Keysight 346A User Manual

Keysight 346A/B/C Noise Source (Including Options 001 and 004)
Instruments with serial prefixes 4124A and above are supplied with the noise source ENR data preloaded on a diskette - see Chapters 3 and 4.
Operating and Service Manual
Notices
CAUTION
WARNING
Copyright Notice
© Keysight Technologies 2001 - 2018 No part of this manual may be repro-
duced in any form or by any means (including electronic storage and retrieval or translation into a foreign language) without prior agreement and written consent from Keysight Technol­ogies as governed by United States and international copyright laws.
Manual Part Number
00346-90148
Edition
Edition 4, July 26, 2018
Printed in:
Printed in Malaysia
Published by:
Keysight Technologies Bayan Lepas Free Industrial Zone, 11900 Penang, Malaysia
Technology Licenses
The hard ware and/or software described in this document are fur­nished under a license and may be used or copied only in accordance with the terms of such license.
Declaration of Conformity
Declarations of Conformity for this product and for other Keysight prod­ucts may be downloaded from the Web. Go to http://www.keysight.com/
go/conformity. You can then search by
product number to find the latest Dec­laration of Conformity.
U.S. Government Rights
The Software is “commercial computer software,” as defined by Federal Acqui­sition Regulation (“FAR”) 2.101. Pursu­ant to FAR 12.212 and 27.405-3 and Department of Defense FAR Supple­ment (“DFARS”) 227.7202, the U.S. government acquires commercial com­puter software under the same terms by which the software is customarily provided to the public. Accordingly, Keysight provides the Software to U.S. government customers under its stan­dard commercial license, which is embodied in its End User License Agreement (EULA), a copy of which can be found at http://www.keysight.com/
find/sweula. The license set forth in the
EULA represents the exclusive authority by which the U.S. government may use, modify, distribute, or disclose the Soft­ware. The EULA and the license set forth therein, does not require or per­mit, among other things, that Keysight: (1) Furnish technical information related to commercial computer soft­ware or commercial computer software documentation that is not customarily provided to the public; or (2) Relinquish to, or otherwise provide, the govern­ment rights in excess of these rights customarily provided to the public to use, modify, reproduce, release, per­form, display, or disclose commercial computer software or commercial com­puter software documentation. No additional government requirements beyond those set forth in the EULA shall apply, except to the extent that those terms, rights, or licenses are explicitly required from all providers of commercial computer software pursu­ant to the FAR and the DFARS and are set forth specifically in writing else­where in the EULA. Keysight shall be under no obligation to update, revise or otherwise modify the Software. With respect to any technical data as defined by FAR 2.101, pursuant to FAR
12.211 and 27.404.2 and DFARS
227.7102, the U.S. government acquires no greater than Limited Rights as defined in FAR 27.401 or DFAR
227.7103-5 (c), as applicable in any technical data.
Warranty
THE MATERIAL CONTAINED IN THIS DOCUMENT IS PROVIDED “AS IS,” AND IS SUBJECT TO BEING CHANGED, WITHOUT NOTICE, IN FUTURE EDITIONS. FURTHER, TO THE MAXIMUM EXTENT PERMITTED BY APPLICABLE LAW, KEYSIGHT DIS­CLAIMS ALL WARRANTIES, EITHER EXPRESS OR IMPLIED, WITH REGARD TO THIS MANUAL AND ANY INFORMA­TION CONTAINED HEREIN, INCLUD­ING BUT NOT LIMITED TO THE IMPLIED WARRANTIES OF MER­CHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE. KEYSIGHT SHALL NOT BE LIABLE FOR ERRORS OR FOR INCIDENTAL OR CONSE­QUENTIAL DAMAGES IN CONNECTION WITH THE FURNISHING, USE, OR PERFORMANCE OF THIS DOCUMENT OR OF ANY INFORMATION CON­TAINED HEREIN. SHOULD KEYSIGHT AND THE USER HAVE A SEPARATE WRITTEN AGREEMENT WITH WAR­RANTY TERMS COVERING THE MATE­RIAL IN THIS DOCUMENT THAT CONFLICT WITH THESE TERMS, THE WARRANTY TERMS IN THE SEPARATE AGREEMENT SHALL CONTROL.
Safety Information
A CAUTION notice denotes a hazard. It calls attention to an operating proce­dure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not pro­ceed beyond a CAUTION notice until the indicated conditions are fully understood and met.
A WARNING notice denotes a hazard. It calls attention to an operating proce­dure, practice, or the like that, if not correctly performed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice until the indicated conditions are fully understood and met.
2 Keysight 346A/B/C Operating and Service Manual

Environmental Conditions

The instrument is designed for indoor use and in an area with low condensation. The table below shows the general environmental requirements for this instrument.
Environmental condition Requirement
Temperature
Humidity <95% RH
Altitude
Operating condition – 0 °C to 55 °C Storage/shipment condition – –55 °C to 75 °C
Operating condition – <4600 m (15000 ft) Storage/shipment condition – <15300 m (50000 ft)
Keysight 346A/B/C Operating and Service Manual 3

Safety Symbols

The following symbols on the instrument and in the manual indicate precautions which must be taken to maintain safe operation of the instrument.
The Instruction Documentation Symbol. The product is marked with this symbol when it is necessary for the user to refer to the instructions in the supplied documentation.
This symbol indicates that a device, or part of a device, may be susceptible to electrostatic discharges (ESD) which can result in damage to the product. Observe ESD precautions given on the product, or its user documentation, when handling equipment bearing this mark.
4 Keysight 346A/B/C Operating and Service Manual

Regulatory Information

MSIP-REM-Kst-
xxxxxxxxxxxxxx
The 346A/B/C Noise Source (including Options 001 and 004) complies with the following Electromagnetic Compatibility (EMC) compliances:
– IEC 61326-1/EN 61326-1
– Canada: ICES/NMB-001
– Australia/New Zealand: AS/NZS CISPR11
In order to preserve the EMC performance of the product, any cable which becomes worn or damaged must be replaced with the same type and specification.

Regulatory markings

The CE mark is a registered trademark of the European Community. This CE mark shows that the product complies with all the relevant European Legal Directives.
ICES/NMB-001 indicates that this ISM product complies with the Canadian ICES-001.
Cet appareil ISM est conforme a la norme NMB-001 du Canada.
ISM GRP.1 Class A indicates that this is an Industrial Scientific and Medical Group 1 Class A product.
Keysight 346A/B/C Operating and Service Manual 5
The RCM mark is a registered trademark of the Australian Communications and Media Authority.
This instrument is Class A suitable for professional use and is for use in electromagnetic environments outside of the home.
This symbol indicates the time period during which no hazardous or toxic substance elements are expected to leak or deteriorate during normal use. Forty years is the expected useful life of the product.
This instrument complies with the WEEE Directive (2002/96/EC) marking requirement. This affixed product label indicates that you must not discard this electrical or electronic product in domestic household waste.

Waste Electrical and Electronic Equipment (WEEE) Directive 2002/ 96/EC

This instrument complies with the WEEE Directive (2002/96/EC) marking requirement. This affixed product label indicates that you must not discard this electrical or electronic product in domestic household waste.

Product category

With reference to the equipment types in the WEEE directive Annex 1, this instrument is classified as a “Monitoring and Control Instrument” product.
The affixed product label is as shown below.
Do not dispose in domestic household waste.
To return this unwanted instrument, contact your nearest Keysight Service Center, or visit http://about.keysight.com/en/companyinfo/environment/takeback.shtml for more information.

Sales and Technical Support

To contact Keysight for sales and technical support, refer to the support links on the following Keysight websites:
www.keysight.com/find/mta
(product-specific information and support, software and documentation updates)
www.keysight.com/find/assist
(worldwide contact information for repair and service)
6 Keysight 346A/B/C Operating and Service Manual

Table of Contents

Environmental Conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
Safety Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4
Regulatory Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
Regulatory markings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
Waste Electrical and Electronic Equipment (WEEE) Directive 2002/96/
EC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Product category . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6
Sales and Technical Support . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6
1Introduction
Product Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16
2Installation
Handling Precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .20
Initial Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Service and Recalibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .21
Instruments Covered by this Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Equipment available but not supplied . . . . . . . . . . . . . . . . . . . . . . . . 22
Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .22
Mating connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
3 Specifications
General Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26
Excess Noise Ratio expanded uncertainty (U(Y)) . . . . . . . . . . . . . . . 27
Supplemental characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .29
Physical specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
4Operating Guide
Operating Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Recommended test equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . .33
Keysight 346A/B/C Operating and Service Manual 7
Operator’s check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Operator’s Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Performance Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Adjustments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Replaceable Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
Principles of operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
Repair . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
Returning the noise source for calibration . . . . . . . . . . . . . . . . . . . . 44
5 Manual Changes
Manual Changes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
Change A . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
6 Using the ENR Data Diskette or CD
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
Loading ENR Data from Diskette . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
Retrieving ENR Data from CD . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
Saving the ENR Data to the Noise Figure Analyzer’s Internal
Memory . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
Copying the ENR Data to the Noise Figure Analyzer’s Internal
Memory . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
7 ENR File Format
Format Details . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
Comment records . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
Header fields . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
ENR data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
Examples . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
Example 1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
Example 2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
A Caring for Connectors
8 Keysight 346A/B/C Operating and Service Manual
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .66
Connector part numbers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .66
Handling and storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .66
Visual Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .67
Obvious defects and damage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .67
Mating plane surfaces . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .67
Precision 7 mm connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .68
Sexed connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .69
Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .70
Compressed air . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .70
Cleaning alcohol . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70
Precision 7 mm connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .71
Cleaning interior surfaces . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
Drying connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .72
Mechanical Inspection: Connector Gages . . . . . . . . . . . . . . . . . . . . . . .73
Mechanical Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
Precision 7 mm connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .74
Sexed connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .75
50 Ω Type-N connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .75
75 Ω Type-N connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .76
Using Connector Gages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .77
Inspecting and cleaning the gage . . . . . . . . . . . . . . . . . . . . . . . . . . .77
Zeroing the gage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .77
Measuring connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .78
Making Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .80
Align connectors carefully . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .80
To make a preliminary connection . . . . . . . . . . . . . . . . . . . . . . . . . .81
Final connection using a torque wrench . . . . . . . . . . . . . . . . . . . . . .81
Disconnection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .83
Adapters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .84
Principles of Microwave Connector Care . . . . . . . . . . . . . . . . . . . . . . . .86
Keysight 346A/B/C Operating and Service Manual 9
THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.
10 Keysight 346A/B/C Operating and Service Manual

List of Figures

Figure 1-1 A partial view of typical calibration label . . . . . . . . . . .16
Figure 3-1 Characteristic ENR plot versus cardinal frequency
Figure 4-1 Typical noise figure measurement test setup . . . . . . .32
Figure 4-2 Operator’s check test setup 1 . . . . . . . . . . . . . . . . . . .34
Figure 4-3 Operator’s check test setup 2a . . . . . . . . . . . . . . . . . .35
Figure 4-4 Operator’s check test setup 2b . . . . . . . . . . . . . . . . . .36
Figure 4-5 Typical example of power reading with Noise
Figure 4-6 Typical example of power reading with Noise
Figure 4-7 Preparing for parts removal/replacement . . . . . . . . . .42
Figure 4-8 Simplified diagram of the noise source . . . . . . . . . . . .43
Figure A-1 Precision 7 mm connector . . . . . . . . . . . . . . . . . . . . . .68
Figure A-2 Precision 3.5 mm connectors . . . . . . . . . . . . . . . . . . . .69
Figure A-3 SMA connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .69
Figure A-4 Cleaning interior surfaces . . . . . . . . . . . . . . . . . . . . . .72
Figure A-5 Type-N connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . .75
Figure A-6 Using the connector gage . . . . . . . . . . . . . . . . . . . . . .78
points . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .28
Source On . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Source Off . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .38
Keysight 346A/B/C Operating and Service Manual 11
THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.
12 Keysight 346A/B/C Operating and Service Manual

List of Tables

Table 1-1 Model characteristics of individual noise sources . . . .17
Table 2-1 Connectors that can be mated with the noise
sources . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .23
Table 3-1 346A/B/C specifications . . . . . . . . . . . . . . . . . . . . . . . 26
Table 3-2 Characteristic ENR (U(Y)) specification . . . . . . . . . . . .28
Table 3-3 Supplemental characteristics . . . . . . . . . . . . . . . . . . .29
Table 3-4 Physical specifications . . . . . . . . . . . . . . . . . . . . . . . . .29
Table 4-1 Recommended test equipment for operator’s
checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .33
Table 4-2 Replaceable parts list for 346A . . . . . . . . . . . . . . . . . .41
Table 4-3 Replaceable parts list for 346B and 346C . . . . . . . . . .41
Table 4-4 Replaceable labels list . . . . . . . . . . . . . . . . . . . . . . . . .41
Table 5-1 Manual changes by serial number . . . . . . . . . . . . . . . .46
Table 7-1 Mandatory header fields . . . . . . . . . . . . . . . . . . . . . . .57
Table 7-2 Optional header fields . . . . . . . . . . . . . . . . . . . . . . . . .57
Table A-1 Recommended connector gages . . . . . . . . . . . . . . . . .73
Table A-2 Recommended torque settings . . . . . . . . . . . . . . . . . . 82
Table A-3 Adapters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .85
Table A-4 Principles of microwave connector care . . . . . . . . . . .86
Keysight 346A/B/C Operating and Service Manual 13
THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.
14 Keysight 346A/B/C Operating and Service Manual
Keysight 346A/B/C Noise Source Operating and Service Manual

1 Introduction

Product Overview 16
This chapter provides an overview of the Keysight 346A/B/C Noise Source.
15
1Introduction

Product Overview

The Keysight 346A/B/C noise source produces noise output (power-on) when +28 V is applied. When it is off, there is residual noise due to thermal agitation in the noise source (power-off). These two noise levels are used to measure the gain and added noise of the device under test, and consequently, its noise figure.
The Excess Noise Ratio (ENR) for each noise source has been measured at major frequencies and recorded on a label attached to the noise source (see Figure 1-1). ENR references power-on to the noise power that exists at 290 Kelvin (17 °C). In addition, a separate calibration sheet showing the complex reflection coefficient in both the on and off states is included with each instrument.
All three models are provided with a BNC female connector for power input. The output connector is a male APC-3.5 on the standard noise sources. Type-N connectors are available as options for the Keysight 346A and 346B noise sources. In addition to these general characteristics, certain characteristics apply to the specific models. These characteristics are listed in Tab le 1-1.
Figure 1-1 A partial view of typical calibration label
16 Keysight 346A/B/C Operating and Service Manual
Introduction 1
Table 1-1 Model characteristics of individual noise sources
Keysight 346A Keysight 346B Keysight 346C
Reflection coefficient differential between on and off states is specified to be no greater than
0.01.
Specified output noise spectrum from 10 MHz to 18 GHz.
Nominal ENR is 6 dB over the specified frequency range.
Not directly usable for Keysight 8970A IF attenuator calibration (Special Function 33). 10 dB of gain is required.
Not directly usable for Keysight 8970B IF attenuator calibration. Refer to “Calibration, IF Attenuators” in the Keysight 8970B Operating Manual.
Reflection coefficient differential between on and off states is not specified, but is less than 0.1 typically.
Specified output noise spectrum from 10 MHz to 18 GHz.
Nominal ENR is 15 dB over the specified frequency range.
Suitable for Keysight 8970A IF attenuator calibration (Special Function 33).
Suitable for Keysight 8970B IF attenuator calibration. Refer to “Calibration, IF Attenuators” in the Keysight 8970B Operating Manual.
Reflection coefficient differential between on and off states is not specified, but is less than 0.1 typically.
Specified output noise spectrum from 10 MHz to 26.5 GHz.
Nominal ENR is 15 dB over the specified frequency range.
Suitable for Keysight 8970A IF attenuator calibration (Special Function 33).
Suitable for Keysight 8970B IF attenuator calibration. Refer to “Calibration, IF Attenuators” in the Keysight 8970B Operating Manual.
Keysight 346A/B/C Operating and Service Manual 17
1Introduction
THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.
18 Keysight 346A/B/C Operating and Service Manual
Keysight 346A/B/C Noise Source Operating and Service Manual

2 Installation

Handling Precautions 20 Initial Inspection 20 Service and Recalibration 21 Instruments Covered by this Manual 21
This chapter provides you important information on how to check and prepare your instrument for operation.
19
2Installation
CAUTION
CAUTION

Handling Precautions

Do not disassemble the noise source. The diode module is static sensitive and can be damaged or the calibration can be altered.
Do not drop the noise source. Dropping can damage the unit or alter the calibration.
Proper connector care is essential. See “Operator’s Maintenance” on page 39 for more information.

Initial Inspection

1 Unpack and inspect the shipping container and its contents thoroughly to
ensure that nothing was damaged during shipment. If the shipping container or cushioning material is damaged, the contents should be checked both mechanically and electrically.
– Check for mechanical damage such as scratches or dents.
– Procedures for checking electrical performance are given under “Operator’s
check” on page 33.
2 If the contents are damaged or defective, contact your nearest Keysight
Technologies Service and Support Office. Refer to “Sales and Technical
Support” on page 6 of this manual. Keysight Technologies will arrange for
repair or replacement of the damaged or defective equipment. Keep the shipping materials for the carrier’s inspection.
3 If you are returning the instrument under warranty or for service, repackaging
the instrument requires original shipping containers and material or their equivalents. Keysight Technologies can provide packaging materials identical to the original materials. Refer to “Sales and Technical Support” on page 6 of this manual for the Keysight Technologies nearest to you. Attach a tag indicating the type of service required, return address, model number, and serial number. Mark the container FRAGILE to insure careful handling. In any correspondence, refer to the instrument by model number and serial number.
20 Keysight 346A/B/C Operating and Service Manual

Service and Recalibration

If your 346A/B/C requires service or repair, contact the nearest Keysight office for information on where to send it. The performance of the 346A/B/C can only be verified by specially-manufactured equipment and calibration standard from Keysight. The recommended interval for recalibration is 12 months.

Instruments Covered by this Manual

The noise sources covered by this manual have a two-part serial number. The first four digits and letter constitute the serial number prefix. The last five digits form a sequential suffix that is unique to each noise source. The prefix is the same for all noise sources of a particular configuration. It will change when a design modification occurs. The contents of this manual apply directly to those instruments having the same serial number prefixes listed on the title page.
A noise source manufactured after the printing of this manual may have a serial number prefix which is not listed on the title page. This unlisted serial number prefix indicates the noise source is different than those documented in this manual. If manual changes are needed, the manual for this newer noise source is accompanied by a Manual Changes supplement. The supplement contains “change information” that explains how to adapt this manual to the newer noise source.
In addition to change information, the supplement may contain information for correcting errors in the manual. The supplement is identified with the manual print date and part number, both of which appear on the manual title page.
Installation 2
For information concerning a serial number prefix that is not listed on the title page in the Manual Change supplement, contact your nearest Keysight Technologies office.
Keysight 346A/B/C Noise Source which was manufactured before the printing of this manual will have a serial number prefix lower than the one listed on the title page. Manual changes for these older models are found in Chapter 5, “Manual
Changes”.
Keysight 346A/B/C Operating and Service Manual 21
2Installation
NOTE
NOTE
Keysight Technologies are migrating their existing product serial numbers to a new format. All new products use this new format. The format is the first two letters signify the country of manufacture, for example, US representing the USA, the next four numbers, the serial number prefix, and the last five numbers form a sequential suffix that is unique to each product. For example, US123456789, where 1234 is the prefix and 56789 is the suffix. The two letters MY represent Malaysia.

Equipment available but not supplied

The following equipment is available from Keysight Technologies for use with the noise sources:
– 8710-1766: 3/4" torque wrench (APC-7)
– 5060-0344: 9/16" torque wrench (APC-3.5)
Keysight Model 346B Noise Sources with serial prefixes of 2037A and below, have an APC-3.5 connector of a different configuration. The 9/16" torque wrench will not fit these older models.

Options

The Keysight 346A and 346B Noise Sources are available with the following output connector options:
– Option 001, Type-N (male)
– Option 004, Type-N (female)
Since the Type-N connectors do not operate up to 26.5 GHz, no output connector options are available for the Keysight 346C.
22 Keysight 346A/B/C Operating and Service Manual

Mating connectors

The noise sources can be mated with other instrumentation having the connectors listed in Table 2-1.
Table 2-1 Connectors that can be mated with the noise
Configuration Mating connector
Input: all units
Output: standard
Opt 001
Opt 004
[a] Must comply with U.S. Military Standard MII,C-39012.
sources
BNC male
– APC-3.5 female – SMA female
Type-N female
Type-N male
[a]
[a]
[a]
Installation 2
Keysight 346A/B/C Operating and Service Manual 23
2Installation
THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.
24 Keysight 346A/B/C Operating and Service Manual
Keysight 346A/B/C Noise Source Operating and Service Manual

3 Specifications

General Specifications 26
This chapter provides the specifications of the 346A/B/C Noise Source.
25
3 Specifications

General Specifications

Specifications

The specifications in Table 3-1 are performance standards or limits against which the noise source may be tested. These specifications for the noise source when used with a Noise Figure Analyzer are ONLY valid if the analyzer has been allowed to meet its specified warm up time of 60 minutes.
ENR expanded uncertainty analysis and supplemental characteristics are not specifications but are typical characteristics included as additional information.
Tab le 3-1 346A/B/C specifications
Specification 346A 346B 346C
Frequency range 10 MHz to 18 GHz 10 MHz to 18 GHz 10 MHz to 26.5 GHz
Operating temperature 0 °C to 55 °C 0 °C to 55 °C 0 °C to 55 °C
Excess Noise Ratio (ENR)
[a]
range
Maximum Standing Wave Ratio (SWR)
[for source ON/OFF states]
Reflection Coefficient (Rho) (r)
[for source ON/OFF
[b]
states]
Impedance 50 Ω nominal 50 Ω nominal 50 Ω nominal
Maximum reverse power 1 watt 1 watt 1 watt
Power required
Connectors
[c]
4.5 to 6.5 dB 14 to 16 dB 12 to 17 dB
<1.3:1 (0.01 to 0.03 GHz) <1.15:1 (0.03 to 5.0 GHz) <1.25:1 (5.0 to 18.0 GHz)
0.13 (0.01 to 0.03 GHz)
0.07 (0.03 to 5.0 GHz)
0.11 (5.0 to 18.0 GHz)
28 ±1 V (60 mA peak, 30 mA
average for source ON)
Output: APC-3.5 (male)
Input: Bias: BNC (f)
standard
[d]
<1.3:1 (0.01 to 0.03 GHz) <1.15:1 (0.03 to 5.0 GHz) <1.25:1 (5.0 to 18.0 GHz)
0.13 (0.01 to 0.03 GHz)
0.07 (0.03 to 5.0 GHz)
0.11 (5.0 to 18.0 GHz)
28 ±1 V (60 mA peak, 30 mA
average for source ON)
Output: APC-3.5 (male)
Input: Bias: BNC (f)
standard
[d]
<1.25:1 (0.01 to 18.0 GHz) <1.35:1 (18.0 to 26.5 GHz)
0.11 (0.01 to 18.0 GHz)
0.15 (18.0 to 26.5 GHz)
28 ±1 V (45 mA)
Output: APC-3.5 (male)
Input: Bias: BNC (f)
standard
[d]
26 Keysight 346A/B/C Operating and Service Manual
Specifications 3
[a] ENR values are given at cardinal frequency points over the frequency range of each noise source. Calibrated values at cardinal frequencies
printed on each noise source label and on a separate report supplied with each Keysight 346A/B/C Noise Source. Specifications are valid at calibration temperature only.
[b] Maximum change in complex reflection coefficient between source ON and source OFF at all frequencies for 346A only: 0.01.
[c] For correct connector usage, refer Tab le A -2 for the torque settings.
[d] Also mates with female SMA connectors. See option information for other connector styles.

Excess Noise Ratio expanded uncertainty (U(Y))

ENR values are given at cardinal frequency points over the frequency range of each noise source. These ENR values are printed on the noise source label and on a separate printed report supplied with the 346A/B/C Noise Source. Included in the printed report is the measurement uncertainty for each cardinal frequency point.
A significant proportion of the expanded uncertainty (U(Y)) is based on the uncertainties provided by the United States National Institute of Standards and Technology (NIST) and/or The United Kingdom National Physics Laboratories (NPL). Keysight therefore reserve the right to change the overall expanded uncertainties based on changes in uncertainty values within the National Standards Institutes.
Uncertainties are valid at the ambient temperature of 23 °C ±1 °C (296 K) only.
A typical characteristic plot of ENR (U(Y)) versus each cardinal frequency point is shown in Figure 3-1. The uncertainties given are typical — refer to the printed calibration report for the actual uncertainties for your noise source.
Keysight 346A/B/C Operating and Service Manual 27