the signal by simply noting its distance from the ground symbol.
If the channel is AC coupled, the DC component of the signal is removed, allowing you to use greater sensitivity to display the AC component of the signal.
1 Press the desired channel key. 2 In the Channel menu, press the Coupling softkey to select the input channel
coupling:
DC — DC coupling is useful for viewing waveforms as low as 0 Hz that do not
have large DC offsets.
AC — AC coupling is useful for viewing waveforms with large DC offsets.
AC coupling places a 10 Hz high-pass filter in series with the input waveform that removes any DC offset voltage from the waveform.
Note that Channel Coupling is independent of Trigger Coupling. To change trigger coupling see "To select the trigger coupling" on page 124.

To specify bandwidth limiting

1 Press the desired channel key. 2 In the Channel menu, press the BW Limit softkey to enable or disable bandwidth
limiting.
50 Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide
Page 51
Vertical Controls 3
When bandwidth limit is on, the maximum bandwidth for the channel is approximately 20 MHz. For waveforms with frequencies below this, turning bandwidth limit on removes unwanted high frequency noise from the waveform. The bandwidth limit also limits the trigger signal path of any channel that has BW Limit turned on.

To change the vertical scale knob's coarse/fine adjustment setting

1 Push the channel's vertical scale knob (or press the channel key and then the
Fine softkey in the Channel menu) to toggle between fine and coarse
adjustment of the vertical scale.
When Fine adjustment is selected, you can change the channel's vertical sensitivity in smaller increments. The channel sensitivity remains fully calibrated when Fine is on.
The vertical scale value is displayed in the status line at the top of the display.
When Fine is turned off, turning the volts/division knob changes the channel sensitivity in a 1-2-5 step sequence.

To invert a waveform

1 Press the desired channel key. 2 In the Channel menu, press the Invert softkey to invert the selected channel.
When Invert is selected, the voltage values of the displayed waveform are inverted.
Invert affects how a channel is displayed. However, when using basic triggers, the oscilloscope attempts to maintain the same trigger point by changing trigger settings.
Inverting a channel also changes the result of any math function selected in the waveform Math menu or any measurement.

Setting Analog Channel Probe Options

1 Press the probe's associated channel key.
Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide 51
Page 52
3 Vertical Controls
2 In the Channel menu, press the Probe softkey to display the Channel Probe
menu.
The Channel Probe menu lets you select additional probe parameters such as attenuation factor and units of measurement for the connected probe.
The Probe Check softkey guides you through the process of compensating passive probes (such as the N2140A, N2142A, N2862A/B, N2863A/B, N2889A, N2890A, 10073C, 10074C, or 1165A probes).
See Also "To specify the channel units" on page 52
"To specify the probe attenuation" on page 52
"To specify the probe skew" on page 53

To specify the channel units

1 Press the probe's associated channel key. 2 In the Channel menu, press Probe. 3 In the Channel Probe menu, press Units; then, select:
Volts — for a voltage probe.
Amps — for a current probe.
Channel sensitivity, trigger level, measurement results, and math functions will reflect the measurement units you have selected.

To specify the probe attenuation

The probe attenuation factor must be set properly for accurate measurement results.
52 Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide
Page 53
To set the probe attenuation factor:
1 Press the channel key. 2 Press the Probe softkey until you have selected how you want to specify the
attenuation factor, choosing either Ratio or Decibels.
3 Turn the Entry knob to set the attenuation factor for the connected probe.
When measuring voltage values, the attenuation factor can be set from 0.100:1 to 10000:1 in a 1-2-5 sequence.
When measuring current values with a current probe, the attenuation factor can be set from 10.0 V/A to 0.0001 V/A.
When specifying the attenuation factor in decibels, you can select values from
-20 dB to 80 dB.
If Amps is chosen as the units and a manual attenuation factor is chosen, then the units as well as the attenuation factor are displayed.

To specify the probe skew

Vertical Controls 3
When measuring time intervals in the nanoseconds (ns) range, small differences in cable length can affect the measurement. Use Skew to remove cable-delay errors between any two channels.
1 Probe the same point with both probes. 2 Press one of the probes associated channel key. 3 In the Channel menu, press Probe. 4 In the Channel Probe menu, press Skew; then, select the desired skew value.
Each analog channel can be adjusted ±100 ns for a total of 200 ns difference.
The skew setting is not affected by pressing [Default Setup] or [Auto Scale].
Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide 53
Page 54
3 Vertical Controls
54 Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide
Page 55
Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes
User's Guide

4 Analog Bus Display

You can display a bus made up of the analog channel inputs and, on 2-channel oscilloscope models, the external trigger input. Any of the input channels can be assigned to the bus.
To display an analog bus:
1 Press the [Analyze] key. 2 Press Features; then, select Analog Bus. 3 Press Features again (or push the Entry knob) to enable the analog bus display. 4 To assign channels to the bus, press the Channel softkey, turn the Entry knob to
select the channel, and then press Channel again to include or exclude the channel in the bus.
5 To specify the analog bus value number base, press Base and turn the Entry
knob to select Hex or Binary.
6 To specify the channel threshold voltage levels, which determines 0 and 1 bit
levels in the bus value, use the Source softkey and the Entry knob to select a channel, then use the Threshold softkey and Entry knob to specify its threshold voltage.
You can push the Entry knob to toggle between Source and Threshold softkey selection.
55
Page 56
4 Analog Bus Display
The bus values display appears at the bottom of the graticule, above any serial bus waveforms (if present). Channel 1 is the least significant bit and channel 4 is the most significant bit.
56 Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide
Page 57
Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes
User's Guide

5 FFT Spectral Analysis

FFT is used to compute the fast Fourier transform of analog input signals. FFT takes the digitized time record of the specified source and transforms it to the frequency domain. When the FFT function is selected, the FFT spectrum is plotted on the oscilloscope display as magnitude in dBV versus frequency. The readout for the horizontal axis changes from time to frequency (Hertz) and the vertical readout changes from volts to dB.
Use the FFT function to find crosstalk problems, to find distortion problems in analog waveforms caused by amplifier non-linearity, or for adjusting analog filters.
To display a FFT waveform: 1 Press the [FFT] key.
Source 1 — selects the source for the FFT.
Settings — displays the FFT Settings menu.
Span — specifies the frequency range represented by the
width of the display. Divide span by 10 to calculate the frequency scale per division.
Center — specifies the frequency at the center vertical grid
line of the display.
57
Page 58
5 FFT Spectral Analysis
NOTE
2 Press the Settings softkey to display additional FFT settings.
Source 1 — selects the source for the FFT.
Window— selects a window to apply to your FFT input
signal:
Hanning — window for making accurate frequency
measurements or for resolving two frequencies that are close together.
Flat Top — window for making accurate amplitude
measurements of frequency peaks.
Rectangular — good frequency resolution and amplitude
accuracy, but use only where there will be no leakage effects. Use on self-windowing waveforms such as pseudo-random noise, impulses, sine bursts, and decaying sinusoids.
Blackman Harris — window reduces time resolution
compared to a rectangular window, but improves the capacity to detect smaller impulses due to lower secondary lobes.
Vertical Units — lets you select Decibels or VRMS as the units for the FFT
vertical scale.
Auto Setup — sets the frequency Span and Center to values that will cause
the entire available spectrum to be displayed. The maximum available frequency is half the FFT sample rate, which is a function of the time per division setting. The FFT resolution is the quotient of the sampling rate and the number of FFT points (f
/N). The current FFT Resolution is displayed.
S
Scale and offset considerations
If you do not manually change the FFT scale or offset settings, when you turn the horizontal scale knob, the span and center frequency settings will automatically change to allow optimum viewing of the full spectrum.
If you do manually set scale or offset, turning the horizontal scale knob will not change the span or center frequency settings, allowing you see better detail around a specific frequency.
Pressing the FFT Auto Setup softkey will automatically rescale the waveform and span and center will again automatically track the horizontal scale setting.
58 Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide
Page 59
FFT Spectral Analysis 5
3 To make cursor measurements, press the [Cursors] key and set the Source
softkey to Math N.
Use the X1 and X2 cursors to measure frequency values and difference between two frequency values (ΔX). Use the Y1 and Y2 cursors to measure amplitude in dB and difference in amplitude (ΔY).
4 To make other measurements, press the [Meas] key and set the Source softkey to
Math N.
You can make peak-to-peak, maximum, minimum, and average dB measurements on the FFT waveform. You can also find the frequency value at the first occurrence of the waveform maximum by using the X at Max Y measurement.
To adjust the FFT waveform scale and offset, use the multiplexed scale and position knobs next to the [Math] key to re-size and re-position the math waveform.
The following FFT spectrum was obtained by connecting a 2.5 V, 100 kHz square wave to channel 2. Set the horizontal scale to 50 µs/div, vertical sensitivity to 1 V/div, Units/div to 20 dBV, Offset to -40.0 dBV, Center frequency to 500 kHz, frequency Span to 1 MHz, and window to Hanning.
See Also "FFT Measurement Hints" on page 60
Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide 59
Page 60
5 FFT Spectral Analysis
NOTE
"FFT Units" on page 61
"FFT DC Value" on page 61
"FFT Aliasing" on page 62
"FFT Spectral Leakage" on page 63
"Units for Math Waveforms" on page 67

FFT Measurement Hints

The number of points acquired for the FFT record can be up to 65,536, and when frequency span is at maximum, all points are displayed. Once the FFT spectrum is displayed, the frequency span and center frequency controls are used much like the controls of a spectrum analyzer to examine the frequency of interest in greater detail. Place the desired part of the waveform at the center of the screen and decrease frequency span to increase the display resolution. As frequency span is decreased, the number of points shown is reduced, and the display is magnified.
While the FFT spectrum is displayed, use the [Math] and [Cursors] keys to switch between measurement functions and frequency domain controls in FFT menu.
FFT Resolution
The FFT resolution is the quotient of the sampling rate and the number of FFT points (fS/N). With a fixed number of FFT points (up to 65,536), the lower the sampling rate, the better the resolution.
Decreasing the effective sampling rate by selecting a greater time/div setting will increase the low frequency resolution of the FFT display and also increase the chance that an alias will be displayed. The resolution of the FFT is the effective sample rate divided by the number of points in the FFT. The actual resolution of the display will not be this fine as the shape of the window will be the actual limiting factor in the FFTs ability to resolve two closely space frequencies. A good way to test the ability of the FFT to resolve two closely spaced frequencies is to examine the sidebands of an amplitude modulated sine wave.
For the best vertical accuracy on peak measurements:
• Make sure the probe attenuation is set correctly. The probe attenuation is set from the Channel menu if the operand is a channel.
60 Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide
Page 61

FFT Units

FFT Spectral Analysis 5
• Set the source sensitivity so that the input signal is near full screen, but not clipped.
• Use the Flat Top window.
• Set the FFT sensitivity to a sensitive range, such as 2 dB/division.
For best frequency accuracy on peaks:
• Use the Hanning window.
• Use Cursors to place an X cursor on the frequency of interest.
• Adjust frequency span for better cursor placement.
• Return to the Cursors menu to fine tune the X cursor.
For more information on the use of FFTs, please refer to Keysight Application Note 243, The Fundamentals of Signal Analysis at
http://literature.cdn.keysight.com/litweb/pdf/5952-8898E.pdf. Additional
information can be obtained from Chapter 4 of the book Spectrum and Network Measurements by Robert A. Witte.
0 dBV is the amplitude of a 1 Vrms sinusoid. When the FFT source is channel 1 or channel 2 (or channel 3 or 4 on 4-channel models), FFT units will be displayed in dBV when channel units is set to Volts and channel impedance is set to 1 MΩ.
FFT units will be displayed in dBm when channel units is set to Volts and channel impedance is set to 50Ω.
FFT units will be displayed as dB for all other FFT sources or when a source channel's units has been set to Amps.

FFT DC Value

The FFT computation produces a DC value that is incorrect. It does not take the offset at center screen into account. The DC value is not corrected in order to accurately represent frequency components near DC.
Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide 61
Page 62
5 FFT Spectral Analysis
NOTE

FFT Aliasing

When using FFTs, it is important to be aware of frequency aliasing. This requires that the operator have some knowledge as to what the frequency domain should contain, and also consider the sampling rate, frequency span, and oscilloscope vertical bandwidth when making FFT measurements. The FFT resolution (the quotient of the sampling rate and the number of FFT points) is displayed on-screen when the FFT menu is open.
Nyquist Frequency and Aliasing in the Frequency Domain
The Nyquist frequency is the highest frequency that any real-time digitizing oscilloscope can acquire without aliasing. This frequency is half of the sample rate. Frequencies above the Nyquist frequency will be under sampled, which causes aliasing. The Nyquist frequency is also called the folding frequency because aliased frequency components fold back from that frequency when viewing the frequency domain.
Aliasing happens when there are frequency components in the signal higher than half the sample rate. Because the FFT spectrum is limited by this frequency, any higher components are displayed at a lower (aliased) frequency.
The following figure illustrates aliasing. This is the spectrum of a 990 Hz square wave, which has many harmonics. The horizontal time/div setting for the square wave sets the sample rate and results in a FFT resolution of 1.91 Hz. The displayed FFT spectrum waveform shows the components of the input signal above the Nyquist frequency to be mirrored (aliased) on the display and reflected off the right edge.
62 Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide
Page 63

Figure 9 Aliasing

FFT Spectral Analysis 5
Because the frequency span goes from 0 to the Nyquist frequency, the best way to prevent aliasing is to make sure that the frequency span is greater than the frequencies of significant energy present in the input signal.

FFT Spectral Leakage

The FFT operation assumes that the time record repeats. Unless there is an integral number of cycles of the sampled waveform in the record, a discontinuity is created at the end of the record. This is referred to as leakage. In order to minimize spectral leakage, windows that approach zero smoothly at the beginning and end of the signal are employed as filters to the FFT. The FFT menu provides four windows: Hanning, Flat Top, Rectangular, and Blackman-Harris. For more information on leakage, see Keysight Application Note 243, The Fundamentals of Signal Analysis at
http://literature.cdn.keysight.com/litweb/pdf/5952-8898E.pdf.
Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide 63
Page 64
5 FFT Spectral Analysis
64 Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide
Page 65
Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes
User's Guide

6 Math Waveforms

To display math waveforms / 66 To perform a math function on an arithmetic operation / 67 To adjust the math waveform scale and offset / 67 Units for Math Waveforms / 67 Math Operators / 68 Math Transforms / 70 Math Filters / 74
Math functions can be performed on analog channels. The resulting math waveform is displayed in light purple.
You can use a math function on a channel even if you choose not to display the channel on-screen.
You can:
• Perform an arithmetic operation (like add, subtract, or multiply) on analog input channels.
• Perform a transform function (like FFT) on an analog input channel.
• Perform a transform function on the result of an arithmetic operation.
65
Page 66
6 Math Waveforms
TIP

To display math waveforms

1 Press the [Math] key on the front panel to display the waveform Math menu.
2 If f(t) is not already shown on the Function softkey, press the
Function sofkey and select f(t): Displayed.
3 Use the Operator softkey to select an operator or transform.
For more information on the operators, transforms, or filters, see:
"Math Operators" on page 68
"Math Transforms" on page 70
"Math Filters" on page 74
4 Use the Source 1 softkey to select the analog channel on
which to perform math. You can rotate the Entry knob or repetitively press the Source 1 softkey to make your selection. If you choose a transform function (FFT) the result is displayed.
5 If you select an arithmetic operator, use the Source 2 softkey
to select the second source for the arithmetic operation. The result is displayed.
6 To re-size and re-position the math waveform, see "To adjust the math
waveform scale and offset" on page 67.
Math Operating Hints
If the analog channel or math function is clipped (not fully displayed on screen) the resulting displayed math function will also be clipped.
Once the function is displayed, the analog channel(s) may be turned off for better viewing of the math waveform.
The vertical scaling and offset of each math function can be adjusted for ease of viewing and measurement considerations.
The math function waveform can be measured using [Cursors] and/or [Meas].
66 Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide
Page 67

To perform a math function on an arithmetic operation

NOTE
To perform an FFT or low-pass filter math function on an arithmetic operation (add, subtract, or multiply):
1 Press the Function softkey and select g(t): Internal. 2 Use the Operator, Source 1, and Source 2 softkeys to set up an arithmetic
operation.
3 Press the Function softkey and select f(t): Displayed. 4 Use the Operator softkey to select an FFT or low-pass filter math function. 5 Press the Source 1 softkey and select g(t) as the source. Note that g(t) is only
available when you select an FFT or low-pass filter math function in the previous step.

To adjust the math waveform scale and offset

Use the multiplexed scale and position knobs next to the [Math] key to re-size and re-position the math waveform.
Math Waveforms 6
Math Scale and Offset are Set Automatically
Any time the currently displayed math function definition is changed, the function is automatically scaled for optimum vertical scale and offset. If you manually set scale and offset for a function, select a new function, then select the original function, the original function will be automatically rescaled.
See Also "Units for Math Waveforms" on page 67

Units for Math Waveforms

Units for each input channel can be set to Volts or Amps using the Units softkey in the Channel Probe menu. Units for math function waveforms are:
Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide 67
Page 68
6 Math Waveforms

Math Operators

Math function Units
add or subtract V or A
multiply
FFT Magnitude dB (decibels) or V RMS.
FFT Phase degrees or radians
A scale unit of U (undefined) will be displayed for math functions when two source channels are used and they are set to dissimilar units and the combination of units cannot be resolved.
Math operators perform arithmetic operations (add, subtract, or multiply) on analog input channels.
"Add or Subtract" on page 68
"Multiply or Divide" on page 69
2
, A2, or W (Volt-Amp)
V

Add or Subtract

When you select add or subtract, the Source 1 and Source 2 values are added or subtracted point by point, and the result is displayed.
You can use subtract to make a differential measurement or to compare two waveforms.
If your waveforms' DC offsets are larger than the dynamic range of the oscilloscope's input channels you will need to use a differential probe instead.
68 Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide
Page 69
Figure 10 Example of Subtract Channel 2 from Channel 1
Math Waveforms 6
See Also "Units for Math Waveforms" on page 67

Multiply or Divide

When you select the multiply or divide math function, the Source 1 and Source 2 values are multiplied or divided point by point, and the result is displayed.
The divide by zero case places holes (that is, zero values) in the output waveform.
Multiply is useful for seeing power relationships when one of the channels is proportional to the current.
Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide 69
Page 70
6 Math Waveforms
Figure 11 Example of Multiply Channel 1 by Channel 2
See Also "Units for Math Waveforms" on page 67

Math Transforms

Math transforms perform a transform function (FFT) on an analog input channel or on the result of an arithmetic operation.
"FFT Magnitude, FFT Phase" on page 70

FFT Magnitude, FFT Phase

Using the Fast Fourier Transform (FFT), the FFT (Magnitude) math function displays the magnitudes of the frequency content that makes up the source waveform, and the FFT (Phase) math function shows the phase relationships of the frequency content. The FFT takes the digitized time record of the specified source and transforms it to the frequency domain.
The source of the FFT math functions can be analog input channels or an arithmetic operation g(t).
70 Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide
Page 71
Math Waveforms 6
The horizontal axis of FFT math functions is frequency (Hertz). For the FFT (Magnitude) math function, the vertical axis is in decibels or V RMS. For the FFT (Phase) math function, the vertical axis is in degrees or radians.
Use the FFT (Magnitude) function to find crosstalk problems, to find distortion problems in analog waveforms caused by amplifier non-linearity, or for adjusting analog filters.
To display a FFT waveform: 1 Press the [Math] key, press the Function softkey and select f(t), press the Operator
softkey and select FFT (Magnitude) or FFT (Phase).
More — displays the FFT settings menu.
Auto Setup — sets the frequency Span and Center to values
that will cause the entire available spectrum to be displayed. The maximum available frequency is half the FFT sample rate, which is a function of the time per division setting. The FFT resolution is the quotient of the sampling rate and the number of FFT points (f
current FFT Resolution is displayed.
/N). The
S
Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide 71
Page 72
6 Math Waveforms
2 Press the More softkey to display the FFT settings.
Source 1 — selects the source for the FFT. (See "To perform
a math function on an arithmetic operation" on
page 67 for information about using g(t) as the source.)
Span — sets the overall width of the FFT spectrum that you
see on the display (left to right). Divide span by 10 to calculate the number of Hertz per division. It is possible to set Span above the maximum available frequency, in which case the displayed spectrum will not take up the whole screen. Press the Span softkey, then turn the Entry knob to set the desired frequency span of the display.
Center — sets the FFT spectrum frequency represented at
the center vertical grid line of the display. It is possible to set the Center to values below half the span or above the maximum available frequency, in which case the displayed spectrum will not take up the whole screen. Press the Center softkey, then turn the Entry knob to set the desired center frequency of the display.
Window— selects a window to apply to your FFT input signal:
Hanning — window for making accurate frequency measurements or for
resolving two frequencies that are close together.
Flat Top — window for making accurate amplitude measurements of
frequency peaks.
Rectangular — good frequency resolution and amplitude accuracy, but use
only where there will be no leakage effects. Use on self-windowing waveforms such as pseudo-random noise, impulses, sine bursts, and decaying sinusoids.
Blackman Harris — window reduces time resolution compared to a
rectangular window, but improves the capacity to detect smaller impulses due to lower secondary lobes.
Vertical Units — For FFT (Magnitude), you can select Decibels or VRMS. For
FFT (Phase), you can select Degrees or Radians.
Use the knobs for the [Math] key to adjust the FFT waveform vertical scale and offset.
72 Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide
Page 73
Math Waveforms 6
NOTE
Scale and offset considerations
If you do not manually change the FFT scale or offset settings, when you turn the horizontal scale knob, the span and center frequency settings will automatically change to allow optimum viewing of the full spectrum.
If you do manually set scale or offset, turning the horizontal scale knob will not change the span or center frequency settings, allowing you see better detail around a specific frequency.
Pressing the FFT Auto Setup softkey will automatically rescale the waveform and span and center will again automatically track the horizontal scale setting.
3 To make cursor measurements, press the [Cursors] key and set the Source
softkey to Math: f(t).
Use the X1 and X2 cursors to measure frequency values and difference between two frequency values (ΔX). Use the Y1 and Y2 cursors to measure amplitude in dB and difference in amplitude (ΔY).
4 To make other measurements, press the [Meas] key and set the Source softkey to
Math: f(t).
You can make peak-to-peak, maximum, minimum, and average dB measurements on the FFT waveform. You can also find the frequency value at the first occurrence of the waveform maximum by using the X at Max Y measurement.
The following FFT spectrum was obtained by connecting a 4 V, 75 kHz square wave to channel 1. Set the horizontal scale to 50 µs/div, vertical sensitivity to 1 V/div, Units/div to 20 dBV, Offset to -60.0 dBV, Center frequency to 250 kHz, frequency Span to 500 kHz, and window to Hanning.
Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide 73
Page 74
6 Math Waveforms
See Also "To perform a math function on an arithmetic operation" on page 67
"FFT Measurement Hints" on page 60
"FFT Units" on page 61
"FFT DC Value" on page 61
"FFT Aliasing" on page 62
"FFT Spectral Leakage" on page 63
"Units for Math Waveforms" on page 67

Math Filters

You can use math filters to create a waveform that is the result of a low-pass filter on an analog input channel or on the result of an arithmetic operation.
"Low Pass Filter" on page 74

Low Pass Filter

The low-pass filter function applies the filter to the selected source waveform and display the resut in the math waveform.
74 Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide
Page 75
Math Waveforms 6
NOTE
The low-pass filter is a 4th order Bessel-Thompson filter.
Use the Bandwidth softkey to select the filter's -3 dB cutoff frequency.
The ratio of the input signal's Nyquist frequency and the selected -3 dB cutoff frequency affects how many points are available in the output, and under some circumstances, there are no points in the output waveform.
Figure 12 Example of Low Pass Filter
Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide 75
Page 76
6 Math Waveforms
76 Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide
Page 77
Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes
User's Guide

7 Reference Waveforms

To save a waveform to a reference waveform location / 77 To display a reference waveform / 78 To scale and position reference waveforms / 79 To adjust reference waveform skew / 79 To display reference waveform information / 80 To save/recall reference waveform files to/from a USB storage device / 80
Analog channel or math waveforms can be saved to one of two reference waveform locations in the oscilloscope. Then, a reference waveform can be displayed and compared against other waveforms. One reference waveform can be displayed at a time.
You can adjust the vertical scale and offset of reference waveforms. There is also a skew adjustment for reference waveforms. Reference waveform scale, offset, and skew information can optionally be included on the oscilloscope display.
Analog channel, math, or reference waveforms can be saved to a reference waveform file on a USB storage device. You can recall a reference waveform file from a USB storage device into one of the reference waveform locations.

To save a waveform to a reference waveform location

1 Press the [Analyze] key. 2 Press Features; then, select R1 or R2. 3 Press Features again (or push the Entry knob) to turn on the reference waveform. 4 Press Save/Clear.
77
Page 78
7 Reference Waveforms
NOTE
5 In the Save/Clear menu, press the Source softkey and turn the Entry knob to
6 Press the Save to R1/R2 softkey to save the waveform to the reference waveform
Reference waveforms are non-volatile — they remain after power cycling or performing a default setup.
select the source waveform.
location.
To clear a reference
waveform location
1 Press the [Analyze] key. 2 Press Features; then, select R1 or R2. 3 Press Features again (or push the Entry knob) to turn on the reference waveform. 4 Press Save/Clear. 5 In the Save/Clear menu, press the Clear R1/R2 softkey to clear the reference
waveform location.
Reference waveforms are also cleared by a Factory Default or Secure Erase (see
Chapter 20, “Save/Recall (Setups, Screens, Data),” starting on page 215).

To display a reference waveform

1 Press the [Analyze] key. 2 Press Features; then, select R1 or R2. 3 Press Features again (or push the Entry knob) to turn on the reference waveform.
78 Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide
Page 79
One reference waveform can be displayed at a time.
Reference Waveforms 7
See Also "To display reference waveform information" on page 80

To scale and position reference waveforms

1 Display the desired reference waveform (see "To display a reference
waveform" on page 78).
2 Press the Scale softkey and turn the Entry knob to adjust the reference
waveform vertical scale.
3 Press the Offset softkey and turn the Entry knob to adjust the reference
waveform vertical offset.

To adjust reference waveform skew

Once reference waveforms are displayed, you can adjust their skew. 1 Display the desired reference waveform (see "To display a reference
waveform" on page 78).
Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide 79
Page 80
7 Reference Waveforms
2 Press the Skew softkey and turn the Entry knob to adjust the reference
waveform skew.

To display reference waveform information

1 Press the [Analyze] key. 2 Press Features; then, select R1 or R2. 3 Press Save/Clear. 4 In the Save/Clear menu, press the Display Info softkey to enable or disable
reference waveform information on the oscilloscope display.
5 Press the Transparent softkey to enable or disable transparent information
backgrounds.
This setting is also used for other oscilloscope information on the display, like mask test statistics, etc.

To save/recall reference waveform files to/from a USB storage device

Analog channel, math, or reference waveforms can be saved to a reference waveform file on a USB storage device. See "To save reference waveform files to a
USB storage device" on page 220.
You can recall a reference waveform file from a USB storage device into one of the reference waveform locations. See "To recall reference waveform files from a USB
storage device" on page 224.
80 Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide
Page 81
Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes
User's Guide

8 Serial Bus Decode/Trigger

Depending on the oscilloscope model, the following hardware-accelerated serial decode and trigger features are available:
Serial decode and trigger type: Available on:
CAN (Controller Area Network) DSOX1200-Series models
I2C (Inter-IC) DSOX1200-Series and EDUX1052A/G
models
LIN (Local Interconnect Network) DSOX1200-Series models
SPI (Serial Peripheral Interface) DSOX1200-Series models
UART (Universal Asynchronous Receiver/Transmitter) protocols including RS232 (Recommended Standard 232)
Triggering on
Serial Data
See Also Chapter 25, “CAN Triggering and Serial Decode,” starting on page 275
In some cases, such as when triggering on a slow serial signal (for example, I2C, SPI, CAN, LIN, etc.), to stabilize the waveform display, it may be necessary to switch from the Auto trigger mode to the Normal trigger mode to prevent the oscilloscope from auto-triggering. You can select the trigger mode by pressing the [Trigger] key, then the Mode softkey.
Also, the threshold voltage level must be set appropriately for each source channel. The threshold level for each serial signal can be set in the Signals menu (press [Analyze] > Features, select Serial Bus, then press the Signals softkey).
Chapter 26, “I2C Triggering and Serial Decode,” starting on page 285
Chapter 27, “LIN Triggering and Serial Decode,” starting on page 293
DSOX1200-Series and EDUX1052A/G models
81
Page 82
8 Serial Bus Decode/Trigger
Chapter 28, “SPI Triggering and Serial Decode,” starting on page 301
Chapter 29, “UART/RS232 Triggering and Serial Decode,” starting on page 313
82 Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide
Page 83
Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes
User's Guide

9 Display Settings

To adjust waveform intensity / 83 To set or clear persistence / 85 To clear the display / 86 To select the grid type / 86 To adjust the grid intensity / 87 To add an annotation / 87 To freeze the display / 89

To adjust waveform intensity

You can adjust the intensity of displayed waveforms to account for various signal characteristics, such as fast time/div settings and low trigger rates.
Increasing the intensity lets you see the maximum amount of noise and infrequently occurring events.
Reducing the intensity can expose more detail in complex signals as shown in the following figures.
1 Press the [Intensity] key to illuminate it.
This key is located just below the Entry knob.
2 Turn the Entry knob to adjust the waveform instensity.
Waveform intensity adjustment affects analog channel waveforms only (not math waveforms, reference waveforms, etc.).
83
Page 84
9 Display Settings

Figure 13 Amplitude Modulation Shown at 100% Intensity

Figure 14 Amplitude Modulation Shown at 40% Intensity

84 Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide
Page 85

To set or clear persistence

With persistence, the oscilloscope updates the display with new acquisitions, but does not immediately erase the results of previous acquisitions. All previous acquisitions are displayed with reduced intensity. New acquisitions are shown in their normal color with normal intensity.
Waveform persistence is kept only for the current display area; you cannot pan and zoom the persistence display.
To use persistence:
1 Press the [Display] key. 2 In the Display menu, press Persistence. 3 In the Persistence menu, press Persistence; then, turn the Entry knob to select
between:
Off — turns off persistence.
When persistence is off, you can press the Capture Waveforms softkey to perform a single-shot infinite persistence. A single acquisition's data is displayed with reduced intensity, and it remains on the display until you clear persistence or clear the display.
∞ Persistence — (infinite presistence) Results of previous acquisitions are
never erased.
Display Settings 9
Use infinite persistence to measure noise and jitter, to see the worst-case extremes of varying waveforms, to look for timing violations, or to capture events that occur infrequently.
Variable Persistence — Results of previous acquisitions are erased after a
certain amount of time.
Variable persistence gives you a view of acquired data that is similar to analog oscilloscopes.
When variable persistence is selected, press the Time softkey and use the Entry knob to specify the amount of time that previous acquisitions are to be displayed.
The display will begin accumulating multiple acquisitions.
4 To erase the results of previous acquisitions from the display, press the Clear
Persistence softkey.
Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide 85
Page 86
9 Display Settings
The oscilloscope will start to accumulate acquisitions again.
5 To return the oscilloscope to the normal display mode, turn off persistence;
then, press the Clear Persistence softkey.
Turning off persistence does not clear the display. The display is cleared if you press the Clear Display softkey or if you press the [Auto Scale] key (which also turns off persistence).
For another method of seeing worst-case extremes of varying waveforms, see
"Glitch or Narrow Pulse Capture" on page 136.

To clear the display

1 Press [Display] > Clear Display.
You can also configure the [Quick Action] key to clear the display. See "Configuring
the [Quick Action] Key" on page 246.

To select the grid type

When the Video trigger type is selected (see "Video Trigger" on page 108), and the vertical scaling of at least one displayed channel is 140 mV/div, the Grid softkey lets you select from these grid types:
Full — the normal oscilloscope grid.
mV — shows vertical grids, labeled on the left, from -0.3 V to 0.8 V.
IRE — (Institute of Radio Engineers) shows vertical grids in IRE units, labeled on the left, from -40 to 100 IRE. The 0.35 V and 0.7 V levels from the mV grid are also shown and labeled at the right. When the IRE grid is selected, cursor values are also shown in IRE units. (Cursor values via the remote interface are not in IRE units.)
The mV and IRE grid values are accurate (and match Y cursor values) when the vertical scaling is 140 mV/division and the vertical offset is 245 mV.
To select the grid type:
1 Press [Display]. 2 In the Display menu, press Grid.
86 Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide
Page 87
3 In the Grid menu, press the Grid softkey; then, turn the Entry knob to select
the grid type.

To adjust the grid intensity

To adjust the display grid (graticule) intensity:
1 Press [Display]. 2 In the Display menu, press Grid.
3 In the Grid menu, press the Intensity softkey; then, turn the Entry knob to
change the intensity of the displayed grid.
The intensity level is shown in the Intensity softkey and is adjustable from 0 to 100%.
Each major vertical division in the grid corresponds to the vertical sensitivity shown in the status line at the top of the display.
Each major horizontal division in the grid corresponds to the time/div shown in the status line at the top of the display.
Display Settings 9

To add an annotation

You can add an annotation to the upper left corner of the oscilloscope's display. The annotation is useful for documentation purposes, to add notes before capturing screens.
To add an annotation:
1 On the oscilloscope's front panel, press [Display]. 2 In the Display menu, press Annotation. 3 In the Annotation menu, press Annotation to enable the annotation. 4 Press Edit. 5 In the Edit Annotation menu:
Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide 87
Page 88
9 Display Settings
NOTE
Use the Spell, , , and Delete Character softkeys to enter the annotation
text:
Spell — press this softkey and turn the Entry knob to select the character
at the current position.
— press this softkey to enter characters and move the cursor to the
next character position.
— press this softkey to enter characters and move the cursor to the
previous character position.
Delete Character — press the or softkeys until the desired character
is highlighted; then, press this softkey to delete the character.
You can use a connected USB keyboard instead of using the Spell (and other) character editing softkeys.
Use the Clear softkey to delete all annotation characters.
Press OK to save the annotation edits.
88 Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide
Page 89
Display Settings 9
Back
6 Press the Back key to return the the Annotation menu. 7 Press the Text Color softkey and turn the Entry knob to select the annotation
color.
You can choose white, red, or colors that match analog channels, digital channels, math waveforms, reference waveforms, or markers.
8 Press the Background softkey and turn the Entry knob to select the annotation
background:
Opaque — the annotation has a solid background.
Inverted — the annotation's foreground and background colors are switched.
Transparent — the annotation has a transparent background.
See Also "To save BMP or PNG image files" on page 217
"To print the oscilloscope's display" on page 227

To freeze the display

To freeze the display without stopping running acquisitions, you must configure the [Quick Action] key. See "Configuring the [Quick Action] Key" on page 246.
Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide 89
Page 90
9 Display Settings
1 Once the [Quick Action] key has been configured, press it to freeze the display. 2 To un-freeze the display, press [Quick Action] again.
Manual cursors can be used on the frozen display.
Many activities, such as adjusting the trigger level, adjusting vertical or horizontal settings, or saving data will un-freeze the display.
90 Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide
Page 91
Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes
User's Guide

10 Labels

To turn the label display on or off / 91 To assign a predefined label to a channel / 92 To define a new label / 93 To load a list of labels from a text file you create / 94 To reset the label library to the factory default / 95
You can define labels and assign them to each analog input channel, or you can turn labels off to increase the waveform display area.

To turn the label display on or off

1 Press [Display] > Labels > Display.
This turns on labels for the displayed analog channels. Labels are displayed at the left edge of the displayed traces.
The figure below shows an example of displayed labels.
91
Page 92
10 Labels
2 To turn the labels off, press the Display softkey again.

To assign a predefined label to a channel

1 Press [Display] > Labels > Edit. 2 Press the Channel softkey, then turn the Entry knob or successively press the
Channel softkey to select a channel for label assignment.
92 Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide
Page 93
Labels 10
The figure above shows the list of channels and their default labels. The channel does not have to be turned on to have a label assigned to it.
3 Press the Library softkey, then turn the Entry knob or successively press the
Library softkey to select a predefined label from the library. 4 Press the Apply New Label softkey to assign the label to your selected channel. 5 Repeat the above procedure for each predefined label you want to assign to a
channel.

To define a new label

1 Press [Display] > Labels > Edit. 2 Press the Channel softkey; then, turn the Entry knob or successively press the
softkey to select a channel for label assignment.
The channel does not have to be turned on to have a label assigned to it. If the
channel is turned on, its current label will be highlighted. 3 Press the Spell softkey; then, turn the Entry knob to select the first character in
the new label.
Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide 93
Page 94
10 Labels
NOTE
Turning the Entry knob selects a character to enter into the highlighted position
shown in the "New label =" line and in the Spell softkey. Labels can be up to ten
characters in length. 4 Press the Enter softkey to enter the selected character and to go to the next
character position. 5 You may position the highlight on any character in the label name by
successively pressing the Enter softkey. 6 To delete a character from the label, press the Enter softkey until the letter you
want to delete is highlighted, then press the Delete Character softkey.
You can use a connected USB keyboard instead of using the Spell (and other) character editing softkeys.
7 When you are done entering characters for the label, press the Apply New Label
softkey to assign the label to the selected channel.
When you define a new label, it is added to the nonvolatile label list.
Label Assignment
Auto-Increment
When you assign a label ending in a digit, such as ADDR0 or DATA0, the oscilloscope automatically increments the digit and displays the modified label in the "New label" field after you press the Apply New Label softkey. Therefore, you only need to select a new channel and press the Apply New Label softkey again to assign the label to the channel. Only the original label is saved in the label list. This feature makes it easier to assign successive labels to numbered control lines and data bus lines.

To load a list of labels from a text file you create

It may be convenient to create a list of labels using a text editor, then load the label list into the oscilloscope. This lets you type on a keyboard rather than edit the label list using the oscilloscope's controls.
You can create a list of up to 75 labels and load it into the oscilloscope. Labels are added to the beginning of the list. If more than 75 labels are loaded, only the first 75 are stored.
To load labels from a text file into the oscilloscope:
94 Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide
Page 95
Labels 10
NOTE
NOTE
1 Use a text editor to create each label. Each label can be up to ten characters in
length. Separate each label with a line feed. 2 Name the file labellist.txt and save it on a USB mass storage device such as a
thumb drive.
3 Load the list into the oscilloscope using the File Explorer (press [Utility] > File
Explorer).
Label List Management
When you press the Library softkey, you will see a list of the last 75 labels used. The list does not save duplicate labels. Labels can end in any number of trailing digits. As long as the base string is the same as an existing label in the library, the new label will not be put in the library. For example, if label A0 is in the library and you make a new label called A12345, the new label is not added to the library.
When you save a new user-defined label, the new label will replace the oldest label in the list. Oldest is defined as the longest time since the label was last assigned to a channel. Any time you assign any label to a channel, that label will move to the newest in the list. Thus, after you use the label list for a while, your labels will predominate, making it easier to customize the instrument display for your needs.
When you reset the label library list (see next topic), all of your custom labels will be deleted, and the label list will be returned to its factory configuration.

To reset the label library to the factory default

Pressing the Default Library softkey will remove all user-defined labels from the library and set the labels back to the factory default. Once deleted, these user-defined labels cannot be recovered.
1 Press [Utility] > Options > Preferences. 2 Press the Default Library softkey.
This will delete all user-defined labels from the library and set the labels in the
library back to the factory default. However, this does not default the labels
currently assigned to the channels (those labels that appear in the waveform
area).
Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide 95
Page 96
10 Labels
NOTE
Defaulting labels without erasing the default library
Pressing [Default Setup] sets all channel labels back to the default labels but does not erase the list of user-defined labels in the library.
96 Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide
Page 97
Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes
User's Guide

11 Triggers

Adjusting the Trigger Level / 99 Forcing a Trigger / 99 Edge Trigger / 99 Pulse Width Trigger / 102 Pattern Trigger / 104 Rise/Fall Time Trigger / 106 Setup and Hold Trigger / 107 Video Trigger / 108 Serial Trigger / 118
A trigger setup tells the oscilloscope when to acquire and display data. For example, you can set up to trigger on the rising edge of the analog channel 1 input signal.
You can adjust the vertical level used for analog channel edge detection by turning the Trigger Level knob.
In addition to the edge trigger type, you can also set up triggers on pulse widths or video signals. On DSOX1200-Series models, you can additionally set up triggers on patterns, rise/fall times, or setup and hold violations.
97
Page 98
11 Triggers
Pre-Trigger Buffer
Post-Trigger Buffer
Acquisition Memory
Trigger Event
You can use any input channel or the EXT TRIG input BNC (see "External Trigger
Input" on page 127) as the source for most trigger types.
Changes to the trigger setup are applied immediately. If the oscilloscope is stopped when you change a trigger setup, the oscilloscope uses the new specification when you press [Run/Stop] or [Single]. If the oscilloscope is running when you change a trigger setup, it uses the new trigger definition when it starts the next acquisition.
You can use the [Force] key to acquire and display data when triggers are not occurring.
You can use the [Trigger] key to set options that affect all trigger types (see
Chapter 12, “Trigger Mode/Coupling,” starting on page 121).
You can save trigger setups along with the oscilloscope setup (see Chapter 20, “Save/Recall (Setups, Screens, Data),” starting on page 215).
Triggers - General
Information
A triggered waveform is one in which the oscilloscope begins tracing (displaying) the waveform, from the left side of the display to the right, each time a particular trigger condition is met. This provides stable display of periodic signals such as sine waves and square waves, as well as nonperiodic signals such as serial data streams.
The figure below shows the conceptual representation of acquisition memory. You can think of the trigger event as dividing acquisition memory into a pre-trigger and post-trigger buffer. The position of the trigger event in acquisition memory is defined by the time reference point and the delay (horizontal position) setting (see
"To adjust the horizontal delay (position)" on page 39).
98 Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide
Page 99

Adjusting the Trigger Level

T
NOTE
You can adjust the trigger level for a selected analog channel by turning the Trigger Level knob.
You can push the Trigger Level knob to set the level to the waveform's 50% value. If AC coupling is used, pushing the Trigger Level knob sets the trigger level to about 0 V.
The position of the trigger level for the analog channel is indicated by the trigger level icon (if the analog channel is on) at the far left side of the display. The
value of the analog channel trigger level is displayed in the upper-right corner of the display.
The line trigger level is not adjustable. This trigger is synchronized with the power line supplied to the oscilloscope.
You can also change the trigger level of all channels by pressing [Analyze] > Features and then selecting Trigger Levels.
Triggers 11

Forcing a Trigger

The [Force] key causes a trigger (on anything) and displays the acquisition.
This key is useful in the Normal trigger mode where acquisitions are made only when the trigger condition is met. In this mode, if no triggers are occurring (that is, the "Trig'd?" indicator is displayed), you can press [Force] to force a trigger and see what the input signals look like.
In the Auto trigger mode, when the trigger condition is not met, triggers are forced and the "Auto?" indicator is displayed.

Edge Trigger

The Edge trigger type identifies a trigger by looking for a specified edge (slope) and voltage level on a waveform. You can define the trigger source and slope in this menu. The trigger type, source, and level (if applicable) are displayed in the upper-right corner of the display.
Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide 99
Page 100
11 Triggers
1 On the front panel, in the Trigger section, press the [Trigger] key. 2 In the Trigger menu, press the Trigger Type softkey. 3 Press the Trigger Type softkey; then, turn the Entry knob to select Edge. 4 Select the trigger source:
Analog channel, 1 to the number of channels
External — triggers on the rear panel EXT TRIG input signal.
Line — triggers at the 50% level of the rising or falling edge of the AC power
source signal.
WaveGen — triggers at the 50% level of the rising edge of the waveform
generator output signal. (Not available when the DC or Noise waveforms are selected.)
WaveGen Mod (FSK/FM) — when waveform generator FSK or FM modulation is
used, triggers at the 50% level of the rising edge of the modulating signal.
You can choose a channel that is turned off (not displayed) as the source for the
edge trigger.
The selected trigger source is indicated in the upper-right corner of the display
next to the slope symbol:
1 through 4 = analog channels.
E = External trigger input.
L = Line trigger.
W = Waveform generator.
5 Press the Slope softkey and select rising edge, falling edge, alternating edges,
or either edge (depending on the selected source). The selected slope is
displayed in the upper-right corner of the display.
100 Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide
Loading...