KERN & SOHN SD 30N70 User Manual

Sauter GmbH
Hi Lt
Lo dT
Lo Pe
P.OFF
HoldT
rS232
Tieringerstr. 11-15 D-72336 Balingen E-Mail: info@sauter.eu
Instruction Manual
Model: SD, Spring Tester
Thank you for buying a SAUTER Spring Tester. We hope you are pleased with your high quality force gauge with its big functional range. If you have any queries, wishes or helpful suggestions, do not hesitate to call our service number.
Working Conditions
1.
10°C to 30°C / 15% up to 80% humidity
2.
Operation
a. Display
b. Operating keys
ON / OFF:
(1) Measuring Result (2) Measuring Units (3) Activation of PRINT Function (4) Indication of power charging status (5) PEAK or AUTO-PEAK Mode (6) Average value of stored peak values (7) Force direction (8) Occupancy of storing spaces (9) AVERAGE- or Saving Mode
ON / OFF key (For ON, press 1 sec.)
SD
Tel: +49-[0]7433- 9976-174 Fax: +49-[0]7433-9976-285 Internet: www. sauter.eu
UNIT:
- Press shortly: Select unit: N, kg or lb
- Press for 2 sec.: Display return
ZERO:
Three functions:
- Zeros the measuring result (Tara function)
- Cleans the peak value (in Peak mode)
- Saves a setting (in SET mode)
SET:
- 1 x Press: Upper Limit [HidT / Hi Lt]. To change press: ▲ or ▼ (see section c.)
- 1 x Press: Lower Limit [LodT]. To change press ▲ or ▼ (see section c.)
- 1 x Press: Minimum limit to save peak values in the instrument’s memory. [LE.SET / Lo Pe]. Please see section „Min Limit Peak Save“. To change press: ▲ or ▼ (Only in „Peak-Mode“ active) (see section g.)
- 1 x Press: Auto-Off Function. Turns the instrument off after a here defined time period in sec. [P.OFF]. To change press: ▲ or ▼ (Only in „Battery-Mode“ active)
- 1 x Press: Peak-Freeze-Time [PE.2E / A.PE / HoldT]: Time period in which a peak value is being shown in the display in sec. To change press: ▲ or ▼
- To the printer (Print)
1 x Press: Saves the Settings
Back Light:
SD-BA-e-1112 1
Sauter GmbH
Tieringerstr. 11-15 D-72336 Balingen E-Mail: info@sauter.eu
Instruction Manual
PEAK:
Three functions
- Track mode (continuous measurement)
- Peak mode (capture of maximum values)
- Auto-Peak mode, same as Peak-mode, only without the „Min limit peak save“ function
MEMORY:
Saves the peak values to calculate the average value (please see section “Memory mode”
DELETE Function
Deletes stored peak values (only in „Memory mode“ active)
PRINT:
Sends the stored peak values to a PC or Printer (please see section „Data Output)
c. High / Low limit function
LEDs to display OK / NOT OK Tests
Lower than lower limit OK OK ▲ Higher than higher limit
This function allows efficient testing of OK / NOT OK measurements of similar or identical testing objects. A lower and an upper limit value can be defined. The instrument compares the individual measuring results with these limit values and shows the OK or NOT OK result by green or red light diode and by sound. To set these limit values, please see the SET Menu in section „Operating keys“.
d. Measurement (Track Modus)
Display (1) shows the continuous force in a defined direction (6)
To zero the display, press:
SD
Tel: +49-[0]7433- 9976-174 Fax: +49-[0]7433-9976-285 Internet: www. sauter.eu
e. Peak-Hold Function (Peak Mode)
Please press:
f. Auto-Peak-Hold-Function (Auto-Peak Mode)
Please press:
g. Min limit peak save
This function allows to eliminate unwanted „Pre-Peak values“ that are lower than the main peak value (Fp). The “Min limit peak save” value (Fo) takes care, that these “Pre-Peak values” are not saved.
The „Min limit peak save“ function is only in the Peak­Mode possible.
To set this Min limit value, please see the SET Menu in section „Operating keys“.
h. Memory mode and average value
peak values)
Saving peak values in the instrument
Activating the „AUTO PEAK Function“ by PEAK key
Deactivating the „Average Function“ by MEMORY
key
Now, all peak values are stored automatically in the
instrument
To browse through the stored values, please use the
▲ or ▼ keys. (The values will be shown in the upper display segment)
By pressing the MEMORY key, the average value of
the stored peak values can be shown (in the upper display segment)
To delete every stored value, press the▼-key in the
AVERAGE-Mode
(from up to 10
3. Thread of the fine adjustment screw
The following is valid for small Spring Testers (SD 10, 20,30) and those of medium size (SD 50, 100, 200, 300,
500):