Weighing range (max) 1.5 kg 3 kg 6 kg
Readability (d) 0.2 g 0.5 g 1 g
Reproducibility 0.2 g 0.5 g 1 g
Linearity ±0.4 g ±1 g ±2 g
Recommended adjustment
weight, not added (class)
Stabilization time (typical) 2 sec.
Units kg
Warm-up tim e 30 min.
Electric Supply
Rechargeable battery
operation
1500 g
(M1)
Input voltage: 110V-230V AC
Power supply unit: 12 V, 500 mA
Service life: background illumination ON: 30 h
Background illumination off: 50 hours
Loading time 12 h
WTB
3K -3N
3 kg
(M1)
WTB
6K -3N
6 kg
(M1)
Auto-Off (rechargeable
battery)
Display type LCD with background illumination
Operating temperature -10° C …. + 40° C
Humidity of air 0 % - 80 % (non-condensing)
Dimensions of weighing plate
(Stainless steel) (mm)
Casing dimensions 286 x 316 x 126.5
Weight kg (net) 3.2
IP protection IP65
15 min., 5 min., 3 min., off
Character height 2.5 cm
262 x 202
WTB-N_WTB-NM-BA-e-1421 4
KERN WTB
10K -3N
Weighing range (max) 15 kg 30 kg
Readability (d) 2 g 5 g
Reproducibility 2 g 5 g
Linearity ±4 g ±10 g
Recommended adjustment
weight, not added (class)
Stabilization time (typical) 2 sec.
Units kg
Warm-up tim e 30 min.
Electric Supply
Service life: background illumination ON: 30 h
Rechargeable battery
operation
15 kg
(M1)
Input voltage: 110V-230V AC
Power supply unit : 12 V, 500 mA
Background illumination off: 50 hours
Loading time: 12 h
WTB
30K -3N
30 kg
(M1)
Auto-Off (rechargeable
battery)
Display type LCD, digit height 25 mm
Operating temperature -10° C …. + 40° C
Humidity of air 0 % - 80 % (non-condensing)
Dimensions of weighing plate
(Stainless steel) (mm)
Casing dimensions 286 x 316 x 126.5
Weight kg (net) 3.2
IP protection IP65
15 min., 5 min., 3 min., off
262 x 202
5 WTB-N_WTB-NM-BA-e-1421
KERN WTB
1K-4NM
Weighing range (max) 1.5 kg 3 kg 6 kg
Readability (d) 0.5g 1 g 2 g
Reproducibility 0.5 g 1 g 2 g
Linearity ±0.5 g ±1 g ±2 g
Recommended adjustment
weight, not added (class)
Stabilization time (typical) 2 sec.
Units kg
Warm-up tim e 30 min.
Electric Supply
Rechargeable battery
operation
1500 g
(M3)
Input voltage: 110V-230V AC
Power supply unit: 12 V, 500 mA
Service life: background illumination ON: 30 h
Background illumination off: 50 hours
Loading time 12 h
WTB
3K-3NM
3 kg
(M3)
WTB
6K-3NM
6 kg
(M3)
Auto-Off (rechargeable
battery)
Display type LCD with background illumination
Operating temperature -10° C …. + 40° C
Humidity of air 0 % - 80 % (non-condensing)
Dimensions of weighing plate
(Stainless steel) (mm)
Casing dimensions 286 x 316 x 126.5
Weight kg (net) 3.2
IP protection IP65
15 min., 5 min., 3 min., off
Character height 2.5 cm
262 x 202
WTB-N_WTB-NM-BA-e-1421 6
KERN WTB
10K-3NM
Weighing range (max) 15 kg 30 kg
Readability (d) 5 g 10 g
Reproducibility 5 g 10 g
Linearity ±5 g ±10 g
Recommended adjustment
weight, not added (class)
Stabilization time (typical) 2 sec.
Units kg
Warm-up tim e 30 min.
Electric Supply
Service life: background illumination ON: 30 h
Rechargeable battery
operation
15 kg
(M3)
Input voltage: 110V-230V AC
Power supply unit: 12 V, 500 mA
Background illumination off: 50 hours
Loading time: 12 h
WTB
30K-2NM
30 kg
(M3)
Auto-Off (rechargeable
battery)
Display type LCD, digit height 25 mm
Operating temperature -10° C …. + 40° C
Humidity of air 0 % - 80 % (non-condensing)
Dimensions of weighing plate
(Stainless steel) (mm)
Casing dimensions 286 x 316 x 126.5
Weight kg (net) 3.2
IP protection IP65
Should the balance not display exactly zero
despite empty scale pan, press the
button. The balance will be set to zero after a
short standby time.
Stability display Scales are in a steady state
Net weight display Net weight will be displayed
Capacity display
Rechargeable battery
Display Weighing with
tolerance range
Power supply connected Lights with power supply via power supply unit,
Display Weighing unit kg Displayed weight in kg
Triangle above the capacity indicator appears
when the battery capacity is almost exhausted
Triangle above "+": upper limit value
Triangle above "-": lower limit value
then the battery is being charged
WTB-N_WTB-NM-BA-e-1421 10
Switch-over units
2.2 Keyboard overview
Key Description Function i n m enu
ON/OFF-switch Turn on/off
+
TARE button
Tare balance
Weighing with
tolerance range
Call up menu
Change to the next menu
item /parameter
Zero setting key Set balance to zero Select menu item /
parameter
Push TARE + reset
button
simultaneously
11 WTB-N_WTB-NM-BA-e-1421
3 Basic Information (General)
3.1 Proper use
The balance you purchased is intended to determine the weighing value of material
to be weighed. It is intended to be used as a “non-automatic balcance”, i.e. the
material to be weighed is manually and carefully placed in the centre of the weighing
pan.. As soon as a stable weighing value is reached the weighing value can be read.
3.2 Improper Use
Do not use balance for dynamic weighings. In the event that small quantities are
removed or added to the material to be weighed, incorrect weighing results can be
displayed due to the “stability compensation“. (Example: Slowly draining fluids from a
container on the balance.)
Do not leave permanent load on the weighing pan. This may damage the measuring
system.
Impacts and overloading exceeding the stated maximum load (max) of the balance,
minus a possibly existing tare load, must be strictly avoided. Balance may be
damage by this.
Never operate balance in explosive environment. The serial version is not explosion
protected.
The structure of the balance may not be modified. This may lead to incorrect
weighing results, safety-relat ed faul ts and destruction of the bala nce.
The balance may only be used according to the described conditions. Other areas of
use must be released by KERN in writing.
3.3 Warranty
Warranty claims shall be voided in case
Our conditions in the operation manual are ignored
The appliance is used outside the described uses
The appliance is modified or opened
Mechanical damage or damage by media, liquids, natural wear and tear
The appliance is improperly set up or incorrectly electrically connected
The measuring system is overloaded
3.4 Monitoring of Test Resource s
In the framework of quality assurance the measuring-related properties of the
balance and, if applicable, the testing weight, must be checked regularly. The
responsible user must define a suitable interval as well as type and scope of this test.
Information is available on KERN’s home page (www.kern-sohn.com with regard to
the monitoring of balance test substances and the test weights required for this. In
KERN’s accredited DKD calibration laboratory test weights and balances may be
calibrated (return to the national standard) fast and at moderate cost.
WTB-N_WTB-NM-BA-e-1421 12
Carefully read this operation manual before setup and commissioning,
German is binding.
4 Basic Safety Precautions
4.1 Pay attention to the instr uc t ions in the Operation Manual
even if you are already familiar with KERN balances.
All language versions contain a non-binding translation. The original
4.2 Personnel training
The appliance may only be operated and maintained by trained personnel.
5 Transport and storage
5.1 Testing upon acceptance
When receiving the appliance, please check packaging immediately, and the
appliance itself when unpacking for possible visible damage.
5.2 Packaging / return transport
Keep all parts of the original packaging for a possibly required return.
Only use original packaging for returning.
Prior to dispatch disconnect all cables and remove loose/mobile
parts.
Reattach possibly supplied transport securing devices.
Secure all parts such as the glass wind screen, the weighing
platform, power unit etc. against shifting and damage.
13 WTB-N_WTB-NM-BA-e-1421
6 Unpacking, Setup and Commissioning
6.1 Installation Site, Locat ion of Use
The balances are designed in a way that reliable weighing results are achieved in
common conditions of use.
You will work accurately and fast, if you select the right location for your balance.
Therefore, observe the following for the installation site:
Place scales on a stable, even surface;
Avoid extreme heat as well as temperature fluctuation caused by installing
next to a radiator or in the direct sunlight;
Protect the balance against direct draughts due to open windows and
doors;
Avoid jarring during weighing;
Protect the balance against high humidity, vapours and dust;
Do not expose the device to extreme dampness for longer periods of time.
Non-permitted condensation (condensati on o f air humidi ty on the
appliance) may occur if a cold appliance is taken to a considerably warmer
environment. In this case, acclimatize the disconnected appliance for ca.
2 hours at room temperature.
Avoid static charge of goods to be weighed or weighing container.
Major display deviations (incorrect weighing results) may be experienced should
electromagnetic fields (e.g. due to mobile phones or radio equipment), static
electricity accumulations or instable power supply occur. Change location or remove
source of interference.
WTB-N_WTB-NM-BA-e-1421 14
6.2 Unpacking and placing
Carefully remove the balance from the packaging, remove plastic cover and setup
balance at the intended workstation.
1
Weighing pan
2
Locking screw
3
Carrier weighing plate
4
Shim
5
Display
6
Bubble level
7
Keyboard
8
Footscrews
9
Rechargeable battery compartment cover
10
Screws of battery compartment cover
15 WTB-N_WTB-NM-BA-e-1421
Level balance with foot screws until the air bubble of the water balance is in the
prescribed circle.
Pressure compensation screw:
1 Rechargeable battery
compartment
2 Mains port
3 Pressure compensation screw
[1]
On the lower side of the balance there is a pressure compensation screw [1] which
has to be opened for the weighing process. Fasten the screw before cleaning the
balance.
WTB-N_WTB-NM-BA-e-1421 16
Membrane for degree o f
6.2.1 Scope of delivery
Serial accessories:
Balance
Weighing pan
Carrier weighing plate
Fastening screw of the carrier weighing plate
Operating instructions
Hexagonal socket wrench
protection IP65
6.3 Rechargeable battery operation ( opt ional)
The rechargeable battery is charged via the delivered power supply.
The operating time of the rechargeable battery with background illumination is 30h,
without that it is 50 hrs. Charging time until complete recharging 12h.
In the menu you can activate the AUTO-OFF function [time off], see chap. 9.2.
According to menu settings, the balance switches automatically off in order to spare
the battery.
Insert rechargeable battery:
Remove any
adherent moisture
from the balance
Rotate both levers
(1) 90 degrees to the
left
17 WTB-N_WTB-NM-BA-e-1421
Remove battery compartment
cover and the foam pieces
Connecting the rechargeable
battery
Watch the colours:
Red on red !
Black on black!
Insert the rechargeable battery
in a manner that it cannot slip
(fasten with a rubber foam mat)
Ensure that the cables are
not squeezed.
WTB-N_WTB-NM-BA-e-1421 18
Close the rechargeable battery
cover
Rotate both levers (1) 90 degrees to
the right
If in the display the triangle appears above the capacity display , the
rechargeable battery is nearly exhausted. Connect power pack, the rechargeable
battery is loaded.
Avoid excess pressure on the balance in order to avoid damaging
it, particularly as it is supported by the weighing plate.
6.3.1 Mains connection during rechargeable battery operation
During the rechargeable battery operation ensure that the mains
connection is covered with a rubber cap.
Only by this the type of protection IP65 is ensured.
19 WTB-N_WTB-NM-BA-e-1421
6.4 Initial Commissioning
In order to obtain exact results with the electronic balances, your balance must have
reached the operating temperature (see warming up time chap. 1).
For this warm-up period the scale must be connected to the power supply (batteries).
The accuracy of the balance depends on the local acceleration of gravity.
Strictly observe hints in chapter Adjustment.
6.5 Type of protection IP65
The KERN WTB meets the requirements for IP65protection type.
Suitable for a short-time contact with liquids. For cleaning use damp cloth. Dust-tight.
7 Adjustment
As the acceleration value due to gravity is not the same at every location on earth,
each balance must be coordinated - in compliance with the underlying physical
weighing principle - to the existing acceleration due to gravity at its place of location
(only if the balance has not already been adjusted to the location in the factory). This
adjustment process must be carried out for the first commissioning, after each
change of location as well as in case of fluctuating environment temperature. To
receive accurate measuring values it is also recommended to adjust the balance
periodically in weighing operation.
7.1 Adjustment
Carry out adjustment as close as possible to the maximum load of the scales )see
chap. 1 „Technical data“). The accuracy of the adjustment weight must correspond
approximately to or, if possible, be better than, the readability d of the balance.
Information about test weights you will find in the internet under http://www.kern-
sohn.com
Procedure when adjusting:
Observe stable environmental conditions. A warming up time (see chapter 1) is
required for stabilization.
WTB-N_WTB-NM-BA-e-1421 20
Display
Operation
necessary, change value by .
7.1.1 Adjustment non verifiable models
Start balance by pressing
While the balance carries out a self test (00…- 99…), press
and keep pressed until „F1 CAL“ appears on the
display.
Press ;
in the display appears „UnLod“. Wait for the triangle above
(example)
the stability symbol, then press , the value of the lastused adjustment weight is displayed. Exp: „6 kg“. If
Place the adjustment w eig ht
Wait for the triangle above the stability symbol
Press
While the balance carries out a self-test, take away the
adjustment weight
The balance changes over to zero display.
Now the process of adjustment is completed.
If the display shows an error message or an incorrect value, repeat adjustment
process. If the error message remains, please contact your dealer.
21 WTB-N_WTB-NM-BA-e-1421
Display
Operation
7.1.2 Adjustment of verifiable models
In verified balances the menu is locked by the verification switch.
To disable the access lock, destroy the seal and actuate the adjustment
switch. Position of the adjustment switch see chapter 7.2.1
Attention:
After destruction of the seal the balance must be re-verified by an
authorised agency and a new verification wire/seal mark fitted before it can
be reused for applications subject to verification.
(example)
Start balance by pressing .
While the balance carries out a self test (00…- 99…), press
and keep pressed until „F1 CAL“ appears on the
display.
Actuate the adjustment switch on the lower side of the
balance
Press
in the display appears „UnLod“.
Press anew
The weighing value is displayed.
Press to enter the value of the adjustment weight
(see chapter 1)
Place the adjustment w eig ht
Press
While the balance carries out a self-test, take away the
adjustment weight
The balance changes over to zero display.
Now the process of adjustment is completed.
WTB-N_WTB-NM-BA-e-1421 22
7.2 Verification
General introduction:
According to EU directive 2009/23/EC balances must be officially verified if they are
used as follows (legally controlled area):
a) For commercial transactions if the price of goods is determined by weighing.
b) For the production of medicines in pharmacies as well as for analyses in the
medical and pharmaceutical laboratory.
c) For official purposes
d) For manufacturing final packages
In cases of doubt, please contact your local trade in standard.
Verification notes:
An EU type approval exists for balances described in their technical data as
verifiable. If a balance is used where obligation to verify exists as described above, it
must be verified and re-verified at regular intervals.
Re-verification of a balance is carried out according to the respective national
regulations. The validity for verification of balances in Germany is e.g. 2 years.
The legal regulation of the country where the balance is used must be observed!
Verification of the balance is invalid without the seal.
The seal marks attached on verified balances point out that the balance may
only be opened and serviced by trained and authorised specialist staff. If the
seal mark is destroyed, verification looses its validity. Please obs er v e all
national laws and legal regulations. In Germany a re-verification will be
necessary.
23 WTB-N_WTB-NM-BA-e-1421
7.2.1 Adjustment switch and seals
After a verification the balance is sealed at the indicated positions.
Position of the official seals:
1. Verification wire fastening
2. Verification wire fastening
3. Self-destroying seal mark
4. Cover of adjustment switch
5. Ad ju stment switch
6. Self-destroying seal mark
WTB-N_WTB-NM-BA-e-1421 24
container.
8 Operation
8.1 Weighing
Start balance by pressing .
The balance will carry out a self-test
When the weight display shows "0.0" and the triangle symbol
above the stability display appears, the scale is ready
for weighing.
However, you can reset the weighing scale to zero by pressing
the key.
Switch off balance using .
The "0.0" display disappears and the scales are switched off.
8.2 Taring
The tare weight of any preloads can be deducted by pressing a button so that the
actual weight of the weighed material is displayed in subsequent weighings.
Put on weighing receptacles and press .
(example)
The zero display appears, and above the reset symbol
the stability symbol and the net weight symbol the
triangle appears.
The weight of the container is now internally saved.
Place goods to be weighed in the weighing container.
The net weight of the goods to be weighed is displayed.
25 WTB-N_WTB-NM-BA-e-1421
(example)
The weight of the weighing container will be displayed as a
minus number (=gross weight) after removing the weighing
0.0
The tare weight is saved until it is deleted. Remove the load
from the balance and press . The zero display appears
and the triangle above the net weight symbol
vanishes.
Gross weight:
Press as long as the weighing container and the
object to be weighed are on the weighing plate.
Remove the object to be weighed as well as the weighing
container.
The gross weight is displayed as a negative value.
8.3 Background illumination
In weighing mode press for about 3 seconds, "bK-AU"
appears.
Press to select background illumination automatically
off („bK-AU“) and background illumination off („bK-oF“).
Press to confirm selected setting.
The balance changes into weighing mode
WTB-N_WTB-NM-BA-e-1421 26
8.4 Weighing with tolerance range
You can set an upper or lower limit when weighing with tolerance range and thus
ensure that the weighed load remains exactly within the set tolerance limits.
Settings:
In weighing mode press for about 3 seconds, the
display for entering the upper limit appears. The left digit
blinks. A triangle appears above the "+" symbol.
Moreover, the background of the screen turns red.
Enter upper limit value, the active position blinks.
Mit Correct the numerical value by , change the
(example)
decimal using . If the last digit is entered, confirm input
by , the upper limit value is hereby saved.
The display changes to the prompt for entering the lower limit
value.
The colour of the display then changes to orange.
(example)
Now enter the lower limit value as described above and
confirm input using . The lower limit value is thus
entered. The balance changes into weighing mode.
Audio signal:
The acoustic signal depends on the settings in menu block „F9 BEP“.
Options:
• BEP 0
Acoustic signal turned off
• BEP 1
• BEP 2
An acoustic signal sounds when load is within tolerance limits
An acoustic signal sounds when load is beyond tolerance limits
27 WTB-N_WTB-NM-BA-e-1421
Optical signals:
The triangle above the "+" - or the "-" symbol indicates whether the sample weighed
is within the tolerance range:
Goods to be weighed above tolerance limit
Goods to be weighed within tolerance range
Goods to be weighed below tolerance limit
Traffic light function:
The indicator lights up in red, green and orange, depending on the area in which your
sample is:
red Goods to be weighed above tolerance limit
green Goods to be weighed within tolerance range
orange Goods to be weighed below tolerance limit
Weighing with tolerance range
Tare when using a weighing container
Put on goods to be weighed, tolerance control is started
Load below specified
tolerance
orange background
Load within specified
tolerance
green background
Load exceeds specified
tolerance
red background
WTB-N_WTB-NM-BA-e-1421 28
9 The menu
9.1 Navigation in the menu
In verified balances the menu is locked by the verification switch.
Switch on the scales by and during the balance’s self-
test, press until "F1 CAL" is displayed.
Use to change to the next menu item
Select menu item using
Use to change between the individual parameters
Select the desired parameter using
29 WTB-N_WTB-NM-BA-e-1421
Menu item
Function
Max load
9.2 Menu overview (locked in verifiable models)
Adjustment process.
3000
6000
dual
1.5 kg
3 kg
6 kg
15 kg
25 kg
30 kg
not documented
3 min
5 min
15 min
off
Resolution
Please do not change default
setting
Do not change the default
setting
Auto-OFF function,
adjustable between off, 3, 5
and 15 minutes
not documented
SP 7.5
SP 15
SP 30
SP 60
Display speed
Do not change the default
setting
WTB-N_WTB-NM-BA-e-1421 30
Multi-Tare function:
Acoustic signal in tolerance weighing mode
bEP 0
Acoustic signal turned off
bEP 1
Acoustic signal when weighed material is
within the tolerance range
bEP 2
Acoustic signal when weighed material is
outside the tolerance range
Return to weighing mode
0 tare Multi-Tare function disabled
P tare Multi-Tare function enabled
31 WTB-N_WTB-NM-BA-e-1421
9.3 Set auto-OFF function
Switch on the scales by and during the balance’s self-
test, press until "F1 CAL" is displayed.
Press repeatedly until „F5 oFF“ appears
Select function using
Use to change between the individual parameters
Select the desired parameter using
“F5 oFF“ appears, now the function is enabled
Press repeatedly until “bACK“ appears
Press
The balance carries out a restart and changes to weighing
mode
WTB-N_WTB-NM-BA-e-1421 32
9.4 Multi-Tare function
The scale has the ability to tare several times in a row. For this purpose, select the
menu item "F8 tm" and set the parameter "P tare".
10 Error messages
Display Description Remedy
Zero range exceeded Unload the balance
A/D converter outside range U nl oad the bal a nce; check if
the weighing is correctly
placed and screwed on
33 WTB-N_WTB-NM-BA-e-1421
displayed weight is permanently
ng value is obviously
11 Instant help
In case of an error in the program process, briefly turn off the balance and disconnect
from power supply. The weighing process must then be restarted from the beginning.
Fault Possible cause
The displayed weight does not glow. T he balance is not switched on.
Batteries are inserted incorrectly or empty
No batteries inserted.
The
changing
The weighi
wrong
Should other error messages occur, switch balance off and then on again. If the error
message remains inform manufacturer.
Draught/air movement
Table/floor vibrations
Weighing pan has contact with other
objects.
Electromagnetic fi elds / stati c c har g i ng
(choose different location/switch off
interfering device if possible)
The display of the balance is not at zero
Adjustment is no longer correct.
Great fluctuations in temperature.
The balance is on an uneven surface.
Electromagnetic fi elds / stati c c har g i ng
(choose different location/switch off
interfering device if possible)
WTB-N_WTB-NM-BA-e-1421 34
12 Service, maintenance, disposal
12.1 Cleaning
Before cleaning, please remove batteries from the appliance
Please do not use aggressive cleaning agents (solvents or similar agents), but a
cloth dampened with mild soap suds. Ensure that no liquid penetrates into the device
and wipe with a dry soft cloth.
Loose residue sample/powder can be removed carefully with a brush or manual
vacuum cleaner.
Spilled weighing goods must be removed immediately.
12.2 Service, maintenance
The appliance may only be opened by trained service technicians who are authorized
by KERN.
Before opening, disconnect from power supply.
12.3 Disposal
Disposal of packaging and appliance must be carried out by operator according to
valid national or regional law of the location where the appliance is used.
35 WTB-N_WTB-NM-BA-e-1421